ASI Automated ASI RPR Test for Syphilis on the ASI Evolution
K182391 · Arlington Scientific, Inc. (Asi) · GMQ · Nov 30, 2018 · Microbiology
Device Facts
Record ID
K182391
Device Name
ASI Automated ASI RPR Test for Syphilis on the ASI Evolution
Applicant
Arlington Scientific, Inc. (Asi)
Product Code
GMQ · Microbiology
Decision Date
Nov 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3820
Device Class
Class 2
Intended Use
The ASI Automated RPR (rapid plasma reagin) Test for Syphilis, for use on the ASI Evolution Automated Syphilis Analyzer, is a qualitative and semi-quantitative nontreponemal flocculation test for the detection of reagin antibodies in human serum and plasma as a screening test for serological evidence of syphilis. All reactive RPR test samples should be further tested with a treponemal test. The ASI Automated RPR Test for Syphilis is for professional use only. The test is intended to be used for in vitro diagnostic testing and blood donor screening. The ASI Evolution is intended to be used as a fully automated analyzer to objectively interpret the results of the ASI Automated RPR test for Syphilis. The ASI Evolution is designed to provide standardized test interpretation and to provide for storage, retrieval, and transmittal of the test results. It is intended to be acquired, possessed and used only by health care professionals. The ASI Evolution analyzer, in conjunction with the ASI Automated RPR Test for Syphilis is intended to be used for in vitro diagnostic testing and blood donor screening.
Device Story
Automated nontreponemal flocculation test for reagin antibodies; uses carbon-based antigen (cardiolipin, lecithin, cholesterol, charcoal). System automates sample/reagent handling, mixing, and timing in 48-well microtiter plates. Internal CCD camera captures high-resolution images via light reflectance; embedded software analyzes agglutination patterns. Instrument reports qualitative (reactive/non-reactive) and semi-quantitative endpoint titers. Used in clinical laboratories by healthcare professionals. Output provides objective, standardized results; replaces subjective visual interpretation. Benefits include reduced human error, standardized reporting, and efficient high-throughput processing.
Clinical Evidence
Clinical performance evaluated under K173376 using 1068 prospective serum samples, 10 retrospective serum samples, 1003 retrospective plasma samples, 143 syphilis patients, and 250 pregnant women. Analytical validation for semi-quantitative claims performed using a randomized, blinded panel of 10 human serum samples (titers 1:1 to 1:256) tested 80 times each. Results showed 100% agreement within one dilution of known titers (95% CI: 95.49% - 100%) and 100% agreement for non-reactive samples.
Technological Characteristics
Nontreponemal macroscopic flocculation test. Reagents: cardiolipin, lecithin, cholesterol, activated charcoal, phosphate buffer, sodium azide. Instrument: ASI Evolution (integrated digital particle analyzer). Sensing: CCD camera, light reflectance. Connectivity: PC-based software for data storage/retrieval. Software: Pattern recognition algorithm. Sterilization: N/A (reagents).
Indications for Use
Indicated for professional use in the detection of reagin antibodies in human serum and plasma as a screening test for serological evidence of syphilis, including blood donor screening. For prescription use only.
Regulatory Classification
Identification
Treponema pallidum nontreponemal test reagents are devices that consist of antigens derived from nontreponemal sources (sources not directly associated with treponemal organisms) and control sera (standardized sera with which test results are compared) used in serological tests to identify reagin, an antibody-like agent, which is produced from the reaction of treponema microorganisms with body tissues. The identification aids in the diagnosis of syphilis caused by microorganisms belonging to the genus Treponema and provides epidemiological information on syphilis.
Predicate Devices
Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System (K150358)
Related Devices
K173376 — ASI Evolution · Arlington Scientific, Inc. (Asi) · Jun 14, 2018
K201438 — ASI Automated RPR (rapid plasma reagin) Test for Syphilis, for use on the ASI Evolution Automated Syphilis Analyzer · Arlington Scientific, Inc. (Asi) · Oct 21, 2020
K111356 — ASIMANAGER - AT · Arlington Scientific, Inc. · Nov 9, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Indications for Use (Describe)
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{1}------------------------------------------------
Arlington Scientific, Inc.
## 5.0 510(k) Summary
#### 5.1 Preparation Date: 10/15/2018
## Submitted By
David Binks MT (ASCP), MBA COO/Vice President Regulatory Affairs Arlington Scientific, Inc. 1840 North Technology Dr. Springville, UT 84663 Phone 801-489-8911 / Fax 801-489-5552
- 5.2 Trade Name – ASI Automated RPR Test for Syphilis/ASI Evolution
Regulation section: (21 CFR 866.3820) Treponema pallidum nontreponemal test reagents
Classification: Class II
Product Code: GMQ
Panel: Microbiology
Previous ASI Evolution Clearance – K173376
- 5.3 Predicate Device(s) - Gold Standard Diagnostics AIX1000 RPR Automated Test System (K150358)
{2}------------------------------------------------
Arlington Scientific, Inc.
