BIOPLEX 2200 RUBELLA & CMV IGM KIT ON THE BIOPLEX 2200 MULTI ANALYTE DETECTION SYSTEM
Applicant
Bio-Rad Laboratories
Product Code
LFX · Microbiology
Decision Date
Dec 3, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3510
Device Class
Class 2
Intended Use
The BioPlex™ 2200 Rubella and CMV IgM kit is a multiplex flow immunoassay intended for the qualitative detection of IgM antibodies to Rubella and Cytomegalovirus (CMV) in human serum and potassium EDTA or sodium heparin plasma. The BioPlex 2200 Rubella and CMV IgM kit is intended for use with the Bio-Rad BioPlex 2200 System. This kit is intended as an aid in the diagnosis of a current or recent Rubella and/or CMV infection, in individuals suspected of having one of the respective disease states including women of child bearing age. This assay is not FDA cleared or approved for use in testing (screening) blood or plasma donors. Performance characteristics for the Rubella and CMV IgM assays have not been evaluated in immunosupressed or organ transplant individuals. Performance characteristics of this kit have not been established for use in neonatal screening or for use at point of care facilities. The BioPlex® 2200 Rubella and CMV IgM Calibrator Set is intended for the calibration of the BioPlex 2200® Rubella and CMV IgM Reagent Pack. The BioPlex® 2200 Rubella and CMV IgM Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 instrument and BioPlex 2200® Rubella and CMV IgM Reagent Pack in the clinical laboratory. The performance of the BioPlex 2200® Rubella and CMV IgM Control Set has not been established with any other Rubella and CMV IgM assays.
Device Story
Multiplex flow immunoassay system for simultaneous detection of Rubella and CMV IgM antibodies. Input: patient serum or plasma sample. Process: sample, diluent, and antigen-coated dyed beads incubated; anti-human IgM-phycoerythrin (PE) conjugate added; bead mixture passed through detector. Detector identifies beads via dye fluorescence and quantifies captured antibody via PE fluorescence. Includes Internal Standard, Serum Verification, and Reagent Blank beads for quality control. Output: Relative Fluorescence Intensity (RFI) converted to antibody index (AI) values. Used in clinical laboratories by trained personnel. Results aid clinicians in diagnosing current or recent infections, supporting patient management decisions.
Clinical Evidence
Clinical performance evaluated via prospective and retrospective studies. Prospective study (N=700) compared results against commercial immunoassays; Rubella IgM positive percent agreement 66.7%, negative 95.6%; CMV IgM positive 53.8%, negative 97.7%. Retrospective study (N=336 presumptive positives) showed 96.3% positive agreement for Rubella and 91.3% for CMV. Reproducibility studies (CLSI EP5-A2) showed total precision CVs ranging from 2.3% to 14.2%. Matrix comparison (N=20) confirmed equivalence between serum and plasma (r ≥ 0.98).
Technological Characteristics
Multiplexed flow immunoassay using paramagnetic microbeads infused with fluorescent dyes. Antigens: Rubella and CMV lysates. Detection: Phycoerythrin-conjugated anti-human IgM. Energy: Fluorescence-based detection. Connectivity: Integrated with BioPlex 2200 System. Preservatives: ProClin 300, sodium azide, sodium benzoate. Form factor: Reagent pack with bead set, conjugate, and diluent.
Indications for Use
Indicated for qualitative detection of IgM antibodies to Rubella and CMV in human serum, potassium EDTA, or sodium heparin plasma. Intended as an aid in diagnosis of current or recent infection in individuals suspected of having these diseases, including women of childbearing age. Not for blood/plasma donor screening, neonatal screening, or point-of-care use. Not evaluated in immunosuppressed or organ transplant patients.
Regulatory Classification
Identification
Rubella virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubella virus in serum. The identification aids in the diagnosis of rubella (German measles) or confirmation of a person's immune status from past infections or immunizations and provides epidemiological information on German measles. Newborns infected in the uterus with rubella virus may be born with multiple congenital defects (rubella syndrome).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) National Committee for Clinical Laboratory Standards':
(i) 1/LA6 “Detection and Quantitation of Rubella IgG Antibody: Evaluation and Performance Criteria for Multiple Component Test Products, Speciment Handling, and Use of the Test Products in the Clinical Laboratory, October 1997,”
(ii) 1/LA18 “Specifications for Immunological Testing for Infectious Diseases, December 1994,”
(iii) D13 “Agglutination Characteristics, Methodology, Limitations, and Clinical Validation, October 1993,”
(iv) EP5 “Evaluation of Precision Performance of Clinical Chemistry Devices, February 1999,” and
(v) EP10 “Preliminary Evaluation of the Linearity of Quantitive Clinical Laboratory Methods, May 1998,”
(2) Centers for Disease Control's:
(i) Low Titer Rubella Standard,
(ii) Reference Panel of Well Characterized Rubella Sera, and
(3) World Health Organization's International Rubella Standard.
K080008 — BIOPLEX 2200 SYSTEM TORC IGG KIT, CALIBRATOR SET AND CONTROL SET · Bio-Rad Laboratories, Inc. · Feb 23, 2009
K170509 — BioPlex 2200 ToRC IgM, BioPlex 2200 ToRC IgM Calibrator Set, BioPlex 2200 ToRC IgM Control Set · Bio-Rad Laboratories · May 19, 2017
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K092587
B. Purpose for Submission:
New device
C. Measurand:
IgM antibodies specific to Rubella and Cytomegalovirus (CMV).
D. Type of Test:
Multiplex flow immunoassay (multiplexed fluoromagnetic bead assay)
E. Applicant:
Bio-Rad Laboratories Inc.
