H. PYLORI QUIK CHEK

K181379 · Techlab, Inc. · LYR · Aug 21, 2018 · Microbiology

Device Facts

Record IDK181379
Device NameH. PYLORI QUIK CHEK
ApplicantTechlab, Inc.
Product CodeLYR · Microbiology
Decision DateAug 21, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3110
Device ClassClass 1

Intended Use

The TECHLAB® H. PYLORI QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the qualitative detection of Helicobacter pylori specific antigen in a single use cassette. It is intended for use with human fecal specimens to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori antigen following treatment. The test can be used with unpreserved fecal specimens and fecal specimens preserved in transport media from patients suspected of H. pylori infection. Testing of patients to demonstrate loss of H. pylori antigen following treatment should be performed no sooner than 4 weeks after completion of the treatment regimen. Test results should be taken into consideration by the physician in conjunction with the patient history and symptoms.

Device Story

Rapid membrane enzyme immunoassay for qualitative detection of H. pylori antigen in human fecal specimens. Device uses single-use cassette with reaction window containing immobilized antibodies. Input: unpreserved or transport-media-preserved stool samples. Principle: lateral flow immunochromatographic assay; H. pylori antigen binds to HRP-coupled antibodies; visual appearance of blue lines on 'C' (control) and 'T' (test) sides indicates presence of antigen. Used in clinical settings; operated by laboratory personnel. Results interpreted visually by clinician to aid diagnosis of H. pylori infection or confirm eradication post-treatment. Benefits include rapid, point-of-care-compatible diagnostic confirmation of infection status.

Clinical Evidence

Prospective clinical study (n=122) compared device to composite reference method (CRM: histology/rapid urease test). Sensitivity 97.0% (95% CI: 84.7-99.5%), specificity 100% (95% CI: 95.9-100%). Post-therapy study (n=9) showed 100% sensitivity. Retrospective study (n=200) compared to FDA-cleared ELISA, showing 98.9% PPA and 97.2% NPA. Bench testing confirmed 100% reproducibility and no prozone/hook effect at 10x clinical concentrations.

Technological Characteristics

Rapid membrane enzyme immunoassay (ELISA). Components: single-use cassette, HRP-conjugated antibodies, tetramethylbenzidine substrate. Analyte: H. pylori fecal antigen. Detection: visual colorimetric (blue lines). Storage: 2-8°C. Analytical sensitivity (LoD): 16.08 ng/mL (unpreserved), 13.01-19.96 ng/mL (transport media). No cross-reactivity with common intestinal pathogens.

Indications for Use

Indicated for patients suspected of H. pylori infection, including those with dyspepsia, gastritis, or peptic ulcer symptoms. Applicable for initial diagnosis and post-treatment monitoring (at least 4 weeks post-therapy). No specific age or gender contraindications.

