The GEMS-H is a robotic exoskeleton that fits orthotically on the wearer's waist and thighs, outside of clothing. The device is intended to help assist ambulatory function in rehabilitation institutions under the supervision of a trained healthcare professional for the following population: Individuals with stroke who have gait deficits and exhibit gait speeds of at least 0.4 m/s and are able to walk at least 10 meters with assistance from a maximum of one person. The trained healthcare professional must successfully complete a training program prior to use of the device. The device is not intended for sports.
Device Story
GEMS-H is a lightweight (4.7 lbs) robotic exoskeleton worn around waist and thighs; assists hip flexion and extension during gait. Input data includes hip joint angle and torque output via integrated sensors. Device uses two motors powered by a rechargeable battery; assist settings are adjusted by a trained physical therapist via a tablet PC application. Used in rehabilitation institutions; therapist monitors torque in real-time. Output provides assistive torque to thighs via support frames; helps improve ambulatory function and gait speed in stroke patients. Failsafe feature disables motor torque, rendering device passive if needed.
Clinical Evidence
Prospective, single-arm, open-label study (N=41) in stroke patients. Primary endpoint: safety (AEs) and effectiveness (improvement ≥ 0.14 m/s in 10MWT gait speed). Results: 0.8% device-related AE rate; mean gait speed improvement +0.12 m/s (p<0.0001) without device, +0.16 m/s (p<0.0001) with device. 6MWT improved by 53.28 m (p<0.0001).
Indicated for stroke patients with gait deficits, gait speeds ≥ 0.4 m/s, and ability to walk ≥ 10 meters with max one-person assistance. Contraindicated for patients with heterotopic ossification or severe osteoporosis.
Regulatory Classification
Identification
A powered lower extremity exoskeleton is a prescription device that is composed of an external, powered, motorized orthosis that is placed over a person's paralyzed or weakened limbs for medical purposes.
Special Controls
In combination with the general controls of the FD&C Act, the Powered Exoskeleton is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Elements of the device materials that may contact the patient must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility/interference (EMC/EMI), electrical safety, thermal safety, mechanical safety, battery performance and safety, and wireless performance, if applicable.
(3) Appropriate software verification, validation, and hazard analysis must be performed.
(4) Design characteristics must ensure geometry and materials composition are consistent with intended use.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must include:
(i) Mechanical bench testing (including durability testing) to demonstrate that the device will withstand forces, conditions, and environments encountered during use;
(ii) Simulated use testing (
*i.e.,* cyclic loading testing) to demonstrate performance of device commands and safeguard under worst case conditions and after durability testing;(iii) Verification and validation of manual override controls are necessary, if present;
(iv) The accuracy of device features and safeguards; and
(v) Device functionality in terms of flame retardant materials, liquid/particle ingress prevention, sensor and actuator performance, and motor performance.
(6) Clinical testing must demonstrate a reasonable assurance of safe and effective use and capture any adverse events observed during clinical use when used under the proposed conditions of use, which must include considerations for:
(i) Level of supervision necessary, and
(ii) Environment of use (
*e.g.,* indoors and/or outdoors) including obstacles and terrain representative of the intended use environment.(7) A training program must be included with sufficient educational elements so that upon completion of training program, the clinician, user, and companion can:
(i) Identify the safe environments for device use,
(ii) Use all safety features of device, and
(iii) Operate the device in simulated or actual use environments representative of indicated environments and use.
(8) Labeling for the Physician and User must include the following:
(i) Appropriate instructions, warning, cautions, limitations, and information related to the necessary safeguards of the device, including warning against activities and environments that may put the user at greater risk.
(ii) Specific instructions and the clinical training needed for the safe use of the device, which includes:
(A) Instructions on assembling the device in all available configurations;
(B) Instructions on fitting the patient;
(C) Instructions and explanations of all available programs and how to program the device;
(D) Instructions and explanation of all controls, input, and outputs;
(E) Instructions on all available modes or states of the device;
(F) Instructions on all safety features of the device; and
(G) Instructions for properly maintaining the device.
(iii) Information on the patient population for which the device has been demonstrated to have a reasonable assurance of safety and effectiveness.
