Alinity m SARS-CoV-2 AMP Kit (09N78-096); Alinity m SARS-CoV-2 CTRL Kit (09N78-086)
Applicant
Abbott Molecular
Product Code
QQX · Microbiology
Decision Date
Dec 6, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3981
Device Class
Class 2
Indications for Use
Alinity m SARS-CoV-2 is a real-time in vitro reverse transcription polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System for the qualitative detection of nucleic acid from SARS-CoV-2 from patients with signs and symptoms of COVID-19 in nasopharyngeal (NP) swab and anterior nasal swab (ANS) specimens. Results are for the detection and identification of SARS-CoV-2 RNA. Alinity m SARS-CoV-2 assay is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA; clinical correlation with patient history and other diagnostic information is necessary to determine patient infection status. Positive results do not rule out bacterial infection or co-infection with other viruses. Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information.
Device Story
Alinity m SARS-CoV-2 is an automated, qualitative RT-PCR assay for SARS-CoV-2 RNA detection. Input: nasopharyngeal or anterior nasal swab specimens in transport media. Operation: Alinity m System performs automated sample preparation via magnetic microparticle technology; RT-PCR assembly; amplification; and real-time fluorescence detection. The assay targets RdRp and N genes using a dual-target design with a single fluorophore. System is a random-access analyzer; executes all steps automatically without user intervention. Output: qualitative result (Detected/Not Detected). Used in clinical laboratories by trained personnel. Results aid COVID-19 diagnosis when combined with clinical/epidemiological data. Benefits: high-throughput, automated workflow reduces manual processing errors and improves diagnostic efficiency.
Clinical Evidence
Two prospective clinical studies evaluated performance against a composite comparator (CC) of multiple EUA molecular assays. Study 1 (NPS in UVT, n=535) showed 96.3% PPA and 95.2% NPA. Study 2 (ANS in UVT, n=759) showed 100.0% PPA and 99.7% NPA. Study 2 (ANS in UCK, n=766) showed 97.9% PPA and 97.9% NPA. Analytical studies included LoD determination (55 GE/mL for UCK), precision (within-lab SD 0.888-1.031), reproducibility, interference testing, and cross-reactivity/microbial interference panels. All studies support substantial equivalence.
Technological Characteristics
Real-time RT-PCR assay; targets RdRp and N genes. Automated magnetic microparticle-based nucleic acid extraction. Fluorescence-based detection. Random-access analyzer (Alinity m System). Reagents include primers/probes for SARS-CoV-2 and exogenous internal control. Prescription use only.
Indications for Use
Indicated for qualitative detection of SARS-CoV-2 RNA in nasopharyngeal and anterior nasal swab specimens from symptomatic patients suspected of COVID-19. For prescription use.
Regulatory Classification
Identification
A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
*e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
*e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
*e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
*e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
*i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.
Predicate Devices
cobas SARS-CoV-2 Qualitative for use on the cobas 6800/8800 Systems (K213804)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K241580
B Applicant
Abbott Molecular
C Proprietary and Established Names
Alinity m SARS-CoV-2
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QQX | Class II | 21 CFR 866.3981 - Device to Detect and Identify Nucleic Acid Targets in Respiratory Specimens From Microbial Agents That Cause The SARS-CoV-2 Respiratory Infection and Other Microbial Agents When in A Multi-Target Test | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
The purpose of this submission is to show that the Alinity m SARS-CoV-2 assay is substantially equivalent to the Roche cobas SARS-CoV-2 for use on the cobas 6800/8800 Systems assay (K213804) and to obtain clearance for the Alinity m SARS-CoV-2 assay.
### B Measurand:
Severe Acute Respiratory Syndrome Coronavirus-2 (SARS-CoV-2) RNA
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
C Type of Test:
Qualitative Real Time Polymerase Chain Reaction (RT-PCR)
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Alinity m SARS-CoV-2 is a real-time *in vitro* reverse transcription polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System for the qualitative detection of nucleic acid from SARS-CoV-2 from patients with signs and symptoms of COVID-19 in nasopharyngeal (NP) swab and anterior nasal swab (ANS) specimens.
Results are for the detection and identification of SARS-CoV-2 RNA. Alinity m SARS-CoV-2 assay is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA; clinical correlation with patient history and other diagnostic information is necessary to determine patient infection status. Positive results do not rule out bacterial infection or co-infection with other viruses.
Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
All steps of the Alinity m SARS-CoV-2 assay procedure are executed on the Alinity m System
IV Device/System Characteristics:
A Device Description:
The Alinity m SARS-CoV-2 assay consists of 2 reagent kits:
- Alinity m SARS-CoV-2 AMP Kit (09N78-096) is comprised of 2 types of multi-well trays: Alinity m SARS-CoV-2 AMP TRAY 1 and Alinity m SARS-CoV-2 ACT TRAY 2. The intended storage condition for the Alinity m SARS-CoV-2 AMP Kit is $-15^{\circ}\mathrm{C}$ to $-25^{\circ}\mathrm{C}$.
- Alinity m SARS-CoV-2 CTRL Kit (09N78-086) consists of negative controls and positive controls, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m SARS-CoV-2 Control Kit is $-15^{\circ}\mathrm{C}$ to $-25^{\circ}\mathrm{C}$.
K241580 - Page 2 of 38
{2}
The Alinity m SARS-CoV-2 assay may utilize the following for collection and transport of anterior nasal swab specimens:
- Abbott Universal Collection Kit (09N92-030) consists of one Transport Tube with a solid cap containing 1.65 mL Specimen Transport Buffer and one sterile Specimen Collection Swab. The intended storage condition for the Abbott Universal Collection Kit is 15°C to 30°C.
- Abbott Universal Collection Kit II (09N92-040) consists of one Transport Tube with a pierceable cap containing 1.65 mL Specimen Transport Buffer, one sterile Specimen Collection Swab, and one absorbent pad. The intended storage condition for the Abbott Universal Collection Kit is 15°C to 30°C.
Alinity m SARS-CoV-2 is a real-time *in vitro* reverse transcription polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System for the qualitative detection of nucleic acid from SARS-CoV-2 in specimens collected from patients with signs and symptoms of COVID-19.
The steps of the Alinity m SARS-CoV-2 assay consist of sample preparation, RT-PCR assembly, amplification/detection, and result reporting. All stages of the Alinity m SARS-CoV-2 assay procedure are executed automatically by the Alinity m System. No intermediate processing or transfer steps are performed by the user. The Alinity m System is designed to be a random-access analyzer that can perform the Alinity m SARS-CoV-2 assay in parallel with other Alinity m assays on the same instrument.
SARS-CoV-2 RNA from specimens are extracted automatically on-board the Alinity m System using the Alinity m Sample Prep Kit 2, Alinity m Lysis Solution, and Alinity m Diluent Solution. The Alinity m System employs magnetic microparticle technology to facilitate nucleic acid capture, wash and elution. The resulting purified nucleic acids are then combined with the liquid unit-dose activation reagent, liquid unit-dose amplification reagents, and Alinity m Vapor Barrier Solution, and transferred by the instrument to an amplification/detection module for reverse transcription, PCR amplification, and real-time fluorescence detection.
The Alinity m SARS-CoV-2 amplification reagents include primers and probes that amplify and detect an exogenous internal control (containing an armored RNA sequence). Amplification and detection of the internal control demonstrates proper sample processing. The internal control is used to demonstrate assay validity.
Assay controls are tested to help ensure that instrument and reagent performance remain satisfactory. During each control event, a negative control and a positive control are processed through sample preparation and RT-PCR procedures that are identical to those used for specimens. Assay controls are used to demonstrate proper sample processing and assay validity. Each Alinity m SARS-CoV-2 CTRL kit contains 12 vials (1.3 mL fill volume) of Negative Control and 12 vials (1.3 mL fill volume) of Positive Control.
## Specimen Collection
Nasopharyngeal or anterior nasal swab specimens collected in viral transport media (BD UVT or Copan UTM). The Alinity m SARS-CoV-2 assay may utilize the following for collection and transport of anterior nasal swab specimens:
K241580 - Page 3 of 38
{3}
- Abbott Universal Collection Kit (09N92-030)
- Abbott Universal Collection Kit II (09N92-040)
## B Principle of Operation:
The Alinity m SARS-CoV-2 assay utilizes real-time polymerase chain reaction (PCR) to amplify and detect genomic RNA sequences of SARS-CoV-2 from nasopharyngeal and anterior nasal swab specimens. The Alinity m SARS-CoV-2 is designed to target two highly conserved sequences within the SARS-CoV-2 genome: the RNA-dependent RNA polymerase (RdRp) and Nucleocapsid (N) genes. The RdRp and N probes utilized in the SARS-CoV-2 dual-target design are labeled with the same fluorophore. Therefore, amplification of both SARS-CoV-2 targets is detected as a single fluorescent signal during real-time PCR. The steps of the Alinity m SARS-CoV-2 assay consist of sample preparation, PCR assembly, amplification/detection, and result calculation and reporting. All steps of the Alinity m SARS-CoV-2 assay procedure are executed automatically by the Alinity m System. The Alinity m System is designed to be a random-access analyzer that can perform the Alinity m SARS-CoV-2 assay in parallel with other Alinity m assays on the same instrument.
## C Instrument Description Information:
1. Instrument Name: Alinity m System.
2. Specimen Identification: Specimen identification information is entered either manually or via barcode.
3. Specimen Sampling and Handling: The samples may be loaded on the system in any order. The system pipettor robot dispenses and aspirates liquids, as appropriate for each reaction. Sample handling and reagent transport is performed by a handler robot.
