APTIMA COMBO 2 ASSAY (PANTHER SYSTEM)

K132251 · Hologic / Gen-Probe Incorporated · LSL · Oct 17, 2013 · Microbiology

Device Facts

Record IDK132251
Device NameAPTIMA COMBO 2 ASSAY (PANTHER SYSTEM)
ApplicantHologic / Gen-Probe Incorporated
Product CodeLSL · Microbiology
Decision DateOct 17, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3390
Device ClassClass 2

Intended Use

The APTIMA Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal urogenital disease using the PANTHER System as specified. On the PANTHER System, the assay may be used to test the following specimens from symptomatic and asymptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens, clinician-collected gynecological specimens collected in the PreservCyt Solution, patient-collected vaginal swab specimens,¹ and male urine specimens. ¹ Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use.

Device Story

Nucleic acid amplification test (NAAT) for qualitative detection/differentiation of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) rRNA; utilizes target capture, transcription-mediated amplification (TMA), and dual kinetic assay (DKA). Input: clinician-collected swabs (endocervical, vaginal, male urethral), PreservCyt liquid Pap, patient-collected vaginal swabs, male urine. Process: rRNA isolated via magnetic microparticles; specific target amplification; chemiluminescent DNA probes labeled with acridinium ester; luminometer measures photon signals (Relative Light Units). Operation: performed on PANTHER System in clinical labs; automated processing. Output: RLU measurements interpreted by software against established cutoffs to provide CT/GC results. Clinical decision-making: aids diagnosis of urogenital chlamydial/gonococcal disease; enables targeted patient treatment.

Clinical Evidence

Two prospective, multicenter clinical studies evaluated performance. Study 1 included 580 males and 1332 females; Study 2 included 1492 males. Primary endpoints were sensitivity and specificity compared to an infected status algorithm using reference NAATs. For male urine, CT sensitivity was 95.2% (95% CI: 91.3-97.4%) and specificity was 99.8% (95% CI: 99.4-99.9%). GC sensitivity was 98.7% (95% CI: 92.9-99.8%) and specificity was 99.7% (95% CI: 99.3-99.9%).

Technological Characteristics

Uses Target Capture (TC), Transcription-Mediated Amplification (TMA), and Hybridization Protection Assay (HPA). Detection via chemiluminescent acridinium ester-labeled DNA probes. Instrument: PANTHER System. Connectivity: Automated. Software: Embedded firmware for instrument control and signal processing. Sterilization: Not applicable (reagents).

Indications for Use

Indicated for symptomatic and asymptomatic individuals, including males and females, for the qualitative detection and differentiation of Chlamydia trachomatis and Neisseria gonorrhoeae rRNA in clinician-collected endocervical, vaginal, and male urethral swabs, PreservCyt liquid Pap specimens, patient-collected vaginal swabs, and male urine.

