The Gemore OTC TENS, Model GM310PP/GM321PP/GM330PP, is intended for temporary relief of pain associated with sore and aching muscles in the low back and/or upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
Device Story
Gemore OTC TENS (Models GM310PP/GM320PP/GM321PP/GM330PP) is a dual-channel transcutaneous electrical nerve stimulator for pain relief. Device consists of a stimulus generator and electrodes applied to user's skin. It delivers electrical current to stimulate nerves; intensity is user-adjustable within fixed limits. Operation modes are pre-programmed with fixed pulse width, rate, and frequency; timer and amplitude are adjustable. Device is intended for OTC use. Healthcare providers or patients use the device to manage muscle pain; output affects clinical decision-making by providing non-pharmacological pain relief. Benefits include temporary relief of muscle soreness and aching.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 10993-5, and ISO 10993-10. Software verification performed per FDA guidance. No clinical data provided.
Technological Characteristics
Dual-channel TENS; ABS plastic housing; 9V (6F22) battery power; Type BF isolation. Output: Pulsed monophasic waveform, rectangular shape. Connectivity: Standalone. Sterilization: Not specified. Software: Microprocessor-controlled with automatic no-load trip and automatic shut-off. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for temporary relief of pain associated with sore and aching muscles in the low back and/or upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
OTC TENS For Low Back Pain Relief/ Model GM310PP/GM320PP/GM321PP/GM330PP (K060222)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community or interconnectedness. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 12, 2015
Gemore Technology Co., Ltd. Boden Lai Official Correspondent 11 Fl., No.29-5, Sec.2, Chung Cheng E. Rd., Tan Shui, New Taipei City, 251 TW
Re: K150681
Trade/Device Name: Gemore OTC TENS, Models GM310PP/GM320PP/GM321PP/GM330PP Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator Regulatory Class: Class II Product Code: NUH Dated: March 16, 2015 Received: March 17, 2015
Dear Boden Lai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications For Use
510(k) Number (if known): K150681
Device Name: Gemore OTC TENS, Model GM310PP/GM320PP/GM321PP/GM330PP
Indications For Use:
The Gemore OTC TENS, Model GM310PP/GM321PP/GM330PP, is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities
Prescription Use (Part 21 CFR 801 Subpart D)
OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K 150681 .
- Submitter's Identifications: 1.
Company Name : GEMORE TECHNOLOGY CO., LTD. Contact person: Boden S.P. Lai 11FL., NO. 29-5, Sec. 2, Chung Cheng E. RD., Tan Shui, New Taipei City, Taiwan Address : TEL No. : 886-2-8809-1799 Fax No. : 886-2-8809-1781 boden@gemore.com.tw E-mail address : Date of Summary Preparation: April 27, 2015.
- 2. Name of the Device:
Trade/Device Name: Gemore OTC TENS, Model: GM310PP/GM320PP/GM321PP/GM330PP. Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator. Regulatory Class: II Product Code: NUH
- Information of the 510(k) Cleared Device (Predicate Device): 3. K133723: OTC Patch/ Model WL-2406. K091757: OTC TENS For Arm & Leg Pain Relief/ Model WL-2407
- 4. Device Description:
The Gemore OTC TENS. model GM310PP/GM321PP/GM321PP/GM330PP is a dual channel transcutaneous electrical nerve stimulator used for pain relief by applying an electrical current to electrodes, which are attached on the user's skin. The output and waveform characteristic is fixed for every operation mode, only the intensity is adjustable within specified limit.
The Gemore OTC TENS, model GM310PP/GM320PP/GM321PP/GM330PP, consist mainly of two parts: the stimulus generator, electrode. The stimulus generates the output current specified as the innut of controller. The output port transmits the output current to the electrode, which is attached to the user's skin so as to transmit this stimulus current for pain relief.
The stimulation modes for Gemore OTC TENS is pre-program modes width, pulse rate, frequency. The timer and amplitude for each program are adjustable. This operation way is considered the simplification from the pre-program modes of comparison clear model, OTC TENS For Low Back Pain Relief/ Model GM310PP/GM320PP /GM321PP/ GM330PP (K060222). Every operation mode of Gemore OTC TENS, model GM310PP/GM320PP/ GM321PP/GM330PP has its individual stimulation operation cycle.
