Byteflies Kit

K192549 · Byteflies NV · OMC · Jun 22, 2020 · Neurology

Device Facts

Record IDK192549
Device NameByteflies Kit
ApplicantByteflies NV
Product CodeOMC · Neurology
Decision DateJun 22, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2

Indications for Use

The Byteflies Kit is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, and transmit electrical activity of the brain by placing non-invasive electrodes on the head of patients. It acquires, records and transmits two channels of electroencephalogram (EEG) data. The medical use of data acquired by the Byteflies Kit is to be performed under the direction and interpretation of a licensed medical professional. The Byteflies Kit does not provide any diagnostic conclusions about the patient's condition.

Device Story

Wearable medical device for continuous, non-invasive EEG recording; consists of Sensor Dot, Sensor Patch, and Docking Station. Sensor Dot measures up to 2 bipolar EEG channels; stores data for up to 24 hours. Sensor Patch interfaces Sensor Dot with user-supplied disposable electrodes via DIN 42802 connectors. Docking Station charges Sensor Dots and facilitates data transfer to computer via local WiFi. Used in home, clinical, or research environments; configured by healthcare professionals. Device records raw sample-level EEG data; no diagnostic conclusions provided. Healthcare providers download and review data to assist in clinical decision-making. Benefits include continuous monitoring capability outside traditional clinical settings.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-26), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993).

Technological Characteristics

Wearable EEG monitor; 2-channel bipolar acquisition. Components: Sensor Dot (rechargeable Li-Ion battery), Sensor Patch (DIN 42802 connectors), Docking Station (AC/DC power). Connectivity: Local WiFi for data download. Biocompatibility: ISO 10993 compliant. Electrical safety: IEC 60601-1, 60601-1-11, 60601-2-26. EMC: IEC 60601-1-2. Software: Firmware for data acquisition and storage.