(iv) Pertinent non-clinical testing information (
*e.g.,* EMC, battery longevity).(v) A detailed summary of the clinical testing including:
(A) Adverse events encountered under use conditions,
(B) Summary of study outcomes and endpoints, and
(C) Information pertinent to use of the device including the conditions under which the device was studied (
*e.g.,* level of supervision or assistance, and environment of use (*e.g.,* indoors and/or outdoors) including obstacles and terrain).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
April 21, 2022
Samsung Electronics Co., Ltd. % Kyoungju Kim Consultant MDLab Inc. Room 804, 161-17 Magokjungang-ro, Gangseo-gu, Seoul South Korea, 07788
Re: K213452
Trade/Device Name: GEMS-H Regulation Number: 21 CFR 890.3480 Regulation Name: Powered lower extremity exoskeleton Regulatory Class: Class II Product Code: PHL Dated: March 23, 2022 Received: March 24, 2022
Dear Kyoungju Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Device Name GEMS-H
Indications for Use (Describe)
The GEMS-H is a robotic exoskeleton that fits orthotically on the wearer's waist and thighs, outside of clothing. The device is intended to help assist ambulatory function in rehabilitations under the supervision of a trained healthcare professional for the following population:
· Individuals with stroke who have gait deficits and exhibit gait speeds of at least 0.4 m/s and are able to walk at least 10 meters with assistance from a maximum of one person.
The trained healthcare professional must successfully complete a training program prior to use of the device. The device is not intended for sports.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-family: DejaVu Sans, sans-serif;"> ✔ </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-family: DejaVu Sans, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Samsung Electronics Co., Ltd.
### 510(k) Summary
### Submitter
Samsung Electronics Co., Ltd. Minhyung Lee, Ph.D. 129, Samsung-Ro, Yeongtong-Gu, Suwon-Si, Gyeonggido, 16677 South Korea Email: mhyung.lee@samsung.com Phone: +82-2-6147-3805 Fax: +82-2-6147-7980
### Official Correspondent
MDLab Inc. Kyoungju Kim, Consultant. Room 804, 161-17, Magokjungang-ro, Gangseo-gu, Seoul South Korea. 07788 Email: kj.kim@mdlab.co.kr Phone: +82-10-2264-5341 Fax: 82-2-3664-0830
### Device Information
- Trade Name: GEMS-H ●
- Common Name: Powered exoskeleton
- Classification Name: Powered lower extremity exoskeleton
- Product Code: PHL
- Panel: Neurology
- Regulation Number: 21 C.F.R. § 890.3480
- Device Class: Class II ●
- Date prepared: 04/11/2022 ●
### Predicate Device
K181294, Honda Walking Assist Device (by Honda Motor Company, Ltd.)
### Device Description
The GEMS-H is a lightweight, robotic exoskeleton designed to help assist ambulatory function of stroke patients who meet the assessment criteria, in rehabilitations under the supervision of a trained healthcare professional. The GEMS-H device provides assistance to the patient during hip flexion and extension.
The device is worn over clothing around the wearer's waist and fastened with Velcro straps to assists hip flexion and extension. The device weighs 4.7 lbs (2.1 kg) and has two motors that run on a single rechargeable battery. The device is equipped with joint angle and electrical current sensors to monitor hip joint angle and torque output, respectively.
The assist torque is transmitted to the wearer's thighs via thigh support frames. A trained healthcare professional, who operates the device, can change assist settings through software that runs on the tablet PC.
{4}------------------------------------------------
### Indications for use
The GEMS-H is a robotic exoskeleton that fits orthotically on the wearer's waist, and thighs outside of clothing. The device is intended to help assist ambulatory function in rehabilitation institutions under the supervision of a trained healthcare professional for the following population:
- Individuals with stroke who have gait deficits and exhibit gait speeds of at least 0.4 m/s and are able to walk at least 10 meters with assistance from a maximum of one person.
The trained healthcare professional must successfully complete a training program prior to use of the device. The device is not intended for sports.
### Substantial Equivalence Comparison
The subject and predicate device (K181294) have the same intended use, and nearly identical indications for use, and are similar in design, technology, functions, and principle of operation.
Both devices are intended to help assist ambulatory function in individuals with stroke who have gait deficits and exhibit gait speeds of at least 0.4 m/s and are able to walk at least 10 meters with assistance from a maximum of one person. Both are worn around the wearer's waist and thighs and assist with hip joint flexion and extension. They also have two motors that run on a single pack of rechargeable batteries. They are equipped with angle and current sensors to monitor hip joint angle and torque output respectively. Both devices provide an assistive torque that is transmitted to the wearer's thighs via thigh frames. Both devices can change assist settings through software that runs on a touchscreen tablet device.
The differences in technological characteristics between the subject and predicate are as follows:
- . Range of Motion: The GEMS-H has a smaller range of motion than predicate device. Since the GEMS-H's range of motion falls within the predicate's range, this difference does not raise different safety or effectiveness questions.
- Device Weight: The GEMS-H weighs less than the predicate due to its design. The difference in weight does not raise different questions of safety or effectiveness.
- Battery Specification: The GEMS-H can be operated from a minimum of 1 hour to a maximum of . 2 hours, while the predicate device only allows 1 hour of continuous operation. This difference does not raise different questions of safety or effectiveness.
