ELECSYS HBSAG IMMUNOASSAY, ELECSYS HBSAG CONFIRMATORY, AND PRECICONTROL HBSAG
Applicant
Roche Diagnostics Corp.
Product Code
LOM · Microbiology
Decision Date
Jun 1, 2001
Decision
APWD
Regulation
21 CFR 866.3172
Device Class
Class 2
Indications for Use
For the in vitro test qualitative detection of hepatitis B surface antigen (HBsAg) in human serum or plasma (sodium heparin, EDTA-K3 and sodium citrate). Assay results, in conjunction with other serological and clinical information may be used for the laboratory diagnosis of individuals with acute or chronic hepatitis B. This assay may be used to screen for hepatitis B infection in pregnant women to identify neonates who are at high risk of acquiring HBV during the perinatal period. This assay is intended only for use on the Elecsys® 2010 immunoassay analyzer, and is based on electrochemiluminescence immunoassay technology (“ECLIA”).
Device Story
The Elecsys HBsAg Immunoassay is an automated in vitro diagnostic test for detecting hepatitis B surface antigen in human serum or plasma. It utilizes electrochemiluminescence immunoassay (ECLIA) technology on the Elecsys 2010 analyzer. The process involves a sandwich complex formation using biotinylated and ruthenium-labeled monoclonal antibodies specific to the HBsAg 'a' region determinant. These complexes bind to streptavidin-coated microparticles, which are magnetically captured on an electrode. Voltage application induces chemiluminescent emission, measured by a photomultiplier. The Elecsys software automatically calculates results by comparing the signal to a calibrated cutoff. A confirmatory test using specific antibody neutralization is required for repeatedly reactive samples. The system is operated by laboratory personnel in clinical settings. By providing accurate HBsAg detection, the device aids clinicians in diagnosing HBV infection and identifying high-risk pregnant patients, facilitating timely intervention to prevent perinatal transmission.
Clinical Evidence
Clinical performance evaluated in a multi-center prospective study (n=1445) including first-time blood donors and at-risk populations. Sensitivity and specificity compared to FDA-licensed reference methods. Overall agreement was high; 100% positive agreement (6/6) and 99.9% negative agreement (111/111) in at-risk subjects. Confirmatory test performance showed 98% agreement with reference neutralization assays. Bench testing included analytical sensitivity (LOD 0.028 IU/mL), matrix effects, endogenous interference, carryover, and high-dose hook effect studies.
Technological Characteristics
Electrochemiluminescence immunoassay (ECLIA) on Elecsys 2010 analyzer. Reagents include streptavidin-coated microparticles, biotinylated anti-HBsAg monoclonal antibodies, and ruthenium-labeled anti-HBsAg monoclonal antibodies. Sample types: serum, sodium heparin, EDTA-K3, sodium citrate plasma. Automated signal processing via photomultiplier. Shelf life: 11 months (immunoassay), 9 months (confirmatory), 12 months (controls). Storage: 2-8°C.
Indications for Use
Indicated for qualitative detection of HBsAg in human serum or plasma to aid in diagnosis of acute or chronic hepatitis B. Indicated for screening pregnant women to identify neonates at high risk of perinatal HBV acquisition. No known contraindications.
Regulatory Classification
Identification
A qualitative hepatitis B virus (HBV) antigen assay is identified as an in vitro diagnostic device intended for prescription use for qualitative use with human serum, plasma, or other matrices that aids in the diagnosis of chronic or acute HBV infection. HBV surface antigen (HbsAg) is also used for screening of HBV infection in pregnant women to identify neonates who are at risk of acquiring hepatitis B during perinatal period. The assay is not intended for screening of blood, plasma, cells, or tissue donors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate:
(A) The specimen types for which the device has been cleared, and that use of this assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results.
(B) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other populations where assay performance may be affected.
(C) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures.
(D) Detection of HBV antigens indicates a current infection with hepatitis B virus but does not differentiate between acute or chronic infection. False reactive HbsAg result may occur for up to 2 weeks after vaccination with HbsAg containing vaccine.
(E) Current methods for the detection of hepatitis B antigens may not detect all potentially infected individuals. A non-reactive assay result does not exclude the possibility of exposure to or infection with hepatitis B virus. A non-reactive assay result in individuals with prior exposure to hepatitis B may be due to but not limited to antigen levels below the detection limit of this assay or lack of antigen reactivity to the antibodies in this assay. HBV mutants lacking the ability to produce antigens have been reported. These may occur as “escape” mutants in the presence of anti-HBV antibodies and such patients may be infectious.
(F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay.
(2) Design verification and validation must include the following:
(i) A detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, design of the capture antibody(ies), external controls, and computational path from collected raw data to reported result (
*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable to the detection method and device design.(ii) For devices with assay calibrators, the design and composition of all primary, secondary, and subsequent quantitation standards used for calibration as well as their traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard). In addition, analytical testing must be performed following the release of a new lot of the standard material that was used for device clearance or approval, or when there is a transition to a new calibration standard.(iii) Documentation and characterization (
*e.g.,* supplier, determination of identity, purity, and stability) of all critical reagents (including description of the capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iv) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance.
(v) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability.
(vi) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range.
(vii) All stability protocols, including acceptance criteria.
(viii) Final release assay results for each lot used in clinical studies.
(ix) Reproducibility study data that includes the testing of three independent production lots.
(x) Detailed documentation of analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including, the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility) including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, prominent mutants/variants detection (
*e.g.,* for HbsAg), specimen stability, reagent stability, and cross-genotype antigen detection sensitivity, when appropriate.(xi) Analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays.
(xii) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(xiii) Detailed documentation and results from a clinical study. Performance must be analyzed relative to an FDA cleared or approved HBV antigen assay or a comparator that FDA has determined is appropriate. This study must be conducted using appropriate patient samples, with an appropriate number of HBV reactive and non-reactive samples in applicable risk and disease categories, and any applicable confirmatory testing. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas. This study must be conducted in the appropriate settings by the intended users to demonstrate clinical performance.
Related Devices
P160019 — Elecsys HBsAg II/Elecsys HBsAg Confirmatory Test/ PreciControl HBsAg II · Roche Diagnostics, Inc. · Dec 23, 2016
P030049 — ADVIA CENTAUR HBSAG READY PACK REAGENTS/CONFIRMATORY READY PACK REAGENTS/QUALITY CONTROL MATERIAL · Siemens Healthcare Diagnostics · May 26, 2005
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS
1. General Information
1.1. Name and Address of Applicant
Roche Diagnostics Corporation
9115 Hague Road
Indianapolis, IN 46256 USA
Establishment 1823260
1.2. Device Trade Name(s):
Elecsys® HBsAg Immunoassay
Elecsys® HBsAg Confirmatory Test
Elecsys® PreciControl HBsAg
1.3. Device Generic Names:
Hepatitis B Surface Antigen (HBsAg) Assay
Hepatitis B Surface Antigen (HBsAg) Control
Hepatitis B Surface Antigen (HBsAg) Confirmatory Kit
1.4. PMA Number:
P990012
1.5. Date of Panel Recommendation:
Pursuant to Section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not the subject of an FDA Immunology Devices Advisory Panel meeting because the information in the PMA substantially duplicated information previously reviewed by this Panel.
1.6. Date of Notice of Approval to Applicant:
June 1, 2001
2. Indications for Use
2.1. Elecsys® HBsAg Immunoassay
For the in vitro test qualitative detection of hepatitis B surface antigen (HBsAg) in human serum or plasma (sodium heparin, EDTA-K3 and sodium citrate).
