Alinity i Rubella IgG

K243168 · Abbott Laboratories · LFX · Jun 20, 2025 · Microbiology

Device Facts

Record IDK243168
Device NameAlinity i Rubella IgG
ApplicantAbbott Laboratories
Product CodeLFX · Microbiology
Decision DateJun 20, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3510
Device ClassClass 2

Intended Use

The Alinity i Rubella IgG assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of IgG antibodies to rubella virus in human serum, serum separator, and plasma tubes (lithium heparin, lithium heparin separator, and tripotassium EDTA) on the Alinity i system. The Alinity i Rubella IgG assay is to be used as an aid in the determination of immune status to rubella in individuals including women of child-bearing age. The Alinity i Rubella IgG assay has not been cleared for use in screening blood, plasma, or tissue donors. The performance of this device has not been established for cord blood or neonatal samples. Likewise, performance has not been established for populations of immunocompromised or immunosuppressed individuals.

Device Story

Automated two-step chemiluminescent microparticle immunoassay (CMIA) for quantitative detection of anti-rubella IgG. Input: human serum or plasma samples. Process: sample, rubella virus-coated paramagnetic microparticles, and assay diluent incubated; anti-rubella IgG binds to microparticles; wash cycle; anti-human IgG acridinium-labeled conjugate added and incubated; wash cycle; Pre-Trigger and Trigger solutions added. Output: chemiluminescent signal (RLU) measured by system optics, directly proportional to anti-rubella IgG concentration. Used in clinical laboratories on the Alinity i system. Results aid clinicians in assessing rubella immune status, particularly for women of child-bearing age.

Clinical Evidence

Clinical performance evaluated against a composite comparator (3 FDA-cleared assays) using 1,433 US specimens and additional OUS samples. PPA and NPA calculated relative to medical decision point (10 IU/mL). Routine order (US) PPA 95.36%, NPA 97.62%. Pregnant females (US) PPA 95.24%, NPA 96.49%. CDC reference panel (n=100) showed 93.9% PPA and 100% NPA. Matrix equivalency confirmed for serum, serum separator, lithium heparin, lithium heparin separator, and K3 EDTA tubes.

Technological Characteristics

CMIA technology; paramagnetic microparticles coated with rubella virus antigen; acridinium-labeled anti-human IgG conjugate. Automated on Alinity i system. Analytical measuring interval: 1.0–500.0 IU/mL; extended interval: 500.0–1300.0 IU/mL. Reagents include TRIS buffer, surfactants, and preservatives (ProClin 950, sodium azide). Calibration storage: 30 days.

Indications for Use

Indicated for the quantitative determination of IgG antibodies to rubella virus in human serum and plasma to aid in the determination of immune status, including in women of child-bearing age. Not for use in screening blood, plasma, or tissue donors. Performance not established for cord blood, neonatal samples, or immunocompromised/immunosuppressed individuals.

