K230267 · Neumodx Molecular, Inc. · QEP · Dec 22, 2023 · Microbiology
Device Facts
Record ID
K230267
Device Name
NeuMoDx CT/NG Assay 2.0
Applicant
Neumodx Molecular, Inc.
Product Code
QEP · Microbiology
Decision Date
Dec 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3393
Device Class
Class 2
Indications for Use
The NeuMoDx CT/NG Assay 2.0, as implemented on the NeuMoDx 96 Molecular System and NeuMoDx 288 Molecular System, is an automated, qualitative test for the direct detection and differentiation of Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (NG) DNA as an aid in the diagnosis of chlamydial and gonococcal urogenital disease in symptomatic and asymptomatic individuals. The Assay utilizes real-time Polymerase Chain Reaction (PCR) and may be used to test male and female urine, and self-collected vaginal swab specimens (collected in a clinical setting).
Device Story
Automated in vitro diagnostic test for direct detection/differentiation of Chlamydia trachomatis and Neisseria gonorrhoeae DNA; utilizes real-time PCR and fluorogenic TaqMan probes; employs paramagnetic bead technology for nucleic acid extraction from urine and vaginal swab specimens. Implemented on NeuMoDx 96 and 288 Molecular Systems; provides complete automation from lysis to result reporting in 8 hours. Systems include XPCR modules, liquid/cartridge/plate workstation, and PC interface. Healthcare providers use results as an aid in diagnosis of urogenital disease; automated interpretation reduces manual error; enables rapid clinical decision-making for patient management.
Clinical Evidence
Prospective, multi-center study (14 sites) with 4017 participants (1825 males, 2192 females). Compared NeuMoDx CT/NG Assay 2.0 results against Patient Infected Status (PIS) established by two FDA-cleared NAATs. Results across all specimen types (SCVS, FU, MU) showed high sensitivity (92.3%-100%) and specificity (99.5%-100%) for both CT and NG. Invalid/unresolved rates were low (0.2% unresolved).
Technological Characteristics
Automated real-time PCR system. Materials: unitized reagent test strips, extraction plates with dried reagents, lysis buffer (chaotropic salts), wash/release reagents. Sensing: fluorogenic TaqMan probes. Connectivity: standalone/networked NeuMoDx 96/288 systems. Software: automated result interpretation based on Ct, EP, and EPR thresholds. Sterilization: not specified.
Indications for Use
Indicated for the qualitative detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in symptomatic and asymptomatic individuals. Patient population includes males and females (age 15-77). Specimens include male/female urine and self-collected vaginal swabs (collected in clinical settings).
Regulatory Classification
Identification
A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.
Special Controls
A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) must comply with the following special controls: (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended. (2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device: alternatively, the sample collection device must be cleared in a premarket submission as a part of this device. (3) The 21 CFR 809.10(b) labeling must include: (i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens; (ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including, but not limited to. Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, with-in lab precision, and reproducibility, as appropriate; (iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing. (iv) Limiting statements indicating that: (A)a negative test result does not preclude the possibility of infection; (B) the test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician; (C) reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe procedures in any one of these steps can lead to incorrect results; and (D)if appropriate (e.g., recommended by CDC, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type. (v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that: (A)negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present; (B) detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora; (C) detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and (D) therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy. (4) Design verification and validation must include: (i) Detailed device description documentation, including, but not limited to, methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported result). (ii) Detailed documentation of analytical studies including but not limited to, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, with-in lab precision, and reproducibility, as appropriate. (iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses. (iv) A detailed description of the impact of any software, including, but not limited to, software applications and hardware-based devices that incorporate software, on the device's functions.
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate;
(iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) Limiting statements indicating that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) Reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) If appropriate (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type; and(v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that:
(A) Negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present;
(B) Detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora;
(C) Detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and
(D) Therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy.
(4) Design verification and validation must include:
(i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (
*e.g.,* how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies, including, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate.
(iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses.
(iv) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.
Predicate Devices
Aptima Combo 2 Assay (Panther System) (k190515)
Related Devices
K173887 — cobas CT/NG for use on cobas 6800/8800 systems · Roche Molecular Systems, Inc. · Mar 21, 2018
K202408 — cobas CTNG for use on cobas 6800/8800 systems · Roche Molecular Systems, Inc. · Jan 21, 2021
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K230267
B Applicant
NeuMoDx Molecular, Inc.
C Proprietary and Established Names
NeuMoDx CT/NG Assay 2.0
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QEP | Class II | 21 CFR 866.3393 - Device To Detect Nucleic Acids From Non-Viral Microorganism(S) Causing Sexually Transmitted Infections And Associated Resistance Marker(S) | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
NeuMoDx Molecular, Inc submitted a Traditional 510(k) to obtain market clearance for the distribution and/or use of a new device called NeuMoDx CT/NG Assay 2.0, on the NeuMoDx 96 Molecular System and NeuMoDx 288 Molecular System.
Measurand:
Chlamydia trachomatis and/or Neisseria gonorrhoeae DNA.
B Type of Test:
Nucleic acid amplification assay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The NeuMoDx CT/NG Assay 2.0, as implemented on the NeuMoDx 96 Molecular System and NeuMoDx 288 Molecular System, is an automated, qualitative test for the direct detection and differentiation of *Chlamydia trachomatis* (CT) and/or *Neisseria gonorrhoeae* (NG) DNA as an aid in the diagnosis of chlamydial and gonococcal urogenital disease in symptomatic and asymptomatic individuals. The Assay utilizes real-time Polymerase Chain Reaction (PCR) and may be used to test male and female urine, and self-collected vaginal swab specimens (collected in a clinical setting).
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For in vitro diagnostic use.
IV Device/System Characteristics:
A Device Description:
The NeuMoDx CT/NG Assay 2.0 is an automated *in vitro* diagnostic test which aims at providing direct detection of CT and/or NG DNA from asymptomatic and symptomatic patient specimens.
- The assay utilizes real-time polymerase chain reaction (PCR) for the amplification of CT and/or NG DNA and fluorogenic target-specific TaqMan probes for the detection of the amplified DNA.
- Results are interpreted with pre-established decision criteria.
- The assay is implemented on NeuMoDx 96 Molecular System (N96) and NeuMoDx 288 (N288) Molecular System PCR instruments which deliver complete automation from initial steps of specimen lysis to final steps of amplicon detection and aims to provide complete sets of results in 8 hours.
- Each of the instruments, N96 and N288, consists of XPCR modules, liquid/cartridge/plate workstation, extraction plate module, liquid reagent module, barcode scanner, PC with touch screen interface, power system and enclosure. These systems also utilize sample racks, rack trays and other accessories.
The NeuMoDx System consists of assay-specific test strip, consumables, reagents, accessories, analyzers, and software. The following components, consumables and reagents are required with NeuMoDx CT/NG Assay 2.0
K230267 - Page 2 of 24
{2}
- NeuMoDx CT/NG Test Strip 2.0 (included with assay) - contains ambient-temperature stable PCR reagents which are required for the amplification of CT/NG and SPC1 nucleic acids.
- NeuMoDx Extraction Plate required for cell lysis, protein degradation, nucleic acid binding, and process monitoring.
- NeuMoDx Cartridge incorporates a proprietary microfluidic design and facilitates nucleic acid extraction and purification, as well as PCR amplification and detection.
- NeuMoDx Lysis buffer 2 helps lyse bacterial targets in human urogenital specimens.
- NeuMoDx Wash Reagent.
- NeuMoDx Release Reagent.
## B Principle of Operation:
The NeuMoDx CT/NG Assay 2.0 is a nucleic acid-based test using real-time polymerase chain reaction (PCR) technology to determine the presence of CT and/or NG. A proprietary paramagnetic bead technology is employed for efficient nucleic acid capture from lysed urine and swab specimens. Following extraction, DNA is subjected to real-time PCR to determine the presence of CT and NG. detection, and reporting of results.
## C Instrument Description Information:
1. Instrument Name:
- NeuMoDx 288 Molecular System
- NeuMoDx 96 Molecular System
2. Specimen Identification:
By hand-held barcode reader or automated barcode reader and positional checks.
3. Specimen Sampling and Handling:
Fully automated
4. Calibration:
Not Applicable
5. Quality Control:
- Internal Sample Processing Control: This monitors the efficacy of the entire test process including nucleic acid extraction and real-time PCR amplification.