4. Calibration: Not Applicable
5. Quality Control: Alinity assays utilize quality controls to periodically monitor the performance of assay reagents and of the Alinity m System. A QC is defined as the set of assay controls that are, when valid, necessary to allow reporting of specimen results. A QC in this context consists of a negative control and positive control(s). Each assay control is processed through the same sample extraction and PCR procedure used for specimens.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
cobas SARS-CoV-2 Qualitative for use on the cobas 6800/8800 Systems.
K241580 - Page 4 of 38
{4}
B Predicate 510(k) Number(s):
K213804
C Comparison with Predicate(s):
The Alinity m SARS-CoV-2 assay has the same general intended uses as the predicate device cobas SARS-CoV-2 Qualitative for use on the cobas 6800/8800 Systems.
| Device & Predicate Device(s): | K241580 | K213804 |
| --- | --- | --- |
| Device Trade Name | Alinity m SARS-CoV-2 Assay | Roche cobas SARS-CoV-2 Qualitative for use on the cobas 6800/8800 Systems |
| Regulation Number and Product Code | 21 CFR 866.3981; QQX | 21 CFR 866.3981; QQX |
| Device Class | II | II |
| Technology/Detection | Real-Time Reverse Transcription- Polymerase Chain Reaction (RT-PCR) | Real-Time Reverse Transcription- Polymerase Chain Reaction (RT-PCR) |
| Instrument System | Alinity m System | cobas 6800/8800 Systems |
| General Device Characteristic Similarities | | |
| Intended Use/Indications for Use | Alinity m SARS-CoV-2 is a real-time in vitro reverse transcription polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System for the qualitative detection of nucleic acid from SARS-CoV-2 from patients with signs and symptoms of COVID-19 in nasopharyngeal (NP) swab and anterior nasal swab (ANS) specimens. Results are for the detection and identification of SARS-CoV-2 RNA. Alinity m SARS-CoV-2 assay is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiologic, and laboratory findings. Positive results are indicative of the presence of SARS-CoV-2 RNA; clinical correlation with patient history and other diagnostic | cobas SARS-CoV-2 Qualitative for use on the cobas 6800/8800 Systems is a real-time RT-PCR test intended for the qualitative detection of nucleic acids from SARS-CoV-2 in nasal and nasopharyngeal specimens collected from symptomatic individuals suspected of COVID-19 by their healthcare provider. Results are for the detection of SARS-CoV-2 RNA. Positive results are indicative of the presence of SARS-CoV-2 RNA; clinical correlation with patient history and other diagnostic information is necessary to determine patient infection status. Positive results do not rule out bacterial infection or co- infection with other pathogens. Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for |
K241580 - Page 5 of 38
{5}
K241580 - Page 6 of 38
| | information is necessary to determine patient infection status. Positive results do not rule out bacterial infection or co-infection with other viruses.
Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information. | patient management decisions. Results are meant to be used in conjunction with clinical observations, patient history, recent exposures and epidemiological information, and laboratory data, in accordance with the guidelines provided by the relevant public health authorities. cobas SARS-CoV-2 is intended for use by qualified clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and on the use of the cobas 6800/8800 Systems. |
| --- | --- | --- |
| Conditions for use | For prescription use | For prescription use |
| Assay Type | Qualitative | Qualitative |
| Specimen Types | •Nasopharyngeal swab
•Anterior nasal swab | •Nasopharyngeal swab
•Nasal swab |
| Assay Targets | SARS-CoV-2 RNA | SARS-CoV-2 RNA |
| Assay Steps | All steps of the Alinity SARS-CoV-2 assay procedure are executed automatically by the Alinity m System.
No intermediate processing or transfer steps are performed by the user. | All steps of the cobas SARS-CoV-2 qualitative assay procedure are executed automatically by the cobas 6800/8800 Systems.
No intermediate processing or transfer steps are performed by the user. |
| Principles of the Procedure | Alinity m SARS-CoV-2 is a real-time in vitro reverse transcription polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System for the qualitative detection of nucleic acid from SARS-CoV-2 in NPS and ANS specimens.
The steps of the Alinity m SARS-CoV-2 assay consist of sample preparation, RT-PCR assembly, amplification/detection, and result reporting. All stages of the Alinity m SARS-CoV-2 assay procedure are executed | Selective amplification of RNA Internal Control is achieved by the use of non- competitive sequence specific forward and reverse primers which have no homology with the coronavirus genome. A thermostable DNA polymerase enzyme is used for amplification.
The cobas SARS-CoV-2 master mix contains detection probes which are specific for the coronavirus type SARS-CoV-2, members of the Sarbecovirus subgenus, and the RNA Internal Control nucleic acid. The coronavirus and RNA |
{6}
K241580 - Page 7 of 38
| | automatically by the Alinity m System. No intermediate processing or transfer steps are performed by the user. The Alinity m System is designed to be a random-access analyzer that can perform the Alinity m SARS-CoV-2 assay in parallel with other Alinity m assays on the same instrument. | Internal Control detection probes are each labeled with unique fluorescent dyes that act as a reporter. Each probe also has a second dye which acts as a quencher. When not bound to the target sequence, the fluorescent signals of the intact probes are suppressed by the quencher dye. During the PCR amplification step, hybridization of the probes to the specific single-stranded DNA template results in cleavage of the probe by the 5' to 3' exonuclease activity of the DNA polymerase resulting in separation of the reporter and quencher dyes and the generation of a fluorescent signal. With each PCR cycle, increasing amounts of cleaved probes are generated and the cumulative signal of the reporter dye increases concomitantly. Each reporter dye is measured at defined wavelengths, which enables simultaneous detection and discrimination of the amplified coronavirus target and the RNA Internal Control. The master mix includes deoxyuridine triphosphate (dUTP), instead of deoxythimidine triphosphate (dTTP), which is incorporated into the newly synthesized DNA (amplicon). Any contaminating amplicons from previous PCR runs are destroyed by the AmpErase enzyme [uracil-N-glycosylase], which is included in the PCR mix, when heated in the first thermal cycling step. However, newly formed amplicons are not destroyed since the AmpErase enzyme is |
| --- | --- | --- |
| | SARS-CoV-2 RNA from specimens are extracted automatically on-board the Alinity m System using the Alinity m Sample Prep Kit 2, Alinity m Lysis Solution, and Alinity m Diluent Solution. The Alinity m System employs magnetic microparticle technology to facilitate nucleic acid capture, wash and elution. The resulting purified nucleic acids are then combined with the liquid unit-dose activation reagent, liquid unit-dose amplification reagents, and Alinity m Vapor Barrier Solution, and transferred by the instrument to an amplification/detection module for reverse transcription, PCR amplification, and real-time fluorescence detection. | |
{7}
K241580 - Page 8 of 38
| | | inactivated once exposed to temperatures above 55°C. |
| --- | --- | --- |
| Instrumentation System Components | Alinity m System: High-throughput, fully integrated laboratory automation system which utilize real-time PCR technology | cobas 6800/8800 Systems: High-throughput, fully integrated laboratory automation systems which utilize real-time PCR technology |
| Sample Preparation Instrument Components | Automated liquid handling and robotic manipulation platform. | Automated liquid handling and robotic manipulation platform |
| Amplification and Detection Instrument Components | The Amp-Detect units of the Alinity m System are comprised of PCR thermal cycler/fluorescence reader modules that automate the steps for real-time PCR. | The Analytic Modules of the cobas 6800/8800 Systems are used for amplification and detection of nucleic acid using real-time PCR, which is carried out by employing fluorescence spectroscopy. |
| Sample Extraction Technology | • Non-specific nucleic acid capture with magnetic microparticles
• Magnetic microparticles are washed to remove unbound sample components including potential inhibitors
• The bound nucleic acids are eluted and transferred to the lyophilized master mix.
An internal control (IC) is taken through the entire sample preparation and real-time PCR procedure along with the specimens, calibrators, and controls to demonstrate proper sample processing and IC validity | • Nucleic acid capture with magnetic glass particles
• Magnetic glass particles are washed to remove unbound substances and impurities, such as denatured protein, cellular debris and potential PCR inhibitors.
• The purified nucleic acids are eluted from the glass particles with elution buffer.
An internal control (i.e., DNA-QS) is taken through the sample preparation procedure along with the specimens for monitoring the sample preparation and PCR amplification process. |
| Amplification Controls | Assay specific IC. The IC primer/probe set amplifies and detects an exogenous RNA sequence unrelated to the SARS-CoV-2 target sequences | Selective amplification of RNA Internal Control is achieved by the use of non-competitive sequence specific forward and reverse primers which have no homology with the coronavirus genome. |
| Detection Procedure | • Optical detection of stimulated fluorescence.
The fluorescence reader monitors real-time | • Optical detection of stimulated fluorescence.
The Analytic Module monitors real-time |
{8}
K241580 - Page 9 of 38
| | fluorescence during every PCR amplification cycle. | fluorescence during every PCR amplification cycle. |
| --- | --- | --- |
| Detection Chemistry | • Fluorescence labeled, single stranded, target-specific probes.
• Detection of the DNA amplicon is achieved using nucleic acid (DNA:DNA) hybridization.
• SARS-CoV-2 and Internal control (IC) probes are labeled with different fluorophores allowing SARS- CoV-2 to be distinguished from the IC amplicons. | Paired reporter and quencher fluorescence labeled probes (TaqMan Technology) using fluorescence resonance energy transfer (FRET) |
| Assay Controls | • Internal Control (IC)
• Negative Control
• Positive Control | •Sample processing Control (IC)
•Negative control
•Positive control |
| General Device Characteristic Differences | | |
| Specimen Collection and Transport | Nasopharyngeal samples must be collected in viral transport media. Anterior nasal specimens must be collected in viral transport medium, universal transport media, or Abbott Universal Collection Kit, or Abbott Universal Collection Kit II. | Nasopharyngeal swab and nasal swab, specimens collected in viral transport medium or universal transport media.