Regulatory Classification

Identification

Rubella virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubella virus in serum. The identification aids in the diagnosis of rubella (German measles) or confirmation of a person's immune status from past infections or immunizations and provides epidemiological information on German measles. Newborns infected in the uterus with rubella virus may be born with multiple congenital defects (rubella syndrome).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) National Committee for Clinical Laboratory Standards': (i) 1/LA6 “Detection and Quantitation of Rubella IgG Antibody: Evaluation and Performance Criteria for Multiple Component Test Products, Speciment Handling, and Use of the Test Products in the Clinical Laboratory, October 1997,” (ii) 1/LA18 “Specifications for Immunological Testing for Infectious Diseases, December 1994,” (iii) D13 “Agglutination Characteristics, Methodology, Limitations, and Clinical Validation, October 1993,” (iv) EP5 “Evaluation of Precision Performance of Clinical Chemistry Devices, February 1999,” and (v) EP10 “Preliminary Evaluation of the Linearity of Quantitive Clinical Laboratory Methods, May 1998,” (2) Centers for Disease Control's: (i) Low Titer Rubella Standard, (ii) Reference Panel of Well Characterized Rubella Sera, and (3) World Health Organization's International Rubella Standard.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION June 20, 2025 Abbott Laboratories Laura Fraczek Regulatory Affairs Senior Specialist 100 Abbott Park Road Abbott Park, Illinois 60064 Re: K243168 Trade/Device Name: Alinity i Rubella IgG Regulation Number: 21 CFR 866.3510 Regulation Name: Rubella Virus Serological Reagents Regulatory Class: Class II Product Code: LFX Dated: May 22, 2025 Received: May 22, 2025 Dear Laura Fraczek: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243168 - Laura Fraczek Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243168 - Laura Fraczek Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JORGE L. MUNOZ -S Digitally signed by JORGE L. MUNOZ -S Date: 2025.06.20 13:41:14 -04'00' Jorge L. Munoz, Ph.D. Deputy Branch Chief Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K243168 | | | Device Name Alinity i Rubella IgG | | | Indications for Use (Describe) The Alinity i Rubella IgG assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of IgG antibodies to rubella virus in human serum, serum separator, and plasma tubes (lithium heparin, lithium heparin separator, and tripotassium EDTA) on the Alinity i system. The Alinity i Rubella IgG assay is to be used as an aid in the determination of immune status to rubella in individuals including women of child-bearing age. The Alinity i Rubella IgG assay has not been cleared for use in screening blood, plasma, or tissue donors. The performance of this device has not been established for cord blood or neonatal samples. Likewise, performance has not been established for populations of immunocompromised or immunosuppressed individuals. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Alinity i Rubella IgG K243168 510(k) Summary Page 1 of 18 510(k) Summary This summary of the 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. I. 510(k) Number K243168 II. Applicant Name Abbott Laboratories Department 09AA 100 Abbott Park Road Abbott Park, IL 60064 Primary contact person for all communications: Laura Fraczek, Regulatory Affairs Manager Abbott Diagnostics Division Telephone Number: (224) 668-8852 Fax Number: (224) 667-4836 Secondary contact person for all communications: Jacob Richards, Associate Director, Regulatory Affairs Abbott Diagnostic Division Telephone Number: (224) 668-5877 Fax Number: (224) 667-4836 Date summary prepared: June 19, 2025 {5} Alinity i Rubella IgG K243168 510(k) Summary Page 2 of 18 # III. Device Name Alinity i Rubella IgG ## Reagents Trade Name: Alinity i Rubella IgG Reagent Kit Device Classification: Class II Classification Name: Rubella virus serological reagents Governing Regulation: 21 CFR 866.3510 Code: LFX ## Calibrator Trade Name: Alinity i Rubella IgG Calibrators Device Classification: Class II Classification Name: Rubella virus serological reagents Governing Regulation: 21 CFR 866.3510 Code: LFX ## Controls Trade Name: Alinity i Rubella IgG Controls Device Classification: Class II Classification Name: Rubella virus serological reagents Governing Regulation: 21 CFR 866.3510 Code: LFX # IV. Predicate Device bioMérieux VIDAS RUB IgG assay (K080766) # V. Description of Device ## Reagents The kit configuration of the Alinity i Rubella IgG Reagent Kit is described below. | List Number (LN) | 04W0222 | | --- | --- | | Tests