CorDx Tyfast COVID-19 Ag Rapid Test; CorDx COVID-19 Ag Test
K240728 · CorDx, Inc. · QYT · Jun 21, 2024 · Microbiology
Device Facts
Record ID
K240728
Device Name
CorDx Tyfast COVID-19 Ag Rapid Test; CorDx COVID-19 Ag Test
Applicant
CorDx, Inc.
Product Code
QYT · Microbiology
Decision Date
Jun 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3984
Device Class
Class 2
Attributes
Pediatric
Intended Use
The CorDx Tyfast COVID-19 Ag Rapid Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. The CorDx Tyfast COVID-19 Ag Rapid Test does not differentiate between SARS-CoV and SARS-CoV-2. All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment. Positive results do not rule out co-infection with other respiratory pathogens. This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek follow up care from their healthcare provider. The performance characteristics for SARS-CoV-2 were established from September, 2023, to December, 2023, when SARS-CoV-2 Omicron was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected.
Device Story
Lateral flow immunoassay for qualitative detection of SARS-CoV-2 nucleocapsid protein antigen; utilizes anterior nasal swab specimens. Test strip in plastic cassette contains colloidal-gold conjugated monoclonal antibodies; sample extract migrates along membrane; viral antigen captured by specific monoclonal antibody at test line (T). Procedural control line (C) confirms proper wicking. Visually read by user in home setting; results available in 10 minutes. Negative results are presumptive; serial testing recommended for symptomatic individuals to improve sensitivity. Assists in identifying potential COVID-19 infection; guides users to seek professional medical follow-up if symptoms persist or worsen.
Clinical Evidence
Prospective clinical study of 740 symptomatic subjects (0-7 days post-symptom onset). Compared to 510(k)-cleared RT-PCR. Performance in 693 subjects (0-5 DPSO): PPA 85.6% (95% CI: 78.1-90.8%), NPA 99.5% (95% CI: 98.5-99.8%). Analytical studies included LoD (1x10^4 TCID50/ml), inclusivity testing against variants (Alpha, Gamma, Beta, Delta, Omicron), and interference/cross-reactivity panels.
Technological Characteristics
Lateral flow immunochromatographic membrane assay. Components: sample pad, reagent pad (colloidal-gold conjugated monoclonal antibody), reaction membrane (secondary antibody), absorbing pad. Plastic cassette housing. No electronic components. Sterilization via ethylene oxide (ISO 11135, ISO 10993-7).
Indications for Use
Indicated for rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigen in anterior nasal swab specimens from symptomatic individuals. For non-prescription home use by individuals aged 14+ testing themselves, or adults testing individuals aged 2+.
Regulatory Classification
Identification
For the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen in individuals 2 years and older.
K231795 — QuickVue COVID-19 Test · Quidel Corporation · Mar 22, 2024
K251289 — WELLlife COVID-19 Antigen Test Rx · Guangzhou Wondfo Biotech Co., Ltd. · Oct 22, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
June 21, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
CorDx, Inc. % Jinjie Hu President and Principal Consultant Axteria Biomed Consulting Inc. 8040 Cobble Creek Circle Potomac, Maryland 20854
Re: K240728
Trade/Device Name: CorDx Tyfast COVID-19 Ag Rapid Test; CorDx COVID-19 Ag Test Regulation Number: 21 CFR 866.3984 Regulation Name: Over-the-counter test to detect SARS-CoV-2 from clinical specimens Regulatory Class: Class II Product Code: QYT Dated: March 18, 2024 Received: March 18, 2024
Dear Jinjie Hu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Silke Digitally signed by Silke Schlottmann -S Schlottmann -S Date: 2024.06.21 10:54:26 -04'00 Silke Schlottmann, Ph.D. Deputy Assistant Director Bacteriology Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K240728
#### Device Name
CorDx Tyfast COVID-19 Ag Rapid Test; CorDx COVID-19 Ag Rapid Test
Indications for Use (Describe)
The CorDx Tyfast COVID- 19 Ag Rapid Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19.
