VIRAZYME INFLUENZA ID TEST FOR INFLUENZA TYPES A AND B VIRUSES
Applicant
Zymetx, Inc.
Product Code
GNX · Microbiology
Decision Date
Sep 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3330
Device Class
Class 1
Indications for Use
The ViraZyme® Influenza ID test is a direct specimen test indicated for use in the qualitative detection of both influenza types A and B virus from throat swab specimens. The ViraZyme® Influenza ID test may be used when a patient is suspected of having symptoms of an influenza-like illness. These symptoms can include, but are not limited to the following: fever of 38.5℃, sore throat, headache, myalgia, rhinitis, vomiting, chills, malaise, and cough. A positive ViraZyme® result would indicate the presence of influenza type A or B virus. A negative result is considered presumptive and should be confirmed by culture. The ViraZyme® Influenza ID test does not detect influenza C, and is indicated for in Vitro Diagnostic Use only.
Device Story
Endogenous viral-encoded enzyme assay (EVEA) for influenza A and B detection; utilizes throat swab specimens. Principle: viral neuraminidase hydrolyzes modified N-acetylneuraminic acid (Neu5Ac) substrate coupled to chromogen; cleavage releases free chromogen, precipitating as blue color on filter device. Used in physician offices, clinics, laboratories, and hospitals; operated by physicians, nurses, or technicians. Visual output (blue precipitate) interpreted by healthcare provider to identify presence of influenza A or B. Negative results require culture confirmation. Benefits: rapid qualitative assessment of influenza-like illness.
Clinical Evidence
Clinical study of 157 throat swab specimens compared ViraZyme test to viral isolation/culture. Results: 62.2% sensitivity for influenza A/B (65.3% for A, 57.6% for B); 98.7% specificity (99.1% for A, 99.2% for B). Reproducibility testing at seven sites (physician offices, clinics, labs, hospitals) showed 100% correlation.
Technological Characteristics
Endogenous viral-encoded enzyme assay (EVEA). Employs modified Neu5Ac substrate coupled to chromogen. Manual filter-based precipitation method. Standalone diagnostic test kit.
Indications for Use
Indicated for qualitative detection of influenza A and B viruses in patients with influenza-like symptoms (e.g., fever, sore throat, headache, myalgia, cough). For in vitro diagnostic use.
Regulatory Classification
Identification
Influenza virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to influenza in serum. The identification aids in the diagnosis of influenza (flu) and provides epidemiological information on influenza. Influenza is an acute respiratory tract disease, which is often epidemic.
Related Devices
K982429 — ZSTATFLU TEST FOR INFLUENZA TYPES A AND B VIRUSES · Zymetx, Inc. · Aug 25, 1998
K063689 — BD DIRECTIGEN EZ FLU A+B TEST · Becton, Dickinson & CO · Mar 7, 2007
K950838 — VIRAZYME CULTURE CONFIRMATION SCREEN FOR INFLUENZA AND PARAINFLUENZA VIRUSES · Zymetx, Inc. · Jun 17, 1996
K042472 — BD DIRECTIGEN EZ FLU A + B · Becton, Dickinson & CO · Jul 7, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
KY71494
ViraZyme® Influenza ID Test for Influenza Types A and B Virus Appendix D - Safety and Effectiveness Summary
SEP 1 0 1997
# Appendix D Safety and Effectiveness Summary for ViraZyme® Influenza ID Test for Influenza Types A and B Virus
This information is provided as a summary of the safety and effectiveness of the ZymeTx, ViraZyme® Influenza ID Test for Influenza Types A and B Viruses. For more detailed information please refer to the product package insert.
The ViraZyme® Influenza ID Test for Influenza Types A and B Viruses is an endogenous viralencoded enzyme assay (EVEA) and is intended for use in the qualitative determination of influenza types A and B from throat swab specimens. The ViraZyme® Influenza ID Test is not intended for the detection of influenza C.
Influenza types A and B virus possess surface glycoproteins with neuraminidase activity, that hydrolyze substrates which contain alpha-ketosidically linked N-acetylneuraminic acid (Neu5Ac). A modified Neu5Ac molecule has been synthesized and coupled to a chromogen to produce the neuraminidase substrate. In the presence of influenza types A and B virus the chromogenic substrate is then cleaved by the action of viral neuraminidase, releasing a free chromogen. This free chromogen precipitates to produce a blue color. The blue precipitate is then concentrated and collected from the reaction mixture onto a filter device.
### PERFORMANCE CHARACTERISTICS
Clinical studies were performed using the ViraZyme® Influenza ID Test for Influenza Types A and B Virus at seven separate locations throughout the United States. A portion of these sites conducted studies comparing the ViraZyme® Influenza ID Test results obtained from standard culture confirmation with monoclonal antibodies. All seven of the sites conducted reproducibility studies to determine that the ViraZyme® Influenza ID Test would perform similarly in various physician offices, laboratories, clinics and hospital settings.
