X8 System - Sleep Profiler (SP40), X8 System - Sleep Profiler PSG2 (SP29), X8 System - Stat X8 (XS29)
K152040 · Advanced Brain Monitoring, Inc. · OMC · Oct 30, 2015 · Neurology
Device Facts
Record ID
K152040
Device Name
X8 System - Sleep Profiler (SP40), X8 System - Sleep Profiler PSG2 (SP29), X8 System - Stat X8 (XS29)
Applicant
Advanced Brain Monitoring, Inc.
Product Code
OMC · Neurology
Decision Date
Oct 30, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The X8 System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, transmit, and display physiological signals from adult patients. All X8 models (SP40, SP29, and XS29) acquire, record, transmit, and/or display electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals. Model SP29 additionally includes a nasal pressure transducer and cannula (for airflow), thoracic and abdomen respiratory effort, and pulse rate and oxyhemoglobin saturation from the finger. The X8 system only acquires and displays physiological signals; no claims are being made for analysis of the acquired signals with respect to the accuracy, precision, and reliability.
Device Story
Battery-powered headset worn by patient; acquires physiological signals (EEG, EOG, ECG, EMG, accelerometer, acoustical, photoplethysmographic); SP29 model adds nasal pressure, respiratory effort, and pulse oximetry. Device records data to internal memory/SD card; transmits via Bluetooth or USB to PC or cloud-based server. Used in home, clinic, or research environments; self-applied by patient or applied by technician. Device Manager software facilitates data transfer and formatting; X-Series Basic-Utility Software provides GUI for signal presentation, impedance measurement, and marker placement. Output is raw physiological signal data in EDF format for review by healthcare providers using third-party software. Device provides visual/auditory alerts for signal quality. Benefits include remote or clinical physiological monitoring without requiring device removal during patient movement.
Clinical Evidence
Prospective study (n=11) using Model SP29. Primary endpoints: signal quality compared to Compumedics Somte (K072201) and user ability to self-apply. Results: <10% of airflow events and <20% of respiratory effort events were inferior to predicate (actual: 0% airflow, 4.0% thorax, 1.3% abdomen). 91% of subjects acquired at least one full night of data. >90% valid data obtained for oximetry, nasal pressure, and respiratory effort. 90% of subjects reported audio alerts did not significantly disrupt sleep.
Technological Characteristics
Battery-powered (3.7V Li-ION) headset; Black ABS enclosure. Sensors: EEG/EOG/ECG/EMG electrodes, nasal cannula, RIP belts, Bluetooth oximeter. Connectivity: Bluetooth, USB, SD card. Software: Device Manager and X-Series Basic-Utility. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-26, IEC 62133, ISO 14971. Biocompatibility identical to predicate X4. Cleaning: alcohol-based sanitizer/isopropyl alcohol for sensors/enclosure, dish soap for straps.
Indications for Use
Indicated for adult patients in home, healthcare, or clinical research settings for the acquisition, recording, transmission, and display of physiological signals (EEG, EOG, ECG, EMG, respiratory, and oximetry).
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 30, 2015
Advanced Brain Monitoring, Inc. % Adrienne Lenz Member Pathway Regulatory Consulting, LLC W324S3649 County Road E Dousman, Wisconsin 53118
Re: K152040
Trade/Device Name: X8 System - Sleep Profiler (SP40), X8 System - Sleep Profiler PSG2 (SP29), X8 System - Stat X8 (XS29) Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMC, OLV Dated: July 17, 2015 Received: July 22, 2015
Dear Adrienne Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152040
#### Device Name
X8 System - Sleep Profiler (SP40), X8 System - Sleep Profiler PSG2 (SP29), X8 System - Stat X8 (XS29)
#### Indications for Use (Describe)
The X8 System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, transmit, and display physiological signals from adult patients. All X8 models (SP40, SP29, and XS29) acquire, record, transmit, and/or display electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals. Model SP29 additionally includes a nasal pressure transducer and cannula (for airflow), thoracic and abdomen respiratory effort, and pulse rate and oxyhemoglobin saturation from the finger. The X8 system only acquires and displays physiological signals; no claims are being made for analysis of the acquired signals with respect to the accuracy, precision, and reliability.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
DATE: September 24, 2015
### SUBMITTER:
Advanced Brain Monitoring 2237 Faraday Avenue, Suite 100 Carlsbad, CA 92008 T 760.720.0099 F 760.720.3337
PRIMARY CONTACT PERSON:
Adrienne Lenz, RAC Member Pathway Regulatory Consulting, LLC T 262-290-0023
### SECONDARY CONTACT PERSON:
Dan Levendowski President and Co-founder Advanced Brain Monitoring, Inc.
