IMMULITE TOXOPLASMA IGM, MODEL LKTM1 AND LKTM2, IMMULITE 2000 TOXOPLASMA IGM, MODEL L2KTM2 AND L2KTM6

K012075 · Diagnostic Products Corp. · LGD · Jan 10, 2002 · Microbiology

Device Facts

Record IDK012075
Device NameIMMULITE TOXOPLASMA IGM, MODEL LKTM1 AND LKTM2, IMMULITE 2000 TOXOPLASMA IGM, MODEL L2KTM2 AND L2KTM6
ApplicantDiagnostic Products Corp.
Product CodeLGD · Microbiology
Decision DateJan 10, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3780
Device ClassClass 2

Indications for Use

IMMULITE® Toxoplasma IgM - For in vitro diagnostic use with the IMMULITE® Analyzers - for the presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactive Toxoplasma gondii infection. This product has not been cleared/approved by the FDA for blood/plasma donor screening. IMMULITE® 2000 Toxoplasma IgM - For in vitro diagnostic use with the IMMULITE® 2000 Analyzers - for the presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® 2000 Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactive Toxoplasma gondii infection. This product has not been cleared/approved by the FDA for blood/plasma donor screening.

Device Story

IMMULITE and IMMULITE 2000 Toxoplasma IgM are automated, solid-phase, two-step chemiluminescent enzyme immunoassays. The device uses a polystyrene bead coated with partially purified Toxoplasma gondii antigen. Patient serum or plasma is incubated with the bead; Toxoplasma-specific IgM binds to the antigen. After a centrifugal wash, an alkaline phosphatase-labeled anti-human IgM antibody is added. A second incubation and wash follow. A phosphate ester of adamantyl dioxetane substrate is added; hydrolysis by alkaline phosphatase produces light. A luminometer measures photon output, which is proportional to the amount of bound IgM. Results are determined by comparing patient signal to a cutoff (adjustor) material. Used in clinical laboratories by technicians to aid physicians in diagnosing acute or recent Toxoplasma infection, particularly in pregnant women. The output provides a qualitative result that informs clinical decision-making regarding infection status.

Clinical Evidence

Clinical studies conducted at multiple sites in the U.S. compared IMMULITE/IMMULITE 2000 Toxoplasma IgM against predicate ELISAs. Studies included pregnant women and symptomatic/asymptomatic subjects. For IMMULITE, agreement with Kit A was 97.4% (all subjects) and 97.6% (pregnant); agreement with Kit B was 95.5% (all) and 95.2% (pregnant). For IMMULITE 2000, agreement with Kit A was 98.7% (all) and 98.8% (pregnant). CDC panel testing showed 97.9% and 99.0% total agreement for IMMULITE and IMMULITE 2000, respectively. Precision studies (within-run and total CVs) and interference testing (bilirubin, lipemia, hemoglobin) were performed.

Technological Characteristics

Solid-phase, two-step chemiluminescent enzyme immunoassay. Solid phase: polystyrene bead coated with partially purified Toxoplasma gondii antigen. Energy source: automated luminometer for photon detection. Substrate: phosphate ester of adamantyl dioxetane. Analyzers: IMMULITE and IMMULITE 2000. Connectivity: automated system. Sterilization: not specified. Software: embedded firmware for instrument control and signal processing.

Indications for Use

Indicated for presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. Used as an aid in diagnosis of acute, recent, or reactivated Toxoplasma infection when performed with Toxoplasma IgG assay. Not for blood/plasma donor screening.

