K161027 · Cadwell Industries, Inc. · OMA · Nov 8, 2016 · Neurology
Device Facts
Record ID
K161027
Device Name
Cadwell AmpliScan
Applicant
Cadwell Industries, Inc.
Product Code
OMA · Neurology
Decision Date
Nov 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Cadwell AmpliScan is a software-only device indicated for use with electroencephalographic (EEG) data from Cadwell Arc application software. Cadwell AmpliScan is distributed solely for use with Cadwell Arc software. The Cadwell AmpliScan device is for prescription use only by qualified medical practitioners, trained in Electroencephalography, who will exercise professional judgement when using the information. This device does not provide any diagnostic conclusion about the patient's condition to the user. Cadwell AmpliScan uses electroencephic (EEG) data to calculate and display a quantitative aEEG measure. This quantitative measure should always be interpreted by the user in conjunction with review of the original EEG waveforms. The aEEG quantitative measure of Cadwell AmpliScan is intended to monitor the state of the brain.
Device Story
Software-only device; operates within Cadwell Arc application software. Inputs: stored EEG data from Cadwell Arc. Transformation: applies Maynard and Prior Cerebral Function Monitor (CFM) algorithm; includes input filtering, signal compression, and low-pass filtering. Output: quantitative aEEG measure displayed to user. Used in clinical settings by neurologists/trained practitioners. Provider interprets aEEG alongside original EEG waveforms to monitor brain state. Benefits: provides trended brain activity visualization to assist clinical monitoring.
Clinical Evidence
No clinical data. Bench testing only. Software verification and validation conducted per FDA guidance for moderate level of concern software. Equivalence demonstrated by comparing outputs of subject device and predicate using identical input data.
Technological Characteristics
Software-only; implements Maynard and Prior CFM algorithm. Features: input filtering, 0.5s time constant low-pass filter, linear (0-10 uV) and log10 (10-100 uV) signal compression. Integrated into Cadwell Arc software environment.
Indications for Use
Indicated for use by qualified medical practitioners trained in Electroencephalography to monitor brain state via quantitative aEEG measures derived from EEG data. Prescription use only. No specific age or gender restrictions. Not for diagnostic conclusions.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
Persyst 12 EEG Review and Analysis Software (K132306)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a wing-like shape above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 8, 2016
Cadwell Industries, Inc. John Cadwell Jr. Electrical Engineer 909 N. Kellog St. Kennewick, Washington 99336
Re: K161027
Trade/Device Name: Cadwell AmpliScan Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMA Dated: August 12, 2016 Received: October 3, 2016
Dear Mr. Cadwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
## Michael J. Hoffmann -A
for
Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161027
Device Name Cadwell AmpliScan
### Indications for Use (Describe)
1) Cadwell AmpliScan is a software-only device indicated for use with electroencephalographic (EEG) data from Cadwell Arc application software. Cadwell AmpliScan is distributed solely for use with Cadwell Arc software.
2) The Cadwell AmpliScan device is for prescription use only by qualified medical practitioners, trained in Electroencephalography, who will exercise professional judgement when using the information.
3) This device does not provide any diagnostic conclusion about the patient's condition to the user.
4) Cadwell AmpliScan uses electroencephic (EEG) data to calculate and display a quantitative aEEG measure. This quantitative measure should always be interpreted by the user in conjunction with review of the original EEG waveforms. The aEEG quantitative measure of Cadwell AmpliScan is intended to monitor the state of the brain.
Type of Use (Select one or both, as applicable)
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------------------|
| <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510k Summary
### 1) Submitter:
Cadwell Industries, Inc. 909 N. Kellogg Street Kennewick, Washington 99336
Phone: 509-735-6481
Contact Person: John A Cadwell Jr. Date Prepared: August 12, 2016
### 2) Device:
Trade Name of Device: Cadwell AmpliScan Common or Usual Name: Cadwell AmpliScan aEEG software Classification Name: Electroencephalograph (21 CFR 882.1400) Regulatory Classification: Class II Product Code: OMA
### 3) Predicate Device:
Persyst 12 EEG Review and Analysis Software (K132306) is a predicate. This predicate has not been subject to a design-related recall.
Olympic CFM6000 (K031149) is a reference predicate. This reference predicate has not been subject to a design-related recall.
### 4) Device Description:
Cadwell AmpliScan is a software-only device distributed solely for use with the application software commonly known and marketed as Cadwell Arc software. Cadwell Ampliscan software is installed with installation of Arc software, and does not require installation or removal separate from the Arc application.