per cartridge | 100 | | Number of cartridges per kit | 2 | | Tests per kit | 200 | | Microparticles | 6.6 mL | | Conjugate | 6.1 mL | | Assay Diluent | 10.4 mL | {6} - Microparticles: Partially purified rubella virus coated microparticles in TRIS buffer with surfactant. Minimum concentration: 0.12% solids. Preservatives: ProClin 950 and sodium azide. - Conjugate: Anti-human IgG (mouse, monoclonal) acridinium-labeled conjugate in MES buffer with protein (bovine) stabilizer and surfactant. Minimum concentration: 16 ng/mL. Preservatives: antimicrobial agents. - Assay Diluent: TRIS buffer with protein (caprine, bovine, murine) stabilizers and surfactant. Preservatives: ProClin 300 and ProClin 950. ## Calibrators The Alinity i Rubella IgG Calibrators are described below. - Calibrator A: Contains recalcified human plasma with protein (ovine) stabilizer. - Calibrator B – F: Contain recalcified human plasma (reactive for IgG antibodies to rubella virus [anti-rubella IgG]) with protein (ovine) stabilizer. - Preservatives: ProClin 950 and sodium azide. The target concentrations for the calibrators are provided in the following table. | Calibrator | Quantity | Anti-Rubella IgG Concentration (IU/mL) | | --- | --- | --- | | Calibrator A | 1 x 3.0 mL | 0.0 | | Calibrator B | 1 x 3.0 mL | 5.5 | | Calibrator C | 1 x 3.0 mL | 16.5 | | Calibrator D | 1 x 3.0 mL | 82.5 | | Calibrator E | 1 x 3.0 mL | 275.0 | | Calibrator F | 1 x 3.0 mL | 550.0 | The Alinity i Rubella IgG Calibrator B through F are referenced to the World Health Organization (WHO) International Standard (RUBI-1-94) for Anti-Rubella Immunoglobulin at each concentration level. Alinity i Rubella IgG K243168 510(k) Summary Page 3 of 18 {7} # Controls The Alinity i Rubella IgG Controls are described below. - Negative Control: Contains recalcified human plasma with protein (ovine) stabilizer. - Positive Control 1 and 2: Contain recalcified human plasma (reactive for anti-rubella IgG) with protein (ovine) stabilizer. - Preservatives: ProClin 950 and sodium azide. The target concentrations and ranges for the controls are provided in the table below. | Control | Quantity | Anti-Rubella IgG | | | --- | --- | --- | --- | | | | Concentration (IU/mL) | Range (IU/mL) | | Negative Control | 1 x 8.0 mL | 0.0 | < 4.4 | | Positive Control 1 | 1 x 8.0 mL | 30.7 | 18.5 – 42.9 | | Positive Control 2 | 1 x 8.0 mL | 348.8 | 209.3 – 488.3 | The Alinity i Rubella IgG positive controls are traceable to the WHO International Standard (RUBI-1-94) for Anti-Rubella Immunoglobulin at each concentration. # Biological Principles of the Procedure The Alinity i Rubella IgG assay is an automated, two-step immunoassay for the quantitative determination of anti-rubella IgG in human serum and plasma using chemiluminescent microparticle immunoassay (CMIA) technology. Sample, partially purified rubella virus-coated paramagnetic microparticles, and assay diluent are combined and incubated. The anti-rubella IgG present in the sample bind to the rubella virus coated microparticles. The mixture is washed. Anti-human IgG acridinium-labeled conjugate is added to create a reaction mixture and incubated. Following a wash cycle, Pre-Trigger and Trigger Solutions are added. Alinity i Rubella IgG K243168 510(k) Summary {8} The resulting chemiluminescent reaction is measured as a relative light unit (RLU). There is a direct relationship between the amount of anti-rubella IgG in the sample and the RLU detected by the system optics. ## VI. Intended Use of the Device The Alinity i Rubella IgG assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of IgG antibodies to rubella virus in human serum, serum separator, and plasma tubes (lithium heparin, lithium heparin separator, and tripotassium EDTA) on the Alinity i system. The Alinity i Rubella IgG assay is to be used as an aid in the determination of immune status to rubella in individuals including women of child-bearing age. The Alinity i Rubella IgG assay has not been cleared for use in screening blood, plasma, or tissue donors. The performance of this device has not been established for cord blood or neonatal samples. Likewise, performance has not been established for populations of immunocompromised or immunosuppressed individuals. ## VII. Comparison of Technological Characteristics The Alinity i Rubella IgG assay (subject device) utilizes a CMIA methodology for the quantitative determination of IgG antibodies to rubella virus and is intended for use on the Alinity i system. The similarities and differences between the subject device and the predicate device are presented in the following tables. Alinity i Rubella IgG K243168 510(k) Summary Page 5 of 18 {9} Assay Similarities | Characteristics | Subject Device Alinity i Rubella IgG (K243168) | Predicate Device bioMérieux VIDAS RUB IgG (K080766) | | --- | --- | --- | | Intended Use and Indications for Use | The Alinity i Rubella IgG assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of IgG antibodies to rubella virus in human serum, serum separator, and plasma tubes (lithium heparin, lithium heparin separator, and tripotassium EDTA) on the Alinity i system. The Alinity i Rubella IgG assay is to be used as an aid in the determination of immune status to rubella in individuals including women of child-bearing age. The Alinity i Rubella IgG assay has not been cleared for use in screening blood, plasma, or tissue donors. The performance of this device has not been established for cord blood or neonatal samples. Likewise, performance has not been established for populations of immunocompromised or immunosuppressed individuals. | The VIDAS® RUB IgG (RBG) assay uses Enzyme Linked Fluorescent Assay (ELFA) technology on the VIDAS® automated instruments for the in vitro quantitative and qualitative measurement of IgG antibodies to rubella virus in human serum. The VIDAS® RUB IgG assay is intended as an aid in the determination of immune status to rubella. The performance of this device has not been established for screening of cord blood, or for neonatal samples. Likewise, performance characteristics of the assay have not been established for immunocompromised or immunosuppressed individuals. | | Interpretation of Results | Nonreactive: < 5.0 IU/mL Grayzone/Equivocal: 5.0 to < 10.0 IU/mL Reactive: ≥ 10.0 IU/mL | Negative: < 5 IU/mL Equivocal: 5 to < 10 IU/mL Positive: ≥ 10 IU/mL | | Antigen and Antibody Used | • Rubella virus antigen • Anti-human IgG antibody (mouse, monoclonal) | • Rubella virus antigen (Strain MR 383) • Anti-human IgG antibodies (mouse, monoclonal) | Alinity i Rubella IgG K243168 510(k) Summary {10} Assay Differences | Characteristics | Subject Device Alinity i Rubella IgG (K243168) | Predicate Device bioMérieux VIDAS RUB IgG (K080766) | | --- | --- | --- | | Calibrator(s) | 6 Calibrators | 1 Calibrator | | Control(s) | 3 (1 Negative, 2 Positives) | 2 (1 Negative, 1 Positive) | | Type of Specimen | Serum and Plasma | Serum | | Methodology | Chemiluminescent microparticle immunoassay (CMIA) | Enzyme-linked fluorescent assay (ELFA) | | Components | Microparticles – Partially purified rubella virus coated microparticles in TRIS buffer with surfactant. Minimum concentration: 0.12% solids. Preservatives: ProClin 950 and sodium azide. Conjugate – Anti-human IgG (mouse, monoclonal) acridinium-labeled conjugate in MES buffer with protein (bovine) stabilizer and surfactant. Minimum concentration: 16 ng/mL. Preservatives: antimicrobial agents. Assay Diluent – TRIS buffer with protein (caprine, bovine, murine) stabilizers and surfactant. Preservatives: ProClin 300 and ProClin 950. | RBG SPRs – Solid Phase Receptacles (SPR) coated with rubella antigen (Strain MR 383). Ready to use. Conjugate – Alkaline phosphatase labeled monoclonal anti-human IgG antibodies (mouse); 0.09% sodium azide. Sample Diluent and Wash Solution – TRIS buffer (50 mmol/L) pH 7.4; protein and chemical stabilizers; 0.09% sodium azide. | | Calibration Storage | Maximum of 30 days | Maximum of 14 days | Alinity i Rubella IgG K243168 510(k) Summary Page 7 of 18 {11} VIII. Summary of Nonclinical Performance A. Within-Laboratory Precision (20-Day) A 20-day within-laboratory precision study was performed based on guidance from CLSI EP05-A3.* Testing was conducted using 3 lots of the Alinity i Rubella IgG reagents, 3 lots of the Alinity i Rubella IgG Calibrators, 3 lots of the Alinity i Rubella IgG Controls, and 1 instrument. Three controls and 7 recalcified human plasma panels (representing serum matrix) were tested in 3 replicates at 2 separate times per day on 20 different days using 3 reagent lot/calibrator lot combinations, where a unique reagent lot and a unique calibrator lot are paired. The performance is shown in the following table. | Sample | n | Mean (IU/mL) | Repeatability (Within-Run) | | Between-Run | | Between-Day | | Between-Lota | | Overall Within-Laboratoryb | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | Negative Control | 360 | 0.0 | 0.06 | NAc | 0.00 | NAc | 0.00 | NAc | 0.00 | NAc | 0.06 | NAc | | Positive Control 1 | 360 | 30.4 | 1.04 | 3.4 | 0.39 | 1.3 | 0.37 | 1.2 | 0.94 | 3.1 | 1.50 | 4.9 | | Positive Control 2 | 360 | 342.7 | 18.14 | 5.3 | 7.58 | 2.2 | 4.26 | 1.2 | 8.55 | 2.5 | 21.86 | 6.4 | | Panel 1 | 359d | 0.0 | 0.03 | NAc | 0.01 | NAc | 0.02 | NAc | 0.02 | NAc | 0.04 | NAc | | Panel 2 | 360 | 