The NeuroStar TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Device Story
Computerized, electromechanical system delivering non-invasive, repetitive transcranial magnetic stimulation (rTMS). System components: mobile console with GUI, gantry-supported treatment coil, head support system with laser-guided alignment, and SenStar Treatment Link. SenStar provides contact sensing, magnetic field quality control, and surface field cancellation. TMS TrakStar system manages patient/treatment data. Operated by clinicians in inpatient/outpatient settings. Device generates brief, rapidly alternating magnetic pulses to induce electrical currents in cerebral cortex. Clinician uses output to guide treatment sessions for MDD. Benefits include non-invasive therapeutic stimulation for treatment-resistant depression.
Clinical Evidence
Prospective, multi-site, randomized, double-blind, sham-controlled trial (OPT-TMS, N=197). Patients (aged 18-70) with moderate-to-severe MDD and 1-4 failed antidepressant trials. Primary endpoint: remission (HAMD24 <3 or two consecutive <10). Active TMS showed statistically significant advantage over sham (13.4% vs 5.0% remission; p=0.0173; OR 4.05). Secondary measures (MADRS, CGI-S, IDS-SR) also favored active TMS. Safety profile showed no deaths or seizures; most common AEs were headache and application site pain, not significantly different from sham.
Technological Characteristics
Electromechanical TMS system. Delivers brief (185 usec nominal) pulsed magnetic fields. Includes mobile console, gantry, treatment coil, and laser-guided head support. SenStar Treatment Link provides contact sensing and field monitoring. Software-controlled GUI. Stand-alone data management system (TMS TrakStar).
Indications for Use
Indicated for adult patients with Major Depressive Disorder who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{0}------------------------------------------------
# NEURONETICS
# 510(k) Summary
# NeuroStar® TMS Therapy System
| 510(k) Owner: | Neuronetics, Inc.<br>3222 Phoenixville Pike<br>Malvern, PA 19355<br>Phone: 610-640-4202<br>Fax: 610-640-4206 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Judy P. Ways, Ph.D.<br>Vice President,<br>Regulatory Affairs and Quality Assurance<br>Neuronetics, Inc.<br>3222 Phoenixville Pike<br>Malvern, PA 19355<br>Phone: 610-981-4107<br>Fax: 610-640-4206 |
| Date Prepared: | 27 March 2014 |
| Proprietary Name: | NeuroStar® TMS Therapy System |
| Common Name: | Transcranial Magnetic Stimulator |
| Classification Name: | Transcranial Magnetic Stimulation System [21 CFR 882.5805, Product<br>Code OBP] |
| Predicate Device: | NeuroStar TMS Therapy® System [K061053/K083538/K130233] |
# Device Description:
The NeuroStar TMS Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation using brief duration (185 usec (nominal)) rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation or TMS. NeuroStar TMS Therapy is a non-invasive tool for the stimulation of cortical neurons for the treatment of adult patients with Major Depressive Disorder (MDD) who have failed to receive satisfactory improvement from prior antidepressant medication as described under Intended Use. The NeuroStar System is used for patient treatment by prescription only under the supervision of a licensed physician. It can be used in both inpatient and outpatient settings including physician's offices and clinics, and hospitals.
{1}------------------------------------------------
# K133408
The NeuroStar TMS Therapy System is an integrated system consisting of a combination of hardware, software, accessories and consumable supplies. It includes a Mobile Console which houses the electronics, includes a software controlled graphical user interface, and gantry that supports the Treatment Coil. The ferromagnetic Treatment Coil delivers the TMS Therapy™. The Head Support System provides accurate positioning of the Treatment Coil using a laserguided alignment system. A single-use device, the SenStar® Treatment Link, which is applied to the Treatment Coil, provides contact sensing to monitor contact of the treatment coil with the patient's head throughout a treatment session, quality control by monitoring the magnetic field level prior to patient treatment, surface field cancellation to reduce stimulation of the scalp, and acts as a hygiene barrier from patient. The TMS TrakStar™ practice data management system consists of a stand-alone computer and data management software that facilitates recording and retrieval of patient and treatment information and communication of data among multiple NeuroStar TMS Systems.
# Intended Use:
The NeuroStar TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
# Technological Characteristics and Substantial Equivalence:
The subject device, NeuroStar TMS System, has the following similarities to the predicate TMS device (NeuroStar TMS Therapy System, K061053/K083538/K130233):
- . Principles of operation
- Design for delivery of Transcranial Magnetic Stimulation (TMS) .
