APRO 55 Catheter

K234115 · Alembic, LLC · DQY · Mar 15, 2024 · Cardiovascular

Device Facts

Record IDK234115
Device NameAPRO 55 Catheter
ApplicantAlembic, LLC
Product CodeDQY · Cardiovascular
Decision DateMar 15, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The APRO 55 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 55 Catheter is also indicated for use as a conduit for retrieval devices.

Device Story

Single-lumen, braid and coil reinforced catheter; used to facilitate insertion/guidance of interventional devices into peripheral and neurovascular systems; functions as conduit for retrieval devices. Operated by clinicians under fluoroscopic guidance; introduced through guide catheter/sheath over guidewire. Features hydrophilic coating on distal shaft for navigation; radiopaque marker at distal tip for visualization. Provides mechanical access to target vasculature; enables delivery of therapeutic devices. Benefits patient by allowing minimally invasive access to peripheral and neurovascular sites for interventional procedures.

Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical bench testing, including visual/dimensional analysis, particulate testing, kink resistance, hub air leakage, torque strength, burst pressure, leakage, tensile strength, tip buckling force, and simulated use in anatomical neurovascular models. Biocompatibility testing performed per ISO 10993-1.

Technological Characteristics

Single-lumen catheter; 304V stainless steel braid/coil reinforcement; PTFE/Tecoflex composite liner; nylon hub; thermoplastic elastomer strain relief; platinum/iridium marker band. Hydrophilic coating. Sterilized via ethylene oxide (ISO 11135). Dimensions: 0.066" OD, 0.055" ID. Complies with ISO 10555-1 and ISO 80369-7.

