IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control Unit
K192289 · Irras USA, Ltd. · JXG · Nov 15, 2019 · Neurology
Device Facts
Record ID
K192289
Device Name
IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control Unit
Applicant
Irras USA, Ltd.
Product Code
JXG · Neurology
Decision Date
Nov 15, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The use of IRRAflow® CNS System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.
Device Story
System monitors ICP and drains CSF; consists of control unit, tube set, and catheter. Input: CSF pressure via catheter. Operation: Gravity-driven drainage; peristaltic pump and pinch valve in control unit automate flow based on user settings; allows bolus injections to clear catheter clogs. Output: Real-time ICP display, drainage control. Used in clinical settings by healthcare providers. Provider views ICP on control unit interface; adjusts drainage parameters to manage patient ICP. Benefits: Automated drainage and clog clearance; continuous monitoring; reduces manual intervention.
System includes control unit, silicone catheter (9Fr, 400mm), and disposable tube set. Sensing via integrated pressure transducers. Energy source: electrical (battery backup included). Connectivity: standalone. Sterilization: E-beam (catheter) and Ethylene Oxide (tube set). Software: rule-based control for pump/valve. Standards: ISO 10993, ISO 11135, ISO 11137, ISO 14971, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8.
Indications for Use
Indicated for patients requiring intracranial pressure (ICP) monitoring and external drainage of intracranial fluid to reduce ICP.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2019
IRRAS USA, Inc. Niloufa Insanally Head of Regulatory Affairs 11975 El Camino Real, 3rd Floor San Diego, California 92130
Re: K192289
Trade/Device Name: IRRAflow® CNS System Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG, GWM Dated: August 21, 2019 Received: August 23, 2019
Dear Niloufa Insanally:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D. Director (Acting) DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192289
Device Name IRRAflow® CNS System
Indications for Use (Describe)
The use of IRRAflow® CNS System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="text-align:left;"><span> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> |
|---------------------------------------------------------------------------------------------------|
| <div style="text-align:left;"><span> Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Applicant: IRRAS USA, Inc. 11975 El Camino Real, Suite 304 San Diego, CA 92130 U.S.A Phone Number: 858-247-7033 Fax number: 866-575-1002 FDA Registration #: 3013508628
Primary Contact Personnel:
Vinny Podichetty VP Clinical, Regulatory Affairs and Quality Phone Number: 949-356-3094 Fax Number: 866-575-1002 Email: Vinny.podichetty@jirras.com
Secondary Contact Personnel:
| Kleanthis G. Xanthopoulos, Ph.D., | |
|-----------------------------------|----------------------------------|
| President and CEO | |
| Phone Number: | 858-336-8093 |
| Fax Number: | 866-575-1002 |
| Email: | Kleanthis.xanthopoulos@irras.com |
| Date prepared: | August 21, 2019 |
|----------------|-----------------------------------------------|
| Trade name: | IRRAflow® CNS System |
| Common Name: | CSF Drainage System with ventricular catheter |
### Primary Classification:
| Name: | Central Nervous System Fluid Shunt and Component |
|---------------|--------------------------------------------------|
| Product Code: | JXG |
| Regulation: | 21 CFR 882.5550 |
Secondary Classification: Name: Intracranial Pressure Monitoring Device Product Code: GWM Regulation: 21 CFR 882.1620
Predicate and Reference Device(s): IRAflow® CNS System (K171880)
{4}------------------------------------------------
# DEVICE DESCRIPTION
The IRRAflow® CNS System is an intracranial pressure (ICP) monitoring and drainage system. The IRRAflow® CNS System consists of an IRRAflow Control Unit and two sterile disposable parts, the IRRAflow Tube Set and the IRRAflow Catheter.
The drainage flow of cerebrospinal fluid (CSF) into the IRRAflow Catheter is uni-directional and gravity-driven; there is no recirculation of the CSF. A parallel line from the saline infusion bag is used in case clearance at the tip of the catheter is required. The IRRAflow Tube Set has a cassette that clicks on to the IRRAflow Control Unit and aligns the tubing against a peristaltic pump and pinch valve. An aspiration bag is attached to the Control Unit tape measure, defining the height of the bag relative to the catheter's tip position in the patient's head and thus controlling the speed of drainage. The tubing and catheter can be disconnected and connected by standard Luer-Lock connectors. Settings can be changed via the user interface on the Control Unit.
The default mode provides drainage and measuring ICP, allowing single bolus injections when indicated. The bolus injections allow the catheter to be flushed when it becomes clogged. CSF or intracranial fluid samples can be taken from the aspiration port.