This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
The CorDx Tyfast COVID- 19 Ag Rapid Test does not differentiate between SARS-CoV and SARS-CoV-2.
All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment. Positive results do not rule out co-infection with other respiratory pathogens.
This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek follow up care from their healthcare provider.
The performance characteristics for SARS-CoV-2 were established from September, 2023, when SARS-CoV-2 Omicron was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### CorDx, Inc.
### 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
- CorDx, Inc. SUBMITTER: 9540 Waples St. # C, San Diego, CA 92121
- Contact person: Jinjie Hu, Axteria BioMed Consulting Inc. 8040 Cobble Creek Circle Potomac, MD 20854 Tel: 1-(301)814-4985 Email: jinjie.hu@axteriabiomed.com
Date prepared: March 22, 2024
#### Purpose of Submission:
To obtain a substantial equivalence for the CorDx Tyfast COVID-19 Ag Rapid Test (also known as CorDx COVID-19 Ag Test in EUA)
{4}------------------------------------------------
#### CorDx, Inc.
- A. 510(k) Number: K240728
- B. Purpose for Submission: New Device
- C. Measure: Antigen of SARS-CoV-2
#### D. Type of Test: Qualitative chromatographic immunoassay
## E. Applicant:
CorDx, Inc.
- F. Proprietary and Established Names: CorDx Tyfast COVID-19 Ag Rapid Test
#### G. Regulatory Information:
- 1. Regulation: 21 CFR 866.3984 - Over-The-Counter Test to Detect SARS-Cov-2 From Clinical Specimens
- 2. Classification: Class II
- 3. Product Code: QYT
- 4. Panel: Microbiology
{5}------------------------------------------------
#### CorDx, Inc.
#### H. Intended Use:
#### 1. Intended use(s):
The CorDx Tyfast COVID- 19 Ag Rapid Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19.
This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
The CorDx Tyfast COVID- 19 Ag Rapid Test does not differentiate between SARS-CoV and SARS-CoV-2.
All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment. Positive results do not rule out co-infection with other respiratory pathogens.
This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek follow up care from their healthcare provider.
The performance characteristics for SARS-CoV-2 were established from September, 2023, to December, 2023, when SARS-CoV-2 Omicron was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected.
{6}------------------------------------------------
#### CorDx, Inc.
The CorDx Tyfast COVID-19 Ag Rapid Test is intended for non-prescription self-use and/or as applicable an adult lay user testing another person 2 years of age or older in a nonlaboratory setting.
- 2. Special condition for use statement(s): This device is intended for over-the-counter (OTC) use.
- 3. Special instrument requirements: None
#### I. Device Description:
The CorDx Tyfast COVID-19 Ag Rapid Test is a rapid, immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect nucleocapsid protein from SARS-CoV-2 from anterior nasal swab specimens. This test does not differentiate between SARS-CoV and SARS-CoV-2. The test kit includes the: test cassette, swab, tube with sample processing solution, tube holder (back of the box) and the Quick Reference Instructions (QRI).
The test strip enclosed in a cassette housing is comprised of the following components: sample pad, reagent pad, reaction membrane, and absorbing pad. The reagent pad contains colloidal-gold conjugated with a monoclonal antibody against the nucleocapsid protein of SARS-CoV-2; the reaction membrane contains the secondary antibody for the nucleocapsid protein of SARS-CoV-2. The whole strip is fixed inside a plastic cassette.
When the sample extract is added into the sample well, conjugates dried onto the reagent pad are dissolved and migrate along with the sample. If SARS-CoV-2 nucleocapsid antigen is present in the sample, a complex formed between the anti-SARS-2 conjugate and the viral antigen will be captured by the specific anti-SARS-2 monoclonal antibody coated on the test line region (T). Absence of the test line (T) suggests a negative result. To serve as a procedural control, a red line will always appear in the control line region (C) indicating that proper volume of sample has been added and membrane wicking has occurred.