Regulatory Classification

Identification

Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 1 0 2002 # 012075 510(k) Summary of Safety and Effectiveness IMMULITE and IMMULITE 2000 Toxoplasma IgM This summary of safety and effectiveness information has been prepared in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92. | Name:<br>Address: | Diagnostic Products Corporation<br>5700 West 96th Street<br>Los Angeles, California 90045-5597 | |-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Telephone Number:<br>Facsimile Number: | (310) 645-8200<br>(310) 645-9999 | | Contact Person: | Edward M. Levine, Ph.D.<br>Director of Clinical Affairs | | Date of Preparation: | January 8, 2002 | | Device Name:<br>Trade: | IMMULITE® Toxoplasma IgM<br>IMMULITE® 2000 Toxoplasma IgM | | Catalog Number: | LKTM1 (100 tests), LKTM2 (200 tests)<br>L2KTM2 (200 tests), L2KTM6 (600 tests) | | Common: | Reagent system for the detection of IgM antibodies to<br>Toxoplasma gondii in human serum | | Classification: | Class II device, 83-LGD (21CFR 866.3780) | | Manufacturer: | Diagnostic Products Corporation<br>5700 West 96th Street<br>Los Angeles, CA 90045<br>(The Quality System of Diagnostic Products Corporation is<br>registered to ISO 9001:1994) | | Establishment Registration<br>Number | DPC's Registration Number 2017183 | | Substantially Equivalent<br>Predicate Device: | Vidas (ELFA) Toxo IgM (K923166)<br>Zeus Scientific Toxo IgM ELISA Test System (K913787) | | Description of Devices: | IMMULITE and IMMULITE 2000 Toxoplasma IgM are<br>clinical devices for use with their respective IMMULITE<br>and IMMULITE 2000 Automated Immunoassay Analyzers. | {1}------------------------------------------------ ### Intended Use of the Device: IMMULITE® Toxoplasma IgM is a solid-phase chemiluminescent enzyme immunoassay for in vitro diagnostic use with the IMMULITE® Analyzer - for the presumptive qualitative detection of IgM antibodies to Toxoplasma in human serum or plasma (EDTA or heparinized), particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactivated Toxoplasma infection. IMMULITE® 2000 Toxoplasma IgM is a solid-phase chemiluminescent enzyme immunoassay for in vitro diagnostic use with the IMMULITE® 2000 Analyzer – for the presumptive qualitative detection of IgM antibodies to Toxoplasma in human serum or plasma (EDTA or heparinized), particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® 2000 Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactivated Toxoplasma infection. ### Performance Equivalence: Diagnostic Products Corporation (DPC) asserts that IMMULITE Toxoplasma IgM and IMMULITE 2000 Toxoplasma IgM produce substantially equivalent results to other commercially marketed assays, such Vidas (ELFA) Toxo IgM or Zeus Scientific Toxo IgM ELISA Test System. ### Technology Comparison: Provided below is a comparison of DPC's IMMULITE and IMMULITE 2000 Toxoplasma IgM technology vs. the Zeus Scientific Toxo IgM ELISA Test System and Vidas Toxo IgM technology. IMMULITE Toxoplasma IgM is a solid-phase, two-step chemiluminescent enzyme immunoassay. The solid phase, a polystyrene bead enclosed within an IMMULITE Test Unit, is coated with partially purified Toxoplasma gondii antigen. Prediluted patient sample (1-in-21 dilution) and a protein-based buffer are simultaneously introduced into the Test Unit, and incubated for approximately 30 minutes at 37℃ with intermittent agitation. During this time. Toxoplasma -specific IgM in the sample binds to the Toxoplasma antigen-coated bead. Unbound serum is then removed by a centrifugal wash. An alkaline phosphatase-labeled anti-human IgM antibody is introduced, and the Test Unit is incubated for approximately another 30-minute cycle. The unbound enzyme conjugate is removed