### INDICATIONS FOR USE
The use of IRRAflow 0 CNS System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.
# SUBSTANTIAL EQUIVALENCE
The IRRAflow® CNS system's intended use, technological characteristics and principles of operation are the same as the predicate device.
Comparison of these Monitoring Systems is provided in table 1 below.
{5}------------------------------------------------
| Items | PREDICATE | IRRAflow® CNS System<br>(JXG), (GWM)<br>(New) | Equivalence |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| | IRRAflow® CNS System (K171880)<br>(JXG), (GWM) | | |
| Primary Product | JXG | JXG | Same |
| Primary Regulation<br>Number: | 21 CFR 882.5550 | 21 CFR 882.5550 | Same |
| Secondary Product | GWM | GWM | Same |
| Secondary Regulation<br>Number: | 21 CFR 882.1620 | 21 CFR 882.1620 | Same |
| Indications for Use | The use of IRRAflow® CNS System<br>is indicated when intracranial<br>pressure monitoring is required and<br>for externally draining intracranial<br>fluid as a means of reducing<br>intracranial pressure in patients where<br>an external drainage and monitoring<br>system is needed for ≤ 24 hours. | The use of IRRAflow® CNS System<br>is indicated when intracranial<br>pressure monitoring is required and<br>for externally draining intracranial<br>fluid as a means of reducing<br>intracranial pressure in patients where<br>an external drainage and monitoring<br>system is needed. | See<br>discussion<br>below |
| Injection/ CSF<br>Sampling Ports | Yes | Yes | Same |
| Unidirectional<br>Flow of Drained Fluid | Yes | Yes | Same |
| Fluid Injection<br>Capability | Yes | Yes | Same |
| Attaches to separate,<br>commercially<br>available EVD<br>Catheter | Yes<br>The IRRAflow system attaches to<br>IRRAflow Catheter which is an EVD<br>Catheter part of the complete system. | Yes<br>The IRRAflow system attaches to<br>IRRAflow Catheter which is an EVD<br>Catheter part of the complete system. | Same |
| Sterile Disposable<br>tubing set | Yes | Yes | Same |
| CSF Drainage Bag | Yes | Yes | Same |
| Gravity drainage of<br>Method to control<br>gravity drainage of<br>CSF | Yes<br>Automated adjustment based on user<br>settings via a stepper-motor<br>controlled, tube-pinching mechanism<br>to either compress or release the<br>compliant drainage tubing contained<br>within the sterile, disposable<br>Cartridge. | Yes<br>Automated adjustment based on user<br>settings via a stepper-motor<br>controlled, tube-pinching mechanism<br>to either compress or release the<br>compliant drainage tubing contained<br>within the sterile, disposable<br>Cartridge. | Same |
| Pressure Transducer<br>for ICP Measurement | Yes<br>(The IRRAflow system integrates<br>transducers into its design for<br>measurement and visual display of<br>ICP) | Yes<br>(The IRRAflow system integrates<br>transducers into its design for<br>measurement and visual display of<br>ICP) | Same |
| Items | PREDICATE<br>IRRAflow® CNS System (K171880)<br>(JXG), (GWM) | IRRAflow® CNS System<br>(JXG), (GWM)<br>(New) | Equivalence |
| Software-based,<br>Powered Console for<br>User Interface, User<br>Settings and Alarm<br>Adjustments, Data<br>Storage and Display,<br>and Alarms for ICP | Yes | Yes | Same |
| Method to account for<br>location of ventricles<br>via patient head<br>position | Reference marks on the device to<br>allow for the system to be aligned<br>with patient's head positioning. | Reference marks on the device to<br>allow for the system to be aligned<br>with patient's head positioning. | Same |
| Measured Pressure<br>Range | -80 mmHg to +100 mmHg | -80 mmHg to +100 mmHg | Same |
| Displayed ICP | Yes | Yes | Same |
| Battery Back-up | Yes | Yes | Same |
# Table 1 – Substantial Equivalence Comparison of the Monitoring Systems
{6}------------------------------------------------
### Discussion of differences in Table 1 IFU:
| PREDICATE | IRRAflow CNS system (New) | Discussion: |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IRRAflow CNS system (K171880) | | |
| Indications for Use: | | |
| The use of IRRAflow® CNS System<br>is indicated when intracranial<br>pressure monitoring is required and<br>for externally draining intracranial<br>fluid as a means of reducing<br>intracranial pressure in patients<br>where an external drainage and<br>monitoring system is needed for ≤ 24<br>hours. | The use of IRRAflow® CNS System<br>is indicated when intracranial<br>pressure monitoring is required and<br>for externally draining intracranial<br>fluid as a means of reducing<br>intracranial pressure in patients<br>where an external drainage and<br>monitoring system is needed. | The difference in the indications for<br>us in the duration of monitoring.<br>The indications for use for the<br>predicate (K171880) was ≤ 24<br>hours. Biocompatibility testing has<br>been conducted and provided with<br>this submission that the duration of<br>use can be > 24 hours but less than or<br>equal to 5 days. |
# EVD Catheter:
### SUBSTANTIAL EQUIVALENCE
The IRRAflow® CNS System's Catheter technological characteristics and principles of operation are similar to the IRRAflow® CNS System Catheter (K171880). There have been no changes to the system catheter. The materials, design and principle of operations remain the same.