Page 1 of 22
8
{1}
Assay results, in conjunction with other serological and clinical information may be used for the laboratory diagnosis of individuals with acute or chronic hepatitis B. This assay may be used to screen for hepatitis B infection in pregnant women to identify neonates who are at high risk of acquiring HBV during the perinatal period.
This assay is intended only for use on the Elecsys® 2010 immunoassay analyzer, and is based on electrochemiluminescence immunoassay technology (“ECLIA”).
## 2.2. Elecsys® PreciControl HBsAg
For the quality control of the Elecsys HBsAg Immunoassay and Elecsys HBsAg Confirmatory Test when testing human serum. The performance of the PreciControl HBsAg has not been established with any other HBsAg assay.
PreciControl HBsAg is not a calibrator and should not be used for assay calibration.
## 2.3. Elecsys® HBsAg Confirmatory Assay
For the in vitro qualitative confirmation of the presence of hepatitis B surface antigen in human serum or plasma (sodium heparin, EDTA-K3 and sodium citrate) found repeatedly reactive by the Elecsys HBsAg Immunoassay.
This assay is intended only for use on the Elecsys® 2010 immunoassay analyzer. The performance of the Elecsys HBsAg Confirmatory Test has not been established with any other HBsAg immunoassay.
## 3. Device Description
### 3.1. Principle of Device Methodology
#### 3.1.1. The Elecsys® HBsAg Immunoassay:
Is used for the measurement of hepatitis B surface antigen (HBsAg) in human serum and plasma (sodium heparin, EDTA-K3, sodium citrate). The assay is based on the principles of electrochemiluminescence.
In the first incubation, HBsAg in the sample, a biotinylated monoclonal HBsAg-specific antibody and a monoclonal HBsAg-specific antibody labeled with a ruthenium complex react to form a “sandwich” complex. After the addition of streptavidin-coated microparticles the complex becomes bound to the solid phase via interaction of biotin and streptavidin. After incubation the reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with a buffer. Application of a voltage to the electrode then induces chemiluminescent emission that is measured by a photomultiplier. Results are determined automatically by the Elecsys software by comparing the
Page 2 of 22
{2}
electrochemiluminescence signal obtained from the sample with the cutoff value previously obtained by HBsAg calibration. The total duration of the assay is 18 minutes.
### 3.1.2. The Elecsys HBsAg Confirmatory Test:
Uses the principle of specific antibody neutralization to confirm the presence of HBsAg. Samples found to be repeatedly reactive in the Elecsys HBsAg Immunoassay test are treated in parallel with the confirmatory reagent (containing anti-HBs) and control reagent. During incubation the excess anti-HBs antibodies in the confirmatory reagent neutralize any HBsAg in the sample. In the subsequent Elecsys HBsAg immunoassay test this leads to a reduction in the cutoff index (COI) value (signal of sample/COI) in comparison to the value originally obtained for the sample.
### 3.1.3. The Elecsys PreciControl HBsAg:
Contains human serum in the negative and positive concentration range that are used for monitoring the accuracy of Elecsys HBsAg immunoassay.
### 3.2. Kit Configuration and Component
#### 3.2.1. The Elecsys® HBsAg Immunoassay is composed of five reagents:
- The M reagent consists of Streptavidin coated microparticles (“beads”) in HEPES (4-(2-hydroxyethyl)-1-piperazine-ethanesulfonic acid) buffer with preservative.
- The R1 reagent, Anti-HBsAg-Ab-biotin, consists of purified, biotinylated anti-HBsAg antibody (mouse monoclonal directed against the “a” region determinant) in a phosphate buffer solution with preservative.
- The R2 reagent, Anti-HBsAg-Ab~Ru(bpy)$_{3}^{2+}$, consists of purified anti-HBsAg antibodies (mouse monoclonal directed against the “a” region determinant) labeled with ruthenium complex in a phosphate buffer solution with preservative.
- Cal 1, Negative Calibrator for five calibrations, consists of human serum, non-reactive for HBsAg, anti-HBs, anti-HCV, anti-HIV-1 and anti-HIV-2 with preservative.
- Cal 2, Positive Calibrator for five calibrations, consists of HBsAg 0.5 IU/mL (subtype ad) in human serum, non-reactive for anti-HBs, anti-HCV, anti-HIV-1 and anti-HIV-2 with preservative.
#### 3.2.2. The Elecsys PreciControl HBsAg contains two reagents
Page 3 of 22
{3}
- PreciControl 1, PC HBsAg 1, consists of human serum, non-reactive for HBsAg, anti-HBs, anti-HCV, anti-HIV-1 and anti-HIV-2 with preservative.
- PreciControl 2, PC HBsAg 2, consists of HBsAg (human, unspecified “a” subtype) at approximately 0.2 IU/mL in human serum, non-reactive for anti-HBs, anti-HCV, anti-HIV-1 and anti-HIV-2 with preservative.
### 3.2.3. The Elecsys HBsAg Confirmatory Test contains two reagents
- The confirmatory reagent, bottle 1, consists of anti-HBs (human) ≥ 200,000 IU/mL in human serum, non-reactive for HBsAg, anti-HCV, anti-HIV-1 and anti-HIV-2 with preservative.
- The control reagent, bottle 2, consists of human serum, non-reactive for HBsAg, anti-HBs, anti-HCV, anti-HIV-1 and anti-HIV-2 with preservative.
## 4. Contraindications
There are no known contraindications for the Elecsys® HBsAg Immunoassay.
## 5. Warnings and Precautions
For *in vitro* diagnostic used only.
Warnings and precautions are stated in the attached product labeling.
## 6. Alternative Practices and Procedures
There are currently several FDA approved and licensed in vitro diagnostic tests for serological markers of hepatitis B virus (HBV) infection when used in conjunction with a patient’s medical history, clinical examination, and other findings can be used for diagnostic purposes.
## 7. Prior Marketing History
The Elecsys® HBsAg Immunoassay, Elecsys HBsAg Confirmatory Test, and Elecsys PreciControl HBsAg has been marketed worldwide since 1998. The following list represents the countries where these devices have been marketed.
Page 4 of 22
//
{4}
| Argentina | Asia Pacific | Athens | Australia | Austria | Bahrain |
| --- | --- | --- | --- | --- | --- |
| Belarus | Belgium | Bosnia & Herzegovina | Brazil | Brazil | Budapest |
| Bulgaria | Canada | Cyprus | Egypt | Estonia | France |
| Germany | Hong Kong | Hungary | Iceland | Israel | Italy |
| Japan | Johannesburg | Jordan | Korea | Lebanon | Libyan Arab Jamahiriya |
| Lima | Lithuania | Luxembourg | Malta | Marokko | Mexico |
| Montevideo | Morocco | Moscow | Netherlands | New Zealand | Nigeria |
| Oman | Polska | Prague | Quito | Romania | Saudi Arabia |
| Slovenia | Spain | Spain | Switzerland | Syrian Arab Republic | Thailand |
| Tunisia | Turkey | UK | United Arab Emirates | Uruguay | Venezuela |
The device has not been withdrawn from marketing in any country for reasons relating to the safety and effectiveness of the device.
## 8. Potential Adverse Effects of the Device on Health
As an *in vitro* diagnostic, there is no direct adverse effect of the Elecsys HBsAg immunoassay test system on the health of the patient.
