The SnugKap is intended for medical purposes for use on infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional Plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads by applying mild pressure to prominent regions of the infant's cranium in order to improve cranial symmetry and/or shape. The device is also indicated for adjunctive use for infants from 3 to 18 months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic- and scaphocephalic-shaped heads.
Device Story
SnugKap is a custom-fitted cranial orthosis for infants; treats nonsynostotic positional plagiocephaly and post-surgical synostosis. Input: 3D scan file of patient's head. Process: scan file digitally modified by orthotist to address anatomical abnormalities; device additively manufactured (3D printed) based on case plan. Output: custom cranial orthosis. Used in clinical settings; operated by orthotists. Principle of operation: static equilibrium; applies mild pressure to prominent cranial regions to improve symmetry/shape. Worn 23 hours/day after 3-day ease-in period. Benefits: improves cranial symmetry and shape in infants.
Clinical Evidence
Bench testing only. Studies included worst-case determination, strength testing (band), durability (open-close test), pressure testing, biocompatibility (sensitization, cytotoxicity, irritation), software validation, and 3D printing validation. No clinical data presented.
Indicated for infants 3-18 months old with moderate to severe nonsynostotic positional plagiocephaly (plagiocephalic, brachycephalic, or scaphocephalic shapes) or post-surgical synostosis with residual cranial deformities. Contraindications not specified.
Regulatory Classification
Identification
A cranial orthosis is a device that is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Special Controls
*Classification.* Class II (special controls) (prescription use in accordance with § 801.109 of this chapter, biocompatibility testing, and labeling (contraindications, warnings, precautions, adverse events, instructions for physicians and parents)).
In addition to the general controls of the act, the Dynamic Orthotic Cranioplasty - DOC™ Band is subject to the following special controls in order to provide reasonable assurance of the safety and effectiveness: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include (a) contraindications for the use of the device on infants with synostosis or with hydrocephalus; (b) warnings indicating the need: (i) to evaluate head circumference measurements and neurological status at intervals appropriate to the infant’s age and rate of head growth, and to describe steps that should be taken in order to reduce the potential for restriction of cranial growth and possible impairment of brain growth and development; (ii) to evaluate the skin at frequent intervals, e.g., every three to four hours, and to describe steps that should be taken if skin irritation or breakdown occurs; (c) precautions indicating the need: (i) to additionally treat torticollis, if the positional plagiocephaly is associated with torticollis; (ii) to evaluate device fit and to describe the steps that should be taken in order to reduce the potential for restriction of cranial growth, possible impairment of brain growth and development and skin irritation and/or breakdown; (iii) to evaluate the structural integrity of the device and to describe the steps that should be taken to reduce the potential for the device to slip out of place and cause asphyxiation or trauma to the eyes or skin; (d) adverse events, i.e., skin irritation and breakdown that have occurred with the use of this device; (e) clinician’s instructions for casting the infant, for fitting the device, and for care and use of the device; and (f) parents’ instructions for care and use of the device. (3) The materials must be assessed for biocompatibility with testing appropriate for long term direct skin contact.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2021
HeadStart, Ltd. % Dave McGurl Director, Regulatory Affairs MCRA, LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001
Re: K193383
Trade/Device Name: SnugKap Regulation Number: 21 CFR 882.5970 Regulation Name: Cranial Orthosis Regulatory Class: Class II Product Code: MVA Dated: May 7, 2021 Received: May 7, 2021
Dear Dave McGurl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193383
Device Name SnugKap
#### Indications for Use (Describe)
The SnugKap is intended for medical purposes for use on infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional Plagiocephaly, infants with plagiocephalic-, brachycephalic-, and scaphocephalicshaped heads by applying mild pressure to prominent regions of the infant's cranium in order to improve cranial symmetry and/or shape. The device is also indicated for adjunctive use for infants from 3 to 18 months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic- and scaphocephalic-shaped heads.
