The NightWare digital therapeutic is indicated to provide vibrotactile feedback on an Apple Watch based on an analysis of heart rate and motion during sleep for the temporary reduction of sleep disturbance related to nightmares in adults 22 years or older who suffer from nightmare disorder or have nightmares from posttraumatic stress disorder (PTSD). It is intended for home use.
Device Story
NightWare Kit uses Apple Watch sensors to monitor heart rate and motion during sleep; data processed to detect physiological signs of nightmares; device provides vibrotactile feedback via Apple Watch to disrupt nightmares without waking patient. Intended for home use by patients; prescription required. Output informs patient sleep quality; intended as adjunct therapy, not replacement for medication. Benefits include temporary reduction of sleep disturbance.
Clinical Evidence
Clinical performance testing required to evaluate therapy efficacy for sleep disturbance, monitoring for worsening of condition-specific symptoms, and assessment of sleepiness using validated measures.
Technological Characteristics
General purpose computing platform (Apple iPhone, Apple Watch). Sensors: heart rate, accelerometer, gyroscope. Connectivity: WiFi for server data transmission. Software: digital therapy application. Stimulation: vibrotactile feedback. Operation: home-use, prescription-only.
Indications for Use
Indicated for adults 22+ with nightmare disorder or PTSD-related nightmares to temporarily reduce sleep disturbance.
Regulatory Classification
Identification
A digital therapy device to reduce sleep disturbance for psychiatric conditions is a prescription device that is intended to provide stimulation using a general purpose computing platform to reduce sleep disturbance in patients who experience this symptom due to psychiatric conditions such as nightmare disorder or post-traumatic stress disorder.
Special Controls
In combination with the general controls of the FD&C Act, the digital therapy device to reduce sleep disturbance for psychiatric conditions is subject to the following special controls:
*(b) Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing under the labeled conditions for use must evaluate the following:
(i) The ability of the device to provide therapy for patients with sleep disturbance due to psychiatric conditions, using a validated measure;
(ii) Worsening of any condition-specific symptoms using a validated measure for assessment of the particular condition; and
(iii) Increase in symptoms of disturbed sleep or sleepiness using a validated measure.
(2) Software must clearly describe all features and functions of the software implementing the digital therapy. Software verification, validation, and hazard analysis must also be provided.
(3) The labeling must include the following:
(i) Patient and physician labeling must include instructions for use, including images that demonstrate how to interact with the device;
(ii) Patient and physician labeling must list the minimum operating system and general purpose computing requirements that support the software of the device;
(iii) Patient and physician labeling must include a warning that the digital therapy device is not intended for use as a stand-alone therapeutic device;
(iv) Patient and physician labeling must include a warning that the digital therapy device does not represent a substitution for the patient's medication; and
(v) Physician labeling must include a summary of the clinical performance testing conducted with the device.
Submission Summary (Full Text)
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR NIGHTWARE KIT (APPLE IPHONE, APPLE WATCH, APPLE IPHONE CHARGING CABLE, APPLE WATCH CHARGING CABLE)
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Digital therapy device to reduce sleep disturbance for psychiatric conditions. A digital therapy device to reduce sleep disturbance for psychiatric conditions is a prescription device that is intended to provide stimulation using a general purpose computing platform to reduce sleep disturbance in patients who experience this symptom due to psychiatric conditions such as nightmare disorder or post traumatic stress disorder (PTSD).
NEW REGULATION NUMBER: 21 CFR 882.5705
CLASSIFICATION: Class II
PRODUCT CODE: QMZ
#### BACKGROUND
DEVICE NAME: NightWare Kit (including Apple iPhone, Apple Watch, Apple iPhone Charging Cable, Apple Watch Charging Cable)
SUBMISSION NUMBER: DEN200033
DATE DE NOVO RECEIVED: May 27, 2020
#### SPONSOR INFORMATION:
NightWare, Inc 153 Ashley Road Hopkins, Minnesota 55343
#### INDICATIONS FOR USE
The NightWare digital therapeutic is indicated to provide vibrotactile feedback on an Apple Watch based on an analysis of heart rate and motion during sleep for the temporary reduction of sleep disturbance related to nightmares in adults 22 years or older who suffer from nightmare disorder or have nightmares from posttraumatic stress disorder (PTSD). It is intended for home use.
{1}------------------------------------------------
## LIMITATIONS
The sale, distribution, and use of the NightWare device are restricted to prescription use in accordance with 21 CFR 801.109.