- O Output stimulation parameters (pulse width, frequency, train duration, inter-train interval, etc.)
- Materials .
The proposed change for the NeuroStar TMS Therapy System in this submission is a modification to the current FDA-cleared Indication for Use to expand the indicated population in major depression to adult patients who have failed to benefit from one or more prior antidepressant medications in the current episode. The proposed change is supported by information submitted in this premarket notification and with the following rationale:
- Device is substantially equivalent to the FDA-cleared NeuroStar TMS Therapy System: 1. The NeuroStar TMS Therapy System that is the subject of the premarket notification is the same device cleared by the FDA under 510(k)s K061053, K083538 and K130233.
- Device new clinical data supports the revised Intended Use: New clinical data from a 2. randomized, controlled trial (OPT-TMS RCT) demonstrate the safety and efficacy of the NeuroStar TMS Therapy System in the treatment of patients with major depression as described in the revised Intended Use.
{2}------------------------------------------------
Changes to the device labeling are to include the revised Intended Use and new clinical data. No other changes are made to the device or labeling. Therefore, the NeuroStar TMS Therapy System with the change proposed in the premarket notification is substantially equivalent to the predicate device.
# Clinical Performance:
# Randomized Controlled Trial OPT-TMS
The OPT-TMS randomized controlled trial (RCT) (ClinicalTrials.gov Identifier NCT00149838) was conducted to evaluate the safety and efficacy of NeuroStar TMS Therapy in adult patients (N=197) with moderate to severe major depressive disorder and who failed to benefit from 1-4 adequate antidepressant medication trials as defined using the Antidepressant Treatment History Form (ATHF) (George, et al., 2010).
After determination of protocol eligibility and medication washout, patients participated in a multi-site, parallel group, double-blind, sham-controlled, randomized comparison of active NeuroStar TMS Therapy and sham treatment for a fixed trial period of 3 weeks. At the end of this period, patients who showed a criterion level of improvement were eligible to continue with their blinded, randomized assignment for up to 3 additional weeks, based on twice weekly determinations of clinical progress in a duration-adaptive phase.
Treatment was according to the standard treatment protocol described in the NeuroStar TMS Therapy System labeling. In addition to standard blinding methods, an "active" sham coil was used for blinding against TMS-related stimulation adverse effects. A post-study survey was conducted at the conclusion of the study to assess study blinding which indicated that blinding was effectively maintained.
Enrolled patients were aged 18-70, and experiencing a DSM-IV defined major depressive episode with moderate to severe depression as determined by a baseline HAMD24 ≥20 with a 5 year limit on the duration of the current depressive episode. Patients had a moderate to severe level of current treatment resistance, with at least 1 but no more than 4 ATHF level 3 (i.e., minimum labeled dose, 4 weeks duration) treatment attempts without clinical benefit, or had intolerance to 3 or more antidepressant medications. Excluded were patients who were unlikely to show response according to the TMS literature (presence of psychosis, comorbid and prominent anxiety disorder or substance abuse), and those patients for whom TMS treatment might be unsafe (e.g. those with prior head trauma, seizures, suicidal intent). Because this was an antidepressant medication-free study conducted in outpatients only, patients who could not tolerate being tapered from medications were excluded.
Demographic and clinical features of the enrolled population were not statistically significantly different between the active TMS and sham TMS treatment groups and were similar for patients studied in randomized controlled trial Study 101 (O'Reardon, et al., 2007). Roughly half of the population was female and the average age was 47 years. Patients had moderate to severe major depression by symptom measures. Current episode treatment resistance averaged 1.6 failed research-quality adequate treatment trials (verified by ATHF criteria), which translates
・
{3}------------------------------------------------
approximately to 3 to 6 clinical antidepressant medication attempts in the current episode. During their lifetime, patients had failed 3.3 research-adequate treatment trials (approximately 9 clinical attempts).