by a centrifugal wash. Substrate is then added, and the Test Unit is incubated for a further 10 minutes. The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to vield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving {2}------------------------------------------------ precision by providing a window for multiple readings. The bound complex -- and thus also the photon output, as measured by the luminometer -- is related to the presence of Toxoplasma InM in the sample. A result is then obtained by comparing the patient result to the response of the cutoff material (i.e., the adjustor). IMMULITE 2000 Toxoplasma IgM is a solid-phase, two-step chemiluminescent enzyme immunoassay. The solid phase, a polystyrene bead added to an IMMULITE 2000 Reaction Tube, is coated with partially purified Toxoplasma gondii antigen. Prediluted patient sample (1-in-20 dilution) and a protein-based buffer are simultaneously introduced into the Reaction Tube, and incubated for approximately 30 minutes at 37℃ with intermittent agitation. During this time. Toxoplasma -specific IgM in the sample binds to the Toxoplasma antigen-coated bead. Unbound serum is then removed by a centrifyigal wash. An alkaline phosphatase-labeled anti-human IgM antibody is introduced, and the Reaction Tube is incubated for approximately another 30-minute cycle. The unbound enzyme conjugate is removed by a centrifugal wash. Substrate is then added, and the Reaction Tube is incubated for a further 5 minutes. The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to yield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex -- and thus also the photon output, as measured by the luminometer -- is related to the presence of Toxoplasma IgM in the sample. A result is then obtained by comparing the patient result to the response of the cutoff material (i.e., the adjustor) Zeus Scientific Toxo IgM ELISA Test System is an enzyme-linked immunosorbent assay (ELISA) designed to detect IgM class antibodies to T. gondii antigen. Patient serum samples to be assayed for antibody are first diluted and incubated in microwells coated with T. gondii antigen. Any antigen specific IgM antibody in the sample will bind to the immobilized antigen. The plate is washed to remove unbound antibody and other serum components. Peroxidase conjugated goat anti-human IgM is added to the wells and the plate is incubated. The conjugate will react with IgM antibody immobilized on the solid phase in step 2. The wells are washed to remove unreacted conjugate. The microtiter wells containing immobilized peroxidase conjugate are incubated with peroxidase substrate solution. Hydrolysis of the substrate by peroxidase produces a color change. After a period of time, the reaction is stopped and the color intensity of the solution is measured photometrically. The color intensity of the solution is directly related to the antibody concentration in the test sample. Vidas Toxo IgM is an enzyme-linked fluorescent immunoassay (ELFA) performed in an automated Vidas instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR) serves as the solid phase as well as the pipettor for the assay. The SPR is coated with goat anti-u chain {3}------------------------------------------------ antibodies. The Vidas TXM assay configuration prevents nonspecific reactions with the SPR. Reagents for the assay are in the sealed TXM Reagent Strips. After a sample dilution step, the sample is cycled in and out of the SPR for a specific length of time. IgM antibodies present in the specimen will bind to the anti-u chain antibodies coating the interior of the SPR. Unbound sample components are washed away. An Immunocomplex of T. gondii antigen and mouse monoclonal anti-P30 antibodies conjugated with alkaline