CDC HUMAN INFLUENZA VIRUS REAL-TIME RT-PCR DIAGNOSTIC PANEL
K111507 · Centers for Disease Control and Prevention · OQW · Aug 23, 2011 · Microbiology
Device Facts
Record ID
K111507
Device Name
CDC HUMAN INFLUENZA VIRUS REAL-TIME RT-PCR DIAGNOSTIC PANEL
Applicant
Centers for Disease Control and Prevention
Product Code
OQW · Microbiology
Decision Date
Aug 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3332
Device Class
Class 2
Intended Use
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR Instrument in conjunction with clinical and epidemiological information: - For qualitative detection of influenza virus type A or B viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture. - For determination of the subtype of seasonal human influenza A virus as seasonal A/H1, A/H3, and/or A/H1pdm09 from viral RNA in upper respiratory tract clinical specimens (including NPS, NS, TS, NA, NW and NPS/TS) and lower respiratory tract specimens (including BAL, BW, TA, sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture. - For the presumptive identification of virus in patients who may be infected with influenza A subtype A/H5 (Asian Lineage) from viral RNA in human respiratory specimens and viral culture in conjunction with clinical and epidemiological risk factors. - To provide epidemiologic information for surveillance of circulating influenza viruses.
Device Story
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel is an in vitro diagnostic kit for qualitative detection and subtyping of influenza viruses. Input: viral RNA extracted from human respiratory specimens (swabs, aspirates, washes, BAL, sputum, tissue) or viral culture. The device uses oligonucleotide primers and dual-labeled TaqMan hydrolysis probes in an rRT-PCR assay performed on the ABI 7500 Fast Dx Real-Time PCR instrument. The assay targets conserved regions of the matrix (M), nucleoprotein (NP), non-structural (NS), and hemagglutinin (HA) genes. Output: fluorescence growth curves indicating presence/absence of specific influenza targets (InfA, InfB, H1, H3, H5a, H5b, pdmInfA, pdmH1) and RNase P (extraction control). Used in clinical/public health laboratories by trained personnel. Results assist clinicians in patient management and provide epidemiological surveillance data. Positive results for novel/emerging strains (e.g., H5) require immediate referral to CDC for confirmation.
Clinical Evidence
Clinical performance was established via prospective studies during the 2006-2007 and 2009-2010 respiratory seasons. Studies compared the panel against gold standard viral culture (shell vial) followed by IFA/DFA and/or bi-directional sequencing. Sensitivity and specificity were calculated for various analytes (InfA, InfB, H1, H3) across multiple specimen types (TS, NA, NPS/TS). Sensitivity for InfA ranged from 100% (TS, NA, NPS/TS) to 100% (isolates); specificity ranged from 56.1% to 100%. Data aggregated from K080570 and K101564.
Technological Characteristics
Real-time RT-PCR assay using oligonucleotide primers and dual-labeled TaqMan hydrolysis probes. Targets conserved regions of M, NP, NS, and HA genes (100-200 nucleotides). Requires ABI 7500 Fast Dx Real-Time PCR instrument. Includes lyophilized primer/probe sets and liquid positive controls (PIPC, H5VC, HSC). Compatible with various automated RNA extraction kits (Roche MagNA Pure, Qiagen QIAamp/QIAcube, bioMérieux NucliSENS easyMAG).
Indications for Use
Indicated for qualitative detection and differentiation of influenza A and B virus RNA in respiratory specimens (upper/lower tract) and viral culture from patients with signs/symptoms of respiratory infection. Used for subtyping seasonal A/H1, A/H3, A/H1pdm09, and presumptive identification of A/H5 (Asian lineage). Prescription use only.
Regulatory Classification
Identification
Reagents for detection of specific novel influenza A viruses are devices that are intended for use in a nucleic acid amplification test to directly detect specific virus RNA in human respiratory specimens or viral cultures. Detection of specific virus RNA aids in the diagnosis of influenza caused by specific novel influenza A viruses in patients with clinical risk of infection with these viruses, and also aids in the presumptive laboratory identification of specific novel influenza A viruses to provide epidemiological information on influenza. These reagents include primers, probes, and specific influenza A virus controls.
Special Controls
The device is classified as Class II under regulation 21 CFR 866.3332 with special controls. The special control guidance document "Reagents for Detection of Specific Novel Influenza A viruses" will be available shortly.
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Reagents for Detection of Specific Novel Influenza A Viruses.” See § 866.1(e) for information on obtaining this document.
(2) The distribution of these devices is limited to laboratories with experienced personnel who have training in standardized molecular testing procedures and expertise in viral diagnosis, and appropriate biosafety equipment and containment.
