PureLift is intended for facial stimulation and indicated for over-the-counter cosmetic use.
Device Story
Hand-held facial stimulation device; applies low-level electrical impulses to strategic facial locations via two fixed, chrome-plated spherical probes. User-operated; powered by 3.7V rechargeable battery. Device features LCD screen, power button, and intensity adjustment (levels 1-10). Used with electroconductive media. Probes glide over skin to deliver impulses. Intended for OTC cosmetic use. Benefits include facial stimulation. Device includes automatic shut-off and timer (10 minutes).
Clinical Evidence
No clinical data. Bench testing only. Included oscilloscope tracings, vibration, temperature, mechanical strength, drop, ball, acoustic, and accessible parts testing. Biocompatibility testing (cytotoxicity, sensitization, irritation) per ISO 10993-5 and -10. Electrical safety and EMC testing per IEC 60601-1, 60601-1-2, 60601-1-11, and 60601-2-10.
Indicated for over-the-counter cosmetic facial stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
August 28, 2019
Xtreem Pulse. LLC % Rhonda Alexander, Ph.D. Sr. Consultant, Regulatory Strategy IUVO Consulting, LLC PO Box 56436 Virginia Beach, Virginia 23456
Re: K190269
Trade/Device Name: PureLift Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: August 26, 2019 Received: August 27, 2019
Dear Dr. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vivek Pinto, Ph.D. Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190269
Device Name PureLift
Indications for Use (Describe)
PureLift is intended for facial stimulation and indicated for over-the-counter cosmetic use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### l. SUBMITTER
XTREEM PULSE LLC 353 W. 29 ST. SUITE 3 NEW YORK, NY 10001 UNITED STATES
| Contact Person: | Rhonda Alexander, DSL |
|-----------------|-------------------------------------|
| | Sr. Consultant, Regulatory Strategy |
| | IUVO Consulting, LLC |
| | (757) 582-4337 |
| | ralexander@iuvoconsulting.com |
Date Prepared: 27 August 2019
#### II. DEVICE
| Name of Device: | PureLift |
|----------------------|-----------------------------------------------------------|
| Classification Name: | Stimulator, Transcutaneous Electrical, Aesthetic Purposes |
| Regulatory Class: | Class II |
| Product Code: | NFO |
| Regulation Number: | 21 CFR 882.5890 |
#### III. PREDICATE DEVICE
NuFace Nodel NU-4003 Manufacturer: Carol Cole Company K Number: K072260
Reference Device: Rejuvenique Model RJV10 K Number: K011935 Manufacturer: Salton, Inc.
## IV. DEVICE DESCRIPTION
The PureLift is a hand-held device intended to apply electrical impulses to strategic locations on the face. The PureLift probes are designed for optimal contact with the face. The device continually alternates between the positive and negative probes and allows the user to adjust the settings for personalized comfort level by pressing the up/down button. The intensity starts at (1) and continues to (10).
The device measures 13.4″ L x 4.8″ W x 4.3″ D. Its outer case is injection molded of thermoplastic resin and the probes consist of chrome-plated spheres. The device, powered by a 3.7-volt battery, produces low-level current that is transmitted through the two fixed, smooth spherical probes. To turn the device
{4}------------------------------------------------
on, the power button is pushed. Then LCD screen will be displayed, indicating the unit is ready for use. Users then follow the instructions for use. The two probes gently glide over the skin to deliver low-level electrical impulses to strategic locations on the face.
PureLift is intended to be used with a legally marketed electroconductive media.
The PureLift unit contains a power supply and rechargeable battery. The enclosure is made of medical grade biocompatible plastics and the output contacts (Probes) consist of chrome-plated spheres.
## V. INDICATIONS FOR USE
PureLift is intended for facial stimulation and indicated for over-the-counter cosmetic use.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Elements of<br>Comparison | PureLift | Predicate Device<br>NuFace<br>(K072260) | Reference Device<br>Rejuvenique<br>(K011935) |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Device Name, Model | PureLift | NuFace® | Rejuvenique® |
| Indications for Use | PureLift is intended for<br>facial stimulation and<br>indicated for over-the-<br>counter cosmetic use. | The NuFace® Facial<br>Toning Device is<br>intended for facial<br>stimulation and is<br>indicated for over-the-<br>counter cosmetic use. | The Rejuvenique System<br>is indicated for cosmetic<br>use. |
