K111247 · Orthotic Care Services, Llp · MVA · Aug 15, 2011 · Neurology
Device Facts
Record ID
K111247
Device Name
CRANIAL REMOLDING ORTHOSIS
Applicant
Orthotic Care Services, Llp
Product Code
MVA · Neurology
Decision Date
Aug 15, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5970
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Cranial Remolding Orthosis is prescribed for use on infants between 3-18 months with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic, scaphocephalic, and brachycephalic-shaped heads, by applying pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape. The Cranial Remolding Orthosis is also prescribed for use following minimally invasive surgery for infants between 3-18 months with craniosynostosis who still need moderate to severe correction of plagiocephalic-shaped, brachycephalic-shaped, and/or scaphocephalic-shaped heads.
Device Story
Custom-fabricated cranial remolding orthosis; treats infants 3-18 months with moderate-to-severe positional plagiocephaly or post-surgical craniosynostosis. Device input: 3D digital scan of infant head via OMEGA Scanner (laser imaging). Fabrication: 3D model modified by technician; 5-axis routing machine carves positive mold. Device structure: 3/16" copolymer plastic outer shell; 1/2" hypoallergenic polyethylene foam (Pelite, Plastazote, or Aliplast) liner. Operation: Helmet maintains total contact over prominent (bossed) areas while providing voids over flattened areas to redirect cranial growth. Used in clinical settings; prescribed by physicians or qualified professionals. Benefits: Improves cranial symmetry and proportion through passive pressure application.
Clinical Evidence
No clinical trials performed. Substantial equivalence based on identical materials, design, and manufacturing processes to the predicate device. Biocompatibility assessment conducted per ISO 10993-11, concluding no new testing required due to established history of use for these materials in cranial orthotics.
Technological Characteristics
Custom-molded cranial orthosis. Materials: 3/16" copolymer plastic shell, 1/2" hypoallergenic polyethylene foam (Pelite, Plastazote, or Aliplast) liner, Dacron straps, chafe buckles, and rivets. Sensing: 3D laser scanning (OMEGA Scanner). Manufacturing: 5-axis routing of positive mold. Connectivity: Standalone. Sterilization: Not specified (non-sterile).
Indications for Use
Indicated for infants 3-18 months with moderate to severe nonsynostotic positional plagiocephaly (plagiocephalic, scaphocephalic, or brachycephalic shapes) or post-minimally invasive surgery for craniosynostosis requiring correction of these head shapes.
Regulatory Classification
Identification
A cranial orthosis is a device that is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Special Controls
*Classification.* Class II (special controls) (prescription use in accordance with § 801.109 of this chapter, biocompatibility testing, and labeling (contraindications, warnings, precautions, adverse events, instructions for physicians and parents)).
In addition to the general controls of the act, the Dynamic Orthotic Cranioplasty - DOC™ Band is subject to the following special controls in order to provide reasonable assurance of the safety and effectiveness: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include (a) contraindications for the use of the device on infants with synostosis or with hydrocephalus; (b) warnings indicating the need: (i) to evaluate head circumference measurements and neurological status at intervals appropriate to the infant’s age and rate of head growth, and to describe steps that should be taken in order to reduce the potential for restriction of cranial growth and possible impairment of brain growth and development; (ii) to evaluate the skin at frequent intervals, e.g., every three to four hours, and to describe steps that should be taken if skin irritation or breakdown occurs; (c) precautions indicating the need: (i) to additionally treat torticollis, if the positional plagiocephaly is associated with torticollis; (ii) to evaluate device fit and to describe the steps that should be taken in order to reduce the potential for restriction of cranial growth, possible impairment of brain growth and development and skin irritation and/or breakdown; (iii) to evaluate the structural integrity of the device and to describe the steps that should be taken to reduce the potential for the device to slip out of place and cause asphyxiation or trauma to the eyes or skin; (d) adverse events, i.e., skin irritation and breakdown that have occurred with the use of this device; (e) clinician’s instructions for casting the infant, for fitting the device, and for care and use of the device; and (f) parents’ instructions for care and use of the device. (3) The materials must be assessed for biocompatibility with testing appropriate for long term direct skin contact.
{0}------------------------------------------------
K111247
## Orthotic Care Services, LLP
AUG 1 5 2011
## 510(k) Summary of Effectiveness and Safety
The following summary is provided pursuant to 510(k) summaries specified in 21 CFR 807.92(a) of the Federal Food, Drug, and Cosmetic Act.