Nasal swab specimens may also be collected in cobas PCR Media Uni Swab Sample Kit, cobas PCR Media Dual Swab Sample Kit, cobas PCR Media Kit (and 100 tube PCR Media Kit), or 0.9% Physiological Saline. |
| Results Reporting | • Not Detected, Negative
• xx.xx CN, Positive | • Target 1 and Target 2 Positive, Detected
• Target 1 Positive and Target 2 Negative, Detected
• Target 1 Negative and Target 2 Positive, Presumptive Positive
• Targets 1 and 2 Negative, Not Detected |
{9}
VI Standards/Guidance Documents Referenced:
Class II Special Controls as per 21 CFR 866.3981.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
a. Within-Laboratory Precision
Within-laboratory precision was evaluated at a single site for the Alinity m SARS-CoV-2 assay using Alinity m SARS-CoV-2 reagents and the Alinity m System with a total of three panels (negative samples, positive samples spiked with SARS-CoV-2 at 2X LoD, and 5X LoD) in a negative Simulated Nasal Matrix (SNM). Positive panel members were prepared by diluting gamma-irradiated SARS-CoV-2 virus in SNM. Un-spiked SNM was used as the SARS-CoV-2 negative panel member. Three operator/instrument/lot combinations were tested for each panel member using Alinity m SARS-CoV-2 Amplification Reagent Kits on five days with two runs per day and four replicates per run, resulting in a total of 120 replicates per panel member (40 replicates per operator/instrument/lot combination). Each Alinity m SARS-CoV-2 AMP Kit reagent lot was assigned to a specific Alinity m System and each operator was assigned to one lot/Alinity m System combination. Precision study results are summarized in Table 1.
Table 1. Within-Laboratory Precision Study Results
| Target Concentration | Tested | Positive | Rate | 95% CI |
| --- | --- | --- | --- | --- |
| Negative^{a} | 119 | 0 | 100.0% | (96.9%,100.0%) |
| 2X LoD^{b} (Low positive) | 119 | 118 | 99.2% | (95.4%,99.9%) |
| 5X LoD (Moderate Positive) | 120 | 120 | 100.0% | (96.9%,100.0%) |
a One replicate of negative sample was invalid and excluded from analysis. The sample was not retested as the minimum sample size of 90 was achieved.
b One replicate of low positive sample was invalid and excluded from analysis. The sample was not retested as the minimum sample size of 90 was achieved.
Low positive panel members were positive at 99.2% (118/119) while moderate positive panel members were positive at 100% (120/120) for the spiked analyte. The negative panel members were 100% negative.
b. Within-Laboratory Precision – Testing Specimens in Universal Collection Kits Buffer
An additional precision study was carried out to assess precision by testing specimens in SNM in the Abbott Universal Collection Kits (UCKs) transport buffer. Precision was evaluated for the Alinity mSARS-CoV-2 assay by testing three panel members (negative samples, positive samples spiked with SARS-CoV-2 at 2X LoD, and 5X LoD) prepared in a negative matrix simulating clinical specimens collected in the Universal Collection
K241580 - Page 10 of 38
{10}
Kit or Universal Collection Kit II that contains transport buffer. Positive Panel Members were prepared by diluting gamma irradiated SARS-CoV-2 virus in SARS-CoV-2 negative Simulated Nasal Matrix (SNM) described previously. Un-spiked SNM in the UCKs transport buffer was used as the SARS-CoV-2 negative panel member. Target concentrations for each panel member and testing outline are summarized in Table 2.
Table 2. Precision Panel Members (Specimens in SNM in the Universal Collection Kits Transport Buffer) and Testing Plan
| Panel Member | Target Concentration | Number of Instruments/Operators | Number of Days | Runs per Day | Replicates per Run | Total Replicates |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | Negative | 3 | 5 | 2 | 3 | 90 |
| 2 | 2X LoD (Near LoD) | 3 | 5 | 2 | 3 | 90 |
| 3 | 5X LoD (Moderate Positive | 3 | 5 | 2 | 3 | 90 |
The precision study design comprised one sample preparation reagent and one amplification reagent pair on three instrument pairs by three operators with a minimum of six replicates per run and one testing run each day for a minimum of five days on each instrument/operator combination to ensure a minimum of 90 valid replicates per panel member. Results are summarized in Table 3.
Table 3. Precision Study Results for Specimens in SNM in the Universal Collection Kit Transport Buffer
| Panel Member | Target Concentration | Tested | Positive | Rate | 95% CI |
| --- | --- | --- | --- | --- | --- |
| 1 | Negative | 90 | 0 | 100% | (95.9%,100%) |
| 2 | 2X LoD (Near LoD) | 90 | 90 | 100% | (95.9%,100%) |
| 3 | 5X LoD (Moderate Positive) | 90 | 90 | 100% | (95.9%,100%) |
All spiked panel members were at $100\%$ positivity (90/90). The negative panel members were $100\%$ negative. The results of the study demonstrate acceptable assay variability testing samples collected in Universal Collection Kit buffer.
# c. Reproducibility
Reproducibility of the Alinity m SARS-CoV-2 assay was evaluated at three external testing sites by testing a three-member panel prepared in SNM. The two positive panel members consisted of one moderate positive panel member containing gamma irradiated SARS-CoV-2 virus at approximately 5X LoD and one low positive panel member containing gamma irradiated SARS-CoV-2 virus at approximately 2X LoD. Negative SNM was used as a negative panel member.
K241580 - Page 11 of 38
{11}
A total of three Alinity m SARS-CoV-2 AMP Kit lots were used. Each of the three external sites tested two Alinity m SARS-CoV-2 AMP Kit lots, on five non-consecutive days for each lot. Four replicates of each panel member were tested on each of five days. Each of the three external sites used different lots of Alinity m SARS-CoV-2 CTRL Kits and Alinity m Sample Prep Kit 2. Reproducibility results are summarized in Table 4.
Table 4. Reproducibility Study Summary Results
| | | | | | Within-Run | | Between-Run | | Between Day | | WithinLaboratoryc | | BetweenSite/Instrument | | Totald | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| PanelMember | Na | Nb | Agreement | MeanCN | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| ModeratePositive(5x LoD) | 120 | 120 | 100% | 34.27 | 0.62 | 1.8 | 0.00 | 0.0 | 0.54 | 1.6 | 0.535 | 1.6 | 0.00 | 0.0 | 0.82 | 2.4 |
| LowPositive(2x LoD) | 119 | 119 | 100% | 35.78 | 0.74 | 2.1 | 0.38 | 1.1 | 0.28 | 0.8 | 0.681 | 1.9 | 0.17 | 0.5 | 0.89 | 2.5 |
| Negative | 119 | 119 | 100% | | - | - | - | - | - | - | - | - | - | - | - | - |
Total number of valid replicates.
${}^{b}$ Replicates with positive result interpretation for positive panels and negative result interpretation for negative panel. Number of replicates used in the Mean and SD calculation for the positive panel members.
a Within-laboratory includes Within-Run, Between-Run, and Between-Day Components.
Total includes Within-Run, Between-Run, Between-Day, and Between-Instrument Components.
Both the Low Positive and Moderate Positive Panels were positive $100\%$ of the time. The negative panel was negative $100\%$ of the time. The results of the study demonstrate acceptable assay reproducibility.
# 2. Linearity:
A Linearity Study is not applicable because the Alinity m SARS-CoV-2 assay is a qualitative assay.
# 3. Analytical Reactivity (Inclusivity):
The inclusivity of the Alinity m SARS-CoV-2 Assay for the detection of SARS-CoV-2 was evaluated using a combination of in silico analysis of publicly available sequence information and laboratory (wet) testing of seven isolates of SARS-CoV-2 from six different geographical regions. Seven different strains/isolates were tested at 3X the assay's LoD and were prepared in negative nasopharyngeal matrix with either purified genomic RNA from the SARS-CoV-2 isolates or inactivated virus. Each strain/isolate was tested in five replicates and results are presented in Table 5.
Table 5. SARS-CoV-2 Inclusivity Wet Testing Results.
| SARS-CoV-2 Strain (isolate) | Concentrationa | Tested (N) | Positive (N) | Negative (N) | Detection Rate |
| --- | --- | --- | --- | --- | --- |
| USA-AZ1/2020 | 165 GE/mL | 5 | 5 | 0 | 100% (5/5) |
| USA-CA3/2020 | 165 GE/mL | 5 | 5 | 0 | 100% (5/5) |
K241580 - Page 12 of 38
{12}
| Hong Kong/VM20001061/2020 | 165 GE/mL | 5 | 5 | 0 | 100% (5/5) |
| --- | --- | --- | --- | --- | --- |
| USA-IL1/2020 | 165 GE/mL | 5 | 5 | 0 | 100% (5/5) |
| Italy-INMI1 | 165 GE/mL | 5 | 5 | 0 | 100% (5/5) |
| USA/CA_CDC_5574/2020 (B.1.1.7) | 0.027 TCID50/mL | 5 | 5 | 0 | 100% (5/5) |
| hCoV-19/South Africa/KRISP-K005325/2020 (B.1.351) | 0.027 TCID50/mL | 5 | 5 | 0 | 100% (5/5) |
$^{\mathrm{a}}$ GE/mL = Genome Equivalent/mL or TCID50/mL = Median Tissue Culture Infectious Dose/mL
## In Silico
The inclusivity of the Alinity m SARS-CoV-2 Assay was evaluated using *in silico* analysis of the primers and probes for the ability to detect various strains of the SARS-CoV-2 virus. Primer and probe sequences for SARS-CoV-2 detection were analyzed for homology with sequences available in the Global Initiative on Sharing All Influenza Data (GISAID, as of October 11, 2023) and National Center for Biotechnology Information (NCBI, available as of October 10, 2023) databases and the potential impact of any mismatches and the ability to detect SARS-CoV-2 were evaluated.
