Treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Device Story
MagVita TMS Therapy System is a computerized, electromechanical device for non-invasive transcranial magnetic stimulation (TMS). It uses a capacitor-based circuit to discharge current through a figure-of-eight electromagnetic coil, generating magnetic fields that induce electrical currents in the cerebral cortex. The system is operated by clinicians in a clinical setting to treat Major Depressive Disorder. The clinician uses a marking apparatus and patient head fixation (chair, vacuum pillow) to position the coil over the left dorsolateral prefrontal cortex (DLPFC). The device delivers repetitive pulse trains at 10Hz to modulate cortical excitability. By inducing sufficient current to excite neurons, the system aims to provide therapeutic benefit to patients resistant to antidepressant medication. The clinician monitors the procedure and adjusts stimulation parameters based on the patient's motor threshold.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical performance testing of components (previously cleared under K061645, K071821, and K091940) and reliance on the clinical evidence established for the predicate device.
Technological Characteristics
Electromechanical TMS system; figure-of-eight air-core coil; liquid cooling. Biphasic waveform; 10Hz repetition rate; 290µs pulse width. Standards: ISO 13485:2012, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2. Integrated system includes stimulator, trolley, coil, marking apparatus, and patient chair with vacuum fixation.
Indications for Use
Indicated for treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2015
Tonica Elektronik A/S Lise Terkelsen Regulatory Affairs/ Quality Assurance Specialist Lucernemarken 15 DK-3520 Farum, Denmark
Re: K150641
Trade/Device Name: MagVita TMS Therapy System Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP Dated: March 10, 2015 Received: March 13, 2015
Dear Lise Terkelsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -SD/△
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150641
Device Name MagVita TMS Therapy System
Indications for Use (Describe)
The Mag Vita TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
# 510(k) Summary
# Submitter's Information
| Name of 510(k) owner: | Tonica Elektronik A/S<br>Lucernemarken 15<br>DK-3520 Farum, Denmark |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone: | +45 4499 8444 |
| Fax: | +45 4499 1544 |
| Contact person: | Lise Terkelsen<br>Email: lise.terkelsen@tonica.dk |
| Preparation date: | July 24, 2015 |
| Trade name: | MagVita TMS Therapy System |
| Common name: | Transcranial Magnetic Stimulator |
| Classification name: | Repetitive Transcranial Magnetic Stimulator for treatment of<br>Major Depressive Disorder [21 CFR 882.5805, Product Code<br>OBP] |
| Classification: | Class II Medical Device |
| Predicate Device: | NeuroStar TMS Therapy Systems (K083538, K133408)<br>21 CFR 882.5805, Repetitive Transcranial Magnetic<br>Stimulation<br>Product code: OBP<br>Device Class: II |
{4}------------------------------------------------
## Device description
The MagVita TMS Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic fields to induce electrical currents directed at regions of the cerebral cortex. The Mag Vita TMS Therapy System is indicated for
## Treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode
Transcranial magnetic stimulation (TMS) is a non-invasive technique for stimulating brain and neural tissue. The principle of magnetic stimulation is implicit in Faraday's law. The pulses of current are generated with a circuit containing a capacitor connected to the stimulating coil. With the capacitor charged to a certain level, the conducting state will cause the discharging of the capacitor through the coil. A magnetic field is generated proportional to this current. The rapid change in the magnetic field induces a current in conducting materials e.g. the body tissue. If the current induced in the human body is of sufficient amplitude and duration, it will excite neurons.
In the MagVita TMS therapy system TMS pulses are applied repetitively at a frequency of 10Hz. Such stimulation has been shown to be as effective as the predicate device in modulating cortical excitability. The observed and documented increase in cortical excitability after high frequency (10Hz) repetitive transcranial magnetic stimulation has been shown to persist beyond the duration of the train of stimulation. Repetitive Magnetic stimulation with the MagVita TMS therapy system is applied to the human brain on the left dorsolateral prefrontal cortex (DLPFC).
The MagVita TMS Therapy System is an integrated system consisting of the following components:
- MagPro Stimulator and Trolley .
- 。 MagPro family
- 。Trolley with holding arrangements
- Coil for MT determination .
- 。 Coil C-B60
- Marking apparatus for locating treatment area .
- o Marking plate for Coil C-B60
- 。 Pen for marking, Cap, Ruler
- Patient head fixation .
- 。 Treatment Chair
- o Vacuum Pump and Vacuum pillow
- 。Super Flexible Arm mounted on the trolley
- Coil for Depression Treatment .
- 。 Coil Cool-B65 with Coil Cooler unit
- . Isolation Transformer
{5}------------------------------------------------
#### Intended Use/Indication for Use:
Treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
#### Standards:
The MagVita TMS Therapy System has been tested and complies with the following standards
- ISO 13485:2012 .
- IEC60601-1 .
- IEC60601-1-1 .
- IEC60601-1-2 .