phosphatase is cycled in and out of the SPR and will attach to the human anti-T. gondii IgM bound to the SPR wall. A final wash step removes unbound conjugate. A fluorescent substrate is introduced into the SPR. Enzyme remaining on the wall of the SPR will catalyze the conversion of the substrate to the fluorescent product. The intensity of the fluorescence is measured by the optical scanner in the instrument. #### Expected Values Individuals infected with the Toxoplasma organism will typically exhibit detectable levels of IgM antibody immediately before or soon after the onset of symptoms.2 IgM titers normally decline within four to six months, but may persist at low levels up to a year. 4 Patients with active toxoplasma chorioretinitis usually have undetectable levels of IgM. The prevalence of Toxoplasma infection can vary depending on a number of factors such as age, gender, geographical location, socio-economic status, race, type of test used, specimen collection and handling procedures, and clinical and epidemiological history of individual patients. There are approximately 3000 cases of congenital toxoplasmosis reported per year, with an average of 0.6 cases per 1000 pregnancies in the United States. #### IMMULITE Toxoplasma IgM Studies with presumed healthy, asymptomatic subjects and individuals suspected of acute toxoplasma viral infection were conducted at two clinical sites in the southern and northeastern United States. The study in the southern United States consisted of 172 specimens from 93 pregnant women and 79 individuals with various conditions. IMMULITE Toxoplasma IgM tests on these samples vielded the following results: | Subjects | Total<br>n | Positive | | Negative | | Indeterminate | | |----------|------------|----------|-----|----------|-----|---------------|----| | | | n | % | n | % | n | % | | Pregnant | 93 | 22 | 24% | 64 | 69% | 7 | 8% | | Various | 79 | 14 | 18% | 62 | 78% | 3 | 4% | | All | 172 | 36 | 21% | 126 | 73% | 10 | 6% | {4}------------------------------------------------ Image /page/4/Figure/0 description: The image is a bar graph showing frequency on the y-axis and ratio on the x-axis. The frequency ranges from 0 to 70, while the ratio ranges from 0.0 to 5.0. The bar graph shows that the frequency is highest at a ratio of 0.0, with a frequency of around 60, and then decreases as the ratio increases. ## Observed signal/cutoff ratios for all samples The study in the northeastern United States consisted of 168 specimens from 91 pregnant women and 77 individuals with various conditions. IMMULITE Toxoplasma IgM tests on these samples yielded the following results: | Subjects | Total<br>n | Positive | | Negative | | Indeterminate | | |----------|------------|----------|-----|----------|-----|---------------|----| | | | n | % | n | % | n | % | | Pregnant | 91 | 20 | 22% | 65 | 71% | 6 | 7% | | Various | 77 | 14 | 18% | 60 | 78% | 3 | 4% | | All | 168 | 34 | 20% | 125 | 74% | 9 | 5% | Observed signal/cutoff ratios for all samples Image /page/4/Figure/5 description: This image is a bar graph that shows frequency on the y-axis and ratio on the x-axis. The x-axis ranges from 0.0 to 5.0, and the y-axis ranges from 0 to 90. The bar at ratio 0.0 has a frequency of approximately 84, while the bar at ratio 1.0 has a frequency of approximately 21. {5}------------------------------------------------ ### IMMULITE 2000 Toxoplasma IgM ININGEFFE 2000 renthy, asymptomatic subjects and individuals suspected of acute toxoplasma viral infection were conducted at one clinical sites in the southern United States. The study consisted of 172 specimens from 93 pregnant women and 79 individuals with various study consisted or 172 proplasma IgM tests on these samples yielded the following results: | Subjects | Total<br>n | Positive | | Negative | | Indeterminate | | |----------|------------|----------|-----|----------|-----|---------------|-----| | | | n | % | n | % | n | % | | Pregnant | 93 | 23 | 25% | 61 | 66% | 9 | 10% | | Various | 79 | 16 | 20% | 59 | 75% | 4 | 5% | | All | 172 | 39 | 23% | 120 | 70% | 13 | 8% | 80 70 60 ૨૦ requenc 40 30 20 10 0 5.0 0.0 1.0 #### Observed signal/cutoff ratios for all samples #### Performance Characteristics #### Clinical Performance In a clinical study in the southern United States, a total of 172 frozen samples from apparently healthy male and female subjects, pregnant women and patients suspected of being toxoplasma IgM positive were tested by IMMULITE Toxoplasma IgM and by a commercially available enzyme immunoassay (Kit A- Zeus Scientific Toxo IgM ELISA Test System). These samples were also tested for Toxoplasma IgG. The IMMULITE Toxoplasma IgM results were compared to the results of Kit A. Comparison for all subjects | Kit A | Positive | Indeterm | Negative | |----------------------------------------------------------|----------|----------|----------| | Positive | 36 | 9 | 4 | | Indeterm | 0 | 1 | 6 | | Negative | 0 | 0 | 116 | | Positive Agreement: 90.0% (36/40, 95% CI: 76.3% - 97.2%) | | | | | Negative Agreement: 100% (116/116, 95% CI: 96.9% - 100%) | | | | | Agreement: 97.4% (152/156, 95% CI: 93.6% - 99.3%) | | | | {6}------------------------------------------------ Comparison for pregnant subjects | Kit A | Pos | Ind | Neg | |----------------------------------------------------------|-----|-----|-----| | Pos | 22 | 7 | 2 | | Ind | 0 | 0 | 4 | | Neg | 0 | 0 | 58 | | Positive Agreement: 91.7% (22/24, 95% CI: 73.0% - 99.0%) | | | | | Negative Agreement: 100% (58/58, 95% CI: 93.8% - 100%) | | | | | Agreement: 97.6% (80/82, 95% CI: 91.5% - 99.7%) | | | | In another clinical study in the northeastern United States, a total of 168 frozen samples from apparently healthy male and female subjects, pregnant women and patients suspected of being toxoplasma IgM positive were tested by IMMULITE Toxoplasma IgM and another commercially available enzyme immunoassay (Kit B - Vidas Toxo IgM). These samples were also tested for Toxoplasma IgG. The IMMULITE Toxoplasma IgM results were compared to the results of Kit B. Comparison for all subjects | Kit B | Positive | Indeterm | Negative | |------------------------------------------------------------|----------|----------|----------| | Positive | 28 | 4 | 1 | | Indeterm | 0 | 0 | 2 | | Negative | 6 | 5 | 122 | | Positive Agreement: 96.6% (28/29, 95% CI: 82.2% - 99.9%) | | | | | Negative Agreement: 95.3% (122/128, 95% CI: 90.1% - 98.3%) | | | | | Agreement: 95.5% (150/157, 95% CI: 91.0% - 98.2%) | | | | Comparison for pregnant subjects | | Pos | Ind | Neg | |----------------------------------------------------------|-----|-----|-----| | Pos | 16 | 4 | 0 | | Ind | 0 | 0 | 1 | | Neg | 4 | 2 | 64 | | Positive Agreement: 100% (16/16, 95% CI: 79.4% - 100%) | | | | | Negative Agreement: 94.1% (64/68, 95% CI: 85.6% - 98.4%) | | | | | Agreement: 95.2% (80/84, 95% CI: 88.3% - 98.7%) | | | | In a clinical study in the southern United States, a total of 172 frozen samples from normal male and female subjects, pregnant women and patients suspected of being toxoplasma IgM positive were tested by IMMULITE 2000 toxoplasma IgM, and by a commercially available enzyme immunoassays (Kit A- Zeus Scientific Toxo IgM ELISA Test System). These samples were also tested for Toxoplasma IgG. The IMMULITE 2000 Toxoplasma IgM results were compared to the results of Kit A. {7}------------------------------------------------ #### Comparison for all subjects | Kit A | Positive | Indeterm | Negative | |----------------------------------------------------------|----------|----------|----------| | Positive | 38 | 9 | 2 | | Indeterm | 1 | 3 | 3 | | Negative | 0 | 1 | 115 | | Positive Agreement: 95.0% (38/40, 95% CI: 83.1% - 99.4%) | | | | | Negative Agreement: 100% (115/115, 95% CI: 96.8% - 100%) | | | | | Agreement: 98.7% (153/155, 95% CI: 95.4% - 99.8%) | | | | Comparison for pregnant subjects: | | | IMMULITE 2000 Toxoplasma IgM | | | | |----------------------------------------------------------|-----|-------------------------------------|-----|--|--| | Kit A | Pos | Ind | Neg | | | | Pos | 23 | 7 | 1 | | | | Ind | 0 | 2 | 2 | | | | Neg | 0 | 0 | 58 | | | | Positive Agreement: 95.8% (23/24, 95% CI: 78.9% - 99.9%) | | | | | | | Negative Agreement: 100% (58/58, 95% CI: 93.8% - 100%) | | | | | | | Agreement: | | 98.8% (81/82, 95% CI: 93.4% - 100%) | | | | In a study at DPC, IMMULITE 2000 Toxoplasma IgM was compared to IMMULITE Toxoplasma IgM on 291 samples: | | Positive | Indeterm | Negative | |----------------------------------------------------------|----------|----------|----------| | Positive | 15 | 0 | 0 | | Indeterm | 3 | 0 | 0 | | Negative | 0 | 5 | 268 | | Positive Agreement: 100% (15/15, 95% CI: 78.2% - 100%) | | | | | Negative Agreement: 100% (268/268, 95% CI: 98.6% - 100%) | | | | | Agreement: 100% (283/283, 95% CI: 98.7% - 100%) | | | | Indeterminate results were excluded from calculations. #### Performance Data Precision (Serum): Precision studies for IMMULITE Toxoplasma IgM assay were conducted at three different sites: in-house at DPC (Site 1) and at two sites in the southern and northeasten United States (Sites 2 and 3). At Site 1, samples were assayed in duplicate over the course of 20 days, two runs per day, for a total of 40 runs and 80 replicates. (See "Site 1" table). At Sites 2 and 3, samples were assayed in triplicate over the course of 5 days, one run per day, for a total of 5 runs and 15 replicates. (See "Site 2" and "Site 3" tables). The means, within-run and total CVs were calculated by the Analysis of Variance. Results are expressed as a signal-to-cutoff ratio. Precision statistics are summarized below. {8}------------------------------------------------ | | | Within-Run | | Total | | |---|-------|------------|------|-------|-------| | | Mean | SD | CV | SD | CV | | 1 | 3.03 | 0.241 | 8.0% | 0.337 | 11.1% | | 2 | 1.42 | 0.110 | 7.8% | 0.155 | 10.9% | | 3 | 0.928 | 0.055 | 5.9% | 0.089 | 9.5% | | 4 | 0.579 | 0.033 | 5.7% | 0.055 | 9.4% | | 5 | 0.25* | | | | | ### IMMULITE Toxoplasma IgM Precision - Serum (ratio): Site 1 * Consistently at a very low ratio #### IMMULITE Toxoplasma IgM Precision - Serum (ratio): Site 2 | | Mean | Within-Run | | Total | | |---|--------|------------|------|-------|------| | | | SD | CV | SD | CV | | 1 | 3.30 | 0.184 | 5.6% | 0.203 | 6.2% | | 2 | 1.56 | 0.113 | 7.2% | 0.110 | 7.1% | | 3 | 0.980 | 0.068 | 6.9% | 0.063 | 6.4% | | 4 | 0.650 | 0.032 | 4.9% | 0.030 | 4.6% | | 5 | 0.284* | | | | | * Consistently at a very low ratio #### IMMULITE Toxoplasma IgM Precision - Serum (ratio): Site 3 | | Within-Run | | | Total | | |---|------------|-------|------|-------|------| | | Mean | SD | CV | SD | CV | | 1 | 3.10 | 0.110 | 3.5% | 0.152 | 4.9% | | 2 | 1.38 | 0.099 | 7.2% | 0.097 | 7.0% | | 3 | 0.910 | 0.029 | 3.2% | 0.056 | 6.2% | | 4 | 0.600 | 0.022 | 3.7% | 0.030 | 5.0% | | 5 | 0.266* | - | - | - | - | * Consistently at a very low ratio Precision (Serum): Precision studies for IMMULITE 2000 Toxoplasma IgM assay were conducted at two different sites: in-house at DPC (Site 1) and in the southern United States (Site 2). At both sites, samples were assayed in triplicate over the course of 5 days, one run per day, for a total of 5 runs and 15 replicates. (See "Site 1" and "Site 2" tables). The