Regulatory Classification

Identification

Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identify Neisseria spp. from cultured isolates. Additionally, some of these reagents consist of Neisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence of Neisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Neisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K132251 ## 510(k) SUMMARY ## APTIMA Combo 2® Assay (on the PANTHER® System) Attached is a 510(k) summary as described in 21 CFR 807.92 ## Sponsor Information ## Submitted By: | Name: | | |----------|--| | Address: | | Hologic | Gen-Probe Incorporated 10210 Genetic Center Drive San Diego, CA 92121 (858) 410-8000 **OCT** 17 **2013** Company Contact: Contact: Phone: Fax: Email: Jody J. Fleming Regulatory Affairs Manager 858-410-8634 858-410-7876 jody.fleming@gen-probe.com Date Prepared: July 18, 2012 General Information | Trade Name: | APTIMA Combo 2® Assay | |-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common or Usual Name: | Ribosomal RNA (rRNA) target-amplified nucleic acid probe<br>test for the <i>in vitro</i> diagnostic detection of <i>Chlamydia trachomatis</i> and/or <i>Neisseria gonorrhoeae</i> | | Classification Names: | DNA Probe, Nucleic Acid Amplification, Chlamydia<br>DNA Reagents, Neisseria | {1}------------------------------------------------ ## APTIMA Combo 2® Assay Device Description DNA Probe, Nucleic Acid Amplification, Chlamydia Medical Specialty Microbiology Product Code MKZ Device Class 1 Regulation number 866.3120 Device Description DNA Reagents, Neisseria Medical Specialty Microbiology Product Code LSL Device Class 2 Regulation number 866.3390 ## Substantially Equivalent Device: APTIMA Combo 20 Assay (PANTHER® System); K111409 This premarket application is to clear the APTIMA Combo 2 Assay for use on the PANTHER System with the male urine specimen type. #### Device Description The APTIMA Combo 2 Assay combines the technologies of target capture, transcriptionmediated amplification (TMA), and dual kinetic assay (DKA). Specimens are collected and transferred into their respective specimen transport tubes. The transport solutions in these tubes release the rRNA targets and protect them from degradation during storage. When the APTIMA Combo 2 Assay is performed in the laboratory, the target rRNA molecules are isolated from specimens by use of capture oligomers via target capture that utilizes magnetic microparticles. The capture oligomers contain sequences complementary to specific regions of the target molecules as well as a string of deox yadenosine residues. A separate capture oligomer is used for each target. During the hybridization step, the sequence specific regions of the capture oligomers bind to specific regions of the target molecules. The capture oligomer:target complex is then captured out of solution by decreasing the temperature of the reaction to room temperature. This temperature reduction allows hybridization to occur between the deoxyadenosine region on the capture oligomer and the poly-deoxythymidine Page 2 of 35 {2}------------------------------------------------ molecules that are covalently attached to the magnetic particles. The microparticles, including the captured target molecules bound to them, are pulled to the side of the reaction vessel using magnets and the supernatant is aspirated. The particles are washed to remove residual specimen matrix that may contain amplification reaction inhibitors. After the target capture steps are completed, the specimens are ready for amplification. Target amplification assays are based on the ability of complementary oligonucleotide primers to specifically anneal and allow enzymatic amplification of the target nucleic acid strands. The APTIMA Combo 2 Assay replicates a specific region of the 23S rRNA from CT and a specific region of the 16S rRNA from GC via DNA intermediates. A unique set of primers is used for each target molecule. Detection of the rRNA amplification product sequences (amplicon) is achieved using nucleic acid hybridization. Single-stranded chemiluminescent DNA probes, which are complementary to a region of each target amplicon, are labeled with different acridinium ester molecules. The labeled DNA probes combine with amplicon to form stable RNA:DNA hybrids. The Selection Reagent differentiates hybridized from unhybridized probe, eliminating the generation of signal from unhybridized probe. During the detection step, light emitted from the labeled RNA:DNA hybrids is measured as photon signals in a luminometer, and are reported as Relative Light Units (RLU). In DKA, differences in the kinetic profiles of the CT and GC labeled probes allow for the differentiation of signal; kinetic profiles are derived from measurements of photon output during the detection read time. The chemiluminescent detection for CT signal has very rapid kinetics and has the "flasher" kinetic type. The chemiluminescent detection reaction for GC signal is relatively slower and has the "glower" kinetic type. Assay results are determined by a cut-off based on the total RLU and the kinetic curve type. {3}------------------------------------------------ #### Intended Use The APTIMA Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal urogenital disease using the PANTHER System as specified. On the PANTHER System, the assay may be used to test the following specimens from symptomatic and asymptomatic individuals: clinician-collected endocervical, vaginal and male urethral swab specimens, clinician-collected gynecological specimens collected in the PreservCyt Solution, patient-collected vaginal swab specimens, and male urine specimens. Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use. {4}------------------------------------------------ ## Comparison to Predicate A comparison of the APTIMA Combo 2® Assay (PANTHER® System) with the addition of the male urine specimen type to the predicate APTIMA Combo 2® Assay (PANTHER® System) (K111409) is summarized below. | Item | Predicate Device | Test Device<br>(with Male Urine Claim) | |------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K111409 | K132251 | | Trade Name | APTIMA Combo 2®<br>Assay | APTIMA Combo 2® Assay | | Instrument | PANTHER® System | PANTHER® System | | Model Number | 303094 and 302923 | 303094 and 302923 | | Device Class | II | II | | Regulation Specialty | Microbiology | Microbiology | | Qualitative /Quantitative<br>Assay | Qualitative | Qualitative | | Function | Detection and differentiation of<br>rRNA from Chlamydia<br>trachomatis and Neisseria<br>gonorrhoeae | Same | | Indications For Use /<br>Intended Use | The APTIMA Combo 2 Assay is<br>a target amplification nucleic acid<br>probe test that utilizes target<br>capture for the in vitro qualitative<br>detection and differentiation of<br>ribosomal RNA (rRNA) from<br>Chlamydia trachomatis (CT)<br>and/or Neisseria gonorrhoeae<br>(GC) to aid in the diagnosis of<br>chlamydial and/or gonococcal<br>urogenital disease using the<br>PANTHER System as specified.<br>On the PANTHER System, the<br>assay may be used to test the<br>following specimens from<br>symptomatic and asymptomatic<br>individuals: clinician-collected<br>endocervical, vaginal and male<br>urethral swab specimens. | The APTIMA Combo 2 Assay is<br>a target amplification nucleic acid<br>probe test that utilizes target<br>capture for the in vitro qualitative<br>detection and differentiation of<br>ribosomal RNA (rRNA) from<br>Chlamydia trachomatis (CT)<br>and/or Neisseria gonorrhoeae<br>(GC) to aid in the diagnosis of<br>chlamydial and/or gonococcal<br>urogenital disease using the<br>PANTHER System as specified.<br>On the PANTHER System, the<br>assay may be used to test the<br>following specimens from<br>symptomatic and asymptomatic<br>individuals: clinician-collected<br>endocervical, vaginal and male<br>urethral swab specimens. | | Item | Predicate Device | Test Device<br>(with Male Urine Claim) | | Indications For Use /<br>Intended Use<br>(Continued) | clinician-collected gynecological<br>specimens collected in the<br>PreservCyt Solution, patient-<br>collected vaginal swab<br>specimens.1<br><br>Patient-collected vaginal swab<br>specimens are an option for<br>screening women when a pelvic<br>exam is not otherwise indicated.<br>The vaginal swab specimen<br>collection kit is not for home use. | clinician-collected gynecological<br>specimens collected in the<br>PreservCyt Solution, patient-<br>collected vaginal swab<br>specimens1, and male urine<br>specimens.<br><br>Patient-collected vaginal swab<br>specimens are an option for<br>screening women when a pelvic<br>exam is not otherwise indicated.<br>The vaginal swab specimen<br>collection kit is not for home use. | | Specimen Types | Female specimens:<br>• Vaginal swab<br>• Endocervical swab<br>• ThinPrep.in PreservCyt<br>solution | Female specimens:<br>• Vaginal swab<br>• Endocervical swab<br>• ThinPrep in PreservCyt<br>solution | | | Male Specimens:<br>• Urethral Swab | Male Specimens:<br>• Urethral Swab<br>• Urine | | | Swabs:<br>After collection, transport and<br>store swab in transport tube at 2-<br>30°C and test within 60 days. If<br>longer storage is desired, freeze at<br>-20°C to -70°C for up to 365<br>days. | Swabs:<br>Same | | Specimen<br>Transport/Storage | ThinPrep Liquid Pap in<br>PreservCyt:<br>Transport and store in PreservCyt<br>solution at 2-30°C for up to 30<br>days. After transfer to APTIMA<br>specimen transfer tube, store at<br>15-30°C for 14 days or store at<br>2-8°C for 30 days. If longer<br>storage is desired, freeze at -20°C<br>to -70°C for up to 365 days | ThinPrep Liquid Pap in<br>PreservCyt:<br>Same | | Item | Predicate Device | Test Device<br>(with Male Urine Claim) | | Specimen<br>Transport/Storage<br>(Continued) | | Urine:<br>After collection, transport the<br>processed urine specimens in the<br>APTIMA urine specimen<br>transport tube at $2°C$ to $30°C$ and<br>store at $2°C$ to $30°C$ until tested.<br>Processed urine specimens should<br>be assayed with the APTIMA<br>Combo 2 Assay within 30 days<br>of collection. If longer storage is<br>needed, freeze at $-20°C$ to $-70°C$<br>for up to 12 months after<br>collection. | | Type of Assay | Nucleic Acid Amplification Test | Same | | Technology* | Target Capture (TC),<br>Transcription-Mediated<br>Amplification (TMA),<br>Hybridization Protection Assay<br>(HPA) | Same | | Detection Format | HPA which provides relative<br>light units (RLUs) that are<br>assessed against an established<br>assay cutoff | Same | {5}------------------------------------------------ {6}------------------------------------------------ *A number of Gen-Probe APTIMA assays (that are based on TC, TMA, and HPA technologies) have been cleared by FDA including: the APTIMA Combo 2 Assay (K060652), the APTIMA CT A (K061413) and the APTIMA GC Assay (K061509) for use on the automated TIGRIS System and the semiautomated DTS Systems. {7}------------------------------------------------ ### Performance Data #### Brief Description of Non-Clinical Data The following analytical studies were conducted to support clearance of the male urine claim for the APTIMA Combo 2 Assay on the PANTHER System. ### Analytical Sensitivity Study Chlamydia trachomatis analytical sensitivity (limit of detection) was determined by testing dilutions of CT organisms in the APTIMA Combo 2 Assay. The analytical sensitivity claim for the assay is 1 IFU/assay (7.25 IFU/swab, 9.75 IFU/mL PreservCyt Solution liquid Pap, 5.0 IFU/mL