- . Actuator specifications: The maximum torque of the GEMS-H is larger than that of the predicate. However, even though the motors of the GEMS-H are capable of producing a maximum of 12 Nm± 15%, the maximum torque will not be reached in all cases and for all patients. The trained physical therapist can monitor the torque in real time and change the setting on the tablet PC to control the generated maximum torque. In addition, any potential risk associated with the higher maximum torque is mitigated through the exclusion criteria that exclude patients with severe osteoporosis (as determined by a physician) from study participation. Likewise, the proposed labeling contains a contraindication for patients with heterotopic ossification or severe osteoporosis. As a result, this difference in maximum torque does not raise different questions of safety or effectiveness.
{5}------------------------------------------------
### Samsung Electronics Co., Ltd.
The following table summarizes the similarities and differences between the subject and predicate devices.
| | Subject Device | Primary Predicate Device | Differences/Similarity |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA 510(k) No. | K213452 | K181294 | |
| Company | Samsung Electronics Co., Ltd. | Honda Motor Company | |
| Trade/Device<br>Name | GEMS-H | Honda Walking Assist Device | |
| Regulation No. | 21 CFR 890.3480 | 21 CFR 890.3480 | Identical |
| Regulation Name | Powered exoskeleton | Powered exoskeleton | Identical |
| Regulatory Class | Class 2 | Class 2 | Identical |
| Product Code | PHL | PHL | Identical |
| Indication for use | The GEMS-H is a robotic exoskeleton that fits<br>orthotically on the wearer's waist and thighs,<br>outside of clothing. The device is intended to<br>help assist ambulatory function in<br>rehabilitation institutions under the<br>supervision of a trained healthcare<br>professional for the following population:<br>· Individuals with stroke who have gait deficits<br>and exhibit gait speeds of at least 0.4 m/s and<br>are able to walk at least 10 meters with<br>assistance from a maximum of one person.<br>The trained healthcare professional must<br>successfully complete a training program prior<br>to use of the device. The device is not intended<br>for sports. | The Honda Walking Assist Device is a robotic<br>exoskeleton that fits orthotically on the user's<br>waist and thigh, outside of clothing. The device<br>is intended to help assist ambulatory function<br>in rehabilitation institutes under the<br>supervision of a trained healthcare professional<br>for the following population:<br>• Individuals with stroke who have gait<br>deficits and exhibit gait speeds of at least 0.4<br>m/s and are able to walk at least 10 meters<br>with assistance from a maximum of one<br>person.<br>The trained healthcare professional must<br>successfully complete a training program prior<br>to use of the device. The devices are not<br>intended for sports. | Identical |
| Body coverage | Worn around the waist & thighs | Worn around the waist & thighs | Identical |
| Mobility aid | Optional (e.g., walker, cane, harness) | Optional (e.g., walker, cane) | Similar |
| Device limit on<br>patient's gait speed | None | None | Identical |
| Type of Surface for<br>training | Smooth, cement, carpet | Smooth, cement, carpet | Identical |
| | Subject Device | Primary Predicate Device | Differences/Similarity |
| Height of Patient | 61 in to 75 in (1.55 m to 1.91 m) | 55 in to ~79 in (1.4 m to 2.0 m) | Similar |
| Weight of Patient | 99 lbs (45 kg) up to 220 lbs (100 kg) and BMI< 35 kg/ m² | <220 lbs (100 kg) | Similar |
| Control Method | Handheld interface to allow physical therapist<br>to control and operate the device remotely | Handheld interface for physical therapist | Identical |
| Device Range of<br>Motion | Hip: 100° flexion to 40° extension | Hips: 113° flexion to 47° extension | Different. The subject device has a smaller<br>motion range than the predicate device. This<br>difference does not raise different safety or<br>effectiveness questions. |
| Device Weight | 4.7 lbs (2.1 kg) | 5.95 lbs (2.7 kg) | Similar. The GEMS-H weighs less than the<br>predicate device; the difference in weight does<br>not raise different questions of safety or<br>effectiveness. |
| Battery<br>specifications | • Rechargeable lithium ion.<br>• 21.6V, 2,950 mAh,<br>• 2 hr charge time<br>• Minimum 2 Hr. of continuous walking* (with<br>gain = 7, delay = 0.25 s)<br>• Minimum 1 Hr. of continuous walking* (with<br>gain = 8, delay = 0.25 s)<br>* at a walking speed of about 1.9 ~ 2.5 MPH<br>(0.8 ~ 1.1 m/s) | • Rechargeable lithium ion,<br>• 22.2 V, 1 A-h,<br>• 2 hr charge time<br>• 1 hr continuous operation | Similar.<br>The GEMS-H can be operated from a<br>minimum of 1 hour to a maximum of 2 hours,<br>while the predicate device only allows 1 hour<br>of continuous operation. This difference does<br>not raise different questions of safety or<br>effectiveness. |