K230267 - Page 3 of 24
{3}
- External Controls: External controls are not provided with the assay. The package insert recommends a commercial source of external controls.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Aptima Combo 2 Assay (Panther System)
B Predicate 510(k) Number(s):
K190515
C Comparison with Predicate(s):
| Device & Predicate Device(s): | k230267 | k190515 |
| --- | --- | --- |
| Device Trade Name | NeuMoDx CT/NG Assay 2.0 | Aptima Combo 2 Assay (Panther System) |
| General Device Characteristic Similarities and Differences | | |
| Intended Use/Indications for Use | The NeuMoDx CT/NG Assay 2.0, as implemented on the NeuMoDx 96 Molecular System and NeuMoDx 288 Molecular System, is an automated, qualitative test for the direct detection and differentiation of Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (NG) DNA as an aid in the diagnosis of chlamydial and gonococcal urogenital disease in symptomatic and asymptomatic individuals. The Assay utilizes real-time Polymerase Chain Reaction (PCR) and may be used to test male and female urine, and self-collected vaginal | The Aptima Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (NG) to aid in the diagnosis of chlamydial and/or gonococcal disease using the Panther System as specified. On the Panther System, the assay may be used to test the following specimens from symptomatic and asymptomatic individuals: clinician-collected endocervical, vaginal, throat, rectal and male urethral swab specimens, |
K230267 - Page 4 of 24
{4}
K230267 - Page 5 of 24
| | swab specimens (collected in a clinical setting). | clinician-collected gynecological specimens collected in the PreservCyt Solution, patient- collected vaginal swab specimens^{1}, and female and male urine specimens.
^{1}Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The Aptima Multitest Swab Specimen Collection Kit has not been evaluated for home use. |
| --- | --- | --- |
| Assay principle | Real-time PCR | Target Capture (TC), Transcription- Mediated Amplification (TMA), Hybridization Protection Assay (HPA) |
| Instrumentation | NeuMoDx 288 Molecular System and NeuMoDx 96 Molecular System | Panther System |
| Analyte targets | CT cryptic plasmid DNA
CT OMP-putative porin gene
NG OPC- opacity gene | Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) rRNA |
| Amplification technology | Real-time PCR | Target Capture (TC), Transcription- Mediated Amplification (TMA), Hybridization Protection Assay (HPA) |
| Test interpretation | Automated test interpretation | Same |
| Controls | External positive and negative controls
Internal sample process control | External positive and negative controls |
| Organism detected | Chlamydia trachomatis and/or Neisseria gonorrhoeae | Same |
| Specimen type | **Female specimens:**
• Vaginal self- | **Female specimens:**
• Vaginal swab |
{5}
| | collected swab
(collected in a
clinical setting)
• Urine
**Male Specimens:**
• Urine | • Endocervical swab
• Gynecological specimens
in PreservCyt solution
• Urine
• Throat swab
• Rectal swab
**Male Specimens:**
• Urethral swab
• Urine
• Throat swab
• Rectal swab |
| --- | --- | --- |
VI Standards/Guidance Documents Referenced:
- 21 CFR 866.3393: Device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s)
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Within-Laboratory Precision
The within-laboratory precision of the NeuMoDx CT/NG 2.0 Assay was evaluated at NeuMoDx internal facility using N96 and N288 systems. Panel members were created by spiking CT and NG into pooled negative, urine specimen and UVT-transport medium for swab samples. Six panel members for each organism were prepared in negative matrix. Testing was performed using three of each instrument over 12 days with 2 runs per day and 3 replicates of each CT/NG panel member per run. The results were evaluated for overall % agreement with expected results. Additionally, the standard deviation and % coefficient of variation (CV) was calculated for each component of variability (repeatability, between-run, between days, within systems, between systems, within laboratories and overall/total) for each panel member are shown in tables 1 and 2.
K230267 - Page 6 of 24
{6}
Table 1. NeuMoDx CT/NG Assay 2.0 Within-Laboratory Precision for CT
| | | Vaginal Swabs | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Overall | | | | SD (% CV) | | | | | |
| Panel Member (Conc) | | No. of Correct/Total No. of Tested | Avg Ct | % Agreement ** | 95% CI | Repeatability | Between Run | Between Day | Within System | Between System | Within Lab |
| Moderate CT | N96 | 216/216 | 29.7 | 100 | 98.2-100.0 | 0.614 (2.1) | 0.191 (0.6) | 0.0 (0.0) | 0.643 (2.2) | 0.104 (0.3) | 0.651 (2.2) |
| | N288 | 216/216 | 29.6 | 100 | 98.2-100.0 | 0.733 (2.5) | 0.0 (0.0) | 0.303 (1.0) | 0.793 (2.7) | 0.099 (0.3) | 0.799 (2.7) |
| Low CT* | N96 | 432/432 | 31.7 | 100 | 99.1-100.0 | 0.771 (2.4) | 0.196 (0.6) | 0.0 (0.0) | 0.795 (2.5) | 0.115 (0.4) | 0.803 (2.5) |
| | N288 | 431/431 | 31.7 | 100 | 99.1-100.0 | 0.771 (2.4) | 0.08 (0.3) | 0.193 (0.6) | 0.799 (2.5) | 0.133 (0.4) | 0.810 (2.6) |
| High Negative CT | N96 | 36/218 | N/A | 16.5 | 12.7-22.3 | N/A | | | | | |
| | N288 | 31/216 | N/A | 14.4 | 10.3-19.6 | N/A | | | | | |
| True Negative CT | N96 | 215/216 | N/A | 99.5 | 97.4-99.9 | N/A | | | | | |
| | N288 | 217/217 | N/A | 100 | 98.3-100.0 | N/A | | | | | |
| | | Urine | | | | | | | | | |
| | | Overall | | | | SD (% CV) | | | | | |
| Panel Member | | No. of Correct/Total No. of Tested | Avg Ct | % Agree Ment ** | 95% CI | Repeatability | Between Run | Between Day | Within System | Between System | Within Lab |
| Moderate CT | N96 | 216/216 | 30.1 | 100 | 98.2-100.0 | 0.545 (1.8) | 0.300 (1.0) | 0.101 (0.3) | 0.630 (2.1) | 0.094 (0.3) | 0.637 (2.1) |
| | N288 | 215/215 | 29.9 | 100 | 98.2-100.0 | 0.702 (2.3) | 0.079 (0.3) | 0.117 (0.4) | 0.716 (2.4) | 0.158 (0.5) | 0.734 (2.5) |
| Low CT* | N96 | 432/432 | 32.0 | 100 | 99.1-100.0 | 0.648 (2.0) | 0.294 (0.9) | 0.108 (0.3) | 0.720 (2.2) | 0.056 (0.2) | 0.722 (2.3) |
| | N288 | 216/216 | 31.9 | 100 | 98.2-100.0 | 0.641 (2.0) | 0.096 (0.3) | 0.128 (0.4) | 0.666 (2.1) | 0.0 (0.0) | 0.666 (2.1) |
| High Negative CT | N96 | 31/215 | N/A | 14.4 | 10.2-20.0 | N/A | | | | | |
| | N288 | 35/216 | N/A | 16.2 | 11.7-22.0 | N/A | | | | | |
| True Negative CT | N96 | 214/215 | N/A | 99.5 | 97.4-99.9 | N/A | | | | | |
| | N288 | 215/215 | N/A | 100 | 98.2-100.0 | N/A | | | | | |
Note: Repeatability = Between Replicates, Avg= Average, SD= Standard Deviation, CV= Coefficient of Variation, CI= Confidence Interval, * Low level data pooled from two panel members, ** Agreement with expected results.