The Cadwell AmpliScan software-only device applies the Amplitude-Integrated EEG (aEEG) algorithm referred to as the Cerebral Function Monitor (CFM) to stored data within the Cadwell Arc software, and stores and displays the results.
{4}------------------------------------------------
# 1 = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
## 510k Summary
### 5) Indications for Use:
- 1) Cadwell AmpliScan is a software-only device indicated for use with electroencephalographic (EEG) data from Cadwell Arc application software. Cadwell AmpliScan is distributed solely for use with Cadwell Arc software.
- 2) The Cadwell AmpliScan device is for prescription use only by qualified medical practitioners, trained in Electroencephalography, who will exercise professional judgement when using the information.
- 3) This device does not provide any diagnostic conclusion about the patient's condition to the user.
- Cadwell AmpliScan uses electroencephalographic (EEG) data to calculate and display a 4) quantitative aEEG measure. This quantitative measure should always be interpreted by the user in conjunction with review of the original EEG waveforms. The aEEG quantitative measure of Cadwell AmpliScan is intended to monitor the state of the brain.
{5}------------------------------------------------
# VELL
### 6) Comparison of Technological Characteristics with the Predicate Device:
A comparison is made between Cadwell AmpliScan and the aEEG functionality of the predicate.
Cadwell AmpliScan and the predicate device both utilize the Maynard and Prior "Cerebral Function Monitor (CFM)" implementation to perform amplitude-integrated EEG analysis on stored patient data and display the results. The CFM requires input filtering, signal compression, low pass filtering, and specific display characteristics. Given that the Cadwell AmpliScan and the predicate implement the required analysis on input data in the same way, the respective outputs of Cadwell AmpliScan and the predicate are equivalent.
The following table illustrates technological characteristics of Cadwell AmpliScan and the predicate.
| | Persyst 12 EEG<br>Review and Analysis<br>Software (K132306) | Cadwell AmpliScan<br>aEEG |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Device Type | Software-only | SAME |
| aEEG | YES | SAME |
| Product Codes | OMA, ORT, OLT | OMA |
| aEEG Algorithm | Cerebral Function Monitor<br>Maynard and Prior | SAME |
| Input Filter | Per CFM, Maynard and Prior | SAME |
| Signal Compression | Linear 0-10 uV, Log10 10-<br>100uV | SAME |
| Low Pass Filter<br>Time Constant | 0.5 S | SAME |
| Vertical Display | Linear 0-10 uV, Log10 10-<br>100uV | SAME |
| Horizontal Display | Vertical line display from min<br>to max within each epoch | SAME |
| aEEG Indications<br>for Use | "The aEEG functionality...is<br>intended to monitor the state<br>of the brain" | SAME |
| Supported Input<br>Data | "As recorded by OEM EEG<br>equipment being used<br>to acquire the EEG signal"<br>including: Cadwell, Natus,<br>Compumedics, Nihon-<br>Kohden, and others. | CADWELL ARC EEG Data |
{6}------------------------------------------------
## 510k Summary
The following technological differences exist between the subject and predicate devices:
- . The predicate device can utilize stored EEG data from a wide range of sources, including Cadwell, Natus, Compumedics, Nihon-Kohden and others. The Cadwell AmpliScan limits the supported input data to EEG data stored in the Cadwell Arc environment.
- The predicate device is a standalone software product that can be integrated into a . variety of vendor's application software. The Cadwell AmpliScan software is integrated only into Cadwell Arc application software.
No new issues of safety or effectiveness are introduced by the differences.
### 7) Performance Data:
The following performance data was provided in support of the substantial equivalence determination:
Software Verification and Validation Testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for the device was considered moderate level of concern.
In the Substantial Equivalence Discussion, a comparison of outputs from Cadwell AmpliScan and the predicate with like input data demonstrate the resulting equivalence of analysis and display.
Given that the Cadwell AmpliScan is a software-only device, biocompatibility, electrical safety and electromagnetic compatibility, mechanical testing, bench testing, clinical testing, and animal testing were not applicable.
### Summary:
Based on the device performance as documented in this submission. Cadwell AmpliScan was found to have a safety and effectiveness profile that is similar to the predicate device.
### 8) Conclusions:
The Cadwell AmpliScan is substantially equivalent to predicate device in design, intended use, technology, and function. The included data supports the safety and efficacy of the device, and software verification and validation demonstrate Cadwell AmpliScan performance is comparable to the predicate for the intended use.