The Pylori IgG ELISA kit is an Enzyme-Linked Immunosorbent Assays (ELISA) for the qualitative determination of IgG antibodies in human serum to Helicobacter pylori antigen. The Wampole Pylori IgG ELISA assay may be used as an aid in the diagnosis of Helicobacter pylori infection in persons with gastrointestinal symptoms. For In Vitro Diagnostic Use Only.
Device Story
ELISA test kit for qualitative detection of IgG antibodies to H. pylori in human serum; utilizes purified H. pylori antigen coated on microtiter wells. Patient serum added; if specific IgG present, binds to antigen. After incubation/washing, enzyme-labeled anti-human IgG conjugate added; binds to captured antibody. Substrate solution added; enzyme presence triggers color change. Color intensity measured photometrically; provides indirect measurement of specific antibody concentration. Used in clinical laboratory settings by trained personnel. Results aid clinicians in diagnosing H. pylori infection in symptomatic patients, supporting clinical decision-making regarding treatment for gastrointestinal conditions.
Clinical Evidence
Clinical evaluation via masked testing of 371 serum samples from patients with various gastrointestinal diagnoses (gastritis, ulcers, dyspepsia, esophagitis) and normal controls. Compared against biopsy (culture or stain). Sensitivity: 96.4% (95% CI: 94.1%–98.8%); Specificity: 96.1% (95% CI: 92.3%–99.9%); Agreement: 96.4% (95% CI: 94.4%–98.3%). Precision evaluated via intra- and inter-assay testing (n=30). Cross-reactivity testing performed against C. jejuni, C. fetus, and Borrelia burgdorferi, showing no significant cross-reactivity.
Technological Characteristics
Enzyme-Linked Immunosorbent Assay (ELISA) using purified H. pylori antigen immobilized on solid-phase microtiter wells. Detection via enzyme-labeled anti-human IgG conjugate and substrate-induced colorimetric reaction. Photometric measurement of color intensity. In vitro diagnostic kit format.
Indications for Use
Indicated for qualitative detection of IgG antibodies to H. pylori in human serum as an aid in diagnosing H. pylori infection in patients with gastrointestinal symptoms (e.g., gastritis, gastric/duodenal ulcers, non-ulcer dyspepsia, esophagitis). No specific age or gender restrictions.
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
Predicate Devices
Biopsy (culture or stain)
Reference Devices
Pylori Stat
Related Devices
K990462 — IS H. PYLORI IGG ELISA TEST SYSTEM · Columbia Bioscience, Inc. · Sep 3, 1999
K110745 — HELICOBACTER PYLORI ELISA IGG TEST KIT · Gold Standard Diagnostics · Mar 2, 2012
K991820 — HELICOBACTER PYLORI IGG ELISA TEST SYSTEM. AN ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA)FOR THE DETECTION OF IGG ANTIBODI · Zeus Scientific, Inc. · Aug 12, 1999
K973508 — PYLORI DETECT IGG · Micro Detect, Inc. · Mar 10, 1998
Submission Summary (Full Text)
{0}
K961221
JUL 30 1996
Summary of Safety and Effectiveness Information Pylori IgG ELISA Test Kit
I. Immuno Probe Inc.
1306 Bailes Lane, Suite F
Frederick, Maryland 21701
Contact person: William Boteler
Telephone: 301-695-7920
Date of preparation: March 28, 1995
II. Description of Device
The Pylori IgG ELISA kit is an Enzyme-Linked Immunosorbent Assays (ELISA) for the qualitative determination of IgG antibodies in human serum to *Helicobacter pylori* antigen. The Wampole Pylori IgG ELISA assay may be used as an aid in the diagnosis of *Helicobacter pylori* infection in persons with gastrointestinal symptoms. For In Vitro Diagnostic Use Only.
The Pylori IgG ELISA test is an enzyme linked immunosorbent assay to detect IgG antibodies to *Helicobacter pylori*. Purified *Helicobacter pylori* antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.
III. Predicate Device
The Pylori IgG ELISA test is substantially equivalent to biopsy. Equivalence is demonstrated by the following comparative results:
{1}
# Performance Characteristics
## A. Evaluation of Pylori IgG ELISA Sensitivity and Specificity Relative to Biopsy
The Pylori IgG ELISA is a modification of Pylori Stat. Pylori Stat was originally evaluated by masked testing 386 serum from five geographically different areas, with biopsy with stain or culture results for *H. pylori*. The serum were from patients with random gender and various ages with the following clinical diagnosis: gastritis, gastric ulcer, duodenal ulcer, non-ulcer dyspepsia, esophagitis and normal. Table 1 illustrates the sensitivity and specificity of Pylori Stat to Biopsy.
### Table 1
Pylori Stat IgG ELISA Sensitivity and Specificity
| Biopsy* | Pylori Stat | | | | Total |
| --- | --- | --- | --- | --- | --- |
| | + | + | eq | - | |
| + | 261 | 12 | 4 | 277 | |
| - | 2 | 2 | 105 | 109 | |
| Total | 263 | 14 | 109 | 386 | |
Sensitivity = 261/265 = 98.5% 95% Confidence interval = 97.0% - 100%
Specificity = 105/107 = 98.1% 95% Confidence interval = 95.5% - 100%
Agreement = 366/372 = 98.4% 95% Confidence interval = 97.1% - 99.7%
* Culture or stain
Equivocals were not included in the above calculations.
The 95% confidence intervals were calculated using the normal method.