K230267 - Page 7 of 24
{7}
Table 2. NeuMoDx CT/NG Assay 2.0 Within-Laboratory Precision for NG
| | | Vaginal Swabs | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Overall | | | | SD (% CV) | | | | | |
| Panel Member | | No. of Correct/Total No. of Tested | Avg Ct | % Agreement ** | 95% CI | Repeatability | Between Run | Between Day | Within System | Between System | Within Laboratory |
| Moderate NG | N96 | 216/216 | 27.3 | 100 | 98.2-100 | 0.528 (1.9) | 0.166 (0.6) | 0.0 (0.0) | 0.553 (2.0) | 0.014 (0.1) | 0.554 (2.0) |
| | N288 | 216/216 | 27.2 | 100 | 98.2-100 | 0.643 (2.2) | 0.128 (0.4) | 0.108 (0.4) | 0.664 (2.3) | 0.049 (0.2) | 0.666 (2.3) |
| Low NG* | N96 | 432/432 | 29.1 | 100 | 99.1-100 | 0.538 (1.8) | 0.155 (0.5) | 0.101 (0.3) | 0.569 (2.0) | 0.116 (0.4) | 0.581 (2.0) |
| | N288 | 432/432 | 29.1 | 100 | 99.1-100 | 0.643 (2.2) | 0.128 (0.4) | 0.108 (0.4) | 0.664 (2.3) | 0.049 (0.2) | 0.666 (2.3) |
| High Negative NG | N96 | 124/215 | N/A | 57.7 | 51.0-64.1 | N/A | | | | | |
| | N288 | 121/216 | N/A | 56.0 | 49.3-62.5 | N/A | | | | | |
| True Negative NG | N96 | 216/216 | N/A | 100 | 98.2-100 | N/A | | | | | |
| | N288 | 216/216 | N/A | 100 | 98.2-100 | N/A | | | | | |
| | | Urine | | | | | | | | | |
| | | Overall | | | | SD (% CV) | | | | | |
| Panel Member | | No. of Correct/Total No. of Tested | Avg Ct | % Agreement | 95% CI | Repeatability | Between Run | Between Day | Within System | Between System | Within Laboratory |
| Moderate NG | N96 | 216/216 | 28.4 | 100 | 98.2-100.0 | 0.503 (1.8) | 0.337 (1.2) | 0.0 (0.0) | 0.605 (2.1) | 0.095 (0.3) | 0.612 (2.1) |
| | N288 | 216/216 | 28.3 | 100 | 98.2-100.0 | 0.491 (1.7) | 0.191 (0.7) | 0.0 (0.0) | 0.527 (1.9) | 0.026 (0.1) | 0.528 (1.9) |
| Low NG* | N96 | 432/432 | 30.0 | 100 | 99.1-100.0 | 0.648 (2.0) | 0.294 (0.9) | 0.0 (0.0) | 0.614 (2.0) | 0.073 (0.2) | 0.618 (2.1) |
| | N288 | 431/432 | 30.0 | 99.8 | 98.2-100.0 | 0.680 (2.3) | 0.0 (0.0) | 0.210 (0.7) | 0.712 (2.4) | 0.134 (0.4) | 0.724 (2.4) |
| High Negative NG | N96 | 69/215 | N/A | 32.1 | 26.0-38.8 | N/A | | | | | |
| | N288 | 70/216 | N/A | 32.4 | 26.5-39.0 | N/A | | | | | |
| True Negative NG | N96 | 216/216 | N/A | 100 | 98.2-100.0 | N/A | | | | | |
| | N288 | 215/215 | N/A | 100 | 97.8-100.0 | N/A | | | | | |
Note: Repeatability = Between Replicates, Ave= Average, SD= Standard Deviation, CV= Coefficient of Variation, CI= Confidence Interval, * Low level data pooled from two panel members, ** Agreement with expected results.
K230267 - Page 8 of 24
{8}
Using urine as a representative matrix, the precision of the NeuMoDx CT/NG 2.0 Assay was evaluated at 1X and 2X of the confirmed LoD. Low positive CT and NG samples were processed on each of five days, along with a negative control sample, one on each N96 and N288 system, using three lots of the NeuMoDx CT/NG Test Strip 2.0. Each panel consisted of two replicates of CT at 1x LoD, two replicates of NG at 1x LoD, two replicates of 2x LoD of CT and NG together, and two negative control samples, resulting in 60 data points per target level per system across five days, for a total of 120 replicates per level across both systems. Summary of results are shown below in Tables 3 and 4.
Table 3. Precision of NeuMoDx CT/NG Assay 2.0 with Low-Positive CT and NG in Urine Samples
| URINE | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| System Level | Reagent Lot | CT (Serovar D) | | | | | NG (B5025) | | | | |
| | | n | Avg Ct | % Agreement (95% CI)** | SD | % CV | n | Avg Ct | % Agreement (95% CI)** | SD | % CV |
| N96 | 1x LoD CT | 59 * | 33.37 | 96.6% (88.5 -99.1%) | 1.01 | 3.04 | N/A | | | | |
| | 1x LoD NG | N/A | | | | | 60 | 32.68 | 100% (94.0-100.0%) | 0.66 | 2.01 |
| | 2x LoD CT/ NG | 60 | 33.05 | 100% (94.0- 100.0%) | 0.84 | 2.55 | 60 | 31.89 | 100% (94.0-100.0%) | 0.54 | 1.71 |
| N288 | 1x LoD CT | 60 | 33.70 | 100% (94.0-100.0%) | 0.62 | 1.85 | N/A | | | | |
| | 1x LoD NG | N/A | | | | | 59 * | 32.66 | 100% (93.9-100.0%) | 0.48 | 1.46 |
| | 2x LoD CT/NG | 60 | 33.05 | 100% (94.0-100.0%) | 0.66 | 2.00 | 60 | 32.01 | 100% (94.0-100.0%) | 0.50 | 1.57 |
Note: Avg= Average, SD= Standard Deviation, CV= Coefficient of Variation
*1 replicate was invalid, ** Agreement with expected results
Table 4. Precision of NeuMoDx CT/NG Assay 2.0 with Low-Positive CT and NG in Urine Samples by Reagent Lot
| URINE | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| System | Level | Reagent Lot | CT (Serovar D) | | | | | NG (B5025) | | | | |
| | | | n | Avg Ct | % Agreement (95% CI)** | SD | % CV | n | Avg Ct | % Agreement (95% CI)** | SD | % CV |
| N96 | 1x LoD | 1 | 20 | 33.36 | 90.0% (69.9-97.2%) | 1.28 | 3.83 | 20 | 32.86 | 100% (83.9-100.0%) | 0.48 | 1.45 |
| | | 2 | 20 | 33.21 | 100% (83.9-100.0%) | 1.03 | 3.11 | 20 | 32.53 | 100% (83.9-100.0%) | 0.60 | 1.84 |
| | | 3 | 19 * | 33.54 | 100% (83.2- 100.0%) | 0.68 | 2.04 | 20 | 32.75 | 100% (83.9-100.0%) | 0.79 | 2.40 |
| | 2x LoD | 1 | 20 | 33.17 | 100% (83.9-100.0%) | 0.86 | 2.59 | 20 | 31.87 | 100% (83.9-100.0%) | 0.52 | 1.62 |
| | | 2 | 20 | 32.68 | 100% (83.9-100.0%) | 0.73 | 2.22 | 20 | 31.69 | 100% | 0.58 | 1.83 |
K230267 - Page 9 of 24
{9}
| | | | | | | | | | | (83.9-100.0%) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 3 | 20 | 33.32 | 100% (83.9-100.0%) | 0.84 | 2.51 | 20 | 32.09 | 100% (83.9-100.0%) | 0.48 | 1.51 |
| N288 | 1x LoD | 1 | 20 | 33.79 | 100% (83.9-100.0%) | 0.7 | 2.07 | 20 | 32.79 | 100% (83.9-100.0%) | 0.37 | 1.13 |
| | | 2 | 20 | 33.51 | 100% (83.9-100.0%) | 0.47 | 1.41 | 19* | 32.42 | 100% (83.2-100.0%) | 0.48 | 1.49 |
| | | 3 | 20 | 33.81 | 100% (83.9-100.0%) | 0.66 | 1.56 | 20 | 32.75 | 100% (83.9-100.0%) | 0.50 | 1.53 |
| | 2x LoD | 1 | 20 | 32.97 | 100% (83.9-100.0%) | 0.66 | 1.99 | 20 | 31.84 | 100% (83.9-100.0%) | 0.54 | 1.68 |
| | | 2 | 20 | 33.00 | 100% (83.9-100.0%) | 0.45 | 1.36 | 20 | 32.00 | 100% (83.9-100.0%) | 0.36 | 1.13 |
| | | 3 | 20 | 33.86 | 100% (83.9- 100.0%) | 0.75 | 2.20 | 20 | 32.18 | 100% (83.9- 100.0%) | 0.55 | 1.71 |
Note: Avg= Average, SD= Standard Deviation, CV= Coefficient of Variation, *1 replicate was invalid, ** Agreement with expected results.
