Sofia Influenza A+B FIA on Sofia 2

K162438 · Quidel Corporation · PSZ · Apr 14, 2017 · Microbiology

Device Facts

Record IDK162438
Device NameSofia Influenza A+B FIA on Sofia 2
ApplicantQuidel Corporation
Product CodePSZ · Microbiology
Decision DateApr 14, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3328
Device ClassClass 2

Intended Use

The Sofia Influenza A+B FIA employs immunofluorescence to detect influenza A and influenza B viral nucleoprotein antigens in direct nasal swab, nasopharyngeal swab, and nasopharyngeal aspirate/wash specimens and nasopharyngeal swab and nasopharyngeal aspirate/wash specimens in transport media from symptomatic patients. This qualitative test is intended for use as an aid in the rapid differential diagnosis of acute influenza A and influenza B viral infections. The test is not intended to detect influenza C antigens. A negative test is presumptive and it is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative results do not preclude influenza virus infections and should not be used as the sole basis for treatment or other patient management decisions. This test is intended for professional and laboratory use. The Sofia Influenza A+B FIA may be used with Sofia or Sofia 2.

Device Story

Sofia 2 is a microprocessor-controlled, desktop-sized immunofluorescence analyzer for detecting influenza A and B viral nucleoproteins. Input: patient samples (nasal/nasopharyngeal swabs or aspirates/washes) processed in reagent tubes to disrupt virus particles, then applied to test cassettes. Operation: cassettes inserted into Sofia 2 for automated scanning of test strips via UV light source and fluorescent tag detection. Modes: 'Walk Away' (automated timing) or 'Read Now' (manual timing). Output: qualitative results (Positive/Invalid) displayed on touchscreen or printed. Used in professional/laboratory settings by clinicians/technicians. Device aids rapid differential diagnosis; negative results require confirmation by culture or molecular assay. Benefits include faster clinical decision-making and point-of-care testing capability.

Clinical Evidence

Bench testing only. Studies compared Sofia 2 to the predicate Sofia system, including Limit of Detection (LoD), precision, reproducibility (intra/inter-operator and intra/inter-laboratory), and method comparison using clinical sample panels. Results confirmed equivalent qualitative performance between the two systems. Early read capability in 'Walk Away' mode allows positive detection as early as 3 minutes depending on viral load.

Technological Characteristics

Immunofluorescence-based detection of viral nucleoproteins. System uses UV light source to excite fluorescent tags on test strips. Dimensions: 19.7 cm x 11.4 cm x 12.7 cm; weight ~2.5 lbs. Power: 100-240 VAC or rechargeable lithium polymer battery. Connectivity: USB port for external printer, optional network printer. User interface: 4-inch color LCD touchscreen. Software: microprocessor-controlled with method-specific algorithms for image processing and result interpretation.

Indications for Use

Indicated for symptomatic patients requiring rapid differential diagnosis of acute influenza A and B viral infections. Specimens include direct nasal swabs, nasopharyngeal swabs, and nasopharyngeal aspirate/wash (fresh or in transport media). Not for influenza C detection.

