INCSTAR CYTOMEGALOVIRUS IGM CAPTURE ELISA ASSAY MODEL 8530
Device Facts
| Record ID | K960434 |
|---|---|
| Device Name | INCSTAR CYTOMEGALOVIRUS IGM CAPTURE ELISA ASSAY MODEL 8530 |
| Applicant | Incstar Corp. |
| Product Code | LFZ · Microbiology |
| Decision Date | Oct 17, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.3175 |
| Device Class | Class 2 |
Intended Use
The INCSTAR CMV IgM Capture ELISA Kit contains instructions and materials for the qualitative and/or semi-quantitative detection of IgM antibodies to cytomegalovirus in human serum by reverse capture enzyme-linked immunosorbent assay (ELISA) technique. When performed according to instructions, the CMV IgM Capture ELISA test can be used as an aid in the diagnosis of current or recent active CMV infection. The evaluation of paired CMV IgM sera can also aid in determining the stage of active CMV infection.
Device Story
In vitro diagnostic ELISA kit; detects IgM antibodies to cytomegalovirus (CMV) in human serum. Principle: reverse capture ELISA; patient serum incubated in microtiter wells coated with mouse monoclonal anti-human IgM; patient IgM captured; anti-CMV monoclonal antibody conjugated to horseradish peroxidase binds to captured anti-CMV IgM; chromogen added for colorimetric detection. Absorbance measured at 450 nm; proportional to anti-CMV IgM concentration. Used in clinical laboratories by trained personnel. Results aid clinicians in diagnosing current/recent active CMV infection and staging infection via paired sera analysis.
Clinical Evidence
Clinical study of 529 serum samples from 474 individuals (healthy, immunocompromised, transplant, congenital CMV, other illnesses). Compared INCSTAR CMV IgM Capture ELISA to BioWhittaker CMV CAP-M ELISA. Results (95% CI): relative sensitivity 88-97%, relative specificity 92-97%, overall agreement 92-96%. Discrepant analysis performed using Gull CMV IgM ELISA (K903807).
Technological Characteristics
Reverse capture ELISA; microtiter well solid phase; mouse monoclonal anti-human IgM (μ chain specific) capture antibody; horseradish peroxidase-conjugated anti-CMV monoclonal antibody detection system; chromogenic substrate; 450 nm absorbance measurement.
Indications for Use
Indicated for the qualitative and/or semi-quantitative detection of IgM antibodies to cytomegalovirus in human serum to aid in the diagnosis of current or recent active CMV infection in patients, including healthy donors, immunocompromised hosts, transplant patients, and congenital CMV cases.
Regulatory Classification
Identification
Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.
Predicate Devices
- BioWhittaker CMV CAP-M ELISA test (K896295)
Reference Devices
- Gull CMV IgM ELISA (K903807)
Related Devices
- K955361 — INCSTAR CYTOMEGALOVIRUS IGG ELISA ASSAY · Incstar Corp. · Jul 22, 1996
- K955360 — INCSTAR CYTOMEGALOVIRUS IGG FAST ELISA ASSAY · Incstar Corp. · Jul 22, 1996
- K963289 — INCSTAR TOXOPLASMA IGM CAPTURE ELISA ASSAY · Incstar Corp. · May 1, 1997
- K982485 — SERAQUEST CMV IGM · Quest Intl., Inc. · Jan 13, 1999
- K142133 — ELECYS CMV IGM IMMUNOASSAY; ELECYS PRECICONTROL CMV IGM · Roche Diagnostics · Oct 24, 2014