The Cepheid® Xpert Flu Assay is an automated, multiplex real-time RT-PCR assay intended for the in vitro qualitative detection and differentiation of influenza A, influenza B and 2009 H1N1 influenza viral RNA. The Xpert Flu Assay uses nasal aspirates/washes and nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. The Xpert Flu Assay is intended as an aid in the diagnosis of influenza. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Performance characteristics for influenza A were established during the 2009-2010 influenza season when 2009 H1N1 influenza was the predominant influenza A virus in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Device Story
Xpert Flu Assay is an automated, multiplex rRT-PCR diagnostic test. Input: nasal aspirates/washes or nasopharyngeal swabs collected in Universal Transport Medium. Process: specimen loaded into single-use disposable cartridge; GeneXpert Instrument System (Dx, Infinity-48, or Infinity-80) performs automated sample purification, nucleic acid amplification, and real-time detection. Cartridge contains all reagents, valve system, and I-CORE thermocycler module. Output: qualitative results for Flu A, Flu B, and 2009 H1N1 in ~75 minutes. Used in clinical laboratory settings; operated by trained personnel. Software interprets fluorescent signals via embedded algorithms to provide tabular/graphic results. Benefits: rapid, automated, closed-system testing reducing contamination risk and providing timely diagnostic aid for respiratory infections.
Clinical Evidence
No clinical studies conducted for this modification. Analytical performance evaluation (reproducibility/precision) compared GeneXpert Dx and Infinity-80 systems using a 10-specimen panel across 12 days with two operators. Results demonstrated comparable performance between systems with high total agreement.
Indicated for patients with signs and symptoms of respiratory infection to detect and differentiate influenza A, influenza B, and 2009 H1N1 viral RNA from nasal aspirates/washes and nasopharyngeal swabs. Aid in diagnosis; not for sole basis of treatment decisions.
Regulatory Classification
Identification
Reagents for detection of specific novel influenza A viruses are devices that are intended for use in a nucleic acid amplification test to directly detect specific virus RNA in human respiratory specimens or viral cultures. Detection of specific virus RNA aids in the diagnosis of influenza caused by specific novel influenza A viruses in patients with clinical risk of infection with these viruses, and also aids in the presumptive laboratory identification of specific novel influenza A viruses to provide epidemiological information on influenza. These reagents include primers, probes, and specific influenza A virus controls.
Special Controls
The device is classified as Class II under regulation 21 CFR 866.3332 with special controls. The special control guidance document "Reagents for Detection of Specific Novel Influenza A viruses" will be available shortly.
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Reagents for Detection of Specific Novel Influenza A Viruses.” See § 866.1(e) for information on obtaining this document.
(2) The distribution of these devices is limited to laboratories with experienced personnel who have training in standardized molecular testing procedures and expertise in viral diagnosis, and appropriate biosafety equipment and containment.
{0}------------------------------------------------
## 510(k) Summary
As required by 21 CFR Section 807.92(c).
MAY 1 8 2012
Submitted by:
Cepheid 904 Caribbean Drive Sunnyvale, CA 90489 Phone number: (408) 400-8230 Fax number: (408) 541-6439
Russel K. Enns, Ph.D.
Contact:
Date of Preparation:
Trade name:
Device:
Xpert® Flu
Xpert Flu Assay
May 7, 2012
Common name :: Type of Test:
Classification:
Classification name:
Regulation number:
Automated, multiplex real-time reverse transcriptionpolymerase chain reaction (rRT-PCR) assay intended for the in vitro qualitative detection and differentiation of influenza A, influenza B and 2009 H1N1 influenza viral RNA.
Class II
Respiratory viral panel multiplex nucleic acid assay
866.3980/Respiratory viral panel multiplex nucleic acid assay and 866.3332/Reagents for detection of specific novel influenza A viruses
| Product code: | OQW, OCC, OOI |
|-------------------|-----------------------------------|
| Panel: | Microbiology (83) |
| Predicate Device: | Xpert® Flu Assay, 510(k) #K103766 |
### Device Description:
The Cepheid Xpert® Flu Assay is a rapid, automated in vitro diagnostic test for qualitative detection and differentiation of influenza B, and influenza A, subtype 2009 H1N1 from nasal aspirates/washes (NA/W) and nasopharyngeal (NP) swab specimens from patients with signs and symptoms of respiratory infection. The assay is performed on the Cepheid GeneXpert Instrument Systems, which consist of the GeneXpert Dx System and the GeneXpert Infinity-48 System. The Device is being modified with this 510(k), to add the GeneXpert Infinity-80 System as an additional instrument system for use with the Xpert Flu Assay.
{1}------------------------------------------------
As a result of this modification, the Intended Use of the modified Xpert Flu Assay has not changed. Further, the fundamental scientific technology has not changed as a result of the modification to use the GeneXpert Infinity-80 System with the Xpert Flu Assay.