#### DEVICE:
TRADE NAME: X8 System (Sleep Profiler, Sleep Profiler PSG2, X8 Stat)
COMMON/USUAL NAME: X8
CLASSIFICATION NAMES: 882.1400 Electroencephalograph
REVIEW PANEL: Neurology
PRODUCT CODE: OMC
PREDICATE DEVICE(S):
K130013 X4 System
{4}------------------------------------------------
## DEVICE DESCRIPTION:
The X8 System is indicated for acquiring, recording/storing, transmitting, and displaying physiological data in patients. It can be used by patients in the home, healthcare facility, or clinical research environment. Patients can move within their home or healthcare environment without having to remove the device (e.g. walk to the restroom).
The X8 System is comprised of the X8 device which is worn on the patient's head and body, accessories, the Device Manager software, and the X-Series Basic-Utility Software. The study records are saved to the PC in EDF format and are available for analysis by third party software applications, such as Persyst Reveal (K011397).
The X8 System combines hardware, firmware, and software to acquire physiological signals. It acquires physiological data through a battery powered headset worn by the patient and provides a flexible platform for applying sensors and acquiring signals from multiple locations on the head or body, transmitting and recording the signals, and providing visual and auditory indications to ensure high quality data are obtained.
Model SP40 Sleep Profiler is applied by the patient to acquire and record electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals during sleep. This model utilizes the X8 Sleep Profiler Strip.
Model SP29 Sleep Profiler PSG2 is applied by the patient to acquire and record electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals during sleep. This model additionally includes a nasal pressure transducer and cannula (for airflow), thoracic and abdomen respiratory effort, and pulse rate and oxyhemoglobin saturation from the finger. This model utilizes the X8 Sleep Profiler Strip.
Model XS29 X8 Stat is applied by a technician to acquire, record, transmit, and/or display electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals acquired during non-sleep conditions. This model utilizes either the X8 Midline or the X8 Referential strip.
The Device Manager software application provides a means to communicate with the X8 Device, transfer study data, and format a device. The software transfers data saved in the memory of the X8 device using either a PC application or a web-based data entry application operating in a cloud server environment.
{5}------------------------------------------------
The X-Series Basic-Utility Software acquires, presents, and stores physiological signals from the X8 Device. The software has a modular architecture that allows the users to interact using either the Graphical User Interface (GUI) provided with the installation, or programmatically via a Software Development Kit. Additional functionalities provided by X-Series Basic-Utility Software include impedance measurements, custom markers, and interface with the Bluetooth Receiving Dongle.
## INTENDED USE:
The X8 System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, transmit, and display physiological signals from adult patients. All X8 models (SP40, SP29, and XS29) acquire, record, transmit, and/or display electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals. Model SP29 additionally includes a nasal pressure transducer and cannula (for airflow), thoracic and abdomen respiratory effort, and pulse rate and oxyhemoglobin saturation from the finger. The X8 system only acquires and displays physiological signals; no claims are being made for analysis of the acquired signals with respect to the accuracy, precision, and reliability.
