K033999 · Physiometrix, Inc. · OLW · Feb 6, 2004 · Neurology
Device Facts
Record ID
K033999
Device Name
SEDLINE WITH FRONTAL PSI
Applicant
Physiometrix, Inc.
Product Code
OLW · Neurology
Decision Date
Feb 6, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The Sedline System is indicated for use in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. It is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents.
Device Story
Sedline System monitors brain state via real-time EEG signal acquisition and processing. Input: 4-channel EEG signals from PSArray2 electrodes (F7, F8, Fp1, Fp2, AFZ). Processing: PC-based CPU executes proprietary PSI algorithm to compute Patient State Index (PSI), Electromyograph (EMG), Artifact (ART), and Suppression Ratio (SR). Output: Real-time EEG waveforms and processed parameters displayed on monitor; data archived for review. Usage: OR, ICU, EEG lab; operated by clinicians. Benefits: Provides objective assessment of anesthetic agent effects on brain state. Hardware includes monitor, patient module (analog-to-digital converter), patient cable, and PSArray2 electrode set. System performs automatic power-up self-tests.
Clinical Evidence
Bench testing only. Validation included software, mechanical, and electrical validation testing, and EMC testing in accordance with UL 2601, CSA 22.2 No. 601-1, IEC 601-1, and IEC 601-2-26 standards.
Technological Characteristics
PC-based EEG monitor; 4-channel acquisition; proprietary PSI algorithm. Materials: flexible substrate for PSArray2. Connectivity: shielded cabling. Power: AC or internal battery. Standards: UL 2601, CSA 22.2 No. 601-1, IEC 601-1, IEC 601-2-26. Dimensions: 9" W x 8" H x 6" D monitor.
Indications for Use
Indicated for patients in the operating room, ICU, or clinical research laboratory requiring brain state monitoring via EEG signal acquisition and processing to assess the effects of anesthetic agents.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
FEB - 6 2004
### · 510(K) SUMMARY
Date: February 4, 2004 Company: Physiometrix, Inc. Five Billerica Park 101 Billerica Avenue N. Billerica, MA 01862 www.physiometrix.com Contact: Dawn E. Frazer Vice President Regulatory Affairs & Quality Assurance (978) 670-2422 x243 dfrazer@physiometrix.com Sedline with Frontal PSI Subject Device: Class II, 21 CFR Part 882.1400, Electroencephalograph Classification: The Sedline System is indicated for use in the operating room (OR), intensive Intended Use: care unit (ICU), and clinical research laboratory. It is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. Sedline is an EEG monitor designed for use in the OR, ICU, EEG laboratory and Description: for clinical research. It provides the ability to acquire and display real-time EEG waveforms, process the real time EEG data using digital signal processing techniques, display the processed EEG data in several different formats, and archive the real-time or processed EEG data for future review. The Sedline System consists of five components, the monitor, PSI algorithm, amplifier, patient cable and PSArray2 EEG Electrode Set. The device performs automatic self tests upon power up to ensure that the monitor and its
components are functioning properly.
S
page ix
KO33999
{1}------------------------------------------------
#### Monitor
The monitor provides signal processing and display capabilities for the 4 channels of real-time EEG data acquired from the preamplifier.
The monitor dimensions are 9" W x 8" H x 6" D. The color display area is 3.8" H x 5.2" W. In addition to the display area, the front panel is configured with a number of buttons to allow for configuration of the display and data acquisition settings. The monitor includes a base and a clamp that can be used to secure the unit on a flat surface or on a pole (respectively).
The processor is a PC-based CPU that processes the EEG data, calculates the processed parameters and displays the real-time EEG data and processed data. Processed parameters include Electromyograph (EMG), Artifact (ART), Suppression Ratio (SR), and the Patient State Index (PSI).
The monitor is powered through either standard wall outlet through a power cable or by the internal back-up battery power,
#### PSI Algorithm
The PSI algorithm is used to convert EEG acquired from four active (F7, F8, Fp1 and Fo2), ground (Fp2) and reference (AFZ) into a proprietary index, the PSI. The PSI is an EEG variable that is related to the effect of anesthelic agents.
#### Patient Module
The patient module is an electrically isolated, low noise, high gain, analog to digital signal converter that can process up to 4 channels of real-time data.
The preamplifier dimensions are 4.25" W x 1.75" D x 5.5" H. The patient module includes a clamp that can be used to secure the unit. The clamp can accommodate a pole, bed rail or bed sheet with a thickness of up to 1". The preamplifier is connected to the patient and monitor through flexible, shielded cabling.
#### Patient Cable
The patient cable is used to connect the patient module to the PSArray . The cable is 6 feet in length and terminated on one end with a 12 pin round connector design to mate with the palient module, and on the other end with a connector designed to mate with the PSArray".
#### PSArray
The PSArray2 is an EEG Electrode Set designed for use with the Sedline System. It is comprised of six EEG electrodes that are located in a flexible substrate.
Note that this 510(k) does not encompass the PSArray2. This device is the subject of an already cleared pre-market notification, K020670.
{2}------------------------------------------------
# Physiometrix Sedline with Frontal PSI
510(K) Notification
·
:
| Predicate Device: | PSA4000 EEG Monitor with Frontal PSI (K020671) |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Similarities: | The subject device is similar to the predicate device in the following way(s): |
| | a. Both systems are EEG monitors. |
| | b. Both systems provide a variety of processed parameters, EMG, SR, ART and PSI. |
| | c. Both systems include the PSI, a proprietary computed EEG variable that is related to the affects of anesthetics. |
| | d. Both conduct self tests at start up to assure that the device is operating. |
| | e. Both systems use the same EEG Electrode Set, PSArray². |
| Differences: | The subject device is different from the predicate device in the following way(s): |
| | a. Different hardware as required to reduce size and weight of the monitor. |
| | b. EMG trend display optional on Trend View. |
| | c. Added DSA View |
| Test Results: | The following tests have been conducted in order to verify and validate the device: software, mechanical and electrical validation testing and EMC testing. |
| | The PSA4000 has been testing in accordance with the following standards. |
| | • UL 2601 |
| | • CSA 22.2 No. 601-1 |
| | • IEC 601-1 |
| | • IEC 601-2-26 |
| | • FDA Reviewer Guidance for Premarket Notification Submissions, Section 7, Electromagnetic Compatibility dated November 1993. |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings. The bird is positioned above a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Physiometrix c/o Ms. Dawn E. Frazer Vice President, Regulatory Affairs and Quality Assurance Five Billerica Park 101 Billerica Avenue N. Billerica, Massachusetts 01862
APR - 9 2012
Re: K033999
Trade/Device Name: Sedline with Frontal PSI Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLW, OMC, ORT, OLT Dated (Date on orig SE ltr): December 23, 2003 Received (Date on orig SE ltr): January 8, 2004
Dear Ms. Frazer:
This letter corrects our substantially equivalent letter of February 6, 2004.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Dawn E. Frazer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
<http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm>.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman. M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
for
N
N
E
E
O
O
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
| 510(k) Number (if known) | K033999 |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Sedline with Frontal PSI |
| Indications For Use | The Sedline System is indicated for use in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. It is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. |
| Prescription Use<br>(Per 21 CFR 801.109) | X |
OR
| Over-The-Counter Use<br>(21 CFR 807 Subpart C) | |
|------------------------------------------------|--|
|------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
uriam C. Provost
(Evision Sign-Off)
Division of General, Restorative and Neurological Devices
Page 1 of 1
" " " } N ; " : " . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.