ARCHITECT CORE-M REAGENT KIT/CALIBRATORS/CONTROLS

P060035 · Abbott Laboratories · SEI · Nov 6, 2007 · Microbiology

Device Facts

Record IDP060035
Device NameARCHITECT CORE-M REAGENT KIT/CALIBRATORS/CONTROLS
ApplicantAbbott Laboratories
Product CodeSEI · Microbiology
Decision DateNov 6, 2007
DecisionAPRL
Regulation21 CFR 866.3173
Device ClassClass 2
AttributesPediatric

Indications for Use

The ARCHITECT CORE-M assay is a chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of IgM antibody to hepatitis B core antigen (IgM anti-HBc) in human adult and pediatric serum or plasma (dipotassium EDTA, lithium heparin, and sodium heparin) and neonatal serum. A test for IgM anti-HBc is indicated as an aid in the diagnosis of acute or recent hepatitis B virus (HBV) infection in conjunction with other laboratory results and clinical information.

Device Story

The ARCHITECT CORE-M assay is a two-step chemiluminescent microparticle immunoassay (CMIA) performed on the ARCHITECT i System. It uses paramagnetic microparticles coated with anti-human IgM (mouse, monoclonal) to capture IgM from patient serum or plasma. In the second step, acridinium-labeled recombinant hepatitis B virus core antigen (rHBcAg) binds to the captured IgM anti-HBc. Pre-trigger and trigger solutions are added to initiate a chemiluminescent reaction, measured as relative light units (RLUs). The system compares sample RLUs to a stored calibration curve to determine qualitative results (reactive/nonreactive). Used in clinical laboratories by trained personnel to aid in diagnosing acute or recent HBV infection. Results are interpreted by clinicians alongside other serological markers and clinical information to guide patient management.

Clinical Evidence

Multi-center study evaluated 2,159 specimens (adults, pediatric, acute HBV, pre-selected positives). Compared ARCHITECT CORE-M to a comparator IgM anti-HBc assay. In Population One (US), positive percent agreement was 96.77% (30/31) and negative percent agreement was 98.43% (1,694/1,721). In Population Two (Vietnam), negative percent agreement was 99.64% (276/277). Acute HBV specimens showed 100% (14/14) positive agreement. Pediatric population showed 100% positive agreement (8/8) and 99.15% negative agreement (116/117). Precision and reproducibility established across sites, days, and reagent lots.

Technological Characteristics

CMIA technology. Components: anti-human IgM (mouse, monoclonal) coated paramagnetic microparticles in TRIS buffer; acridinium-labeled recombinant HBcAg conjugate in succinate buffer. Requires ARCHITECT i System analyzer, pre-trigger (hydrogen peroxide), trigger (sodium hydroxide), and wash buffer. Compatible with serum (glass/plastic) and plasma (dipotassium EDTA, lithium heparin, sodium heparin). Automated, networked analyzer platform.

Indications for Use

Indicated for qualitative detection of IgM anti-HBc in human adult, pediatric, and neonatal serum or plasma to aid in diagnosis of acute or recent HBV infection. No known contraindications.

Regulatory Classification

Identification

A hepatitis B virus (HBV) antibody assay is identified as an in vitro diagnostic device intended for prescription use in the detection of antibodies to HBV in human serum, plasma, or other matrices, and as a device that aids in the diagnosis of HBV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis B infection. Results from assays may be qualitative or quantitative, such as quantitative anti-HBs. In addition, results from an anti-HBc IgM (IgM antibodies to core antigen) assay indicating the presence of anti-HBc IgM are indicative of recent HBV infection. Anti-HBs (antibodies to surface antigen) assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or when vaccination status is unknown. The assay is not intended for screening of blood, plasma, cells, or tissue donors. The assay is intended as an aid in diagnosis in conjunction with clinical findings and other diagnostic procedures.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include: (i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors. (ii) A detailed explanation of the principles of operation and procedures for performing the assay. (iii) A detailed explanation of the interpretation of results. (iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate: (A) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other special populations where assay performance may be affected. (B) Detection of HBV antibodies to a single viral antigen indicates a present or past infection with hepatitis B virus, but does not differentiate between acute, chronic, or resolved infection. (C) The specimen types for which the device has been cleared, and that use of the assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results. (D) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures. (E) A non-reactive assay result may occur early during acute infection, prior to development of a host antibody response to infection, or when analyte levels are below the limit of detection of the assay. (F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay. (v) For devices intended for the quantitative detection of HBV antibodies (anti-HBs), in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, labeling required under § 809.10(b) of this chapter must include: (A) The assay calibrators' traceability to a standardized reference material that FDA has determined is appropriate ( *e.g.,* a recognized consensus standard) and the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), linearity, and precision to define the analytical measuring interval.(B) Performance results of the analytical sensitivity study testing a standardized reference material that FDA has determined is appropriate ( *e.g.,* a recognized consensus standard).(2) Design verification and validation must include the following: (i) Detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, and design of the antigen(s) and capture antibody(ies) sequences, rationale for the selected epitope(s), degree of amino acid sequence conservation of the target, and the design and composition of all primary, secondary and subsequent standards used for calibration. (ii) Documentation and characterization ( *e.g.,* supplier, determination of identity, and stability) of all critical reagents (including description of the antigen(s) and capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iii) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance. (iv) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability. (v) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range. (vi) All stability protocols, including acceptance criteria. (vii) When applicable, analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays. (viii) Analytical performance studies and results for determining the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility), including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, specimen stability, reagent stability, and cross-genotype antibody detection sensitivity, when appropriate. (ix) For devices intended for the detection of antibodies for which a standardized reference material (that FDA has determined is appropriate) is available, the analytical sensitivity study and results testing the standardized reference material. Detailed documentation of that study and its results must be provided, including the study protocol, study report, testing results, and all statistical analyses. (x) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review. (xi) Detailed documentation of clinical performance testing from a clinical study with an appropriate number of HBV reactive and non-reactive samples in applicable risk categories and conducted in the appropriate settings by the intended users. Performance must be analyzed relative to an FDA cleared or approved HBV antibody assay or a comparator that FDA has determined is appropriate. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas. (3) For any HBV antibody assay intended for quantitative detection of anti-HBV antibodies, the following special controls, in addition to those special controls listed in paragraphs (b)(1) and (2) of this section, also apply: (i) Detailed documentation of the metrological calibration traceability hierarchy to a standardized reference material that FDA has determined is appropriate. (ii) Detailed documentation of the following analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including upper and lower limits of quantitation (UloQ and LloQ, respectively), linearity using clinical samples, and an accuracy study using the recognized international standard material.

Submission Summary (Full Text)