The possibility of erroneous test results due to test malfunctions or operator errors exists. An erroneously elevated test result could theoretically result in unnecessary treatment, however, as the Elecsys HBsAg Confirmatory Test is required for all repeatedly reactive test results, the likelihood of a false positive is low.
An erroneously nonreactive test result may lead to misdiagnosis, lack of treatment and a possible worsening of the patient’s condition, as well as potential for the spread of infection from the patient to uninfected persons.
It is recognized that presently available methods for the detection of hepatitis B surface antigen may not detect all potentially infected individuals. A nonreactive test result does not exclude the possibility of exposure to or infection with hepatitis B. Non-reactive results in individuals with prior exposure to hepatitis B may be due to antigen levels being below the detection limits of this assay or lack of antigen reactivity to the antibodies used in this assay. A falsely negative result for samples containing HBV mutants cannot be ruled out using the Elecsys HBsAg immunoassay. Testing using alternative methodologies may be warranted if signs, symptoms, and risk factors are indicative of viral hepatitis.
A false, transient positive result may occur in individuals recently vaccinated for hepatitis B because of its presence in the vaccine.
As with all tests containing monoclonal mouse antibodies, erroneous findings may be obtained from samples taken from patients who have been treated with monoclonal mouse antibodies or have received them for diagnostic purposes. In rare cases, interference due to extremely high titers of antibodies to streptavidin and ruthenium can occur.
Page 5 of 22
{5}
The risk of incorrect test results is inherent with all in vitro diagnostic products. Therefore, the above potential risks are not unusual in the laboratory setting. Appropriate warnings for each of these risks are contained in the labeling and package insert instructions. Standard good laboratory practices are considered sufficient to minimize risks to the end user.
## 9. Summary of Non-Clinical Studies
All studies were performed using the Elecsys 2010 Immunoassay Analyzer.
## 9.1. Analytical Sensitivity
The Elecsys HBsAg Immunoassay calibrators are standardized against the WHO 1st International Standard (subtype ad, Code 80/549, 1985). Paul Erlich Institute (PEI) standards were also measured by the Elecsys HBsAg Immunoassay and compared with the WHO standard.
The PEI reference material 1A (1992) showed the closest fit to the WHO reference material (1 E/ml PEI (1A) = 1.23 IU/ml WHO). Two analytical sensitivity studies were conducted during the non-clinical PMA studies to assess the sensitivity based upon the above assay standardization.
In the first study serial dilutions of reference standard materials were tested. The limit of detection was interpolated as the HBsAg concentration at which the test signal equaled the cutoff value (COI=1.0). Using the WHO 1st International Standard (subtype ad, 1985), the Elecsys HBsAg Immunoassay limit of detection (LOD) was 0.028 IU/mL (95% CI = 0.024 – 0.032 IU/mL). Similar results were obtained with the PEI standard (PEI reference material No.1, subtype ad), where the Elecsys limit of detection was 0.010 E/mL (95% CI = 0.009 – 0.011 E/mL).
## 9.2. Seroconversion Panels
A total of nine seroconversion panels from commercial vendors were tested at one of the clinical sites by the Elecsys HBsAg immunoassay and the reference HBsAg assay. The following table presents a summary of the Elecsys HBsAg immunoassay test system results for the nine panels in comparison to the reference method. Elecsys results were comparable to the FDA-licensed reference test.
Page 6 of 22
/3
{6}
| Panel No. | Days to HBsAg Reactor | Results of Dose | Difference in Days to HBsAg |
| --- | --- | --- | --- |
| Reference | HBgAg Assay | EIcsys HBgAg Assay | Relative Result** (Reference = EIcsys) |
| 01005 | 16 | 16 | 0 |
| 11004 | 26 | 26 | 0 |
| 40565L | 5 | 5 | 0 |
| 51005 | 0 | 0 | 0 |
| 21469D | 0 | 0 | 0 |
| 22663D | 17 | 17 * | 0 |
| PHM902 | 71 | 71 | 0 |
| PHM907 | 50 | 50 | 0 |
| PHM919 | 19 | 19 ** | 0 |
| PHM920 | 26 | 26 | 0 |
* Initial reactive results only, quantity not sufficient to perform confirmatory with either assay.
** Sample was reactive on Day 14 by Elecsys only, but without confirmation.
## 9.3. Antibody Specificity
A search of the “BLAST” database (http://www.ncbi.nlm.nih.gov/blast) for protein sequence homology to other than HBV was performed for the two mouse monoclonal antibodies (Mabs) used in the Elecsys 2010 HBsAg immunoassay. The biotinylated Mab was found to be highly specific to a conformational epitope of HBsAg “a” region determinant while the ruthenylated-Mab showed no relevant sequence homology with any known virus other than its designated target. The results of the search are interpreted as evidence that cross-reactivity by the kit Mabs to other non-HBV viruses is highly remote.
## 9.4. Matrix Effects:
Studies were conducted to verify the types of blood collection tubes that can be used with the Elecsys HBsAg Immunoassay. EDTA-K₃ plasma, sodium heparin plasma, sodium citrate plasma, and serum containing gel separator (serum separator tube, SST) were evaluated compared to serum. The format was to test 10 unspiked and 10 spiked with varying levels of HBsAg. The results of these studies demonstrated that serum collected using standard sampling tubes or tubes containing separating gel (SST’s) may be used in these assays. In addition plasma collected with sodium heparin, EDTA-K₃, or sodium citrate may also be used.
## 9.5. Endogenous Interference
Samples with abnormally elevated levels of hemoglobin, lipids, bilirubin, total protein, heparin, and biotin were simulated using patient samples (negative and HBsAg positive) spiked with the endogenous analyte of interest and compared to controls. The concentrations evaluated are summarized in the following table.
Page 7 of 22
{7}
| Endogenous Substance | Concentration Evaluated | Solvent | Reference Range |
| --- | --- | --- | --- |
| Hemoglobin | 1600 mg/dl | Serum | 0.5 - 5.0 mg/dl |
| Lipids | 1500 mg/dl | distilled/deionized water | 10 - 190 mg/dl |
| Bilirubin | 30 mg/dl | 0.1 N NaOH | 0.1 to 1.2 mg/dl |
| Total Protein | 12 g/dl | Serum | 6.0 - 7.8 g/dl |
| Heparin | 10 U/ml | Serum | 0.05 - 1.0 U/ml |
| Biotin | 50 ng/ml | 10 mM K₂PO₄ | 0.06 - 0.43 ng/ml |
The results of this study demonstrated that samples containing hemoglobin concentrations up to 1600 mg/dl, triglyceride levels up to 1500 mg/dl, bilirubin levels up to 30 mg/dl, total protein levels up to 12 g/dl, heparin levels up to 10 U/ml, or biotin levels up to 50 ng/ml may be tested accurately with the Elecsys HBsAg Immunoassay.
## 9.6. Carryover Study
A study using the Elecsys HBsAg Immunoassay was done to evaluate the effect of a sample highly reactive for HBsAg on a following negative sample. Two patient samples highly reactive for HBsAg were evaluated in this study. Sample 1, highly reactive for HBsAg (approximately 3000 COI) was tested in three sample cups (H1, H2, H3) followed by replicates of an HBsAg negative pool tested in five sample cups (L1, L2, L3, L4, L5). This sequence was repeated nine times in the same run for a total of ten high-low series. The study was repeated using a second HBsAg positive sample (Sample 2) with a COI of approximately 1900. The same testing sequence was performed as described for Sample 1. The results from this study showed that clinically significant carryover was not observed on the Elecsys HBsAg Immunoassay when testing highly positive HBsAg samples.