Indications for Use

Indicated for facilitating insertion and guidance of interventional devices into peripheral and neurovascular blood vessels and as a conduit for retrieval devices in patients requiring such procedures.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 15, 2024 Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Alembic, LLC % Prithul Bom Most Responsible Person Regulatory Technology Services 1000 Westgate Drive, Suite #510k Saint Paul. Minnesota 55114 ### Re: K234115 Trade/Device Name: APRO 55 Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, QJP Dated: March 6, 2024 Received: March 6, 2024 ### Dear Prithul Bom: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Naira Muradyan -S Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K234115 Device Name APRO 55 Catheter Indications for Use (Describe) The APRO 55 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 55 Catheter is also indicated for use as a conduit for retrieval devices. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="font-family: Arial, sans-serif;"><b> <span style="font-size: 10.0pt;">☑</span></b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;"><b> <span style="font-size: 10.0pt;">☐</span></b></span> Over-The-Counter Use (21 CFR 801 Subpart C) | |-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| |-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary #### 510(k) Number: K234115 This 510(k) Summary is provided in accordance with the requirements of 21 CFR §807.92. #### Submitter information 1) Date Prepared: | Submitter: | Alembic, LLC<br>627 National Ave.<br>Mountain View, CA 94043 | |------------|-----------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Lisa Yen<br>Director of Regulatory and Quality<br>Telephone Number: (650) 388-5087<br>Fax: (650) 390-0107<br>Email: lyen@alembicllc.com | March 13, 2024 ### 2) Device Name and Classification | Trade/Proprietary Name: | APRO® 55 Catheter | |-------------------------|-----------------------------------------------------------------------------------------------------| | Common Name: | Percutaneous Catheter | | Classification Name: | Percutaneous Catheter, 21 CFR 870.1250<br>Catheter, Percutaneous, Neurovasculature, 21 CFR 870.1250 | | Regulatory Class: | Class II | | Product Codes: | DQY, QJP | | Review Panel: | Cardiovascular, Neurology | #### 3) Legally Marketed Predicate Device and Reference Device | Predicate Device: | K151667 AXS Catalyst Distal Access Catheter | |-------------------|--------------------------------------------------------| | Reference Device: | K223545 APRO 70 Catheter and Alembic Aspiration Tubing | #### 4) Device Description The APRO 55 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 55 Catheter is designed to facilitate the insertion and guidance of interventional devices into peripheral and neuro vasculature. Using standard catheterization techniques under fluoroscopic guidance, the APRO 55 Catheter is introduced through a guide catheter or guide sheath and over a guidewire into the target vasculature. The distal segment of the catheter shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal tip of the catheter for visualization under fluoroscopy. The APRO 55 Catheter is provided with an Introducer Sheath. {4}------------------------------------------------ #### 5) Indications for Use The APRO 55 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 55 Catheter is also indicated for use as a conduit for retrieval devices. #### 6) Technological Characteristics Comparison Alembic has demonstrated that the APRO 55 Catheter is substantially equivalent to the predicate device based on the similarity in materials, similarity in design concept, and the same fundamental operating principles. A comparison of the APRO 55 Catheter with the predicate device and reference device is summarized in Table 1 below. | Category | Subject Device<br>APRO 55 Catheter | Predicate Device<br>AXS Catalyst Distal Access<br>Catheter | Reference Device<br>APRO 70 Catheter and<br>Alembic Aspiration Tubing | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K234115 | K151667 | K223545 | | Regulatory Class | Class II, 21 CFR 870.1250,<br>DQY, QJP | Class II, 21 CFR 870.1250,<br>DQY | Class II, 21 CFR 870.1250,<br>NRY | | Indications for<br>Use | The APRO 55 Catheter is<br>indicated for use in<br>facilitating the insertion and<br>guidance of appropriately<br>sized interventional devices<br>into a selected blood vessel in<br>the peripheral and<br>neurovascular systems. The<br>APRO 55 Catheter is also<br>indicated for use as a conduit<br>for retrieval devices. | The AXS Catalyst Distal<br>Access Catheter is indicated for<br>use in facilitating the insertion<br>and guidance of appropriately<br>sized interventional devices into<br>a selected blood vessel in the<br>peripheral and neurovascular<br>systems. The AXS Catalyst<br>Distal Access Catheter is also<br>indicated for use as a conduit<br>for retrieval devices. | The APRO 70 Catheter with an<br>aspiration pump and the Alembic<br>Aspiration Tubing is intended for<br>use in the revascularization of<br>patients with acute ischemic stroke<br>secondary to intracranial large<br>vessel occlusive disease (within the<br>internal carotid, middle cerebral -<br>M1 and M2 segments, basilar, and<br>vertebral arteries) within 8 hours of<br>symptom onset. Patients who are<br>ineligible for intravenous tissue<br>plasminogen activator (IV t-PA) or<br>who fail IV t-PA therapy are<br>candidates for treatment.