#### Non-Clinical performance data:
The non-clinical performance testing of the components of the Mag Vita TMS Therapy System has been tested as required, and cleared by the FDA earlier on K061645, K071821 and K091940.
These tests along with the supportive clinical information from the predicate device demonstrate that the MagVita TMS Therapy System is as safe and effective as the predicate device for use in treatment of Major Depressive Disorder.
#### Substantial equivalence:
The MagVita TMS Therapy System is substantially equivalent to the predicate device (Neurostar TMS Therapy® System). The MagVita TMS Therapy System and the predicate device have identical intended use /indication for use, and the technological characteristics are very similar such that they in our view can be considered equivalent.
The MagVita TMS Therapy System and the predicate device are both indicated for
## Treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Design of the MagVita TMS Therapy System is similar to the predicate device as both systems apply Transcranial Magnetic Stimulation as repetitive pulse trains at 10Hz delivered as brief rapidly alternating magnetic fields to induce electrical currents over the prefrontal cortex.
Both the MagVita TMS Therapy System and the predicate device have the same components consisting of TMS stimulator with software, electromagnetic coil and an articulated arm for positioning of the treatment coil. The operational procedures including system setup, patient preparations, motor threshold determination, coil positioning and treatment with predefined treatment stimulation parameters are essentially the same.
{6}------------------------------------------------
| Area | MagVita TMS Therapy System | NeuroStar TMS Therapy System<br>Neuronetics Inc. (K083538,K133408) |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance | Waveforms: Biphasic.<br>Frequency: 0.1 -30 pulses per<br>second or 0.1 -100 pulses,<br>depending on model<br>Preset range of % MT: 0% -140%<br>Recommended standard treatment:<br>Stimulation Intensity: 120% MT<br>(MT=Motor Threshold intensity)<br>Repetition rate:10 Hz<br>Train duration: 4 sec<br>Interval between pulses: 26 sec<br>Numbers of pulses/ session: 3000<br>Output Stimulation Parameters:<br>Available Stimulation Amplitude<br>in Standard Motor Threshold<br>(SMT) units<br>Amplitude Range: 0 - 1.7 SMT<br>Pulse width:<br>290 $\mu$ s (±5%), Biphasic sinusoid<br>waveform.<br>Frequency Range:<br>0.1-30 pps (±2%) or 0.1-100 pps,<br>depending on model | Waveforms: Biphasic<br>Frequency: 0.1 -30 pulses per<br>second.<br>Preset range of % MT: 80% -140%<br>Recommended standard treatment:<br>Stimulation Intensity: 120% MT<br>(MT=Motor Threshold intensity)<br>Repetition rate:<br>10 Hz<br>Train duration:<br>4 sec<br>Interval between pulses: 26 sec<br>Numbers of pulses/session: 3000<br>Output Stimulation Parameters:<br>Available Stimulation Amplitude<br>in Standard Motor Threshold<br>(SMT) units<br>Amplitude Range: 0.22 - 1.6 SMT<br>Pulse width:<br>185 $\mu$ s (±10%), Biphasic sinusoid<br>waveform.<br>Frequency Range:<br>0.1-30 pps (±2%) |
| Coil<br>Configuration | Figure-of-eight coil<br>Air core | Figure-of-eight coil<br>Ferromagnetic core |
| Cooling | Liquid cooling | Ferrofluidic cooling |
| Standards<br>met | Company complies with ISO<br>13485:2012. | Company complies with ISO<br>13485:2003 |
| Electrical<br>safety | Complies with IEC60601-1,<br>IEC60601-1-1 and IEC60601-1-2 | Complies with UL60601-1 and<br>UL60601-1-2 |
## Characteristics of the Device as Compared to Predicate Device*
*For a more comprehensive comparison of devices please refer to section 12 Substantial Equivalence Comparison
{7}------------------------------------------------
## Conclusion:
The above comparison, demonstrates and supports the substantial equivalency of the MagVita TMS Therapy System to the NeuroStar TMS Therapy System
The indication for use, the target population, the dosage, the treatment procedure, the treatment position and all relevant protocol parameters (intensity, repetition rate, number of pulses in a train, numbers of trains, number of treatment sessions) are identical for the MagVita TMS Therapy System and the predicate device NeuroStar TMS Therapy System.
The transducer design (figure-of-eight) are equivalent and the realized magnetic properties of the MagVita TMS Therapy System and the predicate device are substantial equivalent for the two coils.
The reliability of the positioning method used by the MagVita TMS Therapy System is, based on the direct relationship of the underlying cortical brain anatomy to the patient's scalp, as is the method used in the predicate device. The method for identifying the correct treatment position in the Mag Vita TMS Therapy System is at least as effective as the method employed by the predicate device.
The MagVita TMS Therapy System does not introduce any new safety considerations in comparison to the predicate device.
All other identified differences between the two systems are minor and without any known impact on safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.