## 9.7. High Dose Hook Effect
Studies were run to assess the potential for a high dose hook effect (prozone effect). Samples with highly elevated levels of HBsAg were diluted with HBsAg negative serum. The diluted and undiluted samples were tested by the Elecsys HBsAg Immunoassay. A hook effect was observed for the Elecsys HBsAg Immunoassay, however samples with HBsAg concentrations up to 2,000,000 IU/mL were detected as reactive by the Elecsys HBsAg Immunoassay and did not produce false negative test results.
## 9.8. Establishment of Cutoff
The cutoff for the Elecsys 2010 HBsAg immunoassay was established by testing a 175-member panel of well characterized HBsAg specimens with three kit lots followed by testing on two kits lots with over 2,800 specimens including subjects at various risk for HBV infection, sensitivity, specificity and seroconversion panels and dilution series of both international reference materials and HBsAg positive samples. Using these data, the cutoff value that allowed for optimal discrimination between negative and positive specimens was determined.
{8}
Page 9 of 22
# 9.9. Stability Studies
To assess the real-time stability, whole kit samples were randomly selected from the individual lots of finished product. The kits and reagents were stored at the recommended storage temperature of $2-8^{\circ}\mathrm{C}$, in temperature-controlled cabinets, for the duration of the ongoing stability studies. Temperatures in the storage cabinets were checked at regular intervals. The test measurement intervals started with the production date of the last kit reagent in the released kits, and continued at approximate 1, 3, 6, 9, 12, 15, 18, 24, and 30 month time intervals. Key stability parameters monitored for the Elecsys HBsAg Immunoassay were analytical sensitivity, results of internal control samples and the assay “Test Dynamic” (defined as quotient of the COI of Cal 2 over the COI of Cal 1).
Based upon the results of real-time stability studies, the current shelf life for the Elecsys HBsAg Immunoassay is eleven (11) months, nine (9) months for the Elecsys HBsAg Confirmatory Test, and twelve (12) months for the Elecsys HBsAg PreciControls when the reagents are stored at $2-8^{\circ}\mathrm{C}$.
The stability of kit reagents after temperature stress conditions was examined in several studies using different stress models. The results from the temperature stress studies indicated stability for all Elecsys HBsAg Immunoassay, Confirmatory Test, and PreciControl reagents for at least 1 week at $35^{\circ}\mathrm{C}$. The recommended storage temperature for all the reagents is $2-8^{\circ}\mathrm{C}$.
Additional opened reagent and on-board stability studies were conducted which demonstrated the following:
- The Elecsys HBsAg Immunoassay and Elecsys HBsAg PreciControl kits may be used for 12 weeks once opened, when stored at $2-8^{\circ}\mathrm{C}$,
- The Elecsys HBsAg Confirmatory Test may be used for 8 weeks once opened, when stored at $2-8^{\circ}\mathrm{C}$,
- The Elecsys HBsAg Immunoassay reagent kit is stable on-board the Elecsys 2010 for 4 weeks,
- An Elecsys HBsAg Immunoassay calibrator set may be used for up to 5 calibration events and may stored on-board the Elecsys 2010 for up to 5 hours at $32^{\circ}\mathrm{C}$.
- An Elecsys HBsAg PreciControl set can be used for up to 7 quality control procedures and may be stored on-board the Elecsys 2010 for up to seven hours at $32^{\circ}\mathrm{C}$.
The results from the sample stability study demonstrated that serum samples containing HBsAg are stable for at least ten days when stored at $2-8^{\circ}\mathrm{C}$. The results of a second ongoing long-term sample stability study found that serum samples are stable for at least six months when stored at $-20^{\circ}\mathrm{C}$. In a third sample stability study (sample freeze/thaw experiment), it was shown that samples may be frozen at least six times without affecting Elecsys HBsAg Immunoassay results.
/6
{9}
Results from calibration stability studies found that a single Lot Calibration can be used for one month with multiple reagent packs, as long as the same reagent lot is used. A calibration on an individual reagent pack is stable for 7 days when using the same reagent kit stored on the analyzer.
## 9.10. Analytical Specificity
Patient samples with various disease states were evaluated to determine the analytical specificity of the Elecsys HBsAg Immunoassay. Comparisons were made between the Elecsys HBsAg Immunoassay final results and the determined HBV status using FDA licensed HBsAg testing. All three sites participated in this evaluation; however the analysis has been made with the data from all sites combined.
The table below summarizes the Elecsys HBsAg Immunoassay final test results compared to the HBV status for all sites combined.
Analytical Specificity Test Results: All Sites
| Elecsys® HBsAg Immunoassay | Neg | Neg | Pos | Pos | Total Samples |
| --- | --- | --- | --- | --- | --- |
| HBV Status | Neg | Pos | Neg | Pos | |
| Other Viral Hepatitis Infections | 46 | 0 | 0 | 0 | 46 |
| Other Infectious Diseases | 87 | 1 | 0 | 8 | 96 |
| Non-Viral Liver Diseases | 24 | 0 | 0 | 0 | 24 |
| Autoimmune Diseases | 19 | 0 | 0 | 1 | 20 |
| High Risk Populations | 30 | 0 | 0 | 1 | 31 |
| Post Influenza Vaccination | 5 | 0 | 0 | 0 | 5 |
| Other Diseases | 10 | 0 | 0 | 0 | 10 |
| Total | 221 | 1 | 0 | 10 | 232 |
The following sections present additional details to the table above by category.
Other Viral Hepatitis Infections: Elecsys Final Results
| Elecsys HBsAg Immunoassay | Neg | Neg | Pos | Pos | Total Samples |
| --- | --- | --- | --- | --- | --- |
| HBV Status | Neg | Pos | Neg | Pos | |
| Hepatitis A Infection | 4 | 0 | 0 | 0 | 4 |
| Hepatitis C Infection | 25 | 0 | 0 | 0 | 25 |
| Hepatitis A/ Hepatitis C Co-infection | 6 | 0 | 0 | 0 | 6 |
| Hepatitis E Infection | 11 | 0 | 0 | 0 | 11 |
| Total | 46 | 0 | 0 | 0 | 46 |
For patients with other viral hepatitis infections, 100% agreement was observed between Elecsys HBsAg Immunoassay Final Results and the HBV status (46/46).
Other Infectious Diseases: Elecsys Final Results
Page 10 of 22
{10}
| Elecsys HBsAg Immunoassay | Neg | Neg | Pos | Pos | Total Samples |
| --- | --- | --- | --- | --- | --- |
| HBV Status | Neg | Pos | Neg | Pos | |
| Cytomegalovirus | 10 | 0 | 0 | 0 | 10 |
| Epstein-Barr Virus | 10 | 0 | 0 | 0 | 10 |
| Herpes Simplex Virus | 13 | 1 | 0 | 0 | 14 |
| Parvovirus B19 | 7 | 0 | 0 | 3 | 10 |
| Rubella Virus | 10 | 0 | 0 | 0 | 10 |
| Human Immunodeficiency Virus | 5 | 0 | 0 | 4 | 9 |
| Syphilis | 10 | 0 | 0 | 0 | 10 |
| Toxoplasmosis | 12 | 0 | 0 | 1 | 13 |
| Escherichia coli | 10 | 0 | 0 | 0 | 10 |
| Total | 87 | 1 | 0 | 8 | 96 |
One sample from a patient with Herpes Simplex Virus (HSV) was discrepant between the Elecsys HBsAg Immunoassay and the HBV status. The Elecsys HBsAg Immunoassay results for this sample were borderline negative. The sample was confirmed positive with the reference HBsAg assay. Of the other HBV serological tests, Anti-HBc was positive; HBV DNA test was also positive. Liver function testing was not performed. This patient was a male, age 38, race unknown, with concurrent HIV infection. No other clinical information was available; a subsequent sample was not available for testing.
