ADVIA Centaur Toxoplasma M (Toxo M)

K142826 · Siemens Healthcare Diagnostics, Inc. · LGD · Dec 22, 2015 · Microbiology

Device Facts

Record IDK142826
Device NameADVIA Centaur Toxoplasma M (Toxo M)
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeLGD · Microbiology
Decision DateDec 22, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3780
Device ClassClass 2

Intended Use

The ADVIA Centaur Toxoplasma M (Toxo M) assay is an IgM antibody capture microparticle direct chemiluminometric in vitro diagnostic immunoassay intended for the qualitative detection of IgM antibodies to Toxoplasma gondii in serum or plasma (EDTA, heparin) using the ADVIA Centaur and ADVIA Centaur XP systems. The ADVIA Centaur Toxo M assay is used to measure IgM antibody against T. gondii which is presumptive of an acute, recent, or reactivated toxoplasma infection. Any measurement of IgM antibody to T. gondii must be performed in conjunction with the determination of IgG antibody to T. gondii.

Device Story

The ADVIA Centaur Toxo M assay is an automated in vitro diagnostic immunoassay for detecting IgM antibodies to Toxoplasma gondii. It uses an immunoglobulin class-capture sandwich principle on ADVIA Centaur and ADVIA Centaur XP systems. The system dispenses 10 µL of patient sample and 340 µL of paramagnetic particles coated with mouse anti-human IgM; after incubation, it adds a tracer reagent containing T. gondii antigen complexed with acridinium-ester-labeled anti-p30 monoclonal antibody. Following further incubation and wash steps, acid and base reagents are added to trigger a chemiluminescent reaction. The system measures relative light units (RLUs) to determine an Index Value, classifying samples as reactive, equivocal, or nonreactive. The assay is performed in clinical laboratories by trained personnel. Results assist clinicians in identifying acute or reactivated toxoplasmosis when interpreted alongside IgG antibody status.

Clinical Evidence

Bench testing only. Method comparison study of 1,111 specimens (985 patient, 126 panel) compared the modified assay to the predicate. Patient specimens showed 99.4% positive percent agreement (PPA) and 99.8% negative percent agreement (NPA). Panel specimens showed 98.1% PPA and 100.0% NPA. Precision met acceptance criteria (≤15% total CV). Cross-reactivity and endogenous interference studies showed no significant impact on assay performance.

Technological Characteristics

Immunoglobulin class-capture sandwich immunoassay; chemiluminometric detection. Components: paramagnetic particles coated with mouse anti-human IgM; Lite Reagent with T. gondii antigen and acridinium-labeled mouse anti-p30 F(ab)2 fragment. Buffer systems: Tricine (pH 8.0) with Tween-20, sm-BSA, and mouse IgG. Automated on ADVIA Centaur/XP systems. Standards: CLSI EP05-A2, CLSI EP07-A2, EN 13640, ANSI/AAMI/ISO 14971.

Indications for Use

Indicated for qualitative detection of IgM antibodies to Toxoplasma gondii in serum or plasma (EDTA, heparin) to aid in the presumptive diagnosis of acute, recent, or reactivated toxoplasma infection. Must be used in conjunction with IgG antibody testing. Not for blood, plasma, or tissue donor screening.