#### J. Substantial Equivalence Information:
- 1. Predicate device name(s): ACON Laboratories, Inc. K230828, Flowflex COVID-19 Antigen Home Test
- 2. Predicate K number(s): K230828
- 3. Comparison with predicate
{7}------------------------------------------------
## CorDx Tyfast COVID-19 Ag Rapid Test
### CorDx, Inc.
| Parameter | CorDx Tyfast COVID-19 Ag<br>Rapid Test | Predicate (K230828)<br>Flowflex COVID-19 Antigen<br>Home Test | Similarity<br>or<br>Difference |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Intended Use | The CorDx Tyfast COVID- 19 Ag<br>Rapid Test is a visually read<br>lateral flow immunoassay device<br>intended for the rapid, qualitative<br>detection of SARS-CoV-2 virus<br>nucleocapsid protein antigen<br>directly in anterior nasal swab<br>specimens from individuals with<br>signs and symptoms of<br>COVID-19.<br>This test is for non-prescription<br>home use by individuals aged 14<br>years or older testing themselves,<br>or adults testing individuals aged<br>2 years or older.<br>The CorDx Tyfast COVID- 19 Ag<br>Rapid Test does not differentiate<br>between SARS-CoV and SARS-CoV-2.<br>All negative results are<br>presumptive. Symptomatic<br>individuals with an initial negative<br>test result must be re-tested once<br>between 48 and 72 hours after the<br>first test using an antigen test or a<br>molecular test for SARS-CoV-2.<br>Negative results do not preclude<br>SARS-CoV-2 infections or other<br>pathogens and should not be used<br>as the sole basis for treatment.<br>Positive results do not rule out co-<br>infection with other respiratory<br>pathogens. | The Flowflex COVID-19 Antigen<br>Home Test is a visually read lateral<br>flow immunoassay device intended<br>for the rapid, qualitative detection<br>of SARS-CoV-2 virus<br>nucleocapsid protein antigen<br>directly in anterior nasal swab<br>specimens from individuals with<br>signs and symptoms of COVID-<br>19 within the first 6 days of<br>symptom onset.<br>This test is for non-prescription<br>home use by individuals aged 14<br>years or older testing themselves,<br>or adults testing individuals aged 2<br>years or older.<br>The Flowflex COVID-19 Antigen<br>Home Test does not differentiate<br>between SARS-CoV and SARS-CoV-2.<br>All negative results are<br>presumptive. Symptomatic<br>individuals with an initial negative<br>test result must be re-tested once<br>between 48 and 72 hours after the<br>first test using either an antigen<br>test or a molecular test for SARS-CoV-2. Negative results do not<br>preclude SARS-CoV-2 infections<br>or other pathogens and should not<br>be used as the sole basis for<br>treatment.<br>Positive results do not rule out co-<br>infection with other respiratory<br>pathogens. | Same |
| Parameter | CorDx Tyfast COVID-19 Ag<br>Rapid Test | Predicate (K230828)<br>Flowflex COVID-19 Antigen<br>Home Test | Similarity<br>or<br>Difference |
| | This test is not a substitute for<br>visits to a healthcare provider or<br>appropriate follow-up and should<br>not be used to determine any<br>treatments without provider<br>supervision. Individuals who test<br>negative and experience continued<br>or worsening COVID-19 like<br>symptoms, such as fever, cough<br>and/or shortness of breath, should<br>seek follow up care from their<br>healthcare provider. | This test is not a substitute for<br>visits to a healthcare provider or<br>appropriate follow-up and should<br>not be used to determine any<br>treatments without provider<br>supervision. Individuals who test<br>negative and experience<br>continued or worsening<br>COVID-19 like symptoms, such<br>as fever, cough and/or shortness<br>of breath, should seek follow up<br>care from their healthcare<br>provider. | |
