TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities. EMS: The device is designed to be used to stimulate healthy muscles in order to improve or facilitate muscle performance.
Device Story
Portable, battery-powered (3.7V) TENS and EMS device; 4-channel output; delivers programmed electrical pulses via adhesive electrode pads to skin. User selects from 24 operation modes and 20 intensity levels; LCD displays program, level, time, and channel. Used in home healthcare environment by adult patients. Device transforms electrical energy into biphasic, rectangular pulses; current transferred through lead wires/snap buttons to conductive hydrogel pads. Output parameters (voltage, current, frequency, pulse duration) are software-controlled. Benefits include non-invasive pain relief and muscle performance improvement. Healthcare providers/patients use device to manage muscle soreness or facilitate muscle training.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (ANSI/AAMI/ES 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing for nerve/muscle stimulators (IEC 60601-2-10). Biocompatibility of patient-contacting electrode pads is supported by predicate clearance (K092546) and compliance with ISO 10993-5 and ISO 10993-10.
Indicated for adults (age ≥ 18) for temporary relief of pain in shoulder, waist, back, neck, arms, legs, abdomen, and bottom due to exercise or household strain (TENS), and for stimulation of healthy muscles to improve or facilitate muscle performance (EMS).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 8, 2022
Shenzhen Jianfeng Electronic Technology Co. Ltd. Feng Wen General Manager 902, 903 Jialingyu Industrial Building, Da Pu Road Houting Village, Shajing Town, Baoan District Shenzhen, Guangdong 518104 China
Re: K213741
Trade/Device Name: TENS & EMS Device (Model: FM-B2403A) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH. NGX Dated: December 17, 2021 Received: December 17, 2021
Dear Feng Wen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213741
Device Name TENS & EMS Device (Model: FM-B2403A)
Indications for Use (Describe) TENS:
The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.
EMS:
The device is designed to be used to stimulate healthy muscles in order to improve or facilitate muscle performance.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K213741
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
- 1. Type of 510(k) submission: Traditional Date of the summary prepared: Feb-17-2022
- 2. Submitter's Information Submitter: SHENZHEN JIAN FENG ELECTRONIC TECHNOLOGY CO., LTD. Address: 902, 903 Jialingyu Industrial Building, Da Pu Road, Houting Village, Shajing Town, Baoan District, Shenzhen City, Guangdong Province, 518104, China. Contact Person: Feng Wen Email: wenfeng@fzjhealth.com Tel: +86-755-33858361 Fax: +86-755-33858360
Application Correspondent Company: SHENZHEN JIAN FENG ELECTRONIC TECHNOLOGY CO., LTD. Address: 902, 903 Jialingyu Industrial Building, Da Pu Road, Houting Village, Shajing Town, Baoan District, Shenzhen City, Guangdong Province, 518104, China. Contact Person: Feng Wen Title: General Manager Email: wenfeng@fzjhealth.com Tel: +86-755-33858361 Fax: +86-755-33858360
- 3. Subject Device Information
Trade Name: TENS & EMS Device (Model: FM-B2403A) Common Name: TENS, EMS, Stimulator for pain relief; Classification Name: Powered muscle stimulator; Transcutaneous electrical nerve stimulator for pain relief; Nerve muscle stimulator Review Panel: Physical Medicine; Neurology Product Code: NUH, NGX Regulation Number: 21 CFR 882.5890, 21 CFR 890.5850 Device Classification: Class II
- Use: Over-the-Counter Use (OTC)
- 4. The Predicate Device Information
| Basic Information | Predicate Device |
|-------------------|----------------------------------------------------|
| Manufacturer | Shenzhen Jian Feng Electronic Technology Co., Ltd. |
{4}------------------------------------------------
| Device Name | TENS & EMS Device |
|-------------------|------------------------------|
| Model | FM-B2403 |
| 510(K) Number | K202866 |
| Product Code | NUH, NGX |
| Panel Code | Physical Medicine, Neurology |
| Regulation Number | 21 CFR 882.5890 |
| | 21 CFR 890.5850 |
| Regulation Class | Class II |
#### 5. Device Description / Design of Device
The device of the model FM-B2403A is a portable, battery powered (lithium battery 3.7V DC) multifunction device offering both Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (EMS) qualities in one device, it can be used for arm, shoulder, neck, back, waist, abdomen, and leg.
4 channels that effectively transfer your desired choice of programmed electrical pulses directly through electrode adhesive pads to the suggested area of the body where the electrodes are placed, causing minimal muscle contractions. There are 24 modes of operations.
The electrode pads are cleared by FDA, and 510(k) number is K092546 for the electrode pads. They are used as an accessory to the TENS or EMS device unit, which transmits electrical current to patient skin. The electrical current is first transmitted via the lead wire or snap button then transmitted to the conductive gel which is adhered to patient skin. The electrode pads are composed of a cover, connector lead wire or snap button, conductive carbon film, conductive hydrogel, and an electrode carrier liner. It is non-sterile and intended for single adult patient (age ≥ 18) multiple application use.