Regulatory Classification

Identification

Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized design of three human profiles facing right, arranged in a cascading manner. The profiles are black and have a simple, elegant design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profiles. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 22, 2015 SIEMENS HEALTHCARE DIAGNOSTICS, INC. MATTHEW GEE SENIOR MANAGER, REGULATORY AFFAIRS 511 BENEDICT AVENUE TARRYTOWN NY 10591 Re: K142826 Trade/Device Name: ADVIA Centaur Toxoplasma M (Toxo M) Regulation Number: 21 CFR 866.3780 Regulation Name: Toxoplasma gondii Serological Reagents Regulatory Class: II Product Code: LGD Dated: December 3, 2015 Received: December 4, 2015 Dear Mr. Gee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Ribhi Shawar -S For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K142826 Device Name ADVIA Centaur Toxoplasma M (Toxo M) ### Indications for Use (Describe) The ADVIA Centaur Toxoplasma M (Toxo M) assay is an IgM antibody capture microparticle direct chemiluminometric in vitro diagnostic immunoassay intended for the qualitative detection of IgM antibodies to Toxoplasma gondii in serum or plasma (EDTA, heparin) using the ADVIA Centaur and ADVIA Centaur XP systems. The ADVIA Centaur Toxo M assay is used to measure IgM antibody against T. gondii which is presumptive of an acute, recent, or reactivated toxoplasma infection. Any measurement of IgM antibody to T. gondii must be performed in conjunction with the determination of IgG antibody to T. gondii. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------|-----------------------------------------------------------| | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990. The assigned 510(k) Number is: K142826 #### 1. Date Prepared April 9, 2015 #### 2. Purpose for Submission The purpose of this submission is to describe changes to the ADVIA Centaur Toxoplasma M (Toxo M) assay (K010755). #### 3. Measurand Toxoplasma IgM antibodies #### 4. Type of Test Immunoglobulin class-capture chemiluminescence immunoassay #### Applicant Information 5. - Contact: Matthew Gee, M.Sc. Senior Manager, Regulatory Affairs - Address: Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591-5097 Phone: 914-524-2099 Fax: 914-524-3579 - Email: matthew.gee@siemens.com #### Proprietary and Established Names 6. ADVIA Centaur® Toxoplasma M (Toxo M) #### Regulatory Information 7. | Regulation Section: | 21CFR 866.3780; Toxoplasma gondii Serological Reagents | |-----------------------------|--------------------------------------------------------| | Classification: | Class II | | Product Code: | LGD | | Device Classification Name: | Enzyme linked immunoabsorbent assay, Toxoplasma gondii | | Review Panel: | Microbiology (83) | #### 8. Predicate Device Information | Name: | ADVIA Centaur Toxoplasma M (Toxo M) | |----------------|-------------------------------------| | Manufacturer: | Siemens Healthcare Diagnostics Inc. | | 510(K) Number: | K010755 | {4}------------------------------------------------ #### 9. Intended Use #### 9.1 Intended Use The ADVIA Centaur Toxoplasma M (Toxo M) assay is an IgM antibody capture microparticle direct chemiluminometric in vitro diagnostic immunoassay intended for the qualitative detection of IgM antibodies to Toxoplasma gondii in serum or plasma (EDTA, heparin) using the ADVIA Centaur and ADVIA Centaur XP systems. The ADVIA Centaur Toxo M assay is used to measure IgM antibody against T. gondii which is presumptive of an acute, recent, or reactivated toxoplasma infection. Any measurement of IgM antibody to T. gondii must be performed in conjunction with the determination of IgG antibody to T. gondii. #### 9.2 Indications for Use Same as Intended Use #### 9.3 Special Conditions for Use Statement(s) The detection of toxoplasma IgM in a given specimen, as determined by assays from different manufacturers, can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the toxoplasma IgM assay used. Values obtained with different assay methods cannot be used interchangeably. The reported IgM level cannot be correlated to an endpoint titer. This assay is not intended for use in screening blood, plasma, or tissue donors. The effectiveness of this assay for use in screening blood, plasma, or tissue donors has not been established. #### 9.4 