3.0 | 0.14 | NAc | 0.04 | NAc | 0.04 | NAc | 0.19 | NAc | 0.24 | NAc | | Panel 3 | 358d | 8.0 | 0.33 | NAc | 0.08 | NAc | 0.01 | NAc | 0.34 | NAc | 0.49 | NAc | | Panel 4 | 360 | 12.6 | 0.47 | 3.7 | 0.17 | 1.4 | 0.17 | 1.4 | 0.51 | 4.0 | 0.73 | 5.8 | | Panel 5 | 360 | 25.1 | 1.05 | 4.2 | 0.21 | 0.8 | 0.00 | 0.0 | 0.76 | 3.0 | 1.31 | 5.2 | | Panel 6 | 360 | 234.8 | 10.98 | 4.7 | 5.52 | 2.3 | 0.00 | 0.0 | 8.22 | 3.5 | 14.79 | 6.3 | | Panel 7 | 360 | 398.1 | 29.74 | 7.5 | 7.18 | 1.8 | 0.00 | 0.0 | 9.19 | 2.3 | 31.94 | 8.0 | a Alinity i Rubella IgG reagent lot and Alinity i Rubella IgG calibrator lot are confounded, and the confounding effect is represented by between-lot. b Overall within-laboratory variability contains repeatability (within-run), between-run, between-day, and between-lot variance components. c Not applicable d In cases where n &lt; 360, replicate(s) were excluded due to an instrument error and no results were reported. Alinity i Rubella IgG K243168 510(k) Summary Page 8 of 18 {12} B. Reproducibility Study (5-Day) A 5-day reproducibility study was conducted at 3 US sites based on guidance from CLSI EP05-A3* using 3 controls and 7 recalcified human plasma panels (representing serum matrix). Four replicates per sample were evaluated in 2 runs per day over 5 days. Testing was conducted using 3 lots of the Alinity i Rubella IgG reagents, 2 lots of the Alinity i Rubella IgG Calibrators, and 2 lots of the Alinity i Rubella IgG Controls at each of the 3 testing sites. | Sample | n | Mean (IU/mL) | Repeatability | | Between-Run | | Between-Day | | Between-Site | | Between-Lota | | Reproducibilityb | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | Negative Control | 359d | 0.0 | 0.00 | NAc | 0.00 | NAc | 0.00 | NAc | 0.00 | NAc | 0.00 | NAc | 0.00 | NAc | | Positive Control 1 | 359d | 29.4 | 0.89 | 3.0 | 0.26 | 0.9 | 0.35 | 1.2 | 0.13 | 0.5 | 0.23 | 0.8 | 1.03 | 3.5 | | Positive Control 2 | 360 | 349.1 | 15.63 | 4.5 | 5.03 | 1.4 | 14.66 | 4.2 | 4.01 | 1.1 | 9.33 | 2.7 | 24.25 | 6.9 | | Panel 1 | 360 | 0.0 | 0.02 | NAc | 0.00 | NAc | 0.01 | NAc | 0.01 | NAc | 0.01 | NAc | 0.02 | NAc | | Panel 2 | 359d | 2.8 | 0.10 | NAc | 0.07 | NAc | 0.04 | NAc | 0.07 | NAc | 0.01 | NAc | 0.14 | NAc | | Panel 3 | 360 | 7.6 | 0.33 | NAc | 0.04 | NAc | 0.10 | NAc | 0.14 | NAc | 0.09 | NAc | 0.38 | NAc | | Panel 4 | 360 | 12.1 | 0.32 | 2.6 | 0.25 | 2.1 | 0.18 | 1.5 | 0.06 | 0.5 | 0.13 | 1.1 | 0.47 | 3.9 | | Panel 5 | 360 | 24.8 | 0.69 | 2.8 | 0.59 | 2.4 | 0.27 | 1.1 | 0.11 | 0.5 | 0.30 | 1.2 | 1.00 | 4.0 | | Panel 6 | 360 | 245.2 | 10.53 | 4.3 | 6.31 | 2.6 | 5.68 | 2.3 | 1.62 | 0.7 | 7.21 | 2.9 | 15.41 | 6.3 | | Panel 7 | 356d | 410.9 | 22.94 | 5.6 | 11.15 | 2.7 | 20.94 | 5.1 | 8.39 | 2.0 | 4.33 | 1.1 | 34.32 | 8.4 | a Alinity i Rubella IgG reagent lot and Alinity i Rubella IgG calibrator lot are confounded, and the confounding effect is represented by between-lot. b Reproducibility contains repeatability, between-run, between-day, between-site, and between-lot variance components. c Not applicable d In cases where n &lt; 360, replicate(s) were excluded due to an instrument error and no results were reported. C. Lower Limits of Measurement A study was performed based on guidance from CLSI EP17-A2. Testing was conducted using 3 lots of the Alinity i Rubella IgG reagents on each of Alinity i Rubella IgG K243168 510(k) Summary Page 9 of 18 {13} 2 instruments over a minimum of 3 days. The maximum observed limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) values are summarized below. | | IU/mL | | --- | --- | | LoBa | 0.1 | | LoDb | 0.3 | | LoQc | 0.7 | a The LoB represents the 95th percentile from n ≥ 60 replicates of zero-analyte samples. b The LoD represents the lowest concentration at which the analyte can be detected with 95% probability based on n ≥ 60 replicates of low-analyte level samples. c The LoQ is defined as the lowest concentration at which a maximum allowable precision of 20 %CV and a maximum allowable bias of 15% were independently met. The LoQ was determined from n ≥ 60 replicates of low-analyte level samples. ## D. Linearity A study was performed based on guidance from CLSI EP06, 2nd ed.* This assay is linear across the analytical measuring interval of 1.0 to 500.0 IU/mL. ## E. Analytical Specificity / Interference ### Potentially Interfering Endogenous Substances The Alinity i Rubella IgG assay was evaluated for potential interference caused by endogenous substances based on guidance from CLSI EP07, 3rd ed.,† and CLSI EP37, 1st ed.