| | Device Similarities and Differences | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | ASI Automated RPR Test for Syphilis on the<br>ASI Evolution | Gold Standard Diagnostics AIX1000<br>RPR Automated Test System |
| Intended Use | The ASI Automated RPR (rapid plasma reagin)<br>Test for Syphilis, for use on the ASI Evolution<br>Automated Syphilis Analyzer, is a qualitative<br>and semi-quantitative nontreponemal<br>flocculation test for the detection of reagin<br>antibodies in human serum and plasma as a<br>screening test for serological evidence of<br>syphilis. All reactive RPR test samples should<br>be further tested with a treponemal test.<br>The ASI Automated RPR Test for Syphilis is for<br>professional use only. The test is intended to be<br>used for <i>in vitro</i> diagnostic testing and blood<br>donor screening.<br>The ASI Evolution is intended to be used as a<br>fully automated analyzer to objectively interpret<br>the results of the ASI Automated RPR test for<br>Syphilis. The ASI Evolution is designed to<br>provide standardized test interpretation and to<br>provide for storage, retrieval, and transmittal of<br>the test results. It is intended to be acquired,<br>possessed and used only by health care<br>professionals. The ASI Evolution analyzer, in<br>conjunction with the ASI Automated RPR Test<br>for Syphilis is intended to be used for <i>in vitro</i><br>diagnostic testing and blood donor screening. | The Gold Standard Diagnostics<br>AIX1000 Rapid Plasma Reagin (RPR)<br>Automated Test System is a non-<br>treponemal flocculation test that can<br>qualitatively determine the presence of<br>reagin antibodies in human serum. It<br>may be used to aid in the diagnosis of<br>syphilis when used in conjunction with<br>supplemental treponemal laboratory<br>tests and other clinical information.<br>This test may also be used to detect<br>non-treponemal antibodies in samples<br>serially diluted to establish titer<br>information. This test is not intended<br>for screening blood or tissue donors. |
| Technology<br>Instruments | The ASI Evolution is an integrated digital<br>particle analyzer designed to objectively<br>interpret certain agglutination tests<br>manufactured by Arlington Scientific, Inc. (ASI).<br>The ASI Evolution fully automates the sample<br>and reagent handling steps of the test<br>procedure. Laboratory professionals use the<br>ASI Evolution to provide standardized test<br>interpretation using criteria that define reactive<br>and nonreactive agglutination reactions.<br>The ASI Evolution employs a camera to create<br>a highly sensitive and high-resolution image of<br>the agglutination immunoassay. This image is<br>then analyzed by the proprietary software<br>algorithm to interpret the agglutination pattern.<br>The ASI Evolution further provides tools that<br>enable the creation, storage, retrieval and<br>transmittal of the test results. | The AIX1000 is a fully automated<br>microtiter plate processor that is able<br>to completely perform sample<br>processing steps, including dilutions,<br>dispenses, and incubations. The AIX<br>uses a probe-mounted camera to take<br>high resolution images to provide<br>photographic measurement and<br>analysis of agglutination. The AIX1000<br>software automatically interprets<br>captured well images and reports them<br>as Non-reactive, Reactive, Invalid, or<br>Skipped, based on a complex<br>interpretation algorithm. |
| Technology Reagents | Nontreponemal Macroscopic Flocculation Test | Same |
| Antigen | Cardiolipin, lecithin, and cholesterol with<br>activated carbon | Same |
| Reported Results | Reactive, Nonreactive, endpoint titers | Same |
| Interpretation | Automated | Same |
| Sample Processing | Automated | Same |
| Sample Type | Serum or Plasma | Serum only |
| Controls | Reactive, Weak Reactive, Nonreactive | Reactive, Nonreactive |
| Test card | 48 well plastic test plate | Same |
| Indications for use | Used for <i>in vitro</i> diagnostic testing and blood<br>donor testing. | Used for <i>in vitro</i> diagnostics Testing |
{3}------------------------------------------------
## 5.4 Device Description - The Automated RPR Test for Syphilis performed on the ASI Evolution is an automated nontreponemal flocculation test for the detection of reqain antibodies in human serum and plasma.
The ASI Evolution is an integrated digital particle analyzer designed to objectively interpret certain slide agglutination tests manufactured by Arlington Scientific, Inc. (ASI). The ASI Evolution fully automates the sample and reagent handling steps of the test procedure. Qualitative and semiquantitative tests are performed by laboratory professionals who use the ASI Evolution to provide standardized test interpretation using criteria that define reactive and nonreactive agglutination reactions.
The ASI Evolution employs a camera that uses light reflectance to create a highly sensitive and high-resolution image of the agglutination immunoassay. This image is then analyzed by the proprietary software algorithm to interpret the agglutination pattern.
The ASI Evolution further provides tools that enable the creation, storage, retrieval and transmittal of the test results.
The ASI Automated RPR Test for Syphilis reagents include the following:
CARBON ANTIGEN - 0.003% cardiolipin, 0.020-0.022% lecithin, 0.09% cholesterol, charcoal (activated) as visual enhancer, phosphate buffer, 0.1% sodium azide as preservative and stabilizers.
