K182120 · Omron Healthcare, Inc. · NUH · Jan 10, 2019 · Neurology
Device Facts
Record ID
K182120
Device Name
Maxpower Relief
Applicant
Omron Healthcare, Inc.
Product Code
NUH · Neurology
Decision Date
Jan 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for the relief of pain associated with sore or aching, muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household work activities. When used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis, use the Shoulder mode of stimulation.
Device Story
Small, battery-operated TENS device for OTC pain relief; applies electrical stimulation via electrode pads to lower back, arms, legs, shoulder, or foot. User selects stimulation modes; device delivers regulated current pulses to nerves/muscles. Used in home, clinic, or hospital settings by adult patients. Output affects pain perception; provides symptomatic relief for muscle strain, chronic intractable pain, and arthritis. Device is software-controlled; includes automatic shut-off and no-load trip features. No connection to AC mains or other devices.
Clinical Evidence
No clinical data. Bench testing only, including verification of controls, indicators, battery state, waveforms, and compliance with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-10 standards. Biocompatibility assessed per ISO 10993-1 for limited duration skin contact.
Technological Characteristics
Battery-operated (2x AAA) TENS stimulator; regulated current output; 9 stimulation modes; 1 output channel. Patient-contacting materials (Long Life Pads) identical to predicate. Complies with ES60601-1, IEC 60601-1-2, IEC 60601-2-10, and IEC 60601-1-11. Non-sterile, surface-contacting device.
Indications for Use
Indicated for adults for relief of pain in lower back, arms, legs, shoulder, or foot due to exercise or household strain; includes symptomatic relief/management of chronic, intractable pain and arthritis-related pain using Shoulder mode. Contraindicated for patients with cardiac pacemakers, implanted defibrillators, or other implanted metallic/electronic devices.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure. The FDA part of the logo has the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 10, 2019
Omron Healthcare, Inc. % Paul Dryden Consultant Omron Healthcare, Inc. c/o ProMedic, LLC. 131 Bay Point Dr. NE St. Petersburg, Florida 33704
Re: K182120
Trade/Device Name: Omron Healthcare Maxpower Relief Model PM3032B Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NYN Dated: October 11, 2018 Received: October 12, 2018
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Pamela D. Scott -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K182120
Device Name
Omron Healthcare Maxpower Relief Model PM3032B
Indications for Use (Describe)
This device is intended for the relief of pain associated with sore or aching, muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household work activities. When used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis, use the Shoulder mode of stimulation.
Environments of Use: Clinics, hospital and home environments Patient Population: Adult
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary Page 1 of 10 1/7/2019
| Omron Healthcare, Inc.<br>1925 West Field Court<br>Lake Forest, IL 60045 USA | Tel - 847-247-5626 |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| | Fax- 847-680-6269 |
| Official Contact: | Renee Thornborough – Executive Director QA/RA |
| Proprietary or Trade Name: | Omron Healthcare Maxpower Relief Model PM3032B |
| Common/Usual Name: | Transcutaneous electrical nerve stimulator for pain relief |
| Classification Name/Code:<br>21CFR 882.5890 | NUH - stimulator, nerve, transcutaneous, over-the-counter<br>NYN - stimulator, electrical, transcutaneous, for arthritis<br>Class II |
| Device Name: | Omron Healthcare Maxpower Relief Model PM30321 |
| Predicate Devices: | K141978 - Omron - PM3032<br>K172079 – Omron - Avail |
#### Device Description:
The Omron Healthcare Maxpower Relief Model PM3032B is a small, battery operated TENS device for pain relief intended for OTC use. It complies with ES60601-1, IEC60601-1-2, IEC60601-2-10 and IEC 60601-1-11.
The output modes are intended for application to the following areas: lower back, arms, legs, shoulder or foot. The specifications of each mode will be discussed in greater detail later in this section.
The Omron Healthcare Maxpower Relief Model PM3032B is software controlled which includes all functions, indicators and waveform characteristics.
This device is intended for the relief of pain associated with sore or aching, muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household work activities. When used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis, use the Shoulder mode of stimulation.
The accessories include an electrode cord / cable and electrodes pads (Long Life) which are placed on the specific body part. These are identical to the accessories cleared in K141978.
As above the device is battery powered there is no connection to AC mains supply nor connection to any other device.
