K033274 · Techlab, Inc. · MHI · Nov 4, 2003 · Microbiology
Device Facts
Record ID
K033274
Device Name
GIARDIA II
Applicant
Techlab, Inc.
Product Code
MHI · Microbiology
Decision Date
Nov 4, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 866.3220
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The GIARDIA II test is an enzyme immunoassay for the qualitative detection of Giardia lamblia cyst antigen in human fecal specimens. It is indicated for use with fecal specimens from patients with diarrhea to determine the presence of G. lamblia gastrointestinal infection. This test can be used for fecal specimens submitted for routine clinical testing from adults or children. FOR IN VITRO DIAGNOSTIC USE.
Device Story
The TECHLAB GIARDIA 11 is an enzyme-linked immunosorbent assay (ELISA) for the detection of Giardia lamblia in human fecal specimens. The device utilizes a direct format where the detecting antibody is conjugated to horseradish peroxidase (HRP), simplifying the assay by eliminating the need for an anti-rabbit IgG:HRP conjugate. The test uses a one-component substrate containing tetramethyl benzidine and peroxide, and a 0.6 N sulfuric acid stop solution. The device is used in a laboratory setting by trained personnel. The output is a colorimetric signal measured via optical density (OD) at 450 nm or 450/620 nm, which is compared against established thresholds to determine the presence of Giardia antigens. This aids clinicians in the diagnosis of giardiasis.
Clinical Evidence
No clinical data provided in the document.
Technological Characteristics
Enzyme immunoassay (EIA) for qualitative antigen detection. In vitro diagnostic device.
Indications for Use
Indicated for the detection of Giardia lamblia in human fecal specimens.
Regulatory Classification
Identification
Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Entamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Entamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasite Entamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
Related Devices
K052932 — CRYPTOSPORIDIUM II · Techlab, Inc. · Nov 4, 2005
K020583 — IVD GIARDIA ANTIGEN DETECTION ASSAY MODEL GL-96 · Ivd Research, Inc. · Sep 17, 2002
K982711 — PREMIER GIARDIA · Meridian Diagnostics, Inc. · Nov 25, 1998
K051929 — GIARDIA/CRYPTOSPORIDIUM CHEK · Techlab, Inc. · Nov 17, 2005
K103673 — GIARDIA/ CRYPTOSPORIDIUM QUIK CHEK · Techlab Inc., Corporate Research Center · Aug 18, 2011
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification ODE Review Memorandum
To: THE FILE
RE: DOCUMENT NUMBER K 033274 TECHLAB GIARDIA 11
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Reserved Class I device. The following items are present and acceptable:
1. The name and 510(k) number of the SUBMITTER'S previously cleared device.
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials.
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
The changes identified were as follows:
1. A direct format replaced an indirect format making the test simpler to perform. The detecting antibody is directly labeled with horse radish peroxidase thereby eliminating the anti rabbit IgG:HRP conjugate
2. A two component substrate was replaced by a one component substrate. Both contain the same two chemicals namely tetramethyl benzidine and peroxide.
3. Stop solution was weakened from 1M sulfuric acid to 0.6 N (equiv. to 0.3M) sulfuric acid.
4. A sample diluent (buffered protein solution with 0.02% Thimerosal) is included for use.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and performance characteristics which included a comparison of this device test results to the legally marketed predicate device using 110 positive and 93 negative specimens in one study and 88 positive and 40 negative specimens in another study; reproducibility, freeze-thaw and crossreactivity studies.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
Following are the Acceptance Criteria used to determine a successful validation:
Evaluate each test based upon the following specifications and refer to the diagram following.
Giardia Cysts Standard Curve (Plate 1-strips 1 & 2)
Rows A -D: Must be positive, ≥0.150 OD 450 and/or ≥0.090 OD450/620
Rows E - H: May vary between positive and negative
rCWP1 Antigen Standard Curve (Plate 1-strips 3 & 4)
Rows A - D: Must be positive, ≥0.150 OD 450 and/or ≥0.090 OD450/620
Rows E - H: May vary between positive and negative
Positive Control (Plate 1 - strip 5 Plate 2 - strip 5)
Rows A - H: All wells must be positive, ≥1.000 OD 450/620 and OD450
All wells must display a yellow color
{1}
2
Negative Control (Plate 1-strip 6, Plate 2-strip 4)
Rows A - H: All wells must be negative, < 0.150 on OD 450 and/or < 0.090 OD450/620
All wells must be visually clear
Eight Positive Fecals (Plate 2 - strip 1)
All wells must be positive, ≥0.150 OD 450 and/or ≥0.090 OD450/620
Sixteen Negative Fecals (Plate 2- strips 2 & 3)
All wells must be negative, < 0.150 OD 450 and/or < 0.090 OD450/620
Plate 1
| Giardia Cysts/mL | | rCWP1 Antigen ng/mL | | Pos. Ctr. | Neg. Ctr. | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | |
| 150,000 | 150,000 | 25 | 25 | + | - | | | | | | | |
| 15,000 | 15,000 | 12.5 | 12.5 | + | - | | | | | | | |
| 1,500 | 1,500 | 6.25 | 6.25 | + | - | | | | | | | |
| 750 | 750 | 3.13 | 3.13 | + | - | | | | | | | |
| 375 | 375 | 1.56 | 1.56 | + | - | | | | | | | |
| 188 | 188 | 0.78 | 0.78 | + | - | | | | | | | |
| 94 | 94 | 0.39 | 0.39 | + | - | | | | | | | |
| 47 | 47 | 0.2 | 0.2 | + | - | | | | | | | |
Plate 2
| + Fecals formalin | - Fecals formalin | - Fecals unpreserved | Neg. Ctr. | Pos. Ctr. |
| --- | --- | --- | --- | --- |
| 1 | 2 | 3 | 4 | 5 |
| + F | - F | - F | - | + |
| + F | - F | - F | - | + |
| + F | - F | - F | - | + |
| + F | - F | - F | - | + |
| + F | - F | - F | - | + |
| + F | - F | - F | - | + |
| + F | - F | - F | - | + |
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Statement and the Indications for Use Enclosure.
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular
{2}
3
modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, it is recommended that the device be determined substantially equivalent to the previously cleared device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.