← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K193607

# BOBBY Balloon Guide Catheter (K193607)

_MicroVention, Inc. · DQY · Jul 21, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K193607

## Device Facts

- **Applicant:** MicroVention, Inc.
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** Jul 21, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The BOBBY Balloon Guide Catheter is intended: For use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

## Device Story

BOBBY Balloon Guide Catheter is a co-axial, braid-reinforced, variable stiffness catheter; features distal compliant polyurethane balloon, radiopaque markers, and proximal bifurcated luer hub. Used in peripheral and neurovascular procedures; operated by clinicians. Input: guidewire/catheter through inner lumen; contrast media via syringe through outer lumen for balloon inflation. Output: temporary vascular occlusion; conduit for retrieval devices. Hydrophilic coating on distal end enhances lubricity for advancement. Radiopaque markers facilitate fluoroscopic visualization of balloon position. Distal air-purging system removes air from inflation lumen. Benefits: enables controlled vascular access and occlusion during angiography and interventional procedures.

## Clinical Evidence

Bench testing only. No clinical or animal studies were required. Performance testing included conditioning/aging, visual inspection, dimensional attributes, inflation volume/diameter, burst volume, tip stiffness, torque, force at break, connector compliance, radiopacity, particulate matter, leakage, fatigue, simulated use, kink resistance, lubricity, and lumen collapse. Biocompatibility testing included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, and hemocompatibility (hemolysis, complement activation, thrombogenicity, PTT).

## Technological Characteristics

Co-axial, braid-and-coil reinforced catheter; materials include medical-grade plastics (nylon, PTFE, polyethylene, polyolefin) and stainless steel. Features compliant polyurethane balloon, distal hydrophilic coating, and Pt-Ir radiopaque marker bands. Dimensions: 8Fr (0.110" OD, 0.086" ID), 95 cm length. Sterilized via EtO (SAL 10^-6).

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Concentric Medical, FlowGate2™ ([K153729](/device/K153729.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 21, 2020

MicroVention, Inc. Stephanie Onstot Senior Regulatory Affairs Specialist 1311 Valencia Avenue Tustin, CA 92780

Re: K193607

Trade/Device Name: BOBBY Balloon Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, QJP Dated: June 25, 2020 Received: June 26, 2020

## Dear Stephanie Onstot:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan, Ph.D. Assistant Director (Acting) DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193607

Device Name BOBBY Balloon Guide Catheter

Indications for Use (Describe) The BOBBY Balloon Guide Catheter is intended:

For use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

Type of Use (Select one or both, as applicable)

| <div> <span> <b>                         ☑                     </b>                     Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span>                     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div>                                                       |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY

## 510(k) Owner Information:

| 510(k) Owner:  | MicroVention, Inc.      |                                          |
|----------------|-------------------------|------------------------------------------|
| Address:       | 1311 Valencia Ave       |                                          |
|                | Tustin, California, USA |                                          |
| Telephone:     | 714-247-8000            |                                          |
| Fax:           | 714-247-8014            |                                          |
| Contact Person | Stephanie Onstot        |                                          |
| Telephone      | 714-247-8000 x 8317     |                                          |
| Date Prepared: |                         | June 25, 2020                            |
|                | Device Trade Name:      | BOBBY™ Balloon Guide Catheter            |
|                | Classification:         | Class II                                 |
|                | Classification Name:    | Percutaneous Catheter                    |
|                | Product Code(s):        | DQY, QJP                                 |
|                | Regulation Number(s)    | 870.1250                                 |
|                | Predicate Devices:      | Concentric Medical, FlowGate2™ (K153729) |
|                |                         |                                          |

#### Indication for Use:

The BOBBY Balloon Guide Catheter is intended:

For use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

## Device Description

The BOBBY Balloon Guide Catheter is a co-axial, braid-reinforced, variable stiffness catheter with an external hydrophilic coating. The BOBBY Balloon Guide Catheter incorporates a compliant balloon, radiopaque markers, and a bifurcated luer hub on the proximal end.

The BOBBY Balloon Guide Catheter has an inner lumen through which a guidewire and catheter can be inserted, and a co-axial outer lumen that is used to inflate and deflate the balloon with a syringe filled with contrast media. A bifurcated luer hub is attached to the proximal end of the balloon guide catheter to provide access to both the inner and outer lumens. In addition, a hydrophilic coating is applied to the distal end of the balloon guide catheter to provide a lubricious outer surface for catheter advancement in the vasculature. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic

{4}------------------------------------------------

procedures. The balloon incorporates a distal air-purging system to purge air from the inflation lumen prior to use. The balloon catheter also incorporates radiopaque markers to facilitate fluoroscopic visualization and indication of the balloon position.

