BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel Mini
K241194 · Biofire Diagnostics, LLC · QOF · May 30, 2024 · Microbiology
Device Facts
| Record ID | K241194 |
| Device Name | BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel Mini |
| Applicant | Biofire Diagnostics, LLC |
| Product Code | QOF · Microbiology |
| Decision Date | May 30, 2024 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 866.3981 |
| Device Class | Class 2 |
Intended Use
The BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel Mini is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE® System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19; (Respiratory menu) or in throat swab (TS) specimens from individuals with signs and symptoms of pharyngitis (Sore Throat menu).
Device Story
The device is a multiplexed PCR-based in vitro diagnostic test for infectious disease. It processes nasopharyngeal or throat swab specimens loaded into a single-use pouch. The BIOFIRE SPOTFIRE System performs automated nucleic acid extraction via mechanical lysis and magnetic bead purification, followed by a two-stage nested multiplex PCR. The first stage is a large-volume rt-PCR; the second stage is a singleplex array-based PCR. The system uses Peltier devices for thermal cycling and melt curve analysis to detect signature amplicons. A digital camera captures fluorescent images of the array, and software interprets the results. Used in clinical settings by healthcare providers, the device provides rapid (approx. 15 min) qualitative results to aid in the diagnosis of respiratory infections and pharyngitis. It is intended to be used in conjunction with other clinical and epidemiological information. The device benefits patients by providing rapid, actionable diagnostic information to guide clinical management.
Clinical Evidence
No new performance data were reviewed for this modification. The determination of substantial equivalence is based on design control activities, risk analysis, and verification/validation of software modifications.
Technological Characteristics
Multiplexed nested PCR with melt curve analysis. Instrumentation: BIOFIRE SPOTFIRE System. Reagents: Freeze-dried, room-temperature stable. Connectivity: Standalone system. Software: Automated interpretation of fluorescence/temperature data. Sterilization: Not specified.
Indications for Use
Indicated for the detection and identification of nucleic acid targets in respiratory specimens from microbial agents causing SARS-CoV-2 and other respiratory infections, and Streptococcus spp. nucleic acid-based assay.
Regulatory Classification
Identification
A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
*e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
*e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
*e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
*e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
*i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.
Predicate Devices
- BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel (K232954/CW230018)
Related Devices
- K213954 — BIOFIRE SPOTFIRE Respiratory (R) Panel · Biofire Diagnostics · Feb 3, 2023
- K232954 — BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel · Biofire Diagnostics, LLC · Mar 26, 2024
- K243544 — BIOFIRE SPOTFIRE Respiratory/Sore Throat Panel Mini · Biofire Diagnostics, LLC · Aug 14, 2025
- K143080 — FilmArray Respiratory Panel (RP) for use with Multi-instrument FilmArray System · Biofire Diagnostics, LLC · Feb 17, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
May 30th, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
BioFire Diagnostics, LLC Karli Plenert Directory of Regulatory Affairs 515 Colorow Drive Salt Lake City, Utah 84108
Re: K241194
Trade/Device Name: BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel Mini Regulation Number: 21 CFR 866.3981 Regulation Name: Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The Sars-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test Regulatory Class: Class II Product Code: QOF, OZE, OCC, PGX, NSU Dated: April 29, 2024 Received: April 30, 2024
Dear Karli Plenert:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joseph Briggs -S
Joseph Briggs, Ph.D. Deputy Branch Chief
{2}------------------------------------------------
Viral Respiratory and HPV Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241194
#### Device Name
BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel Mini
#### Indications for Use (Describe)
The BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel Mini is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE® System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19; (Respiratory menu) or in throat swab (TS) specimens from individuals with signs and symptoms of pharyngitis (Sore Throat menu).
The following analytes are identified and differentiated using the SPOTFIRE R/ST Panel Mini:
| <b>Respiratory Menu:</b> |
|-----------------------------|
| Viruses |
| Coronavirus SARS-CoV-2 |
| Human rhinovirus |
| Influenza A virus |
| Influenza B virus |
| Respiratory syncytial virus |
Sore Throat Menu: Viruses Human rhinovirus Influenza A virus Influenza B virus Respiratory syncytial virus Bacteria Streptococcus pyogenes (group A Strep)
Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS/TS specimens during the acute phase of infection. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection and/or pharyngitis is indicative of the presence of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
Negative results in the setting of a respiratory illness and/or pharyngitis may be due to infection with pathogens that are not detected by this test, or a respiratory tract infection that may not be detected by an NPS or TS specimen. Positive results do not rule out co-infection with other organisms. The agent(s) detected by the SPOTFIRE R/ST Panel Mini may not be the definite cause of disease.
Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection and/or pharyngitis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue top half and a yellow-green gradient bottom half. The word "BIOMÉRIEUX" is written in white, sans-serif font across the center of the circle.
# BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel Mini for use on the BIOFIRE® SPOTFIRE® System
# 510(k) Summary BioFire Diagnostics, LLC
#### Introduction:
The purpose of this Special 510(k) submission is to obtain clearance for the BIOFIRE® Respiratory/Sore Throat (R/ST) Panel Mini (SPOTFIRE R/ST Panel Mini). The SPOTFIRE R/ST Panel Mini is compatible with the BIOFIRE® SPOTFIRE® System, a polymerase chain reaction (PCR)-based in vitro diagnostic system for infectious disease testing (K213954).
The SPOTFIRE R/ST Panel Mini is an identical product to the BIOFIRE® Respiratory/Sore Throat (R/ST) Panel (K232954) except it uses modified labeling and modified software to report only five of the 15 targets on the Respiratory Menu and only five of the 14 targets on the Sore Throat Menu.
Modifications to the SPOTFIRE R/ST Mini Panel labeling, which includes changes to the Instructions for Use and Quick Guide, have been made to reflect the change in panel name and reported analytes.
According to the requirements of 21 CFR 807.92, the information included with this submission provides sufficient detail to understand the basis for a determination of substantial equivalence.
#### Submitted by:
BioFire Diagnostics, LLC (bioMérieux) 515 Colorow Drive Salt Lake City, UT 84108
Contact: Karli Plenert, MBA Telephone: 385-414-4985 Email: Karli.Plenert@biomerieux.com
Date submitted: April 29, 2024
## Device Name and Classification:
Trade name: BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel Mini
Primary Requlation Number for Device Classification: 21 CFR 866.3981
Regulation Number: 21 CFR 866.2680
Classification Name: Multi-Target Respiratory Specimen Nucleic Acid Test Including SARS-CoV-2 And Other Microbial Agents
Additional Regulation Numbers: 21 CFR 866.2680, 21 CFR 866.3980, 21 CFR 862.2570
{6}------------------------------------------------
## Predicate Device:
K232954/CW230018 – BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel
#### Intended Use:
The BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel Mini is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE® System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19; (Respiratory menu) or in throat swab (TS) specimens from individuals with signs and symptoms of pharyngitis (Sore Throat menu).
The following analytes are identified and differentiated using the SPOTFIRE R/ST Panel Mini:
| Respiratory Menu | Sore Throat Menu |
|-----------------------------|----------------------------------------|
| Viruses | Viruses |
| Coronavirus SARS-CoV-2 | Human rhinovirus |
| Human rhinovirus | Influenza A virus |
| Influenza A virus | Influenza B virus |
| Influenza B virus | Respiratory syncytial virus |
| Respiratory syncytial virus | Bacteria |
| | Streptococcus pyogenes (group A Strep) |
Table 1. SPOTFIRE R/ST Panel Mini Analytes
Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS/TS specimens during the acute phase of infection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection and/or pharyngitis is indicative of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
Negative results in the setting of a respiratory illness and/or pharyngitis may be due to infection with pathogens that are not detected by this test, or a respiratory tract infection that may not be detected by an NPS or TS specimen. Positive results do not rule out co-infection with other organisms. The agent(s) detected by the SPOTFIRE R/ST Panel Mini may not be the definite cause of disease.
Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection and/or pharyngitis.
## Device Description:
The SPOTFIRE R/ST Panel Mini simultaneously identifies 5 different respiratory viral pathogens in nasopharyngeal swabs (NPS) or 5 different viral and bacterial pharyngitis pathogens in throat swabs (TS) from individuals with signs and symptoms of respiratory tract infections or pharyngitis, respectively, (see Table 1) The SPOTFIRE R/ST Panel Mini is compatible with the BIOFIRE® System, a polymerase chain reaction (PCR)-based in vitro diagnostic system for infectious disease testing. The BIOFIRE System Sottware executes the SPOTFIRE R/ST Panel Mini test and interprets and reports the test results. The SPOTFIRE R/ST Panel Mini was designed to be used in CLIA-waived environments.