<br>The Alembic Aspiration Tubing is<br>intended to connect the APRO 70<br>Catheter to the aspiration pump. | | Materials | | | | | Hub | Nylon | Nylon | Polycarbonate | | Adhesive | Cyanoacrylate | None | Cyanoacrylate | | Strain Relief | Thermoplastic elastomer | Thermoplastic rubber | Thermoplastic elastomer | | Liner | Polytetrafluoroethylene<br>(PTFE)/Tecoflex composite | PTFE | PTFE/Tecoflex composite | | Shaft Coil or<br>Braid | 304V stainless steel braid<br>304V stainless steel coil | Stainless steel with nitinol wire<br>and polymer fiber | 304V stainless steel braid<br>304V stainless steel coil | | Extrusions | Thermoplastic polyurethanes,<br>thermoplastic elastomer | Pebax, Nylon | Thermoplastic polyurethanes,<br>thermoplastic elastomer | | Marker Band | Platinum/ iridium | Platinum/ iridium | Platinum/ iridium | | Adhesive | None | Cyanoacrylate | None | | Coating | Hydrophilic coating, identical | Hydrophilic coating | Hydrophilic coating | ### Table 1 - APRO 55 Catheter Comparison with the Predicate Device and Reference Device {5}------------------------------------------------ | Category | Subject Device<br>APRO 55 Catheter | Predicate Device<br>AXS Catalyst Distal Access<br>Catheter | Reference Device<br>APRO 70 Catheter and<br>Alembic Aspiration Tubing | |------------------------------|--------------------------------------|------------------------------------------------------------|-----------------------------------------------------------------------| | | to reference device | | | | Dimensions | | | | | Proximal Outer<br>Diameter | 0.066 inch | 5.6 F (0.073 inch)<br>6.0 F (0.078 inch) | 0.083 inch | | Distal Outer<br>Diameter | 0.066 inch | 5.3 F (0.069 inch)<br>5.4 F (0.070 inch) | 0.083 inch | | Inner Diameter | 0.055 inch | 0.058 inch, 0.060 inch | 0.070 inch | | Effective Lengths | 125, 137 cm | 115, 132 cm | 125, 132, 135 cm | | Coated Length | 90, 102 cm | 65, 82 cm | 90, 97, 100 cm | | Tip Shape | Straight | Straight | Straight | | Accessories | | | | | Introducer Sheath | Yes | Yes | Yes | | Rotating<br>Hemostasis Valve | None | Yes | None | | Packaging Materials | | | | | Pouch | Nylon/polyethylene/ Tyvek | Nylon/polyethylene/ Tyvek | Nylon/polyethylene/ Tyvek | | Packaging Tube | Polyethylene | Polyethylene | Polyethylene | | Packaging Card | Yes | Yes | Yes | | Shelf Carton | Solid bleached sulfate<br>paperboard | Solid bleached sulfate<br>paperboard | Solid bleached sulfate<br>paperboard | | Other | | | | | Sterilization | Ethylene oxide | Ethylene oxide | Ethylene oxide | | Shelf Life | 6 months | 2 years | 6 months | | Use Conditions | Sterile, single use, disposable | Sterile, single use, disposable | Sterile, single use, disposable | #### Performance Data 7) Alembic performed non-clinical bench, sterility, shelf-life, and biocompatibility testing. The results demonstrate substantial equivalence of the APRO 55 Catheter to the legally marketed predicate device. ### A. Design Verification Testing - Non-Clinical Bench Performance testing was conducted to support the APRO 55 Catheter submission. The results of the design verification and validation testing performed confirm that the APRO 55 Catheter conforms to the predefined specifications and met test acceptance criteria. A summary of the testing is shown in Table 2. | Test | Acceptance Criteria | Conclusion | |-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------| | Visual and Dimensional<br>Characteristics | Catheter meets the visual and dimensional<br>specifications. | The APRO 55 Catheter met the<br>acceptance criteria. | | Particulate | Catheter meets the acceptance criteria.<br>Subject device was evaluated with a<br>predicate device under the same test<br>conditions. | The APRO 55 Catheter met the<br>acceptance criteria. | | Kink Resistance | Catheter shaft shall not kink at clinically<br>relevant radii. | The APRO 55 Catheter met the<br>acceptance criteria. | | Hub Air Leakage | Catheter does not leak air into hub | The APRO 55 Catheter met the | Table 2 – Summary of Non-Clinical Bench Test Results {6}------------------------------------------------ | Test | Acceptance Criteria | Conclusion | |-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------| | Hub Compatibility | assembly with methods specified in ISO<br>10555-1, Annex D. | acceptance criteria. | | Torque Strength | Catheter meets the requirements specified<br>in ISO 80369-7. | The APRO 55 Catheter met the<br>acceptance criteria. | | Dynamic Burst