## Performance Testing Summary:

The BOBBY Balloon Guide Catheter has successfully completed the following relevant performance testing demonstrating that the device is suitable for its intended use.

| Performance Test Summary                                            |                                                                                                                                                                |                                                                                                                                                                                                                      |                                                                       |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Study Name                                                          | Description                                                                                                                                                    | Reference Standard                                                                                                                                                                                                   | Results                                                               |
| Conditioning,<br>Distribution, and Shelf<br>Life Aging Verification | To demonstrate that the product<br>meets the packaging strength and<br>packaging integrity following<br>accelerated aging to a 1-year<br>shelf life equivalent | ASTM F88 Seal<br>Strength of Flexible<br>Barrier materials<br>ASTM F2096 Standard<br>Test Method for<br>Detecting Gross Leaks<br>in Medical Packaging                                                                | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Packaging Visual<br>Inspection                                      | To demonstrate that the product<br>meets the packaging visual<br>inspection requirements given                                                                 | N/A                                                                                                                                                                                                                  | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Visual Surface<br>Requirements                                      | To demonstrate the product<br>satisfies the visual surface<br>requirements                                                                                     | ISO 10555-1:2013<br>Intravascular catheters<br>- Sterile and single-use<br>catheters - Part 1:<br>General requirements                                                                                               | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Dimensional/Physical<br>Attributes Inspection                       | To demonstrate that the product<br>meets the dimensional<br>specifications                                                                                     | ISO 10555-1:2013<br>Intravascular catheters<br>- Sterile and single-use<br>catheters - Part 1:<br>General requirements                                                                                               | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Inflation Volume vs<br>Balloon Diameter                             | To demonstrate that the product<br>meets the inflation volume vs<br>balloon diameter specifications                                                            | In consideration of<br>ISO 10555-4:2013<br>Intravascular Catheters<br>- Sterile and Single-<br>Use Catheters - Part 4:<br>Balloon Dilation<br>Catheters                                                              | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Balloon Burst Volume                                                | To demonstrate that the Balloon<br>is capable of withstanding an<br>injection volume above the<br>maximum fill volume.                                         | In consideration of<br>ISO 10555-4:2013<br>Intravascular Catheters<br>- Sterile and Single-<br>Use Catheters - Part 4:<br>Balloon Dilation<br>Catheters                                                              | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Tip Stiffness                                                       | To demonstrate that the stiffness<br>of the distal end of the product is<br>similar to other marketed devices.                                                 | N/A                                                                                                                                                                                                                  | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Torque Testing                                                      | To demonstrate that the product<br>is capable of 720 degrees of<br>rotation about the central lumen<br>axis without failure.                                   | N/A                                                                                                                                                                                                                  | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Performance Test Summary                                            |                                                                                                                                                                |                                                                                                                                                                                                                      |                                                                       |
| Study Name                                                          | Description                                                                                                                                                    | Reference Standard                                                                                                                                                                                                   | Results                                                               |
| Force at Break                                                      | To demonstrate the product<br>satisfies the force at break<br>requirements                                                                                     | ISO 10555-1:2013<br>Intravascular catheters<br>- Sterile and single-use<br>catheters - Part 1:<br>General requirements                                                                                               | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Small Bore Connector<br>Compliance with<br>Standard                 | To demonstrate that the product<br>meets the requirements                                                                                                      | ISO 80369-7 2016,<br>Small-bore connectors<br>for liquids and gases in<br>healthcare applications<br>- Part 7, Connectors<br>for intravascular or<br>hypodermic<br>applications.                                     | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Radiopacity                                                         | To determine the radiopaque<br>characteristics of the device.                                                                                                  | ISO 10555-1:2013<br>Intravascular catheters<br>- Sterile and single-use<br>catheters - Part 1:<br>General requirements<br>ASTM F640-12<br>Standard Test Methods<br>for Determining<br>Radiopacity for<br>Medical Use | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Particulate, Coating<br>Integrity                                   | This study was conducted to<br>determine the quantity and size of<br>particles generated during<br>simulated use                                               | USP <788> Particulate<br>Matter in Injections                                                                                                                                                                        | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Freedom from Liquid<br>Leakage                                      | To demonstrate that the product<br>meets the liquid leakage<br>requirements given in ISO<br>10555- 1.                                                          | ISO 10555-1:2013<br>Intravascular catheters<br>- Sterile and single-use<br>catheters - Part 1:<br>General requirements                                                                                               | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Hub Aspiration Air<br>Leakage                                       | To demonstrate that the product<br>meets the hub aspiration air<br>leakage requirements given in<br>ISO 10555-1.                                               | ISO 10555-1:2013<br>Intravascular catheters<br>- Sterile and single-use<br>catheters - Part 1:<br>General requirements                                                                                               | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Balloon Fatigue Test                                                | To demonstrate that there is no<br>degradation of the Balloon after<br>repetitive inflation cycles.                                                            | In consideration of<br>ISO 10555-4:2013<br>Intravascular Catheters<br>- Sterile and Single-<br>Use Catheters - Part 4:<br>Balloon Dilation<br>Catheters                                                              | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Simulated Use                                                       | Simulated use under in vitro<br>conditions in a cerebral vascular<br>model                                                                                     | In consideration of<br>ISO 10555-4:2013<br>Intravascular Catheters<br>- Sterile and Single-<br>Use Catheters - Part 4:<br>Balloon Dilation<br>Catheters                                                              | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Performance Test Summary                                            |                                                                                                                                                                |                                                                                                                                                                                                                      |                                                                       |
| Study Name                                                          | Description                                                                                                                                                    | Reference Standard                                                                                                                                                                                                   | Results                                                               |
| Flexural Fatigue                                                    | To demonstrate that the product<br>does not lose structural integrity<br>when used in the tortuous path<br>model.                                              | ISO 10555-1:2013<br>Intravascular catheters<br>- Sterile and single-use<br>catheters - Part 1:<br>General requirements                                                                                               | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Kink Resistance                                                     | To demonstrate that the device<br>has similar kink resistance<br>compared to the predicate device.                                                             | N/A                                                                                                                                                                                                                  | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Lubricity and durability of<br>the hydrophilic coating              | To demonstrate that the<br>hydrophilic coating is lubricious<br>and durable.                                                                                   | N/A                                                                                                                                                                                                                  | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Guidewire lumen burst<br>pressure (Static, dynamic)                 | To demonstrate that the device<br>does not burst below rated burst<br>pressure.                                                                                | ISO 10555-1:2013<br>Intravascular catheters<br>- Sterile and single-use<br>catheters - Part 1:<br>General requirements                                                                                               | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Balloon Deflation Time                                              | To demonstrate that the device<br>has similar balloon deflation time<br>compared to the predicate device.                                                      | ISO10555-4:2013<br>Intravascular catheters<br>- Sterile and single-use<br>catheters – Part 4:<br>Balloon dilation<br>catheters                                                                                       | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |
| Lumen Collapse                                                      | To demonstrate that the<br>guidewire lumen does not<br>collapse under aspiration.                                                                              | N/A                                                                                                                                                                                                                  | Pass<br>All samples met the pre-<br>determined acceptance<br>criteria |