As there were thirteen other samples positive for HSV that were nonreactive by the Elecsys HBsAg Immunoassay, there does not appear to be a generalized specificity issue of the Elecsys HBsAg Immunoassay with HSV positive samples. Furthermore, of the nine samples from patients positive for HIV, all Elecsys HBsAg Immunoassay results were in agreement with the HBV status, also indicating excellent specificity with HIV positive samples. It is unclear why this sample was not detected as positive by the Elecsys HBsAg Immunoassay. This sample is considered to be a false negative by the Elecsys HBsAg Immunoassay.
For patients with other infectious diseases, 99.0% agreement was observed between Elecsys HBsAg Immunoassay Final Results and the HBV status (95/96).
Non-Viral Liver Disease: Elecsys Final Results
| Elecsys HBsAg Immunoassay | Neg | Neg | Pos | Pos | Total Samples |
| --- | --- | --- | --- | --- | --- |
| HBV Status | Neg | Pos | Neg | Pos | |
| Non-viral Liver Disease | 20 | 0 | 0 | 0 | 20 |
| Alcoholic Hepatitis | 4 | 0 | 0 | 0 | 4 |
| Total | 24 | 0 | 0 | 0 | 24 |
For patients with non-viral liver diseases, including alcoholic hepatitis, 100% agreement was observed between Elecsys HBsAg Immunoassay Final Results and the HBV status (24/24).
Autoimmune Diseases: Elecsys Final Results
Page 11 of 22
{11}
| Elecsys HBsAg Immunoassay | Neg | Neg | Pos | Pos | Total Samples |
| --- | --- | --- | --- | --- | --- |
| HBV Status | Neg | Pos | Neg | Pos | |
| Rheumatoid Factor | 9 | 0 | 0 | 1 | 10 |
| Antinuclear antibodies | 10 | 0 | 0 | 0 | 10 |
| Total | 19 | 0 | 0 | 1 | 20 |
For patients with autoimmune diseases, 100% agreement was observed between Elecsys HBsAg Immunoassay Final Results and the HBV status (20/20).
High Risk Populations: Elecsys Final Results
| Elecsys HBsAg Immunoassay | Neg | Neg | Pos | Pos | Total Samples |
| --- | --- | --- | --- | --- | --- |
| HBV Status | Neg | Pos | Neg | Pos | |
| Transplant Recipients | 11 | 0 | 0 | 0 | 11 |
| Chronic Dialysis | 10 | 0 | 0 | 0 | 10 |
| Intravenous Drug Users | 9 | 0 | 0 | 1 | 10 |
| Total | 30 | 0 | 0 | 1 | 31 |
For patients at high risk of acquiring HBV infection, 100% agreement was observed between Elecsys HBsAg Immunoassay Final Results and the HBV status (31/31).
Post Influenza Vaccination: Elecsys Final Results
| Elecsys HBsAg Immunoassay | Neg | Neg | Pos | Pos | Total Samples |
| --- | --- | --- | --- | --- | --- |
| HBV Status | Neg | Pos | Neg | Pos | |
| Post Influenza Vaccination | 5 | 0 | 0 | 0 | 5 |
For patients having recently received a vaccination for influenza, 100% agreement was observed between Elecsys HBsAg Immunoassay Final Results and the HBV status (5/5).
Other Diseases: Elecsys Final Results
| Elecsys HBsAg Immunoassay | Neg | Neg | Pos | Pos | Total Samples |
| --- | --- | --- | --- | --- | --- |
| HBV Status | Neg | Pos | Neg | Pos | |
| Gammopathies | 10 | 0 | 0 | 0 | 10 |
For patients with gammopathies, 100% agreement was observed between Elecsys HBsAg Immunoassay Final Results and the HBV status (10/10).
## 9.11. Reproducibility (Precision)
In a multi-center precision study with a design based on principles contained in the NCCLS draft guideline EP5-T2¹, results from a series of negative and positive samples run on the Elecsys 2010 analyzer at three centers over 20 days demonstrated a within run precision ranging from 1.3 to 7.5% CV. Between day precision, which also included within run and between run components, ranged from 4.6 to 16.3% CV; total precision, which included all precision components, ranged from 6.0 to 22.6% CV.
¹ National Committee for Clinical Laboratory Standards. Evaluation of Precision Performance of Clinical Chemistry Devices - Second Edition; Tentative Guideline. NCCLS document EP5-T2. NCCLS, 771 East Lancaster Avenue, Villanova, Pennsylvania 19085, 1992.
Page 12 of 22
{12}
Reproducibility Study Results for the Elecsys HBsAg Immunoassay
| Panel Member | Mean (COI) | Within Run | | Between Day* | | Total** | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV | SD | CV | SD | CV |
| 1 | 0.419 | 0.031 | 7.5% | 0.069 | 16.3% | 0.095 | 22.6% |
| 2 | 0.767 | 0.052 | 6.7% | 0.063 | 8.2% | 0.105 | 13.7% |
| 3 | 1.78 | 0.043 | 2.4% | 0.109 | 6.1% | 0.149 | 8.3% |
| 4 | 19.13 | 0.254 | 1.3% | 0.897 | 4.7% | 1.139 | 6.0% |
| 5 | 51.25 | 0.672 | 1.3% | 2.435 | 4.7% | 3.157 | 6.2% |
| 6 | 100.9 | 1.276 | 1.3% | 4.673 | 4.6% | 6.031 | 6.0% |
* Includes within run, between run and between day components.
** Includes within run, between run, between day, between site/ lot interaction, between lot and between site components.
Reproducibility of the manual test steps of the Elecsys HBsAg Confirmatory test was determined using three sera of differing HBsAg concentrations (8–10 times per sample with both the control and confirmatory reagents). After a 30-minute period of incubation at 20°C the pretreated samples were determined using Elecsys reagents, calibrators and controls. Representative data for manual sample pretreatment followed by assay on the Elecsys 2010 are shown below.
Intra-Assay Precision After Confirmatory Test Pretreatment
| Sample | Confirmatory Sample | | | Control Sample | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Mean (COI) | SD (COI) | % CV | Mean (COI) | SD (COI) | % CV |
| Human serum, COI < 7.0 | 0.42 | 0.04 | 9.5 | 1.57 | 0.04 | 2.6 |
| Human serum, COI 7.0 - ≤ 30 | 0.40 | 0.02 | 5.0 | 4.85 | 0.10 | 2.1 |
| Human serum, COI > 30 | 1.31 | 0.10 | 7.6 | 1321 | 12.8 | 1.0 |
Intra-Assay Precision Without Confirmatory Test Pretreatment
| Sample | Mean (COI) | SD (COI) | % CV |
| --- | --- | --- | --- |
| Human serum, COI < 7.0 | 1.65 | 0.04 | 2.4 |
| Human serum, COI 7.0 - ≤ 30 | 11.3 | 0.11 | 1.0 |
| Human serum, COI > 30 | 669 | 17.7 | 2.6 |
For the pretreated samples, coefficients of variation (CV) for the confirmatory samples ranged from 5.0 to 9.5%; CVs for the control samples ranged from 1.0 to 2.6%. The samples without the pretreatment procedure had CVs ranging from 1.0 to 2.6%.