urine). However, dilutions of less than 1 IFU/assay tested positive in the APTIMA Combo 2 Assay for the following 12 CT serovars: D, E, F, G, H, I, J, K, L1, L2, L2a and L3 (≥95% positivity was observed in samples containing CT concentrations of 1.89 IFU/mL). Neisseria gonorrhoeae analytical sensitivity (limit of detection) was determined by testing dilutions of GC organisms in the APTIMA Combo 2 Assay. The analytical sensitivity claim for the assay is 50 cells/assay (362 cells/swab, 488 cells/mL PreservCyt Solution liquid Pap, 250 cells/mL urine). However, dilutions of less than 50 cell/assay tested positive in the APTIMA Combo 2 Assay for 30 different strains of GC (≥95% positivity was observed in samples containing GC concentrations of 0.36 CFU/mL). {8}------------------------------------------------ #### Carryover Studies for the Panther System Two studies were conducted to evaluate carryover on the PANTHER System. In the first study, carryover was assessed in multiple runs on three PANTHER Systems with approximately 20% high titer GC samples dispersed between negative samples. The runs included clusters of high positive samples with clusters of negative samples as well as single high positives dispersed within the run. High titer samples were made using GC rRNA spiked into STM to give a final concentration equivalent to 2.5 x 105 CFU/mL. Five runs were performed on each of three PANTHER Systems. Carryover was calculated from a total of 2938 valid negative results. The overall carryover rate from this study was 0% with a 95% confidence interval of 0-0.1%. The second carryover study was conducted on one PANTHER System with high titer GC positive samples (GC rRNA spiked into STM at the equivalent of 2.5 x 10° CFU/mL) alternately processed with negative samples in a checkerboard format. Five checkerboard runs were performed. The overall carryover rate from this study was 0.74% (1/135 negative samples). #### Freeze-Thaw Study Gen-Probe conducted an in-house analytical study that compared the APTIMA Combo 2 Assay (PANTHER System) performance of fresh (non-frozen) versus frozen urines. These data support the use of prospectively-collected frozen specimens to establish the male urine specimen performance claims. Performance of sixty individual urine specimens was assessed after exposure to fresh (non-frozen) and frozen storage conditions. There were no significant differences in sensitivity and specificity between fresh and frozen samples (upper and lower bound of the 95% CI for the difference in sensitivity and specificity was <10%). {9}------------------------------------------------ ## Brief Description of Clinical Data ## Prevalence The prevalence of CT and GC in patient populations depends on risk factors such as age, gender, the presence or absence of symptoms, the type of clinic, and the sensitivity of the test used to detect infections. A summary of the prevalence of three CT and GC disease outcomes, as determined by the APTIMA Combo 2 Assay on the PANTHER System, is shown in Tables 1 and 2 for two multi-center clinical studies by clinical site and overall. {10}------------------------------------------------ able 1: Clinical Study 1. Prevalence of CT and GC Infections as Determined by the APTIMA Combo 2 Assay in Mala Frethral Swab, Vaginal Swab, PreservCyt Solution Liquid Pap, an | Site | MS | | | CVS/PVS | | | PCyt | | | FS | | | |-------------------------------------------------------|------------------|-----------------|-----------------|-------------------|------------------|------------------|------------------|------------------|------------------|------------------|------------------|------------------| | | CT+/GC- | CT-/GC+ | CT+/GC+ | CT+/GC- | CT-/GC+ | CT+/GC+ | CT+/GC- | CT-/GC+ | CT+/GC+ | CT+/GC- | CT-/GC+ | CT+/GC+ | | Prevalence % (# positive/# tested with valid results) | | | | | | | | | | | | | | | 0<br>(-) | 0<br>(-) | 0<br>(-) | 9.9<br>(21/212) | 3.3<br>(7/212) | 3.8<br>(8/212) | 8.9<br>(20/225) | 2.7<br>(6/225) | 3.1<br>(7/225) | 10.4<br>(20/193) | 3.1<br>(6/193) | 3.6<br>(7/193) | | | 13.9<br>(28/202) | 5.9<br>(12/202) | 3.0<br>(6/202) | 8.3<br>(19/230) | 3.9<br>(9/230) | 1.3<br>(3/230) | 8.8<br>(21/239) | 4.6<br>(11/239) | 0.8<br>(2/239) | 8.2<br>(19/231) | 4.8<br>(11/231) | 0.9<br>(2/231) | | | 1.3<br>(1/76) | 1.3 (1/76) | 0.0 (0/76) | 2.7<br>(6/222) | 0.5<br>(1/222) | 0.0<br>(0/222) | 3.1<br>(7/226) | 0.4<br>(1/226) | 0.0<br>(0/226) | 2.7<br>(6/223) | 0.4<br>(1/223) | 0.0<br>(0/223) | | | 24.4<br>(33/135) | 1.5<br>(2/135) | 4.4<br>(6/135) | 11.7<br>(40/342) | 1.5<br>(5/342) | 1.2<br>(4/342) | 10.2<br>(35/342) | 1.5<br>(5/342) | 0.9<br>(3/342) | 11.3<br>(38/337) | 1.8<br>(6/337) | 0.9<br>(3/337) | | | 0<br>(-) | 0<br>(-) | 0 | 4.5<br>(1/22) | 0.0<br>(0/22) | 0.0<br>(0/22) | 4.8<br>(1/21) | 0.0<br>(0/21) | 0.0<br>(0/21) | 4.3<br>(1/23) | 0.0<br>(0/23) | 0.0<br>(0/23) | | | 21.5<br>(28/130) | 5.4<br>(7/130) | 0.8<br>(1/130) | 11.9<br>(13/109) | 3.7<br>(4/109) | 0.9<br>(1/109) | 8.7<br>(10/115) | 1.7<br>(2/115) | 0.9<br>(1/115) | 8.8<br>(10/114) | 1.8<br>(2/114) | 0.9<br>(1/114) | | | 16.7<br>(1/6) | 0.0<br>(0/6) | 0.0<br>(0/6) | 3.2<br>(5/157) | 2.5<br>(4/157) | 0.6<br>(1/157) | 2.5<br>(4/161) | 2.5<br>(4/161) | 0.6<br>(1/161) | 2.6<br>(4/152) | 2.6<br>(4/152) | 0.7<br>(1/152) | | | 16.6<br>(91/549) | 4.0<br>(22/549) | 2.4<br>(13/549) | 8.1<br>(105/1294) | 2.3<br>(30/1294) | 1.3<br>(17/1294) | 7.4<br>(98/1329) | 2.2<br>(29/1329) | 1.1<br>(14/1329) | 7.7<br>(98/1273) | 2.4<br>(30/1273) | 1.1<br>(14/1273) | Pap, PVS = patient-collected vaginal swal Page 11 of 35 {11}------------------------------------------------ Table 2: Clinical Study 1 and Clinical Study 2. Prevalence of CT and GC Infections as Determined by the APTIMA Combo 2 Assay