K214100 · Premier North America, Inc. · NFO · Feb 24, 2022 · Neurology
Device Facts
Record ID
K214100
Device Name
Luminice
Applicant
Premier North America, Inc.
Product Code
NFO · Neurology
Decision Date
Feb 24, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Luminice is intended for body skin stimulation and is indicated for over-the-counter cosmetic use.
Device Story
Luminice is a handheld, battery-powered transcutaneous electrical nerve stimulator (TENS) for cosmetic body skin toning. It uses four fixed stainless steel electrodes to deliver low-level microcurrent electrical impulses to the skin. The device requires the use of a conductive gel. Operation involves pressing an on/off button; the device provides feedback via vibration, red light illumination, and ascending tonal beeps. It features adjustable intensity levels indicated by 1-3 LEDs. The device is intended for home use by the consumer. The microcurrent output is a constant monophasic square wave consisting of alternating bursts of positive and negative pulses. The device includes automatic shut-off and no-load trip safety features. By stimulating the skin, the device is intended to provide cosmetic benefits. It is designed for use on areas of the body other than the face.
Clinical Evidence
Bench testing only. The device underwent electrical safety testing (ANSI/AAMI ES60601-1, IEC 60601-2-10), EMC testing (IEC 60601-1-2), battery safety testing (IEC 62133), and biocompatibility testing (ISO 10993-5, ISO 10993-10). Software validation was performed per FDA guidance.
Technological Characteristics
Handheld device with ABS resin housing and four fixed stainless steel electrodes. Powered by internal rechargeable lithium-ion battery. Delivers microcurrent as a constant monophasic square wave (bursts of 10 positive/10 negative pulses). Frequency: 8.3 Hz. Connectivity: None. Sterilization: Not applicable. Software: Microprocessor-controlled with automatic shut-off and no-load trip. Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 62133, IEC 60601-1-11, ISO 14971.
Indications for Use
Indicated for over-the-counter cosmetic body skin stimulation in adults.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 24, 2022
Premier North America Inc. % Doris Dong Manager Shanghai CV Technology Co., Ltd. Room 903, No.19 Dongbao Road, Songjiang Area Shanghai, Shanghai 201613 China
Re: K214100
Trade/Device Name: Luminice Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: December 22, 2021 Received: December 29, 2021
Dear Doris Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K214100
Device Name Luminice
Indications for Use (Describe)
Luminice is intended for body skin stimulation and is indicated for over-the-counter cosmetic use.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary [As required by 21 CFR 807.92]
#### 1. Submission Information:
| 510(k) Number: | K214100 |
|----------------|----------------------------------------------------------------------------|
| Date prepared: | February 12, 2022 |
| Owner: | Premier North America Inc. |
| | 3301 SW 42ND ST., FORT LAUDERDALE, FL 33312-6828, USA |
| | Tel: 404-4928133-11 |
| Contact: | Doris Dong |
| | [Consultant, from Shanghai CV Technology Co., Ltd.] |
| | Add: Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 China |
| | E-mail: doris.d@ceve.org.cn |
| | Tel: 86 21-31261348 / Fax: 86 21-57712250 |
| 2. Device Description: | |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | Luminice |
| Model: | LUMI536 |
| Regulation Name: | Transcutaneous electrical nerve stimulator for pain relief |
| Device Classification Name: | Stimulator, Transcutaneous Electrical, Aesthetic Purposes |
| Regulation Number: | 21 CFR 882.5890 |
| Product Code: | NFO |
| Device Class: | II |
| Review Panel: | Neurology |
| Device Description: | Luminice is intended for body skin stimulation and is indicated for<br>over-the-counter cosmetic use. Four (4) stainless steel electrodes, fixed on the<br>Luminice device main body, deliver low level electrical impulses (microcurrent)<br>to targeted locations on the body.<br>The Luminice should be used with 510(k) clearance electroconductive media<br>gel, and it's compatible with NuFACE Gel Primer cleared under K161654. |
| Indications for use: | Luminice is intended for body skin stimulation and is indicated<br>for<br>over-the-counter cosmetic use. |
#### 3. Substantial Equivalence to Predicate device:
The proposed Luminice device has the same, or similar, technological characteristics as the NuBODY Skin Toning Device predicate device. The differences between our proposed device and the cleared predicate device listed below do not raise any safety and effectiveness issues.