Regulatory Classification

Identification

Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration". August 21, 2018 TECHLAB. Inc. Donna Link Director Regulatory and Compliance 2001 Kraft Drive Corporate Research Center Blacksburg, Virginia 24060-6358 Re: K181379 Trade/Device Name: H. Pylori Quik Chek Regulation Number: 21 CFR 866.3110 Regulation Name: Campylobacter fetus serological reagents Regulatory Class: Class I Product Code: LYR Dated: May 23, 2018 Received: May 24, 2018 Dear Donna Link: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Ribhi Shawar -S For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181379 Device Name H. PYLORI QUIK CHEK #### Indications for Use (Describe) The TECHLAB H. PYLORI QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the qualitative detection of Helicobacter pylori specific antigen in a single use cassette. It is intended for use with human fecal specimens to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori antigen following treatment. The test can be used with unpreserved fecal specimens and fecal specimens preserved in transport media from patients suspected of H. pylori infection. Testing of patients to demonstrate loss of H. pylori antigen following treatment should be performed no sooner than 4 weeks after completion of the treatment regimen. Test results should be taken into consideration by the physician in conjunction with the patient history and symptoms. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## H. PYLORI QUIK CHEK™ 510(k) SUMMARY ### This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92. ### Applicant/Contact Information: | Date Prepared: | August 16, 2018 | |----------------|---------------------------------------------------------------------------| | Name: | TECHLAB, Inc. | | Address: | 2001 Kraft Drive<br>Corporate Research Center<br>Blacksburg, VA 24060 USA | | Contact Person: | Donna T. Link | |-----------------|-------------------| | Phone Number: | 540-953-1664 | | Email: | dlink@techlab.com | #### 1.1 Manufacturing Facility Address TECHLAB, Inc. 20 Corporate Drive Radford, VA 24141 USA #### 1.2 Product and Trade Name of the Device H. PYLORI QUIK CHEK™ #### 1.3 Common Name or Classification Name H. pylori detection test #### 1.4 Classification and Regulation Class I 21 CFR 866.3110; Campylobacter fetus serological reagents #### 1.5 Product Code LYR – Campylobacter pylori #### 1.6 Panel 83 Microbiology #### 1.7 Reason for Premarket Notification The development of a new rapid membrane enzyme immunoassay for the qualitative detection of H. pylori in a single use cassette. {4}------------------------------------------------ ## Intended Use The TECHLAB® H. PYLORI QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the qualitative detection of Helicobacter pylori specific antigen in a single use cassette. It is intended for use with human fecal specimens to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori antigen following treatment. The test can be used with unpreserved fecal specimens and fecal specimens preserved in transport media from patients suspected of H. pylori infection. Testing of patients to demonstrate loss of H. pylori antigen following treatment should be performed no sooner than 4 weeks after completion of the treatment regimen. Test results should be taken into consideration by the physician in conjunction with the patient history and symptoms. ## Caution: U.S. Federal Law restricts this device to sale by or on the order of a physician ## Explanation It is estimated that half of the global population is infected with H. pylori. The majority of those infected remain asymptomatic and do not require treatment (colonized individuals). A minority of infected individuals develop gastritis, and a fraction of those further develop gastric ulcers or gastric cancer. The diagnosis of H. pylori infection is endoscopy with biopsy - the biopsied tissue is tested for the presence of H. pylori by culture, histology, or rapid urease test. Under current guidelines, endoscopy is still recommended for the diagnosis of H. pylori infection in patients with alarm symptoms (e.g. GI bleeding, sudden weight loss, excessive vomiting, anemia), or patients over the age of 55. However, for younger patients not exhibiting alarm symptoms, non-invasive tests such as the urea breath test (UBT) or fecal antigen test are recommended for diagnosis of H. py/ori infection. Following completion of a treatment regimen of antibiotics and a proton