The primary outcome measure was the clinically significant categorical outcome of remission, defined as HAMD24 total score <3 or 2 consecutive ratings of a HAMD24 total score <10 between 3-6 weeks according to the pre-specified duration adaptive design. There was a statistically significant advantage of NeuroStar TMS Therapy on remission rate as compared to sham control (P=0.0173) in the ITT sample (N=197) (Table 1). There were 13.4% remitters in the TMS arm and 5.0% in the sham arm: the adjusted odds ratio was 4.05 (95% confidence interval (CI), 1.28-12.83).
| Table 1. OPT-TMS Randomized Controlled Trial Primary Outcome (111, N=197) | | | | |
|---------------------------------------------------------------------------|--------------------|-------------------|----------------------------------|----------------------------------|
| Measure<br>HAMD24 | Active TMS<br>N=97 | Sham TMS<br>N=100 | P-Value<br>(Favoring Active TMS) | Adjusted odds<br>ratio (95% CI)§ |
| Remission | 13.4% | 5.0% | 0.0173 | 4.05 (1.28-12.83) |
| Table 1. OPT-TMS Randomized Controlled Trial Primary Outcome (ITT, N=197) | | | |
|---------------------------------------------------------------------------|--|--|--|
| | | | |
\$ Odds ratios was adjusted for site (categorical), age (continuous), duration of current depressive episode, and medication resistance (low vs, high).
Remission occurred as early as week 3, with additional patients achieving remission across weeks 4-5. None of the pre-specified covariates (including the severity of pre-treatment antidepressant resistance) used in the logistic regression model had a statistically significant effect on the outcome in the pre-specified analysis model, thereby establishing efficacy across the ATHF 1-4 study population.
The baseline to endpoint change score outcome using the HAMD24 also favored active TMS to sham treatment (p=0.0588). Baseline to endpoint outcomes for patients treated with active TMS were statistically significant as compared to sham treatment as measured using the MADRS (P=0.0136), CGI-S (P=0.0181) and the patient-rated IDS-SR (P=0.0008). For the categorical endpoints, higher rates of remission were observed for patients receiving active TMS as compared to sham treatment as measured using the MADRS (P=0.0170) and the patient-rated IDS-SR (P=0.1199), and for response (50% improvement from baseline) for all three outcome measures (HAMD24, P=0.0104; MADRS, P=0.0063; IDS-SR, P=0.0145). Response and remission rates were consistent in clinical magnitude on all three outcome measures (i.e., HAMD24, MADRS, and IDS-SR).
The results for the continuous outcome measures are shown in Table 2 with 95% confidence intervals for the treatment difference, and estimates of the standardized effect size. Standard effect sizes range from 0.43 to 0.67, indicating a moderate to large treatment effect for NeuroStar TMS Therapy in a study population with major depression with a broad spectrum of antidepressant medication treatment resistance (ATHF 1-4) in the current episode.
{4}------------------------------------------------
| Outcome<br>Measure | Treatment<br>Group | Baseline<br>Mean (SD) | Baseline<br>N | End of<br>Acute Phase<br>Mean (SD) | End of<br>Acute Phase<br>N | Treatment<br>Effect<br>(95% CI) | Standardized<br>Effect Size | P-Value |
|--------------------|--------------------|-----------------------|---------------|------------------------------------|----------------------------|---------------------------------|-----------------------------|---------|
| HAMD24 | Active | 26.4 (4.9) | 97 | 21.8 (9.2) | 85 | -2.11 | -0.43 | 0.0588 |
| HAMD24 | Sham | 26.6 (4.9) | 100 | 23.5 (7.4) | 93 | (-4.30, 0.08) | | |
| MADRS | Active | 29.6 (6.9) | 97 | 24.8 (11.5) | 85 | -3.41 | -0.51 | 0.0136 |
| MADRS | Sham | 29.9 (6.5) | 100 | 27.9 (9.0) | 93 | (-6.12, -0.71) | | |
| CGI-S | Active | 4.6 (0.7) | 95 | 4.0 (1.2) | 84 | -0.36 | -0.52 | 0.0181 |
| CGI-S | Sham | 4.6 (0.7) | 100 | 4.3 (0.9) | 92 | (-0.65, -0.06) | | |
| IDS-SR | Active | 41.1 (9.1) | 91 | 32.7 (15.3) | 80 | -6.46 | -0.67 | 0.0008 |
| IDS-SR | Sham | 40.5 (10.1) | 96 | 37.1 (14.0) | 90 | (-10.19, -2.74) | | |
Table 2. OPT TMS RCT: Continuous Outcomes Measures (ITT, N=197)
Safety of the NeuroStar TMS Therapy System was assessed at each study visit by review of spontaneously reported adverse events and separate reporting of all serious adverse events. All adverse events are included in Table 3, regardless of the relationship to the assigned treatment condition determined by the investigator.