| | The performance characteristics<br>for SARS-CoV-2 were established<br>from September, 2023, to<br>December, 2023, when SARS-<br>CoV-2 Omicron was dominant.<br>Test accuracy may change as new<br>SARS-CoV-2 viruses emerge.<br>Additional testing with a lab-<br>based molecular test (e.g., PCR)<br>should be considered in situations<br>where a new virus or variant is<br>suspected. | The performance characteristics<br>for SARS-CoV-2 were<br>established from December 2022<br>to<br>March 2023 when SARS-CoV-2<br>Omicron was dominant. Test<br>accuracy may change as new<br>SARS-CoV-2 viruses emerge.<br>Additional testing with a lab-<br>based molecular test (e.g., PCR)<br>should be considered in<br>situations where a new virus or<br>variant is suspected. | |
| Intended Use<br>Environment | Over the counter use/self-testing | Over the counter use/self-testing | Same |
| Intended Use<br>population | Individuals with symptoms of<br>COVID-19 | Individuals with symptoms of<br>COVID-19 | Same |
| Organism<br>detected | SARS-CoV-2 | SARS-CoV-2 | Same |
| Analyte | Nucleocapsid protein antigen from<br>SARS-CoV-2 | Nucleocapsid protein antigen<br>from SARS-CoV-2 | Same |
| Regulation<br>number | 21 CFR 866.3984 | 21 CFR 866.3984 | Same |
| Principle of the<br>Technology | Qualitative lateral flow<br>immunoassay | Qualitative lateral flow<br>immunoassay | Same |
| Sample type | Anterior nasal swab specimens | Anterior nasal swab specimens | Same |
| Detection<br>format | Test cassette, visually read without<br>an instrument. | Test cassette, visually read<br>without an instrument. | Same |
| Assay Result | Qualitative (positive, negative,<br>invalid) | Qualitative (positive, negative,<br>invalid) | Same |
| Parameter | CorDx Tyfast COVID-19 Ag Rapid Test | Predicate (K230828) Flowflex COVID-19 Antigen Home Test | Similarity or Difference |
| Time to Result | Time to result: 10 minutes after the extracted sample is added to the test cassette sample well. | Time to result: 15 minutes after the extracted sample is added to the test cassette sample well. | Different<br>5 minutes less reading time |
| Controls | -Internal procedural controls: Each CorDx Tyfast COVID-19 Rapid Ag Test has a built-in "Control" region which serves as an internal procedural control when a colored line appears in the control line region ("C line"). The "C line” should always appear if the test has been performed correctly. If the "C line" does not appear at 10 minutes, the test result is invalid. It is recommended to review the instructions again and repeat the test with a new sample and a new cassette. If the problem persists, please stop using the product and contact CorDx for technical support.<br>-External controls: a separate external control kit is available for purchase from CorDx. The configurations of the CorDx Tyfast COVID-19 Ag Control Swab Kit from CorDx is: 10 positive swabs and 10 negative swabs per box.<br>The COVID-19 antigen positive control swab is composed of SARS-COV-2 recombinant antigen extract dried onto a swab. The COVID-19 antigen negative control swab is composed of negative sample matrix dried onto a swab. | -Internal procedural controls: The Flowflex Plus COVID-19 Home Test has a built-in internal procedural control. The colored biotin labeled particle will bind to streptavidin on nitrocellulose membrane and the colored mouse monoclonal antibody labeled particle will bind to goat anti-mouse antibody on nitrocellulose membrane to form a colored Control line (C) in the control line region. Formation of the Control line serves as an internal control indicating that proper volume of specimen has been added and membrane wicking has occurred. | Different |
| Reagent Storage | Store at 36~86°F/2~30°C in a place out of direct sunlight and out | Store Flowflex COVID-19 Antigen Home Test between 2-30°C (36-86°F) until use. | Same |