‡ Each substance was evaluated using samples containing anti-rubella IgG at the target ranges of 5.0 to 9.9 IU/mL and 10.0 to 20.0 IU/mL. No significant interference (interference within ± 1.0 IU/mL for samples &lt; 10.0 IU/mL and within ± 10% for samples ≥ 10.0 IU/mL) was observed at the following concentrations. Alinity i Rubella IgG K243168 510(k) Summary Page 10 of 18 {14} | No Significant Interference | | | --- | --- | | Potentially Interfering Substance | Interferent Level | | Conjugated Bilirubin | 40 mg/dL | | Unconjugated Bilirubin | 40 mg/dL | | Hemoglobin | 1000 mg/dL | | Total Protein | 15 g/dL | | Triglycerides | 3000 mg/dL | ## Potentially Interfering Drugs and Other Substances The Alinity i Rubella IgG assay was evaluated for potential interference caused by exogenous substances based on guidance from CLSI EP07, 3rd ed., and CLSI EP37, 1st ed. Each substance was evaluated using samples containing anti-rubella IgG at the target ranges of 5.0 to 9.9 IU/mL and 10.0 to 20.0 IU/mL. ## No significant interference (interference within ± 1.0 IU/mL for samples &lt; 10.0 IU/mL and within ± 10% for samples ≥ 10.0 IU/mL) was observed at the following concentrations. | No Significant Interference | | | --- | --- | | Potentially Interfering Substance | Interferent Level | | Acetaminophen | 250 mg/L | | Acetylsalicylic Acid | 1000 mg/L | | Ascorbic Acid | 300 mg/L | | Biotin | 4250 ng/mL | | Folic Acid | 100 nmol/L | | Ibuprofen | 500 mg/L | ## Potential Cross-Reactivity Potential cross-reactivity for the Alinity i Rubella IgG assay was determined by testing a total of 126 serum specimens from individuals with other medical conditions unrelated to rubella virus, in addition to individual(s) with IgM antibodies to rubella virus. The results are shown in the following table. Alinity i Rubella IgG K243168 510(k) Summary Page 11 of 18 {15} | Category | n | Number of Alinity i Rubella IgG Reactive Results | Number of Alinity i Rubella IgG Grayzone/ Equivocal Results | | --- | --- | --- | --- | | Anti-dsDNA Antibodies | 4 | 0 | 0 | | Anti-nuclear Antibody (ANA) | 5 | 0 | 0 | | Cytomegalovirus (IgG) | 7 | 0 | 0 | | Epstein-Barr Virus (EBV) IgG | 5 | 0 | 0 | | HAMA | 3 | 0 | 1^{a} | | Hepatitis B Antibodies | 10 | 0 | 0 | | Hepatitis C Antibodies | 4 | 0 | 0 | | Herpes Simplex Virus Type 1 (IgG) | 10 | 0 | 0 | | Herpes Simplex Virus Type 2 (IgG) | 5 | 0 | 0 | | Human Immunodeficiency Virus (HIV) | 5 | 0 | 0 | | Human Chorionic Gonadotropin | 5 | 0 | 1^{b} | | Hyper IgG | 5 | 0 | 1^{b} | | Measles (IgG) | 7 | 0 | 0 | | Mumps (IgG) | 4 | 0 | 0 | | Parvovirus B19 (IgG) | 4 | 0 | 0 | | Rheumatoid Factor (RF) | 5 | 0 | 0 | | Rubella (IgM) | 1 | 0 | 0 | | SARS-CoV-2 IgG | 3 | 0 | 0 | | Syphilis | 10 | 0 | 1^{b} | | Systemic Lupus Erythematosus | 8 | 0 | 0 | | Toxoplasmosis (IgG) | 6 | 0 | 0 | | Varicella Zoster Virus (IgG) | 10 | 0 | 0 | | Total | 126 | 0 | 4 | a Positive result obtained on the comparator assay. b Equivocal results obtained on the comparator assay. ## F. Matrix Equivalency A study was performed to evaluate whether specific blood collection tube types are suitable for use with the Alinity i Rubella IgG assay. The matrix collection tube type equivalency study was conducted including 42 donors of reactive (29 donors) and nonreactive (13 donors) samples in 5 types of blood collection tubes (serum, serum separator, lithium heparin plasma, lithium heparin plasma separator, and Alinity i Rubella IgG K243168 510(k) Summary {16} tripotassium EDTA plasma) for use with the Alinity i Rubella IgG assay. Data was analyzed using regression analysis comparing concentrations of all matrices to serum to evaluate any potential bias. All of the blood collection tube types tested are acceptable for use with the Alinity i Rubella IgG assay. Statistical evaluation data are summarized below. | Collection Tube | Slope (95% CI) (IU/mL) | Intercept (95% CI) (IU/mL) | Correlation Coefficient (r) | | --- | --- | --- | --- | | Serum separator | 1.01 (0.98, 1.02) | -0.3 (-0.5, 0.2) | 1.00 | | Lithium heparin plasma | 0.99 (0.97, 1.01) | -0.1 (-0.5, 0.2) | 1.00 | | Lithium heparin plasma separator | 1.00 (0.97, 1.02) | -0.2 (-0.7, 0.1) | 1.00 | | Tripotassium EDTA plasma | 0.96 (0.94, 0.98) | -0.1 (-0.4, 0.0) | 1.00 | # G. CDC Panel Agreement The Centers for Disease Control and Prevention (CDC) Rubella Reference Sera Panel (collected in 2014) was tested using the Alinity i Rubella IgG assay. The Alinity i Rubella IgG assay results were submitted to the CDC for data analysis and for the result interpretation for each sample. The panel consisted of 82 true positive rubella specimens and 18 true negative rubella specimens. Out of the 82 positive specimens, the Alinity i Rubella IgG assay detected 77 as reactive and 5 as equivocal, while all 18 negative specimens were detected as