## Reproducibility
The between-laboratory reproducibility of the NeuMoDx CT/NG Assay 2.0 was conducted at two external sites and one internal site, by two operators on one N96 system and one N288 system, at each of the three sites. The study was conducted over 5 days, with 2 runs per day by 2 operators using individual lots of NeuMoDx strips. Six reproducibility panel members were prepared in pooled negative, urine specimen and UVT-transport medium for swab samples (Table 5). Panels were prepared to include varying LoD for each organism, in combination and one sample negative for both organisms, i.e., unspiked negative urine matrix or UTV for swab matrix. The variability between days, between runs, and within runs was calculated separately for each low and moderate positive panel member and is shown in Tables 5 and 6.
K230267 - Page 10 of 24
{10}
Table 5. NeuMoDx CT/NG Assay 2.0 Multi-site Reproducibility for CT
| Vaginal Swabs | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Overall | | | SD (% CV) | | | | |
| Panel Member | | No. of Correct/ Total No. of Tested | Avg Ct | % Agreement ** | 95% CI | Within Run | Between Run | Between Day | Between Site | Overall |
| Moderate CT | N96 | 90/90 | 29.4 | 100 | 95.9-100.0 | 0.63 (2.1) | 0.15 (0.5) | 0.22 (0.7) | 0.23 (0.8) | 0.69 (2.4) |
| | N288 | 90/90 | 29.4 | 100 | 95.9-100.0 | 0.56 (1.9) | 0.0 (0.0) | 0.0 (0.0) | 0.38 (1.3) | 0.64 (2.2) |
| Low CT* | N96 | 180/180 | 31.2 | 100 | 97.9-100.0 | 1.06 (3.4) | 0.0 (0.0) | 0.0 (0.0) | 0.16 (0.5) | 1.07 (3.4) |
| | N288 | 180/180 | 31.4 | 100 | 97.9-100.0 | 0.60 (1.9) | 0.0 (0.0) | 0.11 (0.4) | 0.35 (1.1) | 0.67 (2.1) |
| High Negative | N96 | 77/90 | N/A | 85.6 | 76.9-91.4 | N/A | | | | |
| | N288 | 70/90 | N/A | 77.8 | 68.2-85.1 | N/A | | | | |
| True Negative | N96 | 90/90 | N/A | 100 | 95.9-100.0 | N/A | | | | |
| | N288 | 90/90 | N/A | 100 | 95.9-100.0 | N/A | | | | |
| Urine | | | | | | | | | | |
| | | | Overall | | | SD (% CV) | | | | |
| Panel Member | | No. of Correct/ Total No. of Tested | Avg Ct | % Agreement ** | 95% CI | Within Run | Between Run | Between Day | Between Site | Overall |
| Moderate CT | N96 | 90/90 | 29.5 | 100 | 95.2-100.0 | 0.51 (1.7) | 0.0 (0.0) | 0.0 (0.0) | 0.24 (0.8) | 0.54 (1.8) |
| | N288 | 90/90 | 29.8 | 100 | 95.9-100.0 | 0.39 (1.3) | 0.0 (0.0) | 0.7 (0.2) | 0.29 (1.0) | 0.46 (1.5) |
| Low CT* | N96 | 180/180 | 31.0 | 100 | 97.9-100.0 | 0.71 (2.3) | 0.12 (0.4) | 0.02 (0.1) | 0.37 (1.2) | 0.78 (2.5) |
| | N288 | 180/180 | 31.4 | 100 | 97.9-100.0 | 0.62 (2.0) | 0.0 (0.0) | 0.0 (0.0) | 0.21 (0.7) | 0.64 (2.0) |
| High Negative | N96 | 88/90 | N/A | 97.8 | 92.3-99.4 | N/A | | | | |
| | N288 | 85/90 | N/A | 94.4 | 87.6-97.6 | N/A | | | | |
| True Negative | N96 | 179/180 | N/A | 99.4 | 96.9-99.9 | N/A | | | | |
| | N288 | 180/180 | N/A | 100 | 97.9-100.0 | N/A | | | | |
Note: Avg= Average, SD= Standard Deviation, CV= Coefficient of Variation.
* Low level data pooled from two panel members, ** Agreement with expected results.
K230267 - Page 11 of 24
{11}
Table 6. NeuMoDx CT/NG Assay 2.0 Multi-site Reproducibility for NG
| Vaginal Swabs | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Overall | | | SD (% CV) | | | | |
| Panel Member | | No. of Correct/ Total No. of Tested | Avg Ct | % Agreement ** | 95% CI | Within Run | Between Run | Between Day | Between Site | Overall |
| Moderate CT | N96 | 90/90 | 27.6 | 100 | 95.9-100.0 | 0.64 (2.3) | 0.0 (0.0) | 0.22 (0.8) | 0.11 (0.4) | 0.67 (2.4) |
| | N288 | 90/90 | 27.7 | 100 | 95.9-100.0 | 0.50 (1.8) | 0.0 (0.0) | 0.19 (0.7) | 0.35 (1.3) | 0.60 (2.2) |
| Low CT* | N96 | 180/180 | 29.4 | 100 | 97.9-100.0 | 0.61 (2.1) | 0.0 (0.0) | 0.0 (0.0) | 0.29 (1.0) | 0.66 (2.2) |
| | N288 | 180/180 | 29.7 | 100 | 97.9-100.0 | 0.54(1.8) | 0.0 (0.0) | 0.04 (0.1) | 0.40 (1.4) | 0.63 (2.1) |
| High Negative | N96 | 42/90 | N/A | 46.7 | 36.7-56.9 | N/A | | | | |
| | N288 | 25/90 | N/A | 27.8 | 19.6-37.8 | N/A | | | | |
| True Negative | N96 | 180/180 | N/A | 100 | 97.9-100.0 | N/A | | | | |
| | N288 | 90/90 | N/A | 100 | 95.9-100.0 | N/A | | | | |
| Urine | | | | | | | | | | |
| | | | Overall | | | SD (% CV) | | | | |
| Panel Member | | No. of Correct/ Total No. of Tested | Avg Ct | % Agreement ** | 95% CI | Within Run | Between Run | Between Day | Between Site | Overall |
| Moderate CT | N96 | 90/90 | 28.3 | 98 | 94.0-99.8 | 0.50 (1.8) | 0.0 (0.0) | 0.15 (0.5) | 0.28 (1.0) | 0.56 (2.0) |
| | N288 | 90/90 | 28.4 | 100 | 95.9-100.0 | 0.56 (1.9) | 0.0 (0.0) | 0.0 (0.0) | 0.38 (1.3) | 0.64 (2.3) |
| Low CT* | N96 | 180/180 | 30.0 | 100 | 97.9-100.0 | 0.51 1.7) | 0.04 (0.1) | 0.0 (0.0) | 0.37 (1.3) | 0.60 (2.0) |
| | N288 | 180/180 | 30.3 | 100 | 97.9-100.0 | 0.62 (2.0) | 0.0 (0.0) | 0.0 (0.0) | 0.21 (0.7) | 0.64 (2.0) |
| High Negative | N96 | 63/90 | N/A | 70.0 | 59.9-78.5 | N/A | | | | |
| | N288 | 85/90 | N/A | 94.4 | 87.6-97.6 | N/A | | | | |
| True Negative | N96 | 180/180 | N/A | 100 | 99.4-100.0 | N/A | | | | |
| | N288 | 180/180 | N/A | 100 | 97.9-100.0 | N/A | | | | |
Note: Avg= Average, SD= Standard Deviation, CV= Coefficient of Variation.
* Low level data pooled from two panel members, ** Agreement with expected results.
## Lot-to-lot reproducibility
Reproducibility between key NeuMoDx CT/NG Assay 2.0 reagent lots was evaluated using a three-member panel of CT and NG targets spiked at varied concentration into negative urine and vaginal swab/ UTV. Two runs of 10 replicates were conducted with each of the 3-reagent lot sets on two N288 systems, the results of which are shown below in Table 7.