{4}------------------------------------------------
#### 510(k) Summary Page 2 of 10 1/7/2019
This submission is for expanding the indications for use to include indications associated with product code NYN. The device is identical to the Maxpower Relief (Model PM3032) cleared under K141978 except the CPU and related components are changed for cost saving purpose. UDI has also added on the main unit and the packaging. There are no other changes (appearance, dimension, materials, waveforms, modes). The accessories are identical to the Maxpower Relief (Model PM3032) cleared under K141978
We have provided the information suggested in Draft Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use from April 5, 2010. A checklist in accordance with this guidance can be found in Section 2 of this submission
Intended User OTC
### Patient Population
Adults
#### Indications for Use:
This device is intended for the relief of pain associated with sore or aching, muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household work activities.
When used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis, use the Shoulder mode of stimulation.
#### Environment of Use:
Clinics, hospital and home environments
#### Contraindications:
Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electric shock, burns, electrical interference, or death.
#### Predicate Device Comparison:
Table 1 outlines the features of the Omron Healthcare Maxpower Relief Model PM3032B and compares it to the predicates that are being used to establish substantial equivalence.
����������������������������������������������������������������������������������������������������������������������������������������������������������������������������
{5}------------------------------------------------
# 510(k) Summary
#### Device Comparison
| | New Device | Primary Predicate | Secondary Predicate | Comment | |
|---------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Feature | Omron Healthcare<br>Maxpower Relief Model | Device<br>PM3032 | Device<br>Omron Avail | | |
| | PM3032B | (MaxPower relief) | K172079 | | |
| | | K141978 | | | |
| Indications for use | This device is intended for the<br>relief of pain associated with<br>sore or aching, muscles of the<br>lower back, arms, legs,<br>shoulder, or foot due to strain<br>from exercise or normal<br>household work activities.<br>When used for the<br>symptomatic relief and<br>management of chronic,<br>intractable pain and relief of<br>pain associated with arthritis,<br>use the Shoulder mode of<br>stimulation.<br>Environments of Use: Clinics,<br>hospital and home<br>environments<br>Patient Population: Adult | This device is intended<br>for the relief of pain<br>associated with sore or<br>aching, muscles of the<br>lower back, arms, legs,<br>shoulder, or foot due to<br>strain from exercise or<br>normal household work<br>activities.<br>Environments of Use:<br>Clinics, hospital and<br>home environments<br>Patient Population: Adult | The Avail is intended for<br>the relief of pain<br>associated with sore or<br>aching muscles of the<br>lower back, arms, legs,<br>shoulders or feet due to<br>strain from exercise or<br>normal household work<br>activities.<br>When used for the<br>symptomatic relief and<br>management of chronic,<br>intractable pain and relief<br>of pain associated with<br>arthritis, use the Tap,<br>Shoulder, Arm or Leg<br>mode of stimulation.<br>Environments of Use:<br>Clinics, hospital and<br>home environments<br>Patient Population: Adult | Identical to<br>Maxpower Relief<br>for temporary<br>relief of pain with<br>sore and aching<br>muscles.<br>Similar to Avail<br>with respect to<br>symptomatic relief<br>and management<br>of chronic,<br>intractable pain,<br>and relief of pain<br>associated with<br>arthritis. | |
| Environments of Use: | Clinics, hospitals and home<br>environments | Clinics, hospitals and<br>home environments | Clinics, hospitals and<br>home environments | Identical | |
| Patient Population | Adult | Adult | Adult | Identical | |
| Classification - Regulation | 21 CFR §882.5890,<br>Transcutaneous electrical | 21 CFR §882.5890,<br>Transcutaneous electrical | 21 CFR §882.5890,<br>Transcutaneous electrical | Identical | |
| Feature | New Device<br>Omron Healthcare<br>Maxpower Relief Model<br>PM3032B | Primary Predicate<br>Device<br>PM3032<br>(MaxPower relief)<br>K141978 | Secondary Predicate<br>Device<br>Omron Avail<br>K172079 | Comment | |