A test is initiated by loading Hydration Solution injection solution injection port of the SPOTFIRE R/ST Panel Mini pouch and NPS or TS specimen, mixed with the provided Sample injection port of the SPOTFIRE R/ST Panel Mini pouch and placing it in the SPOTFIRE System. The reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the SPOTFIRE System Software guides the user through the steps of placing the pouch into the instrument, scanning the sample identification, and initiating the run.
The SPOTFIRE System contains coordinated systems of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid
{7}------------------------------------------------
from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.
Nucleic acid extraction occurs within the SPOTFIRE R/ST Panel Mini pouch using mechanical Ivsis followed by purfication using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the SPOTFIRE System performs a nested multiplex PCR that is executed in two stage. During the first stage, the SPOTFIRE System performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent doublestranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.
The SPOTFIRE System Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the SPOTFIRE R/ST Panel Mini.
## Substantial Equivalence:
The SPOTFIRE R/ST Panel Mini is substantially equivalent to the SPOTFIRE R/ST Panel (K232954/CW230018), which was cleared and CLIA-waived on March 26, 2024, and determined to be a Class II device under the classification code 21 CFR 866.3981.
A comparision of the SPOTFIRE R/ST Panel Mini to the SPOTFIRE R/ST Panel is provided in Table 2.
| Element | Predicate: SPOTFIRE R/ST Panel<br>(K232954/CW230018) | New Device: SPOTFIRE R/ST Panel Mini |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The BIOFIRE® SPOTFIRE® Respiratory/Sore<br>Throat Panel (SPOTFIRE R/ST Panel) is a<br>multiplexed polymerase chain reaction (PCR) test<br>intended for use with the BIOFIRE® SPOTFIRE®<br>System for the simultaneous, qualitative detection<br>and identification of multiple respiratory viral and<br>bacterial nucleic acids in nasopharyngeal swab<br>(NPS) specimens obtained from individuals with<br>signs and symptoms of respiratory tract infection,<br>including COVID-19; (Respiratory menu) or in throat<br>swab (TS) specimens from individuals with signs<br>and symptoms of pharyngitis (Sore Throat menu).<br>The following analytes are identified and<br>differentiated using the SPOTFIRE R/ST Panel:<br><br><b>Respiratory Menu</b><br><i>Viruses</i><br>Adenovirus<br>Coronavirus SARS-CoV-2<br>Coronavirus (seasonal)<br>Human metapneumovirus<br>Human rhinovirus/enterovirus<br>Influenza A virus<br>Influenza A virus A/H1-2009<br>Influenza A virus A/H3<br>Influenza B virus<br>Parainfluenza virus<br>Respiratory syncytial virus<br><br><i>Bacteria</i><br>Bordetella parapertussis<br>Bordetella pertussis<br>Chlamydia pneumoniae | Same except the following analytes are identified and<br>differentiated using the SPOTFIRE R/ST Panel Mini:<br><br><b>Respiratory Menu</b><br><i>Viruses</i><br>Coronavirus SARS-CoV-2<br>Human rhinovirus<br>Influenza A virus<br>Influenza B virus<br>Respiratory syncytial virus<br><br><b>Sore Throat Menu</b><br><i>Viruses</i><br>Human rhinovirus<br>Influenza A virus<br>Influenza B virus<br>Respiratory syncytial virus<br><br><i>Bacteria</i><br>Streptococcus pyogenes (group A Strep) |
| Mycoplasma pneumoniae | | |
| Sore Throat Menu | | |
| | Viruses | |