K230267 - Page 12 of 24
{12}
Table 7. NeuMoDx CT/NG Assay 2.0 Inter-Lot Reproducibility Studies
| Vaginal Swabs | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lots | Panel Member | CT | | | | NG | | | |
| | | % Pos | Avg Ct | Ct SD | Ct % CV | % Pos | Avg Ct | Ct SD | Ct % CV |
| Lot Set #1 | Negative | 0 (0/40) | N/A | | | 0 (0/40) | N/A | | |
| | Moderate CT, Low NG | 100 (40/40) | 29.6 | 0.83 | 2.8 | 100 (40/40) | 28.6 | 0.59 | 2.1 |
| | Low CT, Moderate NG | 100 (40/40) | 31.4 | 0.84 | 2.7 | 100 (40/40) | 27.0 | 0.58 | 2.1 |
| Lot Set #2 | Negative | 0 (0/40) | N/A | | | 0 (0/40) | N/A | | |
| | Moderate CT, Low NG | 100 (40/40) | 29.7 | 0.59 | 2.0 | 100 (40/40) | 28.8 | 0.52 | 1.8 |
| | Low CT, Moderate NG | 100 (40/40) | 31.9 | 0.42 | 1.3 | 100 (40/40) | 27.6 | 0.28 | 1.0 |
| Lot Set #3 | Negative | 0 (0/41) | N/A | | | 0 (0/40) | N/A | | |
| | Moderate CT, Low NG | 100 (40/40) | 29.8 | 0.62 | 2.1 | 100 (40/40) | 28.9 | 0.63 | 2.2 |
| | Low Ct, Moderate NG | 100 (40/40) | 32.0 | 0.46 | 1.4 | 100 (40/40) | 27.6 | 0.41 | 1.5 |
| Total | Negative | 0 (0/121) | N/A | | | 0 (0/120) | N/A | | |
| | Moderate CT, Low NG | 100 (120/120) | 29.7 | 0.69 | 2.3 | 100 (120/120) | 28.8 | 0.59 | 2.1 |
| | Low Ct, Moderate NG | 100 (120/120) | 31.8 | 0.66 | 2.1 | 100 (120/120) | 27.4 | 0.51 | 1.8 |
| Urine | | | | | | | | | |
| Lots | Panel Member | CT | | | | NG | | | |
| | | % Pos | Avg Ct | Ct SD | Ct % CV | % Pos | Avg Ct | Ct SD | Ct % CV |
| Lot Set #1 | Negative | 0 (0/40) | N/A | | | 0 (0/40) | N/A | | |
| | Moderate CT, Low NG | 100 (40/40) | 30.4 | 0.68 | 2.2 | 100 (40/40) | 29.1 | 0.55 | 1.9 |
| | Low CT, Moderate NG | 100 (40/40) | 32.2 | 0.98 | 3.1 | 100 (40/40) | 27.6 | 0.62 | 2.2 |
| Lot Set #2 | Negative | 0 (0/40) | N/A | | | 0 (0/40) | N/A | | |
| | Moderate CT, Low NG | 100 (40/40) | 30.0 | 0.85 | 2.8 | 100 (40/40) | 29.0 | 0.78 | 2.7 |
| | Low CT, Moderate NG | 100 (40/40) | 32.3 | 0.87 | 2.7 | 100 (40/40) | 27.8 | 0.57 | 2.0 |
| Lot Set #3 | Negative | 0 (0/41) | N/A | | | 0 (0/40) | N/A | | |
| | Moderate CT, Low NG | 100 (40/40) | 30.8 | 0.49 | 1.6 | 100 (40/40) | 29.6 | 0.44 | 1.5 |
| | Low Ct, Moderate NG | 100 (40/40) | 32.3 | 0.90 | 2.8 | 100 (40/40) | 27.7 | 0.64 | 2.2 |
K230267 - Page 13 of 24
{13}
| Total | Negative | 0 (0/122) | N/A | | | 0 (0/120) | N/A | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Moderate CT, Low NG | 100 (120/120) | 30.4 | 0.76 | 2.5 | 100 (120/120) | 29.2 | 0.64 | 2.2 |
| | Low Ct, Moderate NG | 100 (120/120) | 32.2 | 0.91 | 2.8 | 100 (120/120) | 27.7 | 0.64 | 2.3 |
Note: Avg= Average, Pos= Positive, SD= Standard Deviation, CV= Coefficient of Variation.
# 2. Linearity:
Not Applicable.
# 3. Analytical Specificity/Interference:
# Cross-reactivity and Interference with other Microorganisms
NeuMoDx CT/NG Assay 2.0 was evaluated for cross-reactivity and microbial interference by spiking pooled urine and swab UVT media with relevant bacteria, fungi, or viral organisms that may be present at the sampling site, such as commensal bacteria that may come into contact with the sampling device or genetically closely related species to CT and/ or NG. The potential for cross-reactivity was evaluated with a panel of 143 microorganisms, using cultured organisms at high titers (0.02-0.03 McFarland standards) prepared in negative, pooled urine matrix and UVT medium (Table 8). These samples were then spiked with CT and NG at $3\mathrm{x}$ LoD. Negative experimental control panels were prepared using negative urine and neat BD UVT, while positive experimental control panels were comprised of negative urine and BD UVT also spiked at $3\mathrm{x}$ LoD. Each panel was tested in triplicate for each matrix type. No false positive results were obtained with CT and NG negative samples demonstrating no cross-reactivity. All pools tested in the presence of CT and NG demonstrated no interference and yielded $100\%$ detection for CT and NG in triplicate for both swab and urine matrix types. The table below lists all the organisms evaluated with the NeuMoDx CT/NG Assay 2.0.
Table 8: Microorganisms Evaluated for Cross-reactivity and Microbial Interference with the NeuMoDx CT/NG Assay 2.0
| Bacteria/Fungi/ Viruses | Bacteria/Fungi/ Viruses | Bacteria/Fungi/ Viruses |
| --- | --- | --- |
| Acinetobacter baumannii | Kingella denitrificans | Peptostreptococcus productus |
| Acinetobacter calcoaceticus | Kingella kingae | Plesiomonas shigelloides |
| Acinetobacter lwoffii | Klebsiella oxytoca | Propionibacterium acnes |
| Actinomyces israelii | Klebsiella pneumoniae | Proteus mirabilis |
| Actinomyces pyogenes | Lactobacillus acidophilus | Proteus vulgaris |
| Aerococcus viridans | Lactobacillus brevis | Providencia stuartii |
| Aeromonas hydrophila | Lactobacillus crispatus | Pseudomonas aeruginosa |
| Alcaligenes faecalis | Lactobacillus jensenii | Pseudomonas fluorescens |
| Bacillus subtilis | Lactobacillus lactis | Pseudomonas putida |
| Bacteriodes fragilis | Lactobacillus oris | Rahnella aquatilis |
K230267 - Page 14 of 24
{14}
K230267 - Page 15 of 24
| Bacteroides caccae | Lactobacillus parabuchnerri | Rhizobium radiobacter |
| --- | --- | --- |
| Bacteroides ureolyticus | Lactobacillus vaginalis | Rhodospirillum rubrum |
| Bergeriella denitrificans | Lactococcus lactis cremoris | Saccharomyces cerevisiae |
| Bifidobacterium adolescentis | Legionella pneumophila | Salmonella minnesota |
| Bifidobacterium breve | Listeria monocytogenes | Salmonella typhimurium |
| Bifidobacterium longum | Micrococcus luteus | Serratia marcescens |
| Brevibacterium linens | Mobiluncus mulieris†† | Staphylococcus aureus |
| Campylobacter jejuni | Moraxella catarrhalis | Staphylococcus epidermidis |
| Candida albicans | Moraxella lacunata | Staphylococcus saprophyticus |
| Candida glabrata | Moraxella osloensis | Streptococcus agalactiae |
| Candida krusei | Morganella morganii | Streptococcus anginosus (group C) |
| Candida parapsilosis | Mycobacterium smegmatis | Streptococcus bovis |
| Candida tropicalis | Mycoplasma genitalium† | Streptococcus dysgalactiae |
| Chlamydia pneumoniae†† | Mycoplasma hominis† | Streptococcus equinus |
| Chromobacterium violaceum | Neisseria cinerea (strain 194) | Streptococcus mitis (F0392) |
| Citrobacter freundii | Neisseria elongata | Streptococcus mutans (serogroup A) |
| Clostridium perfringens | Neisseria flavescens | Streptococcus pneumoniae |
| Corynebacterium genitalium | Neisseria lactamica | Streptococcus pyogenes |
| Corynebacterium xerosis | Neisseria meningitidis (serogroup A) | Streptococcus salivarius |
| Cryptococcus neoformans | Neisseria meningitidis (serogroup B) | Streptococcus sanguinis |
| Deinococcus radiodurans | Neisseria meningitidis (serogroup C) | Streptomyces griseus |
| Derxia gummosa | Neisseria meningitidis (serogroup D) | Trichomonas vaginalis |
| Eikenella corrodens | Neisseria meningitidis (serogroup WL35) | Ureaplasma urealyticum* |
| Elizabethkingia miricola†† | Neisseria meningitidis (serogroup Y) | Veillonella parvula |
| Enterobacter aerogenes | Neisseria mucosa | Vibrio parahaemolyticus |
| Enterobacter cloacae | Neisseria perflava (strain 28) | Weissella paramesenteroides |
| Enterococcus avium | Neisseria perflava (strain 7078) | Yersinia enterocolitica |
| Enterococcus faecalis | Neisseria perflava (strain 831F) | cytomegalovirus** |
{15}
| Enterococcus faecium | Neisseria polysaccharea | epstein-Barr virus** |
| --- | --- | --- |
| Erysipelothrix rhusiopathiae | Neisseria sicca | hepatitis B virus*** |
| Escherichia coli | Neisseria subflava | Flavobacterium meningosepticum |
| Fusobacterium nucleatum | Neisseria weaveri | herpes simplex virus I** |
| Gardnerella vaginalis | Pantoea agglomerans | herpes simplex virus II** |
| Gemella haemolysans | Paracoccus denitrificans | human immunodeficiency virus 1*** |
| Haemophilus ducreyi | Peptoniphilus asaccharolyticus | human papillomavirus 16†† |
| Haemophilus influenzae | Peptostreptococcus anaerobius | human papillomavirus 18†† |
| Neisseria subflava (strain CDN-1.7) | hepatitis C virus*** | |
* Tested at 1 x $10^{6}$ CCU/mL, ** tested at 1 x $10^{6}$ copies/mL, *** tested at 1 x $10^{6}$ IU/mL, † tested at 1 x $10^{6}$ CFU/mL, †† tested at 10 ng/mL purified DNA or RNA, CCU= color changing units, CFU= colony forming units.