Inclusivity was demonstrated by analyzing the sequences of the RdRp and N primer/probe sets for homology with 14,818,776 unambiguous SARS-CoV-2 sequences available in the GISAID database. In addition, an analysis was performed for the sequences that have variant of concern (VOC) and variant of interest (VOI) variant designation. Of the 14,818,776 total available sequences, 13,680,493 have current (as of October 10, 2023) or former VOC, VOI, and variant under monitoring (VUM) designation including Omicron EG.5, Omicron B.1.640, Omicron XBB.1.16, Omicron XBB.1.5, Omicron BA.2.75, Omicron BA.2.86, Omicron CH.1.1, Omicron XBB.1.9.1, Omicron XBB.1.9.2, Omicron XBB.2.3, Omicron XBB, Alpha, Beta, Delta, Gamma, Epsilon, Eta, Iota, Kappa, Lambda, Mu, Theta, and Zeta, among others.
*In silico* analyses were also performed with the full-length SARS-CoV-2 sequences available in the NCBI database. A total of 7,590,332 unambiguous sequences in the NCBI database were analyzed for SARS-CoV-2 inclusivity.
The sequences were evaluated for a mismatch in either RdRp or N gene target regions. The impact of any mismatches on the amplification and detection of each target region was assessed.
## GSAID Database
14,764,951 (99.64%) available sequences in the GISAID database either had no mismatches or had mismatches in one of the assay target regions. These sequences were predicted to have no impact on the detection of SARS-CoV-2. The remaining 53,825 (0.36%) sequences contained at least one mismatch in both of the assay target regions. 53,754 of these sequences were predicted to be unlikely to impact the detection of SARS-CoV-2.
K241580 - Page 13 of 38
{13}
Among the 13,680,493 sequences available in the GISAID database that had VOC and VOI designation, 13,628,903 sequences (99.6%) either had no mismatches in the assay target regions or had mismatches in one of the target regions. One VUM Omicron (XBB.1.9.1+XBB.1.9.1.*), one former VOC Alpha, six former VOC Omicron (B.1.1.529+BA.*) variants, and 59 former VOC Delta variants are potentially impacted. Considering the extremely low prevalence of those potentially impacted, and de-escalated status of most impacted variants, the analysis predicts no impact on the detection of SARS-CoV-2 among former or current VOC, VOI and VUM.
## NCBI Database
Among the 7,590,332 unambiguous sequences, 7,559,687 (99.6%) have either no mismatches in the assay target regions (RdRp and N gene targets) or have mismatches in only one of the target regions. These sequences are predicted not to impact the detection of SARS-CoV-2. The remaining 30,645 (0.4%) isolates contained at least one mismatch in both the RdRp and N gene target regions.
The overall potential impact on SARS-CoV-2 detection is considered likely only when both RdRp and N genes are potentially impacted. Among the 30,645 sequences containing at least one mismatch in both target regions, 30,602 were predicted unlikely to impact the detection of SARS-CoV-2. Among the SARS-CoV-2 sequences in the NCBI database, 43 isolates are predicted to be potentially impacted by mismatches.
The majority of sequences evaluated for SARS-CoV-2 are expected to be detected by the Alinity m SARS-CoV-2 Assay.
A summary of the in-silico analyses is presented in Table 6.
Table 6. Summary of In-Silico Analysis of SARS-CoV-2 Oligo Designs
| | NCBI | | GISAID | |
| --- | --- | --- | --- | --- |
| Total Number of Sequences | 7,590,332 | | 14,818,776 | |
| Mismatch in Both RdRp and N Gene | 30645 | 0.40% | 53825 | 0.36% |
| Potentially Impacteda | 43 | 0.00% | 71 | 0.00% |
a Sequences are considered potentially impacted only when both targets (RdRp and N gene) are likely impacted.
## 4. Analytical Specificity/Interference:
### Cross-Reactivity
This study evaluated the analytical specificity (cross-reactivity) of the Alinity m SARS-CoV-2 Assay in the presence of non-targeted microorganisms that may be found in a respiratory tract clinical specimen. Fifty-three microorganisms listed in Table 7 were evaluated in the study and were inoculated into SARS-CoV-2 negative samples to achieve a final titer of 10⁵ units/mL for viruses and fungi, 10⁶ units/mL for bacteria, and 10% of pooled human nasal wash. Whole microorganisms were used in this study whenever possible. Purified nucleic acid was utilized in cases where the microorganism was not available. The SARS-CoV-2 negative
K241580 - Page 14 of 38
{14}
sample was also tested without addition of any potential cross reactants as the control condition. The negative sample was prepared by pooling SARS-CoV-2 negative nasopharyngeal swabs (NPS) in UTM/VTM. Each test condition and the control condition were evaluated in triplicate with the Alinity m SARS-CoV-2 assay. The results of the study are summarized in Table 8.
Table 7. List of Potential Cross-Reactive Organisms Tested
| Organism Tested | Testing Concentration (Units/mL) |
| --- | --- |
| Human coronavirus 229E1 | 1.00E+05 cp/mL |
| Human coronavirus OC432 | 1.00E+05 TCID50/mL |
| Human coronavirus HKU13 | 1.00E+05 cp/mL |
| Human coronavirus NL631 | 1.00E+05 cp/mL |
| SARS-coronavirus3 | 1.00E+05 cp/mL |
| MERS-coronavirus3 | 1.00E+05 cp/mL |
| Adenovirus Type 52 | 1.00E+05 TCID50/mL |
| Human Metapneumovirus (hMPV)2 | 1.00E+05 U/mL |
| Parainfluenza virus 12 | 1.00E+05 TCID50/mL |
| Parainfluenza virus 22 | 1.00E+05 TCID50/mL |
| Parainfluenza virus 32 | 1.00E+05 TCID50/mL |
| Parainfluenza virus 42 | 1.00E+05 TCID50/mL |
| Influenza A (H1N1)1 | 1.00E+05 cp/mL |
| Influenza A (H3N2)1 | 1.00E+05 cp/mL |
| Influenza B1 | 1.00E+05 cp/mL |
| Respiratory syncytial virus Type B1 | 1.00E+05 cp/mL |
| Enterovirus EV682 | 1.00E+05 TCID50/mL |
| Respiratory syncytial virus Type A1 | 1.00E+05 cp/mL |
| Rhinovirus1 | 1.00E+05 cp/mL |
| Chlamydia pneumoniae4 | 1.00E+06 IFU/mL |
| Haemophilus influenzae4 | 1.00E+06 CFU/mL |
| Legionella pneumophila4 | 1.00E+06 CFU/mL |
| Mycobacterium tuberculosis4 | 1.00E+06 CFU/mL |
| Streptococcus pneumoniae4 | 1.00E+06 CFU/mL |
| Streptococcus pyogenes4 | 1.00E+06 CFU/mL |
| Bordetella pertussis4 | 1.00E+06 CFU/mL |
| Mycoplasma pneumoniae4 | 1.00E+06 CFU/mL |
| Pneumocystis jirovecii (PJP) (2%)4 | N/Aa |
| Pooled human nasal wash (10%)5 | N/Ab |
{15}
Table 8. Cross-Reactivity Study Results
| Organism Tested | Tested (N) | Positive (N) | Negative (N) | Invalid (N) |
| --- | --- | --- | --- | --- |
| Control (no cross-reactant) | 3 | 0 | 3 | 0 |
| Cytomegalovirus (CMV) | 1 | 0 | 1 | 0 |
| Escherichia coli | 1 | 0 | 1 | 0 |
| Lactobacillus (plantarum 17-5) | 1 | 0 | 1 | 0 |
| Lagionella longbeachae | 1 | 0 | 1 | 0 |
| Lactobacillus (plantarum 17-5) | 1 | 0 | 1 | 0 |
a The concentration provided by the vendor is in terms of Ct Range, 23 - 25.
b The sample was tested neat and as a dilution (100% and 2%) in this study.
c The pooled human nasal wash was tested at 1 concentration, 10%, in this study.
d Viral particles
e Viral lysate
f Viral RNA
g Whole microorganism
h Nasal wash
i Synthetic nucleic acid
K241580 - Page 16 of 38
{16}
| Human coronavirus 229E | 3 | 0 | 3 | 0 |
| --- | --- | --- | --- | --- |
| Human coronavirus OC43 | 3 | 0 | 3 | 0 |
| Human coronavirus HKU1 | 3 | 0 | 3 | 0 |
| Human coronavirus NL63 | 3 | 0 | 3 | 0 |
| SARS-coronavirus | 3 | 0 | 3 | 0 |
| MERS-coronavirus | 3 | 0 | 3 | 0 |
| Adenovirus Type 5 | 3 | 0 | 3 | 0 |
| Human Metapneumovirus (hMPV) | 3 | 0 | 3 | 0 |
| Parainfluenza virus 1 | 3 | 0 | 3 | 0 |
| Parainfluenza virus 2 | 3 | 0 | 3 | 0 |
| Parainfluenza virus 3 | 3 | 0 | 3 | 0 |
| Parainfluenza virus 4 | 3 | 0 | 3 | 0 |
| Influenza A (H1N1) | 3 | 0 | 3 | 0 |
| Influenza A (H3N2) | 3 | 0 | 3 | 0 |
| Influenza B | 3 | 0 | 3 | 0 |
| Respiratory syncytial virus Type B | 3 | 0 | 3 | 0 |
| Enterovirus EV68 | 3 | 0 | 3 | 0 |
| Respiratory syncytial virus Type A | 3 | 0 | 3 | 0 |
| Rhinovirus | 3 | 0 | 3 | 0 |
| Chlamydia pneumoniae | 3 | 0 | 3 | 0 |
| Haemophilus influenzae | 3 | 0 | 3 | 0 |
| Legionella pneumophila | 3 | 0 | 3 | 0 |
| Mycobacterium tuberculosis | 3 | 0 | 3 | 0 |
| Streptococcus pneumoniae | 3 | 0 | 3 | 0 |
| Streptococcus pyogenes | 3 | 0 | 3 | 0 |
| Bordetella pertussis | 3 | 0 | 3 | 0 |
| Mycoplasma pneumoniae | 3 | 0 | 3 | 0 |
| Pneumocystis jirovecii (2%) | 3 | 0 | 3 | 0 |
| Pooled human nasal wash | 3 | 0 | 3 | 0 |
None of the potentially cross-reactive viruses or microorganisms evaluated in this study generated a false positive SARS-CoV-2 result when tested with the Alinity m SARS-CoV-2 assay.