Comparison of these Catheters is provided in table 2 below
{7}------------------------------------------------
| Items | PREDICATE<br>IRRAflow® CNS System Catheter<br>(K171880) | IRRAflow® CNS System Catheter<br>(New) | Equivalence |
|---------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------|
| Target Population | Any patient needing removal of<br>intracranial fluids from the brain<br>ventricles | Any patient needing removal of<br>intracranial fluids from the brain<br>ventricles | Same |
| Anatomical Sites | Brain ventricles | Brain ventricles | Same |
| Implant Procedure | Designed to be placed through a<br>prepared opening through the skull<br>and into the brain ventricle | Designed to be placed through a<br>prepared opening through the skull<br>and into the brain ventricle | Same |
| Catheter Size | 9Fr | 9Fr | Same |
| Catheter Length | 400mm | 400mm | Same |
| Catheter Sideports | Yes | Yes | Same |
| Catheter End- hole | Closed | Closed | Same |
| Catheter Depth<br>Markers | Yes | Yes | Same |
| Catheter material | Silicone | Silicone | Same |
| Antimicrobial<br>Agents | None | None | Same |
| Catheter Tip | Radiopaque | Radiopaque | Same |
| Biocompatibility | Tissue contact tested per ISO 10993:<br>Biological Evaluation of<br>Medical Devices | Tissue contact tested per ISO 10993:<br>Biological Evaluation of<br>Medical Devices | Same |
| Cytotoxicity | Acceptable | Acceptable | Same |
| Provided Sterile | Yes | Yes | Same |
| Packaging | Tyvek/polyester pouch | Tyvek/polyester pouch | Same |
| Functional Use | 5 days | 5 days | Same |
| Shelf Life | 18 months | 18 months | Same |
### Table 2 - Substantial Equivalence Comparison - Catheters
{8}------------------------------------------------
### Verification and Validation Documentation:
The IRRAflow® CNS System tests include verification and validation performance testing as well as externals standards testing to demonstrate no new safety and effectiveness issues are raised with this new device. Analyses demonstrate that system accuracy and performance are adequate for the established intended use. In conclusion, the IRRAflow® CNS System is substantially equivalent to the predicate device.
Table 3 below, identifies the testing conducted on the IRRAflow® CNS System to demonstrate substantial equivalence.
