K033051 · Trinity Biotech USA · MJH · Nov 26, 2003 · Microbiology
Device Facts
Record ID
K033051
Device Name
LEGIONELLA PNEUMOPHILA IGG/IGM
Applicant
Trinity Biotech USA
Product Code
MJH · Microbiology
Decision Date
Nov 26, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3300
Device Class
Class 2
Intended Use
The Trinity Biotech Legionella pneumophila IgG/IgM Enzyme-Linked Immunosorbent Assay (ELISA) is intended for the qualitative detection of total antibodies (IgG and IgM) to Legionella pneumophila (Legionella) serogroups 1-6 in serum from patients with clinical suspicion of Legionella Disease. The Legionella pneumophila IgG/IgM ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of total antibodies (IgG and IgM) to Legionella pneumophila serogroups 1-6 in serum from patients with clinical suspicion of Legionella Disease. The assay is not intended to differentiate between the serotypes of Legionella.
Device Story
The Legionella pneumophila IgG/IgM ELISA is an in vitro diagnostic test used in clinical laboratories. It utilizes purified Legionella pneumophila antigens (serogroups 1-6) immobilized on microtiter wells. Patient serum is added; if specific IgG/IgM antibodies are present, they bind to the antigen. After washing, an enzyme-labeled anti-human IgG/IgM conjugate is added, followed by a substrate solution. The resulting color change is measured photometrically. This indirect measurement of antibody concentration assists clinicians in diagnosing Legionella infection. The device is intended for professional use in a clinical laboratory setting.
Clinical Evidence
Performance evaluated via comparison to Legionella IFA at two sites. Site 1 (n=33): 90.0% positive agreement. Site 2 (n=72): 98.53% positive agreement, 98.57% negative agreement. CDC panel (n=31 serum pairs): 93.5% agreement in detecting seroconversions. Precision studies (intra- and inter-assay) conducted per NCCLS EP5 guidelines, showing acceptable CVs across multiple sites and days.
Technological Characteristics
Enzyme-Linked Immunosorbent Assay (ELISA) using purified Legionella pneumophila antigen (serogroups 1-6) coated on solid-phase microtiter wells. Employs enzyme-labeled anti-human IgG/IgM conjugate and substrate for photometric detection. Quantitative output derived from color intensity measurement.
Indications for Use
Indicated for qualitative detection of total IgG/IgM antibodies to Legionella pneumophila serogroups 1-6 in serum from patients with clinical suspicion of Legionella disease.
Regulatory Classification
Identification
Haemophilus spp. serological reagents are devices that consist of antigens and antisera, including antisera conjugated with a fluorescent dye, that are used in serological tests to identify Haemophilus spp. directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Haemophilus and provides epidemiological information on diseases cause by these microorganisms. Diseases most often caused by Haemophilus spp. include pneumonia, pharyngitis, sinusitis, vaginitis, chancroid venereal disease, and a contagious form of conjunctivitis (inflammation of eyelid membranes).
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
K963318 — LEGIONELLA IGG/IGM ELISA TEST SYSTEM · Armkel, LLC · Mar 3, 1997
K970149 — LEGIONELLA ELISA TEST SYSTEM · Zeus Scientific, Inc. · Jun 18, 1997
K991074 — BARTELS LEGIONELLA URINARY ANTIGEN ELISA TEST · Intracel Corp. · Dec 21, 1999
K033064 — MYCOPLASMA IGG · Trinity Biotech USA · Nov 26, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 2 6 2003
K033a51
Summary of Safety and Effectiveness Information Legionella pneumophila IgG/IgM ELISA Test Kit
I. Trinity Biotech 2823 Girts Road Jamestown, NY 14701 Contact person: Bonnie B. DeJoy Telephone: 716-483-3851 Date of preparation: Nov. 20, 2003
II. Description of Device
The Legionella pneumophila IgG/IgM ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of total antibodies (IgG and IgM) to Legionella pneumophila serogroups 1-6 in serum from patients with clinical suspicion of Legionella Disease.