Type of Use (Select one or both, as applicable)
| <div> <input checked="true" type="checkbox"/> <span>Transportation Under CA STC 111.5(a) or Charter</span> </div> |
|-------------------------------------------------------------------------------------------------------------------|
| <div> <input type="checkbox"/> <span>On-Demand Service Under CA STC 111.5(b)</span> </div> |
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K193383
| Device Trade Name: | SnugKap |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | HeadStart, Ltd.<br>801 – 625 5th Ave<br>New Westminster, BC V3M 1X4,<br>Canada<br>Phone: 1 (844) 744 3237 |
| Contact: | Mr. Jason Goodnough<br>Chief Executive Officer<br>HeadStart Medical, Ltd.<br>801 – 625 5th Ave<br>New Westminster, BC V3M 1X4,<br>Canada<br>Phone: 1 (844) 744 3237<br>james@headstartmedical.com |
| Additional Contact: | Mr. Dave McGurl<br>Director, Regulatory Affairs<br>MCRA, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>(202) 552-5797<br>dmcgurl@mcra.com |
| Date Prepared: | June 7, 2021 |
| Classifications: | 21 CFR 882.5970, Cranial Orthosis |
| Class: | II |
| Product Codes: | MVA |
| Primary Predicate: | The subject devices are substantially equivalent to the following<br>primary predicate device and reference devices |
| Manufacturer | Device Name | K Number |
|---------------------------------|-------------------------------------|----------|
| Primary Predicate | | |
| Eastern Cranial Affiliates, LLC | KidCap | K180568 |
| Reference Devices | | |
| Lorica Scientific, LLC | P-POD | K133397 |
| Orthomerica Products, Inc. | STARBand, STARlight, St. Louis Band | K180109 |
| Fairview Orthopedic Laboratory | Molded Cranial Helmet | K012920 |
## Table 1: Primary Predicate Device
{4}------------------------------------------------
#### Indications for Use:
The SnugKap is intended for medical purposes for use on infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional Plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads by applying mild pressure to prominent regions of the infant's cranium in order to improve cranial symmetry and/or shape. The device is also indicated for adjunctive use for infants from 3 to 18 months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic- and scaphocephalic-shaped heads.
## Device Description:
The SnugKap devices are cranial orthoses intended to be utilized on infant patients that suffer from head shape abnormalities. The SnugKap devices are additively manufactured from a 3D scan file of the patient's head. The scan file is digitally modified to address the anatomical/shape abnormalities in the skull. The SnugKap is then 3D printed and fabricated according to the case plan approved by the orthotist. The SnugKap is manufactured from a copolymer made of polylactic acid; additional parts are made of a variety of materials including thermoplastic polyurothane, polypropylene, and Plastazote® (white or pink) polymer foam.
## Predicate Device:
The SnugKap is substantially equivalent to a previously cleared predicate devices and has the same indications for use and function; and similar design and materials. The primary predicate device is the KidCap (K180568). The reference devices are the P-POD. STARBand, and Molded Cranial Helmet (K133397, K180109, and K012920 respectively).
## Performance Testing Summary:
Testing performed on the subject device includes worst case determination, strength testing (band), durability testing (open-close test), and pressure testing. Additional testing included biocompatibility, software validation and 3D printing validation testing. Each of these studies were designed to address risks and demonstrate substantially equivalent performance to predicate devices. Additional details on the testing is provided in the substantial equivalence table below.
## Sterilization and Cleaning:
The SnugKap device is provided clean and non-sterile and is required to be cleaned prior to initial use and cleaned daily after each use. Shelf-life data is not needed to establish maintenance of device performance characteristics over the useful life period, because of the low likelihood of time-dependent material degradation.