{8}------------------------------------------------
## CorDx Tyfast COVID-19 Ag Rapid Test
### CorDx, Inc.
{9}------------------------------------------------
## CorDx Tyfast COVID-19 Ag Rapid Test
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
## CorDx, Inc.
{10}------------------------------------------------
#### CorDx, Inc.
#### K. Standard/Guidance Document referenced (if applicable):
| Document Title | Issued by | Applicable study |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|-------------------|
| Reclassification order for DEN220028 and special controls under 21 CFR 866.3984<br>for Over- the-counter tests to detect SARS-CoV-2 from clinical specimens | FDA | All |
| ISO11135:2014, Sterilization of health care products - Ethylene oxide -<br>Requirements for development, validation and routine control of a sterilization<br>process for medical devices | ISO | Sterility |
| ISO 10993-7, Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide<br>Sterilization Residuals | ISO | Sterility |
| ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and<br>testing within a risk management process | ISO | Biocompatibility |
| EP5-A3 Evaluation of Precision of Quantitative Measurement Procedures | CLSI | Precision |
| EP12-A2 User Protocol for Evaluation of Qualitative Test Performance | CLSI | Precision |
| EP25A Evaluation of Stability of In Vitro Diagnostic Reagents | CLSI | Reagent stability |
| EP37 Supplemental Tables for Interference Testing in Clinical Chemistry | CLSI | Interference |
#### L. Test Principle:
When the sample extract is added into the sample well, conjugates dried onto the reagent pad are dissolved and migrate along with the sample. If SARS-CoV-2 nucleocapsid antigen is present in the sample, a complex formed between the anti-SARS-2 conjugate and the viral antigen will be captured by the specific anti-SARS-2 monoclonal antibody coated on the test line region (T). Absence of the test line (T) suggests a negative result. To serve as a procedural control, a red line will always appear in the control line region (C) indicating that proper volume of sample has been added and membrane wicking has occurred.
#### M. Performance Characteristics
#### 1. Analytical performance:
#### a. Precision Study by Lots
The purpose of the study was to assess lot-to-lot variability of three different lots of the CorDx Tyfast COVID-19 Ag Rapid Test kits. The study included a sample panel consisting of a negative sample, a low positive (2x LoD) sample, and a positive (4x LoD) sample. The positive samples were prepared by spiking UV-inactivated SARS-CoV-2 in pooled negative Nasal Wash Matrix. The study results demonstrated 100% consistent performance of the CorDx Tyfast COVID-19 Ag Rapid Test for the detection of SARS-CoV-2 virus antigen when performed under different conditions of variability (days, runs, operators, kit lots and sample replicates).
| | Test result(#Neg/60, # Pos/60) | | |
|-------|--------------------------------|-------|-------|
| Lot | NEG | 2xLoD | 4xLoD |
| Lot 1 | 60/60 | 60/60 | 60/60 |
| Lot 2 | 60/60 | 60/60 | 60/60 |
#### Table 1. Precision Study Results
{11}------------------------------------------------
#### CorDx Tyfast COVID-19 Ag Rapid Test
#### CorDx, Inc.
| Lot 3 | 60/60 | 60/60 | 60/60 |
|-----------|---------|---------|---------|
| Total | 180/180 | 180/180 | 180/180 |
| Agreement | 100% | 100% | 100% |
#### b. Limit of Detection
#### l ) Limit of Detection in pooled negative Nasal Wash Matrix
UV-inactivated SARS-CoV-2 virus (Isolate: USA-WA1/ 2020) was spiked into pooled negative Nasal Wash Matrix and serially diluted. Each dilution was tested in triplicate on the CorDx Tyfast COVID-19 Ag Rapid Test to find the initial LoD range. The final LoD was confirmed to be 1.25x104 TCID50/mL by testing 20 individual replicates with concentrations at the initial LoD and at concentrations above and below the initial LoD.
#### 2) Limit of Detection in Pooled Negative Swab Matrix
The Limit of Detection (LoD) of the CorDx Tyfast COVID-19 Ag Rapid Test in the Pooled Negative Swab Matrix (PNSM) was established by testing serial dilutions of heat-inactivated SARS-CoV-2 virus (Isolate: USA-WA1/ 2020) spiked in PNSM. The preliminary LoD determined by testing a 10-fold dilution series of five replicates per concentration was confirmed by testing 20 replicates. The confirmed LoD for the CorDx Tyfast COVID-19 Ag Rapid Test in PNSM was 1.00x104 TCID50/mL.