{5}------------------------------------------------

{6}------------------------------------------------

No animal or clinical studies were required to demonstrate substantial equivalence.

## Biocompatibility Testing Summary

Categorized as Externally Communicating Device, Circulating Blood, Limited Contact (< 24 hours), per ISO 10993-1, the following testing was conducted:

| Test Name                    | Test Method                                                                                                                            | Results                                                                                                                                                                    |                                                                                                                              |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity                 | Tested in accordance with ISO 10993-5, Biological<br>Evaluation of Medical Devices – Part 5: Tests for in<br><i>vitro</i> cytotoxicity | Pass<br>Noncytotoxic according to the<br>predetermined acceptance criteria                                                                                                 |                                                                                                                              |
|                              | Sensitization                                                                                                                          | Tested in accordance with ISO 10993-10, Biological<br>Evaluation of Medical Devices – Part 10 Tests for<br>Irritation and Skin Sensitization, Kligman<br>Maximization Test | Pass<br>Did not elicit a sensitization response<br>according to the predetermined<br>acceptance criteria                     |
| Intracutaneous<br>Irritation |                                                                                                                                        | Tested in accordance with ISO 10993-10, Biological<br>Evaluation of Medical Devices – Part 10: Tests for<br>Irritation and Skin Sensitization                              | Pass<br>Test requirements for<br>intracutaneous reactivity were met<br>according to the predetermined<br>acceptance criteria |