The NightWare device is intended for use in adults only.
The NightWare device is not intended to be used as a standalone therapy for PTSD and should be used in conjunction with a patient's prescribed medications and other recommended therapy for PTSD-associated nightmares and nightmare disorder, according to relevant consensus guidelines.
The NightWare device is not intended for use in patients who "act out" nightmares (including in patients who sleepwalk or are violent during nightmares)
The long term safety and effectiveness of the NightWare device have not been established.
The long term effects of the NightWare device use on the sleep architecture have not been established.
Patients who experience drowsiness or daytime sleepiness while using the device should not drive or operate heavy machinery and should contact their healthcare provider.
The "Stress Index" measured by the device is a device-specific measurement that does not represent a clinically validated measure of stress.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
#### DEVICE DESCRIPTION
The NightWare device consists of a software application and the NightWare server. The device uses an Apple Watch and an Apple iPhone for its platform. Patients must use the prescribed iPhone and Apple Watch pre-provisioned by NightWare, Inc.
The watch monitors physiological parameters that indicate that the patient is experiencing a nightmare. Signals from the watch's heart rate sensor, accelerometer, and gyroscope are input to the onboard algorithms on the watch software and sent to the server and processed by NightWare software. The software on the server calculates a device-specific "Stress Index" from watch measurements of heart rate, rotation, and acceleration. When a "Stress Index" threshold is exceeded, the device provides a vibrotactile stimulation on the patient's wrist intended to interrupt the nightmare but not awaken the patient.
Patients need to have at least an intermittent connection to wireless internet through WiFi in order to transmit data to the NightWare server. However, this wireless internet connection is not
{2}------------------------------------------------
needed to receive therapy through the treatment period, as the processing of physiological data and delivery of stimulation occurs on the watch itself.
The "Stress Index" threshold is uniquely calculated for each patient based on an artificial intelligence algorithm. The algorithm differentiates between a patient's normal and abnormal "Stress Index" levels during the night. To create its personalized "Stress Index" threshold, NightWare collects several hours of data during sleep before any intervention is applied. During this initial time, the system monitors the patient's movements and heart rate to delineate the usual low, medium, and high "Stress Index" periods to determine the wearer's usual sleep patterns. The movement and heart rate information is then securely sent to the NightWare server to establish the patient's specific "Stress Index" threshold; when the threshold is reached the vibrotactile stimulation is applied. The "Stress Index" threshold is automatically and periodically updated as needed to accommodate the patient's changing "Stress Index" and thresholds that naturally occur during consistent use.
#### SUMMARY OF NONCLINICAL/BENCH STUDIES
NightWare is a software as a medical device implemented on a general purpose computing platform, and non-clinical/bench testing was generally not needed to demonstrate the safety and effectiveness of the device.
#### SOFTWARE
Software Verification & Validation testing and documentation was provided according to a Moderate Level of Concern and FDA's guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005) to demonstrate that the device software performs as intended. The software documentation provided a clear description of all features and functions of the software implementing the digital therapy.
#### SUMMARY OF CLINICAL INFORMATION
#### Primarv validation study design
A 30-day, double-blind sham-controlled randomized clinical trial of the NightWare Digital Therapeutic was performed study the safety and effectiveness of the device to reduce sleep disturbance related to nightmare disorder associated with post-traumatic stress disorder in adults 22 vears and older. The NightWare device is referred to as the "Active System." The "Sham System" consisted of the same components as the Active System, but the application only monitored the subjects' sleep and did not provide any intervention or feedback to the subject.
#### Study Population
Subjects 22 years of age or older with a prior or current diagnosis of PTSD and current nightmares resulting in sleep disturbance.
{3}------------------------------------------------
# Relevant Inclusion Criteria
- Documented diagnosis of PTSD via American Psychiatric Association PTSD ● diagnostic
criteria in the fourth or fifth edition of its Diagnostic and Statistical Manual of Mental Disorders (DSM-IV, DSM-IV, DSM-IV TR, or DSM-5).
- 22 years of age or older ●
- . Pittsburgh Sleep Quality Index (PSQI) score 10 or more at screening.
- Have repetitive nightmares contributing to disrupted sleep as reported by the subject. ●
- If currently taking Prazosin, must be willing to be tapered off of Prazosin by their ● prescribing provider. Subject must be off Prazosin for 2 weeks prior to enrollment.