#### 3) Limit of Detection by testing the WHO International Standard for SARS-CoV-2 antigen (NIBSC code: 21/368)
This study was designed to determine the Limit of Detection (LoD) for the CorDx Tyfast COVID-19 Ag Rapid Test using the WHO International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in Pooled Negative Swab Matrix (PNSM). The preliminary LoD determined by testing a 2-fold dilution series of three replicates per concentration was tested with an additional 17 replicates, for a total of 20 replicates, to confirm the LoD. Concentrations above and below the preliminary LoD were also tested with 20 replicates to further verify the LoD. The LoD for the CorDx Tyfast COVID-19 Ag Rapid Test was 5.00x102 IU/mL.
#### c. Inclusivity
The inclusivity of the CorDx Tyfast COVID-19 Ag Rapid Test was determined for detecting SARS-CoV-2 Alpha, Beta, Delta and Omicron variants as assessed by its Limit of Detection. Serially diluted Heat-irradiated SARS-CoV-2 Alpha (Lineage B1.1.7), Brazil (Lineage P.1), Beta (Lineage B.1.351), Delta (Lineage B.1.617.2), Omicron (Lineage B.1.1.529, XBB), and B.1595 variants were spiked into Pooled Negative Swab Matrix (PNSM) to determine the LoD for each tested variant using one lot of kit.
Based on the results, the CorDx TyFast COVID-19 Ag Rapid Test detects SARS-CoV-2 variants with LoD at the concentrations as indicated in the table below. The assay can detect
{12}------------------------------------------------
#### CorDx, Inc.
these variants near the LoD of the original SARS-CoV-2 virus and thus displays comparable sensitivity and acceptable inclusivity for these variants tested.
| SARS-CoV-2<br>Variant | Concentration<br>(TCID50/ mL) | SARS-CoV-2<br>Variant | Concentration<br>(TCID50/ mL) |
|-----------------------|-------------------------------|-----------------------|-------------------------------|
| Alpha B1.1.7 | $2.5x10^4$ | Omicron B.1.1.529 | $3.125x10^3$ |
| Brazil P.1 | $2.5x10^4$ | Omicron XBB | $2.5x10^4$ |
| Beta B.1.351 | $2.5x10^4$ | B.1.595 | $3.906x10^2$ |
| Delta B.1.617.2 | $3.125x10^3$ | | |
#### Table 2. LoD Concentrations of SARS-CoV-2 Variants
d. Linearity/assay reportable range:
Not applicable. This is a qualitative device.
#### Cross-reactivity and microbial interference study e.
Cross reactivity (analytical specificity) and microbial interference studies were performed to determine if the CorDx Tyfast COVID-19 Ag Rapid Test reacts with non-SARS-CoV-2 respiratory pathogens and other microorganisms that are likely to be encountered in the clinical sample. Each microorganism was evaluated in the absence and presence of inactivated SARS-CoV-2 virus (3x LoD) to see if false positive and false negative test results may occur. Study results showed that no cross reactivity or interference occurs with the following microorganisms at the concentration tested in the table below.
#### Table 3: Microorganisms and Concentrations Tested in the Study
| Microorganisms | Concentration Tested |
|----------------------------------------------------------------------------------|----------------------|
| Human coronavirus 229E (Inactive) | 1x105 TCID50/mL |
| Human coronavirus 229E (Live) | 2.81x104 TCID50/mL |
| Human coronavirus OC43 (Inactive) | 1x105 TCID50/mL |
| Human coronavirus OC43 (Live) | 1x105 TCID50/mL |
| Human coronavirus NL63 (Inactive) | 1x105 TCID50/mL |
| Human coronavirus NL63 (Live) | 1x105 TCID50/mL |
| MERS-coronavirus (Inactive, UV) | 1x105 TCID50/mL |
| MERS-coronavirus (Inactive, Heat) | 1x105 TCID50/mL |
| SARS-coronavirus<br>(Gamma-irradiated virus in Vero E6 cells in DMEM) (Inactive) | 1x105 PFU/mL |
