The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. And to be used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
Device Story
Wireless Pain Relieve Device (Model LT5011C) is a TENS device for home use by adults. It delivers electrical impulses via self-adhesive electrodes placed on skin near pain sites to modify pain perception. Device is controlled via mobile app (iOS/Android) or remote control using Bluetooth Low Energy. User selects from 23 pre-set TENS programs with adjustable intensity. Device features two output channels, automatic overload/no-load trip, and automatic shut-off. Healthcare providers do not operate the device; patients use it independently to manage pain, potentially reducing reliance on other pain management methods.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-5/10), electrical safety (IEC 60601-1, IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), and output waveform verification (IEC 60601-2-10).
Indicated for adults for temporary relief of pain in shoulder, waist, back, arms, and legs due to exercise or household strain; symptomatic relief and management of chronic, intractable pain; and relief of pain associated with arthritis.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 22, 2018
Shenzhen Dongdixin Technology Co.,Ltd. Siping Yuan R.A. Specialist No. 3 Building Xiliba imang Xusheng Industrial Estate Nansha Shenzhen, China 518108
Re: K180331
Trade/Device Name: Wireless Pain Relieve Device, Model LT5011C Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NYN, NUH Dated: May 15, 2018 Received: May 21, 2018
Dear Siping Yuan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Pamela D. Scott -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180331
Device Name Wireless Pain Relieve Device Model: LT5011C
#### Indications for Use (Describe)
The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. And to be used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
Type of Use (Select one or both, as applicable)
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 201.5 Labeling)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 201.5 Labeling)</label> |
|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is composed of the company name in a bold, sans-serif font, with the "X" in "DIXIN" stylized with an orange line. To the right of the English name is the company name in Chinese characters.
005_510 (k) Summary Version: 1.1
# 510(k) SUMMARY
as required by section 21 CFR 807.92
Wireless Pain Relieve Device
| Date of Submission: | 02/01/2018 |
|---------------------|---------------------------------------------------------------------------------------|
| Submitter's Name: | Shenzhen Dongdixin Technology Co., Ltd |
| Address: | No. 3 Building, Xilibaimang Xusheng Industrial Estate Nanshan, Shenzhen, CHINA 518108 |
| E-mail: | ra@dundex.com |
| Tel: | +86(755) 27652471 |
| FAX: | +86(755) 27652674 |
| Contact: | Siping Yuan |
{4}------------------------------------------------
005 510 (k) Summary Version: 1.1
TECHNOLOGY CO., LTD.
| 1. Proposed Device:<br> | |
|-------------------------|--|
|-------------------------|--|
Device Name: Wireless Pain Relieve Device
Model: LT5011C
| Device classification Name: | Stimulator, Nerve, Transcutaneous, Over-The-Counter |
|-------------------------------|------------------------------------------------------------|
| Regulation Description: | Transcutaneous electrical nerve stimulator for pain relief |
| Regulation Medical Specialty: | Neurology |
| Review Panel: | Neurology |
| Regulation Number: | 882.5890 |
| Product Code: | NUH |
| Device Class: | II |
| Device classification Name: | Stimulator, Electrical, Transcutaneous, For Arthritis |
| Regulation Description: | Transcutaneous electrical nerve stimulator for pain relief |
| Regulation Medical Specialty: | Neurology |
| Review Panel: | Neurology |
| Regulation Number: | 882.5890 |
| Product Code: | NYN |
| Device Class: | II |
#### 2. Predicate Device:
| Legally Marketed Device: | Wireless Pain Relieve Device LT5018C |
|--------------------------|---------------------------------------|
| 510(k) Number: | K173462 |
| Manufacturer: | Shenzhen Dondixin Technology Co., Ltd |
#### 3. Description of Proposed Device:
LT5011C is designed to be used at home, by adults of all genders.
#### Explanation of how the device functions:
The Wireless Pain Relieve Device includes TENS mode. Transcutaneous Electrical Nerve Stimulation (TENS) is a noninvasive, drug free method of controlling pain.
