Simplexa Flu A/B & RSV Direct Gen II, Simplexa Flu A/B & RSV Positive Control Pack

K201505 · Diasorin Molecular, LLC · OCC · Aug 22, 2020 · Microbiology

Device Facts

Record IDK201505
Device NameSimplexa Flu A/B & RSV Direct Gen II, Simplexa Flu A/B & RSV Positive Control Pack
ApplicantDiasorin Molecular, LLC
Product CodeOCC · Microbiology
Decision DateAug 22, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3980
Device ClassClass 2
AttributesPediatric

Intended Use

The DiaSorin Molecular Simplexa™ Flu A/B & RSV Direct Gen II assay is intended for use on the LIAISON® MDX instrument for the in vitro qualitative detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) from human patients with signs and symptoms of respiratory tract infection in conjunction with clinical and epidemiological risk factors. This test is intended for use as an aid in the differential diagnosis of influenza A, influenza B, and RSV viral infections in humans. Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for treatment or other patient management decisions. Performance characteristics for influenza A were established with clinical specimens collected during the 2010/2011 influenza season when 2009 H1N1 influenza and H3N2 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to the state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. DiaSorin Molecular’s Simplexa™ Flu A/B & RSV Positive Control Pack is intended to be used as a control with the Simplexa™ Flu A/B & RSV Direct kit and the Simplexa™ Flu A/B & RSV Direct kit and the Simplexa™ Flu A/B & RSV Direct Gen II kit for use on the LIAISON® MDX instrument. This control is not intended for use with other assays or systems.

Device Story

The Simplexa Flu A/B & RSV Direct Gen II is an in vitro diagnostic assay for the qualitative detection and differentiation of influenza A, influenza B, and RSV RNA. The device utilizes nasopharyngeal swab samples processed on the LIAISON MDX instrument. It employs nucleic acid amplification technology to detect viral RNA. The system provides qualitative results to clinicians to aid in the differential diagnosis of respiratory viral infections. It is intended for use in clinical laboratory settings by trained personnel. The output assists healthcare providers in patient management and treatment decisions, though negative results do not rule out infection. The device includes a positive control pack for quality assurance.

Clinical Evidence

No clinical data provided in the document; the submission relies on substantial equivalence to a previously cleared predicate device.

Technological Characteristics

Real-time RT-PCR; targets Flu A (matrix), Flu B (matrix), RSV (M gene). Uses bi-functional fluorescent probe-primers. Instrument: LIAISON MDX with LIAISON MDX Studio Software. Consumable: Direct Amplification Disc (DAD) with microfluidic channels and laser-activated valves. Detection: 4-channel optical system (FAM, JOE, CR610, Q670). Sterilization: N/A (reagents). Connectivity: Standalone instrument.

Indications for Use

Indicated for human patients with signs and symptoms of respiratory tract infection to aid in the differential diagnosis of influenza A, influenza B, and RSV viral infections using nasopharyngeal swab specimens.

Regulatory Classification

Identification

A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.

Special Controls

*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;” (2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and (3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 22, 2020 DiaSorin Molecular LLC Sharon Young Principal Regulatory Affairs Specialist 11331 Valley View Street Cypress. California 90630 ## Re: K201505 Trade/Device Name: Simplexa Flu A/B & RSV Direct Gen II, Simplexa Flu A/B & RSV Positive Control Pack Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay Regulatory Class: Class II Product Code: OCC, OOI Dated: June 4, 2020 Received: June 5, 2020 ## Dear Sharon Young: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post-marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Steven Gitterman, M.D., Ph.D. Deputy Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201505 #### Device Name Simplexa™ Flu A/B & RSV Direct Gen II and Simplexa™ Flu A/B & RSV Positive Control Pack #### Indications for Use (Describe) Simplexa™ Flu A/B & RSV Direct Gen II The DiaSorin Molecular Simplexa™ Flu A/B & RSV Direct Gen II assay is intended for use on the LIAISON® MDX instrument for the in vitro qualitative detection and differentiation of influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) from human patients with signs and symptoms of respiratory tract infection in conjunction with clinical and epidemiological risk factors. This test is intended for use as an aid in the differential diagnosis of influenza A. influenza B. and RSV viral infections in humans. Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for treatment or other patient management decisions. Performance characteristics for influenza A were established with clinical specimens collected during the 2010/2011 influenza season when 2009 H1N1 influenza and H3N2 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to the state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. #### Simplexa™ Flu A/B & RSV Positive Control Pack The DiaSorin Molecular Simplexa™ Flu A/B & RSV Positive Control Pack is intended to be used as a control with the Simplexa™ Flu A/B & RSV Direct kit and the Simplexa™ Flu A/B & RSV Direct Gen II kit for use on the LIAISON® MDX instrument. This control is not intended for use with other assays or systems. Type of Use (Select one or both, as applicable) | <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-variant: normal;"> <span style="font-weight: normal;"> <span> <span style="white-space: pre;"></span> </span> </span> </span> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-variant: normal;"> <span style="font-weight: normal;"> <span> <span style="white-space: pre;"></span> </span> </span> </span> </span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...