Predicate Devices
CDC Human Influenza Virus Real-time RT-PCR Detection and Characterization Panel (K080570)
CDC Influenza 2009 A (H1N1)pdm Real-time RT-PCR Panel (K101564)
Related Devices
K130551 — CDC HUMAN INFLUENZA VIRUS REAL-TIME RT-PCR DIAGNOSTIC PANEL · Centers for Disease Control and Prevention · May 22, 2013
K123905 — CDC HUMAN INFLUENZA VIRUS REAL-TIME RT-PCR DIAGNOSTIC PANEL · Centers for Disease Control and Prevention · Jan 14, 2013
K133869 — CDC HUMAN INFLUENZA VIRUS REAL-TIME RT-PCR DIAGNOSTIC PANEL, INFLUENZA A/B TYPING KIT · Center For Disease Control and Prevention (Cdc) · Jan 17, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
#### 6. 510(k) Summary
#### Assigned 510(k) number: K111507
#### Submitted by:
Centers for Disease Control and Prevention 1600 Clifton Road NE Atlanta, GA 30333
#### Contact Person:
Hye-Joo Kim, Pharm.D. Associate Director for Regulatory Affairs Office of the Director National Center for Emerging and Zoonotic Infectious Diseases Centers for Disease Control and Prevention 1600 Clifton Road, NE, MS C-12 Atlanta, GA 30333 (404) 639-4643 (office) (404) 639-1275 (fax) hek6@cdc.gov
Date prepared: August 19, 2011
Device Name: CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel
## Common or Usual Name: CDC Flu rRT-PCR Dx Panel
## Regulatory Information:
Regulation Section: Reagents for detection of specific novel influenza A viruses (21 CFR 866.3332) Classification: Class II Product Codes: OQW, NXD, OEP, NSU Panel: Microbiology
#### Predicate Devices:
- CDC Human Influenza Virus Real-time RT-PCR Detection and Characterization Panel (K080570)
- CDC Influenza 2009 A (H1N1)pdm Real-time RT-PCR Panel (K101564)
{1}------------------------------------------------
## Device Description
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel is used in real-time RT-PCR assays (rRT-PCR) on the ABI 7500 Fast Dx Real-Time PCR Instrument.
The CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel is a panel of oligonucleotide primers and dual-labeled hydrolysis (TaqMan®) probes to be used in rRT-PCR for the in vitro qualitative detection and characterization of human influenza viruses from viral RNA in respiratory specimens from patients presenting with influenza-like illness (ILI). Detection of viral RNA not only aids in the diagnosis of illness caused by seasonal and novel influenza viruses in patients with ILI, but also provides epidemiological information on influenza and aids in the presumptive laboratory identification of specific novel influenza A viruses.
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel is based on technology which is used in many molecular diagnostic assays. rRT-PCR assays are one-tube assays that first reverse-transcribe specific regions of RNA into cDNA copies. The cDNA then serves as a template for a polymerase chain reaction that utilizes a thermocyclic heating and cooling of the reaction to logarithmically amplify a specific region of DNA. The probe anneals to a specific internal target sequence located between the target loci of the forward and reverse primers. During the extension phase of the PCR cycle, the 5' nuclease activity of Tag polymerase degrades any probe molecules hybridized to amplified target sequence, causing the reporter dye to separate from the quencher dye, and generating a fluorescent signal. With each cycle, additional reporter dye molecules are cleaved from their respective probes, increasing the fluorescence intensity. Fluorescence intensity is monitored at each PCR cycle. Amplification of targets is reflected by logarithmic increase in fluorescence over time in comparison to background signal.
## Intended Use
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR Instrument in conjunction with clinical and epidemiological information:
- For qualitative detection of influenza virus type A or B viral RNA in upper respiratory . tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture.
- For determination of the subtype of seasonal human influenza A virus as seasonal . A/H1. A/H3. and/or A/H1pdm09 from viral RNA in upper respiratory tract clinical
{2}------------------------------------------------
specimens (including NPS, NS, NA, NW and NPS/TS) and lower respiratory tract specimens (including BAL, BW, TA, sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture.
- For the presumptive identification of virus in patients who may be infected with . influenza A subtype A/H5 (Asian Lineage) from viral RNA in human respiratory specimens and viral culture in coniunction with clinical and epidemiological risk factors.
- To provide epidemiologic information for surveillance of circulating influenza viruses. .
Performance characteristics for influenza were established during a season when influenza viruses A/H1 and A/H3 were the predominant influenza A viruses in circulation and during a season when the A/H1pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses.