{7}------------------------------------------------

| Test Name                                               | Test Method                                                                                                                                                                                                                                                                                                  | Results                                                                                                                                  |
|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Systemic Toxicity:<br>Systemic Injection<br>Test        | Tested in accordance with ISO 10993-11, Biological<br>Evaluation of Medical Devices - Part 11: Tests for<br>Systemic Toxicity                                                                                                                                                                                | Pass<br>Test requirements for systemic<br>toxicity were met according to the<br>predetermined acceptance criteria                        |
| Systemic Toxicity:<br>Material Mediated<br>Pyrogenicity | Tested in accordance with ISO 10993-11, Biological<br>Evaluation of Medical Devices – Part 11: Tests for<br>Systemic Toxicity and USP NF 36:2018 <151><br>Pyrogen Test                                                                                                                                       | Pass<br>Non-pyrogenic, met the<br>predetermined acceptance criteria                                                                      |
| Hemocompatibility:<br>Hemolysis                         | Tested in accordance with ASTM F756-17,<br>Standard Practice for Assessment of Hemolytic<br>Properties of Materials and ISO 10993-4, Biological<br>Evaluation of Medical Devices - Part 4: Selection of<br>Tests for Interactions with Blood. Tests for<br>Hemolytic Properties, Direct and Indirect Methods | Pass<br>Non-hemolytic, met the<br>predetermined acceptance criteria                                                                      |
| Hemocompatibility:<br>Complement<br>Activation          | Tested in accordance with ISO 10993-4, Biological<br>Evaluation of Medical Devices - Part 4: Selection of<br>Tests for Interactions with Blood, SC5b-9<br>Complement Activation                                                                                                                              | Pass<br>Does not activate the complement<br>system, met the predetermined<br>acceptance criteria                                         |
| Hemocompatibility:<br>Thrombogenicity                   | Tested in accordance with ISO 10994-4, Biological<br>Evaluation of Medical Devices - Part 4: Selection of<br>Tests for Interactions with Blood, and ASTM<br>F2888-19, Standard Practice for Platelet Leukocyte<br>Count-An In Vitro Measure for Hemocompatibility<br>Assessment of Cardiovascular Materials  | Pass<br>Demonstrates similar<br>thromboresistance characteristics as<br>the control device, met the<br>predetermined acceptance criteria |
| Hemocompatibility:<br>In Vitro<br>Hemocompatibility     | Tested in accordance with ISO 10993-4, Biological<br>Evaluation of Medical Devices - Part 4: Selection of<br>Tests for Interactions with Blood,<br>Hemocompatibility, Direct Contact Method                                                                                                                  | Pass<br>Not expected to result in adverse<br>effects in vivo, met the<br>predetermined acceptance criteria                               |
| Hemocompatibility:<br>Partial<br>Thromboplastin Time    | Tested in accordance with ISO 10994-4, Biological<br>Evaluation of Medical Devices - Part 4: Selection of<br>Tests for Interactions with Blood and ASTM F2382-<br>18, Standard Test Method for Assessment of<br>Intravascular Medical Device Materials on Partial<br>Thromboplastin Time (PTT)               | Pass<br>Does not have an effect on<br>coagulation of human plasma, met<br>the predetermined acceptance<br>criteria                       |

## Predicate Device Comparison

The following table provides a comparison of the key characteristics of the BOBBY Balloon Guide Catheter to the predicate device.