{0} Page 1 15 # Summary of Safety and Effectiveness Data ## I. GENERAL INFORMATION: **Device Generic Name:** IgM Antibody to Hepatitis B Core Antigen (Anti-HBc IgM) **Device Trade Name:** ARCHITECT® CORE-M™ Reagent Kit ARCHITECT® CORE-M™ Calibrators ARCHITECT® CORE-M™ Controls **Name and Address of Applicant:** Abbott Laboratories Abbott Diagnostics Division 100 Abbott Park Road Abbott Park, IL 60064-3500 **Premarket Approval Application (PMA) Number:** P060035 **Date of Panel Recommendation:** None **Date of Notice of Approval to the Applicant:** November 6, 2007 ## II. INDICATIONS FOR USE: ### Reagent Kit The ARCHITECT CORE-M assay is a chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of IgM antibody to hepatitis B core antigen (IgM anti-HBc) in human adult and pediatric serum or plasma (dipotassium EDTA, lithium heparin, and sodium heparin) and neonatal serum. A test for IgM anti-HBc is indicated as an aid in the diagnosis of acute or recent hepatitis B virus (HBV) infection in conjunction with other laboratory results and clinical information. ### Calibrators The ARCHITECT CORE-M Calibrators are used for the calibration of the ARCHITECT i System when the system is used for the qualitative detection of IgM antibody to hepatitis B core antigen (IgM anti-HBc) using the ARCHITECT CORE-M Reagent Kit. The performance of the ARCHITECT CORE-M Calibrators has not been established with any other IgM anti-HBc assays. {1} Page 2 16 # Summary of Safety and Effectiveness Data ## Controls The ARCHITECT CORE-M Controls are used for monitoring the performance of the ARCHITECT i System when used for the qualitative detection of IgM antibody to hepatitis B core antigen (IgM anti-HBc) in human adult serum and plasma when using the ARCHITECT CORE-M Reagent Kit. The performance of the ARCHITECT CORE-M Controls has not been established with any other IgM anti-HBc assays. ### III. CONTRAINDICATIONS: None known. ### IV. WARNINGS AND PRECAUTIONS: For *in vitro* diagnostic use only. Warnings and precautions for ARCHITECT CORE-M Reagent Kit, ARCHITECT CORE-M Calibrators, and ARCHITECT CORE-M Controls are stated in the respective product labeling. ### V. DEVICE DESCRIPTION: #### Kit Configurations and Components For detection of IgM antibody to hepatitis B core antigen, the ARCHITECT CORE-M Reagent Kit is composed of the following two components: - ARCHITECT CORE-M Microparticles: 1 or 4 Bottle(s) (5.6 mL) Anti-human IgM (mouse, monoclonal) coated microparticles in TRIS buffer with protein (1.0% bovine serum albumin and 2.5% goat IgG) additives. Minimum concentration: 0.12% solids. Preservatives: antimicrobial agents. - ARCHITECT CORE-M Conjugate: 1 or 4 Bottle(s) (5.9 mL) Acridinium-labeled hepatitis B virus core antigen (E. coli, recombinant) conjugate in succinate buffer with protein (2.5% bovine serum albumin and 2.0% bovine calf serum) additives. Minimum concentration: 0.4 µg/mL. Preservatives: antimicrobial agents. {2} Page 3 17 # Summary of Safety and Effectiveness Data In addition, the following components are required for the ARCHITECT CORE-M Reagent Kit: - ARCHITECT i System is an analyzer designed to perform automated immunoassay tests based on the use of CMIA detection technology. - ARCHITECT CORE-M Calibrators, which consists of calibrator 1 and calibrator 2 for the calibration of the instrument. - ARCHITECT CORE-M Controls (or other control material), which consist of a negative control and a positive control. - ARCHITECT i Pre-Trigger Solution contains 1.32% (w/v) hydrogen peroxide. - ARCHITECT i Trigger Solution contains 0.35N sodium hydroxide. - ARCHITECT i Wash Buffer contains phosphate buffered saline solution with preservative. The ARCHITECT CORE-M Calibrators contain: - 1 Bottle (4 mL) of Calibrator 1, which is recalcified IgM anti-HBc negative human plasma. - 1 Bottle (4 mL) of Calibrator 2, which is IgM anti-HBc positive human plasma in recalcified IgM anti-HBc negative human plasma. - ProClin® 950, ProClin 300, and other antimicrobial agents are used as preservatives in Calibrator 1 and Calibrator 2. The ARCHITECT CORE-M Controls contain: - 1 Bottle (4 mL) of Negative Control, which is recalcified IgM anti-HBc negative human plasma. - 1 Bottle (4 mL) of Positive Control, which is IgM anti-HBc positive human plasma in recalcified IgM anti-HBc negative human plasma. - The positive control is blue and contains Acid Blue No. 9 dye. - ProClin 950, ProClin 300, and other antimicrobial agents are used as preservatives in the Negative Control and Positive Control. {3} Page 4 18 # Summary of Safety and Effectiveness Data ## Assay Principle and Format The ARCHITECT CORE-M assay is a two-step immunoassay for the qualitative detection of IgM anti-HBc in human serum and plasma using chemiluminescent microparticle immunoassay (CMIA) technology with flexible assay protocols, referred to as Chemiflex. In the first step, sample is prediluted with wash buffer. The prediluted sample and anti-human IgM (mouse, monoclonal) coated paramagnetic microparticles are combined. Human IgM present in the sample binds to the anti-human IgM (mouse, monoclonal) coated microparticles. After washing, the anti-HBc specific IgM binds to the acridinium-labeled recombinant hepatitis B virus core antigen (rHBcAg) conjugate that is added in the second step. Following another wash cycle, pre-trigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs). A relationship exists between the amount of IgM-anti-HBc in the sample and the RLUs detected by the ARCHITECT i optics. The presence or absence of IgM anti-HBc in the specimen is determined by comparing the chemiluminescent signal in the reaction to the cutoff signal determined from an active ARCHITECT CORE-M calibration curve. ## VI. ALTERNATIVE PRACTICES AND PROCEDURES Patient medical history and thorough physical examinations, including hepatitis serology, determination of liver enzyme levels, and liver biopsy evaluation, will provide further information regarding the status of HBV infection. Alternate procedures for the detection of HBV in human serum and plasma depend on the detection of HBV deoxyribonucleic acid (DNA) by polymerase chain reaction (PCR) assays or nucleic acid testing (NAT), or the detection of HBV antibodies or antigens by commercially available assays that are licensed or approved in the United States. {4} Summary of Safety and Effectiveness Data ## VII. MARKETING HISTORY ARCHITECT CORE-M, List No. 6L23, has not been marketed in any other country. ## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH The ARCHITECT CORE-M assay, together with Calibrators and Quality Control materials, are for *in vitro* diagnostic use, thus there is no direct adverse effect on the patient. Failure of the product to perform as intended or errors in the use of the product may lead to a false result. This assay is used as an aid in the diagnosis of individuals with acute or recent HBV infection in conjunction with other HBV serological markers for the laboratory diagnosis of HBV disease associated with HBV infection. This assay can also be used as an aid in the differential diagnosis in individuals displaying signs and symptoms of hepatitis in whom etiology is unknown. A false nonreactive result does not exclude the possibility of exposure to HBV. A nonreactive result may be due to antibody levels below the detection limits of this assay. Since this assay is used in combination with other HBV assays, a nonreactive result cannot be considered a public health risk, as the individual would be tested with other methodologies if signs and symptoms are indicative of HBV infection. A false reactive result would not be considered a public health risk due to the fact that an individual would be tested with other hepatitis B virus marker assays to define the clinical status of the patient. ## IX. SUMMARY OF NONCLINICAL STUDIES Nonclinical studies were performed at Abbott Laboratories to evaluate the performance characteristics of the ARCHITECT CORE-M assay. The studies are described below. ### Assay Cutoff Determination The presence or absence of IgM antibody to hepatitis B virus core antigen (IgM anti-HBc) in the sample is determined by comparing the chemiluminescent signal in the reaction to the cutoff RLU determined from an active ARCHITECT CORE-M calibration. The ARCHITECT CORE-M assay results are expressed as the ratio of the sample RLU to the cutoff RLU (S/CO). The S/CO is calculated using the equation: $$ \mathrm{S/CO} = \frac{\text{Sample RLU}}{\text{Cutoff RLU}} $$ Page 5 19 {5} Summary of Safety and Effectiveness Data The cutoff signal for the ARCHITECT Anti-HBc IgM assay is calculated by subtracting the Calibrator 1 mean RLU from Calibrator 2 mean RLU and multiplying