F. Proprietary and Established Names:
BioPlex™ 2200 Rubella and CMV IgM Kit
BioPlex™ 2200 Rubella and CMV IgM Calibrator Set
BioPlex™ 2200 Rubella and CMV IgM Control Set
G. Regulatory Information:
| Product code | Classification | Regulation section | Panel |
| --- | --- | --- | --- |
| LFX: Enzyme Linked Immunoabsorbent Assay, Rubella | Class II | 866.3510; Rubella Virus Serological Reagents | Microbiology |
| JIX: Calibrator, multi-analyte mixture | Class II | 862.1150 - Calibrator | Clinical Chemistry |
| JJX: Single (specified) analyte controls (assayed and unassayed) | Class I | 862.1660 - Quality Control Material (assayed and unassayed) | Clinical Chemistry |
Note: The BioPlex™ 2200 Rubella and CMV IgM Kit is a multiplex immunoassay for the detection of IgM antibodies to Rubella and CMV. This device is classified as Class II as described above and no new product code is assigned for this device. The first product code assigned for the device is listed under the regulation section for Rubella reagents.
{1}
The following is the regulation section and product code that is applicable to the other analyte detected by the device subject of this submission.
866.3175 - Cytomegalovirus serological reagents. Class II (Microbiology). Antibody IgM, IF, cytomegalovirus virus (LKQ) and Enzyme Linked Immunoabsorbent Assay, Cytomegalovirus (LFZ)
## H. Intended Use:
### 1. Intended use(s):
#### The BioPlex™ 2200 Rubella and CMV IgM kit
The BioPlex™ 2200 Rubella and CMV IgM kit is a multiplex flow immunoassay intended for the qualitative detection of IgM antibodies to Rubella and Cytomegalovirus (CMV) in human serum and potassium EDTA or sodium heparin plasma.
The BioPlex 2200 Rubella and CMV IgM kit is intended for use with the Bio-Rad BioPlex 2200 System.
This kit is intended as an aid in the diagnosis of a current or recent Rubella and/or CMV infection, in individuals suspected of having one of the respective disease states including women of child bearing age.
This assay is not FDA cleared or approved for use in testing (screening) blood or plasma donors.
Performance characteristics for the Rubella and CMV IgM assays have not been evaluated in immunosupressed or organ transplant individuals. Performance characteristics of this kit have not been established for use in neonatal screening or for use at point of care facilities.
#### The BioPlex 2200 Rubella and CMV IgM Calibrator Set
The BioPlex® 2200 Rubella and CMV IgM Calibrator Set is intended for the calibration of the BioPlex 2200® Rubella and CMV IgM Reagent Pack.
#### The BioPlex 2200 Rubella and CMV IgG Control Set
The BioPlex® 2200 Rubella and CMV IgM Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 instrument and BioPlex 2200® Rubella and CMV IgM Reagent Pack in the clinical laboratory. The performance of the BioPlex 2200® Rubella and CMV IgM Control Set has not been established with any other Rubella and CMV IgM assays.
2
{2}
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
The Bio-Rad BioPlex 2200 System
I. Device Description:
The BioPlex 2200 Rubella and CMV IgM kit is a fully automated multiplexed micro particle bead based immunoassay for the qualitative detection of IgM antibodies to Rubella, and CMV in human serum and EDTA or heparinized plasma using the Luminex flow cytometry technology. It consists of a reagent kit and the Bioplex 2200 Instrument.
The BioPlex 2200 Rubella and CMV IgM Reagent Pack consists of:
1. A bead set containing dyed beads coated with lysates of Rubella, and CMV, an Internal Standard bead (ISB), a Serum Verification bead (SVB), and a Reagent Blank bead (RBB) in buffer with Glycerol and protein stabilizers (bovine), ProClin™ 300 (0.3%) and sodium azide (< 0.1%) as preservatives.
2. Phycoerythrin-conjugated donkey polyclonal anti-human IgM antibody and phycoerythrin -conjugated murine monoclonal anti-human FXIII antibody, in buffer with stabilizers (bovine and equine), ProClin® 300 (0.3%), sodium benzoate (0.1%) and sodium azide (<0.1%) as preservatives.
3. Sample diluent; containing goat anti-human IgG antibody and protein stabilizers (bovine and equine) in buffer, ProClin® 300 (0.3%) and sodium azide (<0.1%) as preservatives.
The device requires other materials which are provided independently from Bio-Rad and these include the BioPlex 2200 Instrument and Software System, the Bioplex 2200 sheath fluid and wash solution.
The BioPlex 2200 Rubella and CMV IgM Calibrator Set consists of three distinct serum based calibrators in three vials. They are made from defibrinated plasma with added known concentrations of Rubella and CMV antibodies from human disease state plasma. The calibrators are used for the qualitative calibration of the device. All calibrators contain ProClin® 300 (0.3%), sodium benzoate (0.1%) and sodium azide (<0.1%) as preservatives.
The BioPlex 2200 Rubella CMV IgM Control Set includes a negative control and a one positive control for Rubella and another for CMV, intended to be used as assayed quality control material with the BioPlex Rubella and CMV IgM Test. All antibodies
3
{3}
are derived from human disease state plasma. All controls contain Amikacin (0.003%), Cycloheximide (C15H23NO4) (0.009%), Amphotericin B (0.002%), Cefotaxime Sodium (0.002%), Ciprofloxacin (0.005%), ProClin® 300 (<0.3%), Sodium benzoate (<0.1%) and sodium azide (<0.1%).
