VersaWrap Nerve Protector is indicated for the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue.
Device Story
VersaWrap Nerve Protector is a thin, flexible, bioresorbable implant; functions as a non-constricting gelatinous interface between injured peripheral nerves and surrounding tissues. Designed to encase the neural environment; prevents adhesions; absorbs over time. Used in surgical settings; placed by surgeons around peripheral nerves. Provides physical barrier; reduces potential for nerve entrapment; benefits patient by supporting nerve healing environment. Material composition: calcium alginate and hyaluronic acid.
Clinical Evidence
No human clinical data. Evidence based on non-clinical bench testing (dimensional, swelling, strength, sterilization, biocompatibility per ISO 10993) and an animal study (rat model). Animal study compared VersaWrap, predicate, and untreated groups across early, mid, and late timepoints. Endpoints included motor/sensory neurological assessments and histopathological evaluation of local/systemic tissue response. Results showed equivalent safety and performance to predicate with no significant adverse events.
Technological Characteristics
Materials: calcium alginate and hyaluronic acid. Form factor: flexible sheet. Sterilization: eBeam (SAL 10^-6 per ISO 11137). Biocompatibility: ISO 10993 compliant (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity).
Indications for Use
Indicated for management of peripheral nerve injuries without substantial loss of nerve tissue.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 14, 2020
Alafair Biosciences, Inc. % Angela Mallery Regulatory Consultant NAMSA 400 Highway 169 South, Suite 500 Minneapolis, Minnesota 55426
Re: K201631
Trade/Device Name: VersaWrap Nerve Protector Regulation Number: 21 CFR 882.5275 Regulation Name: Nerve Cuff Regulatory Class: Class II Product Code: JXI Dated: June 12, 2020 Received: June 16, 2020
Dear Angela Mallery:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam Pierce, Ph.D. Assistant Director (Acting) DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201631
Device Name VersaWrap Nerve Protector
## Indications for Use (Describe)
VersaWrap Nerve Protector is indicated for the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" width="15" x="0" y="0"></rect> <path d="M2 2 L13 13 M2 13 L13 2" stroke="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-family: Arial;"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" width="15" x="0" y="0"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary<br>VersaWrap Nerve Protector | | | | |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| K201631 | | | | |
| Submitted by: | Alafair Biosciences, Inc.<br>6101 W Courtyard Drive<br>Ste. 2-225<br>Austin, TX 78730<br>800.206.5586; info@alafairbiosciences.com | | | |
| Date Prepared: | September 8, 2020 | | | |
| Contact: | Ben Walthall, Ph.D.<br>Chief Regulatory Officer<br>800.206.5586; info@alafairbiosciences.com | | | |
| Product Name | VersaWrap Nerve Protector | | | |
| Common Name | Cuff, Nerve | | | |
| Classification number | 21 CFR 882.5275 | | | |
| Product Code | JXI | | | |
| Primary Predicate: | Integra NeuraWrap Nerve Protector K041620 | | | |
| Reference Device: | Alafair VersaWrap Tendon Protector K200311 | | | |
| Device Description: | VersaWrap Nerve Protector (VersaWrap) is designed to function as an interface between an<br>injured nerve and surrounding tissues and is indicated for use in peripheral nerve injuries<br>where there is no significant loss of nerve tissue.<br>VersaWrap Nerve Protector is a thin, flexible implant, designed to be a non-constricting<br>gelatinous interface encasing peripheral nerves and the neural environment; that starts to<br>absorb after implant.<br>VersaWrap Nerve Protector is designed to be flexible and conformable for placement<br>around a peripheral nerve | | | |
| Indications for Use: | VersaWrap Nerve Protector is indicated for the management of peripheral nerve injuries in<br>which there has been no substantial loss of nerve tissue. | | | |
| Functional and Safety<br>Testing: | Biocompatibility studies demonstrated the Nerve Protector is non-cytotoxic, non-pyrogenic,<br>non-irritating, non-sensitizing, non-toxic, and non-genotoxic. Animal studies demonstrated<br>the device performed as labeled. | | | |
| Comparative<br>Technology<br>Characteristics: | Parameter<br>Device name | Device<br>VersaWrap | Predicate Device<br>NeuraWrap | Reference Device<br>VersaWrap |
| | Company<br>Name | Alafair Biosciences | Integra Lifesciences | Alafair Biosciences |
| | 510(k) # | K201631 | K041620 | K200311 |
| | Material | Calcium alginate and<br>hyaluronic acid | Type 1 bovine<br>collagen and<br>chondroitin-6-sulfate | Calcium alginate and<br>hyaluronic acid |
| | Intended Use | designed to be an<br>interface between the<br>injured area and the<br>surrounding tissue | designed to be an<br>interface between the<br>injured area and the<br>surrounding tissue | designed to be an<br>interface between the<br>injured area and the<br>surrounding tissue |
| | Packaging | Double Pouch | Double Pouch | Double Pouch |
| | Physician<br>Structure | Sheet | Tube | Sheet |
| | Sterilization<br>Method | eBeam | EO | eBeam |
| | Testing was conducted. The specific conclusions drawn from the testing is that the<br>testing demonstrates the proposed device is as safe, as effective, and performs as well<br>as or better than the legally marketed predicate device. | | | |
| | Test/Purpose | Test Method Summary | Discussion of results | |
| Non-Clinical Tests<br>Performed: | Dimensional and<br>visual inspection | Direct measurement of the mass,<br>lengths, width, thickness with<br>micrometers, calipers, pH meter<br>or balance; and visual inspection. | Dimensional analysis<br>was completed to verify<br>the dimensions and<br>visual characteristics and<br>met specifications; | |