{2}
The Pylori IgG ELISA was evaluated by masked testing 371 serum from five geographically different areas, with biopsy with stain or culture results for H. pylori. The serum were from patients with random gender and various ages with the following clinical diagnosis: gastritis, gastric ulcer, duodenal ulcer, non-ulcer dyspepsia, esophagitis and normal. Table 2 illustrates the sensitivity and specificity of Pylori IgG ELISA to Biopsy.
Table 2
Pylori IgG ELISA Relative Sensitivity and Specificity
| Wampole Pylori IgG ELISA | | | | | |
| --- | --- | --- | --- | --- | --- |
| | | + | eq | - | Total |
| Biopsy* | + | 244 | 13 | 9 | 266 |
| | - | 4 | 2 | 99 | 105 |
| | Total | 248 | 15 | 108 | 371 |
| Sensitivity = 244/253 = 96.4% | | | 95% Confidence interval = 94.1% - 98.8% | | |
| Specificity = 99/103 = 96.1% | | | 95% Confidence interval = 92.3% - 99.9% | | |
| Agreement = 343/356 = 96.4% | | | 95% Confidence interval = 94.4% - 98.3% | | |
* Culture or stain positive
Equivocals were not included in the above calculations.
The 95% confidence intervals were calculated using the normal method.
{3}
# B. Evaluation of Pylori IgG ELISA Precision
The precision of the Pylori IgG ELISA was determined by testing six different sera ten times each on three days. The mean coefficients of variation from the intra- and inter-assays are presented in Table 3
Table 3
Pylori IgG ELISA Precision
| | Assay 1 (n=10) | | | Assay 2 (n=10) | | | Assay 3 (n=10) | | | InterAssay (n=30) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | X | SD | CV | X | SD | CV | X | SD | CV | X | SD | CV |
| 1 | 3.13 | 0.209 | 6.68% | 3.05 | 0.145 | 4.75% | 3.13 | 0.210 | 6.71% | 3.10 | 0.188 | 6.06% |
| 2 | 2.08 | 0.151 | 7.26% | 2.15 | 0.156 | 7.26% | 2.01 | 0.127 | 6.32% | 2.08 | 0.151 | 7.26% |
| 3 | 2.31 | 0.258 | 11.2% | 2.29 | 0.115 | 5.02% | 2.24 | 0.173 | 7.72% | 2.28 | 0.187 | 8.20% |
| 4 | 1.17 | 0.184 | 15.7% | 1.47 | 0.148 | 10.1% | 1.30 | 0.179 | 13.8% | 1.31 | 0.207 | 15.8% |
| 5 | 0.06 | 0.018 | 30.0% | 0.12 | 0.014 | 11.7% | 0.11 | 0.056 | 50.9% | 0.08 | 0.031 | 38.8% |
| 6 | 0.10 | 0.016 | 16.0% | 0.14 | 0.020 | 14.3% | 0.10 | 0.067 | 67.0% | 0.12 | 0.023 | 19.2% |
# C. Evaluation of Pylori IgG ELISA with Potentially Cross Reactive Sera.
Pylori IgG ISR values were determined for paired sera from C. jujuni infections and single sera from C. fetus infections. The data in Table 4 shows no rise in antibody for C. jujuni paired sera, and negative responses for C. fetus infections indicating a lack of cross reactivity to these closely related organisms. Serum pairs 1, 3 and 4 demonstrate antibody to H. pylori. The pairs do not show as rise in antibody as would be expected in acute C. jujuni infection, therefore the response is considered to be specific for H. pylori with no cross reaction with C. jujuni. Sera positive for antibodies to Borrelia burgdorferi by ELISA and Western Blot were negative indicating a lack of cross reactivity.
{4}
Table 4
Pylori IgG Results with
Potentially Cross-reactive Sera
| Serum # | Diagnosis* | Pylori IgG ISR |
| --- | --- | --- |
| Acute 1 | C. jejuni Diarrhea | 2.30 |
| Convalescent 1 | | 2.41 |
| Acute 2 | C. jejuni Diarrhea | 2.11 |
| Convalescent 2 | | 2.23 |
| Acute 3 | C. jejuni Diarrhea | 1.23 |
| Convalescent 3 | | 1.20 |
| Acute 4 | C. jejuni Diarrhea | 0.40 |
| Convalescent 4 | | 0.57 |
| Acute 5 | C. jejuni Diarrhea | 0.88 |
| Convalescent 5 | | 0.89 |
| 6 | C. fetus Endocarditis | 0.59 |
| 7 | C. fetus Endocarditis | 0.63 |
| 8 | C. fetus Endocarditis | 0.72 |
| 9 | C. fetus Bacteremia | 0.38 |
| *All cases diagnosed by culture; C. jejuni infection by fecal culture on Campylobacter-specific media, C. fetus infection by blood culture. | | |
| 10. | Borrelia burgdorferi | 0.44 |
| 11. | Borrelia burgdorferi | 0.20 |
| 12. | Borrelia burgdorferi | 0.22 |
| 13. | Borrelia burgdorferi | 0.89 |
| 14. | Borrelia burgdorferi | 0.11 |
{5}
26
15. Borrelia burgdorferi 0.20
16. Borrelia burgdorferi 0.16
17. Borrelia burgdorferi 0.59
18. Borrelia burgdorferi 0.09
19. Borrelia burgdorferi 0.13
All Borrelia burgdorferi sera were positive for antibodies by ELISA and Western Blot. Their clinical histories were suggestive of Lyme Disease.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.