Its output waveform is provided 24 programs and 20 adjustable intensity levels. The LCD screen shows the information of program, level, operating time and channel.
#### 6. Indication for Use
#### TENS:
The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.
#### EMS:
The device is designed to be used to stimulate healthy muscles in order to improve or facilitate muscle performance.
7. Technological characteristics and substantial equivalence
- 7.1 The following table is the basic characteristic:
{5}------------------------------------------------
| Contents | New device | Predicate Device | SE | |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------|
| Trade Name | TENS & EMS Device | TENS & EMS Device | SE | |
| Device Model | FM-B2403A | FM-B2403 | / | |
| 510(k) Number | To Be Assigned | K202866 | / | |
| Manufacturer | Shenzhen Jian Feng Electronic<br>Technology Co., Ltd. | Shenzhen Jian Feng Electronic<br>Technology Co., Ltd. | SE | |
| Regulatory Information | 882.5890, 890.5850 | 882.5890, 890.5850 | SE | |
| Classification | Class II | Class II | SE | |
| Product code | NUH, NGX | NUH, NGX | SE | |
| Panel | Physical Medicine; Neurology | Physical Medicine; Neurology | SE | |
| OTC/RX | OTC | OTC | SE | |
| Intended Use | TENS: The device is designed<br>to be used for temporary relief<br>of pain associated with sore<br>and aching muscles in the<br>shoulder, waist, back, neck,<br>upper extremities (arm), lower<br>extremities (leg), abdomen and<br>bottom due to strain from<br>exercise or normal household<br>work activities.<br>EMS: The device is designed<br>to be used to stimulate healthy<br>muscles in order to improve or<br>facilitate muscle performance. | TENS: The device is designed<br>to be used for temporary relief<br>of pain associated with sore<br>and aching muscles in the<br>shoulder, waist, back, neck,<br>upper extremities (arm), lower<br>extremities (leg), abdomen and<br>bottom due to strain from<br>exercise or normal household<br>work activities.<br>EMS: The device is designed<br>to be used to stimulate healthy<br>muscles in order to improve or<br>facilitate muscle performance. | SE | |
| Apply parts of the body | Shoulder, waist, back, neck,<br>upper extremities (arm), lower<br>extremities (leg), abdomen and<br>bottom | Shoulder, waist, back, neck,<br>upper extremities (arm), lower<br>extremities (leg), abdomen and<br>bottom | SE | |
| Power Source | Built-in 3.7V lithium battery | Built-in 3.7V lithium battery | SE | |
| - Method of Line Current<br>Isolation | BF | BF | SE | |
| - Patient Leakage Current | / | / | SE | |
| - Normal condition | <10μΑ | <10μΑ | SE | |
| - Single fault condition | <50μΑ | <50μΑ | SE | |
| Number of Output Modes | TENS:12, EMS:12 | TENS:19, EMS:5 | SE | |
| Number of Output<br>Channels | 4 | 4 | SE | |
| Synchronous or<br>alternating | Synchronous | Synchronous | SE | |
| Method of Channel<br>Isolation | By electrical circuit and<br>software | By electrical circuit and<br>software | SE | |
| Regulated Current or<br>Regulated Voltage | Regulated voltage control | Regulated voltage control | SE | |
| Software/Firmware/Microp<br>rocessor Control? | Software | Software | SE | |
| | | | | |
| Automatic Overload Trip? | | No | No | SE |
| Automatic No-Load Trip? | | No | No | SE |
| Automatic Shut Off? | | Yes | Yes | SE |
| Patient Override Control? | | Yes | Yes | SE |
| Indicator<br>Display | On/Off Status? | Yes | Yes | SE |
| | Low Battery? | Yes | Yes | SE |
| | Voltage/Current<br>Level? | Yes | Yes | SE |
| Timer Range (minutes) | | 10~80 | 10~80 | SE |
| Compliance with<br>Voluntary Standards? | | Yes.<br>ANSI/AAMI/ES 60601-1<br>IEC60601-1-2<br>IEC 60601-2-10<br>IEC60601-1-11<br>ISO10993-5<br>ISO10993-10 | Yes.<br>ANSI/AAMI/ES 60601-1<br>IEC60601-1-2<br>IEC 60601-2-10<br>IEC60601-1-11<br>ISO10993-5<br>ISO10993-10 | SE |
| Accessories | | Self-adhesive electrodes,<br>electrode wires, adapter, USB<br>cable | Self-adhesive electrodes,<br>electrode wires, adapter, USB<br>cable | SE |
| Compliance* with 21 CFR<br>898? | | Yes | Yes | SE |
| Weight(g) | | 86g | 82 g | SE |
| Dimensions (mm) [D x W<br>H] | | 110*61*24 | 110*60*15 | SE,<br>Note 1 |
| Housing Materials and<br>Construction | | ABS | ABS | SE |
| Biocompatibility | | All user directly contacting<br>materials are compliance with<br>ISO10993-5 and ISO10993-10<br>requirements | All user directly contacting<br>materials are compliance with<br>ISO10993-5 and ISO10993-10<br>requirements | SE |
### Table 1. FM-B2403A
{6}------------------------------------------------
Note 1:
The proposed device FM-B2403A has passed the IEC 60601-1 and AAMI / ANSI ES60601-1 test . The weight, dimensions, appearance of proposed device FM-B2403A has a little different from predicate device K202866, but these differences are insignificant and won't raise any new risk of safety and effectiveness..