# Interference with Substances Potentially Present at the Anatomic Collection Site
The analytical performance of the NeuMoDx CT/NG Assay 2.0 was evaluated in the presence of a panel of potentially interfering substances that may be found in urogenital sites during urine or vaginal swab sample collection. Each potential interferent was spiked into pooled clinical, negative urine and pooled clinical negative vaginal swabs in UVT at clinically relevant concentrations indicated in Table 9. These samples were spiked with CT and NG at $3\mathrm{x}$ LoD in urine or $\leq 3\mathrm{x}$ LoD in pooled negative clinical swabs. All tests generated expected results with $100\%$ detection of CT and NG, thereby demonstrating that the substances listed at the concentrations shown in the table below do not interfere with the performance of NeuMoDx CT/NG Assay 2.0.
Table 9: Endogenous/Exogenous Interfering Substances Tested
| Substances in Urine | CT | NG |
| --- | --- | --- |
| | Positive (pos/total) | Positive (pos/total) |
| 7% Blood | 4/4 | 4/4 |
| PBMCs 10^5 Cells/mL | 4/4 | 4/4 |
| Acidic Urine pH 4.0 | 4/4 | 4/4 |
| Alkaline Urine pH 9.05 | 4/4 | 4/4 |
| Glucose 10.0 mg/mL | 3/3* | 3/3* |
| Albumin 10.5 mg/mL | 4/4 | 4/4 |
| Bilirubin 1.4 mg/dL | 4/4 | 4/4 |
| Progesterone Cream 7.5 mg/mL | 4/4 | 4/4 |
| Vagisil Deodorant Powder 3.6 mg/mL | 4/4 | 4/4 |
| Norforms Deodorant Supp. 2.8 mg/mL | 4/4 | 4/4 |
K230267 - Page 16 of 24
{16}
| Talcum Powder 12.0 mg/mL | 4/4 | 4/4 |
| --- | --- | --- |
| Aspirin 40.0 mg/mL | 4/4 | 4/4 |
| Azithromycin 2.2 mg/mL | 4/4 | 4/4 |
| Doxycycline 4.5 mg/mL | 4/4 | 4/4 |
| Acetaminophen 3.8 mg/mL | 4/4 | 4/4 |
| 5% Seminal Fluid | 4/4 | 4/4 |
*One replicate was IND
| Substances in Blood Concentration/Dosing Plan - Endogenous | CT | NG |
| --- | --- | --- |
| | Positive (pos/total) | Positive (pos/total) |
| Blood 3.5% v/v | 4/4 | 4/4 |
| Mucin (Saturated Swab)* | 4/4 | 4/4 |
| PBMCs 10^5 Cells/mL | 4/4 | 4/4 |
| Interference Samples-Concentration/Dosing Plan - Exogenous | | |
| Seminal Fluid 25.1 mg/mL | 4/4 | 4/4 |
| Progesterone Cream 3.9 mg/mL | 4/4 | 4/4 |
| Vagisil Anti Itch Cream 3.9 mg/mL | 4/4 | 4/4 |
| Clotrimazole Vaginal Cream 6.3 mg/mL | 4/4 | 4/4 |
| Preparation H Hemorrhoidal Cream 6.9 mg/mL | 4/4 | 4/4 |
| Miconazole 37.1 mg/mL | 4/4 | 4/4 |
| Monistat 17.8 mg/mL | 4/4 | 4/4 |
| Abreva Cold Sore Cream 4.1 mg/mL | 4/4 | 4/4 |
*Mucin dosed by swab from a 0.8% stock
4. Assay Reportable Range:
Not applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
1. Internal Control
The full process internal control, Sample Process Control (SPC1), serves as both an extraction and PCR internal control. The primers and probe specific for SPC1 are included in each well of the NeuMoDx CT/NG Test Strip. The SPC1 target itself is encapsulated bacteriophage T7 that is incorporated into each well of the NeuMoDx Extraction Plate. The SPC1 is co-extracted with the DNA from each sample processed and enables monitoring of the performance of the entire test process (nucleic acid isolation and real-time PCR amplification/detection) and key process steps not monitored actively by the NeuMoDx Software. The NeuMoDx CT/NG Assay 2.0 incorporates the results of SPC1 amplification as part of the results processing algorithm.
2. External Controls
External controls are not provided with the NeuMoDx CT/NG Assay 2.0; however, testing of external positive and negative controls is recommended in the assay labeling. External controls should be tested in conformance with local, state, and/or federal regulations or accreditation requirements and each laboratory's standard quality control procedures.
K230267 - Page 17 of 24
{17}
# 3. Calibrators
Not Applicable.
# 4. Stability
The stability of urine and swab specimens was evaluated in an analytical study which used individual contrived positive samples in addition to negatives. Contrived positive CT/NG samples consisted of pooled negative clinical vaginal swabs and negative urine spiked with CT and NG at $\leq 3\mathrm{x}$ LoD. Samples were stored onboard for up to 24 hours (h), or at $2 - 8^{\circ}\mathrm{C}$ for up to 7 days, before being processed separately on the N96 and N288. The positive samples remained positive and negative samples remained negative after storage at $2 - 8^{\circ}\mathrm{C}$ up to 7 days and up to $8\mathrm{h}$ onboard N96 and N288 systems. The study results demonstrated no effect on stability of CT and NG detection and that urine and vaginal swab samples can be stored up to 7 days at $2 - 8^{\circ}\mathrm{C}$ and up to $8\mathrm{h}$ when stored on NeuMoDx systems.
# 6. Detection Limit:
The limit of detection (LoD) of the NeuMoDx CT/NG Assay 2.0 on N96 and N288 was evaluated by testing spiked samples prepared in pooled negative clinical urine and vaginal swab matrix in UVT. The quantified stocks were diluted in pooled urine or swab matrix using a range of concentrations at five different levels (urine: $15\mathrm{EB / mL}$ to $0\mathrm{EB / mL}$ for CT and 2.5 cells/ mL to 0 cells/mL for NG; and vaginal swabs: $20\mathrm{EB / mL}$ to $0\mathrm{EB / mL}$ for CT and 5 cells/ mL to 0 cells/mL for NG). The LoD of the NeuMoDx CT/NG Assay 2.0 was initially estimated by testing separate dilutions of CT (serovar D) elementary bodies (EB) and NG (strain B5025) cells at five levels in both pooled clinical negative urine and vaginal swab matrix. Testing was performed on 20 replicates per level across multiple instruments, over three days, using three key reagent lots. The LoD is defined as the lowest concentration for each strain at which $95\%$ (19/20) of replicates yield a positive result. Further testing included two CT Serovars (D, J) and two NG strains (B5025, NHI 1) for each system. The highest (least sensitive) LoD value was finalized by accepting the target level reaching $\geq 95\%$ positivity for the assay across all system types and serovars/strains and reported as the final claimed LoD (Table 10).