Testing with the influenza H5a and H5b primer and probe sets should not be performed unless the patient meets the most current U.S. Department of Health and Human Services (DHHS) clinical and epidemiologic criteria for testing suspect A/H5 specimens. The definitive identification of influenza A/H5 (Asian lineage) either directly from patient specimens or from virus cultures requires additional laboratory testing, along with clinical and epidemiological assessment in consultation with national influenza surveillance experts.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facilable to receive and culture specimens.
All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
{3}------------------------------------------------
#### Substantial Equivalence Comparison
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel merges two previously FDA-cleared CDC devices and is equivalent to the CDC Human Influenza Virus Real-time RT-PCR Detection and Characterization Panel (K080570) and the CDC Influenza 2009 A (H1N1)pdm Real-time RT-PCR Panel (K101564). These devices were developed to the same specifications and utilize the same real-time RT-PCR technology to detect influenza A and influenza B in human respiratory specimens. The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel further detects and differentiates influenza A and B viruses and characterizes influenza A viruses as A/H1, A/H3 subtypes, A/H1pdm09, and A/H5 using real-time RT-PCR. All three devices utilize the same instrumentation, the Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument. which has been granted marketing clearance by FDA (K082562). The CDC Human Influenza Virus Real-time RT-PCR Detection and Characterization Panel (K080570) and the CDC Influenza 2009 A (H1N1)pdm Real-time RT-PCR Panel (K101564) were granted marketing clearance by the FDA on September 30, 2008, and June 22, 2010, respectively.
| | CDC Human Influenza Virus<br>Real-time PCR Diagnostic<br>Panel | CDC Human Influenza<br>Virus Real-time RT-PCR<br>Detection and<br>Characterization Panel<br>(K080570) | CDC Influenza 2009 A<br>(H1N1)pdm Real-time RT-<br>PCR Panel (K101564) |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The CDC Human Influenza Virus<br>Real-Time PCR Diagnostic Panel<br>is intended for use in Real-time<br>RT-PCR assays on an Applied<br>Biosystems (ABI) 7500 Fast Dx<br>Real-Time PCR Instrument in<br>conjunction with clinical and<br>epidemiological information: 1)<br>for qualitative detection of<br>influenza virus type A or B from<br>viral RNA in upper respiratory<br>tract clinical specimens (including<br>nasopharyngeal swabs, nasal<br>swabs, throat swabs nasal<br>aspirates, nasal washes and dual<br>nasopharyngeal/throat swabs), and<br>lower respiratory tract specimens<br>(including bronchoalveolar<br>lavages, bronchial washes,<br>tracheal aspirates, sputum, and<br>lung tissue) from human patients<br>with signs and symptoms of<br>respiratory infection and/or from<br>viral culture, 2) for determination<br>of the subtype of seasonal human<br>influenza A virus as seasonal<br>A/H1, A/H3, and/or A/H1pdm09<br>from viral RNA in upper<br>respiratory tract clinical<br>specimens (including<br>nasopharyngeal swabs, nasal<br>swabs, throat swabs nasal | The Human Influenza Virus<br>Real-time RT-PCR<br>Detection and<br>Characterization Panel is<br>intended for use in Real-<br>time RT-PCR assays on an<br>ABI 7500 Fast Dx Real-<br>Time PCR Instrument in<br>conjunction with clinical<br>and epidemiological<br>information: for qualitative<br>detection of influenza virus<br>type A or B in symptomatic<br>patients from viral RNA in<br>nasopharyngeal and/or nasal<br>swab specimens, for<br>determination of the<br>subtype of seasonal human<br>influenza A virus, as<br>seasonal A/HI or A/H3, if<br>present, from viral RNA in<br>nasopharyngeal and/or nasal<br>swab specimens, for<br>presumptive identification<br>of virus in patients who may<br>be infected with influenza A<br>subtype A/H5 (Asian<br>lineage) from viral RNA in<br>human respiratory<br>specimens and viral culture<br>in conjunction with clinical<br>and epidemiological risk | The CDC rRT-PCR<br>A(H1N1)pdm09 Flu Panel is<br>intended for use in real-time<br>RT-PCR assays on the<br>Applied Biosystems (ABI)<br>7500 Fast Dx Real-Time<br>PCR Instrument for the in<br>vitro qualitative detection of<br>influenza virus type A and<br>2009 A/H1N1 viral RNA<br>from nasopharyngeal swabs,<br>nasal swabs, throat swabs,<br>nasal aspirates, nasal<br>washes, dual<br>nasopharyngeal / throat<br>swabs and lower respiratory<br>tract specimens from human<br>patients with signs and<br>symptoms of respiratory<br>infection and/or from viral<br>culture, in conjunction with<br>clinical and epidemiological<br>risk factors. |
| | | | |