#### Substantial Equivalence Summary:
Comparative information presented in the 510(k) supports the substantial equivalence of the SnugKap to the primary predicate devices. Comparisons were designed to show the indications, intended use, design, and performance are equivalent between the SnugKap and primary predicate devices. An evaluation table is provided below:
{5}------------------------------------------------
| | Subject Device | Primary<br>Predicate | Reference Devices | | | Comparison |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | SnugKap | KidCap | P-POD | STARBand,<br>STARlight, St. Louis<br>Band | Molded Cranial Helmet | (Subject Device vs.<br>Predicate Devices) |
| 510(k) Number | K193383 | K180568 | K133397 | K180109 | K012920 | N/A |
| Company | HeadStart<br>Ltd. | Eastern Cranial<br>Affiliates, LLC | Lorica Scientific,<br>LLC | Orthomerica Products,<br>Inc. | Fairview Orthopedic<br>Laboratory | N/A |
| Regulation | 21 CFR 882.5970 | 21 CFR 882.5970 | 21 CFR 882.5970 | 21 CFR 882.5970 | 21 CFR 882.5970 | Identical |
| Classification | II | II | II | II | II | Identical |
| Product Code | MVA | MVA | MVA | MVA, OAN | MVA | Identical |
| Indications for<br>Use | The SnugKap is<br>intended for<br>medical purposes<br>for use on infants<br>from 3 to 18<br>months of age,<br>with moderate to<br>severe<br>nonsynostotic<br>positional<br>Plagiocephaly,<br>including infants<br>with<br>plagiocephalic-,<br>brachycephalic-,<br>and<br>scaphocephalic-<br>shaped heads by<br>applying mild<br>pressure to<br>prominent regions<br>of the infant's<br>cranium in order<br>to improve cranial<br>symmetry and/or<br>shape. The device<br>is also indicated<br>for adjunctive use | The KidCap is a<br>cranial orthosis<br>used to treat<br>abnormally shaped<br>craniums in infants<br>between the ages<br>of three (3) months<br>to eighteen (18)<br>months of age. It is<br>designed to address<br>abnormal cranial<br>configurations<br>classified as<br>nonsynostotic<br>positional<br>plagiocephaly and<br>post-operative<br>nonsynostotic<br>plagiocephaly. It<br>includes infants<br>with<br>plagiocephalic,<br>brachycephalic and<br>scaphocephalic<br>patterned head<br>shapes, and post-<br>operative<br>craniosynostosis | The P-POD<br>Helmet is a cranial<br>orthosis device<br>intended for<br>medical purposes<br>to apply static or<br>gentle pressure to<br>prominent regions<br>of an infant's<br>cranium to<br>improve cranial<br>symmetry or<br>shape. The device<br>is intended to treat<br>infants from four<br>to eighteen<br>months of age<br>with moderate to<br>severe non-<br>synostotic<br>positional<br>plagiocephaly,<br>including<br>plagiocephalic-,<br>brachycephalic-,<br>scaphocephalic-<br>shaped heads. | The STARband,<br>STARlight, and St.<br>Louis Band are intended<br>for medical purposes for<br>use on infants from 3 to<br>18 months of age, with<br>moderate-to-severe non-<br>synostotic positional<br>plagiocephaly,<br>including infants with<br>plagiocephalic-,<br>brachycephalic- and<br>scaphocephalic-shaped<br>heads by applying mild<br>pressure to prominent<br>regions of the infant's<br>cranium in order to<br>improve cranial<br>symmetry and/or shape.<br>These devices are also<br>indicated for adjunctive<br>use for infants from 3<br>to 18 months of age<br>whose synostosis has<br>been surgically<br>corrected, but who still<br>have moderate-to-severe<br>cranial deformities | The Molded Cranial<br>Helmet is used to treat<br>children 3-18 months of<br>age for moderate to severe<br>non-synostotic positional<br>plagiocephaly, including<br>infants with<br>plagiocephalic-.<br>brachycephalic- and<br>scaphocephalic-shaped<br>heads. The device is<br>intended for medical<br>purposes to apply pressure<br>to prominent regions of an<br>infant's cranium in order<br>to improve cranial<br>symmetry and/or shape.<br>The device is for use by or<br>on the order of a<br>physician. | Similar:<br>The subject device<br>and predicated<br>devices are cranial<br>orthoses that improve<br>symmetry and cranial<br>shape by using the<br>static equilibrium<br>principle by applying<br>pressure to prominent<br>regions. |