## Microbial Interference
The Alinity m SARS-CoV-2 assay was also evaluated for interference with the organisms listed in Table 9. The organisms were spiked into SARS-CoV-2 positive samples to achieve a final titer of $10^{5}$ units/mL for viruses and fungi, $10^{6}$ units/ml for bacteria, and $10\%$ of pooled human
K241580 - Page 17 of 38
{17}
nasal wash. A SARS-CoV-2 positive sample was also tested without addition of any potential interfering organisms as the control condition.
The SARS-CoV-2 positive samples at 3X LoD were prepared by spiking gamma irradiated SARS-CoV-2 virus into pooled SARS-CoV-2 negative clinical nasopharyngeal swab (NPS) specimens in UTM/VTM. Three replicates of each test condition and the control condition were tested with the Alinity m SARS-CoV-2 assay. The results are summarized in Table 9.
Table 9. Microbial Interference Study Results
| Organism Tested | Tested (N) | Positive(N) | Negative (N) | Invalid (N) |
| --- | --- | --- | --- | --- |
| Control (no cross-reactant) | 3 | 3 | 0 | 0 |
| Human coronavirus 229E | 3 | 3 | 0 | 0 |
| Human coronavirus OC43 | 3 | 3 | 0 | 0 |
| Human coronavirus HKU1 | 3 | 3 | 0 | 0 |
| Human coronavirus NL63 | 3 | 3 | 0 | 0 |
| SARS-coronavirus | 3 | 3 | 0 | 0 |
| MERS-coronavirus | 3 | 3 | 0 | 0 |
| Adenovirus Type 5 | 3 | 3 | 0 | 0 |
| Human Metapneumovirus (hMPV) | 3 | 3 | 0 | 0 |
| Parainfluenza virus 1 | 3 | 3 | 0 | 0 |
| Parainfluenza virus 2 | 3 | 3 | 0 | 0 |
| Parainfluenza virus 3 | 3 | 3 | 0 | 0 |
| Parainfluenza virus 4 | 3 | 3 | 0 | 0 |
| Influenza A (H1N1) | 3 | 3 | 0 | 0 |
| Influenza A (H3N2) | 3 | 3 | 0 | 0 |
| Influenza B | 3 | 3 | 0 | 0 |
| Respiratory syncytial virus Type B | 3 | 3 | 0 | 0 |
| Enterovirus EV68 | 3 | 3 | 0 | 0 |
| Respiratory syncytial virus Type A | 3 | 3 | 0 | 0 |
| Rhinovirus | 3 | 3 | 0 | 0 |
| Chlamydia pneumoniae | 3 | 3 | 0 | 0 |
| Haemophilus influenzae | 3 | 3 | 0 | 0 |
| Legionella pneumophila | 3 | 3 | 0 | 0 |
| Mycobacterium tuberculosis | 3 | 3 | 0 | 0 |
| Streptococcus pneumoniae | 3 | 3 | 0 | 0 |
| Streptococcus pyogenes | 3 | 3 | 0 | 0 |
| Bordetella pertussis | 3 | 3 | 0 | 0 |
{18}
| Mycoplasma pneumoniae | 3 | 3 | 0 | 0 |
| --- | --- | --- | --- | --- |
| Pneumocystis jirovecii (2%) | 3 | 3 | 0 | 0 |
| Pooled human nasal wash | 3 | 3 | 0 | 0 |
| Candida albicans | 3 | 3 | 0 | 0 |
| Pseudomonas aeruginosa | 3 | 3 | 0 | 0 |
| Staphylococcus epidermis | 3 | 3 | 0 | 0 |
| Streptococcus salivarius | 3 | 3 | 0 | 0 |
| Pneumocystis jirovecii (100%) | 3 | 3 | 0 | 0 |
| SARS-coronavirus (Whole org) | 3 | 3 | 0 | 0 |
| MERS-coronavirus (Whole org) | 3 | 3 | 0 | 0 |
| Control (no cross-reactant) | 3 | 3 | 0 | 0 |
| Synthetic Human bocavirus DNA | 3 | 3 | 0 | 0 |
| Cytomegalovirus (CMV) | 3 | 3 | 0 | 0 |
| Epstein-Barr Virus (EBV) | 3 | 3 | 0 | 0 |
| Measles | 3 | 3 | 0 | 0 |
| Mumps Virus | 3 | 3 | 0 | 0 |
| Parechovirus Type 3 | 3 | 3 | 0 | 0 |
| Corynebacterium diphtheriae | 3 | 3 | 0 | 0 |
| Escherichia coli | 3 | 3 | 0 | 0 |
| Lactobacillus gasseri | 3 | 3 | 0 | 0 |
| Moraxella catarrhalis | 3 | 3 | 0 | 0 |
| Neisseria elongata | 3 | 3 | 0 | 0 |
| Neisseria meningitidis | 3 | 3 | 0 | 0 |
| Staphylococcus aureus | 3 | 3 | 0 | 0 |
| Coxiella burnetti | 3 | 3 | 0 | 0 |
| Lactobacillus (plantarum 17-5) | 3 | 3 | 0 | 0 |
| Legionella longbeachae, Long Beach 4 | 3 | 3 | 0 | 0 |
| Aspergillus fumigatus | 3 | 3 | 0 | 0 |
| Chlamydophila psittaci | 3 | 3 | 0 | 0 |
None of the potentially interfering organisms evaluated in this study impacts the detection of SARS-CoV-2 virus when tested with the Alinity m SARS-CoV-2 assay.
## Interfering Substances – Nasopharyngeal Swab (NPS) Specimens in UVT/UTM
An analytical study was performed to assess the potential inhibitory effects of exogenous and endogenous substances that may be found in NPS specimens, on the performance of the Alinity m SARS-CoV-2 assay. A panel of 34 potentially inhibitory exogenous and endogenous substances were added into SARS-CoV-2 low positive samples (3XLoD) and tested using the Alinity m SARS-CoV-2 assay. A SARS-CoV-2 positive sample was also tested without the
K241580 - Page 19 of 38
{19}
addition of any potentially interfering endogenous and exogenous substances (Control Condition).
The SARS-CoV-2 low positive samples at 3X LoD were prepared by spiking gamma irradiated SARS-CoV-2 virus into pooled SARS-CoV-2 negative clinical nasopharyngeal swab (NPS) specimens in UVT/UTM. Three replicates of each test condition and the control condition were tested with the Alinity m SARS-CoV-2 assay. The interfering substances study results are summarized in Table 10.