## 1.) Applicant Information:
#### Submitter and Contact:
Scott Hinshon, CO Orthotic Care Services, LLP 360 Sherman Street, Suite 299 St. Paul, MN 55102 651-291-9000
Summary Date: April 21, 2011
Subject: K11247
#### 2.) Classification and Device Name
Type: Traditional 510(k) Submission
Proprietary Name: Cranial Remolding Orthosis
Common Name: Cranial Orthosis
Classification: Class II, Cranial Orthosis, Code MVA
Classification Name: Cranial Orthosis ( 21 C.F.R. § 882.5970)
3.) Predicate Devices: STARband Cranial Remolding Orthosis (K082950/K011350) 510/k) submitter: Orthomerica Products Inc, 505 31st Street, P.O. Box 2927, Newport Beach, CA 92659, Telephone: (949) 723-4500, Facsimile: (949) 723-4501
{1}------------------------------------------------
## 4.) Device Description
The Cranial Remolding Orthosis is prescribed for use on infants between 3-18 months with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic, scaphocephalic, and brachycephalic-shaped heads by applying pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape. The Cranial Remolding Orthosis is also prescribed for use following minimally invasive surgery for infants between 3-18 months with craniosynostosis who still need moderate to severe correction of plagiocephali-shaped, brachycephalic-shaped, and/or scaphocephalic-shaped heads.
Commonly diagnosed head deformities are positional plagiocephaly, brachycephaly (head is disproportionately wide and flat posteriorly), and scaphocephaly (head is disproportionately long and narrow). In addition, the Cranial Remolding Orthosis can be used following a minimally invasive surgery for infants with craniosynostosis who still need moderate to severe correction of plagiocephaly, brachycephaly and scaphocephaly. They are less common but represent an emerging patient group.
#### Plagiocephaly, Brachycephaly and Scaphocephaly:
This condition is defined as a malformation of the skull resulting in an "oblique", "crooked", or "parallelogram" shaped head. Other similar conditions which are positional in nature are Brachycephaly and Scaphocephaly, and they result in heads that are well out of proportion to the normal width versus length. Medically diagnosed non-synostotic deformational plagiocephaly is typically assessed by a combination of clinical observation, xray, CT scan, and possibly an MRI to rule out craniosynostosis.
#### Craniosynostosis:
This conditions is caused by the premature fusion of one or more cranial sutures of the skull; causing the head to grow into a usual shape. It presents in a dissimilar way to deformational (positional) plagiocephaly. However, in cases where the physician cannot make a definitive diagnosis, patients are referred to specialists such as neurosurgeons or craniofacial surgeons. These specialists will order a test such as a CT scan or MRI to confirm the diagnosis of craniosynostosis.
Following surgery to release the diseased suture, the same principles that guide cranial remolding of deformational head shapes are applicable. A prescription will be provided if the craniofacial surgeon deems that a helmet is medically necessary. In both deformational head shapes and post-surgical head shapes the Cranial Remolding Orthosis is designed to maintain total contact over the areas where growth is not desired while providing space over areas where growth is desired.
#### Causes:
Cranial asymmetries or disproportions are commonly caused by external forces to the head and are influenced by many differing factors that include supine sleeping positions (Back to Sleep campaign in 1992), preferred head orientation, torticollis (limited head rotation), increased incidence of multiple births, low tone (hypotonia), and extended supine positioning.
{2}------------------------------------------------
The device allows for growth of the flattened areas of the infant's skull into the voids within the shell and foam lining. This combination of total contact and relief allows for significant correction of the asymmetrical head shape; thereby giving the infant's skull a more symmetrical and/or proportioned shape.
#### Material Specifications:
The Cranial Remolding Orthosis is fabricated from a 3/16" co-polymer base plastic shell with a ½" foam interface liner. The type of material used on the foam interface liner is Pelite polvethylene foam, Plastazote, or Aliplast foam spacer fills a ½" side opening, opposite the deficit, utilizing the same material as the corresponding foam liner. The anterior inferior edge is trimmed just above the eyebrows, and the sides are trimmed above and around the ears. The posterior inferior is trimmed distally to capture the occiput. A hook and loop closure (also referred to as Velcro within the industry) is attached on the side of the deficit, bridging over the side opening, keeping the device securely in place.
| Materials | 3/16" outer shell of copolymer plastic ½" hypoallergenic polyethylene foam (Pelite, Plastazote or Aliplast) 1 1/2" strap of dacron 1 1/2" chafe buckle Large flange, blind rivet or 89X speedy rivet ½" hypoallergenic polyethylene foam spacer (Pelite or Aliplast) |
|-----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### 5.) Intended Use of Device
The Cranial Remolding Orthosis is prescribed for use on infants between 3-18 months with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic, scaphocephalic, and brachycephalic-shaped heads, by applying pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape. The Cranial Remolding Orthosis is also prescribed for use following minimally invasive surgery for infants between 3-18 months with craniosynostosis who still need moderate to severe correction of plagiocephali-shaped, brachycephalic-shaped, and/or scaphocephalic-shaped heads.