| | Pliability/ Tissue<br>conformance and<br>adherence | The device is placed on a<br>simulated model and graded for<br>visual conformance and<br>adherence. | Visual inspection was<br>conducted and met<br>specifications. | |
| | Swelling | Direct measurement of the device<br>(with micrometers, calipers) when<br>exposed to fluid. | Dimensional analysis<br>was completed and met<br>specifications. | |
| | Strength/Puncture | Direct measurement of the device<br>strength/puncture when measured<br>with a probe. | Testing was completed<br>and met specifications. | |
| | Pouch Seal<br>Integrity/Sterile<br>Barrier | The integrity of the pouch was<br>verified with a visual inspection,<br>peel tester, and burst tester. | Testing was completed<br>and met specifications. | |
| | Sterilization | The e-beam sterilization process<br>was validated to achieve a sterility<br>assurance level (SAL) of 10-6 per<br>ISO 11137 | Testing was completed<br>and met specifications. | |
| | Bacterial<br>endotoxin<br>validation | The bacterial endotoxin validation<br>testing was conducted per USP<br><85> Bacterial Endotoxins and<br>Ph.Eur. Chpt 2.6.14 | Testing was completed<br>and met specification | |
| | Cytotoxicity | The device was evaluated for<br>potential cytotoxic effects using an<br>in vitro mammalian cell culture test.<br>This study was conducted following<br>the guidelines of ISO 10993-5 | The test extract showed no<br>evidence of causing cell<br>lysis or toxicity. | |
| | Sensitization | The device was evaluated for the<br>potential to cause delayed dermal<br>contact sensitization in a guinea pig<br>maximization test. This study was<br>conducted based on the<br>requirements of ISO 10993-10 | The test article showed no<br>evidence of causing delayed<br>contact sensitization. | |
| | Irritation | The device was evaluated for the<br>potential to cause irritation<br>following intracutaneous injection.<br>This study was conducted based on<br>ISO 10993-10 | There was no evidence of<br>significant irritation from<br>the extracts injected<br>intracutaneously. | |
| | Acute systemic<br>toxicity | The device was evaluated for acute<br>systemic toxicity. This study was<br>conducted based on ISO 10993-11. | The sesame oil test article<br>extracts and the sodium<br>chloride test article extracts<br>injected met the passing<br>requirements of the test. | |
| | Pyrogenicity | The device was evaluated for<br>material mediated pyrogenicity in<br>the rabbit. The test was conducted | The total rise of<br>temperatures during the<br>observation period was | |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
| Genotoxicity | The device was evaluated for<br>mutagenic potential, mutagenic<br>changes, and to produce cytogenetic<br>damage | Extract did not cause<br>cytogenetic damage/<br>micronuclei formation.<br>Extracts were well within<br>the limits defined for a<br>negative response and the<br>test article is considered<br>non-mutagenic and<br>considered to be non-<br>mutagenic |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subchronic<br>toxicity | The device was surgically<br>implanted in the subcutaneous<br>tissue of the rat to evaluate potential<br>systemic toxicity and local tissue<br>response at the implantation sites | No evidence of systemic<br>toxicity following<br>subcutaneous implantation.<br>Microscopically, the test<br>article was classified as a<br>non-irritant. |
| Muscle<br>implantation | The purpose of this study was to<br>assess the local tissue effects and<br>absorption profile following<br>implantation in muscle tissue in at<br>multiple timepoints. | At long term timepoints the<br>device was considered to be<br>a non-irritant and a slight<br>irritant at earlier weeks<br>following implantation<br>when compared to the<br>control. At the study mid-<br>point, microscopic<br>evaluation demonstrated the<br>device is bioresorbing as<br>expected. This variable and<br>slight irritation is expected<br>for a device undergoing<br>absorption. |
| Neurotoxicity<br>assessment | A risk analysis was conducted | An analysis was conducted<br>to determine any<br>neurotoxicity risk to the<br>patient |
{7}------------------------------------------------
| Animal Study | The animal study was conducted on multiple injury types for<br>subject device, predicate control device, and untreated groups.<br>Nerve defects were comparable to the clinical use of the device.<br>Data was collected at an early period, a mid-term period; and a<br>late period.<br>The rat is often used in nerve injury models and is suggested in<br>the ISO standards and Organisation for Economic Co-operation<br>and Development guidelines as an acceptable animal model for<br>evaluating the local tissue response of various articles; and the<br>sample size was selected to be large enough to show consistent<br>results and to be able to compare the test, control, and untreated<br>groups.<br>As recommended by FDA, motor and sensory neurological<br>assessments were conducted prior to surgery, and throughout the<br>study. VersaWrap animals and the predicate control animals<br>demonstrated equivalent safety and equivalent performance<br>results; there were no significant events reported in any group.<br>Histopathological evaluations were performed to assess local and<br>systemic tissue effects specific to the test and predicate control<br>articles. Overall, there was no apparent difference in the character<br>or severity of the reactivity to the test or predicate implant. There<br>were no microscopic changes in the nerve at any interval for<br>either test or Integra implants; the microscopic appearance of the<br>nerve and surrounding tissue for both the control and VersaWrap<br>were similar. |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | Based on the results of animal studies, prior in vitro product characterization studies, in vitro and in vivo biocompatibility and performance studies, we conclude the device is as safe as, and substantially equivalent to its predicate devices. |
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.