means, within-run and total CVs were calculated by the Analysis of Variance. Results are expressed as a signal-tocutoff ratio. {9}------------------------------------------------ | | Mean | Within-Run | | Total | | |---|--------|------------|-------|-------|-------| | | | SD | CV | SD | CV | | 1 | 3.90 | 0.081 | 2.1% | 0.094 | 2.4% | | 2 | 2.07 | 0.150 | 7.2% | 0.149 | 7.2% | | 3 | 1.34 | 0.074 | 5.5% | 0.071 | 5.3% | | 4 | 0.810 | 0.092 | 11.4% | 0.082 | 10.1% | | 5 | 0.262* | – | – | – | – | ### IMMULITE 2000 Toxoplasma IgM Precision - Serum (ratio): Site 1 *Consistently at a very low ratio #### IMMULITE 2000 Toxoplasma IgM Precision - Serum (ratio): Site 2 | | | Within-Run | | | Total | | |---|--------|------------|------|--|-------|-------| | | Mean | SD | CV | | SD | CV | | 1 | 3.20 | 0.130 | 4.1% | | 0.138 | 4.3% | | 2 | 1.72 | 0.109 | 6.3% | | 0.113 | 6.6% | | 3 | 1.18 | 0.044 | 3.7% | | 0.078 | 6.6% | | 4 | 0.720 | 0.031 | 4.3% | | 0.104 | 14.4% | | 5 | 0.300* | | | | | | *Consistently at a very low ratio Precision (Plasma): Precision studies for IMMULITE Toxoplasma IgM and IMMULITE 2000 Toxoplasma IgM assays on plasma samples (EDTA and heparin) were conducted at DPC by testing samples in triplicate over the course of 3 days, two runs per day, for a total of 6 runs and 18 replicates. The means, within-run and total CVs were calculated by the Analysis of Variance. Results are expressed as a signal-to-cutoff ratio. Precision statistics are summarized below. ### IMMULITE Toxoplasma IgM Precision - EDTA (ratio): | | | Within-Run | | Total | | |---|-------|------------|------|-------|-------| | | Mean | SD | CV | SD | CV | | 1 | 0.262 | 0.023 | 8.8% | 0.033 | 12.6% | | 2 | 1.11 | 0.052 | 4.7% | 0.069 | 6.2% | | 3 | 1.33 | 0.075 | 5.6% | 0.077 | 5.8% | | 4 | 1.66 | 0.087 | 5.2% | 0.083 | 5.0% | {10}------------------------------------------------ | | Mean | Within-Run | | Total | | |---|-------|------------|------|--------|------| | | | SD | CV | SD | CV | | 1 | 0.259 | 0.0191 | 7.4% | 0.0183 | 7.1% | | 2 | 1.00 | 0.06 | 6.0% | 0.052 | 5.2% | | 3 | 1.21 | 0.097 | 8.0% | 0.084 | 6.9% | | 4 | 1.51 | 0.084 | 5.6% | 0.073 | 4.8% | ### IMMULITE 2000 Toxoplasma IgM Precision - EDTA (ratio): #### IMMULITE Toxoplasma IgM Precision - Heparin (ratio): | | Within-Run | | | Total | | |---|------------|-------|-------|-------|-------| | | Mean | SD | CV | SD | CV | | 1 | 0.67 | 0.077 | 11.5% | 0.081 | 12.1% | | 2 | 1.18 | 0.089 | 7.5% | 0.123 | 10.4% | | 3 | 1.42 | 0.112 | 7.9% | 0.136 | 9.6% | | 4 | 1.71 | 0.137 | 8.0 | 0.161 | 9.4% | #### IMMULITE 2000 Toxoplasma IgM Precision - Heparin (ratio): | | Mean | Within-Run | | | Total | | |---|------|------------|-------|-------|-------|----| | | | SD | CV | | SD | CV | | 1 | 0.65 | 0.068 | 10.5% | 0.07 | 10.8% | | | 2 | 1.2 | 0.061 | 5.1% | 0.057 | 4.8% | | | 3 | 1.32 | 0.073 | 5.5% | 0.071 | 5.4% | | | 4 | 1.54 | 0.084 | 5.5% | 0.078 | 5.1% | | Crossreactivity: Crossreactivity: A study was conducted to evaluate whether the measurement of Toxoplasma IgM antibody is affected by closely related microorganisms. Ninety-three seronegative sera containing antibodies to Varicella Zoster Virus (n=3), Measles (n=10), Cytomegalovirus (CMV) (n=10), Herpes Simplex Virus (n=10), Toxoplasma (n=10), mvcloplasma pneumoniae (n=10), Epstein-Barr Virus (n=10), Syphilis (n=10) and Parvovirus (n=10) and rheumatoid factor (n=10) were tested by IMMULITE and IMMULITE 2000 Toxoplasma IgM and all yielded negative results. #### Conjugated or unconjugated bilirubin: no effect up to 20 mg/dL Interference: Lipemia: no effect of triglycerides up to 3000 mg/dL Hemoglobin: no effect up to 539 mg/dL {11}------------------------------------------------ Anti-coagulants: Twenty-eight blood samples drawn