Special Instrument Requirements ADVIA Centaur and ADVIA Centaur XP #### Device Description 10. The ADVIA Centaur Toxo M assay consists of the following: | Component | Volume | Ingredients | |--------------------------------------|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ADVIA Centaur Toxo M<br>Lite Reagent | 10.0 mL/pack | partially purified <i>T. gondii</i> antigen (~3 µg/mL)<br>complexed with a mouse anti- <i>T.gondii</i> p30<br>monoclonal antibody (F(ab)2 fragment) labeled with<br>acridinium ester in protein buffer with surfactant and<br>preservatives | | ADVIA Centaur Toxo M<br>Solid Phase | 17.0 mL/pack | mouse anti-human IgMµ monoclonal antibody (~24 µg/mL) covalently coupled to paramagnetic particles<br>in protein buffer with surfactant and preservatives | | ADVIA Centaur Toxo M<br>Calibrators | 600 µL/vial | defibrinated recalcified processed human plasma<br>positive for toxoplasma IgM antibodies with<br>preservatives | | ADVIA Centaur Toxo M<br>Controls | 1.5 mL/vial | defibrinated recalcified processed human plasma<br>negative and positive for toxoplasma IgM antibodies<br>with preservatives | Table 1. Summary of Ingredients of the ADVIA Centaur Toxo M Assay Components {5}------------------------------------------------ #### Purpose of the Submission 11. The purpose of this submission is to submit a modification to the ADVIA Centaur Toxo M assay. In addition to manufacturing process changes to improve operational effectiveness, and buffer changes to decrease the likelihood of non-specific binding, the main change to the ADVIA Centaur Toxo M assay is a truncation of the anti-p30 antibody in the Lite Reagent to remove the Fc portion of the molecule. This antibody is used to link the acridinium ester (chemiluminescent tag) to the toxoplasma p30 antigen. It is not involved in the binding of analyte from patient samples (i.e. not a component of reaction mechanism). #### Comparison of Predicate Device and Modified Device 12. Table 3 provides a list the similarities of the currently marketed predicate ADVIA Centaur Toxo M assay (with a whole IgG antibody in the Lite Reagent) and the modified ADVIA Centaur Toxo M assay with a F(ab), fragment in the Lite Reagent. Table 4 provides a list of differences between the predicate and modified devices. | Item | Predicate Device (K010755) | Modified Device | |-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------| | Intended Use | The ADVIA Centaur Toxoplasma M<br>(Toxo M) assay is an IgM antibody<br>capture microparticle direct<br>chemiluminometric in vitro diagnostic<br>immunoassay intended for the<br>qualitative detection of IgM<br>antibodies to Toxoplasma gondii in<br>serum or plasma (EDTA, heparin)<br>using the ADVIA Centaur and ADVIA<br>Centaur XP systems.<br>The ADVIA Centaur Toxo M assay is | Same | | | used to measure IgM antibody<br>against T. gondii which is<br>presumptive of an acute, recent, or<br>reactivated toxoplasma infection.<br>Any measurement of IgM antibody to<br>T. gondii must be performed in<br>conjunction with the determination of<br>IgG antibody to T. gondii. | | | Instrument Platforms | ADVIA Centaur<br>ADVIA Centaur XP | Same | | Methodology | Immunoglobulin class-capture<br>sandwich immunoassay using direct,<br>chemiluminometric technology | Same | | Capture Antibody<br>(Solid Phase) | Mouse anti-human IgMµ<br>monoclonal antibody | Same | | Tracer<br>(Lite Reagent) | Toxoplasma p30 antigen<br>bound to acridinium ester<br>(via mouse anti-T. gondii p30<br>monoclonal antibody) | Same | # Table 2. Similarities of Modified ADVIA Centaur Toxo M Assay and Predicate {6}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness | ltem | Predicate Device (K010755) | Modified Device | |--------------------------------|------------------------------------------------------------------------|-----------------| | Specimen Type | Serum or plasma (EDTA, heparin) | Same | | Sample Volume | 10 µL | Same | | Calibration | 2-point calibration using Toxo M Cal | Same | | Performance<br>Characteristics | Positive Percent Agreement: 99.2%<br>Negative Percent Agreement: 99.2% | Same | # Table 2. Similarities of Modified ADVIA Centaur Toxo M Assay and Predicate # Table 3. Differences Between Modified ADVIA Centaur Toxo M Assay and Predicate | Item | Predicate Device (K010755) | Modified Device | |--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| | Toxoplasma IgM Source<br>(Calibrators, Controls) | Toxoplasma IgM positive<br>human plasma pools | Cell culture supernatant of<br>human anti-toxoplasma IgM<br>monoclonal antibody-producing<br>cells | | Particle Resuspension | Particle resuspension with<br>water | Particle resuspension with Wash<br>1 (phosphate buffered saline) | | Lite Reagent Conjugate* | Ab format = Whole IgG<br>Ab Concentration = 30 ng/mL<br>Conjugate Loading Ratio = 30:1 | Ab format = F(ab)2 fragment<br>Ab Concentration = 12.5 ng/mL<br>Conjugate Loading Ratio = 18:1 | | Solid Phase Buffer | Buffer: Tris (pH =8.0)<br>NaCl: 150 mM<br>Surfactant: CHAPS = 0.1 g/L<br>Blocker: Gelatin = 22.2 g/L<br>Mouse IgG: 25 mg/L<br>EDTA: none | Buffer: Tricine (pH = 8.0)<br>NaCl: 300 mM<br>Surfactant: Tween-20 = 2.2 g/L<br>Blocker: sm-BSA = 10.0 g/L<br>Mouse IgG: 100 mg/L<br>EDTA: 0.7 g/L | | Claimed Measuring<br>Range^ | 0.10–40.00 Index | 0.10–10.00 Index | * The modified antibody in the Lite Reagent recognizes same epitope as the antibody in the cleared device. It is simply used to attach the acridinium ester (chemiluminescent tag) to the toxoplasma p30 antigen. It is not part of the analyte-binding reaction mechanism. #### 13. Standard/Guidance Document References The following recognized standards from Clinical Laboratory Standards Institute (CLSI) were used as a basis of the study procedures described in this submission: - Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Second Edition (CLSI EP05-A2, 2004; Recognition No. 7-110) - l Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition (CLSI EP07-A2, 2005; Recognition No. 7-127) - l Stability testing of in vitro diagnostic reagents (European Committee for Standardization EN 13640:2002; Recognition No. 7-84) - Medical devices - Application of risk management to medical devices (ANSI/AAMI/ISO 14971:2007/(R)2010; Recognition No. 5-70) {7}------------------------------------------------ #### 14. Test Principle The ADVIA Centaur Toxo M assay is an immunoqlobulin class-capture sandwich immunoassay using direct, chemiluminometric technology. The anti-human IgM monoclonal antibody is covalently coupled to paramagnetic particles in the Solid Phase. In the Lite Reagent, the T. gondii antigen is complexed with an anti-p30 monoclonal antibody (F(ab)> fragment) labeled with acridinium ester. Antibody-antigen complexes will form if toxoplasma IgM is present in the sample. ADVIA Centaur systems automatically perform the following steps for the Toxo M assay: - 트 Dispenses 10 µL of sample into a cuvette. - I Dispenses 340 µL of Solid Phase and incubates the mixture for 18 minutes at 37°C. - . Separates the Solid Phase from the mixture and aspirates the unbound reagent. - I Washes the cuvette with Wash 1. - I Dispenses 200 µL Lite Reagent and incubates the mixture for 18 minutes at 37°C. - I Separates the Solid Phase from the mixture and aspirates the unbound reagent. - I Washes the cuvette with Wash 1. - . Dispenses 300 µL each of Acid Reagent and Base Reagent to initiate the chemiluminescent reaction. A direct relationship exists between the amount of toxoplasma IgM activity present in the patient sample and the amount of relative light units (RLUs) detected by the system. A result of reactive (positive) or nonreactive (negative) is determined using an Index Value. #### Performance Characteristics 15. Analytical performance (precision, interference with endogenous substances and potentially interfering agents) and testing with panel samples and patient samples was evaluated using the modified ADVIA Centaur Toxo M assay. The results were comparable to those established for the previous version of the device (currently-marketed predicate). #### 16. Conclusions The results of performance testing and verification activities demonstrate that the design modifications to the ADVIA Centaur Toxo M assay do not impact its safety or effectiveness and do not alter its performance claims or alter its intended use, as described in the labeling. Based on the results of comparative testing, the modified ADVIA Centaur Toxo M assay is substantially equivalent in principle and performance to the currently-marketed predicate device, ADVIA Centaur Toxo M, cleared under 510(k) K010755.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...