## TECHNOLOGY:
The X8 System has similar indications for use and uses the same fundamental technology as the X4 System for most features including the electrophysiological (EEG, EOG, ECG, EMG), wireless acquisition, and actigraphy. Like the X4 System, the X8 system acquires, records and transmits EEG, EOG, ECG, EMG, accelerometer, acoustical and photoplethysmographic signals. Additional respiratory signals acquired by the X8 system use similar technology as reference device ARES Model 610. The technologies used in the X8 System are used in the same manner as the predicate and reference products and do not raise new questions of safety and effectiveness. A comparison of key specifications is presented in the following table:
| Specification | X8 System (New) | Predicate Device X4 System (K130013) | Reference Device ARES Model 610 (K112514) |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient Population | Adults | Adults | Adults |
| Anatomical Sites | Forehead, Head, Chest, Abdomen, Chin and Finger | Forehead, Chest, Chin | Forehead, Chest |
| Environment of Use | Home (data acquisition)<br>Healthcare facility (data acquisition, analysis and reporting)<br>Clinical Research Environment | Home (data acquisition)<br>Healthcare facility (data acquisition, analysis and reporting)<br>Clinical Research Environment | Home (data acquisition)<br>Healthcare facility (data acquisition, analysis and reporting) |
| User Control | ON/OFF | ON/OFF | ON/OFF |
| Specification | X8 System (New) | Predicate Device<br>X4 System (K130013) | Reference Device<br>ARES Model 610 (K112514) |
| Visual Indicator | Green and yellow LED | Green and yellow LED | Green LED |
| Audio Indicator | Speaker. Voice messages alert user to problems during recording. | Speaker. Voice messages alert user to problems during recording. | Speaker. Voice messages alert user to problems during recording. |
| EEG Electrodes | Vermed VersaTrode Sensor 40-4305 (K781430) | Kendall Meditrace Mini 230 (K960968)<br>Kendall Meditrace Mini 533 (K945479) | Kendall Meditrace Mini 130 (K864722) |
| EOG Electrodes | Vermed VersaTrode Sensor 40-4305 (K781430) | Kendall Meditrace Mini 130 (K864722) | Kendall Meditrace Mini 130 (K864722) |
| ECG Electrodes | Vermed VersaTrode Sensor 40-4305 (K781430) | Kendall Meditrace Mini 533 (K945479) | None |
| EMG Electrodes | Vermed NeuroPlus Sensors A10041-60 (K010638) | Kendall Meditrace Mini 130 (K864722) | None |
| Linked Mastoid<br>Sensors | MBS (3BF3) Disposable Ag/Cl sensors with adhesive (K842514) | MBS (3BF3) Disposable Ag/Cl sensors with adhesive (K842514) | None |
| 2-pin Leadwire | 2-pin leadwire with touch-proof connector | 2-pin leadwire with touch-proof connector | 2-pin leadwire with touch-proof connector |
| 3-pin Leadwire | 3-pin leadwire with touch-proof connector | 3-pin leadwire with touch-proof connector | None |
| Respiratory<br>Accessory | Nasal Cannula and nasal pressure sensor<br>Chest and abdomen effort belts | None | Nasal Cannula and nasal pressure sensor<br>Chest effort belt |
| Signals Acquired | Forehead/head EEG Infra-red (IR) optical signal Microphone 3-D actigraphy Optional channel (ECG/EEG/EOG/EMG) Nasal Pressure & cannula Respiratory Effort | Forehead/head EEG Infra-red (IR) optical signal Microphone 3-D actigraphy Optional channel (ECG/EEG/EOG/EMG) | Forehead EEG Red and infra-red (IR) optical signals Microphone 3-D actigraphy Derived EOG and EMG Nasal Pressure & cannula Respiratory Effort |
| Power Supply | 1 x 600 mAH 3.7V Li-ION battery | 1 x 600 mAH 3.7V Li-ION battery | 2 x 250 mAH 3.7V Li-Ion batteries |
| Specification | X8 System (New) | Predicate Device<br>X4 System (K130013) | Reference Device<br>ARES Model 610 (K112514) |
| Battery<br>Charging | Patients - external battery<br>pack<br><br>Technicians - Via USB cable<br>connected to USB port or USB<br>wall charger | Via USB cable connected to<br>USB port or USB wall charger | Via USB cable connected to<br>USB port or USB wall charger |
| Typical Charging<br>Time | 0.5 - 3.0 hours | 0.5 - 3.0 hours | 0.5 - 2.5 hours |
| Acquisition<br>modes | Record or Monitor | Record or Monitor | Record only |