| | Adenovirus<br>Coronavirus (seasonal)<br>Human metapneumovirus<br>Human rhinovirus/enterovirus<br>Influenza A virus<br>Influenza A virus A/H1-2009<br>Influenza A virus A/H3<br>Influenza B virus<br>Parainfluenza virus<br>Respiratory syncytial virus | |
| | Bacteria | |
| | Chlamydia pneumoniae<br>Mycoplasma pneumoniae<br>Streptococcus dysgalactiae (group C/G Strep)<br>Streptococcus pyogenes (group A Strep) | |
| | Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS/TS specimens during the acute phase of infection. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection and/or pharyngitis is indicative of the presence of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. | |
| | Negative results in the setting of a respiratory illness and/or pharyngitis may be due to infection with pathogens that are not detected by this test, or a respiratory tract infection that may not be detected by an NPS or TS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected by the SPOTFIRE R/ST Panel may not be the definite cause of disease. | |
| | Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection and/or pharyngitis. | |
| Specimen Types | Nasopharyngeal swab in transport media<br>or<br>Throat swab in transport media | Same |
| | Viruses | Viruses |
| | Adenovirus<br>Coronavirus SARS-CoV-2<br>Coronavirus (seasonal)<br>Human metapneumovirus<br>Human rhinovirus/enterovirus<br>Influenza A virus<br>Influenza A virus A/H1-2009<br>Influenza A virus A/H3<br>Influenza B virus<br>Parainfluenza virus<br>Respiratory syncytial virus | Coronavirus SARS-CoV-2 (Respiratory Menu only)<br>Human rhinovirus<br>Influenza A virus<br>Influenza B virus<br>Respiratory syncytial virus |
| Organisms detected | Bacteria<br>Chlamydia pneumoniae<br>Mycoplasma pneumoniae<br>Bordetella parapertussis<br>Bordetella pertussis<br>Streptococcus dysgalactiae (group C/G Strep)<br>Streptococcus pyogenes (group A Strep) | Bacteria<br>Streptococcus pyogenes (group A Strep)<br>(Sore Throat Menu only) |
Table 2. Similarities and differences between the SPOTFIRE R/ST Panel and the SPOTFIRE R/ST Panel Mini
{8}------------------------------------------------
{9}------------------------------------------------
| Technological Principles | Highly multiplexed nested nucleic acid amplification<br>test with melt analysis | Same |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Instrumentation | SPOTFIRE System | Same |
| Time to result | About 15 minutes | Same |
| Reagent Storage | Room Temperature | Same |
| Test Interpretation | Automated test interpretation and reporting. User<br>cannot access raw data. | Same |
| Controls | Two controls are included in each reagent pouch to<br>control for sample processing and both stages of<br>PCR and melt analysis. | Same |
| User complexity | Low (CLIA-waived) | Moderate (Intending to seek CLIA Waiver following<br>clearance of the SPOTFIRE R/ST Panel Mini) |
| Panel Software Functions | Defines panel-specific parameters, instrument<br>protocols and report requirements. | Same |
| | Analyzes processed image data (fluorescence and<br>temperature data) and provides test results. | Same |
#### Summary of Performance Data:
The performance data for the SPOTFIRE R/ST Panel Mini is summarized in the B/OF/RE SPOTFIRE Respiratory/Sore Throat Panel Mini Instructions for Use. A summary of the R/ST Panel Mini performance data is also provided below.
#### Clinical Performance
Table 3 and Table 4 provide a summary of the performance of each analyte from all studies performed for NPS and TS specimens, respectively.
