To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
Device Story
Quick Relax Pro is a handheld, battery-powered (1.5V AA) transcutaneous electrical nerve stimulator (TENS) for OTC pain relief. It features a single-mode, single-function design with a rotating disk to manually adjust microcurrent intensity. The device delivers pulsed asymmetrical biphasic waveforms directly to the body via an integrated electrode. It is intended for patient self-use to provide temporary relief of minor muscle pain in the shoulder, waist, back, neck, and extremities. The device lacks software, microprocessors, or automated shut-off features. Users operate the device by applying the electrode to the affected area and adjusting the current level manually. The device provides a non-pharmacological option for managing muscle soreness.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-5, ISO 10993-10, and ISO 10993-12. Product safety testing performed per IEC 60601-1. Electromagnetic compatibility testing performed per IEC 60601-1-2. No clinical data provided.
Technological Characteristics
Handheld TENS device; 1.5V AA battery powered; flame-retardant ABS and stainless steel construction; pulsed asymmetrical biphasic waveform; 11 µSec pulse duration; 30 Hz frequency; Type BF line current isolation; no software/microprocessor; manual rotating disk intensity control.
Indications for Use
Indicated for temporary relief of minor muscle pain in shoulder, waist, back, neck, arms, and legs caused by exercise or household work. For over-the-counter use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
### K131956
### Section 005 – 510(k) Summary
### APR 2 4 2014
.
### Applicant or Sponsor Information
| Name: | Healthcare Innovations, LLC |
|-----------------------------|-----------------------------|
| Address: | 12889 Coco Plum Lane |
| | Naples, FL 34119-8530 |
| Contact: | Dr. John Ricci |
| Phone: | (239) 593 - 9570 |
| Fax: | (239) 593 - 9570 |
| E-mail: | jriccirx@earthlink.net |
| Establishment Registration: | TBD |
| Date of Submission: | 23 June, 2013 |
#### Submission Correspondent (Preparer's) Information
| Name: | Shepard G. Bentley, RAC<br>President, Bentley Biomedical Consulting, LLC<br>Consultant to Healthcare Innovations, LLC |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Address: | 28241 Crown Valley Parkway<br>Suite 510(k)<br>Laguna Niguel, CA 92677 |
| Device Information | |
| Device Common Name: | Transcutaneous electrical nerve stimulator for pain relief<br>intended for over the counter use |
| Device Trade Name: | Quick Relax Pro |
| Device Classification Name: | Stimulator, nerve, transcutaneous, over-the -counter |
(
{1}------------------------------------------------
| Device Class: | Class II | |
|------------------|----------|--------------------|
| Product Code: | NUH | |
| CFR Reference: | 882.5890 | |
| Predicate Device | K121757 | HealthmateForever® |
#### Design and Use of the Device
Hand held, 1.5 volt AA battery powered, single mode, single function device with ten "settings" that are controlled by means of rotating a marked disk that increases the microcurrent when rotated clockwise and decreases the microcurrent when rotated counterclockwise, directly through the device integrated electrode into the indicated region of the body. The purpose of the Quick Relax Pro is to provide a temporary relief of minor muscle pain within the indicated region of the body.
### Technological Characteristics Comparison
| Technical Characteristics | Quick Relax Pro | HealthmateForever® Pro-8AB |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | To be used for temporary<br>relief of pain associated with<br>sore and aching muscles in<br>the shoulder, waist, back,<br>neck, upper extremities (arm),<br>and lower extremities (leg)<br>due to strain from exercise or<br>normal household work<br>activities | To be used for temporary<br>relief of pain associated with<br>sore and aching muscles in<br>the shoulder, waist, back,<br>neck, upper extremities (arm),<br>and lower extremities (leg)<br>due to strain from exercise or<br>normal household work<br>activities |
| Battery Powered | 1.5V DC | 3.7V DC |
| Method of Line Current<br>Isolation | Type BF | Type BF |
| Number of Output Modes | 1 | 8 |
| Waveform | Asymetrical Biphasic | Asymetrical Biphasic |
| Software | No | Yes |
| Functions | Single | Multiple |
{2}------------------------------------------------
| Parameter | | Subject Device | Predicate Device |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|----------------------------------------|
| 510(k) Number | | K131956 | K121757 |
| Device Name and Model | | Quick Relax Pro | HealthmateForever |
| Manufacturer | | Healthcare<br>Innovations, LLC | Healthmate<br>International, LLC |
| Power Source(s) | | Battery, single<br>AA dry cell | Battery, single dry<br>cell |
| -Method of Line Current Isolation | | N/A | N/A |
| -Patient Leakage Current | | N/A | N/A |
| -Normal Condition (μC) | | N/A | N/A |
| -Single Fault Condition (µC) | | N/A | N/A |
| Average DC current through electrodes when device is<br>on but no pulses are being applied (µC) | | N/A (pulses<br>always applied<br>when on) | N/A (pulses always<br>applied when on) |
| Number of Output Modes | | 1 | 2 |
| Number of Output Channels | Synchronous or<br>Alternating?<br>Method of<br>Channel<br>Isolation | N/A | N/A |
| Regulated Current or Regulated<br>Voltage? | | N/A | N/A |
| Software/Firmware/Microprocessor Control? | | No | Yes |
| Automatic Overload Trip? | | No | No |
| Automatic No-Load Trip? | | No | No |
| Automatic Shut Off? | | No | No |