in Male Urine Samples by Clinical Site | Site | Prevalence % (# positive/# tested with valid results) | | | |------|-------------------------------------------------------|------------------|------------------| | | CT+/GC- | CT-/GC+ | CT+/GC+ | | 1 | 6.0<br>(6/100) | 0.0<br>(0/100) | 0.0<br>(0/100) | | 2 | 3.0<br>(2/67) | 3.0<br>(2/67) | 0.0<br>(0/67) | | 3 | 0.0<br>(0/109) | 0.9<br>(1/109) | 0.0<br>(0/109) | | 4 | 13.0<br>(13/100) | 3.0<br>(3/100) | 1.0<br>(1/100) | | 5 | 13.6<br>(17/125) | 5.6<br>(7/125) | 0.0<br>(0/125) | | 6 | 15.1<br>(43/284) | 7.0<br>(20/284) | 2.1<br>(6/284) | | 7 | 1.4<br>(3/212) | 0.9<br>(2/212) | 0.0<br>(0/212) | | 8 | 1.3<br>(1/75) | 0.0<br>(0/75) | 0.0<br>(0/75) | | 9 | 16.7<br>(42/251) | 5.2<br>(13/251) | 3.2<br>(8/251) | | 10 | 20.5<br>(17/83) | 1.2<br>(1/83) | 0.0<br>(0/83) | | 11 | 4.1<br>(6/146) | 0.7<br>(1/146) | 0.7<br>(1/146) | | 12 | 14.3<br>(16/112) | 4.5<br>(5/112) | 2.7<br>(3/112) | | 13 | 8.9<br>(10/112) | 2.7<br>(3/112) | 2.7<br>(3/112) | | 14 | 7.7<br>(2/26) | 0.0<br>(0/26) | 0.0<br>(0/26) | | All | 9.9<br>(178/1802) | 3.2<br>(58/1802) | 1.2<br>(22/1802) | Note: CT and GC prevalence was estimated using symptomatic male urine samples from Clinical Study 2 and asymptomatic male urine samples from both studies. . . {12}------------------------------------------------ Positive and Negative Predictive Values for Hypothetical Prevalence Rates The estimated positive and negative predictive values (PPV and NPV) of the APTIMA Combo 2 Assay for different hypothetical prevalence rates are shown for each specimentype in Table 3. For each specimen type, the PPV and NPV are derived for different hypothetical prevalence rates using the sensitivity and specificity estimates from the two multi-center clinical studies (see Table 4 and Table 7). | | Hypothetical<br>Prevalence (%) | | CT Detection | | GC Detection | |-------------------------------------------------------------------------|--------------------------------|---------|--------------|---------|--------------| | Specimen Type | | PPV (%) | NPV (%) | PPV (%) | NPV (%) | | Clinician-Collected<br>Vaginal Swab/ Patient-<br>Collected Vaginal Swab | 1 | 38.9 | 100 | 70.6 | 100 | | | 2 | 56.3 | 99.9 | 82.9 | 100 | | | 5 | 76.8 | 99.9 | 92.6 | 99.9 | | | 10 | 87.5 | 99.7 | 96.3 | 99.7 | | | 15 | 91.7 | 99.5 | 97.7 | 99.6 | | | 20 | 94.0 | 99.3 | 98.3 | 99.4 | | | 25 | 95.5 | 99.1 | 98.8 | 99.2 | | PreservCyt Solution<br>Liquid Pap | 1 | 100 | 100 | 100 | 100 | | | 2 | 100 | 100 | 100 | 100 | | | 5 | 100 | 99.9 | 100 | 100 | | | 10 | 100 | 99.8 | 100 | 100 | | | 15 | 100 | 99.7 | 100 | 100 | | | 20 | 100 | 99.6 | 100 | 100 | | | 25 | 100 | 99.4 | 100 | 100 | | Female Endocervical<br>Swab | 1 | 58.5 | 100 | 85.8 | 100 | | | 2 | 74.0 | 99.9 | 92.4 | 100 | | | 5 | 88.0 | 99.9 | 96.9 | 100 | | | 10 | 93.9 | 99.7 | 98.5 | 100 | | | 15 | 96.1 | 99.5 | 99.1 | 100 | | | 20 | 97.2 | 99.3 | 99.3 | 100 | | | 25 | 97.9 | 99.1 | 99.5 | 100 | | Male Urethral Swab | 1 | 53.1 | 100 | 100 | 100 | | | 2 | 69.6 | 100 | 100 | 100 | | | 5 | 85.5 | 100 | 100 | 100 | | | 10 | 92.6 | 100 | 100 | 100 | | | 15 | 95.2 | 100 | 100 | 100 | | | 20 | 96.6 | 100 | 100 | 100 | | | 25 | 97.4 | 100 | 100 | 100 | | Male Urine | 1 | 83.6 | 100 | 77.4 | 100 | | | 2 | 91.2 | 99.9 | 87.4 | 100 | | | 5 | 96.4 | 99.7 | 94.7 | 99.9 | | | 10 | 98.2 | 99.5 | 97.4 | 99.9 | | | 15 | 98.9 | 99.2 | 98.4 | 99.8 | | | 20 | 99.2 | 98.8 | 98.8 | 99.7 | | | 25 | 99.4 | 98.4 | 99.1 | 99.6 | ## Table 3: Positive and Negative Predictive Values for Hypothetical Prevalence Rates by Specimen Type Note: APTIMA Combo 2 Assay performance was estimated using vaginal swab, PreservCyt Solution Liguid Pap, female endocervical swab, and mate urethral swab sample results from Clinical Study I , symptomatic male urine samples from Clinical Study 2, and asymptomatic male urine samples from both studies. {13}------------------------------------------------ #### Clinical Study Results Two clinical studies were performed. APTIMA Combo 2 Assay clinical performance was estimated with male urethral swab, vaginal swab, PreservCyt Solution liquid Pap, and endocervical swab specimens in Clinical Study 1, and with male urine specimens in Clinical Study 2. ## Clinical Study 1: Vaginal Swab, PreservCvt Solution Liquid Pap, Female Endocervical Swab, and Male Urethral Swab Specimen Clinical Study 4 A prospective, multicenter clinical study was conducted to establish the performance characteristics of the APTIMA Combo 2 Assay on the PANTHER System. Specimens were collected from symptomatic and asymptomatic men (n=580) and women (n=1332) enrolled from 7 geographically and ethnically diverse US clinical sites, including obstetrics and gynecology, family planning, public health, and STD clinics. Subjects were classified as symptomatic if symptoms were reported by the subject. Subjects were classified as asymptomatic if the subject did not report symptoms. Of the 580 male subjects, none were <18 years of age, 72 were 18 to 20 years of age, 201 were 21 to 25 years of age, and 307 were >25 years of age. Of the 1332 female subjects, 11 were 14 to 15 years of age, 59 were 16 to 17 years of age, 319 were 18 to 20 years of age, 401 were 21 to 25 years of age, and 542 were >25 years of age. Up to 2 specimens were collected from each male subject (1 urethral swab and 1 first-catch urine, in that order) and up to 4 specimens were collected from each female subject (1 first-catch urine, 1 vaginal swab, 1 PreservCyt Solution liquid Pap specimen, and 1 endocervical swab, in that order). All specimens were clinician-collected except urine specimens and approximately half of the vaginal swab specimens, which were collected by the subject at the clinic. Approximately half of the