pump inhibitor (PPI), it is recommended that patients be tested to verify eradication of H. pylori infection. Serum antibody tests are also available, but these are unable to distinguish between past and current infection. By detecting antigen present in fecal specimens, the H. PYLORI QUIK CHEK™ test allows for the non-invasive detection of H. pylori when endoscopy is not required. ## Device Description The H. PYLORI QUIK CHEK™ test utilizes antibodies specific for H. pylori antigen. The Membrane Device contains a Reaction Window with two vertical lines of immobilized antibodies. The test line ("T") contains antibodies specific for H, pylori antigen. The ("C") contains antibodies to horseradish peroxidase (HRP). The Conjugate consists of antibodies to H. pylori antigen coupled to horseradish peroxidase. To perform the test, the sample is added to a tube containing a mixture of Diluent and Conjugate. The diluted sample-conjugate mixture is added to the Sample Well and the device is allowed to incubate at room temperature for 15 minutes. During the incubation, any H. pylori antigen in the sample binds to the antibody-peroxidase conjugate. The antigen-antibody-peroxidase complexes migrate through a filter pad to a membrane where they are captured by the immobilized anti-H. pylori antigen antibodies in the test line. The Reaction Window is subsequently washed with Wash Buffer, followed by the addition of Substrate. After a 10-minute incubation period, the Reaction Window is examined visually for the appearance of vertical blue lines on the "C" and "T" sides of the Reaction Window. A blue line on the "T" side of the Reaction Window indicates a positive result. A positive "C" reaction, indicated by a vertical blue line on the "C" side of the Reaction Window, confirms that the sample and reagents were added correctly, the reagents were active at the time of performing the assay, and that the sample migrated properly through the Membrane Device. It also confirms the reactivity of the other reagents associated with the assay. {5}------------------------------------------------ ### Materials Provided - . Membrane Devices - 25 pouches, each pouch contains 1 device - Conjugate (2.5 mL) Antibody specific for H. pylori antigen coupled to horseradish . peroxidase in a buffered protein solution - Diluent (22 mL) Buffered protein solution with black graduated dropper assembly . - Positive Control (2 mL) H. pylori antigen in a buffered protein solution . - Substrate (3.5 mL) Solution containing tetramethylbenzidine ● - Wash Buffer (12 mL) Buffered solution with white graduated dropper assembly ● - Disposable plastic pipettes (50) Graduated at 25 µL, 100 µL, 200 µL, 300 µL, 400 µL and . 500 µL - . Wooden applicator sticks (50) The predicate device (ImmunoCard STAT!® HpSA) and the H. PYLORI QUIK CHEK™ test both detect H. pylori in fecal specimens and are substantially equivalent in principle. The following table shows a comparison of both devices. | Predicate Device Comparison Table<br>Similarities | | | |---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Item | H. PYLORI QUIK CHEKTM | ImmunoCard STAT!® HpSA (K032222) | | Intended Use | The TECHLAB® H. PYLORI QUIK CHEK™<br>test is a rapid membrane enzyme<br>immunoassay for the qualitative detection of<br><i>Helicobacter pylori</i> specific antigen in a single<br>use cassette. It is intended for use with<br>human fecal specimens to aid in the diagnosis<br>of <i>H. pylori</i> infection and to demonstrate loss<br>of <i>H. pylori</i> antigen following treatment. The<br>test can be used with unpreserved fecal<br>specimens and fecal specimens preserved in<br>transport media from patients suspected of <i>H.</i><br><i>pylori</i> infection. Testing of patients to<br>demonstrate loss of <i>H. pylori</i> antigen following<br>treatment should be performed no sooner<br>than 4 weeks after completion of the<br>treatment regimen. Test results should be<br>taken into consideration by the physician in<br>conjunction with the patient history and<br>symptoms. | ImmunoCard STAT! HpSA is a rapid <i>in</i><br><i>vitro</i> qualitative assay for the detection of<br><i>Helicobacter pylori</i> antigen in human stool.