## J. Substantial Equivalence Information:
### 1. Predicate device name(s):
There is no single predicate device for the panel as a whole; the following is a list of the predicate devices for each of the analytes in the panel. These predicate devices are used as comparators to assess the performance of the device and determine the substantial equivalence. The final classification of the new multiplexed device follows that of the Rubella predicate.
ADVIA Centaur® Rubella IgM
bioMérieux VIDAS® CMV IgM
### 2. Predicate 510(k) number(s):
The 510(k) numbers for the Rubella and CMV predicate devices are K010668 and K933549 respectively.
### 3. Comparison with predicates:
1. Comparison with- ADVIA Centaur® Rubella IgM (K010668)
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicates |
| Intended Use | Intended for the qualitative detection of IgM antibodies to Rubella, and Cytomegalovirus (CMV) | Intended for the qualitative detection of IgM antibodies to the rubella virus in serum or plasma (EDTA, heparin) |
| Indications | As an aid in the diagnosis of a current or recent Rubella and/or CMV infection, in individuals suspected of having one of the respective disease states including women of child bearing age. | As an aid in the presumptive diagnosis of current or recent infection with rubella. |
| Measurand | Rubella and CMV IgM | Rubella IgM |
| Detection | Qualitative detection | same |
| Matrices | Serum, EDTA or Heparinized Plasma | same |
| Controls | Negative Control and Positive Controls specific for rubella and CMV IgM | Negative Control and Positive Control specific for rubella IgM |
{4}
2. Comparison with- bioMerieux VIDAS® CMV IgM - K933549
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Intended for the qualitative detection of IgM antibodies to Rubella, and Cytomegalovirus (CMV) | Intended for the qualitative detection of anti-CMV IgM antibodies in human serum. |
| Indications | As an aid in the diagnosis of a current or recent Rubella and/or CMV infection, in individuals suspected of having one of the respective disease states including women of child bearing age. | As an aid in the diagnosis of cytomegalovirus infection. |
| Measurand | Rubella and CMV IgM | CMV IgM |
| Detection | Qualitative detection | same |
| Controls | Negative Control and Positive Controls specific for rubella and CMV IgM | Negative Control and Positive Controls specific for CMV IgM |
| Signal Detection | Fluorescence | same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Matrices | Serum, EDTA or Heparinized Plasma | Serum |
| Analytes detected | Multiple Analytes | Single Analyte |
| Calibrators | Multiple calibrators | Single Calibrator |
| Technology | Multiplexed flow immunoassay, using Antigen-coated paramagnetic microbead reagent. | Two-step enzyme immunoassay sandwich method with fluorescent detection (ELFA) |
| Solid Phase | Antigen-coated paramagnetic | Antigen-coated solid phase |
{5}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | microbead reagent.
Microbeads are infused with red and infrared fluorescent dyes for bead classification. | receptacles |
## K. Standard/Guidance Document referenced (if applicable):
1. CLSI In-vitro guideline; User Demonstration of Performance for Precision and Accuracy; Approved Guideline (EP15-A).
2. CLSI In-vitro guideline; Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition (EP05-A2).
3. CLSI In-vitro guideline; Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition (EP07-A2).
The following FDA guidance documents were also referenced in this submission:
1. Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm078384.htm
2. Guidance for Industry and FDA Staff - Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm071148.htm
## L. Test Principle:
The BioPlex™ 2200 Rubella and CMV IgM kit uses multiplex flow immunoassay, a methodology that greatly resembles traditional EIA, but permits simultaneous detection and identification of many antibodies in a single tube.
The kit employs a panel of two different populations of dyed beads coated with cell lysates bearing Rubella or CMV antigens to identify the presence of IgM antibodies to Rubella, and CMV antigens as follows.
The BioPlex 2200 System combines an aliquot of patient sample, sample diluent, and bead reagent into a reaction vessel; the mixture is incubated at $37^{\circ}\mathrm{C}$. After a wash cycle, anti-human IgM antibody, conjugated to phycoerythrin (PE), is added to the dyed beads, and this mixture is incubated at $37^{\circ}\mathrm{C}$. The excess conjugate is removed in another wash cycle, and the beads are re-suspended in wash buffer. The bead mixture then passes through the detector. The identity of the dyed beads is determined by the fluorescence of the dyes, and the amount of antibody captured by the antigen is determined by the
{6}
fluorescence of the attached PE. Raw data are reported as relative fluorescence intensity (RFI). Additionally, the ISB beads, SVB beads and a RBB beads are present in each reaction mixture to verify detector response, the addition of serum or plasma to the reaction vessel and the absence of significant non-specific binding in serum or plasma.
The BioPlex 2200 Rubella and CMV IgM Calibrator Set calibrates the instrument using a set of three distinct serum based calibrators. Calibrators are used in a test system to establish points of reference that are used in the determination of qualitative numeric measurement of IgM antibodies to Rubella, and CMV in human serum and EDTA or heparinized plasma. The Calibrator Set should be loaded and assayed at minimum in duplicate every 30 days or with each new Re-agent Pack lot.
The BioPlex 2200 Rubella and CMV IgM Control Set includes a negative control as well as a CMV IgM positive control and a Rubella IgM positive control. The BioPlex Rubella and CMV IgM Positive Controls are manufactured to give positive results, with values above the cut-off for each specific analyte. The BioPlex Rubella and CMV IgM Negative Control are manufactured to give negative results, with values below the cut-off for each specific analyte. The recommended frequency for performing quality control is once every 24-hour testing period. Performing quality control is also necessary after each new assay calibration and certain service procedures. A BioPlex 2200 Rubella and CMV IgM Control Lot Data disk is available to load the necessary value assignment data into the instrument. The result for each of the tested antibodies is expressed as an antibody index (AI). For Rubella and CMV, results of ≤ 0.8 AI are negative, 0.9 and 1.0 AI are equivocal and results of ≥ 1.1 AI are reported as positive.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Separate internal and external reproducibility studies were conducted to evaluate the reproducibility of the BioPlex 2200 Rubella and CMV IgM kit on the BioPlex 2200 instrument.