| SARS-coronavirus (Gamma-irradiated virus in PBS) (Inactive) | 1x107 PFU/mL |
| Adenovirus (e.g., C1 Ad. 71) (Live) | 1x106 TCID50/mL |
| Human Metapneumovirus 3 (hMPV-3) Type B1 (Inactive) | 1x105 TCID50/mL |
| Human Metapneumovirus 3 (hMPV-3) Type B1 (Live) | 1x105 TCID50/mL |
| Parainfluenza virus Type 1 (Inactive) | 1x107 TCID50/mL |
| Parainfluenza virus Type 1 (Live) | 1x105 TCID50/mL |
| Parainfluenza virus Type 2 (Inactive) | 1x105 TCID50/mL |
| Parainfluenza virus Type 2 (Live) | 1x105 TCID50/mL |
| Parainfluenza virus Type 3 (Inactive) | 1x107 TCID50/mL |
{13}------------------------------------------------
## CorDx Tyfast COVID-19 Ag Rapid Test
#### CorDx, Inc.
| Microorganisms | Concentration Tested |
|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Parainfluenza virus Type 3 (Live) | 1x106 TCID50/mL |
| Parainfluenza virus Type 4A (Inactive) | 1x105 TCID50/mL |
| Parainfluenza virus Type 4A (Live) | 1x105 TCID50/mL |
| Influenza A /Perth/16/09 (H3N2) (Inactive) | 1x105 TCID50/mL |
| Influenza A/California/07/09 (H1N1) (Inactive) | 1x105 TCID50/mL |
| Influenza A /Hong Kong/2671/19 (Live) | 1x105 TCID50/mL |
| Influenza A/Indiana/02/ 2020 (H1N1) pdm09 (Live) | 1x107 CEID50/mL |
| Influenza B/Brisbane/60 /08 (Victoria lineage) (Inactive) | For Cross reactivity:<br>4.68 x104 TCID50/mL<br>For Interference:<br>2.34 x104 TCID50/mL |
| Influenza B/Wisconsin/01 /10 (Yamagata lineage) (Inactive) | 1x105 TCID50/mL |
| Influenza B/Washington/ 02/19 (Victoria lineage) (Live) | 1 x104 TCID50/mL |
| Influenza B/Florida/4/ 2006 (Yamagata lineage) (Live) | 1x107 CEID50/mL |
| Enterovirus B111 2015 (Inactive) | 1x106 TCID50/mL |
| Enterovirus Type 71 (Live) | 1x106 TCID50/mL |
| Respiratory syncytial virus (Live) | 1x106 TCID50/mL |
| Rhinovirus Type 1A (Inactive) | 1x105 TCID50/mL |
| Rhinovirus Type 1A (Live) | 1x105 TCID50/mL |
| Haemophilus influenzae type b (Eagan) (Live) | 1x107 CFU/mL |
| Streptococcus pneumoniae Z022 (Live) | 1x107 CFU/mL |
| Streptococcus pyogenes Z018 (Live) | 1x107 CFU/mL |
| Candida albicans Z006 (Live) | 1x107 CFU/mL |
| Pooled human nasal wash -<br>representative of normal respiratory microbial flora | NA |
| Bordetella pertussis A639 (Live) | 1x107 CFU/mL |
| Mycoplasma pneumoniae M129 (Live) | 1x107 CCU/mL |
| Chlamydia pneumoniae TW-183 (Live) | 1x107 IFU/mL |
| Legionella pneumophila Philadelphia (Live) | 1x107 CFU/mL |
| Staphylococcus aureus MRSA; COL (Live) | 1x107 CFU/mL |
| Staphylococcus epidermidis MRSE: PR62A (Live) | 1x107 CFU/mL |
#### In-silico Analysis
Three microorganisms including Human coronavirus HKU1, Mycobacterium tuberculosis and Pneumocystis jirovecii were not commercially available for wet experiments. Thus, in silico sequence homology analyses were performed to predict potential cross-reactivity with these microorganisms.
The sequence analysis results showed that Homology exists between the SARS-CoV-2 Nucleocapsid protein and Human Coronavirus HKU1. BLASTP results showed 36 sequence IDs. Sequence ID AGT17773.1 and AGW27840.1 had the highest alignment score (199) and were 39.1% homologous across 76% of the sequences. While this is relatively low sequence
{14}------------------------------------------------
#### CorDx, Inc.
homology, the possibility of cross-reactivity cannot be ruled out. No significant homology was found between the SARS-CoV-2 Nucleocapsid protein and the Mycobacterium tuberculosis or Pneumocystis jirovecii, suggesting that these two organisms would not crossreact or interfere with the CorDx Tyfast COVID-19 Ag Rapid Test even if they were present in the samples.