Table 10. Claimed LoD of NeuMoDx CT/NG Assay 2.0 on N96 and N288 in CT serovars and NG strains
| Specimen Type | Chlamydia trachomatis LoD (EB/ mL) | Neisseria gonorrhoeae LoD (cells/ mL) |
| --- | --- | --- |
| Urine | 7.08 | 0.68 |
| Vaginal Swabs | 12.54 | 0.87 |
# Analytical Reactivity (Strain Inclusivity):
The ability of NeuMoDx CT/NG Assay 2.0 to detect 13 serovars of CT serovars, and 20 strains of NG was evaluated by spiking each of the serovars/strains into negative, pooled urine matrix and pooled UVT at within $2\mathrm{x}$ claimed LoD concentrations. Each serovar or strain dilution was detected $100\%$ with the NeuMoDx CT/NG 2.0 Assay in a minimum of 3 replicates. The CT serovars and NG strains tested in the study are listed in Table 11.
K230267 - Page 18 of 24
{18}
Table 11: Determination of Inclusivity for CT and NG on NeuMoDx CT/NG Assay 2.0
| Chlamydia trachomatis Serovars | Neisseria gonorrhoeae Strains |
| --- | --- |
| A | Strain 2686 |
| B | MHD 446 [NCTC 10933] |
| Ba | ATCC 9793 |
| C | Strain CDC Ng-98 |
| E* | BDMS 8658 [CIP 104217] |
| F | 83F0120 [DC-83-82] |
| H | Strain NCTC 8375 [B 5025] |
| I | Strain BDMS T4145 [CDC 83015965, Difco A0726] (Spectinomycin resistant) |
| K | Strain ODH 2915 [NCTC 10928] |
| LGV I | Strain 40836 (Uri) |
| LGV II | Strain B-1094 |
| LGV III | Strain MHD 340 [NCTC 10929] |
| G | 83F0091 [HWD 4345] |
| | WHO V [CDC 78-63856] |
| | 76.061782 |
| | F-18 [89-018314, CDC 10,001, P935] |
| | Strain FA1090 (Streptomycin resistant) |
| | 83F0120 [DC-83-82] |
| | Strain C-58 [D-10] |
| | Strain CDC Ng-116 |
| | Strain GC/CB/001 [CDC M-2] |
*Euv (Swedish variant) was also evaluated and tested positive
## Competitive Inhibition:
Due to potential of co-infection of CT and NG, where one target may be at a disproportionately higher concentration than the other (and adversely impact the detection of target at a lower concentration), detection of coinfection of CT at target levels near 3x LoD were chosen in the presence of competing co-infection of 10⁶ cells/mL NG, or for NG at target levels near 3x LoD in the presence of 10⁶ EB/mL CT. Competitive analyses revealed 100% detection of lower concentration pathogen in the presence of high concentration of competing pathogen. The results of the study demonstrated no impact to sensitivity for either target in combinations of high and low levels, respectively.
K230267 - Page 19 of 24
{19}
7. Assay Cut-Off:
Assay result from each test is determined by a cut-off based on the end point ratio (EPR), end point signal (EP), and cycle threshold (Ct). A test may be negative, positive, invalid, or unresolved as determined by these measurements for the detection of CT and NG DNA.
8. Accuracy (Instrument):
Not applicable.
9. Carry-Over:
A two-part analytical study was conducted on N96 and N288 to evaluate potential carryover contamination while performing a within and across run study. High-positive CT/NG samples were prepared with spiking pooled negative urine and neat UVT media at 10⁶ EB/mL for CT and 10⁶ cells/mL for NG. Negative samples were pooled negative urine or UVT without spiking with CT and NG. The first part of the study involved the processing of high positive CT/NG samples alternated with negative samples in an alternating, checkerboard arrangement. The second part of the study involved a full run of high positive samples processed in one run on an instrument followed by a full run of negative samples processed immediately after the high positive run across the same instrument. The study included 4 runs (2 alternating checkerboard, 1 positive followed by a negative run).
The overall carryover rate of the NeuMoDx CT/NG Assay 2.0 for urine and swab specimens on the N288 and N96 was 0% (100% agreement with expected results) when testing high titers of target concentration of CT and NG adjacent to negative samples processed within the same run or in alternate runs.
10. Matrix Equivalence:
A matrix equivalency study was conducted to demonstrate the equivalence of samples contrived in neat BD UVT used in analytical testing to samples collected in clinical negative vaginal swab matrix (collected in BD UVT) when testing with the NeuMoDx CT/NG Assay 2.0. Performance of each matrix was assessed in parallel using moderate positive (5x LoD), low positive (2x LoD), and negative samples.
CT/ NG targets were dosed together in positive samples. Negative samples consisted of only the matrix being evaluated. Each matrix was tested in replicates of n=10 at 5x LoD and n=30 at 2x LoD. Ten replicates were tested in each matrix lacking both analytes, in neat BD UVT and negative clinical vaginal swab matrix (collected in BD UVT).
The 5x LoD and 2x LoD level produced rates of 100% positivity in both the CT and NG targets. Negative samples in both matrices resulted in no amplification of the CT or NG targets, with valid negative results achieved at a rate of 100%. These results demonstrated matrix equivalency.
K230267 - Page 20 of 24
{20}
B Comparison Studies:
1. Method Comparison with Predicate Device:
See Clinical Studies section below.
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
The clinical performance of the NeuMoDx CT/NG Assay 2.0 was evaluated in a prospective, multi-center study conducted at 14 collection sites. The participating testing sites, at geographically diverse locations across the U.S., included family practice, obstetrics/gynecology practice, public health clinics, STD and family planning clinics, college campus and teen clinics, and emergency room/urgent care locations. A single neat urine from each male and female subject and one self-collected vaginal swab (collected in a clinical setting) and two clinician-collected vaginal swabs were collected from each female subject. All evaluable samples were tested with the NeuMoDx CT/NG Assay 2.0 in accordance with package insert instructions for use. The samples were split between three external laboratories and tested on N96 and N288 systems. The comparator testing was performed at a separate reference laboratory.
A total of 4017 participants with a mean age of 31±10 years (median: 29; range:15–77 years) were enrolled in the study, including 1825 males and 2192 females. Among the 1825 males enrolled in the study, five participants were excluded after consent because of not meeting eligibility criteria (n=3), previous enrollment (n=1), or withdrawal owing to insufficient urine collection (n=1), which resulted in 1820 valid male participants. Among the 2192 females enrolled, 36 were excluded after consent, with reasons including ineligibility or unconfirmed eligibility (n=13), withdrawal from the study (n=4), and samples evaluable by the NeuMoDx CT/NG Assay 2.0 not collected (n=19), leaving 2156 females included in the study.
Of the remaining 1820 male participants, a male urine (MU) specimen was collected from each, but 119 specimens were excluded from analysis owing to protocol deviation, collection or instrument device events, and other specimen testing issues which prevented a final result from being obtained by NeuMoDx CT/NG Assay 2.0 or comparator method, which left data from a total of 1701 male subjects included in the study.
Of the 1701 MU specimens, for CT, there were 10 cases where either a valid NeuMoDx CT/NG Assay 2.0 result could not be determined (n=2) or where the PIS result could not be determined (n=8), therefore there were 1691 MU specimens included in the primary analysis of performance for CT. Similarly, for NG, there were 3 cases where a valid NeuMoDx CT/NG Assay 2.0 result could not be determined, therefore there were 1698 MU specimens included in the primary analysis of performance for NG.
Among the evaluable specimens collected from the 2156 female participants, 138 female urine (FU) specimens and 140 self-collected vaginal swab (SCVS) specimens (in a clinical setting) were excluded due to protocol deviations, collection device events, instrument device events, and other events which prevented a final result from being obtained by either NeuMoDx CT/NG Assay 2.0 or comparator, allowing for 2018 and 2016 female subject specimens to be included in the study for FU and SCVS respectively.