| | nerve stimulator for pain<br>relief | nerve stimulator for pain<br>relief | nerve stimulator for pain<br>relief | | |
| Classification - Product Code | Primary: NUH - stimulator,<br>nerve, transcutaneous, over-<br>the-counter<br>Secondary: NYN - stimulator,<br>electrical, transcutaneous, for<br>arthritis. | NUH - stimulator, nerve,<br>transcutaneous, over-the-<br>counter | Primary: NUH -<br>stimulator, nerve,<br>transcutaneous, over-the-<br>counter<br>Secondary: NYN -<br>stimulator, electrical,<br>transcutaneous, for<br>arthritis. | Proposed device<br>and secondary<br>predicate device<br>have the same<br>medical device<br>product code. | |
| Prescriptive | No, OTC | No, OTC | No, OTC | Identical | |
| Contraindications/Warning/Precautions | Contraindications<br>Do not use this device if you<br>have a cardiac pacemaker,<br>implanted defibrillator, or<br>other implanted metallic or<br>electronic device. Such use<br>could cause electric shock,<br>burns, electrical interference,<br>or death. | Contraindications<br>Do not use this device if<br>you have a cardiac<br>pacemaker, implanted<br>defibrillator, or other<br>implanted metallic or<br>electronic device. Such<br>use could cause electric<br>shock, burns, electrical<br>interference, or death. | Contraindications<br>Do not use this device if<br>you have a cardiac<br>pacemaker, implanted<br>defibrillator, or<br>otherimplanted metallic<br>or electronic device. Such<br>use could cause electric<br>shock, burns, electrical<br>interference, or death. | Identical | |
| Single Use | Pads are for single patient use | Pads are for single patient<br>use | Patient-contacting Pads<br>are for single patient use | Identical to<br>MaxPower Relief | |
| Sterility | External contacting device,<br>nonsterile | External contacting<br>device, nonsterile | External contacting<br>device, nonsterile | Identical | |
| Over-the-Counter (OTC) | Yes | Yes | Yes | Identical | |
| | | | | | |
| 1/7/2019 | | | | | |
| Feature | | New Device<br>Omron Healthcare<br>Maxpower Relief Model<br>PM3032B | Primary Predicate<br>Device<br>PM3032<br>(MaxPower relief)<br>K141978 | Secondary Predicate<br>Device<br>Omron Avail<br>K172079 | Comment |
| Power Source(s) | | Two AAA alkaline batteries | Two AAA alkaline<br>batteries | Rechargeable lithium Ion<br>battery | Identical to<br>Maxpower Relief |
| - Method of Line Current Isolation | | N.A.(internal power source) | N.A.(internal power<br>source) | N.A.(internal power<br>source) | Identical |
| Patient Leakage Current - Normal<br>Condition (uA) | | 1 | 1 | <10uA | Identical to<br>Maxpower Relief |
| Patient Leakage Current - Single Fault<br>Condition (uA) | | 8.9 max | 8.9 max | <50uA | Identical to<br>Maxpower Relief |
| Average DC current through electrodes<br>when device is on but no pulse are<br>being applied (uA) | | 0 (uA) | 0 (uA) | 0 (uA) | Identical |
| Number of output Modes | | 9 TENS modes | Same | 9 TENS modes<br>1 Microcurrent mode | Identical to<br>Maxpower Relief |
| Number of output<br>channels | Synchronous or<br>Alternating | 1 ch | 1 ch | 1 ch | Identical |
| | Method of Channel<br>Isolation | None | None | None | Identical |
| | Regulated Current or Regulated Voltage | Regulated Current | Regulated Current | Regulated Current | Identical |
| Software/Firmware/Microprocessor<br>Control | | Microprocessor | Microprocessor | Microprocessor | Identical |
| Automatic Overload Trip | | No | No | No | Identical |
| Automatic No-Load Trip | | Yes | Yes | Yes | Identical |
| Automatic shut Off | | Yes | Yes | Yes | Identical |
| | Feature | New Device<br>Omron Healthcare<br>Maxpower Relief Model<br>PM3032B<br>1/7/2019 | Primary Predicate<br>Device<br>PM3032<br>(MaxPower relief)<br>K141978 | Secondary Predicate<br>Device<br>Omron Avail<br>K172079 | Comment |
| | User over ride control | Yes, Power On/Off button | Yes, Power On/Off<br>button | Yes, Power On/Off<br>button on the device and<br>in the App software | Identical to<br>Maxpower Relief |
| | Indication display | ON/Off status | Yes | Yes on App and LED<br>indicator on main unit | Identical to<br>Maxpower Relief |
| | | Low Battery | Yes | Yes on App | Identical to<br>Maxpower Relief |
| | | Voltage/Current<br>Level | Yes | Yes on App | Identical to<br>Maxpower Relief |
| | Timer Range (minutes)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.