#### Interference Study-other potential interference f.
This study was performed to evaluate interference between the CorDx Tyfast COVID-19 Ag Rapid Test and endogenous or exogenous substances which may be naturally present in respiratory specimens or that may be artificially introduced.
Negative samples and positive samples (viral titer 3x LoD) were tested in triplicate in the presence of the potentially interfering substances. The performance of the CorDx Tyfast COVID-19 Ag Rapid Test was not affected by any of the interfering substances listed in the table below at the tested concentration.
| Potential Interfering Substance | Concentration Tested |
|------------------------------------------------------------------|----------------------|
| No other substance | NA |
| Whole Blood | 4% |
| Mucin | 0.5% |
| Chloraseptic (Methol/Benzocaine) | 1.5 mg/mL |
| Naso GEL (NeilMed) | 5% v/v |
| CVS Nasal Drops (Phenylephrine) | 15% v/v |
| Afrin (Oxymetazoline) | 15% v/v |
| CVS Nasal Spray (Cromolyn) | 15% v/v |
| Zicam | 5% v/v |
| Homeopathic (Alkalol) | 10% v/v |
| Sore Throat Phenol Spray | 15% v/v |
| Tobramycin | 4 $ \mu $ g/mL |
| Mupirocin | 10 mg/mL |
| Fluticasone Propionate | 5% v/v |
| Tamiflu (Oseltamivir Phosphate) | 5 mg/mL |
| Chloraseptic (Menthol/Benzocaine) | 3 mg/mL |
| Gericare Saline Nasal Spray (Sodium chloride with preservatives) | 15% v/v |
| CVS Health Budesonide Allergy Nasal Spray (Budesonide) | 15% v/v |
| Nasonex 24HR Allergy Nasal Spray (Mometasone) | 15% v/v |
| HealthA2Z Fluticasone Propionate Nasal Spray (Fluticasone) | 15% v/v |
| Luffeel Nasal Spray (Luffa opperculata, Sulfur) | 1.25 % |
| Boiron Galphimia glauca | 15% v/v |
| Boiron Histaminum Hydrochloricum | 15% v/v |
Table 4: Potential Interfering Substances and Concentrations Tested in the Study
{15}------------------------------------------------
#### CorDx Tyfast COVID-19 Ag Rapid Test
#### CorDx, Inc.
#### g. High dose hook effect study:
No hook effect was observed when testing specimens containing SARS-CoV-2 (Isolate: USA-WA1/ 2020) viral concentration as high as 4.57x106 TCIDs0/mL.
#### h. Traceability:
The tests were calibrated against the WHO International Standard for SARS-CoV-2 Antigen.
#### Matrix Comparison i.
Matrix equivalency studies between negative clinical nasal swab matrix and the surrogate pooled nasal cavity wash matrix was conducted and demonstrated equivalent performance of the test with both matrices.
#### Sample Stability: j.
The study results confirmed that the dry swab specimen could be stored at room temperature for up to 2 hours after specimen collection by testing contrived samples prepared with the Pooled Nasal Swab Matrix (PNSM) using three kit lots.
#### k. Reagent Stability:
The real time stability study data demonstrated that the CorDx Tyfast COVID-19 Ag Rapid Test kit can be stored under recommended conditions of 2-30℃ for at least 24 months.
#### l. Shipping Stability:
The stability of the test kits under different stress shipping conditions was evaluated by testing 3 lots of test kits that went through the process of storing respectively at 7 different harsh temperature conditions and the process of dropping from various orientations. The kit components were stable under all conditions tested.
#### 2. Clinical Performance:
#### a. Method comparison with predicate device:
Not applicable. See "3. Clinical Studies" for clinical performance.
#### 3. Clinical Studies:
The clinical performance of the CorDx Tyfast COVID-19 Ag Rapid Test was evaluated in a prospective c…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.