into plain, heparinized and EDTA vacutainer tubes were assayed by the IMMULITE 2000 Toxoplasma IgM assay. Results (in S/CO ratio) from the anticoagulant tubes were compared with those from the serum tubes in regression analyses: | Regressions: | Heparin = $0.96 \times (Serum) + 0.05$<br>EDTA = $0.93 \times (Serum) + 0.04$ | r = 0.991<br>r = 0.992 | |---------------------|-------------------------------------------------------------------------------|------------------------| | Means (S/CO ratio): | Serum = 0.80<br>Heparin = 0.82<br>EDTA= 0.79 | | Gel Barrier: Twenty-eight blood samples drawn into plain and SST vacutainer tubes were assayed by the IMMULITE 2000 Toxoplasma IgM assay. Results (in S/CO ratio) from the SST tubes were compared with those from the serum tubes in a regression analysis: | Regressions: $SST = 0.97 \times (Serum) + 0.08$ | $r = 0.994$ | |-------------------------------------------------|-------------| | Means (S/CO ratio): Serum = 0.80 | | | SST = 0.85 | | #### CDC Toxoplasma 1998 Human Serum Panel The following information is from a serum panel obtained from the CDC and tested by DPC. The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC. The panel consists of 32 positive and 65 negative samples. IMMULITE Toxoplasma IgM demonstrated 97.9% total agreement with the CDC results. There was 93.8% agreement with the positive specimens and 100% agreement with the negative specimens. IMMULITE 2000 Toxoplasma IgM demonstrated 99.0% total agreement with the CDC results. There was 96.9% agreement with the positive specimens and 100% agreement with the negative specimens. #### Conclusion: The data presented in this summary of safety and effectiveness is the data that the Food and Drug Administration used in granting DPC substantial equivalence for the IMMULITE Toxoplasma IgM and IMMULITE 2000 Toxoplasma IgM assays. {12}------------------------------------------------ Image /page/12/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wavy lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. The text is written in a sans-serif font and is evenly spaced around the circle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JAN 1 0 2002 Edward M. Levine, Ph.D. Director of Clinical Affairs Diagnostic Products Corporation 5700 West 96th Street Los Angeles, CA 90045-5597 Re: k012075 Trade/Device Name: IMMULITE® and IMMULITE® 2000 Toxoplasma IgM Regulation Number: 21 CFR 866.3780 Regulation Name: Toxoplasma gondii serological reagents Regulatory Class: Class II Product Code: LGD Dated: November 30, 2001 Received: December 3, 2001 #### Dear Dr. Levine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {13}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {14}------------------------------------------------ #### 510(k) Number: Kolao 75 IMMULITE® Toxoplasma IgM and Device Name: IMMULITE® 2000 Toxoplasma IgM ### Indications For Use: IMMULITE® Toxoplasma IgM - For in vitro diagnostic use with the IMMULITE® Analyzers - for the presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactive Toxoplasma gondii infection. This product has not been cleared/approved by the FDA for blood/plasma donor screening. IMMULITE® 2000 Toxoplasma IgM - For in vitro diagnostic use with the IMMULITE® 2000 Analyzers - for the presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® 2000 Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactive Toxoplasma gondii infection. This product has not been cleared/approved by the FDA for blood/plasma donor screening. ### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Woody Dubois (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K012075 \$\checkmark\$ **Prescription Use** **(Per 21 CFR 801.109)** OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...