<br>The stool antigen detection is intended to aid<br>in the diagnosis of <i>H. pylori</i> infection and to<br>demonstrate loss of <i>H. pylori</i> stool antigen<br>following treatment. Conventional medical<br>practice recommends that testing by any<br>method to confirm the loss of antigen be done<br>at least four weeks following completion of<br>therapy. | | Measured<br>analyte | Detection of <i>H. pylori</i> stool antigen | Detection of <i>H. pylori</i> stool antigen | | Target<br>Population | Persons suspected of having<br><i>H. pylori</i> infection | Same | | Type of Test | Qualitative | Same | {6}------------------------------------------------ | Predicate Device Comparison Table<br>Similarities (continued) | | | |---------------------------------------------------------------|------------------------------------------------------------------------|----------------------------------| | Item | H. PYLORI QUIK CHEK™ | ImmunoCard STAT!® HpSA (K032222) | | Controls | Positive and negative control included in kit<br>Internal Control line | Same | | Storage | Refrigerated (2°C – 8°C) | Same | There are no differences between the subject device and the predicate(s) with respect to indications and intended use. | Predicate Device Comparison Table<br>Differences | | | |--------------------------------------------------|----------------------------------------------------------|----------------------------------------------| | ltem | H. PYLORI QUIK CHEKTM | ImmunoCard STAT!® HpSA (K032222) | | Format | Single Use Membrane Cassette<br>25 tests | Single Use Lateral Flow Cassette<br>20 tests | | Specimen Type | Fecal Specimens in Cary-Blair and C&S<br>Transport Media | Unpreserved Fecal Specimen | | Time to Result | 30 minutes | 5 minutes | | Technology | Enzyme Linked Immunoassay (ELISA) | Immunochromatographic (ICT) | | Antibody Format | Polyclonal/Polyclonal | Monoclonal/Monoclonal | ## Summary of Performance Data The performance of the H. PYLORI QUIK CHEK™ test was evaluated at 6 independent sites. Patients were recruited that were undergoing endoscopy as part of routine care. A composite reference method (CRM) comparison was used in the evaluation consisting of rapid urease and histology of the biopsy samples. The following table shows a summary of the clinical performance data. The results of the study show that the H. PYLORI QUIK CHEK™ test exhibity of 97.0% and specificity 100% with CRM biopsy results. {7}------------------------------------------------ ### Age and Gender Distribution Age and gender information was available for 122 patients. The ages ranged from 19 to 82 years. Of the 122 patients tested, 68% were female and 32% were male. No difference in test performance was observed based on patient age or gender. #### Initial Diagnosis H. PYLORI QUIK CHEK™ test versus Composite reference Method (CRM) | N = 122 | CRM<br>Positive | CRM<br>Negative | |----------------------------------|-----------------|-----------------| | H. PYLORI QUIK CHEK™<br>Positive | 32 | 0 | | H. PYLORI QUIK CHEK™<br>Negative | 1* | 89 | | | | 95% Confidence Limits | |-------------|--------|-----------------------| | Sensitivity | 97.0% | 84.7% - 99.5% | | Specificity | 100.0% | 95.9% - 100.0% | *Additional testing with an FDA cleared H. pylori stool antigen teste provided an antigen negative result. ## Post-Therapy For Eradication (post-therapy), there were 9 samples from patients being tested post therapy. The results show that the H. PYLORI QUIK CHEK™ test exhibited a sensitivity of 100% with the composite reference method. | N = 9 | CRM<br>Positive | CRM<br>Negative | | | 95% Confidence Limits | |----------------------------------|-----------------|-----------------|-------------|--------|-----------------------| | H. PYLORI QUIK CHEK™<br>Positive | 9 | 0 | Sensitivity | 100.0% | 70.1% - 100.0% | | H. PYLORI QUIK CHEK™<br>Negative | 0 | 0 | | | | {8}------------------------------------------------ ### Retrospective Sample Study A supplemental retrospective sample study was performed comparing the H. PYLORI QUIK CHEK™ test to an FDA cleared commercial ELISA. For this study, 200 samples (94 positive and 106 negative by the commercial ELISA). There was 98.9% Positive Agreement and 97.2% Negative Agreement of results between the assays. | N = 200 | FDA Cleared Commercial<br>ELISA Positive | FDA Cleared Commercial<br>ELISA Negative | |----------------------------------|------------------------------------------|------------------------------------------| | H. PYLORI QUIK CHEK™<br>Positive | 93 | 3* | | H. PYLORI QUIK CHEK™<br>Negative | 1** | 103 | | | | 95% Confidence Limits | |----------------------------|-------|-----------------------| | Percent Positive Agreement | 98.9% | 94.2% - 99.8% | | Percent Negative Agreement | 97.2% | 92.0% - 