| SPOTFIRE R/ST Panel<br>R Menu Result | Study | Positive Percent Agreement | | | Negative Percent Agreement | | |
|--------------------------------------|-------------|----------------------------|------|------------|----------------------------|------|------------|
| | | TP/(TP + FN) | % | 95%CI | TN/(TN + FP) | % | 95%CI |
| Viruses | | | | | | | |
| Coronavirus SARS-CoV-2 | Prospective | 71/73 | 97.3 | 90.5-99.2% | 1031/1037 | 99.4 | 98.7-99.7% |
| | Archived | 0/0 | - | - | 0/0 | - | - |
| | Contrived | 0/0 | - | - | 0/0 | - | - |
| Human rhinovirus | Prospective | 345/348 | 99.1 | 97.5-99.7% | 695/767 | 90.6 | 88.3-92.5% |
| | Archived | 29/30 | 96.7 | 83.3-99.4% | 439/454 | 96.7 | 94.6-98.0% |
| | Contrived | 0/0 | - | - | 0/0 | - | - |
| Influenza A virus | Prospective | 0/0 | - | - | 1115/1115 | 100 | 99.7-100% |
| | Archived | 58/59 | 98.3 | 91.0-99.7% | 423/423 | 100 | 99.1-100% |
| | Contrived | 0/0 | - | - | 0/0 | - | - |
| Influenza B virus | Prospective | 0/0 | - | - | 1110/1110 | 100 | 99.7-100% |
| | Archived | 30/30 | 100 | 88.6-100% | 28/28 | 100 | 87.9-100% |
| | Contrived | 0/0 | - | - | 0/0 | - | - |
| Respiratory syncytial virus | Prospective | 26/27 | 96.3 | 81.7-99.3% | 1086/1088 | 99.8 | 99.3-100% |
| | Archived | 37/37 | 100 | 90.6-100% | 440/447 | 98.4 | 96.8-99.2% |
| | Contrived | 0/0 | - | - | 0/0 | - | - |
Table 3. R/ST Panel Mini Performance Summary for NPS Specimens (Respiratory Menu; as tested by intended users)
#### Table 4. R/ST Panel Mini Performance Summary for TS Specimens (Sore Throat Menu; as tested by intended users)
| SPOTFIRE R/ST Panel | Study | Sensitivity/PPA | | | Specificity/NPA | | |
|---------------------|-------------|-----------------|------|------------|-----------------|------|------------|
| ST Menu Result | | TP/(TP + FN) | % | 95%CI | TN/(TN + FP) | % | 95%CI |
| Viruses | | | | | | | |
| Human rhinovirus | Prospective | 202/213 | 94.8 | 91.0-97.1% | 619/662 | 93.5 | 91.4-95.1% |
| | Archived | 2/2 | 100 | 34.2-100% | 55/57 | 96.5 | 88.1-99.0% |
| | Contrived | 0/0 | - | - | 0/0 | - | - |
{10}------------------------------------------------
| SPOTFIRE R/ST Panel<br>ST Menu Result | Study | Sensitivity/PPA | | | Specificity/NPA | | | |
|-------------------------------------------|-------------|-----------------|---------|------------|-----------------|---------|------------|------------|
| | | TP/(TP + FN) | % | 95%CI | TN/(TN + FP) | % | 95%CI | |
| Influenza A virus | Prospective | 35/35 | 100 | 90.1-100% | 840/840 | 100 | 99.5-100% | |
| Influenza A virus | Archived | 11/11 | 100 | 74.1-100% | 44/44 | 100 | 92.0-100% | |
| Influenza A virus | Contrived | 93/93 | 100 | 96.0-100% | 332/332 | 100 | 98.9-100% | |
| Influenza B virus | Prospective | 4/4 | 100 | 51.0-100% | 872/872 | 100 | 99.6-100% | |
| Influenza B virus | Archived | 20/20 | 100 | 83.9-100% | 0/0 | - | - | |
| Influenza B virus | Contrived | 47/49 | 95.9 | 86.3-98.9% | 333/333 | 100 | 98.9-100% | |
| Respiratory syncytial virus | Prospective | 21/24 | 87.5 | 69.0-95.7% | 849/851 | 99.8 | 99.1-99.9% | |
| Respiratory syncytial virus | Archived | 2/2 | 100 | 34.2-100% | 56/57 | 98.2 | 90.7-99.7% | |
| Respiratory syncytial virus | Contrived | 49/50 | 98.0 | 89.5-99.6% | 381/381 | 100 | 99.0-100% | |
| Bacteria | | | | | | | | |
| Streptococcus pyogenes<br>(group A Strep) | Prospective | PCR | 209/217 | 96.3 | 92.9-98.1% | 654/660 | 99.1 | 98.0-99.6% |
| | | Culture | 174/177 | 98.3 | 95.1-99.4% | 654/692 | 94.5 | 92.6-96.0% |
| Streptococcus pyogenes<br>(group A Strep) | Archived | 38/39 | 97.4 | 86.8-99.5% | 10/10 | 100 | 72.2-100% | |
| Streptococcus pyogenes<br>(group A Strep) | Contrived | 0/0 | - | | 0/0 | - | | |
The performance measures of sensitivity and spective and Archived Streptococcus analytes for which culture was used as the reference method. Performance measures of PPA and NPA refer to all other analytes, for which molecular assays or known specimen composition were used as comparator methods.