the result by a constant value (i.e., cutoff multiplier factor of 0.75). The product is then added to the Calibrator 1 mean RLU to achieve the cutoff RLU: $$ \text{Cutoff RLU} = \left[ \frac{\text{Calibrator 2 mean RLU} - \text{Calibrator 1 mean RLU}}{\text{Calibrator 1 mean RLU}} \times 0.75 \right] + \text{Calibrator 1 mean RLU} $$ The ARCHITECT Anti-HBc IgM assay results for a total of 1,433 specimens were used in the ROC analysis. The sensitivity was evaluated using 273 diagnosed acute specimens. The specificity of the assay was evaluated based on 920 normal population and a population consisting of normal (n=920) and chronic (n=240). Based on this analysis, the S/CO research value of 4.0 was selected which yielded 100% sensitivity at a specificity of 100% for sensitivity population and 96.4% specificity for normal+chronic population. The S/CO (research value) of 4.0 was translated into the cut-off multiplier of 0.75 in the ARCHITECT CORE-M assay (List no.6L23). The seroconversion detection sensitivity of the ARCHITECT CORE-M assay (List No. 6L23) was verified utilizing the cutoff multiplier of 0.75 and found to be acceptable when compared to the FDA-approved comparator IgM anti-HBc assay. A clinical investigation was performed for the ARCHITECT CORE-M assay (List No. 6L23). The percent agreement between the ARCHITECT CORE-M assay (List No. 6L23) and the comparator IgM anti-HBc assay was evaluated and the data support the selected assay cutoff and grayzone for the ARCHITECT CORE-M assay (List No. 6L23) and thus the following cutoff calculation: $$ \text{Cutoff RLU} = \left[ \frac{\text{Calibrator 2 mean RLU} - \text{Calibrator 1 mean RLU}}{\text{Calibrator 1 mean RLU}} \times 0.75 \right] + \text{Calibrator 1 mean RLU} $$ Page 6 {6} Summary of Safety and Effectiveness Data ## Tube Type Interference A study was conducted to evaluate which anticoagulants (blood collection tube types) are acceptable for use with the ARCHITECT CORE-M assay. Sample sets of human specimens were collected in the control tube type (plastic serum) and the blood collection tube types selected for evaluation. The blood collection tubes for the sample sets were supplemented with IgM anti-HBc positive plasma to prepare high negative samples (targeted to 0.80 S/CO) and low positive samples (targeted to 1.20 S/CO) and were tested. The data support the use of the following blood collection tube types in the ARCHITECT CORE-M assay: ### Glass tubes - Serum ### Plastic tubes - Serum - Serum separator - Dipotassium EDTA - Sodium heparin - Lithium heparin plasma separator On average, the tube types evaluated showed less than a 10% difference when compared to the control tube type (plastic serum). The distribution of the percent differences per tube type is listed in Table 1 below. Table 1 ARCHITECT CORE-M Tube Type Matrix Comparison Distribution of Absolute Percent Differences | Tube Type | Distribution of Absolute Percent Differences | | | | --- | --- | --- | --- | | | < 10% | ≥ 10% to ≤ 20% | > 20% | | Glass Serum | 87.8% (36/41) | 12.2% (5/41) | 0.0% (0/41) | | Plastic Serum Separator | 82.9% (34/41) | 14.6% (6/41) | 2.4% (1/41) | | Plastic Dipotassium EDTA | 80.5% (33/41) | 17.1% (7/41) | 2.4% (1/41) | | Plastic Sodium Heparin | 82.9% (34/41) | 14.6% (6/41) | 2.4% (1/41) | | Plastic Lithium Heparin Plasma Separator | 80.5% (33/41) | 17.1% (7/41) | 2.4% (1/41) | {7} Summary of Safety and Effectiveness Data ## Sample Stability of Serum and Plasma A study was conducted to evaluate the sample storage temperatures and number of freeze/thaw cycles for each blood collection tube type acceptable for use with the ARCHITECT CORE-M assay. Sample sets of human specimens were collected in each of the blood collection tube types and supplemented with IgM anti-HBc positive stock (targeted at 1.1 S/CO). The samples were tested at baseline (time point 1) and after being stored at 2 to 8°C for ≥ 7 days, at 24 to 30°C for ≥ 3 days, or after being subjected to three freeze/thaw cycles. Specimens that were stored at the 24 to 30°C condition and 2 to 8°C condition were tested from the blood collection tubes, as on the clot represents worst-case condition (i.e. specimen contact with the red blood cells). The specimens that were subjected to the freeze/thaw conditions were tested off the clot. The data demonstrate that human serum (collected in glass or plastic tubes, or plastic serum separator tubes) or plasma collected in dipotassium EDTA, lithium heparin plasma separator, or sodium heparin tubes may be used with the ARCHITECT CORE-M assay when: - stored at 2 to 8°C for up to 7 days - stored at 24 to 30°C (room temperature) for up to 3 days - subjected to up to 3 freeze/thaw cycles ## Sample On Board Stability A study was conducted to evaluate samples when stored on the ARCHITECT i System (on board storage) and tested using the ARCHITECT CORE-M assay. High negative samples (targeted to 0.80 S/CO) and low positive samples (targeted to 1.20 S/CO) were tested using one lot of reagents, one lot of calibrators, and one lot of controls on two instruments (one i 2000 and one i 2000SR). Time point 1 consisted of testing the two analyte levels immediately after pipetting the samples. Time point 2 consisted of testing the two analyte levels after being stored on board the instrument for at least 3 hours. The data support sample storage of up to 3 hours on board the ARCHITECT i System when tested using the ARCHITECT CORE-M assay. 22 {8} Page 9 23 # Summary of Safety and Effectiveness Data ## Within-Laboratory Precision (20-day Precision) A 20-day precision study was conducted to evaluate the precision performance of the ARCHITECT CORE-M assay based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP5-A2 and to confirm that the ARCHITECT CORE-M assay can be used on the ARCHITECT i 2000 and i 2000SR. Testing was performed using three ARCHITECT CORE-M reagent lots, three calibrator lots, and one control lot on two instruments (one i 2000 and one i 2000SR). The ARCHITECT CORE-M Negative Control (NC) and Positive Control (PC), high negative panel (targeted to 0.80 S/CO) and low positive panel (targeted to 1.20 S/CO) were assayed in replicates of three at two separate times of day for 20 testing days. The ARCHITECT CORE-M assay demonstrated acceptable precision. The data confirm that the ARCHITECT CORE-M assay can be used on the ARCHITECT i 2000 and i 2000SR. The results are summarized in Table 2 on page 11. {9} Page 10 Summary of Safety and Effectiveness Data Table 2 ARCHITECT CORE-M Precision (20-Day) Overall Precision – Three Reagent Lots | Instrument | Sample | n | Mean S/CO | Within-Run | | Within-Laboratory Precision (Total) | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | SD | %CV | SD | %CV | | 1 | Positive Control | 360 | 3.20 | 0.127 | 4.0 | 0.137 | 4.3 | | | Low positive panel | 359 | 1.21 | 0.049 | 4.0 | 0.052 | 4.3 | | | High negative panel | 359 | 0.83 | 0.032 | 3.9 | 0.035 | 4.2 | | | Negative Control | 360 | 0.04 | 0.005 | N/A | 0.005 | N/A | | 2 | Positive Control | 360 | 3.13 | 0.131 | 4.2 | 0.141 | 4.5 | | | High negative panel | 359 | 0.80 | 0.035 | 4.4 | 0.040 | 5.0 | | | Low positive panel | 358 | 1.18 | 0.052 | 4.4 | 0.057 | 4.8 | | | Negative Control | 356 | 0.03 | 0.005 | N/A | 0.005 | N/A | N/A = not applicable 34 {10} Page 11 25 # Summary of Safety and Effectiveness Data ## Analytical Specificity A study was conducted to evaluate the ARCHITECT CORE-M assay for potential cross-reactivity with specimens from individuals with medical conditions unrelated to HBV infection. Specimens with various medical conditions were tested using the ARCHITECT CORE-M assay and the comparator IgM anti-HBc assay. The final results for each of the specimens were compared between the two assays. Specimen results that were discordant between the two assays were sent to an external reference laboratory for supplemental testing with FDA-approved assays. For the medical conditions evaluated, the ARCHITECT CORE-M assay demonstrated no potential cross-reactivity with specimens from individuals with medical conditions unrelated to HBV. The data are summarized in Table 3 on page 13. {11} Summary of Safety and Effectiveness Data Table 3 ARCHITECT CORE-M versus Comparator IgM anti-HBc Assay Final Results by Category | | Comparator IgM Anti-HBc Assay | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Nonreactive | | | Reactive | | | | | | ARCHITECT CORE-M | | | ARCHITECT CORE-M | | | | Category | Na | NRb | GZb | Rb | NRb | GZb | Rb | | Anti-nuclear antibody (ANA) | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Cytomegalovirus (anti-CMV positive) | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Elevated IgG | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Elevated IgM | 5 | 5 | 0 | 0 | 0 | 0 | 0 | | Epstein-Barr virus (anti-EBV positive) | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | HBV vaccine recipient | 8 | 8 | 0 | 0 | 0 | 0 | 0 | | Hepatitis A virus (anti-HAV IgM positive) | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Hepatitis C virus (anti-HCV positive) | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Herpes simplex virus (anti-HSV positive) IgG | 4 | 4 | 0 | 0 | 0 | 0 | 0 | | Human anti-mouse antibodies (HAMA) positive | 7 | 7 | 0 | 0 | 0 | 0 | 0 | | Human immunodeficiency virus (anti-HIV-1 positive) | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Human immunodeficiency virus (anti-HIV-2 positive) | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Influenza vaccine recipient | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Multiparous female | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Multiple myeloma | 2 | 2 | 0 | 0 | 0 | 0 | 0 | | Mumps virus | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Non-Hodgkin's lymphoma | 6 | 6 | 0 | 0 | 0 | 0 | 0 | | Non-viral liver disease | 12 | 12 | 0 | 0 | 0 | 0 | 0 | | Rheumatoid factor positive | 10 | 9 | 0 | 0 | 1c | 0 | 0 | | Rubella | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Rubeola virus | 9. | 9 | 0 | 0 | 0 | 0 | 0 | | Syphilis | 10 | 10 | 0 | 0 | 0 | 0 | 0 | | Systemic lupus erythematosus (SLE) | 9 | 9 | 0 | 0 | 0 | 0 | 0 | | Toxoplasmosis IgG positive | 9 | 9 | 0 | 0 | 0 | 0 | 0 | | Varicella zoster virus (anti-VZV positive) | 4 | 4 | 0 | 0 | 0 | 0 | 0 | | Yeast Infection | 7 | 7 | 0 | 0 | 0 | 0 | 0 | | Total | 222 | 221 | 0 | 0 | 1 | 0 | 0 | a Number of specimens tested per category b NR = Nonreactive, GZ = Grayzone, R = Reactive c This specimen was tested and determined to be reactive for HBsAg, but did not confirm; negative for total anti-HBc; and positive for anti-HBs. A second FDA-approved IgM anti-HBc assay was performed and the specimen was determined to be negative. 26 {12} Summary of Safety and Effectiveness Data ## Interferences – Bilirubin, Hemoglobin, Total Protein, and Triglycerides A study was conducted to evaluate the susceptibility of the ARCHITECT CORE-M assay to potentially interfering substances based on guidance from the CLSI document EP7-A2. A bilirubin test sample was prepared by supplementing the high negative and low positive samples with bilirubin (conjugated and unconjugated) at &gt; 20 mg/dL (targeted to 22 mg/dL). A hemoglobin test sample was prepared by supplementing the high negative and low positive samples with hemolysate at &gt; 500 mg/dL (targeted to 550 mg/dL). A high protein test sample (&gt; 12 g/dL [targeted to 13.2 g/dL]) was prepared by concentrating a nonreactive, normal protein specimen and supplementing with IgM anti-HBc positive stock to yield two test samples with different analyte levels (0.80 and 1.20 S/CO). A triglyceride test sample was prepared by supplementing the high negative and low positive samples with Liposyn® III at &gt; 3000 mg/dL (targeted to 3300 mg/dL). Reference samples were prepared for each test sample at each analyte level. The reference and test samples were tested. At the concentrations listed below, bilirubin, hemoglobin, total protein, and triglycerides showed less than 10% interference in the ARCHITECT CORE-M assay for high negative samples targeted to 0.80 S/CO (S/CO range: 0.60 to 0.99) and low positive samples targeted to 1.20 S/CO (S/CO range: 1.00 to 1.40): - Bilirubin (≤ 20 mg/dL) - Hemoglobin (≤ 500 mg/dL) - Total Protein (≤ 12 g/dL) - Triglycerides (≤ 3000 mg/dL) Page 13 27 {13} Summary of Safety and Effectiveness Data ## Seroconversion Panels The ability of the ARCHITECT CORE-M assay to detect IgM anti-HBc was evaluated by testing eight seroconversion panels obtained from two commercial vendors. The results were compared to the results of an FDA-approved IgM anti-HBc assay (reference). IgM anti-HBc was detected by ARCHITECT CORE-M coincident with the reference IgM anti-HBc assay in eight panels. The profiles of the eight seroconversion panels were characteristic of an acute HBV infection progressing to eventual recovery and immunity to HBV. ARCHITECT CORE-M detected IgM anti-HBc following detection of HBsAg in all panels during the acute stage of disease. IgM anti-HBc remained detectable over a range of two to ten months in the eight panels. The overall ARCHITECT CORE-M results were consistent with the known serological profile of each panel. ## Neonate Serum A study was conducted to evaluate whether neonate samples may be tested with the ARCHITECT CORE-M assay. Cord blood serum was used as a surrogate for neonate serum. Twenty-one matched cord blood and maternal serum samples were spiked with IgM anti-HBc positive stock to yield a high negative sample (target S/CO 0.80) and a low positive sample (target S/CO 1.20). None of the samples were initially reactive. The data obtained upon spiking are summarized in the following table, showing the amount of bias for the cord blood serum samples from the matched maternal serum samples. For cord blood serum samples with &gt;10% bias, one sample exhibited negative bias and the remaining samples exhibited positive bias when compared to the matched maternal serum samples. The results are presented in Table 4 below. | Analyte Level S/CO | Distribution of % Bias | | | | | --- | --- | --- | --- | --- | | | < 10% | ≥ 10% to < 20% | ≥ 20% to < 30% | ≥ 30% | | 0.80 | 66.7% (14/21) | 28.6% (6/21) | 4.8% (1/21) | 0.0% (0/21) | | 1.20 | 52.4% (11/21) | 38.1% (8/21) | 9.5% (2/21) | 0.0% (0/21) | Page 14 {14} Summary of Safety and Effectiveness Data ## Calibration Curve Storage A study was conducted to evaluate the acceptability of an ARCHITECT CORE-M calibration curve stored on the ARCHITECT i System for a minimum of 30 days and to confirm that the ARCHITECT CORE-M assay can be used on the ARCHITECT i 2000 and i 2000SR. Testing was performed using three ARCHITECT CORE-M reagent lots, three calibrator lots, and one control lot on two instruments (one i 2000 and one i 2000SR). Each reagent lot was matched with a different calibrator lot. Calibrations were performed on each instrument for each reagent lot and the calibration curve generated was stored on each instrument for the duration of the study. The ARCHITECT CORE-M Negative Control and Positive Control were assayed in replicates of three, at two times per day, for a total of 20 time points across a minimum of 31 days. The last time point was performed at 34 days after calibration. The data support the storage of an ARCHITECT CORE-M calibration curve on the ARCHITECT i System for a minimum of 30 days. The data confirm that the ARCHITECT CORE-M assay can be used on the ARCHITECT i 2000 and i 2000SR. ## High Dose Hook Effect A study was conducted to characterize the performance of the ARCHITECT CORE-M assay when used to test specimens containing high levels of IgM anti-HBc that have the potential to cause a high dose hook effect. Two unique stocks of IgM anti-HBc positive human were used for the study, where at least one of the stocks had an S/CO value of &gt;10 after a minimum of a 1:8 dilution. The two IgM anti-HBc positive human plasma stocks were each serially diluted with recalcified IgM anti-HBc negative human plasma and tested on the ARCHITECT i System. The data demonstrate that the ARCHITECT CORE-M assay is not susceptible to interference from specimens with high levels of IgM anti-HBc. Page 15 29 {15} Page 16 30 # Summary of Safety and Effectiveness Data ## Instrument Percent Agreement A study was conducted to confirm that the ARCHITECT CORE-M assay can be used on the ARCHITECT i 2000 and i 2000SR systems. One hundred and one IgM anti-HBc negative specimens and 162 IgM anti-HBc positive specimens were tested on two instruments (one i 2000 and one i 2000SR) using a minimum of two lots of reagents and one lot of calibrators and controls. One replicate of each specimen was tested with the same reagent lot on both instruments. The negative percent agreement was 100.0% with a 95% confidence interval of 96.41% to 100.00%. The positive percent agreement was 100.0% with a 95% confidence interval of 97.75% to 100.00%. The ARCHITECT CORE-M assay demonstrated acceptable agreement between the ARCHITECT i 2000 and i 2000SR. ## Within-Assay Sample Carryover A study was conducted to evaluate the susceptibility of within-assay sample carryover within the ARCHITECT CORE-M assay by comparing the results of a low IgM anti-HBc sample when tested before (protected) and after testing a high IgM anti-HBc sample (unprotected). The difference between the protected low sample and the unprotected low sample mean S/CO values was 0.02 S/CO, indicating