| Actuator<br>Specifications | 2 motors (2 at hip)<br>12 Nm ± 15% max torque | 2 motors (2 at hip)<br>4 Nm max torque | Different.<br>The maximum torque of the GEMS-H is larger<br>than that of the predicate. However, even<br>though the motors of the GEMS-H are capable<br>of producing a maximum of 12 Nm± 15%, this<br>is the maximum allowable torque and is not<br>recommended for all patients under all training<br>conditions. Rather, the trained physical<br>therapist shall monitor the torque in real time<br>and change the setting on the tablet PC to |
| | Subject Device | Primary Predicate Device | Differences/Similarity |
| | | | control the generated maximum torque. In<br>addition, any potential risk associated with the<br>higher maximum torque is mitigated through<br>the exclusion criteria that exclude patients with<br>severe osteoporosis (as determined by a<br>physician) from study participation. Likewise,<br>the proposed labeling contains a<br>contraindication for patients with heterotopic<br>ossification or severe osteoporosis. As a result,<br>this difference in maximum torque does not<br>raise different questions of safety or<br>effectiveness. |
| Training Program | Yes | Yes | Identical |
| Certification<br>Program | Yes | Yes | Identical |
| Feedback | Visual & auditory feedback on both the<br>handheld controller & device | Visual & auditory feedback on both the<br>handheld controller & device | Identical |
| Fall Detection and<br>Mitigation | None | None | Identical |
| Failsafe Feature | Motor torque disables; device becomes<br>passive | Motor torque disables; device becomes passive | Identical |
| Operating<br>Temperature | 32°F to 86°F (0°C to 30°C) | 32 °F to 86 °F (0 °C to 30 °C) | Identical |
| Operating | 30% to 85% RH | 30% to 85% | Identical |
| Patient population | Individuals with stroke who have gait deficits<br>and exhibit gait speeds of at least 0.4 m/s and<br>are able to walk at least 10 meters with<br>assistance from a maximum of one person. | Individuals with stroke who have gait deficits<br>and exhibit gait speeds of at least 0.4 m/s and<br>are able to walk at least 10 meters with<br>assistance from a maximum of one person. | Identical |
{6}------------------------------------------------
### Samsung Electronics Co., Ltd.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "SAMSUNG" in a bold, sans-serif font. The letters are all capitalized and are a dark blue color. The background is white, which makes the text stand out. The logo is simple and recognizable.
### Samsung Electronics Co., Ltd.
{8}------------------------------------------------
### Non-Clinical Testing
The subject device was tested to the following standards.
- Electrical safety testing according to ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, ● C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- Electromagnetic compatibility (EMC) testing according to IEC 60601-1-2:2014
- Usability engineering test according to IEC 60601-1-6:2013
- Vibration testing and thermal shock testing
- Battery safety testing according to IEC 62133-2:2017 Edition 1.0
- Software Verification and validation testing according to IEC 62304:2015 ●
- . Radio frequency wireless technology according to the procedure given in FCC Rules Part 15 Subpart B and FCC Rules Part 15 Subpart C, 15.247
- ANSI IEEE C63.27-2017 American National Standard for Evaluation of Wireless Coexistence
- Particulate resistance (IP20) according to IEC 60529:2013, Edition 2.2
- . Biocompatibility Testing such as In Vitro Cytotoxicity (ANSI AAMI ISO 10993-5:2009/(R)2014), Skin Sensitization (ANSI AAMI ISO 10993-10:2010/(R)2014) and Intracutaneous Reactivity (ANSI AAMI ISO 10993-10:2010/(R)2014)
- . Performance testing such as cyclic loading testing vibration testing, thermal shock testing, driving part durability testing, Velcro testing and worst-case frame testing
- Other Bench testing such as Safety according to the Range of Motion, Max. angular velocity, Max. Torque Test, Static Torque Tracking Performance, Providing Assistive Torque Test at DOC Walking Mode for Walking and Stopping, Generation of assistive torque patterns according to the change in option during DOC mode test, Continuous use time and Noise.
### Clinical Testing
Samsung conducted a prospective, single arm, interventional, open-label, single center study with the GEMS-H in stroke patients. The study enrolled 53 subjects, of whom 41 completed the entire protocol. Each subject completed 18 indoor sessions of training with the GEMS-H in the outpatient clinic with a licensed physical therapist.
A summary of the clinical trial is provided below.
| Title | Safety and Effectiveness on Functional Mobility Following Samsung Gait<br>Enhancing and Motivating System-Hip (GEMS-H) Device Training in Sub-Acute<br>and Chronic Stroke: A Pivotal Study | |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Protocol Number | 00210372…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.