Table 10. Interfering Substances Study Results - NPS Specimens in UVT/UTM
| Potentially Interfering Substance | Active Ingredient | Level | Included Samples (N) | Negative (N) | Positive (N) | Invalid IC (N) |
| --- | --- | --- | --- | --- | --- | --- |
| Control | NA | NA | 3 | 0 | 3 | 0 |
| Blood (human) | NA | 10% (v/v) | 6a | 0 | 5 | 0 |
| Throat Lozenges, Oral Anesthetic and Analgesic - Cepacol | Benzocaine, Menthol | 5 mg/mL | 3 | 0 | 3 | 0 |
| Mucin | Purified Bovine mucin protein | 5 mg/mL | 3 | 0 | 3 | 0 |
| Mucin | Purified Porcine mucin protein | 5 mg/mL | 6b | 0 | 5 | 0 |
| Antibiotic, Nasal Ointment - Bactroban | Mupirocin | 5 mg/mL | 3 | 0 | 3 | 0 |
| Nasal Spray-Afrin | Oxymetazoline | 10% (v/v) | 3 | 0 | 3 | 0 |
| Anti-Viral Drug - Relenza | Zanamivir | 5 mg/mL | 3 | 0 | 3 | 0 |
| Anti-Viral Drug - Veklury | Remdesivir | 27.0 μM | 3 | 0 | 3 | 0 |
| Antibacterial Systemic (Tobramycin) | Tobramycin | 4 μg/mL | 3 | 0 | 3 | 0 |
| Nasal Gel /Homeopathic Allergy Relief Medicine - Zicam | Galphimia glauca, Histaminum, hydrochloricum, Luffa operculata, Sulfur | 10% (v/v) | 6c | 0 | 5 | 0 |
| FluMist d | Live intranasal influenza virus | 10% (v/v) | 3 | 0 | 3 | 0 |
| Nasal Corticosteroid -Flonase Sensimist | Fluticasone Furoate | 10% (v/v) | 3 | 0 | 3 | 0 |
| Nasal Corticosteroid - QVAR | Beclomethasone | 2% (v/v) | 3 | 0 | 3 | 0 |
| Corticosteroid - Dexamethasone | Dexamethasone | 0.2 mg/mL | 3 | 0 | 3 | 0 |
| Nasal Corticosteroid - Flunisolide | Flunisolide | 2% (v/v) | 3 | 0 | 3 | 0 |
| Nasal Corticosteroid - Triamcinolone | Triamcinolone | 2% (v/v) | 3 | 0 | 3 | 0 |
K241580 - Page 20 of 38
{20}
K241580 - Page 21 of 38
| Nasal Corticosteroid - Budesonide | Budesonide | 2% (v/v) | 3 | 0 | 3 | 0 |
| --- | --- | --- | --- | --- | --- | --- |
| Nasal Corticosteroid - Mometasone | Mometasone | 2% (v/v) | 3 | 0 | 3 | 0 |
| Toothpaste | Fluoride | 1% (w/v) | 3 | 0 | 3 | 0 |
| Tobacco product | Nicotine | 0.1% (w/v) | 3 | 0 | 3 | 0 |
| Oral rinse Listerine Cool Mint | Ethanol, essential oil | 10% (v/v) | 3 | 0 | 3 | 0 |
| Leukocytes | Leukocytes | 1.1E6 cells/mL | 3 | 0 | 3 | 0 |
| Nasal Decongestant - Phenylephrine | Phenylephrine | 2% (v/v) | 3 | 0 | 3 | 0 |
| Saline nasal mist | Sodium chloride | 2% (v/v) | 3 | 0 | 3 | 0 |
| Nicotine Product | Nicotine | 0.1% (w/v) equal to 0.05 mg/mL | 3 | 0 | 3 | 0 |
| Nasal Spray-Afrin | Oxymetazoline | 15% (v/v) | 3 | 0 | 3 | 0 |
| Chloroseptic Sore Throat Spray | Phenol | 5% (v/v) | 3 | 0 | 3 | 0 |
| Cough Syrup (Wal-Tussin) | Dextromethorphan and guaifenesin | 5% (v/v) | 3 | 0 | 3 | 0 |
| Mucin - Bovine | Purified mucin protein | 5 mg/mL | 3 | 0 | 3 | 0 |
| Lidocaine and Phenylephrine (e.g., Liposomal NUMB250 Spray) | Lidocaine and Phenylephrine | 2.68 mg/mL | 3 | 0 | 3 | 0 |
| Anti-Viral Drug – Tamiflu - Oseltamivir | Oseltamivir | 3.3 mg/mL | 3 | 0 | 3 | 0 |
| Petroleum Jelly (e.g., Vaseline) | Petroleum Jelly | 1% (w/v) | 3 | 0 | 3 | 0 |
| Analgesic Ointment (e.g., Vicks VapoRub) | Camphor-synthetic, eucalyptus oil and menthol ointment | 1% (w/v) | 3 | 0 | 3 | 0 |
| Throat Lozenge, Oral Anesthetic and Analgesic (e.g., Cold Eeze) | Zincuum Gluconicum | 2.5% (w/v) | 3 | 0 | 3 | 0 |
| Human Genomic DNA | N/A | 0.02 mg/mL | 3 | 0 | 3 | 0 |
| Saliva (Human) | N/A | 10% (v/v) | 3 | 0 | 3 | 0 |
a, b One replicate was valid and not detected. Per protocol, the sample was retested in triplicate and all retest samples were valid and detected.
c One replicate was a “no test” (instrument error) and per protocol, the sample was retested in triplicate and all retest samples were valid and detected.
d Flumist was not tested for negative panel due to material availability
{21}
Interfering Substances – Nasal Swab (NS) Samples in the Abbott Universal Collection Kit or Abbott Universal Collection Kit II
An analytical study was performed to evaluate the performance of the Alinity m SARS-CoV-2 assay when testing nasal swab samples prepared using the Abbott Universal Collection Kit or the Abbott Universal Collection Kit II buffer and containing potentially interfering endogenous and exogenous substances.
The SARS-CoV-2 positive sample at 3X LoD was prepared by spiking gamma irradiated SARS-CoV-2 virus into pooled negative clinical NS specimens prepared using the Abbott Universal Collection Kit or the Abbott Universal Collection Kit II. A minimum of three replicates of each test condition were tested along with the control condition on the Alinity m SARS-CoV-2 assay. The results from this study are presented in Table 11.
Table 11. Interfering Substances Study Results – NS Specimens in UCKs Transport Buffer
| Potentially Interfering Substance | Active Ingredient | Level | Included Samples (N) | Negative (N) | Positive (N) | Invalid IC (N) |
| --- | --- | --- | --- | --- | --- | --- |
| Control | NA | NA | 3 | 0 | 3 | 0 |
| Blood (human) | NA | 10% (v/v) | 3 | 0 | 3 | 0 |
| Throat Lozenges, Oral Anesthetic and Analgesic - Cepacol | Benzocaine, Menthol | 5 mg/mL | 3 | 0 | 3 | 0 |
| Mucin - Bovine | Purified mucin protein | 5 mg/mL | 3 | 0 | 3 | 0 |
| Antibiotic, Nasal Ointment - Bactroban | Mupirocin | 5 mg/mL | 3 | 0 | 3 | 0 |
| Nasal Spray-Afrin | Oxymetazoline | 15% (v/v) | 3 | 0 | 3 | 0 |
| Anti-Viral Drug - Relenza | Zanamivir | 5 mg/mL | 3 | 0 | 3 | 0 |
| Anti-Viral Drug - Veklury | Remdesivir | 27.0 μM | 3 | 0 | 3 | 0 |
| Antibacterial, Systemic (Tobramycin) | Tobramycin | 4 μg/mL | 3 | 0 | 3 | 0 |
| Nasal Gel /Homeopathic Allergy Relief Medicine - Zicam | Galphimia glauca, Histaminum, hydrochloricum, Luffa operculata, Sulfur | 10% (v/v) | 3 | 0 | 3 | 0 |
| FluMist | Live intranasal influenza virus | 10% (v/v) | 3 | 0 | 3 | 0 |
| Nasal Corticosteroid-Flonase Sensimist | Fluticasone Furoate | 10% (v/v) | 3 | 0 | 3 | 0 |
| Nasal Corticosteroid - QVAR | Beclomethasone | 2% (v/v) | 3 | 0 | 3 | 0 |
| Corticosteroid - Dexamethasone | Dexamethasone | 0.2 mg/mL | 3 | 0 | 3 | 0 |
| Nasal Corticosteroid - Flunisolide | Flunisolide | 2% (v/v) | 3 | 0 | 3 | 0 |
K241580 - Page 22 of 38
{22}
| Nasal Corticosteroid - Triamcinolone | Triamcinolone | 2% (v/v) | 3 | 0 | 3 | 0 |
| --- | --- | --- | --- | --- | --- | --- |
| Nasal Corticosteroid - Budesonide | Budesonide | 2% (v/v) | 3 | 0 | 3 | 0 |
| Nasal Corticosteroid - Mometasone | Mometasone | 2% (v/v) | 3 | 0 | 3 | 0 |
| Toothpaste | Fluoride | 1% (w/v) | 5^{a} | 0 | 6 | 0 |
| Tobacco product | Nicotine | 0.1% (w/v) | 3 | 0 | 3 | 0 |
| Oral rinse- Listerine Cool Mint | Ethanol, essential oil | 10% (v/v) | 3 | 0 | 3 | 0 |
| Leukocytes | Leukocytes | 1.1E6 cells/mL | 3 | 0 | 3 | 0 |
| Nasal Decongestant - Phenylephrine | Phenylephrine | 2% (v/v) | 3 | 0 | 3 | 0 |
| Saline nasal mist | Sodium chloride | 2% (v/v) | 3 | 0 | 3 | 0 |
| Nicotine Product | Nicotine | 0.03 mg/mL | 3 | 0 | 3 | 0 |
| Chloroseptic Sore Throat Spray | Phenol | 5% (v/v) | 3 | 0 | 3 | 0 |
| Cough Syrup (e.g., Robitussin) | Dextromethorphan and guaifenesin | 5% (v/v) | 3 | 0 | 3 | 0 |
| Lidocaine & Phenylephrine (e.g., Liposomal NUMB250 Spray) | Lidocaine and phenylephrine | 2.68 mg/mL | 3 | 0 | 3 | 0 |
| Anti-Viral Drug – Tamiflu - Oseltamivir | Oseltamivir | 3.3 mg/mL | 3 | 0 | 3 | 0 |
| Petroleum Jelly (e.g., Vaseline) | Petroleum Jelly | 1% (w/v) | 3 | 0 | 3 | 0 |
| Analgesic Ointment (e.g., Vicks VapoRub) | Camphor-synthetic eucalyptus oil & menthol ointment | 1% (w/v) | 3 | 0 | 3 | 0 |
| Throat Lozenge, Oral Anesthetic & Analgesic (e.g., Cold Eeze) | Zincuum gluconicum | 2.5% (w/v) | 3 | 0 | 3 | 0 |
| Human Genomic DNA | NA | 0.02 mg/mL | 3 | 0 | 3 | 0 |
| Saliva (Human) | NA | 10% (v/v) | 3 | 0 | 3 | 0 |
<a> Three replicates from retest and 2 valid “Positive” from initial testing were included in the analysis per the protocol.