Indications for Use

Indicated for patients requiring acquisition and recording of two-channel EEG data in home, healthcare, or clinical research settings. Intended for prescription use under the direction of a licensed medical professional. No specific age or gender contraindications stated.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ June 22, 2020 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Byteflies NV % Benjamin Vandendriessche Chief Medical Officer MedicSense USA Borsbeeksebrug 22, 6th Floor Berchem Antwerpen, BEL 2600 Re: K192549 Trade/Device Name: Byteflies Kit Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMC Dated: March 27, 2020 Received: March 31, 2020 Dear Benjamin Vandendriessche: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K192549 Device Name Byteflies Kit #### Indications for Use (Describe) The Byteflies Kit is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, and transmit electrical activity of the brain by placing non-invasive electrodes on the head of patients. It acquires, records and transmits two channels of electroencephalogram (EEG) data. The medical use of data acquired by the Byteflies Kit is to be performed under the direction of a licensed medical professional. The Bytellies Kit does not provide any diagnostic conclusions about the patient's condition. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------| | <div> <span> 区 Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary #### Contact Details | Applicant | Byteflies NV | |----------------|----------------------------------------| | | Borsbeeksebrug 22, 6th floor | | | 2600 Berchem | | | Belgium | | | Tel: +32 32 94 74 99 | | Contact Person | Benjamin Vandendriessche, PhD | | | Chief Medical Officer | | | benjamin.vandendriessche@byteflies.com | | | +1 (216) 777-9234 | | Date Prepared | March 25, 2020 | #### Device Name | Trade Name | Byteflies Kit | |-------------------|-----------------------------------------------------------| | Common Name | EEG Monitor | | Classification | Reduced-Montage Standard Electroencephalograph (Class II) | | Product Code | OMC | | Regulation Number | 21 CFR 882.1400 | #### Predicate Device NeuroWave Systems Inc, DiscoverEEG System, Model DE-401 (K142834, product code: OMC) #### Description of Device The Byteflies Kit is a wearable medical device for continuous recording of non-invasive physiological signals in healthcare and home settings. The Byteflies Kit is intended to be configured by a trained healthcare professional and consists of 3 main components: - 1. Sensor Dot: a biopotential wearable sensor that measures up to 2 bipolar channels of electroencephalography (EEG). It is powered by a rechargeable battery, can record data for up to 24 hours on a single charge, and has an LED indicator to report the operating {4}------------------------------------------------ status to the user. The sample-level EEG data is continuously stored on the Sensor Dot for later retrieval. - 2. Sensor Patch: the interface between a Sensor Dot and non-invasive biopotential electrodes (not provided). It attaches magnetically to the Sensor Dot and has four DIN 42802 connectors. Two commercial disposable EEG electrodes per Sensor Patch channel are connected to the head of the subject in a reduced EEG montage. The Sensor Dot attached to the Sensor Patch is carried by the patient to continuously measure electrical brain activity. - 3. Docking Station: up to 5 Sensor Dots can connect magnetically to the Docking Station and transfer recorded data to the Docking Station, while charging their batteries. The recorded sample-level EEG signals can then be downloaded from the Docking Station's Management Interface to a computer via a local WiFi network for long-term storage and further review by a healthcare professional. An AC/DC adapter with micro-USB cable supplies power to the Docking Station. ### Intended Use The Byteflies Kit is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, and transmit electrical activity of the brain by placing non-invasive electrodes on the head of patients. It acquires, records and transmits two channels of electroencephalogram (EEG) data. The medical use of data acquired by the Byteflies Kit is to be performed under the direction and interpretation of a licensed medical professional. The Byteflies Kit does not provide any diagnostic conclusions about the patient's condition. ### Substantial Equivalence Comparison The Byteflies Kit is technologically equivalent to other EEG monitoring devices. Both the Byteflies Kit and the predicate device are wearable devices that acquire analog EEG signals in a reduced montage, digitize the sample-level signal, and save them for later interpretation by a healthcare professional. Based on the results from our extensive testing, we believe that our device does not raise new safety or efficacy concerns. The table below summarizes the technological characteristics of the Byteflies Kit in comparison to the predicate device. {5}------------------------------------------------ ### 510(k) Summary # Cyteflies | Features and<br>Characteristics | Byteflies Kit<br>(subject device) | DiscoverEEG System<br>(predicate device) | Comparison | |--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | The Byteflies Kit is<br>intended for prescription<br>use in the home,<br>healthcare facility, or<br>clinical research<br>environment to acquire,<br>record, and transmit<br>physiological signals from<br>patients. It acquires,<br>records, and transmits<br>two channels of<br>electroencephalogram<br>(EEG) data, that are<br>intended to be reviewed<br>by a trained healthcare<br>professional. It only<br>acquires these signals;<br>no diagnostic claims are<br>made about the patient's<br>condition. | The DiscoverEEG System,<br>Model DE-401 is<br>intended to be used for<br>measuring and recording<br>the electrical activity of a<br>subject's brain, obtained<br>by placing noninvasive<br>electrodes on the head.