The Legionella pneumophila IgG/IgM ELISA test is an enzyme linked immunosorbent assay to detect IgG/IgM antibodies to legionella. Purified Legionella pneumophila antigen (serogroups 1, 2, 3, 4, 5, 6) is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation, the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG/IgM is added to each well. If antibody is present, it will bind to the antibody attached to the antigen on the well. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present, the substrate will undergo a color change. After an incubation period, the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.
#### III. Predicate Device
The Legionella pneumonphila IgG/IgM ELISA test is substantially equivalent to BioWhittaker's Legionella STAT test. Equivalence is demonstrated by the following comparative results:
# Performance Characteristics
### % Agreement Positive and % Agreement Negative
The Trinity Biotech Legionella pneumophila IgG/IgM ELISA was evaluated relative to Legionella IFA at two different sites. The first site was a commercial R&D lab located in Marvland. Thirty-three single IFA positive sera, from an outbreak and samples routinely submitted for Legionella testing, were tested. The results of the study are summarized in Table 3.
{1}------------------------------------------------
## Table 3 Comparison of Trinity Biotech Legionella pneumophila IgG/IgM ELISA and Legionella IFA Trinity Biotech Legionella pneumophila IgG/IgM ELISA
| | + | eq | - | Total |
|-------------|----|----|---|-------|
| + ≥ 256 | 27 | 3 | 3 | 33 |
| IFA - < 256 | 0 | 0 | 0 | 0 |
| Total | 27 | 3 | 3 | 33 |
| % Agreement positive | $27/30 = 90.00%$ | 95% Confidence interval = $79.0% - 100%$ |
|----------------------|------------------|------------------------------------------|
| % Agreement negative | NA | 95% Confidence interval = NA |
Equivocals were not included in the above calculations. The 95% confidence intervals were calculated using the normal method.
Please be advised that "% agreement positive" and "% agreement negative" refer to the comparison of this assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease.
The second site was a clinical laboratory in Pennsylvania. Seventy-two prospective serum for Legionella testing were tested. The results of the study are summarized in Table 4.
# Table 4 Comparison of Trinity Biotech Legionella pneumophila IgG/IgM ELISA and Legionella IFA
### Trinity Biotech Legionella pneumophila IgG/IgM ELISA
| | + | eq | - | Total | |
|-----|--------|----|---|-------|----|
| IFA | + ≥256 | 2 | 0 | 0 | 2 |
| | - <256 | 1 | 2 | 67 | 70 |
| | Total | 3 | 2 | 67 | 72 |
% Agreement positive = 67/68 = 98.53% 95% Confidence Interval = 95.6% - 100% % Agreement negative = 69/70 = 98.57% 95% Confidence Interval = 95.7% - 100%
Equivocals were not included in the above calculations.
The 95% confidence intervals were calculated using the normal method.
{2}------------------------------------------------
### Precision
Seven different sera were assayed at two different sites to determine the precision of the assay. An additional three sera were tested at site 1. Each sera was tested ten times each, on three different days at each of the two study sites. The intra- and inter-assay precision for each site is presented in Tables 5 and 6. The inter-site coefficient of variation (CV) for each serum is presented in Table 7.