## Analytical Performance Characteristics
Bench performance (analytical) testing for the SPOTFIRE R/ST Panel Mini was designed to validate the performance of all analytes and to support testing of both sample types, NPS and TS specimens. Table 5 provides an overall summary of the analytical studies, their results, and conclusions.
| Study | Description | Acceptance Criteria | Results | Conclusion |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sample Storage and<br>Handling | The purpose of this study<br>was to establish that<br>accurate test results are<br>observed for all<br>SPOTFIRE R/ST Panel<br>Mini analytes, including<br>sore throat-specific<br>bacteria that had not<br>been previously<br>evaluated, when throat<br>swab (TS) specimens in<br>Amies media (TSa) are<br>stored within the same<br>range of storage<br>conditions previously<br>validated for<br>nasopharyngeal swab<br>(NPS) specimens in liquid<br>media. | For the storage condition<br>to be considered<br>acceptable for each<br>organism, 100% expected<br>positive results were<br>required to be observed<br>in all samples tested. In<br>addition, crossing point<br>(Cp) values were<br>evaluated for each<br>relevant assay and<br>trended across the<br>conditions to assess<br>analyte stability over time. | Positive results were<br>observed in 100% of all<br>TSa samples tested at all<br>conditions evaluated for<br>all SPOTFIRE R/ST<br>Panel Mini analytes. | The SPOTFIRE R/ST<br>Panel provides accurate<br>results when TS<br>specimens are stored in<br>Amies media for up to 4<br>hours at ambient<br>temperature (15-25 °C),<br>up to 3 days at<br>refrigerated temperature<br>(2-8 °C), and up to 30<br>days at frozen<br>temperature (≤ -15 °C).<br>Similar results were<br>previously observed with<br>NPS specimens stored in<br>transport media. |
| Limit of Detection | The purpose of this study<br>was to determine the<br>Limit of Detection (LoD)<br>for all analytes detected<br>by the SPOTFIRE R/ST<br>Panel Mini in two<br>common liquid media,<br>Viral Transport Media<br>(VTM) and Amies media.<br>In addition, testing was<br>performed to assess<br>whether the presence of | The LoD for each<br>SPOTFIRE R/ST Panel<br>Mini analyte was<br>confirmed when positive<br>results were reported in at<br>least 95% (≥19/20) of<br>replicates tested at the<br>LoD (1× LoD), and fewer<br>than 95% (≤18/20) of<br>replicates tested at a | The LoD concentrations<br>for the SPOTFIRE R/ST<br>Panel Mini analytes were<br>confirmed in viable or<br>infectious units (e.g.<br>TCID50/mL, CFU/mL,<br>cells/mL, IFU/mL,<br>CEID50/mL, CCU/mL)<br>and/or nucleic acid<br>copies/mL. The panel<br>accurately detected<br>viruses and bacteria in | The LoD was determined<br>for each analyte detected<br>by the SPOTFIRE R/ST<br>Panel Mini in VTM and/or<br>Amies media, where<br>appropriate, for the<br>Respiratory and Sore<br>Throat menus,<br>respectively. The<br>SPOTFIRE R/ST Panel<br>Mini provides accurate<br>detection results for all |
| Study | Description | Acceptance Criteria | Results | Conclusion |
| | multiple organisms in a<br>sample has an impact on<br>the ability of the panel to<br>detect low level analytes<br>compared to samples<br>containing only one<br>organism. | concentration 10-fold<br>below LoD (0.1x LoD).<br>Equivalent detection of<br>representative analytes in<br>single analyte and multi-<br>analyte samples was<br>determined primarily<br>based on concordance of<br>positive or negative<br>results at each test<br>concentration. | samples contrived in<br>either VTM or Amies<br>media containing one or<br>multiple organisms. | analytes in single or<br>polymicrobial specimens<br>when present at or above<br>the LoD. No adverse<br>effect on the analytical<br>sensitivity of the<br>SPOTFIRE R/ST Panel<br>Mini was observed when<br>evaluating multi-analyte<br>specimens. |
