K023209 · Fasstech · IKN · Oct 10, 2003 · Physical Medicine
Device Facts
Record ID
K023209
Device Name
INSIGHT MILLENNIUM III
Applicant
Fasstech
Product Code
IKN · Physical Medicine
Decision Date
Oct 10, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.1375
Device Class
Class 2
Indications for Use
To measure bilateral differences in surface EMG along the spine To measure surface EMG along the spine during functional tasks To measure bilateral differences in skin temperature along the spine To measure Range of Motion of the three spinal regions To measure patient self-reported pressure sensitivity in joints and muscles To measure Heart Rate To chart patient progress during the course of treatment
Device Story
Insight Millennium III is a non-invasive, multi-modality physiologic monitoring device. Inputs include surface EMG (muscle activity), infrared skin temperature, inclinometer (range of motion), algometer (pressure sensitivity), and IR plethysmograph (heart rate). Sensors connect to an instrument console, which interfaces with a PC via isolated USB. Software displays real-time readings, records data, generates graphs, and prints reports. Used in clinical settings by healthcare providers to monitor spinal physiology and track patient progress during treatment. Device provides objective data to assist clinicians in assessing spinal function and treatment efficacy, potentially benefiting patients through improved monitoring of musculoskeletal and physiological status.
Clinical Evidence
No clinical data provided; bench testing only. Performance specifications for EMG, range of motion, temperature, algometry, and heart rate sensors are documented.
Technological Characteristics
Multi-modality monitor with five sensor types: surface EMG (4 channels, 20-500 Hz), infrared temperature (thermopile), inclinometer, algometer (pressure transducer), and IR plethysmograph. Console features 16-bit A/D converter, isolated USB connectivity, and UL2601-compliant power supply. Materials include flame-retardant ABS and aluminum. Software-based data acquisition and display on PC. Patient isolation via DC-to-DC converter and optically isolated data link.
Indications for Use
Indicated for patients requiring spinal physiological monitoring, including measurement of surface EMG, skin temperature, spinal range of motion, pressure sensitivity in joints/muscles, and heart rate to track treatment progress.
Regulatory Classification
Identification
A diagnostic electromyograph is a device intended for medical purposes, such as to monitor and display the bioelectric signals produced by muscles, to stimulate peripheral nerves, and to monitor and display the electrical activity produced by nerves, for the diagnosis and prognosis of neuromuscular disease.
{0}------------------------------------------------
OCT 1 0 2003
## Insight Millennium III
510(k) Summary
Submitted by:
Company Name: Company Address #1: Company Address #2: Contact Person: Phone Number: Fax Number:
Fasstech 76 Treble Cove Road, #2 North Billerica, MA 01862 Lee Brody 978.663.2800 978.663.0999
(023209
116
ubinasovva
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Submitted on:
September 25, 2002
{1}------------------------------------------------
K023209 2/6
.
ﺍﻟﺘﻨﺎﻭﻳﺔ
#### Section 1: Device Name
Common or Usual Name: Electromyograph, Diagnostic Proprietary Model Name: Insight Millennium III
#### Indications for Use Section 2:
- To measure bilateral differences in surface EMG along the spine 트
- To measure surface EMG along the spine during functional tasks ■
- To measure bilateral differences in skin temperature along the spine
- To measure Range of Motion of the three spinal regions
- To measure patient self-reported pressure sensitivity in joints and muscles ■
- 트 To measure Heart Rate
- 1 To chart patient progress during the course of treatment
#### Device Description Section 3:
The Insight Millennium III is a non-invasive, multi-modality physiologic monitoring device. The Insight Millennium III contains the following five sensor types: (1) surface EMG used to measure muscle activity, (2) infrared temperature sensor used to measure skin temperature, (3) inclinometer used to measure end-point range of motion, (4) algometer used to measure patient self-reported pressure sensitivity, and (5) heart rate sensor to measure heart rate.
#### Hardware
The Insight Millennium III hardware consists of an instrument console and five different sensor types. All five sensor types plug directly into the front panel of the Insight Instrument Console. The Insight Millennium III Instrument Console is powered via a UL2601 listed power supply. The Instrument Console is connected to a personal computer (IBM compatible) via an isolated USB port connection.