| Technological<br>Characteristics | The PureLift is a<br>handheld device<br>intended for facial<br>stimulation. Its outer<br>case is injection molded<br>of thermoplastic resin<br>and the two probes<br>consist of chrome-plated<br>spheres. The device,<br>powered by a 3.7V<br>battery, produces<br>electrical impulses that<br>is transmitted through<br>the two fixed, smooth<br>spherical probes.<br>To turn the device on,<br>the push the on/off<br>button. An LCD screen<br>will be displayed,<br>indicating the unit is<br>ready for use. Users<br>then follow the<br>instructions for use. The<br>two probes gently glide | NuFace® is a Facial<br>Toning Device intended<br>for facial stimulation.<br>The device measures 7"<br>L x 2.5" W x 1" D. Its<br>outer case is injection<br>molded of thermoplastic<br>resin, ABS UL 94 HB, and<br>the output contacts<br>(probes) consist of<br>chrome-plated spheres.<br>The device, powered by<br>a 9-volt battery,<br>produces microcurrent<br>that is Discharged<br>through the two fixed,<br>smooth spherical<br>probes. To turn the<br>device on, the<br>thumbwheel is pushed<br>upwards. A Green LED<br>light will Then<br>illuminate, indicating the<br>unit is ready for use. | |
| Elements of<br>Comparison | PureLift | Predicate Device<br>NuFace<br>(K072260) | Reference Device<br>Rejuvenique<br>(K011935) |
| | over the skin to deliver<br>low-level electrical<br>impulses to strategic<br>locations on the face. | Users then follow the<br>instructions for use. The<br>two probes gently glide<br>over the skin to deliver<br>low-level electrical<br>impulses to strategic<br>locations on the face. | |
| Power Source | One 3.7 V Battery | One 9V Battery | Single 9V Battery |
| Number of output<br>modules | 1 | 1 | 1 |
| Number of output<br>channels | 1 output channel | 1 output channel | 1 output channel |
| Regulated current or<br>regulated Voltage? | Regulated current | Both | Regulated Voltage |
| Software/Firmware/Microprocessor Control? | Yes | Yes | Yes |
| Automatic Shut off? | Yes | No | Yes |
| Patient override control? | No | No | No |
| Indicator Display | Yes | Yes | Yes |
| Timer range | 10 minutes only | No Timer | Fixed 16 minutes |
| Type of protection | Type BF | Type BF | Type BF |
| On/off status | Yes | Yes | Yes |
| Standards Compliance | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10 | IEC 60601-1 |
| Biocompatibility | ISO 10993-5,-10 | ISO 10993-5,-10 | ISO 10993-5,-10 |
| Waveform | Pulsed Monophasic,<br>alternating polarity | Pulsed Monophasic | pulsed biphasic |
| Shape | Rectangular Pulses | Rectangular Pulses | Rectangular (+phase,<br>Spike(-phase) |
| Maximum output<br>voltage | 15Vpp (@500Ω)<br>23Vpp (@2kΩ)<br>34Vpp(@10kΩ) | 158 Mv @ 500Ω<br>780 Mv @ 2ΚΩ<br>2.6 V @ 10ΚΩ | 18.8V @ 500Ω<br>24.8V @ 2kΩ<br>28.0V @ 10kΩ |
| Maximum output<br>current | 7mA (@500Ω)<br>3.2mA(@2kΩ)<br>0.9mA(@10kΩ) | 0.223mA @ 500Ω | 37.6mA @ 500Ω<br>12.4mA @ 2ΚΩ<br>2.8mA @ 10ΚΩ |
| Output tolerance | +/- 1mA | +/- 10% | +/- 10% |
| Pulse Width | 4μs | 112ms | 300μs |
| Frequency (Hz) | 1.37kHz~1.73kHz | 8.39Hz | 8Hz |
| Symmetrical phases | Not multiphasic | Not Multiphasic | No |
| Phase duration | 4μs | Not determined | 300µs (+phase),<br>124.7ms (-phase,<br>exponential) |
| Elements of<br>Comparison | PureLift | Predicate Device<br>NuFace<br>(K072260) | Reference Device<br>Rejuvenique<br>(K011935) |
| Net Charge (μC per<br>pulsetrain) | 0μC per pulse train | 0 µC per pulsetrain | 0µC per pulsetrain |
| Maximum Phase Charge<br>(μC) | 4.52μC @ 500Ω | 18.13 µC @ 500Ω | 11.3μC @500Ω |
| Maximum current<br>Density (mA/cm²) | 6.8mA/cm² @ 500Ω | 0.341mA/cm² @ 500Ω | 46.4mA/cm² |
| Maximum Power<br>Density (μW/cm²) | 23800μW/cm² @ 500Ω | 3.02μW/cm² @ 500Ω | 2310.0μW/cm² |
| Pulses per burst | 30 | 21 | 160 pulses |
| Bursts per second | 2740 ~ 3460 | 9.1 | 1/240<br>(per electrode group) |
| Burst duration | 230μs | 2.3 seconds | 20 seconds |
| Duty cycle | 0.63 ~ 0.80 | 20.9 | 1/12 |
| ON Time (seconds) | Constant | Constant | 20 seconds/electrode<br>group |
| OFF Time (Seconds) | None | None | None |
{5}------------------------------------------------
{6}------------------------------------------------
#### VII. PERFORMANCE DATA
Oscilloscope tracings, Vibration, Temperature, Push, Mold Stress, Markings, Mechanical Strength, Drop, Ball, Acoustic, and Accessible Parts tests were also conducted. The device passed or met the requirements of all testing.
#### Biocompatibility testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- . Irritation
The subject device is considered tissue contacting for a duration of less than 24 hours. The device passed each test, and was found to not be cytotoxic, sensitizing, or irritating.
#### Electrical safety and electromagnetic compatibility (EMC)
Electricalsafety and EMC testing were conducted on the subject device. It was found to comply with the IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC.
{7}------------------------------------------------
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
#### VIII. CONCLUSIONS
The subject device is identical to the predicate in terms of intended use. The technical differences between subject and predicate were addressed using performance data. Thus, it is concluded that the PureLift is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.