# Relevant Exclusion Criteria
- Patient Health Questionnaire-9 (PHQ-9) score greater than or equal to "", with a . score of (14) on question #9 (suicidal ideation) triggering a risk assessment by the study principal investigator or qualified co-investigator
- Responses on the C-SSRS that indicate High Suicide Risk including current suicidal . ideation, which will prompt assessment by a licensed psychologist.
- Cardiovascular conditions and medications: ●
- Uncontrolled atrial fibrillation o
- Current use of varenicline O
- Current use of beta-blockers (unless ophthalmic solutions) o
- o Current use of non-dihydropyridines
- Circadian rhythm disruption on a regular basis (shiftwork) O
- Sleep- and nightmare-related: ●
- Known diagnosis of obstructive sleep apnea (OSA) о
- Diagnosis of an active disorder of arousal from non-rapid eye movement sleep O
- Diagnosis of rapid eye movement sleep behavior disorder O
- Diagnosis of narcolepsy O
- Previous or foreseeable legal proceedings involving nightmares or trauma O
- Nocturia that causes awakening from sleep O
- Known sleep walking O
- Acting out of dreams prior to PTSD trauma O
- Substance Use:
- o Alcohol Use Disorders Inventory Test (AUDIT) score equal to " or higher
- Drug Abuse Screening Test-10 (DAST-10) score greater than O
- Investigator suspicion of nightmares being secondary to substance abuse or o withdrawal
- Diagnosis or investigator suspicion of dementia ●
# Objectives
# Primary Effectiveness Objective
Demonstrate that subjects treated with the Active System show improved sleep as assessed by the Pittsburgh Sleep Quality Index (PSQI) as compared to the Sham System.
# Primary Safety Objectives
{4}------------------------------------------------
Demonstrate that the NW digital therapeutic does not lead to:
- . Worsening of daytime sleepiness as assessed by the Epworth Sleepiness Scale (ESS).
- . Increase in suicidality as assessed by the Columbia Suicide Severity Rating Scale (C-SSRS).
## Secondary Objectives
Demonstrate each of the following for the Active System as compared to Sham System:
- Evaluate the difference between average PSQI-Addendum for PTSD (PSQI-A) at ● baseline (Day 0) and average PSQI-A at Day 30.
- . Evaluate changes in PTSD severity as assessed by the PTSD Checklist (PCL-5, past month version).
- Assess subjective sleep quality with the Likert Scale for Sleep Quality at Days 0, 7, ● and 30 to assess improvement in sleep.
- Demonstrate that there is not an increase in symptoms of depression as assessed by Patient Health Questionnaire-9 (PHQ-9, past month version).
- Evaluate changes in quality of life as assessed by the Functional Outcomes of Sleep . Questionnaire (FOSO-10).
- . Evaluate changes in nightmare severity as measured by the Trauma Related Nightmare Scale (TRNS).
- Evaluate changes in nightmare frequency as assessed by TRNS. ●
- . Evaluate changes in quality of life as assessed by Veterans Rand 12 Quality of Life instrument (VR-12).
#### Outcome measures
#### Primary Efficacy Outcome of Active System as compared to Sham System:
- Change in average PSQI score at baseline (Day 0) and average PSQI score at the . post- enrollment assessment at Day 30.
#### Primary Safety Outcome:
- Comparison between average ESS at baseline and Day 30. .
- Comparison of suicide risk (Low, Medium, High) between C-SSRS at baseline and . Day 30.
#### Secondary and Observational Outcomes:
- Change in average PSQI-A at baseline and Day 30 as compared to sham. .
- . Change in average PCL-5 at baseline and Day 30 as compared to sham.
- Change in average PHO-9 at baseline and Day 30 as compared to sham. ●
- Change in of average TRNS at baseline and Day 30 as compared to sham. ●
- Change in average FOSQ-10 Quality of Life Measure at baseline and Day 30 as compared to sham.
- Change in average VR-12 at baseline and average VR-12 at Day 30 as compared to sham.
# Sites
This study was conducted at the Minneapolis Veterans Affairs Health Care System.
{5}------------------------------------------------
## Results
Enrollment: In the study, 130 potential participants were screened, 70 were enrolled, and 63 adults completed the study procedure. Six patients in the Active arm withdrew or were lost to follow-up, and one in the sham arm withdrew. One patient dropped out of the Active arm after reporting that the device disrupted sleep; none of the other dropouts or losses to follow-up were determined to be directly related to device usage. The study results are displayed in Tables 1 and 2, and the most relevant safety and effectiveness outcomes are discussed below.