| Analytical Reactivity<br>(Inclusivity) | The purpose of this study<br>was to evaluate the<br>analytical reactivity<br>(inclusivity) of the<br>SPOTFIRE R/ST Panel<br>Mini assays. | The primary data<br>evaluated for this study<br>were the reported<br>positive, negative, and<br>uncertain results. The<br>assay reactivity of each<br>isolate was confirmed if<br>positive results were<br>reported for the<br>appropriate analyte in 3/3<br>or 4/5 replicates tested<br>within $10 \times LoD$ . If positive<br>results were reported in<br>fewer than 4/5 replicates,<br>additional testing was<br>performed at $100 \times LoD$ or<br>higher. Isolates that did<br>not yield the expected<br>results at or below $10 \times$<br>LoD were further<br>investigated to determine<br>the cause of the<br>limitation. Isolates with<br>reactivity limitations will<br>be noted in the<br>SPOTFIRE R/ST Panel<br>Mini product literature. | Analytical reactivity<br>testing demonstrated that<br>the SPOTFIRE R/ST<br>Panel Mini can detect and<br>accurately report results<br>for a diverse collection of<br>isolates from a variety of<br>strains, serotypes, and<br>genotypes of species<br>collected over many<br>years and from<br>geographically distinct<br>locations with few<br>limitations. | The SPOTFIRE R/ST<br>Panel Mini detected and<br>accurately reported<br>results for various<br>species, subspecies,<br>serotypes, and genotypes<br>that may be present in<br>nasopharyngeal and<br>throat swab specimens.<br>The following limitation on<br>reactivity was identified<br>and is noted in the device<br>labeling:<br>• Rare strains of<br><i>Streptococcus<br/>pyogenes</i> do not<br>contain the region of<br>the genome targeted<br>by the SPOTFIRE<br>R/ST Panel Mini and<br>are not detected. |
| Analytical Specificity<br>(Exclusivity) | The objective of this study<br>was to assess the<br>analytical specificity<br>(exclusivity) of the<br>SPOTFIRE R/ST Panel<br>Mini by challenging the<br>system with high<br>concentrations of<br>analytes and evaluating<br>the occurrence of<br>unexpected positive<br>results due to assay<br>cross-reactivity. Testing<br>was divided into two<br>categories: on-panel and<br>off-panel. The on-panel<br>exclusivity evaluation<br>assessed the potential for<br>intra-panel cross-<br>reactivity by testing<br>representative organisms<br>that are targeted by the<br>panel. The off-panel<br>exclusivity evaluation<br>assessed the potential for<br>cross-reactivity of panel<br>assays with organisms<br>not detected by the panel. | The primary results<br>evaluated for this study<br>were positive, negative,<br>and uncertain<br>interpretations for each<br>panel analyte. On-panel<br>organisms were expected<br>to have a positive result<br>for the analyte being<br>tested and negative<br>results for all other<br>analytes targeted by the<br>panel. Off-panel<br>organisms were expected<br>to have negative results<br>for all panel analytes,<br>unless otherwise<br>indicated. | Three cross-reactivities<br>were identified by<br>empirical and/or <i>in silico</i><br>evaluations that are<br>predicted to cause<br>inaccurate test results for<br>the SPOTFIRE R/ST<br>Panel Mini. Two of the<br>identified cross-<br>reactivities are due to<br>reactivity between<br>phylogenetic near-<br>neighbors that are rarely<br>observed in human<br>samples. Only one cross-<br>reactivity (HRV/EV assay<br>with select <i>Bordetella</i><br>species) was due to non-<br>specific interactions<br>between assay primers<br>and sequences within the<br>genome of unrelated<br>organisms. | The SPOTFIRE R/ST<br>Panel Mini assays are<br>specific for the intended<br>analytes, with the<br>following limitations that<br>are noted in the device<br>labeling:<br>• The SARS-CoV-2<br>assays can amplify<br>select sequences<br>from closely related<br>sarbecoviruses<br>isolated from bats<br>and pangolin.<br>• Some <i>Bordetella</i><br>species ( <i>B.<br/>pertussis, B.<br/>parapertussis</i> , and <i>B.<br/>bronchiseptica</i> ) will<br>be reported as<br>Human rhinovirus<br>when organisms are<br>present at a high<br>concentration.<br>• Some bovine and<br>canine |
| Study | Description | Acceptance Criteria | Results | Conclusion |
| |…