No interference with the Alinity m SARS-CoV-2 assay was observed in this analytical study when testing samples prepared using the Abbott Universal Collection Kit or Abbott Universal Collection Kit II buffer.
5. **Assay Reportable Range:**
Not applicable; this is a qualitative assay.
6. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
K241580 - Page 23 of 38
{23}
a. Controls
See Section IV.C.Instrument Descriptive Information.5.Quality Control.
b. Sample Stability
Sample Stability - Nasopharyngeal Swabs in UVT/UTM
An analytical study was performed to establish the stability for nasopharyngeal swab (NPS) specimens eluted in UVT/UTM for testing using the Alinity m SARS-CoV-2 assay. Positive samples were prepared by the dilution of gamma-irradiated SARS-CoV-2 viruses at $3 \times$ LoD in pooled SARS-CoV-2 negative clinical nasopharyngeal swab specimens collected in UVT/UTM. Aliquots of the positive sample were tested immediately (control condition) and at later time points after storage as indicated in Table 12.
Table 12. Sample Stability - Nasopharyngeal Swabs
| Storage Condition | Storage Condition | Tested (N) | Positive (N) | Negative (N) | Positive Rate |
| --- | --- | --- | --- | --- | --- |
| A (Control) | Tested immediately | 6 | 6 | 0 | 100% (6/6) |
| B (Test) | A minimum of 80 hours at 5°C ± 3°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 0 | 100% (6/6) |
| C (Test C) | A minimum of 7 days -70°C or colder (with a minimum of 1 freeze/thaw) plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 0 | 100% (6/6) |
| D (Test) | A minimum of 53 hours at room temperature followed by a minimum of 80 hours at 5°C ± 3°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 0 | 100% (6/6) |
| E (Test) | A minimum of 7 days at 5°C ± 3°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 0 | 100% (6/6) |
| F (Test) | A minimum of 80 hours at 5°C ± 3°C followed by a minimum of 7 days at -70°C or colder (with a minimum of 1 freeze/thaw) plus a minimum of 4 hours onboard storage before testing. | 6 | 6 | 0 | 100% (6/6) |
| A (Control) | Tested immediately | 6 | 6 | 0 | 100% (6/6) |
| G (Test) | A minimum of 90 days at -70°C or colder plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 0 | 100% (6/6) |
| A (Control) | Tested immediately | 6 | 6 | 0 | 100% (6/6) |
| H (Test) | A minimum of 2 days at 15°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 0 | 100% (6/6) |
K241580 - Page 24 of 38
{24}
K241580 - Page 25 of 38
# Sample Stability – Nasal Swabs in UVT/UTM
An analytical study was performed to establish stability for nasal swabs (NS) specimens eluted in UVT/UTM for testing using the Alinity m SARS-CoV-2 assay. Positive samples were prepared by the dilution of gamma-irradiated SARS-CoV-2 viruses at 1.5X LoD in pooled SARS-CoV-2 negative clinical nasal swab specimens collected in UVT/UTM. Aliquots of the positive sample were tested immediately (control condition) and at later time points after storage as indicated in Table 13.
Table 13. Stability Storage Conditions – Nasal Swabs in UVT/UTM
| Storage Condition | Storage Condition | Tested (N) | Positive (N) | Negative (N) | Positive Rate |
| --- | --- | --- | --- | --- | --- |
| A (Control) | Tested immediately | 6 | 6 | 6 | 100% (6/6) |
| B (Test) | A minimum of 80 hours at 5°C ± 3°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
| C (Test) | A minimum of 7 days -70°C or colder (with a minimum of 1 freeze/thaw) plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
| D (Test) | A minimum of 53 hours at 23.9°C followed by a minimum of 80 hours at 5°C ± 3°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
| E (Test) | A minimum of 7 days at 5°C ± 3°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
| F (Test) | A minimum of 80 hours at 5°C ± 3°C followed by a minimum of 7 days at -70°C or colder (with a minimum of 1 freeze/thaw) plus a minimum of 4 hours onboard storage before testing. | 6 | 6 | 6 | 100% (6/6) |
| A (Control) | Tested immediately | 6 | 6 | 6 | 100% (6/6) |
| G (Test) | A minimum of 90 days at -70°C or colder plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
| A (Control) | Tested immediately | 6 | 6 | 6 | 100% (6/6) |
| H (Test) | A minimum of 2 days at 15°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
| I (Test) | A minimum of 2 days at 26°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
{25}
Sample Stability - Nasal Swabs in the Abbott Universal Collection Kits
An analytical study was performed to establish stability for nasal swabs (NS) specimens collected in the Abbott Universal Collection Kits transport buffer. Positive samples were prepared by the dilution of gamma-irradiated SARS-CoV-2 viruses at 3x LoD in pooled SARS-CoV-2 negative clinical nasal swab specimens collected in Universal Collection Kits transport buffer.
Aliquots of the positive sample were tested immediately (control condition) and at later time points after storage as indicated in Table 14.
Table 14. Stability Storage Conditions—Universal Collection Kit Transport Buffer
| Storage Condition | Storage Condition | Tested (N) | Positive (N) | Negative (N) | Positive Rate |
| --- | --- | --- | --- | --- | --- |
| A (Control) | Tested immediately | 6 | 6 | 6 | 100% (6/6) |
| B (Test) | A minimum of 80 hours at 5°C ± 3°C plus a minimum of 4 hours onboard storage | 6 | 6 | 6 | 100% (6/6) |
| C (Test) | A minimum of 7 days at -70°C or colder plus a minimum of 4 hours onboard storage | 6 | 6 | 6 | 100% (6/6) |
| D (Test) | A minimum of 53 hours at room temperature (15°C to 30°C) followed by a minimum of 80 hours at 5°C ± 3°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
| E (Test) | A minimum of 7 days at 5°C ± 3°C plus a minimum of 4 hours onboard storage | 6 | 6 | 6 | 100% (6/6) |
| F (Test) | A minimum of 80 hours at 5°C ± 3°C followed by a minimum of 7 days at -70°C or colder plus a minimum of 4 hours onboard storage. | 6 | 6 | 6 | 100% (6/6) |
| G (Test) | A minimum of 7 days at -20°C or colder plus a minimum of 4 hours onboard storage. | 6 | 6 | 6 | 100% (6/6) |
| H (Test) | A minimum of 1 day at 40°C plus a minimum of 4 hours onboard storage. | 6 | 6 | 6 | 100% (6/6) |
| I (Test) | A minimum of 1 day at 45°C plus a minimum of 4 hours onboard storage. | 6 | 6 | 6 | 100% (6/6) |
| J (Test) | A minimum of 40°C for 6 hours, 22°C for 16 hours, 40°C for 2 hours, 35°C for 22 hours, and 40°C for 4 hours plus a minimum of 4 hours onboard storage. | 6 | 6 | 6 | 100% (6/6) |
| K (Test) | A minimum of 2 days at 40°C plus a minimum off 4 hours onboard storage. | 6 | 6 | 6 | 100% (6/6) |
K241580 - Page 26 of 38
{26}
| L (Test) | A minimum of 2 days at 45°C plus a minimum of 4 hours onboard storage. | 6 | 6 | 6 | 100% (6/6) |
| --- | --- | --- | --- | --- | --- |
| A (Control) | Tested immediately | 6 | 6 | 6 | 100% (6/6) |
| M (Test) | A minimum of 90 days at -70°C or colder plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
| A (Control) | Tested immediately | 6 | 6 | 6 | 100% (6/6) |
| N (Test) | A minimum of 2 days at 15°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
| O (Test) | A minimum of 2 days at 26°C plus a minimum of 4 hours onboard storage before testing | 6 | 6 | 6 | 100% (6/6) |
Stability study results for pre-collection storage conditions for the Abbott Universal Collection Kit supports claims for storage at $-20^{\circ}\mathrm{C}$ , $2^{\circ}\mathrm{C}$ to $8^{\circ}\mathrm{C}$ and $31^{\circ}\mathrm{C}$ to $55^{\circ}\mathrm{C}$ for 84 days and storage at $15^{\circ}\mathrm{C}$ to $30^{\circ}\mathrm{C}$ until the expiration date.
# Freeze-Thaw - Nasopharyngeal Swabs in UVT/UTM
A freeze-thaw study was conducted to evaluate specimen stability for the Alinity m SARS-CoV-2 assay using pooled SARS-CoV-2 negative nasopharyngeal swab specimens in UVT/UTM. Three panel members as shown in Table 15 were tested. Positive panel members were prepared by diluting gamma irradiated SARS-CoV-2 viruses in nasopharyngeal swab specimens collected in UVT/UTM. The results of the study are presented in Table 16.
Table 15. Nasopharyngeal Swab Specimens in UVT/UTM
| Panel Member | Target Concentration | Replicates per condition |
| --- | --- | --- |
| 1 | Negative | 10 |
| 2 | 2X LoD | 40 |
| 3 | 5X LoD | 10 |
Table 16. Freeze-Thaw Study - Nasopharyngeal Swab in UVT/UTM Results Summary
| Panel Member | Condition * | Tested (N) | Positive (N) | Negative (N) | Positive Rate |
| --- | --- | --- | --- | --- | --- |
| Negative | Control | 10 | 0 | 10 | 0.0% (0/10) |
| | One freeze/thaw | 10 | 0 | 10 | 0.0% (0/10) |
| | Two freeze/thaws | 10 | 0 | 10 | 0.0% (0/10) |
| 2X LoD | Control | 40 | 40 | 0 | 100.0% (40/40) |
| | One freeze/thaw | 40 | 40 | 0 | 100.0% (40/40) |
| | Two freeze/thaws | 40 | 40 | 0 | 100.0% (40/40) |
K241580 - Page 27 of 38
{27}
| 5X LoD | Control | 10 | 10 | 0 | 100.0% (10/10) |
| --- | --- | --- | --- | --- | --- |
| | One freeze/thaw | 10 | 10 | 0 | 100.0% (10/10) |
| | Two freeze/thaws | 10 | 10 | 0 | 100.0% (10/10) |
* Freezing at -70°C.