#### 6.) Comparison to Predicate Device
The Cranial Remolding Orthosis is a custom orthosis used to treat an infant's abnormally shaped head from three to eighteen months of age, This condition is clinically known as positional or deformational plagiocephaly. The Cranial Remolding Orthosis contains and limits growth of these bossed and protruding areas while promoting growth of the flattened areas of the skull as facilitated by the voids and spaces created within the helmet. The Cranial Remolding Orthosis is only available when prescribed by a physician, a certified pediatric nurse practitioner or other qualified medical professional.
The Cranial Remolding Orthosis' design has been illustrated in Table 5.1 below. There is not a significant difference between the helmets' materials selections, and the processes for fabrication are substantially equivalent. The inner liner is made with hypoallergenic polyethylene foam. These materials can safely come into contact with the skin and not cause redness, rash or irritation. A speedy rivet or pop-rivet will be used to attach the strap and chafe 3
{3}------------------------------------------------
to the Cranial Remolding Orthosis. The rivet system used provides a cosmetically appealing product.
Similarities in the Cranial Remolding Orthosis and the STARband device are design, production, intended use and special controls. Both devices consist of outer plastic shells and an inner foam lining. The materials used and processes of manufacturing the device are handled in an identical manner to the predicate device, incorporating all the safety and standards of practice. The proposed indications of use, biocompatibility, function and effectiveness are equivalent to those presented by the predicate device.
| | Cranial Remolding Orthosis | STARband |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Maintains total contact over areas<br>of bossing or protrusion and<br>creates voids over areas of<br>depression or flattening to redirect<br>cranial growth toward greater<br>symmetry. | Maintains total contact over<br>areas of bossing or protrusion<br>and creates voids over areas of<br>depression or flattening to<br>redirect cranial growth toward<br>greater symmetry. |
| Materials | • 5/32" outer shell of copolymer<br>plastic<br>• ½" hypoallergenic polyethylene<br>foam (Pelite, Plastazote or<br>Aliplast)<br>• 1 1/2" strap of dacron<br>• 1 1/2" chafe buckle<br>• Large flange, blind rivet or 89X<br>speedy rivet<br>• ½" hypoallergenic polyethylene<br>foam spacer (Pelite or Aliplast) | • Outer shell of .156 copoly<br>plastic<br>• An inner liner of ½" Pelite<br>polyethylene foam or ½"<br>Aliplast foam.<br>• A strap of 1 ½" Dacron<br>• A 1 ½" chafe buckle<br>• Large Flange, Blind Rivet<br>• A Gap Block made from ½"<br>firm pelite polyethylene<br>foam<br>• A nylon washer |
| Product design | Custom fabricated cranial<br>remolding orthosis<br>Approximate weight: 6oz. | Custom made cranial orthosis,<br>approx 6oz. in weight |
| Production | Form orthosis from a positive<br>mold of the patient's head:<br>• A positive scanned model is<br>captured using the Willow<br>Wood Omega Scanner.<br>• A 3-dimentional model is<br>carved using a 5-axis routing<br>machine carver in-house<br>and/or a central fabrication site<br>prior to modification and<br>fabrication of the device.<br>• The 3-dimentional model is<br>modified by a technician under<br>the direction of a clinician.<br>Plaster buildups are added to<br>areas of flattening to create<br>symmetry and/or proper<br>proportions of the head. | • Form orthosis from a<br>positive mold of<br>infant's head.<br>• Positive mold is<br>formed based upon<br>measurements of the<br>infant's head taken by<br>the STARscanner,<br>which a 3-dimensional<br>image is made or from<br>a traditional plaster<br>cast.<br>• The 3-dimensional<br>image is used to<br>produce a positive<br>mold using a 5-axis<br>routing machine. |
Table 5.1 – Cranial Remolding Orthosis Design vs. STARband
{4}------------------------------------------------
#### Comparison of Willow Wood OMEGA Scanner vs. STARscanner
The predicate STARScanner and the OMEGA Scanner are equivalent in many ways. They both achieve the same result through slightly different means while maintaining the highest levels of safety for the patient. Both scanners are Class I FDA approved laser imaging devices that provide a 3D digital image of the patient's head that is accurate to .5mm. The both used the same computer OMEGA Tracer Software. The STARScanner uses eight cameras and four lasers to capture the image, while the OMEGA Scanner uses two cameras, one laser, and reflective stickers to capture an image that is just as accurate. The accuracy, reproducibility, and repeatability of the STARScanner and the OMEGA scanner were evaluated by scanning three cylindrical shapes five times and downloading the files three times to the digital image. Standard measurement system statistical process control procedures were used to evaluate the scan rate of error, standard deviation, repeatability, and reproducibility of multiple scans. Both devices met the predetermined acceptable criteria. Both devices are approved by the FDA to be used in scanning an infant. No eye protection is necessary as the FDA as deemed these devices as Class 1 lasers.