| Device Descriptions | Luminice<br>(New Device) | NuBODY Skin Toning Device<br>(Predicate) | Remark |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| 510(k) Number | K214100 | K171588 | -- |
| Regulation number | 21CFR 882.5890 | 21CFR 882.5890 | -- |
| Regulation Name | Transcutaneous electrical nerve<br>stimulator for pain relief | Transcutaneous electrical nerve<br>stimulator for pain relief | Same |
| Regulatory Class | Class II | Class II | Same |
| Classification Name | Stimulator, Transcutaneous | Stimulator, Transcutaneous | Same |
| | Electrical, Aesthetic Purposes | Electrical, Aesthetic Purposes | |
| Product Code | NFO | NFO | Same |
| Regulation<br>Specialty | Medical Neurology | Neurology | Same |
| Intended use | Luminice is intended for body skin stimulation and is indicated for over-the-counter cosmetic use. | NuBODY Skin Toning Device is intended for body skin stimulation and is indicated for over-the-counter cosmetic use. | Same |
| Indications for use | Over-the-Counter cosmetic use | Over-the-Counter cosmetic use | Same |
| Anatomic Sites | Areas of the body other than the face | Areas of the body other than the face | Same |
| Technological<br>Characteristics | The Luminice is a body skin toning device. Its outer case is injection molded ABS resin. The output contacts consist of stainless steel. The Luminice is powered by a rechargeable lithium ion battery. Luminice produces microcurrent that is discharged through four fixed electrodes. To turn the Luminice on, the on/off button is pressed. Ascending tonal beeps and vibration indicate the Luminice device is on. One to three LED lights illuminate indicating the output intensity level and the unit is ready for use. The four electrodes gently glide over the skin to deliver low-level electrical impulses to targeted locations on the body. The Luminice device electrodes are designed for optimal contact with body skin. The Luminice device delivers microcurrent as a constant monophasic square wave comprised of a burst of (10) positive pulses followed by a burst of (10) negative pulses. The microcurrent output continuously alternates between the positive and negative spherical electrodes, and allows the user to adjust the output for a personalized comfort level. The Luminice device | The NuBODY Skin Toning Device is a body skin toning device. Its outer case is injection molded thermoplastic resin. The output contacts consist of chrome-plated spherical electrodes. The NuBODY device is powered by a rechargeable lithium ion battery. NuBODY device produces microcurrent that is discharged through four fixed, smooth spherical electrodes. To turn the NuBODY device on, the on/off button is pressed. Ascending tonal beeps indicate the NuBODY device is on. One to three LED lights illuminate indicating the output intensity level and the unit is ready for use. The four spheres gently glide over the skin to deliver low-level electrical impulses to targeted locations on the body. The NuBODY device spheres are designed for optimal contact with body skin. The NuBODY device delivers microcurrent as a constant monophasic square wave comprised of a burst of (10) positive pulses followed by a burst of (10) negative pulses. The microcurrent output continuously alternates between the positive and negative spherical | Difference<br>Note 1 |
Table 1- General Comparison Table
{4}------------------------------------------------
{5}------------------------------------------------
| requires the use of a conductive<br>gel. To promote proper use and<br>provide feedback to the user,<br>when the device is is contact with<br>the skin the Luminice device will<br>vibrates and turn on red light. | electrodes, and allows the user to<br>adjust the output for a<br>personalized comfort level. The<br>NuBODY device requires the use<br>of a conductive gel. To promote<br>proper use and provide feedback<br>to the user, the NuBODY device<br>beeps to cue the user to relocate<br>the NuBODY device<br>approximately every 5 seconds. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Basic Unit Characteristic | Luminice<br>(New Device) | NuBODY Skin Toning Device<br>(Predicate) | Remark |
|-----------|----------------------------------------------------|-------------------------------------------|-------------------------------------------|----------------------|