#### Software
The Insight Millennium III software displays real-time surface EMG, spinal Range of Motion, skin temperature, pressure, and heart rate readings, allowing the user to ensure that readings are stable prior to data collection. The Insight Millennium software allows the user to: (1) enter patient information, (2) record surface EMG, spinal Range of Motion, skin temperature, pressure and heart readings, (3) graph surface EMG, spinal Range of Motion, skin temperature, pressure and heart rate readings, and (4) print out reports.
{2}------------------------------------------------
#### Section 4: Predicate Device
This section documents the substantial equivalence of the Insight Millennium III to legally marketed devices. Specifically, this section documents the substantial equivalence of Insight Millennium III to the following three products:
| Manufacturer | Predicate Device Name | 510(k) Number |
|-------------------------|-----------------------|---------------|
| Insight Millennium Plus | Fasstech | K011964 |
| Medac System/3 | Davicon | K914925 |
The Insight Millennium III is equivalent to these legally marketed devices in the following ways:
- l The physical characteristics and electrical characteristics (performance characteristics) of the Insight Millennium III are equivalent to the Insight Millennium Plus. The only difference is the addition of a fifth sensor type, the Insight Heart Rate Sensor.
- . The physical characteristics and electrical characteristics (performance characteristics) of the Insight Millennium Heart Rate Sensor is equivalent to the Medac System/3 heart rate sensor
The Insight Millennium III differs from these legally marketed devices in the following ways:
- . The device types listed above have been combined for reasons of user convenience,
This section documents the substantial equivalence of the Insight Millennium III to legally marketed devices. The Insight Millennium III is a modest expansion of the Insight Millennium Plus, a device that we have previously registered with FDA. The Insight Millennium III differs from the Insight Millennium Plus in the following manner:
- 1. Hardware: Addition of a new sensor, The Insight Heart Rate Sensor. The Insight Millennium III Instrument Console is identical to that of the Insight Millennium Plus Console, including patient isolation and PC connection.
- 2. Software: The Insight Millennium III software is extended to include the Insight Heart Rate Sensor.
| Feature | Insight<br>Millennium III | Insight<br>Millennium Plus | Medac System/3 |
|---------------------------------------------------------------|---------------------------|----------------------------|----------------|
| Four Channels of<br>surface EMG | Yes | Yes | N/A |
| Skin | Yes | Yes | N/A |
| Temperature<br>measurement via<br>infrared thermal<br>scanner | | | |
| Range of Motion | Yes | Yes | N/A |
### Predicate Device Comparison Chart
{3}------------------------------------------------
# K023209 4 6
| Sensor | | | |
|-------------------------------------------------|-----|-----|-----|
| Algometer | Yes | Yes | N/A |
| UL-2601 listed<br>power supply | Yes | Yes | N/A |
| Heart-rate<br>measure using IR<br>Phethsmograph | Yes | N/A | Yes |
#### Section 5: Performance Specification
The Insight Millennium III specifications are summarized below:
EMG
| EMG | Electrodes: | 4 ea. Smart Sensors with low-noise<br>preamplifiers integral to electrode assemblies |
|-----------------|-------------------------------|--------------------------------------------------------------------------------------|
| | Calibrated Range: | 0.1 - 999 uV |
| | Input Bias Current: | Less than 2.0 Picoamperes |
| | Differential Input Impedance: | Greater than 1,000,000 Megaohms |
| | Common Mode Rejection: | 150 dB |
| | Bandwidth: | 20-500 Hz (50/60 Hz notch) |
| | Noise: | Less than 0.1 uV (inputs shorted) |
| | Detector: | Log power detector, 250 mS averaging filter. |
| | Controls: | None |
| Range of Motion | | |
| | Range: | 360 degrees |
| | Accuracy: | +/- 1 degree nominal |
| | Controls: | Enter and Skip Buttons |
| | Physical: | Case Material: Impact resistant, flame retardant<br>ABS. |
| | | 3.4"H x 3.5"W x 1.25"D. Weight 6.5 oz. |
| Temperature | | |
| | Calibrated Range: | 55°F - 120°F |
| | Accuracy: | +0.2°F nominal |
| | Sensors: | Two thermopile, fixed 2.5" apart (center-to-<br>center) |