| Adverse Event | NeuroStar TMS<br>N=97<br>N(%) | Sham TMS<br>N=100<br>N(%) |
|---------------------------------------|-------------------------------|---------------------------|
| Headache | 31 (32.0) | 23 (23.0) |
| Discomfort at the site of stimulation | 17 (17.5) | 10 (10.0) |
| Insomnia | 7 (7.2) | 10 (10.0) |
| Worsening of depression or anxiety | 6 (6.2) | 8 (8.0) |
| Gastrointestinal | 6 (6.2) | 3 (3.0) |
| Fatigue | 5 (5.2) | 4 (4.0) |
| Muscle aches | 4 (4.1) | 4 (4.0) |
| Vertigo | 2 (2.1) | 2 (2.0) |
| Skin pain | 1 (1.0) | 1 (1.0) |
| Facial muscle twitching | 0 | 1 (1.0) |
| Other | 20 (20.6) | 14 (14.0) |
Table 3. OPT-TMS RCT: Spontaneous Adverse Events (N=197, ITT)
Adverse events were coded to link similar terms to the same events using the most common terms reported in studies of TMS. The code term of "Other" was used for any event that did not match to the code terms specified. The most frequently reported events were headache and application site pain. These events were more frequently reported in the active TMS group but were not significantly different between treatment arms.
Treatment with NeuroStar TMS Therapy was well tolerated. There were no deaths or seizures during this study. The all-cause discontinuation was 16.3% (N=31). Of these, 19.6% (N=18) were allocated to the NeuroStar TMS treatment group, while 13.3% (N=13) were allocated to
{5}------------------------------------------------
# K133408
No patients reported an adverse event as the primary reason for study sham TMS. discontinuation.
There were two serious adverse events, both without long-term sequelae: one patient in the active TMS group had syncope that the investigator deemed unlikely related to the treatment and one patient in the sham group experienced paranoid ideation, possibly related to the study. One serious adverse event occurred before treatment: a patient's depression worsened, likely owing to medication discontinuation in the washout phase, and this patient was not randomized.
# Substantial Equivalence Statement:
The NeuroStar TMS Therapy System that is the subject of this premarket 510(k) notification is the same (substantially equivalent) device cleared by the FDA under premarket notifications K061053, K083538 and K130233. New clinical data from the OPT-TMS randomized. controlled trial demonstrate the safety and efficacy of the NeuroStar TMS Therapy System in the treatment of adult patients with major depression who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
'The NeuroStar® and NeuroStar TMS Therapy® are registered trademarks of Neuronetics, Inc. TMS Therapy™ and NeuroStar TrakStar™ are trademarks of Neuronetics, Inc.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
March 28, 2014
Neuronetics, Inc. c/o Judy P. Ways, Ph.D. Vice President, Regulatory Affairs and Quality Assurance 31 General Warren Boulevard Malvern, PA 19355
Re: K133408
Trade/Device Name: NeuroStar TMS Therapy System Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP Dated: February 26, 2014 Received: February 27, 2014
Dear Dr. Ways:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{7}------------------------------------------------
Page 2 - Judy P. Ways, Ph.D.
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K133408
# Device Name
NeuroStar TMS Therapy System
#### Indications for Use (Describe)
The NeuroStar TMS Theray System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Carlos L. Pena -S
{9}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{10}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized abstract symbol to the right and text arranged in a circle around it. The text reads "DEPARTMENT IN HEALTH & HUMAN SERVICES - USA" in all capital letters. The symbol appears to be a stylized representation of a human figure or a flowing design.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-6609 Silver Spring, MD 20993-0002
March 28, 2014
Orthoscan, Inc. % Mr. Adam Menzies VP Product Development 8212 E Evans Road SCOTTSDALE AZ 85260
Re: K133174
Trade/Device Name: Orthoscan FD Mini C-Arm Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OXO, MQB Dated: February 25, 2014 Received: February 26, 2014
Dear Mr. Menzies:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce .prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 l CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{11}------------------------------------------------
Page 2-Mr. Menzies
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K133174
Device Name OrthoScan FD Mini C-arm
### Indications for Use (Describe)
The OrthoScan FD Mini C-arm is designed to provide the physician with general fluoroscopic visualization of the patient's extremities including, but not limited to, surgical orthopedic procedures and critical emergency care procedures in hospital, emergency care, critical care, or physician office environments.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.7)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.