7.2 The following table is the output parameters:
Table 3 FM-B2403A
| Contents | Proposed Device | Predicate Device | SE |
|-----------------------------------------------------------|-----------------|------------------|----|
| Device Name, Model | FM-B2403A | FM-B2403 | / |
| Waveform (e.g., pulsed monophasic,<br>biphasic) | biphasic | biphasic | SE |
| Shape (e.g., rectangular, spike,<br>rectified sinusoidal) | rectangular | rectangular | SE |
{7}------------------------------------------------
| Maximum Output Voltage (volts) (+/-<br>10%) | 98@500Ω<br>150@2KΩ<br>165@10KΩ | 97.6@500Ω<br>135@2KΩ<br>157@10KΩ | SE<br>Note 1 |
|--------------------------------------------------------------------------------------------------------|---------------------------------|-------------------------------------|---------------|
| Maximum Output Current (mA) (+/-<br>10%) | 196@500Ω<br>75@2KΩ<br>16.5@10KΩ | 195.2@500Ω<br>67.5@2KΩ<br>15.7@10KΩ | SE<br>Note 1 |
| Pulse Duration (us) | 50-180 | 90 | SE,<br>Note 1 |
| Frequency† (Hz) [or Rate† (pps)] | <199 | < 90.9 | SE<br>Note 1 |
| Net Charge (micro coulombs (µC)<br>per pulse) (If zero, state method of<br>achieving zero net charge.) | 0.001@500Ω | 0.001@500Ω | SE |
| Maximum Phase Charge, (µC) | 35@500Ω | 15.5@500Ω | SE<br>Note 2 |
| Maximum Current Density(mA/cm2) | 0.5645@500Ω | 0.1192@500Ω | SE<br>Note 2 |
| Maximum Power Density(mW/cm2) | 0.003984@500Ω | 0.176@500Ω | SE<br>Note 2 |
| Pulses per burst | 2 | 2 | SE |
| Bursts per second | 1/50 | 1/30 | SE<br>Note 2 |
| Burst duration (ms) | 50 | 45 | SE<br>Note 2 |
| Duty Cycle: Line (b) x Line (%) | 7.2% | 1.6% | SE<br>Note 2 |
| ON Time (seconds) | 1 | 1 | SE |
| OFF Time (seconds) | 1 | 1 | SE |
#### Comparison:
Note 1: There are some differences on the maximum output voltage .maximum Output current. pulse duration, frequency between proposed device and predicate device K202866.All these parameters have passed IEC 60601-2-10 test codes. Therefore, these differences won't raise any new risk of safety and effectiveness.
Note 2:There are some differences on the net charge, maximum current density ,maximum power density, bursts per second, burst duration and duty cycle between the proposed device and the predicate device K202866,but these parameters don't exceed the safety limit and have passed IEC 60601-2-10 test. The maximum average power density <0.25Watts/cm². Therefore, these differences won't raise any new safety and effectiveness risk.
8. Non-clinical studies and tests performance:
Non-clinical tests have been conducted to verify that the transcutaneous electrical nerve stimulator and/or powered muscle stimulator meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the targeted device complies with the following standards:
ANSI/AAMI/ES 60601-1, Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance
IEC60601-1-11, Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC60601-2-10, Medical electrical equipment -- Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
IEC60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests.
{8}------------------------------------------------
The body-contacting components of this device are electrode patches. We have directly purchased the electrode patches from qualified supplier which has obtained FDA clearance with a 510(k) number of K092546, so we have reason to believe that the electrode patches are safe for the users. The electrode patches comply to the following standards:
1) ISO 10993-5, Biological Evaluation of Medical Devices -- Part 5: Tests for In Vitro Cytotoxicity;
2) ISO 10993-10, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
- 9. Clinical Performance Data Not applicable.
- 10. Final conclusion:
After analyzing non-clinical laboratory studies and safety testing data, it can be concluded that the model FM-B2403A is substantially equivalent to the predicate device K202866.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.