#### Scientific concepts that from the basis for the device:
TENS uses tiny electrical impulses sent through the skin to nerves to modify your pain perception. TENS does not cure any physiological problem; it only helps control the pain, this activates the underlying sensory nerves.
For Transcutaneous Electrical Nerve Stimulation (TENS) Self-adhesive electrodes are placed on the skin close to the area of pain. The user can choose 23 pre-set TENS programs. In each program, the intensity can be adjusted.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is composed of the company name in a bold, sans-serif font, with the letters 'X' stylized in orange. To the right of the English name, the company name is also written in Chinese characters.
The Wireless Pain Relieve Device is controlled by means of an APP on a mobile device (phone or table) or remote control. The communication is done via Bluetooth Low Energy. The APP operates on iOS or Android platforms (iOS 8.0 or greater, Android 5.0 or greater).
#### 4. Description of All Device Modification(s)
Compared with the existing device (LT5018C), the following main parts of Wireless Pain Reliever LT5018C have changed:
- 1) Change the dimensional specification and change the hardware based on the new dimension of enclosure. But the fundamental scientific technology does not change.
- 2) Remove the EMS program and indications for use of EMS mode.
- 3) Remove the neck program and indications for use of neck pain.
- Split the cycle program of unmodified device into four programs for users to choose their 4) comfortable programs and add custom program for users to choose their comfortable programs.
Based on the 21CFR820.30 requirement, this is belong to design change and shall have been evaluated according to the standard IEC60601-2-10, and IEC 60601-1-2. We have conducted these standard test and the results is Passed.
### 5. Proposed Device Intended for Use Statement:
Device Name: Wireless Pain Relieve Device, Model: LT5011C
#### Indications for Use:
The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. And to be used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue and yellow. The text "DONGDIXIN" is in large, bold letters, with the "X" in "DIXIN" being yellow. Below "DONGDIXIN" is the text "TECHNOLOGY CO., LTD." in smaller letters, also in blue. To the right of the English text is the company name in Chinese.
#### 6. Technological Characteristics and Substantial Equivalence
Both the IT5011C and Predicate Device utilize the application of electrodes placed on the skin for pain control. The impulses are generated by the device and delivered through electrodes on the skin in direct proximity to the (painful) muscles to be stimulated. Basic technological characteristics, new device vs. Predicate device
| | | New device | Predicate device | S.E. Discussion |
|----|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | 510K# | To be assigned | K173462 | N/A |
| 2 | Device Name and<br>Model | Wireless Pain Relieve device<br>Mode: LT5011C | Wireless Pain Relieve device<br>Mode: LT5018C | N/A |
| 3 | Manufacturer | Shenzhen Dongdixin Technology Co., Ltd | Shenzhen Dongdixin Technology Co., Ltd | Same |
| 4 | Intended for use | TENS: The device is designed to be used for<br>temporary relief of pain associated with<br>sore and aching muscles in the shoulder,<br>waist, back , upper extremities (arm) and<br>lower extremities (leg) due to strain from<br>exercise or normal household work<br>activities. And to be used for the<br>symptomatic relief and management of<br>chronic, intractable pain and relief of pain<br>associated with arthritis. | TENS: The device is designed to be used<br>for temporary relief of pain associated<br>with sore and aching muscles in the<br>shoulder, waist, back, neck, upper<br>extremities (arm) and lower extremities<br>(leg) due to strain from exercise or normal<br>household work activities. And to be used<br>for the symptomatic relief and<br>management of chronic, intractable pain<br>and relief of pain associated with arthritis.<br>EMS: The device is designed to be used<br>for stimulate healthy muscles in order to<br>improve and facilitate muscle<br>performance. | Similar, the intended use of<br>LT5011C is fallen within the<br>intended use of predicate<br>device. |