| User Override Control? | | No | Yes |
| Indicator Display: | On/Off Status?<br>Low Battery?<br>Voltage/Current<br>Level? | Yes<br>No<br>Yes | Yes<br>Yes<br>Yes |
| Timer Range (minutes) | | N/A | Yes |
| Compliance with Voluntary Standards? | | Yes | Yes |
| Compliance with 21 CFR 898? | | N/A | Presumed |
| Weight (lbs., oz.) | | 4.5 oz | 3.9 oz |
| Dimensions (in) [W x H x D] | | 8.75" x 2" x<br>1.75" | 3.75" x 2" x .5" |
| Housing Materials and Construction | | Flame ret. ABS,<br>stainless steel | Metal, LCD |
| Parameter | Subject Device | Predicate Device | |
| Mode or Program Name | Output | Pro-8AB | |
| Waveform (e.g., pulsed monophasic, biphasic) | Pulsed biphasic (asymmetrical unbalanced) | Pulsed biphasic (asymmetrical unbalanced) | |
| Shape (e.g., rectified, spike, rectified sinusoidal) | Rectified sinusoidal (with undershoot) | Rectangular with taper | |
| Maximum Output Voltage (volts) (+/-5%) | 58.4v @ 500Ω<br>220v @ 2KΩ<br>995v @ 10KΩ<br>applies measure prior to EnCadis' clarification - revised calculation is approx 300v @ 10KΩ | 70v @ 500Ω<br>90v @ 2KΩ<br>100v @10KΩ | |
| Maximum Output Current (mA) (+/- 5%) | 117 mA @ 500Ω<br>110 mA @ 2KΩ<br>100 mA @ 10KΩ | 140 mA @ 500Ω<br>45 mA @ 2KΩ<br>10 mA @ 10KΩ | |
| Duration of primary (depolarizing) phase (µSec) | N/A | unknown | |
| Pulse Duration (µSec) | 11 µSec | 10 mSec | |
| Frequency (Hz) | 30 Hz (max amplitude) | 100 Hz | |
| For multiphasic waveforms only:<br>Symmetrical phases?<br>Phase Duration | No<br><br>Positive Phase:<br>11 µSec<br>Negative Phase:<br>2 µSec (@ min amplitude, 500Ω load)<br>25 µSec (@ max amplitude, 10KΩ load) | No<br><br>Various, all are longer than Quick Relax Pro | |
| Net Charge (microCoulombs (µC) per pulse) | 0.78 µC * | Various, higher | |
| Maximum Phase Charge (µC) | 0.82 µC * | 15.66 µC | |
| Maximum Current Density (mA/cm2, r.m.s.) | 3.37 mA/cm2 r.m.s. | 7.2 mA/cm2 @500Ω | |
| Parameter --------------- | | Subject Device | Predicate Device |
| Maximum Average Current (average absolute value)<br>mA | | 0.220 mA | Not measured |
| Maximum Average Power Density (W/cm2) | | 803 µW/cm2 | Not measured |
| Burst Mode | (a) Pulses per<br>burst | N/A | Various |
| | (b) Bursts per<br>second | N/A | Various |
| | (c) Burst<br>duration | N/A | Various |
| | (d) Duty Cycle:<br>Line b x Line c | N/A | Various |
| ON Time (Sec) | | N/A | 1,200 default |
| OFF Time (Sec) | | N/A | N/A |
| Additional Features | | N/A | Unknown |
{3}------------------------------------------------
:
{4}------------------------------------------------
#### Summary of Technological Characteristics Comparison
The foregoing comparison of the technological characteristics of the subject and predicate devices confirms that the devices are substantially equivalent with respect to intended uses, method of powering, physical descriptions, method of isolation and waveforms.
The differences regarding battery power, material descriptions, number of output modes, presence of software and overall numbers of functions confer an added degree of safety to the subject device.
Therefore, the subject device is at least as safe if not safer than the predicate device, and substantially equivalent with regard to the relevant comparators of the devices.
#### Summary of Non-Clinical Testing
The Quick Relax Pro has undergone biocompatibility testing of its tissue-contacting electrode in compliance with ISO 10993-5, ISO 10993-10 and ISO 10993-12. In addition, the Quick Relax Pro has undergone Product Safety testing in compliance to IEC 60601-1 and Electromagnetic Compatibility testing in compliance to IEC 60601-1-2. The results of each of the foregoing sets of test requirements were satisfactory, and serve to demonstrate the safety and effectiveness of the Quick Relax Pro.
The product safety, electromagnetic compatibility and biocompatibility test results provide assurance that the subject device performs as well or better than the predicate device.
{5}------------------------------------------------
The introduction of the Quick Relax Pro into interstate commerce does not raise any new issues of safety or effectiveness.
and the comments of the count
100 - 100 - 100 -
:
END OF 510(K) SUMMARY
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human services.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2014
Healthcare Innovations, LLC c/o Shepard G. Bentley, RAC Bentley Biomedical Consulting, LLC 28241 Crown Valley Parkway, Suite 510(k) Laguna Niguel, CA 92677
Re: K131956
Trade/Device Name: Quick Relax Pro Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: March 4, 2014 Received: March 25, 2014
Dear Mr. Bentley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{7}------------------------------------------------
#### Page 2 - Mr. Shepard Bentley
comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDcvices/Safety/ReportalProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Peña AS
Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K131956
#### Device Name
Quick Relax Pro
#### Indications for Use (Describe)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
Type of Use (Select one or both, as applicable)
□ Prescription Use (Part 21 CFR 801 Subpart D)
[X] Over-The-Counter Use (21 CFR 801 Subpart C)
...
:
#### PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
.
Carlos Pena -S
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page.
{9}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995,
·
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration
-
- ﺎ ﺗﺄ Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.