| | Controls: | Enter button |
| | Physical: | Case Material: Impact-resistant, Aluminum with |
| | | 0.5" ABS Plastic Outer Ring. |
| | Size: | 5.5"L x 3.5"W x 2.5"H. Weight 15 oz. |
| Algometer | | |
| | Calibrated Range: | 0-100 lbs. |
| | Accuracy: | +3% nominal |
510(k) Summary Fasstech Insight Millennium III
{4}------------------------------------------------
K023209 5/6
| | Contact Area:<br>Sensor:<br>Controls:<br>Physical:<br>Size: | 1.0 cm²<br>One pressure transducer attached to a stiff rod.<br>Enter button<br>Case Material: Impact-resistant, Aluminum with<br>3.0" ABS Plastic Stiff Rod.<br>5.5"L x 1.75"W x 2.5"H. Weight 9 oz. |
|--------------------|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Heart Rate Sensor | | |
| | Sensor Type | IR Plethsmograph (attached to finger with<br>Velcro) |
| | Output Voltage: | 5 - 50 mV, typical at rest |
| | Output Impedance: | 1 kΩ, nominal |
| | Weight: | 28 grams |
| | Sensor Size: | 15 x 15 x6.3 mm |
| Instrument Console | | |
| | Inputs:<br>Output:<br>A/D converter:<br>Controls:<br>Power:<br>Physical: | 4 each EMG electrodes<br>1 each inclinometer<br>1 each temperature sensor<br>1 each algometer sensor<br>1 each Heart Rate sensor<br>Isolated USB<br>16 bit, 16 channel<br>None<br>12V, 500 mA UL-2601 listed power supply.<br>Case Material: Impact resistant, flame retardant<br>ABS.<br>3.5"H x 8.375"W x 9"D. Weight 3 lbs. 11 oz. |
#### Section 6: Patient Safety
The Insight Millennium III patient isolation is assured by the following two electrical isolation barriers:
- Medical-Grade Universal Power Supply: This device is a UL2601 compliant AC line to 1. 12 VDC, 28 W converter. This internal power supply is partitioned from all other circuitry via an earthed steel chassis. The power supply accepts 85-264 VAC, and 47-63 Hz via IEC 60320 fused input module.
This supply provides low-level isolated power to the Non-Patient Side of the Insight Millennium III circuitry, as well as power to the DC-to-DC isolation converter that powers to the Patient Side (see #2 below).
- DC-to-DC Patient Isolation Converter is a UL2601 compliant DC-to-DC 2. converter, and meets the dielectric withstand and leakage current requirements of the UL2601 standard for Patient Care Equipment with isolated patient leads.
{5}------------------------------------------------
This converter supplies all power to patient-applied parts and related circuitry. Each of the direct patient-applied parts have individual current limiters for fault condition.
In addition, the Patient Side of the Insight Millennium III is isolated from the host PC as follows:
- Signals are converted from analog voltages to 16 bit Optically Isolated Data Link: 3. digital values by the analog-to-digital converter (ADC). The digital data is sent to and from the USB of the PC across an optically isolated data link.
This optical link is UL2601 compliant, providing the dielectric withstand and low leakage current characteristics specified in UL2601.
#### Section 7: Conclusion
The Insight Millennium III is substantially equivalent to the predicate devices. Furthermore, the device is safe and effective for its intended use.
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
OCT 1 0 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Lee Brody Chief Operating Officer Fasstech 76 Treble Cove Road, Bldg. #2 North Billerica, Massachusetts 01862
Re: K023209
Trade/Device Name: Insight Millennium III Regulation Number: 21 CFR 890.1375, 882.5050 Regulation Name: Diagnostic electromyography, Biofeedback device Regulatory Class: II Product Code: IKN, HCC Dated: September 5, 2003 Received: September 9, 2003
Dear Mr. Brody:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2 - Mr. Lee Brody
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Page
510(k) Number (if known):
Device Name: ____Insight Millennium III
Indications for Use:
To measure bilateral differences in surface EMG along the spine To measure surface EMG along the spine during functional tasks To measure bilateral differences in skin temperature along the spine To measure nange of Motion of the three spinal regions To measure patient self-reported pressure sensitivity in joints and muscles . To measure Heart Rate To chart patient progress during the course of treatment
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.