| Test | Test Method Summary | Results |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility Testing | | |
| Systemic Toxicity,<br>Mediated Pyrogen | The purpose of the study is to determine if a<br>saline extract of the test article causes a febrile<br>response in rabbits. | PASS Clinical Observations: The test<br>article were determined to be non-<br>pyrogenic. |
| Acute Systemic Toxicity,<br>Injection Test | The purpose of the test was to screen test article<br>extracts for potential toxic effects as a result of a<br>single-dose systemic injection in mice. | PASS Clinical Observations: None of<br>the animals on study were observed with<br>abnormal clinical signs indicative of<br>toxicity during the test period. |
| Irritation/Intracutaneous<br>Reactivity Test | The purpose of the test was to determine if any<br>chemicals that may leach or be extracted from the<br>test article were capable of causing local irritation<br>in the dermal tissues of rabbits. | PASS Clinical Observations: None of the animals on study showed<br>abnormal clinical signs during the<br>observation periods. |
| Indirect Hemolysis<br>(Extract) Test | The test is designed to determine the hemolytic<br>properties of a medical device/material. | PASS Clinical Observations: All test<br>method acceptance criteria were met. |
| Cytotoxicity (MEM<br>Elution) Test | The Minimal Essential Media (MEM) Elution test<br>was designed to determine the cytotoxicity of<br>extractable substances. | PASS Clinical Observations: All test<br>method acceptance criteria were met. |
| Sensitization Test | This test was designed to evaluate the allergenic<br>potential or sensitizing capacity of a test article. | PASS Clinical Observations: None of<br>the animals in the study showed<br>abnormal clinical signs. |
| Genotoxicity | The test was designed to screen the products and<br>materials to determine if they cause structural<br>chromosome aberrations in industry standard<br>Chinese Hamster Ovary (CMC) cells. | PASS Clinical Observations: All test<br>method acceptance criteria were met. |
| Implant Study | This study was conducted for test of local effects<br>after implantation. | PASS Clinical Observations: All test<br>method acceptance criteria were met. |
| Subacute Toxicity | The purpose of this study is to closely examine<br>the toxicological hazard and to evaluate and<br>address any risks associated with the biological<br>endpoints of subacute/subchronic and chronic<br>toxicity. | PASS Clinical Observations: All test<br>method acceptance criteria were met. |
| Test | Test Method Summary | Results |
| Bench and Electrical Testing | | |
| IRRAflow® CNS System<br>Verification | The purpose of this test is to document the results of the system verification testing and system<br>regression verification testing. | PASS The system has been shown to<br>comply with the documented<br>requirements for the system. |
| IRRAflow® CNS System<br>Validation | The validation test procedures for the IRRAS<br>IRRAflow CNS system were designed to ensure<br>that the device complies with established<br>requirements. | PASS The system has been shown to<br>comply with the documented<br>requirements for the system. |
| IRRAflow® CNS System<br>Static Analysis | The purpose of this test is to describe the process and results from the static analysis of the<br>IRRAflow CNS system software. | PASS There were no errors encountered<br>when the static analysis was conducted. |
| Basic Safety test | The following tests were conducted: Input Test;<br>Heating Test; Leakage Current Test; Dielectric<br>Voltage Test. | PASS Acceptance criteria has been met. |
| Electromagnetic<br>Compatibility test | The objective of the testing is to determine<br>compliance with IEC 60601-1-2:2014 Class B. | PASS Acceptance criteria has been met. |
| IRRAflow Catheter<br>Performance Test | The purpose of this test is to describe the tensile<br>test results for the IRRAS catheter. | PASS Acceptance criteria has been met. |
| IRRAflow Catheter<br>Torsion and Shear Test | The purpose of this test is to describe the torsion<br>and shear testing results for the IRRAS catheter. | PASS Acceptance criteria has been met. |
| IRRAflow Catheter<br>Drainage Flow Test | The purpose of this test is to describe the drainage<br>testing methods for the IRRAS catheter. | PASS Acceptance criteria has been met. |
| Shelf Life / Package Integrity Testing | | |
| Validation of sterile barrier | The purpose of this test is to describe the<br>procedures for validating that the sterile barrier<br>meets design and standards requirements. | PASS Acceptance criteria has been met. |
| Packaging Peel Test | The purpose of this test is to describe the roll<br>packing seal testing results for the IRRAS<br>catheter and tube set. | PASS Acceptance criteria has been met. |
| Aging Test | The purpose of this test is to describe the<br>objective evidence that the IRRAS catheter and<br>tube set meet the appropriate requirements for<br>transportation and aging. | PASS Acceptance criteria has been met. |
| Sterilization Testing | | |
