Sofia 2 SARS Antigen+ FIA; Sofia 2 SARS Antigen+ FIA Control Swab Set
Device Facts
| Record ID | K233688 |
|---|---|
| Device Name | Sofia 2 SARS Antigen+ FIA; Sofia 2 SARS Antigen+ FIA Control Swab Set |
| Applicant | Quidel Corporation |
| Product Code | QVF · Microbiology |
| Decision Date | Dec 13, 2023 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 866.3982 |
| Device Class | Class 2 |
Indications for Use
The Sofia 2 SARS Antigen+ FIA is a lateral flow immunofluorescent sandwich assay that is used with the Sofia 2 instrument for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of upper respiratory infection (i.e., symptomatic) when testing is started within 6 days of symptom onset. The test is intended for use as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when either: tested at least twice over three days with at least 48 hours between tests; or when tested once, and negative by the Sofia 2 SARS Antigen+ FIA and followed up with a molecular test. The test does not differentiate between SARS-CoV and SARS-CoV-2. A negative test result is presumptive, and it is recommended these results be confirmed by a molecular SARS-CoV-2 assay. Positive results do not rule out co-infection with bacteria or other viruses and should not be used as the sole basis for treatment or other patient management decisions. Performance characteristics for SARS-CoV-2 were established during the 2021-2022 SARS-CoV-2 pandemic when SARS CoV 2 Omicron was the predominant SARS-CoV-2 variant in circulation. When other SARS-CoV-2 virus variant are emerging, performance characteristics may vary. This test is intended for prescription use only and can be used in Point-of-Care settings.
Device Story
Lateral flow immunofluorescent sandwich assay; detects SARS-CoV-2 nucleocapsid protein in anterior nasal swabs. Sample placed in reagent tube for viral disruption; dispensed into test cassette. Nitrocellulose strip contains immobilized monoclonal antibodies; fluorescent micro-particles coated with monoclonal antibodies bind viral antigen. Sofia 2 instrument automatically scans/images test strip; collects/analyzes fluorescence data; reports positive, negative, or invalid results. No visual interpretation possible. Used in point-of-care settings by healthcare professionals. Provides rapid qualitative results to aid clinical diagnosis; negative results require molecular confirmation. Benefits include rapid identification of SARS-CoV-2 to inform patient management.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and technological similarity to the predicate device.
Technological Characteristics
Lateral flow immunofluorescent sandwich assay; nitrocellulose membrane; fluorescent micro-particles; monoclonal antibodies. Instrument-based detection (Sofia 2 reader). Standalone test cassette. No specific material standards cited.
Indications for Use
Indicated for symptomatic individuals with signs of upper respiratory infection within 6 days of symptom onset for qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in anterior nasal swabs. Used as an aid in COVID-19 diagnosis via serial testing (twice over 3 days, 48h apart) or single test followed by molecular confirmation if negative. Prescription use in point-of-care settings.
Regulatory Classification
Identification
The Sofia 2 SARS Antigen+ FIA is a lateral flow immunofluorescent sandwich assay used with the Sofia 2 instrument for the rapid, qualitative detection of SARS-CoV-2 nucleocapsid protein antigens directly in anterior nasal swab specimens from symptomatic individuals. It is intended as an aid in the diagnosis of SARS-CoV-2 infections (COVID-19) in symptomatic individuals when tested at least twice over three days with at least 48 hours between tests.
Predicate Devices
- Sofia 2 SARS Antigen+ FIA; Sofia 2 SARS Antigen+ FIA Control Swab Set (DEN220039)
Related Devices
- DEN220039 — Sofia 2 SARS Antigen+ FIA, Sofia 2 SARS Antigen+ FIA Control Swab Set · Quidel Corporation · Mar 8, 2023
- K233358 — SCoV-2 Ag Detect Rapid Test · InBios International, Inc. · Aug 23, 2024
- K250273 — BinaxNOW COVID-19 Ag Card · Abbott Diagnostics Scarborough, Inc. · Jun 13, 2025