The GeneXpert Instrument Systems automate and integrate sample purification, nucleic acid amplification, and detection of the target sequence in simple or complex samples using real-time PCR and rRT-PCR assays. The GeneXpert Instrument System family comprises a GeneXpert (GX) instrument, GX-I, GX-IV, GX-XVI; a GeneXpert XVI, available with 4, 8, 12 or 16 modules, a GeneXpert Infinity-48 available with 16, 24, 32 or 48 modules, or a GeneXpert Infinity-80 available with 16, 24, 32, 40, 48, 56, 64, 72, or 80 modules. The instrument systems also contain a computer, and preloaded software for running tests and viewing the results. The GeneXpert Infinity Systems contain robotic features for cartridge handling. Each module contains a syringe drive for dispensing fluids, an ultrasonic horn for lysing cells or spores, a valve drive for sample movement, and I-CORE® thermocycler for performing real-time PCR and detection.
All systems require the use of the assay-specific single-use disposable cartridges that hold the PCR reagents and host the PCR process. The patented single-use cartridges contain: (1) eleven chambers for holding sample, reagents, or other materials, (2) a valve body composed of a plunger and syringe barrel, (3) a rotary valve system for controlling the movement of fluids between chambers, (4) an area for capturing, concentrating, washing, and lysing cells, (5) dry real-time PCR reagents, (6) an integrated PCR reaction tube that can be automatically filled by the instrument, and (7) liquid reagents. To eliminate testto-test contamination, all fluids including amplicons, are contained within the disposable cartridge. The instrument never comes into contact with any fluids within the cartridge. Each disposable cartridge is intended to test one sample. Cartridges are not re-usable.
A sample processing control (SPC) and a system control (Probe Check Control) are controls utilized by the GeneXpert Instrument System platform. The SPC is pre-loaded into the GeneXpert cartridge provided with the assay. The SPC is an encapsidated RNA made up of recombinant fragments developed so that there is no homology to the influenza genome. The SPC is present to control for adequate processing of the target viruses and to monitor the presence of inhibitors in the PCR reaction to reduce the possibility of false negative results. The SPC also ensures the PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the PCR reagents are functional. The Probe Check Control verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity and dye stability.
Commercially-available external controls may also be run in accordance with local, state, and federal accrediting organizations, as applicable.
The Xpert Flu Assay includes reagents for the simultaneous detection and differentiation of the target viruses. The primers and probes in the Xpert Flu Assay detect the presence of nucleic acid sequences for influenza A (Flu A), influenza B (Flu B) and influenza A sub-type 2009 H1N1 (2009 H1N1) directly from nasal aspirates/washes (NA/W) and nasopharyngeal (NP) swab specimens collected from patients suspected of having
{2}------------------------------------------------
influenza. The specimens are collected in Universal Transport Medium (UTM) and transported to the GeneXpert area.
The specimen is prepared according to package insert instructions and transferred to the sample chamber (large opening) of the Xpert Flu Assay Cartridge. Reagent 1 (Binding Reagent) is dispensed into the chamber with the small opening of the Xpert Flu Assay Cartridge. The GeneXpert Cartridge is loaded onto the GeneXpert® Instrument System platform, which performs hands-off automated sample processing and real-time PCR for detection of Flu RNA. Summary and detailed test results are obtained in 75 minutes.
The results are interpolated by the GeneXpert Instrument Systems software from measured fluorescent signals and embedded calculation algorithms and are shown in the "View Results" window in tabular and graphic formats. The Xpert Flu Assay provides test results for influenza A, influenza B and influenza A, subtype 2009 H1N1. It also reports if the test is "Invalid," "Error" or "No Result," and instructs the user to repeat the test.
## Device Intended Use:
The Cepheid Xpert® Flu Assay is an automated, multiplex real-time RT-PCR assay intended for the in vitro qualitative detection and differentiation of influenza A, influenza B and 2009 H1N1 influenza viral RNA. The Xpert Flu Assay uses nasal aspirates/washes and nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. The Xpert Flu Assay is intended as an aid in the diagnosis of influenza.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.
Performance characteristics for influenza A were established during the 2009-2010 influenza season when 2009 H1N1 influenza was the predominant influenza A virus in circulation. When other influenza A viruses are emerging, performance characteristics may vary.
If infection with a novel influenza A virus is suspected based on current clinical and enidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
### Substantial Equivalence:
The Xpert Flu Assay with the GeneXpert Infinity-80 Instrument System option is substantially equivalent to the predicate device, the Xpert Flu Assay; 510(k) #K103766, which was cleared on April 21, 2011. The two devices have the same material composition and technological characteristics. The 510(k) supports the addition of the option of performing the assay on the GeneXpert Infinity-80 System. The Intended Use
{3}------------------------------------------------
of the Xpert Flu Assay remains the same. There are no changes to the fundamental scientific technology of the device as a result of the modification.