The internal precision study was conducted in-house by Bio-Rad Laboratories in accordance with CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods.
For the internal reproducibility study, three (3) panels made from serum and plasma (EDTA and heparinized) were assayed two (2) times in two (2) separate daily runs over 20 days (n=80).
The data were analyzed for within-run, between-run, between-day, and total precision and the standard deviation (SD) and percent coefficient of variation (% CV) were calculated.
7
{7}
The within-run precision for rubella IgM samples ranged from 4.3% to 5.5% in a serum matrix, 6.3% to 6.9% for EDTA and 4.8% to 9.5% for heparin plasma matrices. The within-run precision for CMV IgM samples ranged from 5.3% to 9.2% in a serum matrix, 4.9% to 11.8% for EDTA and 5.2% to 12.8% for heparin plasma matrices.
The external reproducibility study was conducted in accordance with CLSI EP15-A2 during the external clinical evaluation at each of three clinical study sites. Three lots of reagent packs, three lots of the BioPlex 2200 Rubella and CMV IgM Calibrator Set and three lots of BioPlex 2200 Rubella and CMV IgM Control Set were evaluated. The sample panels made from serum and plasma (EDTA and heparinized) at varying levels of analytes, were tested in quadruplicate over five (5) days (4 replicates x 1 run x 5 days x 3 testing sites = 60 replicates per panel member).
The data were analyzed for within-run, between-run, between-day, between-site, and total precision and the standard deviation (SD) and percent coefficient of variation (% CV) were calculated.
The within run precision for positive samples greater than or equal to the cut-off (≥ 1.1 AI) in all sample matrices ranged from 2.7% to 13.3% for Rubella IgM and 2.8% to 13.5% for CMV IgM. The total precision for positive samples greater than or equal to the cut-off (≥1.1 AI) ranged from 2.3% to 13.5% for Rubella IgM and 3.0% to 14.2% for CMV IgM.
The mean Antibody Index (AI), standard deviation (SD), and percent coefficient of variation (%CV) for each panel member were calculated. The data for the serum panel is represented below:
BioPlex 2200 Rubella IgM Serum
| Rubella IgM Panel Member | Mean (AI) | Site 1 | | Site 2 | | Site 3 | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Within-Run | Total | Within-Run | Total | Within-Run | Total |
| | | %CV | %CV | %CV | %CV | %CV | %CV |
| Negative 1 | 0.2 | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% |
| Negative 2 | 0.2 | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% |
| High Negative 1 | 0.3 | 14.2% | 14.9% | 9.5% | 9.5% | 0.0% | 0.0% |
{8}
| Rubella IgM Panel Member | Mean (AI) | Site 1 | | Site 2 | | Site 3 | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Within-Run | Total | Within-Run | Total | Within-Run | Total |
| | | %CV | %CV | %CV | %CV | %CV | %CV |
| High Negative 2 | 0.5 | 0.0% | 0.0% | 0.0% | 0.0% | 4.4% | 4.4% |
| Low Positive 1 | 1.0 | 7.2% | 7.7% | 4.9% | 5.5% | 6.4% | 6.7% |
| Low Positive 2 | 1.5 | 4.3% | 5.3% | 4.2% | 4.2% | 6.6% | 6.6% |
| High Positive 1 | 1.8 | 5.5% | 5.5% | 2.8% | 2.8% | 4.3% | 5.1% |
| High Positive 2 | 2.0 | 6.3% | 6.3% | 2.7% | 2.7% | 3.7% | 3.9% |
| Positive Control | 1.4 | 3.4% | 3.4% | 6.4% | 6.7% | 4.2% | 4.4% |
BioPlex 2200 CMV IgM Serum
| CMV IgM Panel Member | Mean (AI) | Site 1 | | Site 2 | | Site 3 | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Within-Run | Total | Within-Run | Total | Within-Run | Total |
| | | %CV | %CV | %CV | %CV | %CV | %CV |
| Negative 1 | 0.2 | 0.0% | 0.0% | 15.1% | 17.8% | 10.9% | 10.9% |
| Negative 2 | 0.3 | 14.8% | 14.8% | 9.5% | 9.5% | 7.3% | 7.3% |
| High Negative 1 | 0.8 | 0.0% | 0.0% | 5.0% | 5.3% | 7.1% | 7.1% |
| High Negative 2 | 0.7 | 11.1% | 11.5% | 8.3% | 8.3% | 6.7% | 7.4% |
| Low Positive 1 | 1.5 | 7.2% | 7.2% | 6.4% | 6.4% | 9.9% | 9.9% |
| Low Positive 2 | 1.4 | 12.7% | 12.7% | 5.6% | 5.6% | 4.8% | 5.5% |
| High Positive 1 | 2.1 | 5.6% | 5.9% | 2.8% | 3.0% | 6.6% | 6.6% |
| High Positive 2 | 2.6 | 6.4% | 8.4% | 3.4% | 3.5% | 4.0% | 4.0% |
| Positive Control | 1.8 | 4.6% | 4.6% | 7.2% | 7.2% | 4.5% | 6.0% |
{9}
b. Linearity/assay reportable range:
Not applicable as this assay is a qualitative assay.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The traceability, stability and expected values for the controls and calibrators were evaluated per the manufacturer's protocols and met the acceptance criteria for all product claims.