## 10. Summary of Clinical Studies
### 10.1. Expected Results
Of 1445 prospective subjects participating in the Elecsys HBsAg clinical study, 41.5% (n = 600) were first time blood donors, asymptomatic for viral hepatitis. All of these subjects were enrolled in Sacramento, CA. The group was Caucasian (61%), African American (10%), Hispanic (2%), Asian (1%), and 26% electing not to provide this information. The group was 58% male and 42% female ranging in age from 17 to 73
Page 13 of 22
{13}
years. There were no confirmed positive results for HBsAg by either the reference or the Elecsys test system among these subjects.
The table below summarizes the Elecsys HBsAg negative and confirmed positive results by age range and gender.
| Age | HBAG Immunoassay System | | Test System | | Test System | |
| --- | --- | --- | --- | --- | --- | --- |
| | Gender | Body | Percent | Neg. | Percent | Total |
| < 10 | Male | 0 | NA | 0 | NA | 0 |
| | Female | 0 | NA | 0 | NA | 0 |
| 10 – 19 | Male | 0 | NA | 177 | 100 | 177 |
| | Female | 0 | NA | 115 | 100 | 115 |
| 20 – 29 | Male | 0 | NA | 71 | 100 | 71 |
| | Female | 0 | NA | 42 | 100 | 42 |
| 30 – 39 | Male | 0 | NA | 45 | 100 | 45 |
| | Female | 0 | NA | 46 | 100 | 46 |
| 40 – 49 | Male | 0 | NA | 35 | 100 | 35 |
| | Female | 0 | NA | 32 | 100 | 32 |
| 50 – 59 | Male | 0 | NA | 16 | 100 | 16 |
| | Female | 0 | NA | 13 | 100 | 13 |
| 60 – 69 | Male | 0 | NA | 2 | 100 | 2 |
| | Female | 0 | NA | 4 | 100 | 4 |
| 70 – 79 | Male | 0 | NA | 1 | 100 | 1 |
| | Female | 0 | NA | 1 | 100 | 1 |
| 80 – 89 | Male | 0 | NA | 0 | NA | 0 |
| | Female | 0 | NA | 0 | NA | 0 |
| 90 – 99 | Male | 0 | NA | 0 | NA | 0 |
| | Female | 0 | NA | 0 | NA | 0 |
| Unknown | Male | 0 | NA | 0 | NA | 0 |
| | Female | 0 | NA | 0 | NA | 0 |
| Total | Male | 0 | NA | 347 | 100 | 347 |
| | Female | 0 | NA | 253 | 100 | 253 |
| | All | 0 | NA | 600 | 100 | 600 |
The 845 remaining subjects were enrolled from populations considered at risk for viral hepatitis due to lifestyle or behavior. Of these, 448 were outpatients of a health screening clinic, 299 were hospitalized patients and 98 were IV drug users. All 98 IV drug users were enrolled in Baltimore, MD. Of the hospitalized and health screening clinic patients, 444 of the subjects were enrolled in Memphis, TN and 303 in Miami, FL. This collective group was African American (26%), Caucasian (19%), Hispanic (5%), Asian (<1%) or other (<1%) with 49% electing not to provide this information. The group was 49% male and 51% female ranging in age from 8 to 94 years. Six (6) of these subjects were confirmed positive by the both the reference and the Elecsys assay test systems. One specimen was confirmed positive by Elecsys but was not repeat reactive by the reference assay. A follow-up specimen taken 28 days later from the same subject was confirmed positive by both assays showing that the first Elecsys result was correct.
Page 14 of 22
{14}
The table below summarizes the Elecsys HBsAg negative and confirmed positive results by age range and gender.
| Age | Cies Elecsys HBsAg Immunoassay Test System | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Gender | Age | Percent | N° | Percent | Total |
| < 10 | Male | 0 | NA | 1 | 100 | 1 |
| | Female | 0 | NA | 0 | NA | 0 |
| 10 – 19 | Male | 0 | NA | 7 | 100 | 7 |
| | Female | 0 | NA | 10 | 100 | 10 |
| 20 – 29 | Male | 1 | 0.7 | 135 | 99.3 | 136 |
| | Female | 3 | 2.5 | 117 | 97.5 | 120 |
| 30 – 39 | Male | 0 | NA | 60 | 100 | 60 |
| | Female | 2 | 3.1 | 62 | 96.9 | 64 |
| 40 – 49 | Male | 0 | NA | 56 | 100 | 56 |
| | Female | 0 | NA | 65 | 100 | 65 |
| 50 – 59 | Male | 0 | NA | 53 | 100 | 53 |
| | Female | 0 | NA | 44 | 100 | 44 |
| 60 – 69 | Male | 0 | NA | 38 | 100 | 38 |
| | Female | 1 | 1.9 | 53 | 98.1 | 54 |
| 70 – 79 | Male | 0 | NA | 33 | 100 | 33 |
| | Female | 0 | NA | 46 | 100 | 46 |
| 80 – 89 | Male | 0 | NA | 12 | 100 | 12 |
| | Female | 0 | NA | 19 | 100 | 19 |
| 90 – 99 | Male | 0 | NA | 1 | 100 | 1 |
| | Female | 0 | NA | 3 | 100 | 3 |
| Unknown | Male | 0 | NA | 14 | 100 | 14 |
| | Female | 0 | NA | 8 | 100 | 8 |
| Total | Not Given | 0 | NA | 4 | 100 | 4 |
| | Male | 1 | 0.7 | 410 | 99.8 | 411 |
| | Female | 6 | 3.9 | 421 | 98.6 | 433 |
| All | 7 | 0.8 | 0.8 | 992 | 99.2 | 945 |
## 10.2. Clinical Performance
A multi-center prospective study was conducted to characterize the performance of the Elecsys HBsAg Immunoassay test system with individuals from defined populations. All subjects were tested using FDA-approved/cleared reference methods in strict accordance with the manufacturer’s package insert instructions. The collection sites for the specimens were located in Sacramento, CA (41.5%), Baltimore, MD (6.8%), Memphis, TN (30.7%) and Miami, FL (21.0%).
Of the 1445 prospective subjects participating in the Elecsys HBsAg clinical study, 41.5% (n=600) were first time blood donors, asymptomatic for viral hepatitis and 845 subjects were at risk of HBV infection due to lifestyle or behavior. Of the 845 at risk subjects, 53.0% (n=448) were outpatients of a health screening clinic, 35.4% (n=299) were hospitalized patients and 11.6%(n=98) were IV drugs users.
Page 15 of 22
{15}
The first time blood donors were Caucasian (61%), African American (10%), Hispanic (2%), Asian (1%) with 26% electing not to provide this information. The group was 58% male and 42% female ranging in age from 17 to 73 years. The at risk subjects were African American (26%), Caucasian (19%), Hispanic (5%), Asian (<1%) or other (<1%) with 49% electing not to provide this information. This group was 49% male and 51% female ranging in age from 8 to 94 years.
The performance of the Elecsys HBsAg immunoassay test system was analyzed relative to the reference HBsAg reported results for all 1445 specimens. Complete testing using FDA approved methods for all 6 HBV serological markers including HBsAg, HBeAg, anti-HBc, anti-HBc IgM, anti-HBe and anti-HBs, thus allowing cross-sectional serological classifications of HBV status, was available for 382 of the subjects.