<br>The DiscoverEEG DE-401<br>is indicated for use in<br>acquiring<br>electroencephalographic<br>(EEG) signals in the OR,<br>ICU, ER, clinical settings<br>and at home and for<br>clinical research. The<br>medical use of data<br>acquired by the<br>DiscoverEEG is to be<br>performed under the<br>direction and<br>interpretation of a<br>licensed medical<br>professional. The<br>DiscoverEEG DE-401<br>does not provide any<br>diagnostic conclusion<br>about the subject's<br>condition. | Signal Type:<br>EQUIVALENT<br>Both devices record a<br>reduced EEG montage<br>via noninvasive<br>electrodes placed on the<br>head<br>Environment:<br>EQUIVALENT<br>Both devices are<br>indicated for use in<br>home, clinical, and<br>research environments<br>Medical Use:<br>EQUIVALENT<br>Both devices record raw<br>EEG data and make this<br>available to a healthcare<br>professional for further<br>interpretation; neither<br>device provides any<br>diagnostic claims about<br>the subject's condition | | Modalities | EEG | EEG | EQUIVALENT | | Environment of<br>Use | Home (data acquisition),<br>Healthcare and Clinical<br>Research Facility (data<br>acquisition, device<br>configuration, and data<br>retrieval). | Operating room,<br>intensive care unit,<br>emergency room, clinical<br>settings and at home<br>where EEG monitoring is<br>used. | EQUIVALENT | | Power Source | Li-Ion battery (Sensor<br>Dot) and a certified Class | Battery | EQUIVALENT<br>The components of both<br>devices that are carried | | | II AC/DC power supply<br>(Docking Station) | | by the patient are battery<br>powered. The Docking<br>Station for the subject<br>device is powered via the<br>included certified Class II<br>power supply; this<br>difference does not raise<br>questions of safety and<br>efficacy as the Docking<br>Station does not have<br>any AC wiring and was<br>subjected to all<br>applicable electrical<br>safety testing. | | System<br>Components | Sensor Dot, Sensor<br>Patch, Docking Station | Electrode (Sensor) Array,<br>Acquisition and Memory<br>Modules, Data Viewer<br>Software | EQUIVALENT<br>Both devices consist of<br>equivalent system<br>components for<br>connecting to<br>noninvasive EEG<br>electrodes, storing, and<br>retrieving data. One<br>difference exists: the<br>Byteflies Kit provides<br>charging and data<br>retrieval functionality via<br>the Docking Station,<br>which is separate from<br>the main wearable device<br>(Sensor Dot) to keep the<br>latter's size and weight<br>down. This is a usability<br>decision that does not<br>raise questions of safety<br>and efficacy. | | Sensing<br>Electrodes | User-supplied disposable<br>EEG electrodes | Silver-silver chloride<br>disposable EEG<br>electrodes | EQUIVALENT<br>Both devices use<br>disposable noninvasive<br>biopotential electrodes.<br>The Byteflies Kit does not<br>include them as part of<br>the device but uses FDA | | | | | cleared DIN 42802 EEG<br>electrodes as accessories<br>which provides more<br>flexibility to the user. This<br>is a usability decision that<br>does not raise questions<br>on safety and efficacy. | | Screen Display<br>Details | Not applicable | Data Viewer Software | EQUIVALENT<br>The Byteflies Kit contains<br>no data viewer software<br>although data files can<br>be imported into any<br>signal review software<br>that can read and<br>visualize EEG data, as is<br>the case for the<br>predicate device. | | Stored EEG data<br>available | Sample-level EEG as text<br>file which can be<br>accessed via the Docking<br>Station | Yes – SD card | EQUIVALENT<br>Sample-level EEG data<br>can be downloaded to a<br>computer for long-term<br>storage and further<br>analysis by a trained<br>healthcare professional. | | EEG Channels | Up to 2 EEG channels | Up to 4 EEG channels | EQUIVALENT | | Contains patient<br>isolation | Sensor Dot is battery<br>powered, thus no<br>connection between<br>patient and mains. The<br>Docking Station is<br>powered by a certified<br>Class II AC/DC medical<br>power supply and<br>contains now AC wiring. | Battery powered, thus no<br>connection between<br>patient and mains. | EQUIVALENT<br>The components of both<br>devices that are carried<br>by the patient are battery<br>powered. The Docking<br>Station for the subject<br>device is powered via the<br>included certified Class II<br>power supply; this<br>difference does not raise<br>questions of safety and<br>efficacy as the Docking<br>Station does not have<br>any AC wiring and was<br>subjected to all | | | | | applicable electrical<br>safety testing. | | Display Interface | Yes, LEDs to ensure<br>proper system<br>connection and<br>operation. | Yes, LEDs to ensure<br>proper system<br>connection and<br>operation. | EQUIVALENT | | Biocompatibility | All external parts were<br>tested according to ISO<br>10993 (Cytotoxicity - ISO<br>10993-5, Sensitization -<br>ISO 10993-10, and Skin<br>Irritation - ISO 10993-10) | Electrode array was<br>tested according to ISO<br>10993 (Cytotoxicity - ISO<br>10993-5, Sensitization -<br>ISO 10993-10, and Skin<br>Irritation - ISO 10993-<br>10); test status for other<br>components unknown | EQUIVALENT<br>All external parts of the<br>Byteflies Kit meet the<br>requirements of ISO<br>10993. | | Electrical Safety<br>and<br>Electromagnetic<br>Compatibility | IEC60601-1, IEC60601-1-<br>2, IEC60601-1-11,<br>IEC60601-1-8, IEC60601-<br>1-6, IEC60601-2-26 | IEC60601-1, IEC60601-1-<br>11, IEC60601-2-26,<br>IEC60601-1-2 | EQUIVALENT<br>The Byteflies Kit was<br>subjected to the same<br>electrical safety and<br>electromagnetic<br>compatibility testing as<br>the predicate device, and<br>was additionally verified<br>against 60601-1-6 and<br>60601-1-8. | | 510(k) # | K192549 | K142834 | NA | | Regulatory | Product Code OMC 21<br>CFR 882.1400 | Product Code OMC and<br>OLT 21 CFR 882.1400 | EQUIVALENT<br>Both devices use the<br>same product code, but<br>the Byteflies Kit does not<br>include any EEG viewing<br>software; this difference<br>does not raise questions<br>on safety and efficacy as<br>both devices allow 3rd<br>party signal viewers to be<br>used. | {6}------------------------------------------------ # Cyteflies {7}------------------------------------------------ # Cyteflies {8}------------------------------------------------ # Cyteflies ## Non-clinical Testing {9}------------------------------------------------ Design and verification tests were performed on the Byteflies Kit as a result of the risk analysis and product requirements. Independent UL-certified laboratory testing demonstrated that the Byteflies Kit meets the requirements of IEC 60601-1:2005 and IEC 60601-1-11:2015 for electrical safety, and IEC 60601-1-2:2014 for electromagnetic compatibility. The Byteflies Kit complies with the particular requirements of IEC 60601-2-26:2012 for the safety of electroencephalographs. Finally, it meets all applicable requirements of ISO 10993:2018, IEC 60601-1-6:2010 and IEC 60601-1-8:2006. ### Clinical Tests The Byteflies Kit is an electroencephalographic device comprised of hardware and software that has been bench tested to assess safety and effectiveness, and to establish substantial equivalence with the predicate device. We believe further clinical data is not required to demonstrate performance of the Byteflies Kit for the indications for use subject to this submission. #### Conclusion Based on the results of the performance testing and substantial equivalence comparison, the Byteflies Kit has the same intended use and principles of operations as the predicate device. Any differences in technological characteristics do not raise questions about safety and effectiveness. The presented information is sufficient to determine that the Byteflies Kit is substantially equivalent to the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...