PreservCyt Solution liquid Pap specimens were collected with a broom-type device and half were collected with a spatula and cytobrush. Samples were prepared for APTIMA testing in accordance with the appropriate APTIMA specimen collection kit package insert instructions. All evaluable samples (567 male urethral swab, 580 male urine, 1319 vaginal swab, 1330 PreservCyt Solution liquid Pap, and 1310 endocervical swab samples) were tested with the APTIMA Combo 2 Assay on the PANTHER System in accordance with package insert <sup>े</sup> This study included testing of male urine samples with the APTIMA Combo 2 Assay on the PANTHER System that were not included in the original performance results due to the low prevalence of GC in the study population. Page 14 of 35 {14}------------------------------------------------ instructions. The samples were split amongst three laboratories (two external laboratories and in-house). Samples with initial invalid, equivocal, or error results were retested. Eighteen (18) male urethral swab, 25 vaginal swab, 1 PreservCyt Solution liquid Pap. and 37 endocervical swab samples had final invalid results and were excluded from the analyses. Most of the invalid results were due to insufficient sample volume. One vaginal swab and 1 endocervical swab had final CT equivocal results and 1 PreservCyt Solution liquid Pap sample and 1 endocervical swab had final GC equivocal results and were excluded from the analyses. Male urethral swab, male and female urine, and PreservCyt Solution liquid Pap samples were tested with cleared nucleic acid amplification tests (NAATs) to establish the infected status. The infected status algorithm used results from two specimen types and two reference NAATs. Subjects were categorized as infected if a positive result occurred in each of the two reference NAATs (see Tables 10, 11, 13, and 14 for the infected status algorithms). For female subjects, if the positive NAAT results occurred only in the urine specimens and not in the PreservCyt Solution liquid Pap specimens, the subject was categorized as infected; however, for the evaluation of the non-urine specimen types, the specimens were considered non-infected. Subjects that could not be categorized as infected or not infected were excluded from the performance analyses, In addition, male urine samples tested with the APTIMA Combo 2 Assay on the PANTHER System were excluded from the performance analyses due to the low prevalence of GC in the study population, particularly in the asymptomatic subjects. #### Clinical Study 2: Male Urine Specimen Clinical Study A prospective, multicenter clinical study was conducted to establish the performance characteristics of the APTIMA Combo 2 Assay on the PANTHER System in male urine specimens. Specimens were collected from symptomatic and asymptomatic men (n=1492) enrolled from 13 geographically and ethnically diverse US clinical research sites, and family planning, public health, men's health, and STD clinics. Subjects were classified as symptomatic if symptoms were reported by the subject. Subjects were classified as asymptomatic if the subject did not report symptoms. Of the 1492 subjects enrolled, 14 were withdrawn. Two specimens were collected from each subject (1 urethral swab and 1 first-catch urine, in that order). The urethral swab specimens were clinician-collected, and urine specimens were collected by the subject at the clinic. Urine specimens from each subject were processed into Page 15 of 35 {15}------------------------------------------------ multiple samples for CT/GC testing with different NAATs in accordance with the instructions in the appropriate specimen collection kit package insert. The male urine samples for APTIMA Combo 2 Assay testing on the PANTHER System were split among three external laboratories. All 1478 male urine samples from non-withdrawn subjects were tested with the APTIMA Combo 2 Assay on the PANTHER System in accordance with the APTIMA Combo 2 Assay package insert instructions. Samples with initial invalid, equivocal, or error results were retested. One male urine sample had a final invalid result and was excluded from the analyses. The invalid result was due to insufficient sample volume. Of the remaining 1477 evaluable male subjects, 46 were 16 to 17 years of age, 155 were 18 to 20 years of age, 524 were 21 to 30 years of age, 279 were 31 to 40 years of age, and 473 were >40 years of age. Male urethral swab and urine samples were tested with cleared NAATs to establish the infected status (see Tables 13 and 15 for the infected status algorithms). The infected status algorithm used urethral swab and urine sample results from one reference CT and GC NAAT and urine sample results from two additional reference CT and GC NAATs to generate four reference results for each analyte. Subjects were categorized as infected if a positive result occurred in at least two of the reference NAATs. Subjects that could not be categorized as infected or not infected were excluded from the performance analyses: 1 subject had an indeterminate CT infected status and was excluded from the performance analyses for detection of CT. #### Chlamydia trachomatis Performance Results Performance characteristics of the APTIMA Combo 2 Assay for CT detection were estimated for each specimen type and are displayed in Tables 4 and 5 combining data from the two clinical studies. Table 6 displays the clinical performance results by individual study. Performance was calculated by comparing PANTHER System results to an infected status algorithm, which differed between the two clinical studies (see Tables 10 through 12 for the CT infected status algorithms). Table 4 shows the sensitivity, specificity, positive value (PPV), and negative predictive value (NPV) of the APTIMA Combo 2 Assay for CT detection and the prevalence of CT (based on the infected status) in each specimen type. {16}------------------------------------------------ | Specimen<br>Type¹ | n | TP | FP | TN | FN | Prev % | Sensitivity %<br>(95% CI)² | Specificity %<br>(95% CI)² | PPV %<br>(95% CI)³ | NPV %<br>(95% CI)³ | |-------------------|------|-----|----|------|----|--------|----------------------------|----------------------------|--------------------|--------------------| | CVS/<br>PVS | 1274 | 104 | 18 | 1149 | 3 | 8.4 | 97.2 (92.1-99.0) | 98.5 (97.6-99.0) | 85.2 (78.8-90.5) | 99.7 (99.3-99.9) | | PCyt | 1311 | 112 | 0 | 1197 | 2 | 8.7 | 98.2 (93.8-99.5) | 100 (99.7-100) | 100 (96.9-100) | 99.8 (99.4-100) | | FS | 1254 | 104 | 8 | 1139 | 3 | 8.5 | 97.2 (92.1-99.0) | 99.3 (98.6-99.6) | 92.9 (87.1-96.7) | 99.7 (99.3-99.9) | | MS | 549 | 100 | 4 | 445 | 0 | 18.2 | 100 (96.3-100) | 99.1 (97.7-99.7) | 96.2 (90.8-98.9) | 100 (99.2-100) | | MU | 1799 | 197 | 3 | 1589 | 10 | 11.5 | 95.2 (91.3-97.4) | 99.8 (99.4-99.9) | 98.5 (95.8-99.7) | 99.4 (98.9-99.7) | 1 'able 4: Performance Characteristics of the APTIMA Combo 2 Assay for CT Detection Cl = onfidence increal, CVS = clinical swah, N = false positive, FF = false positive, FS = paintent = wh, NS = macurial swab, YV = nake collected valia, "N = neagaine, PV = n = true positive. l at architect and, Precedes (1 ligol by, and crime sent see least and Clinical Sudy , symomte mate net semple realls are fon Clinical Study. 2 and architection and essels o PPV 95% CI computed from the exact 95% CI for the positive likelihood ratio, NPV 95% CI computed from the exact 95% CI from the negative likelihood ratio. {17}------------------------------------------------ Table 5 shows the sensitivity, specificity, PPV, and NPV of the APTIMA Combo 2 Assay for CT detection and the prevalence of CT (based on the infected status) in each specimen type by symptom status. CT prevalence was higher in symptomatic men and women. Page 18 of 35 {18}------------------------------------------------ able 5: Performance Characteristics of the APTIMA Combo 2 Assay for CT Detection by Symptom Status | Specimen-<br>Type¹ | Symptom<br>Status | n | TP | FP | TN | FN | Prev<br>% | Sensitivity %<br>(95% CI)² | Specificity %<br>(95% CI)² | PPV %<br>(95% CI)³ | NPV %<br>(95% CI)³ | |--------------------|-------------------|------|-----|----|------|----|-----------|----------------------------|----------------------------|--------------------|--------------------| | CVS/<br>PVS | Sym | 810 | 73 | 8 | 729 | 0 | 9.0 | 100 (95.0-100) | 98.9 (97.9-99.4) | 90.1 (82.3-95.5) | 100 (99.5-100) | | | Asym | 464 | 31 | 10 | 420 | 3 | 7.3 | 91.2 (77.0-97.0) | 97.7 (95.8-98.7) | 75.6 (63.1-86.2) | 99.3 (98.1-99.8) | | PCyt | Sym | 838 | 76 | 0 | 762 | 0 | 9.1 | 100 (95.2-100) | 100 (99.5-100) | 100 (95.4-100) | 100 (99.5-100) | | | Asym | 473 | 36 | 0 | 435 | 2 | 8.0 | 94.7 (82.7-98.5) | 100 (99.1-100) | 100 (91.1-100) | 99.5 (98.5-99.9) | | FS | Sym | 794 | 71 | 5 | 718 | 0 | 8.9 | 100 (94.9-100) | 99.3 (98.4-99.7) | 93.4 (85.9-97.8) | 100 (99.5-100) | | | Asym | 460 | 33 | 3 | 421 | 3 | 7.8 | 91.7 (78.2-97.1) | 99.3 (97.9-99.8) | 91.7 (79.9-98.0) | 99.3 (98.1-99.8) | | MS | Sym | 238 | 59 | 1 | 178 | 0 | 24.8 | 100 (93.9-100) | 99.4 (96.9-99.9) | 98.3 (91.5-100) | 100 (98.0-100) | | | Asym | 311 | 41 | 3 | 267 | 0 | 13.2 | 100 (91.4-100) | 98.9 (96.8-99.6) | 93.2 (82.5-98.5) | 100 (98.7-100) | | MU | Sym | 497 | 85 | 1 | 406 | 5 | 18.1 | 94.4 (87.6-97.6) | 99.8 (98.6-100) | 98.8 (94.1-100) | 98.8 (97.3-99.6) | | | Asym | 1302 | 112 | 2 | 1183 | 5 | 9.0 | 95.7 (90.4-98.2) | 99.8 (99.4-100) | 98.2 (94.1-99.8) | 99.6 (99.1-99.9) | Aya manais, C = caliciencia subjective, PP = dispositive, PP = disposities N.S = mierceda sol, M.S = mac cuecha sol, M.S = mac cuecha val. A.V. = mierceded value, M.S. = pres Male urethral swab, vaginal swab, PreservCyt liquid Pap, and endocervical swab sample results are from Clinical Study 1, symptomatic male urine sample results are from Clinical Study 2, and asymptomatic male urine sample results are from both studies. Scored C! Score Cl PPV 95% Cl computed from the exact 95% Cl for the positive likelihood ratio, NPV 95% C1 computed from the regative likelihood ratio Page 19 of 35 {19}------------------------------------------------ Table 6: Clinical Performance of the AC2 Assay on PANTHER for Detection of Chiamydia Trachomatis in Male Urine, by Study and Symptom Status | Study | Symptom<br>Status | | TP | FP | TN | FN | Prev<br>%) | Sensitivity % (95% CI)' | Specificity %<br>(95% Cl) | (95% Cl)2<br>PPV % | (95% CI)²<br>NPV % | | |---------|-------------------|------|-----|----|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-----------------------|-------------------------|--| | Study | Asymr | 323 | 40 | | 280 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>2 | 13.0 | 95.2 (84.2 to 98. | 99.6 (98.0 to 99.5 | 97.6 (88.3 to 99.9) | 99.3 (97.6 to 99.9) | | | | | | | | | | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | | | | | Study 2 | Asymp | 979 | 72 | | Amazing American American American American American American American American American American American American American American American American American American Amer<br>903 | | 7.7<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 96.0 (88.9 to 98.6) | 99.9 (99.4 to 100.L | 98.6 (93.0 to 100.0) | 99.9<br>99.7 (99.1 to 1 | | | | Sympt | 497 | જરુ | | 406 | | 18. | 94.4 (87.6 to 97.6) | 99.8 (98.6 to 100.t | 98.8 (94.1 to 100.0) | 98.8 (97.3 to 99.6) | | | | ้โจเจ | 1476 | 157 | | 309 | | 11.2 | 95.2 (90.7 to 97.5) | 99.8 (99.4 to 100. | 99.8<br>98.7 (95.7 to | 99.4 (98,8 to 99.7) | | | | | | | | | | | | | | | | Symptomatic