HBV classifications were performed based on the constellation of test results from FDA-approved methodologies for various markers of HBV. Elecsys test results were not considered in these classifications. Presented below are the interpretations of HBV classification made for each of the serological profiles observed.
| HBV Classification | HBeAg | HBeAg | anti-HBc IgM | anti-HBc IgM | anti-HBe | anti-HBs |
| --- | --- | --- | --- | --- | --- | --- |
| Acute | pos | + or - | - | - | - | - |
| Acute | pos | + or - | pos | pos | + or - | - |
| Chronic* | pos > 6mo | | | | | |
| Chronic | pos | + or - | - | pos | + or - | + or - |
| Early Recovery | - | - | pos | pos | + or - | + or - |
| Recovery | - | - | - | pos | pos | + or - |
| Recovered | - | - | - | pos | - | + or - |
| Vaccinated | - | - | - | - | - | pos |
| not previously infected | - | - | - | - | - | - |
| Uninterpretable | pos | - | pos | pos | pos | pos |
| Uninterpretable | - | pos | - | - | - | - |
| Uninterpretable | - | pos | - | - | - | pos |
| Uninterpretable | - | pos | - | pos | - | pos |
| Uninterpretable | - | - | - | - | pos | pos |
| Uninterpretable | - | - | - | - | - | equiv |
* Subjects known, by testing, to have HBsAg persisting for greater than 6 months.
The following table compares the Elecsys HBsAg results with the reference results for the prospective studies with first time blood donors by HBV classification.
{16}
| HBV Classification | Final HBsAg Result Reference Test System | |
| --- | --- | --- |
| | Electro-HBsAg Result | Total |
| Acute | 0 | 0 |
| Chronic | 0 | 0 |
| Early Recovery | 1 | 1 |
| Recovery | 2 | 2 |
| Recovered | 1 | 1 |
| Uninterpretable | 1 | 1 |
| HBV Vaccine Response | 86 | 86 |
| Not Previously Infected | 31 | 31 |
| Incomplete Testing | 478 | 478 |
| Total | 600 | 600 |
The table below compares the Elecsys HBsAg results with the reference results for the prospective studies with subjects at risk for HBV infection due to lifestyle or behavior by HBV classification.
| HBV Classification | Final HBsAg Result Reference Test System | | | | Total |
| --- | --- | --- | --- | --- | --- |
| | Electro-HBsAg Result | Electro-HBsAg Result | Electro-HBsAg Result | Total | |
| Acute | 0 | 0 | 0 | 3 | 3 |
| Chronic | 0 | 0 | 0 | 2 | 2 |
| Early Recovery | 4 | 0 | 0 | 0 | 4 |
| Recovery | 36 | 0 | 0 | 0 | 36 |
| Recovered | 31 | 0 | 0 | 0 | 31 |
| Uninterpretable | 8 | 0 | 0 | 0 | 8 |
| HBV Vaccine Response | 140 | 0 | 0 | 0 | 140 |
| Not Previously Infected | 35 | 1* | 0 | 0 | 36 |
| Incomplete Testing | 584 | 0 | 0 | 1 | 585 |
| Total | 838 | 1 | 0 | 6 | 845 |
* This specimen was tested using the Roche Molecular Systems HBV Monitor Test and yielded a count of >9600 copies of HBV DNA/mL. A follow-up sample from the same patient taken 28 days after the first sample was confirmed positive on both the reference and Elecsys HBsAg immunoassay test systems.
The table below summarizes the percent agreement between the Elecsys HBsAg Immunoassay Test System and the HBsAg reference assay test system with first time blood donors by specimen classification.
{17}
| HBV Classification | Positive Percent Agreement | Positive Percent Agreement | Negative Percent Agreement | 95% Exact Confidence Interval |
| --- | --- | --- | --- | --- |
| Acute | NA | NA | NA | NA |
| Chronic | NA | NA | NA | NA |
| Early Recovery | NA | NA | 100 (1/1) | 2.5 to 100 |
| Recovery | NA | NA | 100 (2/2) | 15.9 to 100 |
| Recovered | NA | NA | 100 (1/1) | 2.5 to 100 |
| Uninterpretable | NA | NA | 100 (1/1) | 2.5 to 100 |
| HBV Vaccine Response | NA | NA | 100 (86/86) | 95.8 to 100 |
| Not Previously Infected | NA | NA | 100 (31/31) | 88.8 to 100 |
| Incomplete Testing | NA | NA | 100 (478/478) | 99.2 to 100 |
| Overall | NA | NA | 100 (600/600) | 99.4 to 100 |
The table below summarizes the percent agreement between the Elecsys HBsAg Immunoassay Test System and the HBsAg reference assay test system with subjects at risk for HBV infection due to lifestyle or behavior by specimen classification. The table also provides the upper and lower 95% Exact Confidence Intervals.
| HBV Classification | Positive Percent Agreement | 95% Exact Confidence Interval | Negative Percent Agreement | 95% Exact Confidence Interval |
| --- | --- | --- | --- | --- |
| Acute | 100 (3/3) | 29.5 to 100 | NA | NA |
| Chronic | 100 (2/2) | 15.9 to 100 | NA | NA |
| Early Recovery | NA | NA | 100 (4/4) | 40.0 to 100 |
| Recovery | NA | NA | 100 (36/36) | 90.3 to 100 |
| Recovered | NA | NA | 100 (31/31) | 88.8 to 100 |
| Uninterpretable | NA | NA | 100 (8/8) | 63.2 to 100 |
| HBV Vaccine Response | NA | NA | 100 (140/140) | 97.4 to 100 |
| Not Previously Infected | NA | NA | 97.2 (35/36) | 85.5 to 100 |
| Incomplete Testing | 100 (1/1) | 2.5 to 100 | 100 (584/584) | 99.3 to 100 |
| Overall | 100 (6/6) | 54.1 to 100 | 99.9 (111/111) | 99.5 to 100 |
The performance of the Elecsys HBsAg Immunoassay test system was studied with archived specimens representing various discrete stages of HBV infection or recovery. The table below compares the Elecsys HBsAg immunoassay test system results with the HBsAg reference assay test system results by specimen classification.
{18}
| HBs-CI application | Elective | HBsAg | Reagent | Reagent | Total |
| --- | --- | --- | --- | --- | --- |
| Acute | 0 | 0 | 0 | 151 | 151 |
| Chronic | 0 | 0 | 0 | 111 | 111 |
| Chronic – HBsAg >6 mo.^{a} | 0 | 0 | 0 | 74 | 74 |
| Early Recovery | 4 | 1^{b} | 0 | 0 | 5 |
| Recovery | 35 | 0 | 0 | 0 | 35 |
| Recovered | 24 | 0 | 0 | 0 | 24 |
| Uninterpretable | 4 | 0 | 0 | 1 | 5 |
| HBV Vaccine Response | 6 | 0 | 0 | 0 | 6 |
| Not Previously Infected | 9 | 0 | 0 | 0 | 9 |
| Total | 12 | 1 | 0 | 307 | 420 |
a Subjects known, by testing, to have HBsAg persisting for greater than 6 months.
b This specimen by the reference method was repeatedly within 10% of cut-off and showed 100% neutralization but was classified as negative in accordance with reference test’s approved package insert instructions since the initial test was below cut-off at 0.94 COI.