# Table 5 Trinity Biotech Legionella pneumophila IgG/IgM ELISA Intra- and Inter-Assay Precision Study 1
| | Assay 1 (n=10) | | | Assay 2 (n=10) | | | Assay 3 (n=10) | | | Inter-Assay(n=30) | | |
|-------|----------------|-------|-------|----------------|-------|-------|----------------|-------|-------|-------------------|-------|--------|
| Sera# | X | SD | CV | X | SD | CV | X | SD | CV | X | SD | CV |
| 1 | 3.17 | 0.138 | 4.35% | 3.55 | 0.235 | 6.62% | 3.41 | 0.349 | 10.2% | 3.42 | 0.305 | 8.92% |
| 2 | 2.44 | 0.244 | 10.0% | 2.66 | 0.267 | 10.0% | 2.41 | 0.127 | 5.27% | 2.50 | 0.247 | 9.88% |
| 3 | 2.49 | 0.322 | 12.9% | 2.78 | 0.240 | 8.63% | 2.81 | 0.332 | 11.8% | 2.70 | 0.327 | 12.1% |
| 4 | 1.22 | 0.180 | 14.8% | 1.36 | 0.131 | 9.63% | 1.16 | 0.125 | 10.8% | 1.25 | 0.164 | 13.1% |
| 5 | 0.50 | 0.051 | 10.2% | 0.56 | 0.042 | 7.50% | 0.53 | 0.041 | 7.74% | 0.53 | 0.050 | 9.43% |
| 6 | 0.18 | 0.025 | 13.9% | 0.21 | 0.023 | 11.0% | 0.20 | 0.031 | 15.5% | 0.20 | 0.030 | 15.0% |
| 7 | 0.28 | 0.039 | 13.9% | 0.34 | 0.046 | 13.5% | 0.33 | 0.048 | 14.6% | 0.32 | 0.051 | 15.9% |
| 8 | 1.02 | 0.051 | 5.00% | 1.13 | 0.039 | 3.45% | 1.19 | 0.044 | 3.70% | 1.11 | 0.084 | 7.57% |
| 9 | 0.85 | 0.053 | 6.24% | 0.92 | 0.025 | 2.72% | 0.99 | 0.043 | 4.34% | 0.92 | 0.069 | 7.50% |
| 10 | 0.96 | 0.067 | 6.98% | 1.05 | 0.056 | 5.33% | 1.11 | 0.094 | 8.47% | 1.03 | 0.122 | 11.20% |
| HPC* | | | | | | | | | | 3.64 | .0402 | 11.05% |
| CAL** | | | | | | | | | | 1.44 | 0.122 | 8.44% |
| LPC* | | | | | | | | | | 1.49 | 0.195 | 13.11% |
| NC* | | | | | | | | | | 0.18 | 0.052 | 28.97% |
* n = 17 ** n = 51
## Table 6
## Trinity Biotech Legionella pneumophila IgG/IgM ELISA Intra- and Inter-Assay Precision
Study 2
| | | Assay 1 (n=10) | | | Assay 2 (n=10) | | | Assay 3 (n=10) | | | Inter-Assay(n=30) | | |
|-------|---------|----------------|-------|-------|----------------|-------|-------|----------------|-------|-------|-------------------|-------|--------|
| Sera# | | X | SD | CV | X | SD | CV | X | SD | CV | X | SD | CV |
| 1 | | 2.80 | 0.246 | 8.79% | 2.66 | 0.165 | 6.20% | 3.08 | 0.245 | 7.95% | 2.85 | 0.272 | 9.54% |
| 2 | | 3.10 | 0.343 | 11.1% | 3.05 | 0.276 | 9.05% | 3.14 | 0.259 | 8.25% | 3.10 | 0.293 | 9.45% |
| 3 | | 3.31 | 0.392 | 11.8% | 3.17 | 0.220 | 6.94% | 3.38 | 0.214 | 6.33% | 3.31 | 0.289 | 8.73% |
| 4 | | 1.10 | 0.135 | 12.3% | 1.15 | 0.131 | 11.4% | 1.18 | 0.142 | 12.0% | 1.16 | 0.138 | 11.9% |
| 5 | | 0.56 | 0.060 | 10.7% | 0.58 | 0.053 | 9.14% | 0.59 | 0.036 | 6.10% | 0.58 | 0.050 | 8.62% |
| 6 | | 0.28 | 0.016 | 5.71% | 0.26 | 0.013 | 5.00% | 0.29 | 0.020 | 6.90% | 0.28 | 0.020 | 7.14% |
| 7 | | 0.29 | 0.018 | 6.21% | 0.28 | 0.020 | 7.14% | 0.31 | 0.023 | 7.42% | 0.29 | 0.023 | 7.93% |
| HPC* | | | | | | | | | | | 3.16 | 0.092 | 2.91% |
| CAL** | | | | | | | | | | | 1.45 | 0.060 | 4.11% |
| LPC* | | | | | | | | | | | 1.68 | 0.190 | 11.29% |
| NC* | | | | | | | | | | | 0.35 | 0.121 | 34.44% |
| | * n = 5 | | | | | | | | | | | | |
{3}------------------------------------------------