| | 510(k) Number | K214100 | K171588 | -- |
| | Device Name, Model | Luminice (LUMI536) | NuBODY Skin Toning Device | -- |
| | Owner | Premier North America Inc | Carol Cole Company (dba NuFACE) | -- |
| | Power Source(s) | Internal rechargeable Lithium ion battery | Internal rechargeable Lithium ion battery | Same |
| | Method of Line Current Isolation | Type BF | Type BF | Same |
| | Patient Leakage Current<br>- Normal Condition (μΑ) | N/A - Battery operated | N/A - Battery operated | Same |
| | - Single Fault Condition (μΑ) | N/A - Battery operated | N/A - Battery operated | Same |
| | Number of Output channels: | 1 | 1 | Same |
| | Synchronous or Alternating? | N/A - 1 Output channel | N/A - 1 Output channel | Same |
| | Method of Channel Isolation | N/A - 1 Output channel | N/A - 1 Output channel | Same |
| | Regulated Current or Regulated Voltage? | Both | Both | Same |
| | Software/Firmware/Microprocessor Control? | Yes | Yes | Same |
| | Automatic Overload Trip? | Not required due to circuit design | Not required due to circuit design | Same |
| | Automatic No-Load Trip? | Yes | Yes | Same |
| | Automatic Shut Off? | Yes | Yes | Same |
| | User Override Control? | Yes | Yes | Same |
| Indicator | On/Off Status? | Yes | Yes | Same |
| Display | Low Battery? | Yes | Yes | Same |
| | Voltage/Current Level? | Yes | Yes | Same |
| | Timer Range (minutes) | Yes (1min) | Yes (5min) | Difference<br>Note 2 |
# Table 2- Basic Unit Characteristic Comparison
{6}------------------------------------------------
| Compliance with Voluntary<br>Standards? | Yes.<br>AAMI/ANSI ES 60601-1,<br>IEC 60601-1-2, IEC<br>60601-2-10, IEC 62133, IEC<br>60601-1-11, ISO 14971 | Yes.<br>IEC 60601-1, IEC 60601-1-2<br>IEC 60529, IEC 60601-2-10<br>ISO 14971, IEC 60601-1-6<br>IEC 62366 | Difference<br>Note 3 |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------|
| Compliance with 21 CFR<br>898? | Yes | Yes | Same |
| Weight (grams) | Approximately 16.5 oz.<br>without power adapter | Approximately 10-14 oz. without<br>power adapter | Difference<br>Note 2 |
| Dimensions (mm) [W x H x<br>D] | Approximately 3.5" x 6.9" x<br>5.9" | Approximately 2.75" x<br>6.5" x 6.0" | |
| Housing Materials &<br>Construction | ABS and stainless steel | Thermoplastic | |
# Table 3- Output Specification Comparison Table
| Output Specifications | Luminice<br>(New Device) | NuBODY Skin Toning Device<br>(Predicate) | Remark |
|----------------------------------------------|----------------------------------------------------------------------------------|---------------------------------------------------------------|----------------------|
| 510(k) Number | K214100 | K171588 | -- |
| Waveform | Monophasic waveform that is<br>delivered in a burst of pulses | Monophasic waveform that is<br>delivered in a burst of pulses | Same |
| Shape | Voltage Modulated Square | Voltage Modulated Square | Same |
| Maximum Output Voltage<br>(volts) | 28VDC±10% @ open circle<br>410mV±10% @ 500Ω<br>1.7V±10% @ 2KΩ<br>8.1V±10% @ 10KΩ | 28 VDC | Same |
| Maximum Output Current<br>(specify units) | 820μΑ @500Ω | 900μΑ @ 500Ω | Difference<br>Note 4 |
| Output Current when not<br>stimulating | <1μΑ | <1μΑ | Same |
| Pulse width (usec) | 60ms | 60ms | Same |
| Max. pulse frequency (Hz)<br>[or Rate (pps)] | 8.3 Hz±10% | 8.3 Hz±10% | Same |
| For interferential modes, only | | | |
| a. Beat Frequency (Hz) | No Beat Frequency | No Beat Frequency | Same |
| For multiphasic waveforms, only | | | |
| a. Symmetrical phases | Not Multiphasic | Not Multiphasic | Same |
| b. Phase Duration (include<br>units) | Not Multiphasic | Not Multiphasic | Same |
| c. (state range, if<br>applicable) | Not Multiphasic | Not Multiphasic | Same |
| d. (both phases, if<br>asymmetrical) | Not Multiphasic | Not Multiphasic | Same |
| Net Charge (µC per pulse) | 49.2μC | 54μC | Difference<br>Note 4 |
| Maximum Phase Charge<br>(mC/Burst) | 0.984 | 1.08 | Note 4 |
{7}------------------------------------------------
| | Maximum Current Density,<br>(mA/cm², r.m.s.) | 0.19mA/cm² | 0.468mA/cm² | |
|---------------|----------------------------------------------|------------|-------------|---------|
| | Maximum Average Power<br>Density, (mW/cm²) | 0.08mW/cm² | 4.18mW/cm² | |