## Safety:
- . Primary safety: Patients in both the Active and Sham arms of the study reported a 1.2 point decrease (less sleepiness) in the Epworth sleepiness scale (ESS). The difference was not statistically significant at the p (b) (4) level.
- . Primary safety: Patients in the Active arm reported a 0.2 point decrease in the C-SSRS, and patients in the Sham arm reported no change in the C-SSRS. The difference was not statistically significant at the p(b) (4) level. Due to a protocol violation, the first 20 patients enrolled in the study were mistakenly not administered the C-SSRS, but this occurred in both study arms. A complete tabulation of C-SSRS results are provided in Table 2.
#### Effectiveness:
- Primary effectiveness: Patients in the active arm had a mean PSQI improvement of . 3.2 points, and patients in the sham arm had a mean improvement of 2.2 points. The difference was not statistically significant at the p (b) (4) level.
- . Secondary/observational effectiveness: For the secondary and observational outcome measures listed above trends toward improvement of the symptoms assessed by the respective outcomes were observed. One of the key secondary outcome measures was the POSI-A, which is a measure of seven sleep disturbance items often reported by adults with PTSD (including sleep disturbance due to anxiety, nervousness, bad dreams, terrors or screaming during sleep). Patients in the active arm had a mean PSQI-A improvement of 3.3 points; patients in the sham arm had a mean improvement of 1.4 points. The difference was not statistically significant at the p level.
#### Adverse events
Two adverse events were reported during the study. Neither was determined to have a likely relation to device use. One patient was hospitalized for suicidality after study enrollment but prior to use of the investigational device. Another patient who was enrolled and using the device was diagnosed with sleep apnea, but the factors contributing to sleep apnea were determined to be likely present before device use began.
{6}------------------------------------------------
| Baseline Outcome | | | |
|---------------------|-----------------|---------------|---------|
| Measure | Active (n = 29) | Sham (n = 34) | p-value |
| CSSRS 1 Month Score | -0.2 (0.8) | 0.0 (1.0) | 0.2943 |
| PSQI | -3.2 (3.7) | -2.2 (2.9) | 0.2606 |
| PHQ-9 | -2.0 (3.7) | -1.0 (3.8) | 0.2719 |
| PCL | -9.9 (13.4) | -6.5 (9.9) | 0.2727 |
| ESS | -1.2 (4.1) | -1.2 (3.1) | 0.9739 |
| TRNS | -4.8 (7.3) | -2.7 (4.4) | 0.1909 |
| FOSQ | 1.4 (2.8) | 0.8 (2.3) | 0.3683 |
| PSQI-A | -3.3 (4.9) | -1.4 (3.5) | 0.0938 |
| VR12 | 2 (5.8) | 1.6 (6.9) | 0.8093 |
Table 1. Change in outcomes at Day 30 (as compared to baseline, Day 0) for Active and Sham arms. Data shown are mean, with standard deviation in parenthesis. P-values represent a twosample t-test.
| | | Baseline | | Completion | |
|----------|--------------------------------------------------------|----------|------|------------|------|
| Score | Description | Active | Sham | Active | Sham |
| 0 | No ideation | (b) (4) | | | |
| 1 | Wish to be dead | | | | |
| 2 | Non-specific active suicidal thoughts | | | | |
| 3 | Active suicidal ideation, no intent to act, no plan | | | | |
| 4 | Active suicidal ideation, some intent to act, no plan | | | | |
| 5 | Active suicidal ideation with specific plan and intent | | | | |
| Total >0 | | | | | |
Table 2: Reports of suicidal ideation (according to the C-SSRS) at baseline and completion (Day 30) for Active and Sham arms
#### Conclusions
No device-related adverse events were observed in the 30 day study.
There was a 0.2 point difference between the Active and Sham arms in the C-SSRS suicidality safety endpoint, and no difference between arms was observed in the ESS daytime sleepiness safety endpoint.
There was a 1 point difference between the Active and Sham arms in the PSQI sleep assessment primary effectiveness endpoint, but the difference was not statistically
{7}------------------------------------------------
significant. There was a " " point difference between the Active and Sham arms in the PSQI-A endpoint, which incorporates a nightmare component and is considered a more relevant assessment of sleep disturbances in PTSD. Considering the low risk of the device, both the beneficial 3.2 point decrease for the PSQI and beneficial 3.3 point decrease in the PSQI-A in the treatment arm were judged to support that the device would be beneficial to this patient population by providing an acceptable adjunct therapy to their treatment plan, in comparison to other medical treatment approaches which may have potentially more risks to the patient.