BioPlex 2200 ToRC IgM Calibrator Set: Calibrator assignment is established from replicate analyses using a master set of calibrators and a specific lot of BioPlex 2200 ToRC IgM Reagent Pack on multiple BioPlex 2200 instruments.
BioPlex 2200 ToRC IgM Control Set: The mean values listed in the package insert for the BioPlex 2200 ToRC IgM Control Set are derived from replicate analyses. It is recommended that individual laboratories establish their own limits for each parameter and use those provided only as guides.
Stability Studies: Stability studies have been performed to determine the open vial and shelf life stability for the BioPlex 2200 ToRC IgM Control and Calibrator Sets. Stability studies were conducted for Rubella, and CMV IgM and Factor XIIIb. The measurement of Factor XIIIb is used to calibrate the serum verification assay, which ensures that serum or plasma is present in the reaction vessels.
d. Detection limit:
Not applicable as this assay is a qualitative assay.
e. Analytical specificity:
Cross-Reactivity
A cross-reactivity study was performed to determine if samples from various disease states and other potentially cross-reacting agents interfere with test results when tested with the BioPlex 2200 Rubella and CMV IgM kit. Samples known to be positive for each of the potential cross-reactants, as determined by FDA cleared devices, were evaluated with the BioPlex 2200 Rubella and CMV IgM assay. All samples were pre-tested by the predicate devices and only those that tested negative by the predicate devices were further evaluated by the BioPlex 2200 Rubella and CMV IgM kit. Table G summarizes the number of samples scored negative by the BioPlex 2200 Rubella and CMV IgM assay within each of the cross-reactant panels. No significant cross-reactivity was observed except for potential cross-reactivity
10
{10}
of Myeloma IgM samples tested with the Rubella and CMV IgM assays, and EBV VCA IgM, Parvovirus B19 IgM and dsDNA samples tested with the CMV IgM assay.
| Potential Cross-Reactant | BioPlex 2200 Rubella and CMV IgM Kit Results | | | |
| --- | --- | --- | --- | --- |
| | N | Rubella IgM | N | CMV IgM |
| ANA Screen | 10 | | 10 | 9 |
| CMV IgM | 10 | 10 | 0 | N/A |
| dsDNA (SLE clinical) | 10 | 10 | 10 | 9b |
| EBV VCA IgM | 10 | 9 | 8 | 6 |
| hCG (pregnancy) | 10 | 10 | 10 | 10a |
| HIV IgG | 10 | 10 | 10 | 10 |
| HSV-1 or 2 IgM | 9 | 8 | 10 | 10b |
| Hypergammaglobulinemia IgM | 10 | 9 | 10 | 10 |
| Influenza (compliment fixation) | 10 | 10 | 10 | 10 |
| Measles IgM | 10 | 10 | 10 | 10 |
| Mumps IgM | 10 | 10 | 10 | 9 |
| Myeloma IgM | 9 | 7 | 7 | 5 |
| Parvovirus B19 IgM | 10 | 10a | 9 | 4 |
| Rheumatoid Factor (Total) | 9 | 9 | 10 | 9a |
| Rubella IgM | 0 | N/A | 10 | 10 |
| T. gondii IgM | 10 | 10 | 10 | 10 |
| VZV IgM | 10 | 10 | 10 | 10 |
a. One BioPlex 2200 equivocal result.
b. Two BioPlex 2200 equivocal results.
Note: The highlighted areas indicate that the potential cross reactivity with either Rubella or CMV IgM could not be ruled out.
## Interfering Substances
Testing for interfering substances was conducted according to CLSI Protocol EP7-A2 (Vol. 25, No. 27). Samples were prepared by blending a pool of negative human serum with samples positive for Rubella or CMV IgM to achieve values of 1.7 to 2.3 AI for Rubella and 1.7 to 2.3 AI for CMV. Interferent or solvent (negative control) was added exogenously at levels indicated in Table H below. Test and control samples were evaluated in replicates of ten. Changes in results ranged from -13.6 to 10.0%. No significant interference was observed in any of the substances tested.
{11}
| Substance | Concentration |
| --- | --- |
| Hemoglobin | ≤ 500 mg/dL |
| Bilirubin (unconjugated) | ≤ 20 mg/dL |
| Bilirubin (conjugated) | ≤ 30 mg/dL |
| Cholesterol | ≤ 500 mg/dL |
| Red Blood Cells | ≤ 0.4% (v/v) |
| Gamma Globulin | ≤ 6 g/dL |
| Triglycerides | ≤ 3300 mg/dL |
| Beta Carotene | ≤ 0.6 mg/dL |
| Protein (total) | ≤ 12 g/dL |
| Ascorbic Acid | ≤ 3 mg/dL |
| Lithium Heparin | ≤ 8000 units/dL |
| Sodium Heparin | ≤ 8000 units/dL |
| EDTA | ≤ 800 mg/dL |
| Sodium Citrate | ≤ 1000 mg/dL |
f. Assay cut-off:
The BioPlex 2200 Rubella and CMV IgM assays report results as an antibody index, (AI). AI values are calculated by using linear regression analysis of a two point assay calibration curve. The cut-off value and assignment of the calibrators are determined by performing concordance and Receiver Operating Characteristic (ROC) analysis, using predicate results as the standard. Analyze-it software is used for the ROC analysis. This software does not employ an equivocal range for the on-test condition. Therefore, predicate equivocal results are not included in the calculations for establishing cut-off. After establishment of the cut-off, an equivocal range is set bracketing the cut-off value. Based on the results, calibrator values were adjusted such that the cut-off values equal to 1.1 AI for Rubella and CMV IgM assays. A total of 586 and 593 patient samples were evaluated to confirm rubella and CMV IgM cut-off values, respectively. ROC Analysis was performed for each analyte using this population of samples.