K230267 - Page 21 of 24
{21}
Of the 2018 FU specimens, there were 11 cases where a valid NeuMoDx CT/NG Assay 2.0 result could not be determined for the CT target and 12 cases where a valid result was not obtained for the NG target; therefore, there were 2007 FU specimens included in the primary analysis for CT and 2006 for NG.
Of the 2016 SCVS specimens, all produced valid results for CT and NG with the NeuMoDx CT/NG Assay 2.0 and comparator test; these subjects are included in the primary data analysis.
## Clinical Sensitivity and Specificity
The clinical performance of the NeuMoDx CT/NG Assay 2.0 was assessed by comparing the test results obtained with the NeuMoDx CT/NG Assay 2.0 to a Patient Infected Status (PIS) for each subject. For female participants, the PIS was established from the results of FU and SCVS in a clinical setting tested by two FDA-cleared NAAT comparator assays. Females were classified as infected if at least one positive result was obtained by each assay. Any other combination of results was considered a non-infected status. Male subjects were classified as infected if both comparator urine NAAT results were positive. If the MU results were conflicting (i.e., one positive and one negative), a third FDA-cleared NAAT method was performed as a tie-breaker to male PIS.
The clinical sensitivity and specificity of the NeuMoDx CT/NG Assay 2.0, when used with SCVS and urine specimens, was calculated in comparison with the PIS along with prevalence rates observed during the clinical study in the tables 12 and 13 below.
Table 12. NeuMoDx CT/NG Assay 2.0 Performance vs PIS for CT Detection
| Specimen Type | Symptom Status | n | TP | FP | TN | FN | Prev. | Sensitivity (95% CI) | Specificity (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| SCVS | Asymp | 1052 | 43 | 2 | 1007 | 0 | 4.1% | 100.0% (91.8%, 100.0%) | 99.8% (99.3%, 99.9%) |
| | Symp | 964 | 58 | 7 | 896 | 3 | 6.3% | 95.1% (86.5%, 98.3%) | 99.2% (98.4%, 99.6%) |
| | All | 2016 | 101 | 9 | 1903 | 3 | 5.2% | 97.1% (91.9%, 99.0%) | 99.5% (99.1%, 99.8%) |
| FU | Asymp | 1054 | 40 | 1 | 1010 | 3 | 4.1% | 93.0% (81.4%, 97.6%) | 99.9% (99.4%, 100.0%) |
| | Symp | 953 | 56 | 3 | 889 | 5 | 6.4% | 91.8% (82.2%, 96.4%) | 99.7% (99.0%, 99.9%) |
| | All | 2007 | 96 | 4 | 1899 | 8 | 5.2% | 92.3% (85.6%, 96.1%) | 99.8% (99.5%, 99.9%) |
| MU | Asymp | 1227 | 102 | 0 | 1123 | 2 | 8.5% | 98.1% (93.3%, 99.5%) | 100.0% (99.7%, 100.0%) |
| | Symp | 464 | 82 | 1 | 377 | 4 | 18.5% | 95.3% (88.6%, 98.2%) | 99.7% (98.5%, 100.0%) |
| | All | 1691 | 184 | 1 | 1500 | 6 | 11.2% | 96.8% (93.3%, 98.5%) | 99.9% (99.6%, 100.0%) |
FU = Female Urine, SCVS = Self-Collected Vaginal Swab (in a clinical setting), MU = Male Urine, Prev. = Prevalence, Asymp = Asymptomatic, Symp = Symptomatic, TP = True Positive, FP = False Positive, TN = True Negative, FN = False Negative.
Table 13. NeuMoDx CT/NG Assay 2.0 Performance vs PIS for NG Detection
| Specimen Type | Symptom Status | n | TP | FP | TN | FN | Prev. | Sensitivity (95% CI) | Specificity (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| SCVS | Asymp | 1052 | 24 | 0 | 1028 | 0 | 2.3% | 100.0% (86.2%, 100.0%) | 100.0% (99.6%, 100.0%) |
| | Symp | 964 | 20 | 2 | 941 | 1 | 2.2% | 95.2% (77.3%, 99.2%) | 99.8% (99.2%, 99.9%) |
| | All | 2016 | 44 | 2 | 1969 | 1 | 2.2% | 97.8% (88.4%, 99.6%) | 99.9% (99.6%, 100.0%) |
| FU | Asymp | 1053 | 22 | 0 | 1029 | 2 | 2.3% | 91.7% (74.2%, 97.7%) | 100.0% (99.6%, 100.0%) |
| | Symp | 953 | 20 | 1 | 931 | 1 | 2.2% | 95.2% (77.3%, 99.2%) | 99.9% (99.4%, 100.0%) |
| | All | 2006 | 42 | 1 | 1960 | 3 | 2.2% | 93.3% (82.1%, 97.7%) | 99.9% (99.7%, 100.0%) |
K230267 - Page 22 of 24
{22}
| MU | Asymp | 1231 | 11 | 1 | 1219 | 0 | 0.9% | 100.0% (74.1%, 100.0%) | 99.9% (99.5%, 100.0%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Symp | 467 | 81 | 1 | 384 | 1 | 17.6% | 98.8% (93.4%, 99.8%) | 99.7% (98.5%, 100.0%) |
| | All | 1698 | 92 | 2 | 1603 | 1 | 5.5% | 98.9% (94.2%, 99.8%) | 99.9% (99.5%, 100.0%) |
FU = Female Urine, SCVS = Self-Collected Vaginal Swab (in a clinical setting), MU = Male Urine, Prev. = Prevalence, Asymp = Asymptomatic, Symp = Symptomatic, TP = True Positive, FP = False Positive, TN = True Negative, FN = False Negative.
## D Clinical Cut-Off:
Not applicable
## E Expected Values/Reference Range:
### Expected Values
The spread of infection with CT and/or NG in patient populations depends on risk factors such as age, gender, type of clinic, and the sensitivity of the test used to detect infections. The observed positivity rate of the NeuMoDx CT/NG Assay 2.0 in females and males during the clinical study are shown below in Table 14.
Table 14. NeuMoDx CT/NG Assay 2.0 Positivity Rate for CT and NG Detection by Clinical Sites
| | Chlamydia trachomatis
% Positivity rate | | | Neisseria gonorrhoeae
% Positivity rate | | |
| --- | --- | --- | --- | --- | --- | --- |
| Study Site | SCVS | MU | FU | SCVS | MU | FU |
| 1 | 9.72 | 15.43 | 8.92 | 4.17 | 17.24 | 4.23 |
| 2 | 11.11 | 5.26 | 0.0 | 0.0 | 10.53 | 0.0 |
| 3 | 3.53 | N/A | 3.53 | 2.35 | N/A | 2.35 |
| 4 | 7.19 | 11.59 | 5.98 | 3.70 | 7.30 | 3.30 |
| 5 | 2.54 | 11.84 | 2.56 | 1.69 | 9.21 | 1.71 |
| 6 | 2.78 | 10.04 | 2.86 | 0 | 7.34 | 0 |
| 7 | 8.73 | 19.55 | 9.21 | 3.06 | 4.55 | 3.08 |
| 8 | 5.99 | 16.67 | 6.06 | 1.80 | 1.67 | 1.82 |
| 9 | 0.88 | 2.12 | 0.59 | 1.18 | 0.42 | 0.88 |
| 10 | 4.61 | 13.74 | 4.57 | 0.46 | 1.53 | 0.46 |
| 11 | 0.0 | 3.45 | 0.0 | 0.0 | 0.0 | 0.0 |
| 12 | 7.69 | 13.64 | 7.69 | 0.0 | 4.55 | 0.0 |
| 13 | 0.0 | 5.71 | 0.0 | 0.0 | 0.0 | 0.0 |
| 14 | 2.25 | 2.99 | 1.14 | 0.0 | 1.49 | 0.0 |
| All sites | 5.46 | 11.01 | 4.98 | 2.28 | 5.54 | 2.14 |
N/A- Samples not enrolled at site.
## Indeterminate or Unresolved Result Rates
Invalid (indeterminate or unresolved) rates were calculated for each of the N96 and N288 systems, and for combined. Among all specimen types, instrument models, targets, and symptom status, 1/5738 (0.0%; 95% CI 0.0% to 0.1%) were Indeterminate (IND) and 12/5738 (0.2%; 95% CI 0.1% to 0.4%) were Unresolved (UNR).
K230267 - Page 23 of 24
{23}
VIII Proposed Labeling:
The labeling is acceptable and supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K230267 - Page 24 of 24
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.