Pressure | Catheter must withstand the minimum<br>required number of rotations without<br>breakage and without kinking compared to<br>legally marketed devices. | The APRO 55 Catheter met the<br>acceptance criteria. | | Liquid Leakage | No damage to catheter with dynamic<br>pressure. | The APRO 55 Catheter met the<br>acceptance criteria. | | Static Burst | Catheter must withstand pressure<br>with methods specified in ISO 10555-<br>1, Annex C. | The APRO 55 Catheter met the<br>acceptance criteria. | | Hub and Shaft Tensile<br>Strength | Catheter must withstand pressures<br>anticipated for clinical use. | The APRO 55 Catheter met the<br>acceptance criteria. | | Tip Tensile Strength | Catheter hub and shaft must meet tensile<br>strength specification. | The APRO 55 Catheter met the<br>acceptance criteria. | | Tip Buckling Force | Catheter tip must meet tip tensile strength<br>specification. | The APRO 55 Catheter met the<br>acceptance criteria. | | Delivery and Retrieval<br>Force | Catheter tip buckling force must be<br>acceptable. Forces were compared to a<br>predicate. | The APRO 55 Catheter met the<br>acceptance criteria. | | Simulated Use | Catheter delivery and retrieval force must<br>be acceptable. Forces were compared to a<br>predicate. | The APRO 55 Catheter met the<br>acceptance criteria. | | | When used per the Instructions for Use<br>with accessory devices in an anatomical<br>neurovascular model, the Catheter must<br>meet functionality specifications. | The APRO 55 Catheter met the<br>acceptance criteria. | ## B. Design Verification Testing - Animal Substantial equivalence was established based on non-clinical bench performance data. Animal testing data were not deemed necessary. ## C. Sterilization and Shelf-Life The APRO 55 Catheter is sterilized using an ethylene oxide sterilization cycle that was verified to a sterility assurance level of 1 x 10-6 in accordance with ISO 11135. Aging studies for the APRO 55 Catheter have established that the subject device and packaging remain functional for the labeled expiration date. Aging studies for packaging integrity, seal strength, and the device functionality were performed and met the acceptance criteria. ## D. Biocompatibility Biocompatibility testing has been completed for the APRO 55 Catheter in accordance with ISO 10993-1 and the device is deemed non-toxic (local or systemic), non-sensitizing, not locally irritating or otherwise harmful. Test results obtained were acceptable for the intended {7}------------------------------------------------ use as shown in Table 3. The thrombogenicity evaluation of the APRO 55 Catheter was supplemented by the prior in vivo thrombogenicity evaluation of the reference device, APRO 70 Catheter, in a good laboratory practice animal study. | Test | Results | Conclusions | |--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------| | Sensitization<br>(Guinea Pig Maximization) | The APRO 55 Catheter did not elicit a<br>sensitization response. | Non-sensitizing | | Irritation/Intracutaneous<br>Reactivity | The APRO 55 Catheter demonstrated no<br>evidence of irritation. | Non-irritant | | Cytotoxicity<br>(MEM Elution, L929 cells) | The APRO 55 Catheter did not elicit a<br>cytotoxic response at 24 hours and 48 hours. | Non-cytotoxic | | Hemolysis - Indirect | There were no significant differences between<br>the test article extract and negative control<br>article results. | Non-hemolytic | | Hemolysis - Direct | There were no differences between the<br>hemolytic index of the test article and the<br>negative control. | Non-hemolytic | | Partial Thromboplastin<br>Time (PTT) | The average clotting time of the test article was<br>greater than vehicle control and negative<br>control. | Acceptable clotting<br>times | | SC5b9 Complement<br>Activation | The Sc5b9 concentration of the test article was<br>statistically less than the positive control and was<br>not statistically higher than the negative control. | Acceptable | | Acute Systemic Toxicity | No weight loss, mortality, or evidence of<br>systemic toxicity from the extract exposure to<br>the mice. | Non-toxic | | Material-Mediated<br>Pyrogenicity | All individual rabbits for both the test article<br>and negative control showed a total rise in<br>temperature of < 0.5 °C and were determined to<br>be nonpyrogenic. | Non-pyrogenic | ### Table 3 - Biocompatibility Test Results ### E. Clinical Testing Substantial equivalence was established based on non-clinical performance data. Human clinical data were not deemed necessary. #### 8) Conclusion Based on the comparison of the technological characteristics and the non-clinical testing, the subject device is found to be substantially equivalent to the predicate device. The differences in technological characteristics do not raise new questions of safety and effectiveness. Testing was conducted to demonstrate that the subject device meets the specifications and performs as intended.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...