that no within-assay sample carryover was present. Therefore, the ARCHITECT CORE-M assay is not susceptible to within-assay sample carryover. {16} Page 17 31 # Summary of Safety and Effectiveness Data ## Reagent, Calibrator, and Control Developmental Stability The developmental stability is an on-going study to establish the stability (shelf-life integrity) of the ARCHITECT CORE-M Reagents, Calibrators, and Controls at the intended storage condition of 2 to 8°C and during on board storage (for reagents only). In addition, the developmental stability includes the in-use and freeze/thaw conditions. The in-use condition for the reagents, calibrators, and controls simulates customer use over time. The freeze/thaw condition for the reagents, calibrators, and controls supports the transport simulation studies described below. Stability testing is performed on three lots of reagents, calibrators, and controls. The developmental stability is scheduled to continue for a maximum of 15 months (with a minimum of 6 months). To date, the above stability conditions meet the stability action limits over the time period tested. ## Reagent Transport Stability A study was conducted to support the stability of the ARCHITECT CORE-M Reagents following simulated transport stress conditions. One lot of the ARCHITECT CORE-M Reagents was tested after being subjected to simulated transport stress. The data support the stability of the ARCHITECT CORE-M Reagents following transport at ambient temperatures. {17} Page 18 32 # Summary of Safety and Effectiveness Data ## Calibrator and Control Transport Stability A study was conducted to support the stability of the ARCHITECT CORE-M Calibrator and Controls following simulated transport stress conditions. One lot each of the ARCHITECT CORE-M Calibrators and Controls was tested after being subjected to simulated transport stress. The data support the stability of the ARCHITECT CORE-M Calibrators and Controls following transport at ambient temperatures. ## ARCHITECT CORE-M Microbial Challenge Characterization A Microbial Challenge Characterization (MCC) evaluation was performed for the ARCHITECT CORE-M Reagents, Calibrators, and Controls, which consisted of an Antimicrobial Effectiveness Testing (AET) evaluation and a Microbial Interference Characterization (MIC) evaluation. The MCC evaluation integrated the results from both AET and MIC, which determined that the product is adequately protected. {18} Summary of Safety and Effectiveness Data ## X. SUMMARY OF CLINICAL STUDIES A multi-center study was conducted to evaluate the efficacy of the ARCHITECT CORE-M assay for the qualitative detection of IgM anti-HBc in human serum and plasma as measured by precision and method comparison. ### System Reproducibility (5-day Precision) A study was conducted to validate the precision performance of the ARCHITECT CORE-M assay based on guidance from the CLSI document EP15-A2.²¹ Three lots of ARCHITECT CORE-M Reagents, Calibrators, and Controls were tested per site. The ARCHITECT CORE-M Negative Control and Positive Control, and a high negative panel member (Panel 1) (targeted to 0.80 S/CO) and low positive panel member (Panel 2) (targeted to 1.20 S/CO) were assayed in replicates of four at two separate times per day for five days. The data are summarized in Table 5 on page 21 and in Table 6 on page 22. Page 19 33 {19} Page 20 34 # Table 5 ## ARCHITECT CORE-M ### System Reproducibility (5-Day Precision) ### All Sites, All Reagent Lots ### Individual Variance Components | Sample | N | Grand Mean S/CO | Within-Run | | Between-Run | | Between-Day | | Total^{a} | | | Between-Lot | | Between-Site | | Overall^{b} | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | %CV Upper CL^{c} | SD | %CV | SD | %CV | SD | %CV | | Panel 1 | 360 | 0.83 | 0.036 | 4.3 | 0.000 | 0.0 | 0.000 | 0.0 | 0.036 | 4.3 | 4.6 | 0.036 | 4.4 | 0.027 | 3.3 | 0.056 | 6.8 | | Panel 2 | 360 | 1.21 | 0.049 | 4.0 | 0.006 | 0.5 | 0.012 | 1.0 | 0.051 | 4.2 | 4.5 | 0.051 | 4.2 | 0.035 | 2.9 | 0.078 | 6.5 | | Negative Control | 360 | 0.03 | 0.005 | 14.0 | 0.000 | 0.0 | 0.001 | 2.6 | 0.005 | 14.3 | 15.2 | 0.001 | 2.8 | 0.002 | 7.2 | 0.005 | 16.1 | | Positive Control | 360 | 3.19 | 0.115 | 3.6 | 0.003 | 0.1 | 0.038 | 1.2 | 0.121 | 3.8 | 4.0 | 0.110 | 3.4 | 0.106 | 3.3 | 0.187 | 5.8 | a Total variability contains within-run, between-run and between-day variance components. b Overall variability contains within-run, between-run, between-day, between-lot, between-site and lot-site interaction variance components. c One-sided upper 95% confidence limit for % CV with degrees of freedom calculated by Satterthwaite's method. {20} Page 21 Summary of Safety and Effectiveness Data # Table 6 ## ARCHITECT CORE-M System Reproducibility (5-Day Precision) All Sites, All Reagent Lots Cumulative Variance Components | Sample | n | Grand Mean S/CO | Within-Run | | Within-Day | | Within-Laboratory Precision (Total) | | Precision with Additional Component of Between-Site^{a} | | Precision with Additional Component of Between-Lot^{a} | | Precision with Additional Components of Site and Lot (Overall)^{a} | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | Positive Control | 360 | 3.19 | 0.115 | 3.6 | 0.115 | 3.6 | 0.121 | 3.8 | 0.160 | 5.0 | 0.163 | 5.1 | 0.187 | 5.8 | | Low Positive Panel | 360 | 1.21 | 0.049 | 4.0 | 0.049 | 4.1 | 0.051 | 4.2 | 0.061 | 5.1 | 0.072 | 5.9 | 0.078 | 6.5 | | High Negative Panel | 360 | 0.83 | 0.036 | 4.3 | 0.036 | 4.3 | 0.036 | 4.3 | 0.045 | 5.4 | 0.051 | 6.2 | 0.056 | 6.8 | | Negative Control | 360 | 0.03 | 0.005 | N/A | 0.005 | N/A | 0.005 | N/A | 0.005 | N/A | 0.005 | N/A | 0.005 | N/A | N/A = not applicable * Includes site-lot interaction variance component. 35 {21} Summary of Safety and Effectiveness Data ## Method Comparison ### Study Overview and Subject Population A prospective multi-center study was conducted to evaluate the ability of the ARCHITECT CORE-M assay to detect IgM anti-HBc antibodies in specimens from an intended use diagnostic population. Of the 2,159 specimens tested and analyzed in the ARCHITECT CORE-M clinical study, 1,207 specimens were from individuals at increased risk of HBV infection due to lifestyle, behavior, occupation, disease state, or known exposure event and 545 specimens were from individuals exhibiting signs and symptoms of hepatitis infection living in the United States (Population One); 94 specimens were from individuals at increased risk of HBV infection due to lifestyle, behavior, occupation, disease state, or known exposure event and 183 specimens were from individuals exhibiting signs and symptoms of hepatitis infection living in Vietnam (Population Two); 14 specimens were from individuals diagnosed with acute HBV infection; 16 specimens were pre-selected IgM anti-HBc positive specimens; and 100 surplus specimens were from a pediatric population. The 2,159 specimens were collected from specimen collection sites or were purchased from specimen vendors. Each specimen was tested using the ARCHITECT CORE-M assay at one of the three clinical testing sites located in Galveston, TX; Hershey, PA; and Milwaukee, WI. Each specimen was also tested with the comparator IgM anti-HBc assay at an external laboratory. Specimens from Population One, Population Two, and the acute and pre-selected IgM anti-HBc populations were also tested with three HBV reference assays. The comparator and reference assays were from a single manufacturer and during the clinical study, all comparator and reference testing was performed according to manufacturer’s instructions. HBV classification was then determined using the results from the HBV reference markers and a modification of the serological criteria established by the National Center of Infectious Disease (CDC) for diagnosing HBV infection, which is presented in Table 7 on page 24. Nineteen unique reference marker patterns are represented. Page 22 36 {22} Summary of Safety and Effectiveness Data Table 7 HBV Classification | HBV Reference Markers | | | | HBV Classification | | --- | --- | --- | --- | --- | | HBsAg | Anti-HBc IgM | Total Anti-HBc | Anti-HBs | | | + | – | – | – | Early Acute | | + | + | + | – | Acute | | + | + | + | I | Chronic | | + | – | + | + | Chronic | | + | – | + | – | Chronic | | + | – | – | + | Chronic | | + | – | + | I | Chronic* | | + | + | + | + | Late Acute/Recovering* | | – | + | + | + | Recovering Acute | | – | + | + | – | Recovering Acute/Undetectable HBsAg | | – | + | – | + | Recovering Acute† | | – | + | – | – | Possible Recovering Acute/Undetectable HBsAg† | | – | + | + | I | Early Recovery* | | – | – | + | + | Immune Due to Natural Infection | | – | – | + | I | Distantly Immune/Anti-HBs Unknown | | – | – | + | – | Distantly