| Sterilization process for<br>the IRRAflow Catheter<br>and IRRAflow Tube Set | This study was conducted to validate the<br>effectiveness of electron beam radiation of<br>IRRAflow Catheter and Ethylene Oxide<br>sterilization of IRRAflow Tube Set. | PASS Acceptance criteria has been met. |
## Table 3 - IRRAflow® CNS System Testing
{9}------------------------------------------------
# Previous 510(k) Clearances:
The following are a list of the clearances for the Products discussed in this document:
| Product | Clearance Number | Date |
|----------------------|------------------|------------|
| IRRAflow® CNS System | K171880 | 07/13/2018 |
{10}------------------------------------------------
# Recognized Consensus Standards:
| Recognized Consensus Standards: | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--------------------------|------------------------|
| The IRRAflow® CNS System was developed in compliance with the following standards: | | | |
| Standard | FR Recognition<br>List Number | FR Recognition<br>Number | Date of<br>Recognition |
| ISO 10993-1 Fourth Edition 2009-10-15, Biological<br>Evaluation Of Medical Devices - Part 1: Evaluation<br>And Testing Within A Risk Management Process<br>[Including: Technical Corrigendum 1 (2010)] | 044 | 2-220 | 07/26/2016 |
| ISO 10993-3 Third Edition 2014-10-1, Biological<br>Evaluation Of Medical Devices - Part 3: Tests For<br>Genotoxicity, Carcinogenicity And Reproductive<br>Toxicity | 044 | 2-228 | 07/26/2016 |
| ISO 10993-4 Third Edition 2017-04 Biological<br>Evaluation Of Medical Devices--Part 4: Selection<br>Of Tests For Interactions With Blood | 047 | 2-248 | 08/21/2017 |
| ISO 10993-5 Third Edition 2009-06-01, Biological<br>Evaluation Of Medical Devices - Part 5: Tests For In<br>Vitro Cytotoxicity | 046 | 2-245 | 12/23/2016 |
| ISO 10993-6 Third Edition 2016-12-01, Biological<br>Evaluation Of Medical Devices -- Part 6: Tests For<br>Local Effects After Implantation | 047 | 2-247 | 08/21/2017 |
| ISO 10993-10 Third Edition 2010-08-01, Biological<br>Evaluation Of Medical Devices - Part 10: Tests For<br>Irritation And Skin Sensitization | 044 | 2-174 | 07/26/2016 |
| ISO 10993-11 Second Edition 2006-08-15,<br>Biological Evaluation Of Medical Devices - Part 11:<br>Tests For Systemic Toxicity | 044 | 2-176 | 07/26/2016 |
| ISO 10993-12 Fourth Edition 2012-07-01,<br>Biological Evaluation Of Medical Devices - Part 12:<br>Sample Preparation And Reference Materials | 044 | 2-191 | 07/26/2016 |
| ISO 10993-12 Fourth Edition 2012-07-01,<br>Biological Evaluation Of Medical Devices - Part 12:<br>Sample Preparation And Reference Materials | 044 | 2-191 | 07/26/2016 |
| ISO 11135 Second Edition 2014, Sterilization Of<br>Health-Care Products - Ethylene Oxide -<br>Requirements For The Development, Validation<br>And Routine Control Of A Sterilization Process For<br>Medical Devices | 041 | 14-452 | 04/04/2016 |
| ISO 11137-1 First Edition 2006-04-15, Sterilization<br>Of Health Care Products - Radiation - Part 1:<br>Requirements For Development, Validation And<br>Routine Control Of A Sterilization Process For<br>Medical Devices [Including: Amendment 1 (2013)] | 041 | 14-428 | 04/04/2016 |
| The IRRAflow® CNS System was developed in compliance with the following standards: | | | |
| Standard | FR Recognition<br>List Number | FR Recognition<br>Number | Date of<br>Recognition |
| ISO 11137-2 Third Edition 2013-06-01, Sterilization<br>Of Health Care Products - Radiation - Part 2:<br>Establishing The Sterilization Dose | 041 | 14-409 | 04/04/2016 |
| ISO 11607-1 First Edition 2006-04-15, Packaging<br>For Terminally Sterilized Medical Devices - Part 1:<br>Requirements For Materials, Sterile Barrier Systems<br>And Packaging Systems [Including: Amendment 1<br>(2014)] | 038 | 14-454 | 01/27/2015 |
| ISO 14971 Second Edition 2007-03-01, Medical<br>Devices - Application Of Risk Management To<br>Medical Devices | 043 | 5-40 | 06/27/2016 |
| 60601-1:2012, Medical Electrical Equipment - Part<br>1: General Requirements for Safety and essential<br>performance | 036 | 19-4 | 07-09-2014 |
| 60601-1-2 Edition 4.0 2014-02, Medical Electrical<br>Equipment - Part 1-2: General Requirements For<br>Basic Safety And Essential Performance - Collateral<br>Standard: Electromagnetic Disturbances -<br>Requirements And Tests | 050 | 19-8 | 09/17/2018 |
| 60601-1-8:2006 & A1:2012, Medical Electrical<br>Equipment -- Part 1-8: General Requirements For<br>Basic Safety And Essential Performance --<br>Collateral Standard: General Requirements, Tests<br>And Guidance For Alarm Systems In Medical<br>Electrical Equip. | 031 | 5-76 | 08/06/2013 |
{11}------------------------------------------------
### Conclusion:
The IRRAflow® CNS System is substantially equivalent to the predicate device (K171880). The IRRAflow® CNS System has the same intended use, technological characteristics, and principles of operation as the predicate devices. The minor differences in the indications for use between the IRRAflow® CNS System and the predicate device raise no new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.