| | aspirates, nasal washes and dual<br>nasopharyngeal/throat swabs), and<br>lower respiratory tract specimens<br>(including bronchoalveolar<br>lavages, bronchial washes,<br>tracheal aspirates, sputum, and<br>lung tissue) from human patients<br>with signs and symptoms of<br>respiratory infection and/or from<br>viral culture, 3) for the<br>presumptive identification of virus<br>in patients who may be infected<br>with influenza A subtype A/H5<br>(Asian Lineage) from viral RNA<br>in human respiratory specimens<br>and viral culture in conjunction<br>with clinical and epidemiological<br>risk factors, and 4) to provide<br>epidemiologic information for<br>surveillance of the circulating<br>influenza viruses. | factors to provide<br>epidemiologic information<br>for surveillance for<br>influenza viruses. | |
| Specimen Types | Nasopharyngeal swabs, nasal<br>swabs, throat swabs, nasal<br>aspirates, nasal washes and dual<br>nasopharyngeal/throat swabs,<br>bronchoalveolar lavages,<br>bronchial washes, tracheal<br>aspirates, sputum, and lung tissue<br>and virus culture | Nasopharyngeal swabs,<br>nasal swabs, and virus<br>culture | Nasopharyngeal swabs,<br>nasal swabs, nasal aspirates,<br>nasal washes, dual collected<br>nasopharyngeal and throat<br>swabs, bronchoalveolar<br>lavages, bronchial washes,<br>and tracheal aspirates, and<br>virus culture. |
| Technology | Real-time RT-PCR | Real-time RT-PCR | Real-time RT-PCR |
| Required<br>Instrumentation | Applied Biosystems 7500 Fast<br>Dx Real-Time PCR Instrument | Applied Biosystems 7500<br>Fast Dx Real- Time PCR<br>Instrument | Applied Biosystems 7500<br>Fast Dx Real- Time PCR<br>Instrument |
| Organism<br>Detected | Universal influenza A viruses,<br>Swine-origin influenza A<br>viruses, Influenza B viruses, and<br>Influenza A subtypes: seasonal<br>A/H1, A/H3, A/H1pdm09, and<br>A/H5 (Asian Lineage) | Universal influenza A virus,<br>subtypes A/H1 and A/H3;<br>Influenza B virus;<br>Influenza A virus, subtype<br>A/H5 (Asian lineage) | Universal influenza A,<br>Swine-Origin Influenza A,<br>and A/H1pdm09 subtype |
| Nucleic Acid<br>Extraction | Yes | Yes | Yes |
| Extraction<br>Method | • QIAamp® DSP Viral RNA<br>Mini Kit, Qiagen Inc.<br>• MagNA Pure Compact -Total<br>Nucleic Acid Kit, Roche<br>Applied Science<br>• MagNA Pure Compact - RNA<br>Isolation Kit, Roche Applied<br>Science<br>• MagNA Pure LC - RNA<br>Isolation Kit II, Roche Applied<br>Science<br>• Qiagen QIAcube with<br>QIAamp® DSP Viral RNA<br>Mini Kit, Qiagen Inc.<br>• NucliSENS® easyMAG®,<br>bioMerieux | • QIAamp® Viral RNA<br>Mini Kit, Qiagen Inc.<br>• RNeasy® Mini Kit,<br>Qiagen, Inc.<br>• MagNA Pure LC RNA<br>Isolation Kit II, Roche<br>Applied Science<br>• MagNA Pure Total<br>Nucleic Acid Kit, Roche<br>Applied Science | • QIAamp® Viral RNA<br>Mini Kit, Qiagen Inc.<br>• MagNA Pure Compact -<br>Total Nucleic Acid Kit,<br>Roche Applied Science<br>• MagNA Pure Compact -<br>RNA Isolation Kit, Roche<br>Applied Science<br>• MagNA Pure LC - RNA<br>Isolation Kit II, Roche<br>Applied Science<br>• Qiagen QIAcube with<br>QIAamp® Viral RNA<br>Mini Kit, Qiagen Inc.<br>• NucliSENS® easyMAG®,<br>bioMerieux |
| Enzyme Master<br>Mix | Invitrogen SuperScript™ III<br>Platinum® One-Step | Invitrogen SuperScript™ III<br>Platinum® One-Step | Invitrogen SuperScript™ III<br>Platinum® One-Step |
| | | | |
{4}------------------------------------------------
100 - 100 -
and the comments of the comments of the country
,
{5}------------------------------------------------
| Quantitative RT-PCR Kits (with<br>or without ROX) | Quantitative RT-PCR Kits<br>(with or without ROX) | Quantitative RT-PCR Kits<br>(with or without ROX) |
|---------------------------------------------------|---------------------------------------------------|---------------------------------------------------|
|---------------------------------------------------|---------------------------------------------------|---------------------------------------------------|
## Performance Characteristics
For analytical and clinical performance characteristics, please refer to previously FDAcleared CDC 510(k) Premarket Notifications:
- 1. K080570 Cleared on September 30, 2008: CDC Human Influenza Virus Real-Time RT-PCR Detection and Characterization Panel
- 2. · K 101564 Cleared on June 22, 2010: CDC Influenza 2009 A (H N N )pdm Real-Time RT-PCR Panel
Additional performance data were collected during the 2010-2011 influenza season. Recent circulating seasonal influenza virus strains can be detected by the CDC Human Influenza Virus Real-Time RT-PCR Detection and Characterization Panel and the CDC Influenza 2009 A (H1N1)pdm Real-Time RT-PCR Panel as demonstrated by analytical testing of the 49 original specimens received from US public health laboratories that contained 24 Influenza A/H3 (49%) and 25 Influenza B (51%).