Immune/ Anti-HBs Not Detected | | – | – | – | + | Immune Due to HBV Vaccination | | – | – | – | I | Unknown | | – | – | – | – | Susceptible | + = Positive/Reactive; – = Negative; I = Indeterminate † Serological marker pattern was not observed during the clinical evaluation. * Three additional serological marker patterns were observed during the clinical evaluation. Page 23 37 {23} Page 24 38 # Summary of Safety and Effectiveness Data The ARCHITECT CORE-M results were compared to the comparator IgM anti-HBc results. For specimens that were discordant between the ARCHITECT CORE-M assay and the comparator IgM anti-HBc assay, supplemental testing was performed to better characterize the specimens. Supplemental testing was performed on different specimen aliquots at external reference laboratories. ## Results by Specimen Classification Following testing with the comparator IgM anti-HBc assay and the three reference HBV assays, specimens from Population One, Population Two, and the acute and pre-selected IgM anti-HBc populations were assigned an HBV classification using the reactive (+) and nonreactive (-) patterns. The 17 unique reference marker patterns observed in the ARCHITECT CORE-M clinical study for Population One are presented in Table 8 on page 26. The 10 unique reference marker patterns observed in the ARCHITECT CORE-M clinical study for Population Two are presented in Table 9 on page 27. Acute status was determined for all of the specimens in the acute and pre-selected IgM anti-HBc populations. {24} Summary of Safety and Effectiveness Data Table 8 HBV Classification for Increased Risk and Signs and Symptoms Population (Population One) | HBV Classification | HBV Reference Markers | | | | | | --- | --- | --- | --- | --- | --- | | | HBsAg | Anti-HBc IgM | Total Anti-HBc | Anti-HBs | N | | Early Acute | + | - | - | - | 8 | | Acute | + | + | + | - | 17 | | Chronic | + | + | + | 1 | 1 | | Chronic | + | - | + | + | 2 | | Chronic | + | - | + | - | 51 | | Chronic | + | - | - | + | 3 | | Recovering Acute | - | + | + | + | 7 | | Recovering Acute/Undetectable HBsAg | - | + | + | - | 2 | | Immune Due to Natural Infection | - | - | + | + | 220 | | Distantly Immune/Anti-HBs Unknown | - | - | + | 1 | 34 | | Distantly Immune/Anti-HBs Not Detected | - | - | + | - | 107 | | Immune Due to HBV Vaccination | - | - | - | + | 351 | | Susceptible | - | - | - | - | 897 | | Late Acute/Recovering | + | + | + | + | 1 | | Chronic | + | - | + | 1 | 3 | | Early Recovery | - | + | + | 1 | 3 | | Unknown | - | - | - | 1 | 45 | | Total | | | | | 1752 | I = Indeterminate {25} Summary of Safety and Effectiveness Data Table 9 HBV Classification for Increased Risk and Signs and Symptoms Population (Population Two) | HBV Classification | HBV Reference Markers | | | | | | --- | --- | --- | --- | --- | --- | | | HBsAg | Anti-HBc IgM | Total Anti-HBc | Anti-HBs | N | | Early Acute | + | - | - | - | 1 | | Chronic | + | - | + | + | 3 | | Chronic | + | - | + | - | 107 | | Chronic | + | - | - | + | 1 | | Immune Due to Natural Infection | - | - | + | + | 67 | | Distantly Immune/Anti-HBs Unknown | - | - | + | 1 | 5 | | Distantly Immune/Anti-HBs Not Detected | - | - | + | - | 12 | | Immune Due to HBV Vaccination | - | - | - | + | 41 | | Susceptible | - | - | - | - | 37 | | Chronic | + | - | + | 1 | 3 | | Total | | | | | 277 | I = Indeterminate ## Comparison of Results The ARCHITECT CORE-M assay results were compared to the comparator IgM anti-HBc assay results for Population One and Population Two. The data are presented in the Table 10 beginning on page 28 for Population One and in Table 11 on page 30 for Population Two. Page 26 {26} Table 10 ARCHITECT CORE-M Results versus Comparator IgM Anti-HBc Results Comparison for Increased Risk and Signs and Symptoms Population (Population One) by HBV Classification | HBV Classification | Comparator IgM Anti-HBc | | | | | | | | | | | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Reactive | | | | | | Negative | | | | | | | | | | ARCHITECT CORE-M | | | | | | ARCHITECT CORE-M | | | | | | | | | | Reactive | | Grayzone | | Nonreactive | | Reactive | | Grayzone | | Nonreactive | | | | | | N | % | N | % | N | % | N | % | N | % | N | % | N | % | | Early Acute | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 8 | 0.46 | 8 | 0.46 | | Acute | 16 | 0.91 | 1^{a} | 0.06 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 17 | 0.97 | | Chronic | 1 | 0.06 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 4^{d} | 0.23 | 55 | 3.14 | 60 | 3.42 | | Recovering Acute | 7 | 0.40 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 7 | 0.40 | | Recovering Acute/Undetectable HBsAg | 2 | 0.11 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00^{e} | 0 | 0.00 | 0 | 0.00 | 2 | 0.11 | | Immune Due to Natural Infection | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 14^{b} | 0.80 | 5^{e} | 0.29 | 201 | 11.47 | 220 | 12.56 | | Distantly Immune/Anti-HBs Unknown | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 1^{f} | 0.06 | 33 | 1.88 | 34 | 1.94 | | Distantly Immune/Anti-HBs Not Detected | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 2^{c} | 0.11 | 1^{f} | 0.06 | 104 | 5.94 | 107 | 6.11 | | Immune Due to HBV Vaccination | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 351 | 20.03 | 351 | 20.03 | | Susceptible | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 897 | 51.20 | 897 | 51.20 | | Late Acute/Recovering | 1 | 0.06 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 1 | 0.06 | | Early Recovery | 3 | 0.17 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 3 | 0.17 | | Unknown | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 45 | 2.57 | 45 | 2.57 | | Total | 30 | 1.71 | 1 | 0.06 | 0 | 0.00 | 16 | 0.91 | 11 | 0.63 | 1694 | 96.69 | 1752 | 100.00 | Continued on next page {27} Page 28 Summary of Safety and Effectiveness Data # Table 10 (Continued) ## ARCHITECT CORE-M Results versus Comparator IgM Anti-HBc Results ### Comparison for Increased Risk and Signs and Symptoms Population (Population One) by HBV Classification a This specimen was tested and determined to be positive for HBeAg and HBV DNA; negative for anti-HBc; and nonreactive by a second FDA-approved IgM anti-HBc assay. b Two specimens were tested and determined to be negative for HBeAg; positive for anti-HBc and HBV DNA; and nonreactive by a second FDA-approved IgM anti-HBc assay. Two specimens were tested and determined to be negative for HBeAg; positive for anti-HBc and HBV DNA; and grayzone by a second FDA approved IgM anti-HBc assay. Four specimens were tested and determined to be negative for HBeAg and HBV DNA; positive for anti-HBc; and nonreactive by a second FDA-approved IgM anti-HBc assay. Five specimens were tested and determined to be negative for HBeAg and HBV DNA; positive for anti-HBc; and grayzone by a second FDA-approved IgM anti-HBc assay. c One specimen was tested and determined to be negative for HBeAg and HBV DNA; positive for anti-HBc; and nonreactive by a second FDA-approved IgM anti-HBc assay. One specimen was tested and determined to be negative for HBeAg, anti-HBc, and HBV DNA; and grayzone by a second FDA-approved IgM anti-HBc assay. d Two specimens were tested and determined to be negative for HBeAg; positive for anti-HBc and HBV DNA; and nonreactive by a second FDA-approved IgM anti-HBc assay. One specimen was tested and determined to be positive for HBeAg and HBV DNA; negative for anti-HBc; and nonreactive by a second FDA-approved IgM anti-HBc assay. One specimen was tested and determined to be positive for HBeAg and HBV DNA; negative for anti-HBc; and grayzone by a second FDA-approved IgM anti-HBc assay. e Four specimens were tested and determined to be negative for HBeAg and HBV DNA; positive for anti-HBc; and nonreactive by a second FDA-approved IgM anti-HBc assay. One specimen was tested and determined to be negative for HBeAg and HBV DNA; positive for anti-HBc; and grayzone by a second FDA-approved IgM anti-HBc assay. f These specimens were tested and determined to be negative for HBeAg and HBV DNA; positive for anti-HBc; and nonreactive by a second FDA-approved IgM anti-HBc assay. 