K082886 · Brainscope Company, Inc. · OLT · Aug 10, 2009 · Neurology
Device Facts
Record ID
K082886
Device Name
ZOOM-100DC
Applicant
Brainscope Company, Inc.
Product Code
OLT · Neurology
Decision Date
Aug 10, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The ZOOM-100DC is used to measure and record the electrical activity of a patient's brain. The ZOOM-100DC is intended to monitor the state of the brain by acquisition and display of electroencephalogram (EEG) signals and by the calculation of standard quantitative EEG (qEEG) parameters.
Device Story
ZOOM-100DC is an electroencephalograph for hospitals and clinics. It acquires EEG signals via standard off-the-shelf electrodes placed on the patient's head. The device digitizes and processes these signals to display raw EEG waveforms and calculate quantitative EEG (qEEG) parameters. The system includes an external patient interface cable, audio cable, Compact Flash card for data storage, and a battery charger. It is operated by healthcare professionals to monitor brain state. The output, consisting of raw waveforms and derived qEEG metrics (absolute/relative power, mean frequency, coherence, asymmetry), assists clinicians in assessing brain electrical activity. The device is powered by a Li-Ion battery.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological comparison and bench testing demonstrating equivalent performance to predicate devices.
Technological Characteristics
Class II electroencephalograph (21 CFR 882.1400). Features 10/20 array with 8 single-ended channels. Bandwidth: 0.5-4000 Hz (raw), 0.5-45/50 Hz (processed). CMRR ≥ 100 dB; input impedance ≥ 10 Meg Ohms. Power: Li-Ion battery. Connectivity: Compact Flash card for data export. Software performs FFT-based qEEG analysis (absolute/relative power, mean frequency, coherence, asymmetry).
Indications for Use
Indicated for monitoring brain state via acquisition/display of EEG signals and calculation of standard qEEG parameters in patients requiring brain electrical activity measurement.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Infinite Biomedical Technologies, LLC Model I-2000 Monitor (K072382)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten alphanumeric string inside of a box. The string is "K082996". The characters are written in a simple, slightly irregular style, and the box is drawn with a single line.
Image /page/0/Picture/1 description: The image shows the logo for Brainscope. Above the logo is the handwritten text "Page 1 of 4-". The logo is black and bolded.
### 510(k) Summary
# AUG 1 0 2009
The following information is provided as required by 21 CFR § 807.87 for BrainScope's 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
BRAINSCOPE COMPANY INC. Sponsor:
> 1717 Rhode Island Avenue 9th Floor Washington, D.C. 20036
| Proprietary Name: | ZOOM-100DC |
|-------------------|---------------------------|
| Common Name: | Electroencephalograph |
| Regulatory Class: | Class II 21 CFR §882.1400 |
| Product Codes: | OLT & OMC |
Predicate Device(s): Nicólet Bravo Multi-Modality System (K991054), Cleveland Medical Crystal-EEG Model 10 (K970672), Brainz Instruments Ltd BRM3 Brain Monitor (K071449) and Infinite Biomedical Technologies, LLC Model I-2000 Monitor (K072382).
### Device Description:
The ZOOM-100DC records, measures and displays Electroencephalographic (EEG) waveforms which are digitized and processed.
1717 Rhode Island Avenue 9th Floor Washington, D.C. 20036
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "Page 2 of 4" written in cursive handwriting. Below this text is the word "BRAINSCOPE" in a bold, sans-serif font. The word is all capitalized and has a horizontal line above it.
### Indications for Use:
The ZOOM-100DC is used to measure and record the electrical activity of a patient's brain. The ZOOM-100DC is intended to monitor the state of the brain by acquisition and display of electroencephalogram (EEG) signals and by the calculation of standard quantitative EEG (qEEG) parameters.