Regulatory Classification

Identification

An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria: (i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method: (A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent. (B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent. (ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method: (A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent. (B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent. (2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies. (3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria: (i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains. (ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate. (iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by: (A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or (B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access. (4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain: (i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus. (ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by: (A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or (B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is a common symbol associated with healthcare and medicine. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 14, 2017 Quidel Corporation Jennifer S. Rial Director, Regulatory Affairs 12544 High Bluff Drive, Suite 200 San Diego CA 92130 Re: K162438 Trade/Device Name: Sofia® Influenza A+B FIA on Sofia 2 Regulation Number: 21 CFR 866.3328 Regulation Name: Influenza virus antigen detection test systems Regulatory Class: II Product Code: PSZ Dated: March 13, 2017 Received: March 14, 2017 Dear Ms. Rial: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Steven R. Gitterman -S Uwe Scherf Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use # 510(k) Number (if known) K162438 Device Name Sofia® Influenza A+B FIA on Sofia 2 ### Indications for Use (Describe) The Sofia Influenza A+B FIA employs immunofluorescence to detect influenza B viral nucleoprotein antigens in direct nasal swab, nasopharyngeal swab, and nasopharyngeal aspirate/wash specimens and nasopharyngeal swab and nasopharyngeal aspirate/wash specimens in transport media from symptomatic patients. This qualitative test is intended for use as an aid in the rapid differential diagnosis of acute influenza B viral infections. The test is not intended to detect influenza C antigens. A negative test is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative results do not preclude influenza virus infections and should not be used as the sole basis for treatment or other patient management decisions. This test is intended for professional and laboratory use. The Sofia Influenza A+B FIA may be used with Sofia or Sofia 2. Performance characteristics for influenza A and B were established during February through March 2011 when influenza viruses A/California 7/2009 (2009 H1N1), A/Perth/16/2009 (H3N2), and B/Brisbane/60/2008 (Victoria-Like) were the predominant influenza viruses in circulation according to the Mortality Weekly Report from the CDC entitled "Update: Influenza Activity--United States, 2010-2011 Season, and Composition of the 2011-2012 Influenza Vaccine." Performance characteristics may vary against other emerging influenza viruses. If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, samples should be collected with appropriate infection sor novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture samples. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------| | <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square shape with rounded corners, divided into two sections. The top section is a gradient of colors, including red, orange, yellow, green, blue, and purple. The bottom section is a smaller square filled with a blue and purple gradient. Below the square is the word "QUIDEL" in a modern, sans-serif font. ### 5. 510(K) SUMMARY {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square shape with a smaller square cut out of the lower right corner. The larger square is filled with a gradient of colors, including red, orange, yellow, green, blue, and purple. The word "QUIDEL" is written in a sans-serif font below the square. #### 5.1. Submitter Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045 #### 5.2. Submission Contact Jennifer S. Rial #### 5.3. Date Prepared August 26, 2016 #### 5.4. Proprietary and Established Names Sofia® Influenza A+B FIA on Sofia 2 #### 5.5. Common Name Same as above #### 5.6. Regulatory Information | Product Code / Name | Class | Regulatory Section | Panel | |--------------------------------------------------------------|-------|--------------------|--------------| | PSZ - Devices detecting influenza a, b, and c virus antigens | II | 21 CFR 866.3328 | Microbiology | | KHO - Fluorometer for clinical use | | | | #### 5.7. Predicate Device Sofia® Influenza A+B FIA on Sofia (K153012, K131606, and K112177) #### 5.8. Device Description The Sofia Influenza A+B FIA employs immunofluorescence technology that is used with Sofia and Sofia 2 to detect influenza virus nucleoproteins. This test allows for the differential detection of influenza A and influenza B antigens. The patient sample is placed in the Reagent Tube, during which time the virus particles in the sample are disrupted, exposing internal viral nucleoproteins. After disruption, the sample is dispensed into the Test Cassette sample well. From the sample migrates through a test strip containing various unique chemical environments. If influenza viral antigen is present, they will be trapped in a specific location. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a stylized square divided into two sections. The larger section is a gradient of colors, including yellow, orange, red, purple, blue, and green. The smaller section is a darker shade of blue. Below the square is the word "QUIDEL" in a sans-serif font. Note: Depending upon the user's choice, the Test Cassette is either placed inside of Sofia or Sofia 2 for automatically timed development (WALK AWAY Mode) or placed on the counter or bench top for a manually timed development and then placed into Sofia 2 to be scanned (READ NOW Mode). Sofia and Sofia 2 will scan the test strip and measure the fluorescent signal by processing the results using method-specific algorithms. Sofia and Sofia 2 will display the test results (Positive, or Invalid) on the screen. The results can also be automatically printed on an integrated printer if this option is selected. Sofia 2 is a microprocessor-controlled device about the size of a desk top telephone and weighs less than 3 pounds. Sofia 2 uses a fluorescent tag that is illuminated by an Ultraviolet (UV) light source to generate specific results. #### 5.9. Intended Use The Sofia Influenza A+B FIA employs immunofluorescence to detect influenza A and influenza B viral nucleoprotein antigens in direct nasal swab, nasopharyngeal swab, and nasopharyngeal aspirate/wash specimens and nasopharyngeal swab and nasopharyngeal aspirate/wash specimens in transport media from symptomatic patients. This qualitative test is intended for use as an aid in the rapid differential diagnosis of acute influenza A and influenza B viral infections. The test is not intended to detect influenza C antigens. A negative test is presumptive and it is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative results do not preclude influenza virus infections and should not be used as the sole basis for treatment or other patient management decisions. This test is intended for professional and laboratory use. The Sofia Influenza A+B FIA may be used with Sofia or Sofia 2. Performance characteristics for influenza A and B were established during February through March 2011 when influenza viruses A/California/7/2009 (2009 H1N1), A/Perth/16/2009 (H3N2), and B/Brisbane/60/2008 (Victoria-Like) were the predominant influenza viruses in circulation according to the Morbidity and Mortality Weekly Report from the CDC entitled "Update: Influenza Activity--United States, 2010-2011 Season, and Composition of the 2011-2012 Influenza Vaccine." Performance characteristics may vary against other emerging influenza viruses. If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, samples should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture samples. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square shape with rounded corners, filled with a gradient of colors resembling a rainbow. A smaller square is cut out from the lower right corner of the larger square. Below the square is the word "QUIDEL" in a stylized, sans-serif font. #### Substantial Equivalence Information: 5.10 - 1. Predicate Device Name: Sofia Influenza A+B FIA on Sofia - 2. Predicate 510(k) Numbers: K153012, K131606, and K112177 - 3. Comparison with Predicate: | Similarities and Differences | | | |---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Item | Sofia Influenza A+B FIA on Sofia | Sofia 2 Influenza A+B FIA on Sofia and Sofia 2 | | Intended Use | The Sofia Influenza A+B FIA employs<br>immunofluorescence to detect influenza A and<br>influenza B viral nucleoprotein antigens in nasal<br>swab, nasopharyngeal swab, and nasopharyngeal<br>aspirate/wash in fresh or transport media<br>specimens from symptomatic patients. This<br>qualitative test is intended for use as an aid in the<br>rapid differential diagnosis of acute influenza A<br>and influenza B viral infections. The test is not<br>intended to detect influenza C antigens. A negative<br>test is presumptive and it is recommended these<br>results be confirmed by viral culture or an FDA-<br>cleared influenza A and B molecular assay.<br>Negative results do not preclude influenza virus<br>infections and should not be used as the sole basis<br>for treatment or other patient management<br>decisions. This test is intended for professional and<br>laboratory use.<br>Performance characteristics for influenza A and B<br>were established during February through March<br>2011 when influenza viruses A/California/7/2009<br>(2009 H1N1), A/Perth/16/2009 (H3N2),<br>and<br>B/Brisbane/60/2008 (Victoria-Like) were<br>the<br>predominant influenza viruses in circulation<br>according to the Morbidity and Mortality Weekly<br>Report from the CDC entitled "Update: Influenza<br>Activity--United States, 2010-2011 Season, and<br>Composition of the 2011-2012 Influenza Vaccine."<br>Performance characteristics may vary against other<br>emerging influenza viruses.<br>If infection with a novel influenza virus is<br>suspected<br>based<br>on<br>current<br>clinical<br>and<br>epidemiological screening criteria recommended<br>by public health authorities, samples should be<br>collected<br>with appropriate infection control<br>precautions for novel virulent influenza viruses<br>and sent to state or local health department for<br>testing. Viral culture should not be attempted in<br>these cases unless a BSL 3+ facility is available to<br>receive and culture samples. | The Sofia Influenza A+B FIA employs<br>immunofluorescence to detect influenza A and<br>influenza B viral nucleoprotein antigens in direct<br>nasal swab, nasopharyngeal swab, and<br>nasopharyngeal aspirate/wash specimens and<br>nasopharyngeal swab and nasopharyngeal<br>aspirate/wash specimens in transport media from<br>symptomatic patients. This qualitative test is<br>intended for use as an aid in the rapid differential<br>diagnosis of acute influenza A and influenza B<br>viral infections. The test is not intended to detect<br>influenza C antigens. A negative test is<br>presumptive and it is recommended these results be<br>confirmed by viral culture or an FDA-cleared<br>influenza A and B molecular assay. Negative<br>results do not preclude influenza virus infections<br>and should not be used as the sole basis for<br>treatment or other patient management decisions.<br>This test is intended for professional and laboratory<br>use.<br>The Sofia Influenza A+B FIA may be used with<br>Sofia or Sofia 2.<br>Performance characteristics for influenza A and B<br>were established during February through March<br>2011 when influenza viruses A/California/7/2009<br>(2009 H1N1), A/Perth/16/2009 (H3N2),<br>and<br>B/Brisbane/60/2008 (Victoria-Like)<br>were<br>the<br>predominant influenza viruses in circulation<br>according to the Morbidity and Mortality Weekly<br>Report from the CDC entitled "Update: Influenza<br>Activity--United States, 2010-2011 Season, and<br>Composition of the 2011-2012 Influenza Vaccine.'"<br>Performance characteristics may vary against other<br>emerging influenza viruses.<br>If infection with a novel influenza virus is<br>suspected based on current clinical and<br>epidemiological screening criteria recommended<br>by public health authorities, samples should be<br>collected with appropriate infection control<br>precautions for novel virulent influenza viruses and<br>sent to state or local health department for testing.<br>Viral culture should not be attempted in these cases<br>unless a BSL 3+ facility is available to receive and<br>culture samples. | | Similarities and Differences | | | | Item | Sofia Influenza A+B FIA on Sofia | Sofia 2 Influenza A+B FIA on Sofia and Sofia 2 | | Calibration Check | Calibration Check required every 30 days or less,<br>as set by the supervisor. A special Calibration<br>Cassette is provided with the Installation Pack. | Same and uses the same Calibration Cassette | | Development Modes | Two basic assay development modes:<br>Walk-Away: User can walk away during the<br>assay cassette development period<br>Read Now: User manually times the assay<br>•<br>cassette development period outside of Sofia,<br>then places cassette in Sofia to image and<br>provide test result. | Same | | Development Time | 15 minutes for Sofia Influenza A+B FIA | Potential for early read in Walk-Away Mode.<br>Sofia 2 will image cassette at 3, 5, 8, 10, and 15<br>minutes until a positive result is received. | | System Components | | | | User interface | 3.5 inch diagonal color LCD display and numeric<br>keypad with function specific buttons | 4 inch color LCD touchscreen display | | User Types | Has 2 distinct security levels; user and supervisor<br>plus a Quidel only service level | Same | | Barcode<br>scanner(sample) | External hand held barcode scanner | Integrated barcode scanner but same functionality | | Barcode<br>scanner(cassette) | Integrated barcode scanner | Same using custom integrated 0.3 MP camera | | Assay / instrument<br>interface | Drawer (electro-mechanical) | Same (manual) | | Determine test type | Instrument scans barcode on cassette | Same | | Power Supply | 100 - 240 VAC, self-switching, or with 4 AA<br>batteries…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...