## Analytical Performance Evaluation:
A reproducibility/precision study was conducted using the Xpert Flu Assay to compare the performance of the GeneXpert Dx and the GeneXpert Infinity-80 Instrument Systems. A panel of 10 specimens with varying concentrations of influenza A, influenza B, and influenza A subtype 2009 H1N1 were tested on 12 different days by two operators. Each operator conducted four runs of each panel specimen per day on each of the two instrument systems (10 specimens x 4 times/day x 12 days x 2 operators x 2 instrument systems). One lot of Xpert Flu Assay was used for the study. Xpert Flu Assays were performed according to the Xpert Flu Assay procedure. The two Instrument Systems were shown to provide comparable results. Results are summarized in Table 7.1.
| Sample ID | GeneXpert Dx | Infinity - 80 | % Total Agreement<br>by Sample |
|--------------------------------|--------------------|--------------------|--------------------------------|
| Negative | 100.0%<br>(96/96) | 100.0%<br>(96/96) | 100.0%<br>(192/192) |
| Flu A<br>moderate positive | 100.0%<br>(96/96) | 100.0%<br>(96/96) | 100.0%<br>(192/192) |
| Flu A<br>low positive | 97.9%<br>(94/96) | 99.0%<br>(95/96) | 98.4%<br>(189/192) |
| Flu A<br>high negative | 93.8%<br>(90/96) | 87.5%<br>(84/96) | 90.6%<br>(174/192) |
| 2009 H1N1<br>moderate positive | 100.0%<br>(96/96) | 100.0%<br>(96/96) | 100.0%<br>(192/192) |
| 2009 H1N1<br>low positive | 97.9%<br>(94/96) | 99.0%<br>(95/96) | 98.4%<br>(189/192) |
| 2009 H1N1<br>high negative | 54.2%<br>(52/96) | 38.7%<br>(36/93)a | 46.6%<br>(88/189) |
| Flu B<br>moderate positive | 97.9%<br>(94/96) | 100.0%<br>(96/96) | 99.0%<br>(190/192) |
| Flu B<br>low positive | 81.3%<br>(78/96) | 88.5%<br>(85/96) | 84.9%<br>(163/192) |
| Flu B<br>high negative | 85.4%<br>(82/96) | 83.3%<br>(80/96) | 84.4%<br>(162/192) |
| % Total Agreement | 90.8%<br>(872/960) | 89.8%<br>(859/957) | 90.3%<br>(1731/1917) |
Table 7.1: Summary of Instrument System Reproducibility Results
2 n=93 because 3/96 Samples vielded indeterminate results on both the first and second attempt.
{4}------------------------------------------------
# Clinical Performance Evaluation:
There were no clinical studies conducted as a result of the modification to the predicate device.
# Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Food and Drug Administration
10903 New Hampshire Avenue Silver Spring, MD 20993
Cepheid c/o Kerry J. Flom, Ph.D. Senior Vice President 904 Caribbean Drive Sunnyvale, CA 94089-1189
MAY 1 8 2012
Re: k120911
Trade/Device Name: Xpert® Flu Assay Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OQW, OCC, OOI Dated: March 23, 2012 Received: March 26, 2012
Dear Dr. Flom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean i hat FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I odolar statutes and sequirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice
{6}------------------------------------------------
## Page 2 - Kerry J. Flom
requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sally Hojvat, M.Sc., Ph.D Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Statement of Indications for Use 6.0
Indications for Use Form
510(k) Number (if known): الاك 1009)
Device Name: Xpert® Flu
Indications for Use:
The Cepheid Xpert® Flu Assay is an automated, multiplex real-time RT-PCR assay intended for the in vitro qualitative detection and differentiation of influenza B and 2009 HIN1 influenza viral RNA. The Xpert Flu Assay uses nasal aspirates/washes and nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. The Xpert Flu Assay is intended as an aid in the diagnosis of influenza.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.
Performance characteristics for influenza A were established during the 2009-2010 influenza season when 2009 HINI influenza was the predominant influenza A virus in circulation. When other influenza A viruses are emerging, performance characteristics may vary.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR .
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety
Xpert Flu Assay 510(k) March 2012
{8}------------------------------------------------
Taman Tr Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
র প্রশ্নীতি পার্টির প্রতিষ্ঠান করে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে করে পারে করে পারে করে পারে করে পারে করে করে করে করে করে করে করে 510(k)
:
·
・
·
・
.
・
.
・
: :
-
.
.
:
·
: ·
:
Xpert Flu Assay 510(k) March 2012
:
·
··
. :
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.