No statistically-significant differences were observed in any of these endpoints. The study enrolled fewer patients than prospectively planned, hence, there is a high likelihood that the study is underpowered to detect a statistically significant difference in these endpoints.
Although the study was likely underpowered to detect a statistically significant difference in endpoints between the Active and Sham arms, the results show trends in improvement in outcomes related to sleep disturbance and more specifically sleep disturbance related to nightmares.
## Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population. The device is indicated for use in adults only.
#### LABELING
The NightWare Instructions for Use are consistent with the clinical data and cover all the hazards and other clinically relevant information that may impact use of the device. The labeling is sufficient and satisfies the requirements of 21 CFR § 801.109 for prescription devices. The sole contraindication listed is for patients who are known to "act out" their nightmares (including sleepwalking or displaying outbursts of violence). Both physician and patient labeling include instructions for use with images and computing requirements. The physician labeling includes a summary of the clinical study discussed in this document.
The following precautionary statements are included within the device labeling:
- Federal law restricts this device to sale by or on the order of a physician. .
- . NightWare is not a standalone therapy for PTSD. The device should be used in conjunction with prescribed medications for PTSD and other recommended therapies for PTSD-associated nightmares and nightmare disorder, according to relevant consensus guidelines.
- NightWare is intended to be used under the supervision of a healthcare provider. ●
- . If daytime sleepiness occurs, the user is instructed to contact their healthcare provider and not to operate heavy machinery or drive.
- If NightWare causes awakenings not associated with nightmares, the user is instructed to contact their healthcare provider.
- NightWare is intended for adults 22 years or older.
{8}------------------------------------------------
- The long term safety and effectiveness of the NightWare device have not been established.
- The long term effects of the NightWare device use on the sleep architecture have not been established.
### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of a digital therapy device to reduce sleep disturbance for psychiatric conditions and the measures necessary to mitigate these risks.
| Identified Risks to Health | Mitigation Measures |
|---------------------------------------------------------------------------|-----------------------------------------------------------|
| Ineffective treatment leading to worsening sleep | Clinical performance testing |
| Ineffective treatment leading to worsening<br>condition-specific symptoms | Clinical performance testing |
| Device software failure leading to delayed<br>access and treatment | Software verification, validation, and hazard<br>analysis |
| Improper device use leading to worsening sleep | Labeling |
#### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the digital therapy device to reduce sleep disturbance for psychiatric conditions is subject to the following special controls:
- (1) Clinical performance testing under the labeled conditions for use must evaluate the following:
- (i) The ability of the device to provide therapy for patients with sleep disturbance due to psychiatric conditions. using a validated measure:
- (ii) Worsening of any condition-specific symptoms using a validated measure for assessment of the particular condition: and
- (iii)Increase in symptoms of disturbed sleep or sleepiness using a validated measure.
- (2) Software must clearly describe all features and functions of the software implementing the digital therapy. Software verification, validation, and hazard analysis must also be provided.
- (3) The labeling must include the following:
- (i) Patient and physician labeling must include instructions for use. including images that demonstrate how to interact with the device;
- (ii) Patient and physician labeling must list the minimum operating system (OS) and general purpose computing requirements that support the software of the device;
- (iii)Patient and physician labeling must include a warning that the digital therapy device is not intended for use as a standalone therapeutic device;
- (iv)Patient and physician labeling must include a warning that the digital therapy device does not represent a substitution for the patient's medication; and
{9}------------------------------------------------
- (v) Physician labeling must include a summary of the clinical performance testing conducted with the device.
#### BENEFIT-RISK DETERMINATION
The risks of the device are based on the data collected in the clinical study described above.
Two adverse events were reported. However, none were determined to be device-related.
One patient reported that the device disturbed their sleep and therefore withdrew from the study. It is possible that the device's vibrotactile stimulation may actually result in sleep fragmentation and increase the arousal index, leading to sleep disruption as a result. The supporting clinical trial did not perform the polysomnography measurements needed to assess this, nor was the study duration sufficient to investigate this effect. However, at 30 days, no change in daytime sleepiness was observed according to the Epworth Sleepiness scale. The NightWare device does not appear to worsen daytime sleepiness or drowsiness in the short term. The long term effects of the device on sleepiness and sleep fragmentation are not known; if there is any evidence of increased sleepiness, the device labeling advises that patients should contact their healthcare provider.