For the device included in this submission, we use the following of our 510(K) legally marketed electrodes: K062675, "Gemore Reusable Self Adhesive Electrode", Wire Series/ model no. FA5050, size 50x50mm, wire type.
With the combination of the main device parts, the device can be placed on the treatment locations as recommended in the user manual for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
{4}------------------------------------------------
#### 5. Intended Use:
The Gemore OTC TENS, Model GM310PP/GM321PP/GM330PP, is intended for temporary relief of pain associated with sore and aching muscles in the low back and/or upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
#### 6. Summary of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes IEC 60601-1, and IEC 60601-1, and IEC 60601-1-2 as well as ISO 10993-5 and ISO 10993-10 requirements. In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.
Summary: The Compliance to applicable voluntary standards as above mentioned indicates that the new device in this submission used the same standards as that of predicate device. Therefore; we consider that the compliance of standards included in our submission is adequate for the determination of substantial equivalence.
### 7. Comparison of Significant device features
| Features | 510(K) Cleared Models | | New Model |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | WL-2406 | WL-2407 | GM310PP/GM320PP/<br>GM321PP/GM330PP |
| 510(K) No. | K133723 | K091757 | Unknown |
| Prescription or OTC | OTC | OTC | OTC |
| Indication for use | temporary relief of pain<br>associated with sore and<br>aching muscles in the low<br>back as well as upper and<br>lower extremities (arm and/<br>or leg) due to strain from<br>exercise or normal House-<br>hold and work activities | temporary relief of pain<br>associated with sore and<br>aching muscles in the upper<br>And lower extremities (arm<br>and/or leg) due to strain from<br>exercise or normal House-<br>hold and work activities | temporary relief of pain<br>associated with sore and<br>aching muscles in the low<br>back as well as upper and<br>lower extremities (arm and/<br>or leg) due to strain from<br>exercise or normal House-<br>hold and work activities |
| FDA product code | NUH | NUH | NUH |
| Electrode Used | Self Adhesive Electrode<br>(13x7 cm)/K082065 | Belt Electrode and/or<br>Self Adhesive Electrode<br>(5 X 5 cm)/K082065 | Self Adhesive Electrode<br>(5x5 cm)/K062675 |
### 8. Significant output characteristics comparison table:
| Comparison feature | | 510(K) Cleared Model | New Model | |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------|-------------------------------------------|-----------------------------------------------|
| | | WL-2406(K133723) | WL-2407(K091757) | GM310PP/GM320PP/<br>GM321PP/GM330PP |
| Net charge | | 0 | 0 | 0 |
| Max. phase charge | | 13 uc | 20.8 uc | 20.8 uc |
| Max. current Density | | 0.549 mA/cm² | 3.2 mA/cm2 | 3.2 mA/cm2 |
| Max. Average | 500 입 | 50 mA | 80 mA | 80 mA |
| current | 2K S2 | 22.5 mA | 30 mA | 40 mA |
| (RMSA) | 10K ක | 7.5 mA | 10 mA | 10 mA |
| Max. Power Density | | 0.001219 Watts/ cm² | 0.00200 Watts/ cm² | 0.00399 Watts/ cm2 |
| Burst Mode | | Yes | Yes | Yes |
| | | Predicate Device | | New Device |
| Mode or Program Name | | WL-2406 | WL-2407 | GM310PP/GM320PP/<br>GM321PP/GM330PP |
| | 510(K) Number | K133723 | K091757 | Unknown |
| | Waveform (e.g., pulsed monophasic, biphasic) | Biphasic | Biphasic | Pulsed monophasic |
| | Shape (e.g., rectangular, spike, rectified<br>sinusoidal) | Rectangular | Rectangular | Rectangular |
| | Maximum Output Voltage (volts)<br>(+/-_20 %) | 25V @500Ω | 40V @500Ω | 40V @500Ω |
| | | 45V @2KΩ | 60V @2KΩ | 80V @2KΩ |
| | | 75V @10KΩ | 100V @10KΩ | 100V @10KΩ |
| | Maximum Output Current (mA)<br>(+/-_20 %) | 50mA @500Ω | 80mA @500Ω | 80mA @500Ω |
