The ECT Cotton Bite Block is a single-use (disposable) oral protector for use during seizures induced by electroconvulsive therapy (ECT), to protect the patient's lips, teeth and tongue. It is placed between the back molars, with the tongue medial, and one side of the bite block sticking out of the patient's mouth to keep the mouth open.
Device Story
Single-use oral protector; used during electroconvulsive therapy (ECT) in hospital OR or ECT suite. Device placed between back molars to prevent dental/oral injury during seizure-induced masseter muscle contraction. Cylindrical cotton roll design allows mouth to remain open for oxygen administration; protects incisors and compromised molars by limiting compression to one side. Physician/clinician-operated. Benefits include cushioning of teeth and maintenance of airway access.
Clinical Evidence
Bench testing only. Compression testing performed at 600 N maximum human bite force; device maintained >7 mm thickness. Biocompatibility testing (sensitization, intracutaneous irritation, cytotoxicity) conducted at WuXi Laboratories; all tests passed with zero scores.
Technological Characteristics
Material: 100% purified medical grade cotton roll bound by cotton silky yarns. Form factor: Cylindrical, 3/4" diameter; available in 2.75" (medium) and 4" (large) lengths. Single-use. Non-sterile (implied).
Indications for Use
Indicated for patients undergoing electroconvulsive therapy (ECT) to protect lips, teeth, and tongue from injury during seizure-induced masseter muscle contraction.
Regulatory Classification
Identification
A bite block is a device inserted into a patient's mouth to protect the tongue and teeth while the patient is having convulsions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 4, 2021
MECTA Corporation Adrian Kettering Vice President 19799 SW 95th Avenue Suite B Tualatin, Oregon 97062
Re: K201309
Trade/Device Name: ECT Cotton Bite Block- size Large, ECT Cotton Bite Block -size Medium Regulation Number: 21 CFR 882.5070 Regulation Name: Bite Block Regulatory Class: Class II Product Code: JXL Dated: March 6, 2021 Received: March 8, 2021
Dear Adrian Kettering:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201309
Device Name ECT Cotton Bite Block
Indications for Use (Describe)
The ECT Cotton Bite Block is a single-use (disposable) oral protector for use during seizures induced by electroconvulsive therapy (ECT), to protect the patient's lips, teeth and tongue. It is placed between the back molars, with the tongue medial, and one side of the bite block sticking out of the patient's mouth to keep the mouth open.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510K Summarv
## Submitter's Information
Name of Sponsor: MECTA Corporation Address: 19799 SW 95th Ave, Suite B, Tualatin, OR 97062 Contact Name: Adrian Kettering Telephone: 503-612-6780 Fax: 503-612-6542 Email: akettering@mectacorp.com
## Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)|
Trade Name/Model: MECTA ECT Cotton Bite Block Classification Name: Neurological Therapeutic Devices, Bite Block Regulation Number: 882.5070 Product Code: JXL Classification Panel: Neurological Devices Device Class: II
## Identification of Predicate Device
The identified predicate within this submission is the Somatics, Inc. Ventil-A Oral Protector that was cleared by FDA through 510K No. K992269 (Decision date: October 25, 1999).
## Description of Device
During the passage of the electrical stimulus during ECT, despite the fact that the patient has been given a systemic muscle relaxant, the masseter muscles contract forcefully, necessitating the use of a bite block or mouth guard. The ECT Cotton Bite Block is intended for single-use only, and is placed between a patient's back molars in order to protect the teeth and tongue during an electroconvulsive therapy (ECT) treatment. Patients who are mentally ill commonly have poor oral hygiene. The placement of the bite block, only on one side of the mouth, allows for any damaged or compromised molars on the opposite side of the mouth to be free from compression. In addition, incisors avoid the risk of being damaged during compression, and the ¾' thick soft cotton provides necessary cushioning for the remaining teeth that are biting down on the cotton roll. One end of the bite block remains outside of the mouth, propping the mouth open for oxygen to be administered, and to ensure that the bite block is staying in place.
Material: The ECT Cotton Bite Block is made of 100% purified medical grade cotton, and is a cylindrical shaped dense cotton roll wrapped in cotton silky yarns.
Sizes: There are two sizes, and size used is determined by the patient's mouth size. The large size is 3/4 inch diameter and 4 inches long, and the medium is ¾ inch diameter and 2 ¾ inches long.
{4}------------------------------------------------
Environment: Electroconvulsive Therapy is administered in a hospital operating room or ECT Suite.
### Intended Use
The ECT Cotton Bite Block is a single-use (disposable) oral protector for use during seizures induced by electroconvulsive therapy (ECT), to protect the patient's lips, teeth and tongue. It is placed between the back molars, with the tongue medial, and one side of the bite block sticking out of the patient's mouth to keep the mouth open.
#### Comparison of Technological Characteristics
The ECT Cotton Bite Block is substantially equivalent to its predicate device in that both bite blocks protect the compression of the back molars when the masseter muscles contract during electroconvulsive therapy, both allow the airway to remain open, and both are single-use. However, the ECT Cotton Bite Block differs in the material used and the shape.
#### Substantial Equivalence
| Comparison<br>Item | Proposed Device:<br>ECT Cotton Bite Block | Predicate Device:<br>Ventil-A Oral Protector |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Shape | A cylindrical shape that offers two sizes, depending upon the size of the patient's mouth. For both, the diameter is ¾", and the two lengths are 2.75" for medium or 4" for large. This shape allows minimal teeth to be affected by compression, especially the more delicate and vulnerable front incisor teeth, and any damaged or compromised molars on one side of the mouth. | A triangular shape with rounded edges and a punched hole in the middle. Thickness is 3/8". Only available in one size. |
| Material | A one-piece bite block made of 100% Cotton roll bound by Cotton Silky Threads. | A one-piece bite block made of closed-cell foam construction. |
#### Performance and Biocompatibility Testing
Compression testing was conducted on the ECT Cotton Bite Block, using the maximum human bite force of 600 N, and the ECT Cotton Bite Block was found to compress enough to cushion the teeth, but did not compress under the lower limit of 7 mm compression.
Extensive biocompatibility testing was conducted at WuXi Laboratories, including sensitization, intracutaneous irritation and cytotoxicity. The MECTA ECT Cotton bite block passed all of these tests in a very straightforward manner with a score of all zeros.
#### Substantial Equivalence Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, MECTA Corporation concludes that:
- The intended use of the MECTA ECT Cotton Bite Block is the same as the predicate device.
{5}------------------------------------------------
- The ECT Cotton Bite Block protects the ECT patient's mouth, with differences in the shape and material in order to protect compromised teeth, which can be common for patients who are mentally ill.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.