Eighteen original lower respiratory specimens were received from US public health laboratories during the 2010-2011 influenza season from hospitalized patients or fatal cases. The CDC Human Influenza Virus Real-Time RT-PCR Detection and Characterization Panel and the CDC Influenza 2009 A (H1N1)pdm Real-Time RT-PCR Panel detected influenza A/H3, A/H1pdm09, and influenza B in bronchoalveolar lavage, bronchial washes, tracheal aspirates, sputum, and lung tissue specimens, verifying detection of recent circulating influenza viruses in the lower respiratory tract.
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird with outstretched wings.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
National Center for Emerging and Zoonotic Infectious Diseases Center for Disease Control and Prevention c/o CAPT, Hve-Joo Kim, Pharm.D. Associate Director for Regulatory Affairs Office of the Director 1600 Clifton Road, N.E. MS-C12 Atlanta, GA 30333
# AUG 23 2011
| Re: | K111507 |
|--------------------|----------------------------------------------------------------|
| Trade/Device Name: | CDC Human Influenza Virus Real-Time<br>RT-PCR Diagnostic Panel |
| Regulation Number: | 21CFR §866.3332 |
| Regulation Name: | Reagents for detection of specific novel influenza A viruses. |
| Regulatory Class: | Class II |
| Product Code: | OQW, NXD, OEP, NSU |
| Dated: | May 31, 2011 |
| Received: | June 1, 2011 |
Dear Ms. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the
{7}------------------------------------------------
Page 2 - CAPT. Hye-Joo Kim
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Taqdtyra
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use Statement
### 510(k) Number (if known): K111507
## Device Name: CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR Instrument in conjunction with clinical and epidemiological information:
- For qualitative detection of influenza virus type A or B viral RNA in upper respiratory t tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture.
- For determination of the subtype of seasonal human influenza A virus as seasonal . A/H1, A/H3, and/or A/H1pdm09 from viral RNA in upper respiratory tract clinical specimens (including NPS, NS, TS, NA, NW and NPS/TS) and lower respiratory tract specimens (including BAL, BW, TA, sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture.
- For the presumptive identification of virus in patients who may be infected with . influenza A subtype A/H5 (Asian Lineage) from viral RNA in human respiratory specimens and viral culture in conjunction with clinical and epidemiological risk factors.
- To provide epidemiologic information for surveillance of circulating influenza viruses. .
Performance characteristics for influenza were established during a season when influenza viruses A/H1 and A/H3 were the predominant influenza A viruses in circulation and during a season when the A/H1pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses.
Testing with the influenza H5a and H5b primer and probe sets should not be performed unless the patient meets the most current U.S. Department of Health and Human Services (DHHS) clinical and epidemiologic criteria for testing suspect A/H5 specimens. The definitive identification of influenza A/H5 (Asian lineage) either directly from patient specimens or from virus cultures requires additional laboratory testing, along with clinical and epidemiological assessment in consultation with national influenza surveillance experts.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
{9}------------------------------------------------
If infection with a novel influenza A virus is suspected based on current clinical and evidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k)
Tawarah Feldboly
Division Sign-Off
**Office of In Vitro Diagnostic Device Evaluation and Safety**
510(k) K111507
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.