Results obtained using a cut-off of 1.0 AI in the BioPlex Rubella and CMV IgM assay ROC analyses demonstrated >98% negative agreement and >98% positive
{12}
agreement for CMV and >95% negative agreement and >84% positive agreement for Rubella. The BioPlex 2200 Rubella and CMV IgM assays met the concordance specifications for both analytes.
## 2. Comparison studies:
a. Method comparison with predicate device:
Comparative Testing: Prospective
Performance of the Rubella and CMV IgM kit was evaluated against corresponding commercially available Rubella and CMV IgM immunoassays. Three clinical sites tested 700 prospective samples submitted for: Rubella IgM (300 - U.S.), and CMV IgM testing (400; 300 U.S. + 100 Europe). Of the 300 samples submitted for Rubella IgM testing, 71 females were pregnant women. Results are shown in the Table below.
Equivocal results obtained on the predicate devices were re-tested on two other commercially available EIA and the consensus result of 2 out of the 3 tests was used in the analysis.
| Test Ordered | BioPlex 2200 Rubella and CMV IgM kit | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | Pos (+) | Eqv | Neg (-) | Total | Pos (+) % Agreement | Neg (-) % Agreement |
| Commercially Available Immunoassay | Rubella IgM | Test Ordered | Pos (+) | 2 | 0 | 1 | 3 | 66.7% (2/3) 95% CI 20.8 – 93.9% | 95.6% (282/295) 95% CI 92.6 – 97.4% |
| | | | Eqv | 0 | 2 | 0 | 2 | | |
| | | | Neg (-) | 10 | 3 | 282 | 295 | | |
| | | | Total | 12 | 5 | 283 | 300 | | |
| | | Pregnant Women* | Pos (+) | 0 | 0 | 0 | 0 | N/A | 96.8% (70/71) 95%CI 92.8% - 99.8% |
| | | | Eqv | 0 | 0 | 0 | 0 | | |
| | Neg (-) | | 1 | 0 | 70 | 71 | | | |
| | Total | | 1 | 0 | 70 | 71 | | | |
| | CMV IgM | Test Ordered | Pos (+) | 7a | 0 | 3a | 10 | 53.8% (7/13) 95%CI 29.1 – 76.8% | 97.7% (377/386) 95%CI 95.6 – 98.8% |
| | | | Eqv | 1 | 1 | 3 | 5 | | |
| | | | Neg (-) | 5b | 3 | 377a | 385 | | |
| | | | Total | 13 | 4 | 383 | 400 | | |
{13}
a. One sample that was equivocal by the predicate device was adjudicated by two out of three FDA cleared devices.
b. Two samples that were equivocal by the predicate device were adjudicated by two out of three FDA cleared devices.
* The pregnant women population is a subset of the test ordered population.
## Comparative Testing: Retrospective
Performance of the Rubella and CMV IgM kit was evaluated against corresponding commercially available Rubella and CMV IgM immunoassays. Three clinical sites tested 107 Rubella (45 females with 89% in the 15 - 45 age group, 49 males and 13 with unknown gender) and 229 CMV (144 females with 84% in the 15 - 45 age group and 85 males) presumptive IgM positive samples. The pregnancy status of the preselected samples was not available. Presumed positive banked samples for Rubella and CMV IgM were further selected by the respective predicate device used for the comparative analysis. Results are shown in the Table below.
| Presumptive Positive for Rubella or CMV IgM | BioPlex Rubella and CMV IgM kit | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Pos(+) | Eqv | Neg(-) | Total | Pos(+) % Agreement |
| Commercially Available Immunoassay | Rubella IgM | Pos(+) | 103 | 1 | 3 | 107 | 96.3%
(103/107)
95% CI
90.3 – 98.5% |
| | | Eqv | 0 | 0 | 0 | 0 | |
| | | Neg(-) | 0 | 0 | 0 | 0 | |
| | | Total | 103 | 1 | 3 | 107 | |
| | CMV IgM | Pos(+) | 209^{a} | 3 | 17^{b} | 229 | 91.3%
(209/229)
95% CI
86.9 – 94.3% |
| | | Eqv | 0 | 0 | 0 | 0 | |
| | | Neg(-) | 0 | 0 | 0 | 0 | |
| | | Total | 209 | 3 | 17 | 229 | |
a. One sample was that was equivocal by the predicate device was adjudicated by two out of three FDA cleared devices.
b. Two samples that were equivocal by the predicate device were adjudicated by two out of three FDA cleared devices.
## b. Matrix comparison:
Matched serum and plasma (potassium EDTA and sodium heparin) samples drawn from 20 individual donors were acquired from commercial sources. All
{14}
samples were evaluated in replicates of 10. Mean plasma values were compared to matched mean serum AI values. Scatter plots comparing the performance of serum samples against potassium EDTA and sodium heparin plasma samples along with the corresponding slopes of regression and correlation coefficient (r) are shown in Figures 1-4. All assays pass the slope specification of $1.0 \pm 0.2$ , intercept specification of $\pm 0.2$ , and correlation coefficient (r) of $\geq 0.98$ .

