The Polaris® Adjustable Pressure Valves are indicated to drain cerebrospinal fluid (CSF) for the management of hydrocephalus.
Device Story
Implantable adjustable pressure valve for hydrocephalus management; shunts CSF from cerebral ventricles to abdominal cavity or right atrium. Features ball-in-cone valve seat for constant intraventricular pressure maintenance. Pressure settings adjustable transcutaneously via manual adjustment kit. Incorporates self-locking rotor-shuttle micro-magnet system; design prevents unintended pressure changes from unidirectional magnetic fields or MRI exposure (up to 3 Tesla). Device used by neurosurgeons in clinical settings. Benefits patients by providing controlled, adjustable CSF drainage to manage intracranial pressure.
Clinical Evidence
Bench testing only. Testing evaluated magnetic field interactions, heating, and artifacts after exposure to a 3-Tesla MRI environment, and verified that such exposure does not alter the functional pressure settings of the valves.
Technological Characteristics
Implantable pressure-adjustable valve; ball-in-cone valve seat mechanism; self-locking rotor-shuttle micro-magnet system for MRI safety. Operates via transcutaneous manual adjustment. MRI compatible up to 3 Tesla.
Indications for Use
Indicated for the management of hydrocephalus in adult and pediatric patients requiring cerebrospinal fluid (CSF) drainage.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Submission Summary (Full Text)
{0}------------------------------------------------
### ATTACHMENT 2. 510(k) SUMMARY
K090342
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### Name, Address, Phone and Fax Number of the Applicant 3.1
Sophysa SA
22 rue Jean Rostand Parc Club Orsay Université 91893 ORSAY Cedex, France Tel: 011-331-69 35 35 00 Fax: 011 331 69 35 36 90
#### Contact Person 3.2
Jean-Christophe Audras 22 rue Jean Rostand Parc Club Orsay Université 91893 ORSAY Cedex, France Tel: 011-331-69 35 35 06
3.3 Date Prepared
February 6, 2009
#### 3.4 Device Name
Polaris® Adjustable Pressure Valve (Models SPV, SPVA, and SPVB), and
Polaris® Adjustable Pressure Valve System (Models SPV-140, SPV-300, and SPV-400)
#### Device Description 3.5
The Polaris® Adjustable Pressure Valve is an implantable device designed for the treatment of hydrocephalus in adult and pediatric patients by shunting, thereby providing continuous, controlled intraventricular pressure and CSF drainage from the cerebral ventricles. Intraventricular pressure is maintained at a constant level by the device's ball- in-cone valve seat design, and the value is pressure-adjustable transcutaneously. Drainage is directed to the abdominal cavity or to the right atrium of the heart.
The basic settings of the Polaris® Valve Models SPV, SPVA, and SPVB are 30, 70, 110, 150 and 200. Adjustments to intermediate pressures are made manually in 40-50 mm Hz0 increments (decrements) in three operating ranges, low pressure (LP =30 mm Hz0), medium pressure (MP = 110 mm H20), and high pressure (HP = 200 mm H20).
AUG 1 1 2009
{1}------------------------------------------------
The basic settings of the Polaris® SPV-140 Valve are 10, 40, 80, 110 and 140 mm H₂0; adjustments to intermediate pressures are made manually in 30-40 mm H20 increments (decrements).
The basic settings of the Polaris® SPV-300 Valve are 50, 100, 150, 220 and 300 mm H₂0; adjustments to intermediate pressures are made manually in 50-80 mm H20 increments (decrements).
The basic settings of the Polaris® SPV-400 Valve are 80, 150, 230, 330 and 400 mm H-0; adjustments to intermediate pressures are made manually in 70-100 mm H20 increments (decrements).
The specific feature of the self-locking rotor-shuttle micro-magnet system of the passive Polaris® SPV Valve is that the adjustment position of each pressure setting cannot be changed by a unidirectional magnetic field. A domestic field or an exposure to MRI attracts the shuttles in the same direction and thus cannot unlock them simultaneously, therefore the rotor cannot be mobilized and the pressure setting remains fixed and constant.
#### 3.6 Indications for Use
The Polaris® Adjustable Pressure Valves are indicated to drain cerebrospinal fluid (CSF) for the management of hydrocephalus.
#### Description of Modifications 3.7
There have been no changes to the device design. The changes involve labeling changes to upgrade the MRI characteristics of the valves for use in an MRI environment from 1.5 Tesla to up to 3 Tesla. Additional changes were made to the IFU to update instructions for the Polaris Adiustment Kit (PAK) handling procedure as well as numerous other minor changes for clarification of to the procedures.
#### Verification & Validation Testing of Changes 3.8
Testing was conducted to evaluate the presence of magnetic filed interactions, heating, and artifacts for the Polaris Adjustable Pressure Valves after exposure to a 3-Tesla MRI environment; and to determine if exposure to a 3-Tesla MRI environment alters the functional aspects of Polaris Adjustable Pressure Valves. The results supported the MRI labeling statements for exposure of the Polaris Adjustable Pressure Valves to a 3-Tesla MRI environment.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-00002
Sophysa SA c/o Jean-Christophe Audras Director, Regulatory Affairs 22 Rue Jean Rostand Parc Club Orsay Université 91893 Orsay Cedex France
# AUG 1 1 2009
Re: K090342
Trade/Device Name: Polaris® Adjustable Pressure Valve (Models SPV, SPVA, and
SPVB), and Polaris® Adjustable Pressure Valve System (Models
SPV-140, SPV-300, and SPV-400)
### Regulation Number: 21 CFR 882.5550
Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: II
Product Code: JXG Dated: June 5, 2009 Received: June 10, 2009
### Dear Mr. Audras:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
### Page 2 - Mr. Jean-Christophe Audras
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Malvina B. Eydelman, M.D
Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K090342
Polaris® Pressure Adjustable Valve System, Models SPV, SPVA, Device Name:_ SPVB, SPV-140, SPV-300, and SPV-400
Indications For Use:
To drain cerebrospinal fluid (CSF) for the management of hydrocephalus.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kristen Bowsher
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number_ K 0 9 0 3 Y Z
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.