There are some differences between the STARScanner and the OMEGA scanner. One of the major differences between the two devices is that the predicate STARScanner is much less portable than the OMEGA scanner is; with the STARScanner the patient laying on a sled type table and has the measurement devise passed over the patients cranium when they are laying in the appropriate position. This requires the infant to lay as still as possible. The OMEGA scanner takes it readings by waving the hand wand around the infant's reflective cap; the patient can be held or sitting up.
#### 6.)Performance Standards and Data
The predicate device designed by Orthomerica (K011350) using the same techniques and materials. It incorporates an outer copolymer shell lined on its interior with a medium durometer cross linked polyethylene foam: Pelite or Volara (AliPlast) lining. (Orthomerica Retrieved April 4, 2011 from http://www.orthomerica.com/products Products, Inc. /cranial/starband.htm).
Description of Orthosis:
- Side-opening helmet .
- Proximal opening .
- 3/16" Copolymer shell .
- 1/2" Polyethylene foam liner .
- 1 ½" Velcro® strap and chafe closure .
The safety of the cranial orthosis has been demonstrated through past biocompatibility assessments which revealed that this type of device is not expected to adversely affect children under intended conditions of wear/use. Though, our Cranial Remolding Orthosis has not undergone its own individual clinical trials with respect to biocompatibility. This is in compliance with ISO 10993-11 in that after review of the production methods and materials involved, "no testing is needed if the material has a demonstrated history of use in a specified role to that of the device under design."
These assessments reveal that the device and the materials used are not expected to adversely affect the infants under the intended conditions of wear.2 The materials are not reported to cause skin irritation or any toxic effects. Furthermore, it is designed to avoid improper migration or harmful levels of pressure. The interior of the device is smooth and poses no threat to the child during application within the normal scope of its intended use.
{5}------------------------------------------------
Also, the proposed device is not in any quantifiable way different from the predicate device with regard to manufacture or materials used, thus when looking at the predicate device we can see that the helmet has been effectively used for years without complications of this sort.
Special controls are required. 21 C.F.R. § 882.5970; 63 Fed. Reg. 40, 650-651 (July 30, 1998)
1 http://www.iso.org/iso/iso_catalogue.htm
2 Littlefield TA, Beals S, Manwaring KH, Pomnatto JK, Joganic EF, Golden KA, Ripley CE. Treatment of Craniofacial Asymmetry with Dynamic Orthotic Cranioplasty, Journal of Cranio facial Surgery, 1988; 11-17.
#### 7.) Summary
റ
The safety and effectiveness data submitted to the FDA establishes that the Cranial Remolding Orthosis is substantially equivalent to applicable predicate devices. Based on the technological characteristics and performance testing the device has been determined to be substantially equivalent to the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a seal or logo. The logo is circular and contains text around the perimeter. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES". In the center of the seal is a stylized image of an eagle or other bird-like figure with three curved lines representing its wings or feathers.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Scott Hinshon, CO Orthotic Care Services, LLP 360 Sherman Street, Suite 299 St. Paul, MN 55102
AUG 1 5 2011
Re: K111247
Trade/Device Name: Cranial Remolding Orthosis Regulation Number: 21 CFR 882.5970 Regulation Name: Cranial Orthosis Regulatory Class: Class II Product Code: MVA Dated: June 17, 2011 Received: June 24, 2011
Dear Mr. Hinshon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{7}------------------------------------------------
Page 2 – Mr. Scott Hinshon, CO
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1118001.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notfication" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolloffre mumber (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Ruchan, M.D.
Melvin R. Feldman, M.D.
Malvina B. Eydelman, M.E Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K111247
## Device Name: Cranial Remodeling Orthosis
## Indications For Use:
The Cranial Remolding Orthosis is prescribed for use on infants between 3-18 months with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic, scaphocephalic, and brachycephalic-shaped heads, by applying pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape. The Cranial Remolding Orthosis is also prescribed for use following minimally invasive surgery for infants between 3-18 months with craniosynostosis who still need moderate to severe correction of plagiocephali-shaped, brachycephalic-shaped, and/or scaphocephalic-shaped heads.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
KRISTEN BOWSHER (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
Page 1 of __ 1________________________________________________________________________________________________________________________________________________________________
510(k) Number_K III 247
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.