# Device Comparison Table:
| | ZOOM-100DC | Bravo Multi-<br>Modality System<br>K991054 | Crystal-EEG Model<br>10 (K970672) | BRM3 Brain Monitor<br>(K071449) | Model I-2000<br>Monitor (K072382) |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | The ZOOM-100DC<br>is used to measure<br>and record the<br>electrical activity of a<br>patient's brain. The<br>ZOOM-100DC is<br>intended to monitor<br>the state of the brain<br>by acquisition and<br>display of<br>electroencephalogram<br>(EEG) signals and by<br>the calculation of<br>standard quantitative<br>EEG (qEEG)<br>parameters. | The Bravo Multi-<br>Modality System is<br>intended to record and<br>display EEG, EP,<br>EMG and TCD data<br>in the clinic and<br>hospital (including<br>the hospital room,<br>operating room,<br>emergency room,<br>intensive care unit,<br>neuro intensive care<br>unit, critical care unit,<br>etc.), and to import<br>and display data from<br>third party monitoring<br>devices such as vital<br>signs monitors. It is<br>intended to aid the<br>diagnosis and<br>monitoring of<br>potential disorders of<br>the central and<br>peripheral nervous<br>system and muscles. | The Crystal-EEG<br>Model 10 is a mobile,<br>intermediate range,<br>wireless EEG system<br>intended to be used<br>for measuring and<br>transmitting<br>electroencephalogram<br>(EEG) signals. | A device used to<br>measure and record the<br>electrical activity of the<br>patient's brain obtained<br>by placing two or more<br>electrodes on the head.<br>The BRM3 Brain<br>Monitor is<br>intended for use to<br>monitor the state<br>of the brain by<br>acquisition of<br>electroencephalog<br>ram (EEG)<br>signals, in clinical<br>environments<br>such as the<br>intensive care<br>unit, and for<br>clinical research.<br>The BRM3 Brain<br>Monitor is<br>intended for use<br>only with neonatal<br>patients, defined<br>as from birth to 28<br>days post-<br>delivery, and<br>corresponding to a<br>post-conceptual<br>age of 24 to 46<br>weeks.<br>The BRM3 Brain<br>Monitor does not<br>provide any<br>diagnostic<br>conclusion about<br>the patient's<br>condition. | The 1-2000 Monitor<br>is intended to be used<br>for measuring and<br>recording the<br>electrical activity of a<br>patient's brain,<br>obtained by placing<br>electrodes on the<br>head.<br>The 1-2000 Monitor<br>is indicated for use in<br>monitoring the state<br>of the brain by<br>acquisition of<br>electroencephalogram<br>(EEG) signals, in<br>research and clinical<br>environments. |
| Modalities | EEG | EEG, Evoked<br>Potential EMG, CSA | EEG | EEG | EEG |
| Environment<br>of Use | Hospitals and Clinics | Hospitals and Clinics | Hospitals and Clinics | Hospitals and Clinics | Hospitals and Clinics |
| Power Source | Li Ion Battery | 120 Volt 60Hz AC | Unknown | Unknown | Unknown |
1717 Rhode Island Avenue 9t Floor Washington, D.C. 20036
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "Page 3 of 4" written in cursive handwriting. Below the page number is the logo for BrainScope. The BrainScope logo is in bold, black font.
| Device Components | ZOOM-100DC<br>External Patient<br>Interface Cable<br>External Audio Cable<br>CF (Compact Flash)<br>Card<br>Battery Charger | PC running Microsoft<br>windows NT<br>Amplifier Unit<br>Patient Interface<br>Cables<br>Evoked Potential Unit<br>Evoked Potential<br>Patient interface<br>Cable | Unknown | Sensor Set (EEG skin<br>electrodes) or Sensor<br>Adaptor Set (electrode<br>adaptor)<br>Data Acquisition Unit<br>(EEG head stage)<br>Data Cable<br>Monitor<br>Power Supply Unit<br>and Roll-Pole. | tablet PC<br>battery powered<br>wireless headbox<br>headbox charger<br>power conditioner | | | | | | | | |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|---------|--|---------|--|---------|--|--|--|
| Electrode<br>Material | N/A | N/A | N/A | N/A | N/A | | | | | | | | |
| | Uses Standard off the<br>Shelf EEG electrodes | Uses Standard off the<br>Shelf EEG electrodes | Uses Standard off the<br>Shelf EEG electrodes | Uses Standard off the<br>Shelf EEG electrodes | Uses Standard off the<br>Shelf EEG electrodes | | | | | | | | |
| Screen<br>Display<br>Details | Displays:<br>Raw EEG Waveform | Displays:<br>Raw EEG Waveform<br>Raw averaged AEP<br>Waveform | Unknown | Displays:<br>Raw EEG Waveform<br>aEEG parameter<br>Signal Quality | Displays:<br>Raw EEG waveforms<br>EEG power<br>spectrum<br>95% Spectral Edge<br>Frequency | | | | | | | | |