CONTROLS (REACTIVE, WEAK REACTIVE, NONREACTIVE) - Human serum or defibrinated plasma (liguid), with 0.1% sodium azide as preservative.
Reagents have two-year expiration dating from date of manufacture. The specific expiration date is located on the label on the vial.
Intended Use – The ASI Automated RPR (rapid plasma reagin) Test for Syphilis, for use on the ASI Evolution Automated Syphilis Analyzer, is a qualitative flocculation test for the detection of reagin antibodies in human serum and plasma as a screening test for serological evidence of syphilis. All reactive RPR test samples should be further tested with a treponemal test. The ASI Automated RPR Test for Syphilis is for professional use only. The test is intended to be used for in vitro diagnostic testing and blood donor screening.
The ASI Evolution is intended to be used as a fully automated analyzer to objectively interpret the results of the ASI Automated RPR Test for Syphilis. The ASI Evolution is designed to provide standardized test interpretation and to provide for storage, retrieval, and transmittal of the test results. It is intended to be acquired, possessed and used only by health professionals. The ASI Evolution analyzer, in conjunction with the ASI Automated RPR Test for Syphilis is intended to be used for in vitro diagnostic testing and blood donor screening.
### ર્સ રે Clinical Data -
Clinical performance of the ASI Automated RPR Test for Syphilis on the ASI Evolution was evaluated under K173376. The study cohort consisted of 1068 prospective serum samples, 10 retrospective serum samples, 1003 retrospective plasma samples, 143 clinically diagnosed syphilis patients, and 250 pregnant women. For additional details and study results, please refer to the K173376 decision summary document.
### 5.6 Analytical Data -
Analytical studies supporting the qualitative detection of reagin antibodies in human plasma and serum by the ASI Automated RPR Test for Syphilis performed on the ASI Evolution were previously submitted and reviewed under K173376. Details pertaining to the study protocols and resulting data are available in the K173376 decision summary document. In addition to the previously conducted studies, validation testing to support the semi-quantitative claim for the determination of reagin antibody endpoint titers in human serum and plasma was performed. A description of the study protocol and a summary of the test results are provided below.
## Endpoint Titer Validation
A randomized and blinded panel of 10 human serum samples with known reagin antibody endpoint titers, as determined by the ASI RPR Card Test for Syphilis on the ASiManager-AT, was tested with the ASI Automated RPR Test for Syphilis on the ASI Evolution using the semi-quantitative test procedure. The reactive samples had titers ranging from 1:1 to 1:256. Each sample panel member was tested a minimum of 80 times, in-house, on at
{4}------------------------------------------------
## Arlington Scientific, Inc.
least five different days by one operator using a single ASI Evolution instrument. To meet the acceptance criteria, all non-reactive samples were required to yield non-reactive test results, while all reactive samples were required to yield test results within one dilution above or below the known titer. The results of the semi-quantitative analysis are shown below.
| Endpoint Titer Results | | | | | | | | | | | | |
|-----------------------------|------------------|---------------|---------|---------|---------|----------|----------|----------|-----------|-----------|-----------|-------------------------------------------------|
| Sample<br>Reactivity | Non-<br>reactive | Neat<br>(1:1) | 1:<br>2 | 1:<br>4 | 1:<br>8 | 1:1<br>6 | 1:3<br>2 | 1:6<br>4 | 1:12<br>8 | 1:25<br>6 | 1:51<br>2 | % Agreement<br>within +/- 1<br>titer (95% C.I.) |
| 06127<br>(Non-<br>reactive) | 80 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 100%<br>(95.49% -<br>100%) |
| N8E23<br>(Non-<br>reactive) | 80 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 100%<br>(95.49% -<br>100%) |
| W6A16R<br>(1:2) | 0 | 7 | 69 | 4 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 100%<br>(95.49% -<br>100%) |
| W8B01R<br>(1:2) | 0 | 0 | 75 | 5 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 100%<br>(95.49% -<br>100%) |
| R7F01R<br>(1:8) | 0 | 0 | 0 | 0 | 76 | 4 | 0 | 0 | 0 | 0 | 0 | 100%<br>(95.49% -<br>100%) |
| R8B01R<br>(1:8) | 0 | 0 | 0 | 32 | 48 | 0 | 0 | 0 | 0 | 0 | 0 | 100%<br>(95.49% -<br>100%) |
# Table 3: Results from Endpoint Titer Validation Study
Non-reactive test results were observed for all 160 non-reactive samples. In addition, all reactive samples produced test results that were within one dilution of the known regain antibody titer for an overall percent agreement of 100%.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.
November 30, 2018
Arlington Scientific, Inc. (ASI) David Binks, COO/VP Regulatory 1840 North Technology Dr. Springville, Utah 84663
Re: K182391
Trade/Device Name: ASI Automated RPR (rapid plasma reagin) Test for Syphilis ASI Evolution Automated Syphilis Analyzer Regulation Number: 21 CFR 866.3820 Regulation Name: Treponema pallidum nontreponemal test reagents Regulatory Class: Class II Product Code: GMQ, JQT Dated: September 4, 2018 Received: September 4, 2018
Dear David Binks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{6}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.