| Operating Time | Hours of Use:<br>0-4 days after charging:<br>Model SP40 Record: 16.0 -<br>18.5<br>Model XS29 Monitor: 10.5 -<br>12.5<br>Model SP29 Record: 11.5 -13.5<br><br>5-10 days after charging:<br>Model SP40 Record: 14.0 -<br>16.5<br>Model XS29 Monitor: 9.5 -<br>12.0<br>Model SP29 Record: 10.0 -<br>13.0 | Hours of Use:<br>0-4 days after charging:<br>Record: 13.0 - 15.5<br>Monitor: 6.5 - 7.8<br><br>5-10 days after charging:<br>Record: 11.5 - 14.8<br>Monitor: 6.0 - 7.3 | Hours of Use:<br>0-4 days after charging:<br>Record: 19.0 hours<br><br>5-10 days after charging:<br>Record: 17.0 hours |
| Data Storage | 8 GB Micro-SDHC memory card<br>or greater capacity | 2 GB Micro-SD memory card | 2 GB Micro-SD memory card |
| File size per 8 hr<br>recording | Standard mode - 72 MB<br>PSG2 Mode - 134 MB | 72 MB | 31.2 MB |
| Dimensions | 2.1" long, 1.5"wide, 0.75" deep | 2.1" long, 1.5"wide, 0.75" deep | 2.5" long x 2" wide x 1" deep |
| Weight | 2.5 ounces with battery | 2.5 ounces with battery | 4 ounces with batteries |
| Cleaning<br>enclosures and<br>EEG Strip | Cleaned and disinfected by<br>rubbing with alcohol-based<br>hand sanitizer and isopropyl<br>alcohol. | Cleaned and disinfected by<br>rubbing with alcohol-based<br>hand sanitizer and isopropyl<br>alcohol. | Cleaned and disinfected by<br>rubbing with alcohol-based<br>hand sanitizer and isopropyl<br>alcohol. |
| Cleaning<br>enclosure strap | Cleaned and disinfected by<br>washing with dish soap. | Cleaned and disinfected by<br>washing with dish soap. | Single patient use. |
| Specification | X8 System (New) | Predicate Device<br>X4 System (K130013) | Reference Device<br>ARES Model 610 (K112514) |
| Nasal Pressure<br>Transducer | Pressure transducer<br>incorporated into airflow<br>adapter,<br>Affixes to USB connector<br>Size: 3.5 (l) x 2.5 (w) x 1.5 (d) cm<br>Weight - 5.5 grams | Not available | Pressure transducer<br>incorporated into main<br>enclosure |
| Oximeter | Bluetooth (BT) used to obtain<br>pulse and SpO2 acquired with<br>wrist oximeter.<br><br>Nonin WristOx2 Wireless Pulse<br>Oximeter (K102350) | Not available | Pulse and oximeter<br>incorporated into main<br>enclosure |
| Respiratory<br>effort belts | Uses 3rd party respiratory<br>induced plethysmography (RIP)<br>thorax and abdomen effort<br>belts<br><br>Philips/Pro-Tech ezRIP<br>Respiratory Effort Sensors<br>(K913395), and Ambu<br>SleepMate RIPmate<br>Respiratory Effort Sensor<br>(K903300) | Not available | Single piezzo chest belt |
| Data transfer<br>from SD card | Native USB | Native USB | Native USB |
| USB data<br>transfer rate | > 250 MB per minute | > 250 MB per minute | > 250 MB per minute |
| Wireless data<br>transfer | Bluetooth | Bluetooth | N/A |
| Maximum<br>Bluetooth<br>wireless transfer<br>distance and<br>rate | Transfer distance 10 meters<br>line of sight, maximum transfer<br>rate 3 Mbaud | Transfer distance 10 meters<br>line of sight, maximum transfer<br>rate 3 Mbaud | N/A |
| Compatibility | Personal computer with<br>2.4GHz (Pentium 4) Processor<br>or better, 2 GB RAM or higher<br>(or equivalent) with Windows<br>7, 8, or Mac iOS Operating<br>System | Personal computer with<br>Pentium 4, 2 GB RAM or higher<br>processor (or equivalent) with<br>Windows 7 Operating System | Personal computer with<br>Pentium 4, 1 GB RAM or<br>higher processor (or<br>equivalent) with Windows XP<br>Operating System |
| Estimated file<br>size per minute | Standard Mode: 150 KB/Min<br>PSG2 Mode: 280 KB/Min | 150 KB/Min | 67 KB/Min |
| File format type | European Data Format (EDF) | European Data Format (EDF) | Proprietary |
| Specification | X8 System (New) | Predicate Device<br>X4 System (K130013) | Reference Device<br>ARES Model 610 (K112514) |
| Presents raw<br>signals during<br>acquisition in<br>monitoring<br>mode | Yes | Yes | Yes |
| Presents<br>previously<br>acquired signals | No | Yes | Yes |
| Specification | X8 System (New) | Predicate Device<br>X4 System (K130013) | Reference Device<br>ARES Model 610 (K112514) |