| Comparison Chart                |                                                                                                                      |                                                                                                                               |                                                                                                                                                                               |  |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Feature                         | Subject Device                                                                                                       | Predicate device                                                                                                              | Comparison                                                                                                                                                                    |  |
|                                 | BOBBY Balloon<br>Guide Catheter<br>K193607                                                                           | 8F FlowGate2™<br>Balloon Guide Catheter<br>K153729                                                                            |                                                                                                                                                                               |  |
| FDA Classification              | Class II                                                                                                             | Class II                                                                                                                      | Same                                                                                                                                                                          |  |
| Product Code(s)                 | DQY, QJP                                                                                                             | DQY                                                                                                                           | Same                                                                                                                                                                          |  |
| Regulation Number               | 870.1250                                                                                                             | 870.1250                                                                                                                      | Same                                                                                                                                                                          |  |
| Regulation Name                 | Percutaneous Catheter                                                                                                | Percutaneous Catheter                                                                                                         | Same                                                                                                                                                                          |  |
| Anatomical Locations            | Peripheral and neuro vasculature                                                                                     | Peripheral and neuro vasculature                                                                                              | Same                                                                                                                                                                          |  |
| Comparison Chart                |                                                                                                                      |                                                                                                                               |                                                                                                                                                                               |  |
| Feature                         | Subject Device<br>BOBBY Balloon<br>Guide Catheter<br>K193607                                                         | Predicate device<br>8F FlowGate™<br>Balloon Guide Catheter<br>K153729                                                         | Comparison                                                                                                                                                                    |  |
| Material                        | Commonly used<br>medical grade plastics<br>(e.g. nylon, PTFE,<br>polyethylene,<br>polyolefin) and<br>stainless steel | Commonly used<br>medical grade plastics<br>(e.g. nylon, PTFE,<br>polyethylene,<br>polyolefin) and<br>stainless steel          | Similar, minor<br>differences do not<br>raise new questions<br>of safety and<br>efficacy, confirmed<br>through<br>biocompatibility and<br>performance testing                 |  |
| Reinforced Catheter<br>Shaft    | Stainless steel braid<br>and coil reinforced<br>shaft                                                                | Stainless steel braid<br>reinforced shaft                                                                                     | Similar, minor<br>differences do not<br>raise new questions<br>of safety and<br>efficacy                                                                                      |  |
| Injection Port                  | Yes                                                                                                                  | Yes                                                                                                                           | Same                                                                                                                                                                          |  |
| Radiopacity                     | Yes, shaft is visible<br>due to distal tip Pt-Ir<br>marker bands                                                     | Yes, shaft material<br>contains barium<br>sulfate, distal tip Pt-Ir<br>marker band                                            | Similar, minor<br>difference does not<br>raise new questions<br>regarding safety and<br>efficacy, both<br>devices are<br>radiopaque                                           |  |
| Radiopaque Marker<br>Bands      | 2                                                                                                                    | 1                                                                                                                             | Similar, minor<br>difference does not<br>raise new questions<br>regarding safety and<br>efficacy, both<br>devices utilize<br>radiopaque marker<br>bands                       |  |
| Compliant Balloon               | Yes, polyurethane                                                                                                    | Yes, silicone                                                                                                                 | Similar, material<br>differences do not<br>raise new questions<br>of safety and<br>efficacy, both<br>materials are used<br>for compliant<br>balloons for<br>intravascular use |  |
| Labeled Shaft Outer<br>Diameter | 0.110 in (2.8 mm)<br>8Fr                                                                                             | 0.106 in (2.7 mm)<br>8Fr                                                                                                      | Similar, minor<br>differences do not<br>raise new questions<br>of safety and<br>efficacy                                                                                      |  |
| Comparison Chart                |                                                                                                                      |                                                                                                                               |                                                                                                                                                                               |  |
| Feature                         | Subject Device<br>BOBBY Balloon<br>Guide Catheter<br>K193607                                                         | Predicate device<br>8F FlowGate2™<br>Balloon Guide Catheter<br>K153729                                                        | Comparison                                                                                                                                                                    |  |
| Labeled Shaft Inner<br>Diameter | 0.086 in (2.18 mm)<br>6.5Fr                                                                                          | 0.084 in (2.1 mm)<br>6.4Fr                                                                                                    | Similar, minor<br>differences do not<br>raise new questions<br>of safety and<br>efficacy                                                                                      |  |
| Effective length                | 95 cm<br>(37.4 in)                                                                                                   | 85 cm, 95cm<br>(33.5, 37.4 in)                                                                                                | Similar, minor<br>differences do not<br>raise new questions<br>of safety and<br>efficacy                                                                                      |  |
| Tip Shape                       | Straight                                                                                                             | Straight                                                                                                                      | Same                                                                                                                                                                          |  |
| Coating                         | Hydrophilic Coating -<br>Distal portion of shaft                                                                     | None                                                                                                                          | Difference does not<br>raise new questions<br>of safety and<br>efficacy.<br>Hydrophilic coatings<br>are commonly used<br>to lubricate vascular<br>catheters.                  |  |
| Internal Construction           | Coaxial lumen                                                                                                        | Coaxial lumen                                                                                                                 | Same                                                                                                                                                                          |  |
| Accessories Supplied            | Peel Away Sheath                                                                                                     | Dilator, Rotating<br>Hemostasis Valve,<br>Tuohy Borst Valve<br>with sideport, Peel<br>Away Sheaths, Luer-<br>Activated Valves | Similar, minor<br>differences do not<br>raise new questions<br>of safety and<br>efficacy.                                                                                     |  |
| How Supplied                    | Sterile, single use                                                                                                  | Sterile, single use                                                                                                           | Same                                                                                                                                                                          |  |
| Sterilization Method            | EtO                                                                                                                  | EtO                                                                                                                           | Same                                                                                                                                                                          |  |
| Sterility Assurance<br>Level    | 10-6                                                                                                                 | 10-6                                                                                                                          | Same                                                                                                                                                                          |  |

{8}------------------------------------------------

{9}------------------------------------------------

## Conclusion:

MicroVention, Inc. concludes through a review of the benchtop assessments, the comparison of the device classification, indications for use, operating principle, technological characteristics, sterility, and biocompatibility that the BOBBY Balloon Guide Catheter is substantially equivalent to the predicate device, FlowGate2™ Balloon Guide Catheter. Any differences are minor and do not raise different questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K193607](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K193607)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