| | | 22.5mA @2KΩ | 30mA @2KΩ | 40mA @2KΩ |
| | | 7.5mA @10KΩ | 10mA @10KΩ | 10mA @10KΩ |
| | Duration of primary phase (usec) | 260 max | 260 max | 260 max |
| | Pulse Duration (usec) | 8333 max | 16666 max | 8333 max |
| | Frequency (Hz) [or Rate (pps)] | 120 max | 60 max | 120 max |
| For | Yes | Yes | Yes | Yes |
| multiphasic<br>waveforms<br>only: | Not applicable | Not applicable | Not applicable | Not applicable |
| | Power Source(s) | 1.5Vx2<br>(AAA Size) | 1.5Vx3<br>(AAA Size) | 9V x 1<br>(6F22 Size) |
| | - Method of Line current Isolation | Type BF | Type BF | Type BF |
| | - Patient Leakage Current | --- | --- | --- |
| | - Normal condition (uA) | Under 0.1 | Under 0.1 | Under 0.1 |
| | - single Fault condition (uA) | Under 0.5 | Under 0.5 | Under 0.5 |
| | Average DC current through electrodes when<br>device is on but no pulses are being applied (uA) | Not applicable | Not applicable | Not applicable |
| | Number of Output Modes | 8 | 8 | 8 |
| Number of Output<br>Channels: | Synchronous<br>Output Coil | Synchronous<br>Output Coil | Synchronous<br>Output Coil | Synchronous<br>Transformer |
| | Regulated Current or Regulated Voltage? | Voltage | Voltage | Current |
| | Software/Firmware/Microprocessor control? | Yes | Yes | Yes |
| | Automatic Overload Trip? | No | No | No |
| | Automatic No-Load Trip? | Yes | Yes | Yes |
| | Automatic Shut Off? | Yes | Yes | Yes |
| | User Override control? | No | No | No |
| | Indicator Display: | Yes | Yes | Yes |
| | | Yes | Yes | Yes |
| | | Yes | Yes | Yes |
| | Timer Range (Minutes) | 10-60 | 5-60 | 10-60 and Continuous |
| | Compliance with Voluntary Standards? | IEC 60601-2-10 | IEC 60601-2-10 | IEC 60601-2-10 |
| Compliance with 21 CFR 898? | | Yes | Yes | Yes |
| Weight (g) including battery | | 51.6 | 125.5 | 140 |
| Dimensions (mm.) [W x H x D] | | 68x60x17.5 | 90×50.8×12.7 | 108×61.5×25 |
| Housing Materials and construction | | ABS | ABS | ABS |
| Pulse per burst | | 9 | 9 | 9 |
| Burst per second | | 2 | 2 | 2 |
| Bust duration | | 260us | 260us | 260us |
| Duty Cycle | | Same for each<br>program | Same for each<br>program | Same for each<br>program |
| Method of achieving zero net charge for net<br>charge/pulse | | Biphasic<br>symmetric<br>wave for each<br>pulse | Biphasic symmetric<br>wave for each pulse | Biphasic<br>asymmetric<br>wave for each pulse |
{5}------------------------------------------------
## 9. Comparison of Unit Characteristics & Output Specification
{6}------------------------------------------------
### 10. Summary for the technology comparison.
Basically the Gemore OTC TENS, Model GM310PP/GM321PP/GM321PP/GM330PP has the same technological characteristics with the predicate device in the product design, material, energy source type, main program mode and the main output waveform...etc. There exists some difference in the device outlook and dimensions. Through the detailed calculation comparison of stimulation output energy for each operation mode(in particular the output current density and power density), we found the output level in each operation mode for Gemore OTC TENS, Model GM310PP/GM320PP/GM321PP/GM330PP and predicate device are very close and within the acceptable range as specified in FDA quidance. So we believe the difference in detailed output parameters does not affect the determination of substantial equivalence.
### 11. Conclusions
The Gemore OTC TENS, Model GM310PP/GM320PP/GM330PP have the same intended use and the similar technological characteristics as the cleared devices. Moreover, verification and validation tests contained in this submission demonstrate that the submitted models could maintain the same safety and effectiveness as that of cleared device.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.