Figure 1. Rubella IgM : Serum vs. EDTA Plasma (N=20)

Figure 2. Rubella IgM : Serum vs. Heparin Plasma (N=20)

Figure 3. CMV IgM : Serum vs. EDTA Plasma (N=20)

Figure 4. CMV IgM : Serum vs. Heparin Plasma (N=20)
# Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
{15}
# Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
# IgM Detection
Rubella and CMV IgM-positive samples were selected and supplemented with matched specific IgG. The sample pools were split and further supplemented with dithiothreitol (DTT) which inactivates IgM activity. The samples were assayed neat and diluted into assay range in replicates of two. IgM was measured using Rubella and CMV IgM kit. The results are shown in below.
BioPlex 2200 Rubella and CMV IgM kit - Rubella IgM Specificity
| Sample | Rubella IgM (AI) Before Treatment | DTT Treatment AI (% recovery) |
| --- | --- | --- |
| Sample1 | 3.6 | 0.2 (5.6%) |
| Sample2 | 3.3 | 0.2 (6.1%) |
| Sample3 | 3 | 0.2 (6.7%) |
| Sample4 | 2.7 | 0.2 (7.4%) |
| Sample5 | 2.6 | 0.2 (7.7%) |
| Sample6 | 2.5 | 0.2 (8%) |
| Sample7 | 2.3 | 0.2 (8.7%) |
| Sample8 | 2.2 | 0.2 (9.1%) |
| Sample9 | 2.2 | 0.2 (9.1%) |
| Sample10 | 2.2 | 0.2 (9.1%) |
BioPlex 2200 Rubella and CMV IgM kit - CMV IgM Specificity
| Sample | CMV IgM (AI)* Before Treatment | DTT Treatment AI (% recovery) |
| --- | --- | --- |
| Sample1 | 460 | 30.2 (6.6%)* |
| Sample2 | 341.5 | 26.4 (7.7%)* |
| Sample3 | 109 | 0.4 (0.4%) |
| Sample4 | 107 | 0.7 (0.7%) |
| Sample5 | 85.5 | 0.7 (0.8%) |
| Sample6 | 37.1 | 0.6 (1.6%) |
| Sample7 | 35.5 | 0.6 (1.7%) |
| Sample8 | 28.1 | 0.3 (1.1%) |
| Sample9 | 23.8 | 0.3 (1.3%) |
| Sample10 | 22.5 | 0.7 (3.1%) |
| *Values derived from dilution | | |
# Seroconversion Testing
# Rubella IgM
{16}
Three Liquichek™ Rubella IgM seroconversion panels obtained from Bio-Rad Laboratories were tested with the BioPlex 2200 Rubella and CMV IgM kit. The results shown in below were compared to a commercial method.
| Panel RP001 | Rubella IgM | |
| --- | --- | --- |
| Day | Commercial Method Index | BioPlex 2200 Rubella and CMV IgM kit AI |
| 0 | 0.20 (Neg) | 0.3 (Neg) |
| 2 | 0.10 (Neg) | 0.2 (Neg) |
| 7 | 0.15 (Neg) | 0.2 (Neg) |
| 9 | 0.00 (Neg) | 0.2 (Neg) |
| 14 | 0.29 (Neg) | 0.3 (Neg) |
| 17 | 0.91 (Eq) | 1.1 (Pos) |
| 21 | 9.57 (Pos) | >4.0 (Pos) |
| 24 | 7.95 (Pos) | >4.0 (Pos) |
| 28 | 1.87 (Pos) | 1.9 (Pos) |
| 31 | 4.92 (Pos) | 3.7 (Pos) |
| 35 | 7.31 (Pos) | 3.1 (Pos) |
| 38 | 3.54 (Pos) | 1.8 (Pos) |
| 42 | 2.43 (Pos) | 1.7 (Pos) |
| 45 | 1.83 (Pos) | 1.3 (Pos) |
| 50 | 1.95 (Pos) | 1.2 (Pos) |
The BioPlex 2200 Rubella and CMV IgM test detected Rubella IgM four days earlier demonstrating greater sensitivity.
| Panel RP011 | Rubella IgM | |
| --- | --- | --- |
| Day | Commercial Method Index | BioPlex 2200 Rubella and CMV IgM kit AI |
| 0 | 0.21 (Neg) | 0.2 (Neg) |
| 3 | 0.00 (Neg) | 0.2 (Neg) |
| 9 | 0.02 (Neg) | 0.2 (Neg) |
| 12 | 0.00 (Neg) | 0.2 (Neg) |
| 16 | 1.64 (Pos) | 1.1 (Pos) |
| 19 | 5.94 (Pos) | >4.0 (Pos) |
| 24 | 8.95 (Pos) | >4.0 (Pos) |
| 27 | 7.69 (Pos) | >4.0 (Pos) |
| 31 | 4.50 (Pos) | 3.7 (Pos) |
| 36 | 2.70 (Pos) | 2.4 (Pos) |
| 39 | 1.78 (Pos) | 1.8 (Pos) |
| 43 | 1.12 (Pos) | 1.2 (Pos) |
| 46 | 0.77 (Neg) | 0.8 (Neg) |
| 50 | 0.43 (Neg) | 0.8 (Neg) |
| 53 | 0.32 (Neg) | 0.6 (Neg) |
| 57 | 0.32 (Neg) | 0.6 (Neg) |
{17}
The BioPlex 2200 Rubella and CMV IgM test showed comparable performance to a commercial method.