| Hard copy of<br>Displayed<br>Data | Yes - through CF<br>(Compact Flash) card | Yes | Unknown | Unknown | Yes | | | | | | | | |
| Channels | 10/20 Array | 10/20 Array | 8 channels | 3 channels | 2 channels | | | | | | | | |
| | 8 single-ended<br>channels<br>corresponding to 8<br>electrodes placed<br>anywhere on the<br>head, including but<br>not limited to, all<br>locations defined by<br>the International<br>10/20 System. (5<br>differential channels<br>can be viewed<br>concurrently). | Up to 16 differential<br>channels | | | | | | | | | | | |
| Real Time<br>EEG Display | Yes | Yes | Yes | Yes | Yes | | | | | | | | |
| Real Time<br>EEG<br>Bandwidth | 0.5-4000 Hz<br>available | 0.5 - 500 Hz<br>Available EEG<br><br>100 - 20kHz<br>Available AEP | Unknown | Unknown | Unknown | | | | | | | | |
| Processed<br>EEG<br>Bandwidth | 0.5 - 45 Hz<br>50Hz | 0.5 - 500 Hz | Unknown | Unknown | Unknown | | | | | | | | |
| Automatic<br>Artifacting | | Yes | | Unknown | | Unknown | | Unknown | | Unknown | | | |
1717 Rhode Island Avenue 9th Floor Washington, D.C. 20036
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "Page 4 of 4" written in cursive. The number 4 is circled. The text indicates that this is the last page of a document.
Image /page/3/Picture/1 description: The image shows the word "BRAINSCOPE" in a bold, sans-serif font. The word is all in uppercase letters and is black. There are horizontal lines above and below the word. The letters are closely spaced together, creating a solid block of text.
| Amplifier<br>Common<br>Mode<br>Rejection<br>Ratio<br>(CMRR) | ≥ 100 dB | ≥ 110 dB | Unknown | Unknown | Unknown |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|---------|---------|---------|
| Amplifier<br>Input<br>Impedance | ≥ 10 Meg Ohms | > 100 Meg Ohms | Unknown | Unknown | Unknown |
| Electrode<br>Impedance<br>Test | Yes | Yes | Unknown | Unknown | Yes |
| EEG Derived<br>Measures | Yes- Derived from<br>FFT* | Yes- Derived from<br>FFT* | N/A | N/A | N/A |
| | 1. Absolute Power<br>a. Monopolar Power<br>b. Bipolar Power<br><br>2. Relative Power<br>a. Relative<br>Monopolar Power<br>b. Relative Bipolar<br>Power<br><br>3.Mean frequency<br>variables (univariate<br>and multivariate)<br>a. Monopolar Mean<br>Frequency<br>b. Bipolar Mean<br>Frequency<br><br>4. Coherence<br>a. Monopolar<br>Coherence<br>b. Bipolar<br>Coherence<br><br>5. Asymmetry<br>a. Monopolar<br>Asymmetry | | | | |
* See comparison of measures in section 12.3.
### Conclusion
The ZOOM-100DC, when compared to its predicate devices, has the same intended use and equivalent technological characteristics. Performance data demonstrate that the device performs equivalently to the predicate devices. The ZOOM-100DC is substantially equivalent.
1717 Rhode Island Avenue 9th Floor Washington, D.C. 20036
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for the Department of Health & Human Services - USA. The logo consists of two parts: a circular text element and a symbolic graphic. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. To the right of the text is a symbol that resembles an abstract eagle or bird in flight, composed of three stylized lines.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Brainscope Company Inc. % Mr. Michael E. Singer 1717 Rhode Island Avenue, 9th Floor Washington, District of Columbia 20036
2009
Re: K082886
Trade/Device Name: ZOOM-100DC Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLT, OMC Dated: July 24, 2009 Received: July 28, 2009
Dear Mr. Singer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
## Page 2 - Mr. Michael E. Singer
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Mark N. Melkerson Director Division of Surgical, Orthopedic,
Sincerely vours.
and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use Statement
K082886 510(k) Number:
Device Name: ZOOM-100DC
### Indications for Use:
The ZOOM-100DC is used to measure and record the electrical activity of a patient's brain. The ZOOM-100DC is intended to monitor the state of the brain by acquisition and display of electroencephalogram (EEG) signals and by the calculation of standard quantitative EEG (qEEG) parameters.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
、
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) (Division Sign-On)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K082886
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.