99.0% | *H. pylori DNA was amplified from the samples with PCR **No H. pylori DNA was amplified from the sample with PCR ### Reproducibility The reproducibility of the H. PYLORI QUIK CHEK™ test was determined using 8 fecal specimens that were coded to prevent identification during testing. Testing was performed at 2 independent laboratories and on-site at TECHLAB, Inc. The samples were tested in triplicate twice a day over a 5day period by multiple technicians at each site using 2 different kit lots. The results were as expected among the different locations, and exhibited an overall percent agreement of 100%. #### Analytical Sensitivity The Limit of Detection (LoD) for the H. PYLORI QUIK CHEK™ test was established at 16.08 ng/ml. in fecal matrix (0.24 ng/test) for Helicobacter pylori antigen using cell lysate antigen prepared from H. pylori strain ATCC 43526. For specimens in Protocol™ Cary Blair media, the LoD was established at 13.01 ng/mL (0.20 ng/test). For specimens in Protocol™ C&S media, the LoD was established at 19.96 ng/mL (0.31 ng/test). #### Analytical Specificity (Cross Reactivity) The H. PYLORI QUIK CHEK™ test was evaluated for cross-reactivity with common intestinal organisms and viruses listed below. None of the organisms or viruses were shown to interfere with the performance of the H. PYLORI QUIK CHEK™ test. Acinetobacter baumannii Borrelia burgdorferi Campylobacter helveticus Campylobacter lari Clostridium bifermentans Edwardsiella tarda Bacillus cereus Campylobacter coli Campylobacter hyointestinalis Campylobacter upsaliensis Clostridium difficile Enterobacter cloacae Bacillus subtilis Campylobacter fetus Campylobacter jejuni Candida albicans Clostridium perfringens Enterococcus faecalis {9}------------------------------------------------ Escherichia coli Escherichia coli ETEC Escherichia coli 0157:H7 (toxigenic) Listeria monocytogenes Prevotella melaninogenica Pseudomonas fluorescens Staphylococcus aureus (Cowan's) Adenovirus Types 2. 40 Echovirus 9, 22 Escherichia coli EIEC Escherichia coli 0157:H7 (non-toxigenic) Haemophilus influenzae Peptostreptococcus anaerobius Proteus vulgaris Salmonella typhimurium Streptococcus aqalactiae Human Coronavirus Enterovirus 70 Escherichia coli EPEC Lactobacillus acidophilus Porphyromonas asaccharolytica Pseudomonas aeruginosa Staphylococcus aureus Yersinia enterocolitica Coxsackievirus B1. B2. B3. B6 Human Rotavirus ## Inclusivity Study The following strains, which include isolates representing described H. pylori populations, were tested for reactivity with the H. PYLORI QUIK CHEK™ test. All strains tested generated a positive result. | ATCC 700392 | ATCC 43526 | ATCC 700824 | |-------------|------------|-------------| | JP26 | ATCC 43504 | ATCC 43579 | ### Interfering Substances (U.S. Formulation) The following substances had no effect on positive H. PYLORI QUIK CHEK™ test results analyzed at the concentrations indicated: Barium sulfate (5% w/v), Benzalkonium Chloride (1% w/v), Ciprofloxacin (0.25% w/v), Ethanol (1% w/v), Hog gastric mucin (3.5% w/v), Human blood (40% v/v), Hydrocortisone (1% w/v), Imodium® (5% v/v), Kaopectate® (5% v/v), Leukocytes (0.05% v/v), Maalox® Advanced (5% v/v), Mesalazine (10% w/v), Metronidazole (0.25% w/v), MiraLax® (3350 PEG)(7% w/v), Mineral Oil (10% w/v), Mylanta® (4.2 mq/mL), Naproxen Sodium (5% w/v), Nonoxynol-9 (1% w/v), Palmitic Acid/Fecal Fat (40% w/v), Pepto-Bismo® (5% v/v), Phenvlephrine (1% w/v), Prilosec OTC® (5 µg/mL), Sennosides (1% w/v), Simethicone (10% w/v), Stearic Acid/Fecal Fat (40% w/v), Tagamet® (5 ug/mL), TUMS® (50 µg/mL), Human Urine (5% v/v), and Vancomycin (0.25% w/v). #### Prozone To ensure that a high concentration of H. pylori antigen does not interfere with a positive reaction in the H. PYLORI QUIK CHEK™ test, high positive samples were prepared by spiking a negative fecal pool at concentrations up to 10 times the highest concentration of antigen observed in a positive clinical specimen. A total of 5 different dilutions of H. pylori antigen was prepared and tested in triplicate. The results demonstrated that there was no overall prozone effect, that elevated levels of antigen did not affect the detection of the antigen. #### CONCLUSION The conclusions drawn from the nonclinical and clinical tests demonstrate that the H. PYLORI QUIK CHEK™ test is safe and effective and substantially equivalent to the predicate device in performance. The information submitted in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...