42 {28} Table 11 ARCHITECT CORE-M Results versus Comparator IgM Anti-HBc Results Comparison for Increased Risk and Signs and Symptoms Population (Population Two) by HBV Classification | HBV Classification | Comparator IgM Anti-HBc | Total | | --- | --- | --- | | Reactive | Negative | | ARCHITECT CORE-M | ARCHITECT CORE-M | | Reactive | Grayzone | Nonreactive | Reactive | Grayzone | Nonreactive | | | | N | % | N | % | N | % | N | % | N | % | N | % | N | % | | Early Acute | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 1 | 0.36 | 1 | 0.36 | | Chronic | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 1^{a} | 0.36 | 113 | 40.79 | 114 | 41.16 | | Immune Due to Natural Infection | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 67 | 24.19 | 67 | 24.19 | | Distantly Immune/Anti-HBs Unknown | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 5 | 1.81 | 5 | 1.81 | | Distantly Immune/Anti-HBs Not Detected | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 12 | 4.33 | 12 | 4.33 | | Immune Due to HBV Vaccination | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 41 | 14.80 | 41 | 14.80 | | Susceptible | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 37 | 13.36 | 37 | 13.36 | | Total | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 0 | 0.00 | 1 | 0.36 | 276 | 99.64 | 277 | 100.00 | Page 29 Summary of Safety and Effectiveness Data 45 a This specimen was tested and determined to be positive for HBcAg and HBV DNA; negative for anti-HBc; and grayzone by a second FDA-approved IgM anti-HBc assay {29} Page 30 44 # Summary of Safety and Effectiveness Data ## Percent Agreement The negative percent agreement and positive percent agreement between the ARCHITECT CORE-M assay and the comparator IgM anti-HBc assay and their corresponding 95% exact confidence intervals were calculated for Population One, Population Two, and the acute and pre-selected IgM anti-HBc populations combined by HBV classification. For the purposes of calculating percent agreement, the ARCHITECT CORE-M grayzone results were assigned to the opposite clinical interpretation than that of the comparator IgM anti-HBc assay result. The percent agreement between ARCHITECT CORE-M and the comparator IgM anti-HBc assay for Population One by HBV classification is presented in Table 12 on page 32. The percent agreement between ARCHITECT CORE-M and the comparator IgM anti-HBc assay for Population Two by HBV classification is presented in Table 13 on page 33. For individuals diagnosed with acute HBV infection and for the pre-selected IgM anti-HBc positive specimens combined, the positive percent agreement between the ARCHITECT CORE-M assay results and the comparator IgM anti-HBc assay results was 100.00% (30/30, with a 95% confidence interval of 88.43% to 100.00%). {30} Summary of Safety and Effectiveness Data Table 12 ARCHITECT CORE-M Results versus Comparator IgM Anti-HBc Results Percent Agreement for Increased Risk and Signs and Symptoms Population (Population One) by HBV Classification | HBV Classification | Positive Percent Agreement (%) | 95% Confidence Interval | Negative Percent Agreement (%) | 95% Confidence Interval | | --- | --- | --- | --- | --- | | Early Acute | N/A | N/A | 100.00 (8/8) | 63.06-100.00 | | Acute | 94.12 (16/17) | 71.31-99.85 | N/A | N/A | | Chronic | 100.00 (1/1) | 2.50-100.00 | 93.22 (55/59) | 83.54-98.12 | | Recovering Acute | 100.00 (7/7) | 59.04-100.00 | N/A | N/A | | Recovering Acute/Undetectable HBsAg | 100.00 (2/2) | 15.81-100.00 | N/A | N/A | | Immune Due to Natural Infection | N/A | N/A | 91.36 (201/220) | 86.84-94.72 | | Distantly Immune/Anti-HBs Unknown | N/A | N/A | 97.06 (33/34) | 84.67-99.93 | | Distantly Immune/Anti-HBs Not Detected | N/A | N/A | 97.20 (104/107) | 92.02-99.42 | | Immune Due to HBV Vaccination | N/A | N/A | 100.00 (351/351) | 98.95-100.00 | | Susceptible | N/A | N/A | 100.00 (897/897) | 99.59-100.00 | | Late Acute/Recovering | 100.00 (1/1) | 2.50-100.00 | N/A | N/A | | Early Recovery | 100.00 (3/3) | 29.24-100.00 | N/A | N/A | | Unknown | N/A | N/A | 100.00 (45/45) | 92.13-100.00 | | Total | 96.77 (30/31) | 83.30-99.92 | 98.43 (1694/1721) | 97.73-98.96 | NA = not applicable Positive % agreement=[No. of ARCHITECT CORE-M reactive results in agreement with the comparator IgM anti-HBc reactive results] x 100% [Total number of comparator IgM anti-HBc reactive results] Negative % agreement=[No. of ARCHITECT CORE-M nonreactive results in agreement with the comparator IgM anti-HBc negative results] x 100% [Total number of comparator IgM anti-HBc negative results] 45 {31} Summary of Safety and Effectiveness Data Table 13 ARCHITECT CORE-M Results versus Comparator IgM Anti-HBc Results Percent Agreement for Increased Risk and Signs and Symptoms Population (Population Two) by HBV Classification | HBV Classification | Positive Percent Agreement (%) | 95% Confidence Interval | Negative Percent Agreement (%) | 95% Confidence Interval | | --- | --- | --- | --- | --- | | Early Acute | N/A | N/A | 100.00 (1/1) | 2.50-100.00 | | Chronic | N/A | N/A | 99.12 (113/114) | 95.21-99.98 | | Immune Due to Natural Infection | N/A | N/A | 100.00 (67/67) | 94.64-100.00 | | Distantly Immune/Anti-HBs Unknown | N/A | N/A | 100.00 (5/5) | 47.82-100.00 | | Distantly Immune/Anti-HBs Not Detected | N/A | N/A | 100.00 (12/12) | 73.54-100.00 | | Immune Due to HBV Vaccination | N/A | N/A | 100.00 (41/41) | 91.40-100.00 | | Susceptible | N/A | N/A | 100.00 (37/37) | 90.51-100.00 | | Total | N/A | N/A | 99.64 (276/277) | 98.01-99.99 | NA = not applicable Negative % agreement=[No. of ARCHITECT CORE-M nonreactive results in agreement with the comparator IgM anti-HBc negative results] x 100% [Total number of comparator IgM anti-HBc negative results] Page 32 46 {32} Summary of Safety and Effectiveness Data ## Percent of Positive Specimens The percent of positive ARCHITECT CORE-M specimens for individuals diagnosed with acute HBV infection was 100.00% (14/14, with a 95% confidence interval of 76.84% to 100.00%). The percent of positive ARCHITECT CORE-M specimens for the pre-selected IgM anti-HBc positive specimens was 100.00% (16/16, with a 95% confidence interval of 79.41% to 100.00%). ## Clinical Performance in a Pediatric Population The performance of the ARCHITECT CORE-M assay in a pediatric population was evaluated by testing 100 surplus specimens from a pediatric population collected in Fall River, MA by a specimen vendor, and from the 125 prospectively-collected pediatric specimens from Population One, Population Two, and pre-selected positive specimens. For the surplus pediatric specimens, the negative percent agreement between the ARCHITECT CORE-M assay results and the comparator IgM anti-HBc assay results was 100.00% (100/100, with a 95% confidence interval of 96.38% to 100.00%). For the prospectively-collected pediatric specimens, the positive percent agreement between the ARCHITECT CORE-M assay results and the comparator IgM anti-HBc assay results was 100.00% (8/8, with a 95% confidence interval of 63.06% to 100.00%) and the negative percent agreement between the ARCHITECT CORE-M assay results and the comparator IgM anti-HBc assay results was 99.15% (116/117, with a 95% confidence interval of 95.33% to 99.98%). Page 33 47 {33} Page 34 48 # Summary of Safety and Effectiveness Data The distribution of the ARCHITECT CORE-M reactive, grayzone, and nonreactive results for the surplus pediatric population and prospectively-collected pediatric population is presented below in Table 14 and Table 15, respectively. ## Table 14 ARCHITECT CORE-M Results by Age Range and Gender for the Surplus Pediatric Population | Age Range | | ARCHITECT CORE-M Result | | | | | --- | --- | --- | --- | --- | --- | | | Gender | Reactive N (%) | Grayzone N (%) | Nonreactive N (%) | Total | | 2 to 12 Years | Female | 0 (0.00) | 0 (0.00) | 25 (100.00) | 25 | | | Male | 0 (0.00) | 0 (0.00) | 25 (100.00) | 25 | | 13 to 18 Years | Female | 0 (0.00) | 0 (0.00) | 32 (100.00) | 32 | | | Male | 0 (0.00) | 0 (0.00) | 18 (100.00) | 18 | | Total | | 0 (0.00) | 0 (0.00) | 100 (100.00) | 100 | ## Table 15 ARCHITECT CORE-M Results by Age Range and Gender for the Prospective Pediatric Population | Age Range | | ARCHITECT CORE-M Result | | | | | --- | --- | --- | --- | --- | --- | | | Gender | Reactive N (%) | Grayzone N (%) | Nonreactive N(%) | Total | | 2 to 12 Years | Male | 1 (100.00) | 0 (0.00) | 0 (0.00) | 1 | | 13 to 18 Years | Female | 1 (14.29) | 0 (0.00) | 6 (85.71) | 7 | | | Male | 1 (16.67) | 0 (0.00) | 5 (83.33) | 6 | | 19 to 21 Years | Female | 1 (1.54) | 0 (0.00) | 64 (98.46) | 65 | | | Male | 5 (10.87) | 0 (0.00) | 41 (89.13) | 46 | | Total | | 9 (7.20) | 0 (0.00) | 116 (92.80) | 125 | {34} Page 35 # Summary of Safety and Effectiveness Data ## Expected Results Of the 2,059 prospectively-collected specimens tested in the ARCHITECT CORE-M clinical study, 1,207 were from individuals living in the United States with increased risk of HBV infection. All 1,207 were at risk for HBV due to lifestyle, behavior, occupation, or a known exposure event but were asymptomatic and reported no current signs or symptoms of hepatitis. The 1,207 increased risk specimens from Population One were collected from specimen collection sites or were purchased from specimen vendors. Testing of these specimens