nonreactive. The CDC performed kit sensitivity (positive percent agreement [PPA]) and kit specificity (negative percent agreement [NPA]) analyses and sent the results to Abbott. The percent agreement of the Alinity i Rubella IgG assay relative to the CDC results was calculated. The PPA was $93.9\%$ with a $95\%$ confidence interval (CI) of $86.51\%$ to $97.37\%$ . The NPA was $100.0\%$ with a $95\%$ CI of $82.41\%$ to $100.00\%$ . Alinity i Rubella IgG K243168 510(k) Summary {17} The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply endorsement of the assay by the CDC. | Alinity i Rubella IgG Interpretation | CDC Interpretation | | Positive % Agreement (95% CI)a | Negative % Agreement (95% CI)a | | --- | --- | --- | --- | --- | | | Positive | Negative | | | | Reactive | 77 | 0 | 93.9 | 100.0 | | Grayzone/Equivocal | 5 | 0 | (77/82) | (18/18) | | Nonreactive | 0 | 18 | (86.51, 97.37) | (82.41, 100.00) | a The 95% CI for PPA and NPA were estimated using the Wilson score method. ## H. CDC Low Titer Standard The ability of the Alinity i Rubella IgG assay to detect and measure low levels of anti-rubella IgG antibodies was verified by testing the CDC Low-Titer Anti-Rubella Human Reference Serum - CDC Biological Standard (catalog number: IS2153 at 21.0 IU/mL). The mean concentration was 18.2 IU/mL for the resuspended CDC low titer standard and 8.6 IU/mL for the 1:2 dilution of the standard. ## IX. Summary of Clinical Performance ### A. Clinical Agreement A clinical method comparison study was conducted to evaluate the clinical performance of the Alinity i Rubella IgG assay based on guidance from CLSI EP12-A2* using specimens collected from individuals included in the following categories. - The Routine Order category was comprised of consecutively collected remnant specimens from individuals (non-pregnant) with a physician’s routine order for anti-rubella IgG, including specimens collected in the US (n=946) and outside of the US (OUS) (n=457). - The Pregnant Females (US) category was comprised of 435 specimens collected from pregnant females in the US. Alinity i Rubella IgG K243168 510(k) Summary Page 14 of 18 {18} - The Supplemental Preselected Negative Pregnant Female category was comprised of specimens that were additionally obtained to supplement the small number of negative subjects in the Pregnant Females (US) category. These specimens were from pregnant females who had a negative anti-rubella IgG test result with the comparator assay. They were collected in the US (n=52) and outside of the US (OUS) (n=59). - The Preselected Negative (OUS) category was comprised of 135 consecutively collected remnant specimens from individuals (non-pregnant) with a physician's routine order for anti-rubella IgG and a negative anti-rubella IgG test result. Demographic information for specimens from the intended use population collected in the US (n=1433) is shown in the table below. | Specimen | Age | Female (n) | Male (n) | Unknown (n) | Total (n) | | --- | --- | --- | --- | --- | --- | | Routine Order (US) (n=946) | ≤ 5 years | 1 | 1 | 0 | 2 | | | 6 to 21 years | 92 | 29 | 0 | 121 | | | 22 to 59 years | 666 | 121 | 3 | 790 | | | ≥ 60 years | 17 | 16 | 0 | 33 | | | Total | 776 | 167 | 3 | 946 | | Pregnant Females (US) (n=435) | ≤ 5 years | 0 | 0 | 0 | 0 | | | 6 to 21 years | 28 | 0 | 0 | 28 | | | 22 to 59 years | 404 | 0 | 0 | 404 | | | ≥ 60 years | 0 | 0 | 0 | 0 | | | Unknown | 3 | 0 | 0 | 3 | | | Total | 435 | 0 | 0 | 435 | | Supplemental Preselected Negative Pregnant Females (US) (n=52) | ≤ 5 years | 0 | 0 | 0 | 0 | | | 6 to 21 years | 5 | 0 | 0 | 5 | | | 22 to 59 years | 47 | 0 | 0 | 47 | | | ≥ 60 years | 0 | 0 | 0 | 0 | | | Total | 52 | 0 | 0 | 52 | | Total (US) (n=1433) | ≤ 5 years | 1 | 1 | 0 | 2 | | | 6 to 21 years | 125 | 29 | 0 | 154 | | | 22 to 59 years | 1117 | 121 | 3 | 1241 | | | ≥ 60 years | 17 | 16 | 0 | 33 | | | Unknown | 3 | 0 | 0 | 3 | | | Total | 1263 | 167 | 3 | 1433 | Alinity i Rubella IgG K243168 510(k) Summary Page 15 of 18 {19} The clinical performance of the Alinity i Rubella IgG assay was evaluated comparing the results of the Alinity i Rubella IgG assay and the results of a composite comparator method comprised of 3 FDA-cleared anti-rubella IgG assays. Specimens with initially equivocal results on the comparator device (5 to &lt; 10 IU/mL) were tested with 2 additional FDA-cleared devices, and the final 2 out of 3 consensus results were used for performance evaluation. The positive percent agreement (PPA) and negative percent agreement (NPA) between the Alinity i Rubella IgG investigational assay and the composite comparator method anti-rubella IgG assays were calculated. The PPA and NPA