Synovasure Alpha Defensin Lateral Flow Test Kit (1 Kit), Synovasure Alpha Defensin Lateral Flow Test Kit (5 Test), Synovasure Alpha Defensin Lateral Flow Test Kit (10 Test), Synovasure Alpha Defensin Flow Test Kit (30 Test), Synovasure Alpha Defensin Control Kit

K212204 · Cd Diagnostics, Inc. · QGN · Aug 11, 2021 · Microbiology

Device Facts

Record IDK212204
Device NameSynovasure Alpha Defensin Lateral Flow Test Kit (1 Kit), Synovasure Alpha Defensin Lateral Flow Test Kit (5 Test), Synovasure Alpha Defensin Lateral Flow Test Kit (10 Test), Synovasure Alpha Defensin Flow Test Kit (30 Test), Synovasure Alpha Defensin Control Kit
ApplicantCd Diagnostics, Inc.
Product CodeQGN · Microbiology
Decision DateAug 11, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.3230
Device ClassClass 2

Indications for Use

The Synovasure Alpha Defensin Lateral Flow Test Kit is a qualitative visually read immunochromatographic assay for the detection of human host response proteins, Alpha Defensins 1-3, in the synovial fluid of adults with a total joint replacement who are being evaluated for revision surgery. The Synovasure Alpha Defensin Lateral Flow Test Kit results are intended to be used in conjunction with other clinical and diagnostic findings as an aid in the diagnosis of periprosthetic joint infection (PJ). The Synovasure Alpha Defensin Lateral Flow Test Kit is not intended to identify the etiology or severity of a PJI. The Synovasure Alpha Defensin Control Kit is used in the Synovasure Alpha Defensin Lateral Flow Test Kit as assayed quality control samples to monitor performance and reliability of the Synovasure Alpha Defensin Lateral Flow Test Kit.

Device Story

Qualitative immunochromatographic assay; detects human Alpha Defensins 1-3 in synovial fluid. Input: synovial fluid sample. Principle: antibodies specific to alpha defensin bind host protein; immobilized on lateral flow strip; detected via colloidal gold reporter; produces visually read colored line. Used in clinical settings by healthcare professionals. Output: qualitative result (positive/negative) used alongside other clinical findings to aid PJI diagnosis. Does not identify infection etiology or severity. Includes control kit for performance monitoring.

Clinical Evidence

No clinical data. Bench testing only; method comparison study performed to demonstrate equivalent performance between the modified cassette geometry and the existing version of the device.

Technological Characteristics

Qualitative immunochromatographic lateral flow assay. Uses colloidal gold reporter for visual detection of Alpha Defensins 1-3. Materials and design features are substantially equivalent to the predicate, with minor modifications to cassette geometry. Standalone, non-electronic, visual read test.

Indications for Use

Indicated for use as an aid in the detection and identification of localized human infections by detecting and measuring non-microbial analytes.

Regulatory Classification

Identification

A device to detect and measure non-microbial analytes to aid in the detection and identification of localized human infections is identified as an in vitro diagnostic device intended for the detection and qualitative measurement, quantitative measurement, or both of one or more non-microbial analytes in human clinical specimens to aid in the assessment, identification, or both of a localized microbial infection when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.

Special Controls

The special controls for this device are:

*Classification.* Class II (special controls). The special controls for this device are:(1) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of human specimens; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device. (2) The labeling required under § 809.10(b) of this chapter must include: (i) An intended use with a detailed description of what the device detects and measures, the type of results provided to the user, the sample type, whether the measure is qualitative and/or quantitative, the clinical indications for the test use, and the specific population(s) for which the device is intended. (ii) A detailed description of the performance characteristics of the device for all intended specimen types from the analytical and clinical studies (as applicable) required under paragraphs (b)(3)(ii) and (iii) of this section. (iii) A detailed explanation of the interpretation of results, including acceptance criteria for evaluating the validity of individual runs ( *e.g.,* assessment of internal and/or external quality controls, as applicable).(iv) The following limiting statements: (A) A statement that a negative test result does not preclude the possibility of infection; (B) A statement that the test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician; (C) A statement that consistent device performance is dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and (D) A statement that details any limitations associated with the samples, as appropriate ( *e.g.,* collected on the day of admission to the intensive care unit).(3) Design verification and validation must include the following: (i) A detailed device description, including as appropriate, all device parts; control elements incorporated into the test procedure; instrument requirements; reagents required but not provided; and the principle of device operation and test methodology, including all preanalytical methods for the processing of specimens and the methodology from obtaining a sample to the result; design of primer/probe sequences; rationale for target analyte selection; and computational path from collected raw data to reported result ( *e.g.,* how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies including analytical sensitivity (Limit of Detection, Limit of Quantitation, and Limit of Blank), inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross-contamination, specimen stability, within-lab precision, reproducibility, and linearity, as applicable. (iii) Detailed documentation and results either from a clinical study, that includes prospective (sequentially collected) samples for each intended specimen type that are representative of the intended use populations and, when determined to be acceptable by FDA, additional characterized clinical samples; or, when determined to be acceptable by FDA, an equivalent sample set. The clinical study must compare the device performance to results obtained from an FDA-accepted reference method and/or FDA-accepted comparator method, as appropriate. Documentation from the clinical studies must include the clinical study protocol ( *e.g.,* the predefined statistical analysis plan), clinical study report, testing results, and results of all statistical analyses.(iv) An evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics ( *e.g.,* age, racial, ethnic, and sex distribution) similar to the intended use population of the device.(v) Documentation of an appropriate end user device training program that will be offered as part of efforts to mitigate the risks of false results, failure to operate the device correctly, and failure to interpret test results correctly. (vi) An appropriate risk mitigation strategy to ensure that the device does not prevent any other device(s) with which it is indicated for use, including incorporated device(s), from achieving their intended use ( *e.g.,* safety and effectiveness of the functions of the indicated device(s) remain unaffected).(vii) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K212204 B Applicant CD Diagnostics Inc C Proprietary and Established Names Synovasure Alpha Defensin Lateral Flow Test Kit (1 Kit), Synovasure Alpha Defensin Lateral Flow Test Kit (5 Test), Synovasure Alpha Defensin Lateral Flow Test Kit (10 Test), Synovasure Alpha Defensin Flow Test Kit (30 Test), Synovasure Alpha Defensin Control Kit ## D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QGN | Class II | 21 CFR 866.3230 - Device To Detect And Measure Non-Microbial Analytes To Aid In The Detection And Identification Of Localized Human Infections | MI - Microbiology | ## II Review Summary: This 510(k) submission contains information/data on modifications made to the submitter's own class II device requiring 510(k). The following items are present and acceptable. 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) 2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device **MODIFICATION(S)**, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the **FUNDAMENTAL SCIENTIFIC TECHNOLOGY** of the modified device has not changed. This change was for modification of sample window geometry on the lateral flow strip cassette. 4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. K212204 - Page 2 of 2
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...