A total of 157 throat swab specimens were collected from field sites during the 1995-96 influenza season between November 11, 1995 to March 29, 1996. Each of these specimens were tested by the ViraZyme® Influenza ID Test and tested by the reference method of viral isolation and culture confirmation with monoclonal antibodies. Six physicians and their nurses and technicians, from two separate physician offices in a Southwest region participated in the collection and testing of ViraZyme® Test. Duplicate throat swab specimens were collected and transported by courier to a Southwestern viral testing laboratory for the culture portion of this testing.
Of the 157 specimens collected and tested, a total of 49/157 were positive by the viral isolation and culture confirmation method for influenza A (31%); 33/157 were positive by the viral isolation and culture confirmation method for influenza B (21%); 1/157 were positive by the viral isolation and culture confirmation method for parainfluenza type 1 (1%); 1/157 were positive by the viral isolation and culture confirmation method for adenovirus (1%); and 73/157 were negative by the viral isolation and culture confirmation method for the respiratory viruses (46%). No parainfluenza type 2 or 3, or respiratory syncytial virus were detected by the viral isolation and culture confirmation method during this study.
{1}------------------------------------------------
# SENSITIVITY AND SPECIFICITY RESULTS Detection of Influenza Types A and B
The ViraZyme® Influenza ID Test detected 51/82 (62.2%) of culture confirmed positives for influenza type A and type B from throat swab specimens. The ViraZyme® Influenza ID Test properly identified 74/75 (98.7%) of culture confirmed negatives for influenza types A and B virus from throat swab specimens.
#### Detection of Influenza Types A
The ViraZyme® Influenza ID Test detected 32/49 (65.3%) of culture confirmed positives for influenza type A and type B from throat swab specimens. The ViraZyme® Influenza ID Test properly identified 107/108 (99.1%) of culture confirmed negatives for influenza type A virus from throat swab specimens.
#### Detection of Influenza Types B
The ViraZyme® Influenza ID Test detected 19/33 (57.6%) of culture confirmed positives for influenza type B from throat swab specimens. The ViraZyme® Influenza ID Test properly identified 123/124 (99.2%) of culture confirmed negatives for influenza type B virus from throat swab specimens.
### VIRAZYME® INFLUENZA ID TEST RESPRODUCIBILITY
There was 100% correlation of results during the reproducibility testing of the ViraZyme® Influenza ID Test at seven physician offices, clinics, research laboratories and hospital settings conducted during the 1996-97 influenza season. We believe that this adequately demonstrates that the ViraZyme® Influenza ID Test will perform similarly in various test environment settings in the hands of various personnel.
### TECHNICAL INFORMATION
For technical information and comments regarding this product, you may contact the ViraZyme® Influenza ID Product Manager at (405) 271-1314.
This Safety and Effectiveness Summary has been provided as a part of the 510(k) notification for the ViraZyme® Influenza ID Test for Influenza Types A and B Virus
Cray D. Shimazaki Ph.D
Craig D. Shimasaki, Ph.D. Vice President of Research
4/23/91
Date
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the eagle in a circular fashion.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
SEP 1 0 1997
Craig D. Shimasaki, Ph.D. ZymeTx, Inc. 800 Research Parkway * Suite 100 Oklahoma City, OK 73104
SEP 10 1997
Re: K971494
Trade Name: VstatFlu™ Test for Influenza Types A and B Virus Regulatory Class: I Product Code: GNX Dated: July 18, 1997 Received: July 21, 1997
Dear Dr. Shimasaki:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in-vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
1 of of a 1 Page:
# 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: ZstatFlu ™ Test for Influenza Types A and B Viruses
#### Indications for use:
The ViraZyme® Influenza ID test is a direct specimen test indicated for use in the qualitative detection of both influenza types A and B virus from throat swab specimens. The ViraZyme® Influenza ID test may be used when a patient is suspected of having symptoms of an influenza-like illness. These symptoms can include, but are not limited to the following: fever of 38.5℃, sore throat, headache, myalgia, rhinitis, vomiting, chills, malaise, and cough. A positive ViraZyme® result would indicate the presence of influenza type A or B virus. A negative result is considered presumptive and should be confirmed by culture. The ViraZyme® Influenza ID test does not detect influenza C, and is indicated for in Vitro Diagnostic Use only.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| Prescription Use:<br>(Per 21 CFR 801.109) | |
|-------------------------------------------|--|
| OR Over-The-Counter Use: | |
| (Optional Format 1-2-96) | |
|-----------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Clinical Laboratory Devices | |
| 510(k) Number | K971494 |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.