| 5 | Power Source | DC 3.7V Li-ion Battery(main device)<br>DC 3.0V, 2 x AAA Batteries (Remote control) | DC 3.7V Li-ion Battery(main device)<br>DC 3.0V, 2 x AAA Batteries (Remote<br>control) | Same |
| | -Method of<br>current isolation | N/A | N/A | Same |
| | - Patient Leakage<br>Current (μΑ) | | | Same |
| | -Normal condition | 0uA | 0uA | |
| | -Single fault<br>condition | 2.0uA | 2.0uA | |
| 6 | Average DC current through electrodes when device is on but no pulses are being applied (µA) | 0 | 0 | Same |
| 7 | Number of Output Modes | 1 (TENS) | 2 (TENS/EMS(including MASSAGE)) | Different, LT5011C only has TENS mode and LT5011C remove indications for use of EMS mode accordingly. |
| 8 | Number of Output Channels | | | |
| | Synchronous or Alternating? | Alternating | N/A | Different, LT5011C has two channels and have been evaluated and passed the test according to the requirement of IEC60601-2-10. |
| | Method of Channel Isolation | By Electrical Circuit and Software | N/A | Different, LT5011C has two channels and have been evaluated and passed the test according to the requirement of IEC60601-1. |
| 9 | Regulated Current or Regulated Voltage? | Current control | Current control | Same |
| 10 | Software/Firmware/Micro processor Control? | Yes | Yes | Same |
| 11 | Automatic Overload Trip | Yes | Yes | Same |
| 12 | Automatic No Load contact Trip | Yes | Yes | Same |
| 13 | Automatic Shut off | Yes | Yes | Same |
| 14 | User Override Control? | Yes<br>Power on/off button on the device<br>Power on/off on the remote control<br>Power on/off in the APP software | Yes<br>Power on/off button on the device<br>Power on/off on the remote control<br>Power on/off in the APP software | Same |
| 15 | Indicator Display: | | | |
| | On/Off Status? | Yes | Yes | Same |
| | Low Battery? | Yes | Yes | Same |
| | Voltage/ Current<br>Level? | Yes | Yes | Same |
| 16 | Timer Range (minutes) | 5-90 minutes | 30 minutes | Different, new device have been<br>evaluated and passed the test<br>according to the requirement of<br>IEC60601-2-10.<br>User can set the treatment time<br>according their needs. |
| 17 | Weight (grams.) | 33grams (Main device)<br>55 grams (Remote control) | 36grams (Main device)<br>65 grams (Remote control) | Different, the new device have<br>been evaluated and passed the<br>testing according to the<br>requirement of IEC60601-1. |
| 18 | Dimensions (cm.) H*W *<br>L | 80.5(L)x45(W)x13.5(H)mm (Main device)<br>127(L)x48.7(W)x24.2(H)mm (Remote control) | 360(L)x59(W)x11.5(H)mm (Main device)<br>115(L)x53(W)x25(H)mm (Remote control) | |
| 19 | Housing Materials &<br>Construction | ABS | ABS | Same |
| 20 | Compliance with 21 CFR<br>898 | Yes | Yes | Same |
| 21 | Compliance with<br>Voluntary Standards? | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10,<br>IEC 60601-1-11<br>ISO10993-5/10 | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10,<br>IEC 60601-1-11<br>ISO10993-5/10 | Same |
| | | New device | Predicate device | S.E. Discussion |
| 1 | 510K# | To be assigned | K173462 | N/A |
| 2 | Device Name or<br>Program Name | Wireless Pain Relief Device<br>Mode:LT5011C | Wireless Pain Relief Device<br>Mode:LT5018C | N/A |
| 3 | Manufacturer | Shenzhen Dongdixin Technology Co., Ltd | Shenzhen Dongdixin Technology Co., Ltd | Same |
| 4 | Waveform | Biphasic | Biphasic | Same |
| 5 | Shape | Rectangular | Rectangular | Same |
| 6 | Max Output Voltage (V) ±20% | | | |
| 7 | 500Ω | 30.5 | 31 | Similar, the new device have<br>been evaluated and passed<br>the test according to the<br>requirement of IEC60601-2-10 |
| 8 | 2kΩ | 65 | 66 | |
| 9 | 10kΩ | 65 | 66 | |
| 10 | Pulse Duration<br>(µsec) | 100~200us | 150~250us | |
| 11 | Frequency (Hz) | 2~100Hz | 2~100Hz | Same |