In this study, 76% of patients reported a lifetime history of suicidal ideation, and suicidal ideation is prevalent within the intended use population. In the supporting validation study, patients in the Active arm at 30 days showed no worsening suicidal ideation on average according to the C-SSRS. The NightWare device therefore does not appear to worsen suicidal ideation during short duration use. Some uncertainty exists in the assessment of suicidal ideation due to the 20 protocol violations that occurred at the beginning of the study; however, these were balanced across the Active and Sham groups. Longer term influence of device use on suicidality are not known, but therapies that reduce sleep disturbance due to nightmares would be expected to lessen suicidality.
The probable benefits of the device are based on the data collected in the clinical study described above.
The 30 day, randomized, controlled, blinded pivotal validation study of 70 patients demonstrated a 3.2 point change in the PSQI for patients in the Active treatment arm and a 2.2 point change for patients in the Sham treatment arm at day 30 as compared to baseline, for a 1.0 point difference in the PSQI between treatment groups. For the PSQI-A, patients in the Active treatment arm showed a greater benefit relative to sham with (b) ( point difference between groups. The PSQI-A is likely a more useful endpoint for the assessment of sleep disturbance in the intended patient population for the NightWare device. The reason for this is that the PSQI-A explicitly assesses aspects of sleep disturbance that are related to dreams and nightmares as well as the anxiety and panic associated with those dreams and nightmares. The NightWare device was judged, in comparison to the potential low risks, to have greater benefit on the PSOI-A than the PSQI, which suggests greater effectiveness for reducing sleep disturbance related to
{10}------------------------------------------------
nightmares in patients with PTSD as opposed to reducing sleep disturbance as measured by the PSOI in a general population.
There is uncertainty in the benefit demonstrated in the clinical study described above. Seven patients did not complete the study, with the majority of them (6) being in the Active treatment group. While one patient explicitly noted that their dropout was due to the device interrupting sleep, the reasons for the other patient dropouts are unclear. The results include numerical differences only between treatment groups and wide confidence intervals, likely due to the study being underpowered as a result of the early termination. The study was not of sufficient duration to assess the longer term effectiveness of the NightWare device. Despite a number of sources of uncertainty, there is evidence of potential benefit due to the improvement in the PSQI-A and PSQI scores as compared to the sham device.
Additional benefit/risk factors include patient tolerance for risk and risk mitigations due to the device's design. Patients are likely to tolerate uncertainty in benefit, because the device is low risk and offers a passive, non-pharmaceutical method to reduce sleep disturbance in a chronic condition. Additionally, patients may easily remove the device if they believe it to be ineffective or that it is interrupting or worsening their sleep and is not mitigating their sleep disturbance due to nightmares.
Finally, as a Breakthrough-designated device, no other cleared or approved devices address the medical need being met by the NightWare device. Although there are other clinical and pharmaceutical options that address this medical need, this device offers a non-pharmaceutical option that can be used in the home setting.
For the reasons described above, the probable benefits of the NightWare device outweigh the probable risks in light of the listed special controls and the general controls.
#### Patient Perspectives
Patient perspectives considered for the NightWare Kit (Apple iPhone, Apple Watch, Apple iPhone Charging Cable, Apple Watch Charging Cable) during the review include:
Three patients were interviewed on their experiences using the device for 5 weeks. All three patients reported benefit from the device use, noting that the device provided a nonpharmaceutical treatment alternative. The patients noted that the pharmaceutical treatment could leave them feeling "in a fog," while the NightWare device did not.
#### Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
The NightWare digital therapeutic is indicated to provide vibrotactile feedback on an Apple Watch based on an analysis of heart rate and motion during sleep, for the
{11}------------------------------------------------
temporary reduction of sleep disturbance related to nightmares in adults 22 years or older who suffer from nightmare disorder or have nightmares from posttraumatic stress disorder (PTSD). It is intended for home use.
the probable benefits outweigh the probable risks for the NightWare Kit (Apple iPhone, Apple Watch, Apple iPhone Charging Cable, Apple Watch Charging Cable). The device provides benefits to this patient population and the risks can be mitigated by the use of general controls and the identified special controls.
## CONCLUSION
The De Novo request for the NightWare Kit (Apple iPhone, Apple Watch, Apple iPhone Charging Cable, Apple Watch Charging Cable) is granted and the device is classified as follows:
Product Code: QMZ Device Type: Digital therapy device to reduce sleep disturbance for psychiatric conditions Regulation Number: 21 CFR 882.5705 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.