| Desktop<br>interface<br>minimum<br>computer<br>requirements | 1. Processor : 2.4 GHz<br>2. Operating System:<br>Windows 7 or 8<br>3. RAM: 2GB<br>4. USB port: 1 | 1. Processor : 2.4 GHz<br>2. Operating System:<br>Windows 7<br>3. RAM: 2GB<br>4. USB port: 1 | 1. Processor : 2.4 GHz<br>2. Operating System:<br>Windows XP<br>3. RAM: 1GB<br>4. USB port: 1 |
| Web interface<br>minimum<br>computer<br>requirements | 1. Processor : Minimum 2.4 GHz<br>2. Operating System:<br>Windows 7 or 8, Mac iOS<br>3. Java version 6 or greater<br>4. RAM: 1GB<br>5. USB port: 1<br>6. Internet connection:<br>constant<br>7. Web browser: Internet<br>Explorer, Firefox, Opera,<br>Chrome, or Safari | 1. Processor : Minimum 2.4 GHz<br>2. Operating System:<br>Windows XP or 7<br>3. Java version 6 or greater<br>4. RAM: 1GB<br>5. USB port: 1<br>6. Internet connection:<br>constant<br>7. Web browser: Internet<br>Explorer, Firefox, Opera, or<br>Chrome | 1. Processor : Minimum 2.4 GHz<br>2. Operating System:<br>Windows XP or 7<br>3. JAVA version 6 or greater<br>4. RAM: 1GB<br>5. USB port: 1<br>6. Internet connection:<br>constant<br>7. Web browser: Internet<br>Explorer, Firefox, or<br>Chrome |
| | | | |
| Web interface<br>server<br>requirements. | Virtual Server:<br>1. Processor: > 2 GHz<br>2. Operating system: Win<br>Server 2008.<br>3. RAM: > 2GB<br>4. Certificates: Signed SSL<br>5. .NET framework:<br>version 2.0 - web server<br>version 3.5 - processing<br>server<br>or Win Server 2008<br>6. Database: SQL server<br>2005 | Virtual Server:<br>1. Processor: > 2 GHz<br>2. Operating system: Win<br>Server 2008.<br>3. RAM: > 2GB<br>4. Certificates: Signed SSL<br>5. .NET framework:<br>version 2.0 - web server<br>version 3.5 - processing<br>server<br>or Win Server 2008<br>6. Database: SQL server 2005 | Virtual server:<br>1. Processor: > 2 GHz<br>2. Operating system: Win<br>Server 2008.<br>3. RAM: > 2GB<br>4. Certificates: Signed SSL<br>5. .NET framework:<br>version 2.0 - web server<br>version 3.5 - processing<br>server<br>or Win Server 2008<br>6. Database: SQL server<br>2005 |
| | Physical Server:<br>1. Processor: > 2 GHz<br>2. Operating system: Win<br>Server 2008 Enterprise<br>edition (for virtual<br>servers).<br>3. RAM: > 2GB | Physical Server:<br>1. Processor: > 2 GHz<br>2. Operating system: Win<br>Server 2008 Enterprise<br>edition (for virtual servers)<br>3. RAM: > 2GB | Physical server:<br>1. Processor: > 2 GHz<br>2. Operating system: Win<br>Server 2008 Enterprise<br>edition (for virtual<br>servers).<br>3. RAM: > 2GB |
| Specification | X8 System (New) | Predicate Device<br>X4 System (K130013) | Reference Device<br>ARES Model 610 (K112514) |
| Computer/portal<br>Security | 1. The person installing the<br>Java applet must have<br>administrator privileges.<br>2. User must have rights to<br>the account and group to<br>access data via the portal.<br>3. The firewall is not<br>configured to prevent<br>input or output<br>communication with<br>access to the server via<br>ports 22 and 30-39. | 1. The person installing the<br>Java applet must have<br>administrator privileges.<br>2. User must have rights to the<br>account and group to<br>access data via the portal.<br>3. The firewall is not<br>configured to prevent input<br>or output communication<br>with access to the server<br>via ports 22 and 30-39. | 1. The person installing the<br>Java applet must have<br>administrator privileges.<br>2. User must have rights to<br>the account and group to<br>access data via the<br>portal.<br>3. The firewall is not<br>configured to prevent<br>input or output<br>communication with<br>access to the server via<br>ports 22 and 30-39. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
# DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
# SUMMARY OF NON-CLINICAL TESTS:
Support for the substantial equivalence of the X8 System was provided as a result of risk management and testing which included comparison of signals to the predicate device, electrical safety, electromagnetic compatibility and software tests. This testing includes conformity to FDA recognized consensus standards and voluntary standards as follows:
| Standard Number | Standard Title |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1:2005 (3rd Edition) | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance |
| IEC 60601-1-2:2014 | Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests |
| IEC 60601-1-11:2010 | Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment |
| IEC 60601-2-26:2012 | Particular requirements for the basic safety and essential performance of electroencephalographs |
| IEC 62133: 2012 | Secondary Cells And Batteries Containing Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications |
| ISO 14971:2007 | Medical devices - Application of risk management to medical devices |
{12}------------------------------------------------
Additional analysis, verification and validation testing confirmed:
- Instructions for disinfecting the reusable parts of the X8 System and the materials used ● in the reusable portions of the X8 System (i.e. Sensor strip forehead, sensor strip head, and enclosure strap) are identical to those used in the previously cleared X4 System (K120047) and X-Series System (K131383). The remaining reusable portions (i.e. Respiratory Effort Belts and Wireless finger pulse oximeter) of the X8 System are cleared on their own by the supplier. The size and shape of the device enclosure of the X8 System is identical to that of the previously cleared X4 System and the material used is the same Black ABS. The Airflow Adapter enclosure is made of the same Black ABS as the device enclosure. Advanced Brain Monitoring evaluated the differences between the X8 System and the X-Series and X4 Systems and concluded that previously performed cleaning validation was also applicable to the X8 System.
- All materials of the X8 System are identical to materials tested for biocompatibility in the predicate X4 System and reference device X-Series System (K131383).
- 0 All features of the X8 System were compliant with the system and software level requirements.
- Signals acquired with the X8 System provide equivalent information as compared to the predicate device that would allow a physician to interpret the signals.
# SUMMARY OF CLINICAL TESTS:
Advanced Brain Monitoring has conducted prospective studies to: A) compare equivalence of signals obtained with X8 System airflow and respiratory effort signals with an FDA cleared device, Compumedics Somte (K072201), and B) demonstrate that high quality signals can be obtained when the X8 System is self-applied with the user instructions. X8 System Model SP29 was used for both portions of the study.
The following endpoints were successfully achieved:
- No more than 10% of the breathing events recorded with the X8-PSG2 airflow signal were inferior to predicate signal. No events recorded with the X8-PSG2 airflow signal (0%) were found inferior to the predicate signal.
- o No more than 20% of the breathing events recorded with the X8-PSG2 respiratory effort will be inferior to the predicate signal. The X8 thorax and abdomen belts were inferior to the predicate in only 4.0% (9/225) and 1.3% (3/232) of the events, respectively.
- At least 80% of subjects will be able to acquire at least one night of data (i.e., the entire period they were in bed). The results demonstrated that 91% (10 of 11 subjects) were able to acquire at least one night of data for the entire night.
- . At least 70% of each night of recording time will be valid across the oximetry, nasal pressure, and effort belt signals. The percentage of good data obtained for oximetry, nasal pressure (airflow) and respiratory effort (thorax and abdomen) exceeded 90% on each night. While not identified as a primary endpoint for the assessment of cardio-
{13}------------------------------------------------
respiratory signal quality, high EEG quality was also obtained in over 90% of the recording time on each night.
- • At least 70% of subjects did not report X8-PSG2 audio alerts (for signal quality) substantially affected (i.e., strongly agreed) their ability to stay asleep. The results demonstrated that 90% (9 of 10 subjects) did not "strongly agree" that the PSG2 made it difficult for them to stay asleep.
## CONCLUSION:
Advanced Brain Monitoring considers the X8 System to be substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.