for each specimen category are shown in the following 3 by 3 table. | Specimen Category | Alinity i Rubella IgG Resulta | Composite Comparator Anti-Rubella IgG Result | | | PPA (95% CI)b | NPA (95% CI)b | | --- | --- | --- | --- | --- | --- | --- | | | | Positive | Equivocal | Negative | | | | Routine Order (US) (n=946) | Reactive | 822 | 2 | 0 | 93.94 (822/875) (92.16, 95.34) | 86.67 (26/30) (70.32, 94.69) | | | Grayzone/Equivocal | 36 | 41 | 2 | | | | | Nonreactive | 4 | 13 | 26 | | | | Routine Order (OUS) (n=457) | Reactive | 378 | 3 | 0 | 97.42 (378/388) (95.32, 98.59) | 88.37 (38/43) (75.52, 94.93) | | | Grayzone/Equivocal | 9 | 26 | 2 | | | | | Nonreactive | 0 | 1 | 38 | | | | Pregnant Females (US) (n=435) | Reactive | 360 | 2 | 0 | 93.02 (360/387) (90.04, 95.16) | 77.78 (14/18) (54.79, 91.00) | | | Grayzone/Equivocal | 17 | 30 | 2 | | | | | Nonreactive | 1 | 9 | 14 | | | | Supplemental Preselected Negative Pregnant Females (US) (n=52) | Reactive | 0 | 0 | 0 | NAc | 100.00 (52/52) (93.12, 100.00) | | | Grayzone/Equivocal | 0 | 0 | 0 | | | | | Nonreactive | 0 | 0 | 52 | | | Alinity i Rubella IgG K243168 510(k) Summary Page 16 of 18 {20} | Specimen Category | Alinity i Rubella IgG Resulta | Composite Comparator Anti-Rubella IgG Result | | | PPA (95% CI)b | NPA (95% CI)b | | --- | --- | --- | --- | --- | --- | --- | | | | Positive | Equivocal | Negative | | | | Supplemental Preselected Negative Pregnant Females (OUS) (n=59) | Reactive | 0 | 0 | 1d | NAc | 98.31 (58/59) (91.00, 99.70) | | | Grayzone/Equivocal | 0 | 0 | 0 | | | | | Nonreactive | 0 | 0 | 58 | | | | Preselected Negative (OUS) (n=135) | Reactive | 0 | 0 | 0 | 0.00 (0/5) (0.00, 43.45) | 99.23 (129/130) (95.77, 99.86) | | | Grayzone/Equivocal | 0 | 0 | 1 | | | | | Nonreactive | 3 | 2 | 129 | | | a The interpretation for Alinity i Rubella IgG results is $\geq 10.0\mathrm{IU / mL}$ for reactive, $5.0\mathrm{to} &lt; 10.0\mathrm{IU / mL}$ for grayzone/equivocal, and $&lt; 5.0\mathrm{IU / mL}$ for nonreactive. b The $95\%$ CI for PPA and NPA were estimated using the Wilson score method. c Not applicable. d One sample was discordant between Alinity i Rubella IgG (19.0 IU/mL) and the comparator (4 IU/mL). After retesting, the results with Alinity i Rubella IgG were 18.4 and 18.5 IU/mL, and the results with the comparator were 24 and 26 IU/mL. An additional assessment of the PPA and NPA between the Alinity i Rubella IgG assay and the composite comparator, relative to the medical decision point of $10\mathrm{IU / mL}$ antibody level considered to provide immunity from Rubella infection, is provided in the following 2 by 2 table. | Specimen Category | Alinity i Rubella IgG Result (IU/mL)a | Composite Comparator Anti-Rubella IgG Result | | PPA (95% CI)b | NPA (95% CI)b | | --- | --- | --- | --- | --- | --- | | | | Positive | Negative or Equivocal | | | | Routine Order (US) (n=946) | ≥ 10.0 | 822 | 2 | 95.36 (822/862) (93.74, 96.57) | 97.62 (82/84) (91.73, 99.34) | | | < 10.0 | 40 | 82 | | | | Routine Order (OUS) (n=457) | ≥ 10.0 | 378 | 3 | 97.67 (378/387) (95.64, 98.77) | 95.71 (67/70) (88.14, 98.53) | | | < 10.0 | 9 | 67 | | | Alinity i Rubella IgG K243168 510(k) Summary {21} | Specimen Category | Alinity i Rubella IgG Result (IU/mL)a | Composite Comparator Anti-Rubella IgG Result | | PPA (95% CI)b | NPA (95% CI)b | | --- | --- | --- | --- | --- | --- | | | | Positive | Negative or Equivocal | | | | Pregnant Females (US) (n=435) | ≥ 10.0 | 360 | 2 | 95.24 (360/378) (92.60, 96.97) | 96.49 (55/57) (88.08, 99.03) | | | < 10.0 | 18 | 55 | | | | Supplemental Preselected Negative Pregnant Females (US) (n=52) | ≥ 10.0 | 0 | 0 | NAc | 100.00 (52/52) (93.12, 100.00) | | | < 10.0 | 0 | 52 | | | | Supplemental Preselected Negative Pregnant Females (OUS) (n=59) | ≥ 10.0 | 0 | 1 | NAc | 98.31 (58/59) (91.00, 99.70) | | | < 10.0 | 0 | 58 | | | | Preselected Negative (OUS) (n=135) | ≥ 10.0 | 0 | 0 | 0.00 (0/3) (0.00, 56.15) | 100.00 (132/132) (97.17, 100.00) | | | < 10.0 | 3 | 132 | | | a The interpretation for Alinity i Rubella IgG results is $\geq 10.0\mathrm{IU / mL}$ for reactive and $&lt; 10.0~\mathrm{IU / mL}$ for grayzone/equivocal and nonreactive. b The $95\%$ CI for PPA and NPA were estimated using the Wilson score method. c Not applicable. # X. Conclusion Drawn from Nonclinical and Clinical Laboratory Studies The results presented in this $510(\mathrm{k})$ premarket notification demonstrate that the subject device (Alinity i Rubella IgG, K243168) performance is substantially equivalent to the predicate assay (bioMérieux VIDAS RUB IgG assay, K080766). Alinity i Rubella IgG K243168 510(k) Summary
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...