| Burst<br>Mode | (a) Pulses per<br>burst | 20 | 20 | Same |
| | (b) Bursts per<br>second | 8.3 | 8.3 | Same |
| | (c) Burst duration | 2.4 s | 2.4s | Same |
| | (d) Duty Cycle:<br>Line (b) x Line | 19.92s | 20.2s | Similar |
| ON Time | | 60ms | 60ms | Same |
| OFF Time | | 60ms | 60ms | Same |
Differences between New device and Predicate Device:
### Note 1:
The proposed device and predicate device have many similarities in technological characteristics except for electrodes, materials and prompt functions. Both of them have passed the biocompatibility tests. For more stability and better contact, the proposed device uses fixed, oval electrodes without sharp horns, and provides more prompt functions than the predicate device (vibration and red contact light prompt). The proposed device has passed the IEC 60601-1, 60601-1-2 tests. Therefore, these differences will not raise any issues of safety or effectiveness.
## Note 2:
The automatic shutdown time limit is slightly different between the proposed device (1min) and the predicate device (5min). Because this design doesn't affect the main functions of device, and the proposed device has passed ES 60601-1. IEC 60601-1-2. IEC 60601-1-11 tests. Therefore, this difference will not raise any issues of safety or effectiveness.
The proposed device is different from the predicate device in housing material, weight, dimensions and appearance. Both of them have passed the biocompatibility tests, IEC 60601-1-2 tests. Therefore, these differences will not raise any issues of safety or effectiveness.
## Note 3:
The tests performed by our proposed device are slightly different from those of the predicate device. IEC 60529 is the standard for device waterproof level testing, and the waterproof level of our device has been evaluated in the AAMI/ANSI ES 60601-1 test. The predicate device performed usability test according to IEC 60601-1-6 and IEC 62366 standards. The proposed device referred the user manual of the predicate device K171588, and we attached the Labeling of the predicate device K171588 in our submission. Therefore, these differences will not raise any new issues of safety or effectiveness.
### Note 4:
The maximum output current, net charge, maximum phase charge of the proposed device are similar to those of the predicate device. And the maximum current density, maximum average power density of the proposed device are lower than those of the predicate device due to they are calculated by different electrode areas. Both of them meet IEC 60601-2-10 tests. Therefore, these differences don't raise any new safety and effectiveness issues.
{8}------------------------------------------------
## 4. Non-clinical Testing:
The conclusions drawn from the non-clinical testing below demonstrate that the Luminice is substantially equivalent to the predicate devices K171588. The Luminice has been tested and conforms to international consensus standards:
Electrical safety:
- ANSI AAMI ES60601-1: 2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD);
## EMC:
- IEC 60601-1-2:2014. Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And ● Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests;
Additional safety testing:
- IEC60601-1-11 Edition 2.0 2015 Medical Electrical Equipment -- Part 1-11: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment;
- IEC 60601-2-10 Edition 2.1 2016-04: Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.
- IEC 62133 Edition 2.0 2012-12 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications [Including: Corrigendum 1 (2013)];
Biocompatibility testing:
- ISO 10993-5:2009/(R) 2014, Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro ● Cytotoxicity. (Biocompatibility)
- ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices Part 10: Tests For ● Irritation And Skin Sensitization. (Biocompatibility)
### Software Validation Testing:
The Luminice's software was tested and validated in accordance with FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices."
### 5. Conclusions
Based upon comparison to the predicate devices, the Luminice has the same intended uses, with similar technological characteristics as predicate device Luminice is substantial Equivalent to the predicate device K171588.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.