The following table summarizes the percent agreement between the Elecsys HBsAg Immunoassay Test System and the HBsAg reference assay test system by specimen classification, and provides the upper and lower 95% Exact Confidence Intervals.
| HBs-CI application | Elective | HBsAg | Reagent | Reagent |
| --- | --- | --- | --- | --- |
| Acute | 100 (151/151) | 97.6 to 100 | NA | NA |
| Chronic | 100 (111/111) | 96.7 to 100 | NA | NA |
| Chronic – HBsAg >6 mo | 100 (74/74) | 95.1 to 100 | NA | NA |
| Early Recovery | NA | NA | 80.0 (4/5) | 28.6 to 99.5 |
| Recovery | NA | NA | 100 (35/35) | 90.0 to 100 |
| Recovered | NA | NA | 100 (24/24) | 85.8 to 100 |
| Uninterpretable | 100 (1/1) | 2.5 to 100.0 | 100 (4/4) | 40.0 to 100 |
| HBV Vaccine Response | NA | NA | 100 (6/6) | 54.3 to 100 |
| Not Previously Infected | NA | NA | 100 (9/9) | 66.5 to 100 |
| Overall | 100 (337/337) | 98.9 to 100 | 98.8 (82/83) | 95.5 to 100 |
## 10.3. Clinical Performance in Pregnant Women
A total of 81 pregnant women were reported from among all subjects in the Elecsys 2010 HBsAg clinical study. All were considered to be at low risk for HBV infection. The table below summarizes the Elecsys 2010 HBsAg final result in comparison to the reference method by age and ethnic group. Information on trimester stage was not available.
{19}
| | | | Reference HBsAg
Assay Final Result | | Elecsys 2010 HBsAg
Final Result | |
| --- | --- | --- | --- | --- | --- | --- |
| Ethnicity | Age | N | Positive | Negative | Positive | Negative |
| African-American | 10 - 19 | 5 | 0 | 5 | 0 | 5 |
| | 20 - 29 | 8 | 1 | 7 | 1 | 7 |
| | 30 - 39 | 2 | 0 | 2 | 0 | 2 |
| | 40 - 49 | 0 | 0 | 0 | 0 | 0 |
| Caucasian | 10 - 19 | 1 | 0 | 1 | 0 | 1 |
| | 20 - 29 | 7 | 0 | 7 | 0 | 7 |
| | 30 - 39 | 8 | 0 | 8 | 0 | 8 |
| | 40 - 49 | 1 | 0 | 1 | 0 | 1 |
| Not Provided | 10 - 19 | 3 | 0 | 3 | 0 | 3 |
| | 20 - 29 | 29 | 1 | 28 | 1 | 28 |
| | 30 - 39 | 15 | 1 | 14 | 1 | 14 |
| | 40 - 49 | 1 | 0 | 1 | 0 | 1 |
| | not given | 1 | 0 | 1 | 0 | 1 |
| Overall | | 81 | 3
(3.7%) | 78
(96.3%) | 3
(3.7%) | 78
(96.3%) |
The table below summarizes the percent agreement between the Elecsys HBsAg Immunoassay Test System and the HBsAg reference assay test system with pregnant women. The table also provides the upper and lower 95% Exact Confidence Intervals.
| Subject | Positive
Percent
Agreement | 95% Exact
Confidence
Interval | Negative
Percent
Agreement | 95% Exact
Confidence
Interval |
| --- | --- | --- | --- | --- |
| Pregnant Women | 100 (3/3) | 29.5 to 100 | 100 (78/78) | 95.4 to 100 |
## 10.4. Elecsys HBsAg Confirmatory Test
The Elecsys HBsAg Confirmatory Test was evaluated with all specimens throughout the Elecsys HBsAg clinical study that had been confirmed positive by the FDA-licensed reference HBsAg confirmatory assay. The table below compares the Elecsys HBsAg Confirmatory Test results with the reference HBsAg confirmed positive result.
Page 20 of 22
{20}
| Section | N | Elecsys HBsAg
Elecsys®
Immunoassay and
Neutralization | Elecsys HBsAg
Conformatory Test
Conformity Positive by
Neutralization |
| --- | --- | --- | --- |
| First time blood donors | 0 | NA | NA |
| Subjects at risk for HBV infection due to lifestyle or behavior | 6 | 6 | 6 |
| Serologically classified acute HBV infection (archived) | 151 | 151 | 151 |
| Serologically classified chronic HBV infection (archived) | 111 | 111 | 111 |
| Chronic HBV infection, HBsAg for > 6 months (archived) | 74 | 74 | 74 |
| Pregnant women | 3 | 3 | 3 |
| Total | 348 | 345 | 345 (98%) |
## 11. Conclusions Drawn from PMA Studies
In multi-centered clinical trials in the United States, the Elecsys HBsAg Immunoassay and Elecsys HBsAg Confirmatory Test were shown to exhibit clinical sensitivity and specificity that correlate well to other commercially available similar licensed devices.
The clinical evaluations provide valid scientific evidence of the clinical and diagnostic use of the automated Elecsys HBsAg Immunoassay for the qualitative determination of HBsAg in human serum and plasma (sodium heparin, EDTA-K3 and sodium citrate). The results from these studies also support the clinical and diagnostic performance of the Elecsys HBsAg Confirmatory test (a neutralization test), and the Elecsys HBsAg PreciControl reagents.
The PMA studies provide reasonable assurance that the Elecsys HBsAg Immunoassay can safely and effectively be used to test individuals at various risk for HBV infection and can be used as an aid in the diagnosis of patients with acute or chronic HBV infection. These studies establish the use of the assay to screen for hepatitis B infection in pregnant women to identify neonates at high risk of acquiring HBV during the perinatal period (see caution in the Risk/Benefit section below).
The studies support the use of Elecsys HBsAg Confirmatory Test as an immunoassay for the in vitro confirmation of the presence of HBsAg in human serum and plasma (sodium heparin, EDTA-K3, sodium citrate) samples found repeatedly reactive when tested with Elecsys HBsAg Immunoassay.
## 12. Safety and Benefit/Risk Analysis
As a diagnostic test, the HBsAg assay involves removal of blood from an individual for testing purposes. The test, therefore, presents no more safety hazard to an individual being tested than other tests where blood is removed.
The benefits to HBV-infected individuals tested by these devices outweigh any potential adverse event or risk to the patient or user due to device malfunction or operator error. The
Page 21 of 22
28
{21}
use of the Elecsys HBsAg Confirmatory test adds a further safety measure for the accurate reporting of confirmed positive test results.
The potential risks seen for *in vitro* diagnostic tests are not unusual in the laboratory setting, and appropriate warnings for these risks are contained in the labeling and package insert instructions for these devices. Standard good laboratory practices are considered sufficient to minimize the risks to the end user.
The clinical studies minimally establish the use of the assay to screen pregnant women to identify neonates at high risk of acquiring HBV during the perinatal period. Users may wish to establish their own specificity for prenatal screening.
## 13. CDRH Decision on application:
FDA issued an approval order on June 1, 2001.
The applicant’s manufacturing facility was inspected on _______________ and found to be in compliance with the devices Good Manufacturing Practice regulation.
## 14. Approval Specifications:
Directions for Use: See labeling
Conditions of Approval: CDRH approval of this PMA is subject to full compliance with the conditions described in the approval order.
Postapproval Requirements and Restrictions: See approval order.
Page 22 of 22
29
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.