| Attribute | Subject Device<br>SnugKap | Primary Predicate<br>KidCap | Reference Devices<br>P-POD | Reference Devices<br>STARBand, STARlight, St. Louis Band | Reference Devices<br>Molded Cranial Helmet | Comparison<br>(Subject Device vs. Predicate Devices) |
| | for infants from 3 to 18 months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic- and scaphocephalic-shaped heads. | management. The KidCap utilizes the principles of static equilibrium to influence the cranial plates while utilizing an anatomical kinetic chain to influence the bones comprising the orbits, cheek structure, and lower mandible. The orthosis is designed to intimately contact the prominences of the expanding cranium but will not initiate a force upon the cranium. | | including plagiocephalic, brachycephalic- and scaphocephalic-shaped heads. | | |
| Use | Prescription Use | Prescription Use | Prescription Use | Prescription Use | Prescription Use | Identical |
| Population | 3 to 18 months | 3 to 18 months | 4 to 18 months | 3 to 18 months | 3 to 18 months | Similar:<br>The P-POD predicate device is limited to a lower age boundary of 4 months instead of 3 months. The rest of the predicate device share the exact same age range of infants. |
| Anatomical Site | Cranium | Cranium | Cranium | Cranium | Cranium | Identical |
| | Subject Device | | Reference Devices | | | Comparison<br>(Subject Device vs.<br>Predicate Devices) |
| Attribute | SnugKap | KidCap | P-POD | STARBand,<br>STARlight, St. Louis<br>Band | Molded Cranial Helmet | |
| Product Design | Cranial orthosis<br>made to<br>individual's<br>specifications | Bivalved 2-piece<br>polymer shell<br>with padded<br>interface | Cranial orthosis<br>made to<br>individual's<br>specifications | Cranial orthosis made to<br>individual's<br>specifications | Cranial orthosis made to<br>individual's specifications | Identical |
| Patient Contact<br>Duration<br>(Wearing Time) | Day 1: 8 hrs./day<br>(2x4 hrs.)<br>Day 2: 8 hrs./day<br>(1x8 hrs.)<br>Day 3: 12 hrs./day<br>(1x12 hrs.)<br>Day 4: 16 hrs./day<br>(1x16 hrs.)<br>Subsequent days:<br>23 hrs./day | N/A | 23 hours/day | 23 hours/day | N/A | Similar:<br>The SnugKap device<br>is meant to be worn<br>continuously for 23<br>hours per day after an<br>ease-in period of three<br>days that follows a<br>recommended<br>progressive use<br>schedule. The P-POD<br>and STARBand<br>equivalent devices,<br>are also meant to be<br>utilized for 23<br>continuous hours per<br>day. The<br>recommended ease-in<br>period will allow the<br>baby to familiarize<br>with the SnugKap and<br>increase comfort. |
| Materials | • Polylactic acid<br>• Plastazote<br>(closed-cell<br>cross-linked<br>polypropylene<br>foam) white or<br>pink<br>• Thermoplastic<br>Polyurothane<br>• Polypropylene | • Durr Plex™<br>• Hypoallergenic<br>Suspension<br>Padding | • Polypropylene<br>(USP Class VI<br>certified)<br>• Polyurethane<br>foam | • Copolymer plastic<br>• Pelite polyethylene<br>foam<br>• Aliplast foam<br>• Clear surlyn<br>• Clear co-polyester<br>• Velcro strap<br>• Chafe buckle | • Plastazote (closed-cell<br>cross-linked<br>polypropylene foam)<br>• Polypropylene outer<br>shell | Similar<br>All devices'<br>shells/bands are<br>manufactured from<br>copolymers. All<br>devices' straps or<br>ancillary parts are<br>manufactured from a<br>variety of<br>polypropylene and |
| Attribute | Subject Device | Primary<br>Predicate | Reference Devices | | | Comparison |
| | SnugKap | KidCap | P-POD | STARBand,<br>STARlight, St. Louis<br>Band | Molded Cranial Helmet | (Subject Device vs.<br>Predicate Devices)…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.