| 12 | Maximum Phase<br>Charge (uC) 500Ω | 24 | 31 | Similar, the new device have<br>been evaluated and passed<br>the test according to the<br>requirement of<br>IEC60601-2-10. |
| 13 | Maximum Current<br>Density 500Ω | 0.295mA/cm² | 0.48mA/cm² | |
| 14 | Maximum Average<br>Current (average<br>absolute value),<br>mA, 500Ω | 1.2 | 1.2 | |
| 15 | Maximum Average<br>Power Density,<br>(mW/cm²).500Ω | 1.74 | 2.88 | |
#### Table 1: Substantial Equivalence Comparison Table
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue and yellow. The text "TECHNOLOGY CO., LTD." is in a smaller font size than the company name. There are also three Chinese characters to the right of the company name.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue and yellow. The text "DONGDIXIN" is in large, bold, blue letters, with a yellow "X" in the middle. Below the company name is the text "TECHNOLOGY CO., LTD." in smaller, blue letters. To the right of the company name are three Chinese characters in blue.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for DONGDIXIN Technology Co., LTD. The logo features the company name in a bold, sans-serif font, with the "X" in "DIXIN" stylized in orange. To the right of the English name is the company name in Chinese characters.
### Table 2: Output Specification TENS mode
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is composed of the company name in a bold, sans-serif font, with the "X" in "DIXIN" stylized in orange. To the right of the English name is the company name in Chinese characters. The overall design is clean and modern.
#### Discussion:
The number of output modes, number of output channels, timer range, weight and dimension of the new device are different from the predicate devices, but the new devices are evaluated and passed the testing according to IEC60601-1 and IEC60601-2-10, this difference does not pose any new questions of safety.
The Max Output Voltage/Current, Pulse Duration, Maximum Phase Charge, Maximum Current Density and Maximum Average Power Density are similar with the predicate devices, but the new devices are evaluated and passed the testing according to IEC60601-2-10, this difference doesn't pose any new questions of safety and effectiveness.
#### 7. Performance Data:
The following performance data are provided in support of the substantial equivalence determination:
#### 7.1 Biocompatibility testing
Compared with unmodified device, the material of electrodes for the new product LT5011C is the same material of predicate device (K173462) that got the 510(k) clearance on 2017. The material of enclosure is ABS, and it is the same as the material of predicate device (K173462) that got the 510(k) clearance on 2017. These materials have the same product process and injection process.
#### 7.2 Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the LT5011C. The system complies with the IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC.
For FCC part 15 RADIO FREQUENCY DEVICES, Subpart C—Intentional Radiators.
#### 7.3 Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern. IEC 62304 was followed.
#### 7.4 Output waveform Testing
For each program, oscilloscope tracing diagrams describing the electrical output waveform was provided to verify the output specifications of the device according to IEC 60601-2-10.
#### 8. Conclusions
The LT5011C has the similar intended use and technological characteristics as the predicate device Wireless Pain Relieve Device LT5018C. Moreover, bench testing and safety report supplied in this submission demonstrates that the difference in the submitted models could
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue, except for a yellow line that crosses the "X" in "DIXIN". To the right of the English text is the company name in Chinese.
005_510 (k) Summary Version: 1.1
maintain the same safety and effectiveness as that of predicate device. In the other words, those engineering difference do not affect the intended use or alter the fundamental scientific technology of the device. Thus, LT5011C is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.