prevalence = incidence * Symptomatic duration Asympt = asymptomatic Score Cl. Score Cl. PPV 95% Cl computed from the exact 95% Cl for Pestive Likelihood Ratio, NPV 95% Cl computed from the exact 95% Cl for Negative L Page 20 of 35 {20}------------------------------------------------ #### Neisseria gonorrhoeae Performance Results Performance characteristics of the APTIMA Combo 2 Assay for GC detection were estimated for each specimen type and are displayed in Tables 7 and 8 combining data from the two clinical studies. Table 9 displays the clinical performance results by individual study. The infected status algorithm differed between the two clinical studies (see Tables 13 through 15 for the GC infected status algorithms). Table 7 shows the sensitivity, specificity, PPV, and NPV of the APTIMA Combo 2 Assay for GC detection and the prevalence of GC (based on the infected status) in each specimen type. {21}------------------------------------------------ | Specimen<br>Type¹ | n | TP | FP | TN | FN | Prev % | Sensitivity %<br>(95% CI)² | Specificity %<br>(95% CI)² | PPV %<br>(95% CI)³ | NPV %<br>(95% CI)³ | |-------------------|------|----|----|------|----|--------|----------------------------|----------------------------|--------------------|--------------------| | CVS/<br>PVS | 1258 | 42 | 5 | 1210 | 1 | 3.4 | 97.7 (87.9-99.6) | 99.6 (99.0-99.8) | 89.4 (78.6-96.1) | 99.9 (99.6-100) | | PCyt | 1293 | 43 | 0 | 1250 | 0 | 3.3 | 100 (91.8-100) | 100 (99.7-100) | 100 (92.1-100) | 100 (99.7-100) | | FS | 1238 | 42 | 2 | 1194 | 0 | 3.4 | 100 (91.6-100) | 99.8 (99.4-100) | 95.5 (85.4-99.4) | 100 (99.7-100) | | MS | 546 | 34 | 0 | 512 | 0 | 6.2 | 100 (89.8-100) | 100 (99.3-100) | 100 (90.2-100) | 100 (99.3-100) | | MU | 1797 | 75 | 5 | 1716 | 1 | 4.2 | 98.7 (92.9-99.8) | 99.7 (99.3-99.9) | 93.8 (86.7-97.8) | 99.9 (99.7-100) | able 7: Performance Characteristics of the APTIMA Combo 2 Assay for GC Detectio Cl = confieresal, CVS = clinicia-collected vab, FN = false nositive, FS = fende ndocemial swal, MC = male unter swal, MCF = mal unter = true positive Vaginal swo. Preces Cy il quid Pop. colocer cicles web sample results are from Clinical Study I, sympondic natures and see from Clicicle Sudy 2, ad espinsionale with collec PPV 95% Cl computed from the exact 95% CI for the positive likelihood ratio, NPV 95% CI computed from the exact 95% CI from the negative likelihood ratio. {22}------------------------------------------------ Table 8 shows the sensitivity, specificity, PPV, and NPV of the APTIMA Combo 2 Assay for GC detection and the prevalence of GC (based on the infected status) in each specimen type by symptom status. GC prevalence was higher in symptomatic men but similar in symptomatic and asymptomatic women. {23}------------------------------------------------ Table 8: Performance Characteristics of the APTIMA Combo 2 Assay for GC Detection by Symptom Status | Specimen Type | Symptom<br>Status | U | TP | FP | TN | FN | % សារ | Sensitivity % (95% CI)² | pecificity % (95% CI)² | PPV % 95% CI)3 | NPV% 95% CI)3 | |----------------|----------------------------|------|-----|----|-------|----|-------|---------------------------|-----------------------------------------------------------------------------------------------------------------------|------------------|-----------------| | CVS/ PVS | Sym | 802 | LT | ना | ﺎ ﻟﻠ | 0 | 3.4 | 00 (87.5-100 | 99.5 (98.7-99.8) | 87.1 (72.6-96.1) | 100 (99.6-100 | | | Asym | 4 રહ | SI | l | 439 | l | ૩.5 | 93.8 (71.7-98.9 | 99.8 (98.7-100 | 93.8 (74.0-99.8) | 99.8 (98.9-100) | | PCyt | Sym | 829 | LT | | 802 | 0 | 3.3 | 00 (87.5-100) | 100 (99.5-100) | 100 (88.0-100) | 100 (99.6-100 | | | Asym | 464 | ા ર | 0 | 448 | 0 | 3.4 | 00 (80.6-100) | 00 (99.1-100 | 100 (81.3-100) | 100 (99.3-100 | | FS | Sym | 785 | 97 | | 758 | 0 | g g | 00 (87.1-100) | 99.9 (99.3-100) | 96.3 (82.4-99.9 | 00 (99.5-100 | | | Asym | 453 | 91 | | 436 | 0 | 3.5 | 00 (80.6-100 | 99.8 (98.7-100 | 94.1 (74.3-99.8) | 100 (99.3-100) | | MS | Sym · | 236 | । દ | 0 | 205 | 0 | 13.1 | 100 (89.0-100) | 00 (98.2-100 | 100 (89.5-100) | 00 (98.3-100 | | | Asym | 310 | ದ | 0 | 307 | 0 | 1.0 | 00 (43.9-100) | 100 (98.8-100 | 100 (44.4-100 | 00 (99.3-100 | | MU | Sym | 497 | 99 | | 430 | 0 | 13.3 | 100 (94.5-100 | 99.8 (98.7-100 | 98.5 (92.3-100 | 100 (99.2-100 | | | Asym | 1300 | 6 | | 1 286 | | 0.8 | 90.0 (59.6-98.2 | 99.7 (99.2-99.9 | 69.2 (45.6-91.7) | 99.9 (99.7-100) | | | m = asymptomatic, Cl = cor | | | | | | | | idence interval, CVS = clinician-collected vaginal swab, FN = false positive, FS = female endocervical swab, MS = mal | | | All Propenia Creaming Cream Comments of Propential Proponies Proponio Productions Marchises Microsoft Marchiteria Marculares Microsoft Marculares (Marcular Processor Microsof Vaginal swab, PreservCyt liquid Pap, endocervical swab, male urethral swab sample results are from Clinical Study 1, symptomatic male urine sample results are from Clinical Study 2, and asymptomatic male urine sample results are from both studies. PPV 95% CI computed from the exact 95% CI for the positive likelihood ratio, NPV 95% CI computed from the exact 95% CI from the negative likelihood ratio. Page 24 of 35 {24}------------------------------------------------ Table 9. Clinical Performance of the AC2 Assay on PANTHER for Detection of Neisseria Gonorrhoeae in Male Urine, by Study an Symptom Status | Symptom<br>Status | n | TP | FP | TN | FN | Prev<br>(%) | Sensitivity %<br>(95% CI)¹ | Specificity %<br>(95% CI)¹ | PPV %<br>(95% CI)² | NPV %<br>(95% CI)² | |-------------------|------|----|----|------|----|-------------|----------------------------|----------------------------|-----…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...