| Panel RP014 | Rubella IgM | |
| --- | --- | --- |
| Day | Commercial Method Index | BioPlex 2200 Rubella and CMV IgM kit AI |
| 0 | 0.00 (Neg) | <0.2 (Neg) |
| 5 | 0.00 (Neg) | <0.2 (Neg) |
| 7 | 0.00 (Neg) | <0.2 (Neg) |
| 12 | 0.00 (Neg) | <0.2 (Neg) |
| 14 | 0.23 (Neg) | 0.7 (Neg) |
| 19 | 2.44 (Pos) | 2.3 (Pos) |
| 21 | 2.55 (Pos) | 2.4 (Pos) |
| 26 | 2.58 (Pos) | 2.2 (Pos) |
| 28 | 2.15 (Pos) | 1.8 (Pos) |
| 33 | 1.5 (Pos) | 1.6 (Pos) |
| 35 | 1.5 (Pos) | 1.2 (Pos) |
| 40 | 1.00 (Pos) | 1.0 (Eq) |
| 42 | 0.76 (Neg) | 0.8 (Neg) |
The BioPlex 2200 Rubella and CMV IgM test showed comparable performance to a commercial method, except for the day 40 sample which scored equivocal in the BioPlex 2200 Rubella and CMV IgM test and positive (at the cutoff) in the comparison test.
# CMV IgM
Bio-Rad Laboratories Liquichek™ CMV IgM Seroconversion panel was assayed with BioPlex 2200 Rubella and CMV IgM kits and a commercial method. The results are shown in below. The BioPlex 2200 Rubella and CMV IgM test was able to detect CMV IgM in all the samples compared to a commercial test which scored equivocal at 59 days.
| Panel RP003 | CMV IgM | |
| --- | --- | --- |
| Day | Commercial Methos Index | BioPlex 2200 Rubella and CMV IgM kit AI |
| 1 | 1.16 (Pos) | >4.0 (Pos) |
| 4 | 1.62 (Pos) | >4.0 (Pos) |
| 8 | 2.44 (Pos) | >4.0 (Pos) |
| 51 | 1.16 (Pos) | 3.4 (Pos) |
| 55 | 1.04 (Pos) | 3.0 (Pos) |
{18}
| 59 | 0.84 (Eq) | 2.1 (Pos) |
| --- | --- | --- |
| 65 | 0.85 (Eq) | 2.3 (Pos) |
| 67 | 0.75 (Eq) | 2.0 (Pos) |
| 72 | 0.68 (Neg) | 1.7 (Pos) |
| 74 | 0.62 (Neg) | 1.5 (Pos) |
| 79 | 0.64 (Neg) | 1.8 (Pos) |
| 84 | 0.54 (Neg) | 1.7 (Pos) |
| 88 | 0.72 (Eq) | 2.1 (Pos) |
| 95 | 0.59 (Neg) | 1.7 (Pos) |
| 99 | 0.65 (Neg) | 1.6 (Pos) |
# 4. Clinical cut-off:
See section (1f).
# 5. Expected values/Reference range:
The observed prevalence for the Rubella and CMV IgM using the BioPlex 2200 Rubella and CMV IgM assay was determined using samples submitted for Rubella (300, US) or CMV IgM (400; 300 US +100 Europe) testing. The results are presented in the tables below. The predictive values of the test are dependent on the prevalence. As rubella incidence decreases, the predicative positive value of rubella IgM results decreases.
Note: Each laboratory should establish frequency distributions for their specific patient populations.
Expected values using the BioPlex 2200 Rubella and CMV IgM kit in test ordered population
| Age | Gender | Rubella IgM (US) | | CMV IgM (US) | | CMV IgM (EU) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Pos/Total | % Prevalence | Pos/Total | % Prevalence | Pos/Total | % Prevalence |
| 0 - 10 | F | 0/13 | 0.0% | 0/4 | 0.0% | 0/0 | N/A |
| | M | 0/11 | 0.0% | 0/6 | 0.0% | 0/2 | 0.0% |
| 11 - 20 | F | 0/44 | 0.0% | 1/19 | 5.3% | 0/4 | 0.0% |
| | M | 0/12 | 0.0% | 0/20 | 0.0% | 0/1 | 0.0% |
| 21 - 30 | F | 4/80 | 5.0% | 1/41 | 2.4% | 1/36 | 2.8% |
| | M | 0/8 | 0.0% | 0/10 | 0.0% | 0/2 | 0.0% |
| 31 - 40 | F | 4/55 | 7.3% | 1/34 | 2.9% | 3/42 | 7.1% |
| | M | 0/8 | 0.0% | 3/17 | 17.6% | 0/1 | 0.0% |
{19}
| 41 - 50 | F | 3/26 | 11.5% | 1/35 | 2.9% | 0/6 | 0.0% |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | M | 1/12 | 8.3% | 0/19 | 0.0% | 0/2 | 0.0% |
| 51 - 60 | F | 0/19 | 0.0% | 1/23 | 4.3% | 0/1 | 0.0% |
| | M | 0/9 | 0.0% | 1/33 | 0.0% | 0/1 | 0.0% |
| 61 - 70 | F | 0/1 | 0.0% | 0/16 | 0.0% | 0/1 | 0.0% |
| | M | 0/1 | 0.0% | 0/14 | 0.0% | 0/0 | N/A |
| 71 + | F | 0/0 | N/A | 0/2 | 0.0% | 0/0 | N/A |
| | M | 0/1 | 0.0% | 0/7 | 0.0% | 0/1 | 0.0% |
| Total | | 12/300* | 4.0%** | 9/300 | 3.0% | 4/100 | 4.0% |
*One of pregnant women (N=71) was Rubella IgM positive.
** Since 2001 the incidence of Rubella in the US has been less than 10/1,000,000 population.
## N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.