was performed at three clinical sites located in Galveston, TX; Hershey, PA; and Milwaukee, WI. A demographic summary of the increased risk population in Population One by race/ethnic group is provided in Table 16 below. | Race/Ethnic Group | N | Percent (%) | | --- | --- | --- | | African American | 396 | 32.81 | | American Indian/Alaska Native | 4 | 0.33 | | Asian | 26 | 2.15 | | Caucasian | 582 | 48.22 | | Hispanic | 176 | 14.58 | | Other | 21 | 1.74 | | Unknown | 2 | 0.17 | | Total | 1207 | 100.00 | {35} Page 36 50 # Summary of Safety and Effectiveness Data Of the 1,207 increased risk subjects in Population One, 645 (53.44%) were female and 562 (46.56%) were male. The age was not reported for one subject. Of the remaining 1,206 subjects, the mean age was 39 years (age range: 17 to 82 years). The ARCHITECT CORE-M assay was reactive in 20 (1.66%) of the individuals in the increased risk population. The number and percent of ARCHITECT CORE-M reactive results observed at each collection location are presented in Table 17 below. ## Table 17 Number and Percent of Reactive Results by Clinical Testing Site for Increased Risk Population (Population One) | Specimen Collection Site/Vendor Location | Percent of Specimen Enrolled at Each Site | Percent of Reactive Results at Each Site | | --- | --- | --- | | Specimen Collection Site 1 Galveston, TX | 34.55 (417/1207) | 1.20 (5/417) | | Specimen Collection Site 2 Dallas, TX | 13.92 (168/1207) | 1.19 (2/168) | | Specimen Collection Site 3 Miami, FL | 8.86 (107/1207) | 0.93 (1/107) | | Specimen Collection Site 4 St. Petersburg, FL | 19.39 (234/1207) | 2.99 (7/234) | | Specimen Collection Site 5 Chicago, IL | 3.89 (47/1207) | 6.38 (3/47) | | Specimen Collection Site 6 Denver, CO | 3.73 (45/1207) | 0.00 (0/45) | | Specimen Vendor 1 Colton, CA | 2.82 (34/1207) | 0.00 (0/34) | | Specimen Vendor 1 Plymouth, MA | 10.02 (121/1207) | 1.65 (2/121) | | Specimen Vendor 1 High Point, NC | 2.82 (34/1207) | 0.00 (0/34) | {36} Summary of Safety and Effectiveness Data The distribution of ARCHITECT CORE-M reactive, grayzone, and nonreactive results by age range and gender is presented in Table 18 below. Table 18 Results by Age Range and Gender for Individuals at Increased Risk Population (Population One) | Age Range | Gender | ARCHITECT CORE-M Result | | | Total | | --- | --- | --- | --- | --- | --- | | | | Reactive N (%) | Grayzone N (%) | Nonreactive N (%) | | | 10 to 19 | Female | 0 (0.00) | 0 (0.00) | 14 (100.00) | 14 | | | Male | 0 (0.00) | 0 (0.00) | 7 (100.00) | 7 | | 20 to 29 | Female | 4 (2.00) | 1 (0.50) | 195 (97.50) | 200 * | | | Male | 2 (1.72) | 1 (0.86) | 113 (97.41) | 116 | | 30 to 39 | Female | 3 (2.00) | 0 (0.00) | 147 (98.00) | 150 | | | Male | 1 (0.66) | 1 (0.66) | 149 (98.68) | 151 | | 40 to 49 | Female | 3 (1.92) | 2 (1.28) | 151 (96.79) | 156 | | | Male | 2 (1.15) | 1 (0.57) | 171 (98.28) | 174 | | 50 to 59 | Female | 2 (2.06) | 0 (0.00) | 95 (97.94) | 97 | | | Male | 1 (1.11) | 1 (1.11) | 88 (97.78) | 90 | | 60 to 69 | Female | 2 (8.33) | 0 (0.00) | 22 (91.67) | 24 | | | Male | 0 (0.00) | 0 (0.00) | 15 (100.00) | 15 | | 70 to 79 | Female | 0 (0.00) | 0 (0.00) | 1 (100.00) | 1 | | | Male | 0 (0.00) | 0 (0.00) | 8 (100.00) | 8 | | 80 to 89 | Female | 0 (0.00) | 0 (0.00) | 3 (100.00) | 3 | | Unknown | Male | 0 (0.00) | 0 (0.00) | 1 (100.00) | 1 | | Total | | 20 (1.66) | 7 (0.58) | 1180 (97.76) | 1207 | {37} Summary of Safety and Effectiveness Data # XI. CONCLUSIONS DRAWN FROM THE STUDIES A multi-center study was conducted to demonstrate that the ARCHITECT CORE-M assay performs as intended in a diagnostic population. Of the 2,159 specimens tested in the ARCHITECT CORE-M clinical study, 1,752 specimens were from individuals at increased risk or with signs and symptoms of a hepatitis infection living in the United States (Population One); 277 specimens were from individuals at increased risk or with signs and symptoms of a hepatitis infection living in Vietnam (Population Two); 14 specimens were from individuals diagnosed with acute HBV infection; 16 specimens were pre-selected IgM anti-HBc positive specimens; and 100 specimens were from pediatric population. The specimens from Population One and Population Two were assigned an HBV classification and the ARCHITECT CORE-M results were compared to the comparator IgM anti-HBc results. The overall positive percent agreement between the ARCHITECT CORE-M assay and the comparator IgM anti-HBc assay was 96.77% (30/31) for Population One. The overall negative percent agreement between the ARCHITECT CORE-M assay and the comparator IgM anti-HBc assay was 98.43% (1,694/1,721) in the same population. The overall negative percent agreement between the ARCHITECT CORE-M assay and the comparator IgM anti-HBC assay was 99.64% (276/277) for Population Two. The overall positive percent agreement between the ARCHITECT CORE-M assay and the comparator IgM anti-HBC assay was 100.00% (14/14) for the diagnosed acute HBV specimens and 100.00% (16/16) for the pre-selected positive specimens. The overall positive percent agreement between the ARCHITECT CORE-M assay and the comparator IgM anti-HBC assay was 100% (8/8) for the prospectively-collected pediatric population and 100.00% (100/100) for the surplus pediatric population. The overall negative percent agreement between the ARCHITECT CORE-M assay and the comparator IgM anti-HBC assay was 99.15% (116/117) for the prospectively-collected pediatric population. Precision and system reproducibility of the ARCHITECT CORE-M assay was established for within-run, within-day, within-lab, and between-sites. Tube Type Interference study results support the use of human serum and plasma (dipotassium EDTA, lithium heparin, and sodium heparin) in the ARCHITECT CORE-M assay. The ability of the ARCHITECT CORE-M assay to detect HBV infections was demonstrated with eight seroconversion panel evaluations. Page 38 52 {38} Page 39 53 # Summary of Safety and Effectiveness Data Studies with cord blood as a surrogate for neonate serum indicate comparable results with maternal serum. The results from both the nonclinical and clinical studies indicate that the ARCHITECT CORE-M assay can be used safely and effectively for the qualitative *in vitro* determination of anti-HBc antibodies in human serum and plasma. The data also support the use of this assay as an aid in the diagnosis of acute or recent hepatitis B virus (HBV) infection in conjunction with other laboratory results and clinical information. ## RISK BENEFIT ANALYSIS As a diagnostic test, the ARCHITECT CORE-M assay involves removal of blood from an individual for testing purposes. This test presents no more of a safety hazard to an individual than is presented to an individual who is having their blood drawn for any other diagnostic evaluation. The benefits to HBV-infected individuals tested by these assays outweigh any potential adverse event or risk to the patient or user due to assay malfunction or operator error. The potential risks encountered with this *in vitro* diagnostic test are not unusual in the clinical laboratory setting. Appropriate warnings for these risks are contained in the labeling and package inserts for these devices. Standard good laboratory practices are considered sufficient to mitigate the risks to the end user. ## SAFETY Based on the results of the nonclinical and clinical laboratory studies, the ARCHITECT CORE-M assay, when used according to the provided directions and in conjunction with other serological and clinical information, should be safe and pose minimal risk to the patient due to false test results. ## EFFECTIVENESS The effectiveness of the ARCHITECT CORE-M assay has been demonstrated for use in determining if IgM antibodies to the core antigen of the hepatitis B virus are present in an individual’s serum or plasma. A reasonable determination of effectiveness of the ARCHITECT CORE-M assay for aiding in the diagnosis of acute or recent HBV infection has been demonstrated. {39} Summary of Safety and Effectiveness Data ## XII. PANEL RECOMMENDATIONS In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Microbiology Advisory Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel. ## XIII. CDRH DECISION FDA issued an approval order on November 6, 2007. The applicant’s manufacturing facility was inspected on 5/2/07 (Abbott Park), &amp; 7/20/07 (Puerto Rico) and found to be in compliance with the Quality Systems Regulation (21 CFR 820). ## XIV. APPROVAL SPECIFICATIONS Directions for use: See the labeling. Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the labeling. Post-approval Requirements and Restrictions: See approval order. Page 40 54
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...