#### Table 7
# Trinity Biotech Legionella pneumophila IgG/IgM ELISA Inter-Site Precision Study Inter Site (n=60)
| Sera # | X | SD | CV | # positive | #equivocal | #negative |
|--------|------|-------|--------|------------|------------|-----------|
| 1 | 3.13 | 0.406 | 13.0% | 60 | 0 | 0 |
| 2 | 2.80 | 0.403 | 14.4% | 60 | 0 | 0 |
| 3 | 3.00 | 0.431 | 14.4% | 60 | 0 | 0 |
| 4 | 1.21 | 0.158 | 13.1% | 43 | 16 | 1 |
| 5 | 0.56 | 0.055 | 9.82% | 0 | 0 | 60 |
| 6 | 0.24 | 0.046 | 19.2% | 0 | 0 | 60 |
| 7 | 0.30 | 0.040 | 13.3% | 0 | 0 | 60 |
| 8* | 1.11 | 0.084 | 7.57% | 19 | 11 | 0 |
| 9* | 0.92 | 0.069 | 7.50% | 0 | 20 | 10 |
| 10* | 1.03 | 0.122 | 11.20% | 8 | 20 | 2 |
| HPC** | 3.42 | 0.401 | 11.75% | 9 | 0 | 0 |
| CAL*** | 1.45 | 0.069 | 4.77% | 27 | 0 | 0 |
| LPC** | 1.56 | 0.240 | 15.46% | 9 | 0 | 0 |
| NC** | 0.27 | 0.124 | 45.68% | 0 | 0 | 9 |
- * n = 30 ** n = 9 *** n = 27 X = Mean SD = Standard Deviation CV = Coefficient of Variation = SD/X x 100
The methods in NCCLS EP5 were utilized for precision parameters.
## IFA Paired Serum Analysis (CDC Panel)
The following information is from a serum panel tested at the Centers for Disease Control (CDC) by IFA and confirmed to be serologically positive for an increase in titer from <1:256 to >1:256. The sera were submitted to CDC for titer confirmation. The results are presented as a means to convey further information on the performance of this assay with a masked serum panel. This does not imply an endorsement of the assay by the CDC.
The panel consisted of thirty-one serum pairs showing a greater than 4-fold increase in IFA titer. Each serum pair was evaluated on the Trinity Biotech Legionella pneumophila IgG/IgM ELISA assay to determine a seroconversion in antibody. Twenty nine pairs had a seroconversion, thus giving a % agreement positive of 29/31 = 93.5% in detecting seroconversions.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, which is a traditional symbol of medicine and healthcare.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 26 2003
Ms. Bonnie B. DeJoy Director, Quality Systems Trinity Biotech USA P.O. Box 1059 Jamestown, NY 14702-1059
Re: k033051
> Trade/Device Name: Legionella pneumophila IgG/IgM ELISA Regulation Number: 21 CFR 866.3300 Regulation Name: Haemophilus spp. serological reagents Regulatory Class: Class II Product Code: MJH Dated: September 17, 2003 Received: September 29, 2003
Dear Ms. DeJoy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number: K033051
Device Name: Trinity Biotech Captia™ Legionella pneumophila IgG/IgM ELISA
Indications For Use: The Trinity Biotech Captia™ Legionella pneumophila IgG/IgM ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of total antibodies (IgG and IgM) to serogroups 1-6 in serum from patients with clinical suspicion of Legionella Disease. The assay is not intended to differentiate between the serotypes of Legionella pneumophila.
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V (Per 21 XFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
Thomas J. O'Brien for SYH
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K033051
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.