All control nasopharyngeal swab in UVT/UTM samples were 100% negative (0/10). At 2X LoD, 100% of the samples were positive for SARS-CoV-2 detection (40/40). At 5X LoD, 100% of the samples were positive for SARS-CoV-2 detection (10/10).
## Freeze-Thaw – Nasal Swabs in UVT/UTM
A freeze-thaw study was conducted to evaluate specimen stability for the Alinity m SARS-CoV-2 assay using pooled SARS-CoV-2 negative nasal swab specimens in UVT/UTM. The 1X LoD and 5X LoD positive panel members were prepared by diluting gamma irradiated SARS-CoV-2 viruses in pooled SARS-CoV-2 negative nasal swab specimens in UVT/UTM. Three panel members as shown in Table 17 were tested and results are presented in Table 18.
Table 17. Nasal Swab Specimens in UVT/UTM
| Panel Member | Target Concentration | Replicates per condition |
| --- | --- | --- |
| 1 | Negative | 10 |
| 2 | 1X LoD | 40 |
| 3 | 5X LoD | 10 |
Table 18. Freeze-Thaw Study – Nasal Swab in UVT/UTM Results Summary
| Panel Member | Condition * | Included Samples (N) | Positive (N) | Negative (N) | Positive Rate | Negative Rate |
| --- | --- | --- | --- | --- | --- | --- |
| Negative | Control | 10 | 0 | 10 | 0.0% (0/10) | 100.0% (10/10) |
| | One freeze/thaw | 10 | 0 | 10 | 0.0% (0/10) | 100.0% (10/10) |
| | Two freeze/thaws | 10 | 0 | 10 | 0.0% (0/10) | 100.0% (10/10) |
| 1 x LoD | Control | 40 | 40 | 0 | 100.0% (40/40) | 0.0% (0/40) |
| | One freeze/thaw | 42^{h} | 40 | 0 | 95.2% (40/42) | 0.0% (0/40) |
| | Two freeze/thaws | 42^{c} | 41 | 0 | 97.6% (41/42) | 0.0% (0/40) |
| 5 x LoD | Control | 10 | 10 | 0 | 100.0% (10/10) | 0.0% (0/10) |
| | One freeze/thaw | 10 | 10 | 0 | 100.0% (10/10) | 0.0% (0/10) |
| | Two freeze/thaws | 10 | 10 | 0 | 100.0% (10/10) | 0.0% (0/10) |
* Freezing at -70°C.
K241580 - Page 28 of 38
{28}
Two samples were valid but not detected. These samples were retested with another 2 aliquots stored at the same temperature and same number of freeze thaw cycles. All retest replicates were valid and detected.
One sample was valid but not detected. This sample was retested with another 2 aliquots stored at the same temperature and same number of freeze thaw cycles. Retest replicate was valid and detected.
# Freeze-Thaw – Nasal Swabs Collected in the Abbott Universal Collection Kits Transport Buffer
A freeze-thaw study was conducted to evaluate specimen stability for the Alinity m SARS-CoV-2 assay using pooled SARS-CoV-2 negative nasal swab specimens in the Abbott Universal Collection Kits transport buffer. The positive panel members were prepared by diluting gamma irradiated SARS-CoV-2 viruses in pooled SARS-CoV-2 negative nasal swab specimens collected in the Abbott Universal Collection Kits transport buffer. Three panel members as shown in Table 19 were tested. The results are presented in Table 20.
Table 19. Nasal Swab Specimens in Abbott Universal Collection Kits Transport Buffer
| Panel Member | Target Concentration | Replicates per condition |
| --- | --- | --- |
| 1 | Negative | 10 |
| 2 | 2X LoD | 40 |
| 3 | 5X LoD | 10 |
Table 20. Freeze-Thaw Study – Nasal Swab in Abbott Universal Collection Kits Transport Buffer Results Summary
| Panel Member | Condition^{a} | Included Samples (N) | Positive (N) | Negative (N) | Positive Rate | Negative Rate |
| --- | --- | --- | --- | --- | --- | --- |
| Negative | Control | 10 | 0 | 10 | 0.0% (0/10) | 100.0% (10/10) |
| | One freeze/thaw | 10 | 0 | 10 | 0.0% (0/10) | 100.0% (10/10) |
| | Two freeze/thaws | 10 | 0 | 10 | 0.0% (0/10) | 100.0% (10/10) |
| 2X LoD | Control | 40 | 40 | 0 | 100.0% (40/40) | 0.0% (0/40) |
| | One freeze/thaw | 40 | 40 | 0 | 100.0% (40/40) | 0.0% (0/40) |
| | Two freeze/thaws | 40 | 40 | 0 | 100.0% (40/40) | 0.0% (0/40) |
| 5X LoD | Control | 10 | 10 | 0 | 100.0% (10/10) | 0.0% (0/10) |
| | One freeze/thaw | 10 | 10 | 0 | 100.0% (10/10) | 0.0% (0/10) |
| | Two freeze/thaws | 10 | 10 | 0 | 100.0% (10/10) | 0.0% (0/10) |
Freezing at - 70°C
# 7. Detection Limit:
Limit of detection (LoD) studies were performed to evaluate the analytical sensitivity of the Alinity m SARS-CoV-2 Assay.
K241580 - Page 29 of 38
{29}
Limit of Detection (LoD) - Nasopharyngeal Swabs in UVT/UTM
The LoD of the Alinity m SARS-CoV-2 assay was determined with the analysis of a dilution series of inactivated SARS-CoV-2 virus in nasopharyngeal swab specimens in UVT/UTM. Dilutions of gamma irradiated SARS-CoV-2 virus were prepared in pooled SARS-CoV-2 negative clinical nasopharyngeal swab (NPS) specimens in UVT/UTM. The preliminary LoD assessment was carried out by testing replicates of three at the following SARS-CoV-2 target concentrations $(\mathrm{TCID}_{50} / \mathrm{mL})$ : 0.0180, 0.0090, 0.0045, and 0. The preliminary LoD of 0.0090 $\mathrm{TCID}_{50} / \mathrm{mL}$ was confirmed by testing 21 replicates. The results are presented in Table 21.
Table 21. LoD in Clinical Nasopharyngeal Swabs in UVT/UTM.
| SARS-CoV-2 Concentration (TCID50/mL)a | Number of Replicates Tested | Number of Replicates Detected | Detection Rate (%) |
| --- | --- | --- | --- |
| 0.0045 | 21 | 7 | 33.3% |
| 0.0090 | 21 | 21 | 100.0% |
| 0.0180 | 21 | 20 | 95.2% |
a $\mathrm{TCID}_{50} / \mathrm{mL} =$ Median Tissue Culture Infectious Dose/mL
# Limit of Detection (LoD) - Nasal Swabs in UVT/UTM
The LoD of the Alinity m SARS-CoV-2 assay was determined with the analysis of a dilution series of inactivated SARS-CoV-2 virus with nasal swab specimens in UVT/UTM. Dilutions of gamma irradiated SARS-CoV-2 virus were prepared in pooled SARS-CoV-2 negative clinical nasal swab (NS) specimens in UVT/UTM. The preliminary LoD assessment was conducted by testing replicates of three at the following SARS-CoV-2 target concentrations $(\mathrm{TCID}_{50} / \mathrm{mL})$ : 0.0180, 0.0090, 0.0045, and 0. The preliminary LoD was 0.0090 $\mathrm{TCID}_{50} / \mathrm{mL}$ . LoD confirmation was carried out by testing replicates of 21 at the preliminary LoD and two additional levels. The LoD was confirmed to be 0.0180 $\mathrm{TCID}_{50} / \mathrm{mL}$ . Results are presented in Table 22.
Table 22. LoD in Clinical Nasal Swabs in UVT/UTM
| SARS-CoV-2 Concentration (TCID50/mL)a | Number of Replicates Tested | Number of Replicates Detected | Detection Rate (%) |
| --- | --- | --- | --- |
| 0.0045 | 21 | 9 | 42.9% |
| 0.0090 | 21 | 17 | 81.0% |
| 0.0180 | 21 | 21 | 100.0% |
a $\mathrm{TCID}_{50} / \mathrm{mL} =$ Median Tissue Culture Infectious Dose/mL
# Limit of Detection (LoD) - Nasal Swabs in the Abbott Universal Collection Kits Transport Buffer
The LoD of the Alinity m SARS-CoV-2 assay was determined with the analysis of a dilution series of inactivated SARS-CoV-2 virus with nasal swab specimens collected in
K241580 - Page 30 of 38
{30}
the Abbott Universal Collection Kits. Dilutions of gamma irradiated SARS-CoV-2 virus were prepared in pooled SARS-CoV-2 negative clinical nasal swab (NS) specimens collected in the Abbott Universal Collection Kits transport buffer. A five-member panel with the following target concentrations was tested in replicates of three in the range finding study: 0, 13.5, 27, 55, and $109\mathrm{GE / mL}$ . The three highest concentrations were repeated with 20 replicates each. The lowest concentration with a detection rate of $95\%$ or greater was $55\mathrm{GE / mL}$ . Results are presented in Table 23.
Table 23. LoD in Clinical Nasal Swabs in Abbott Universal Collection Kits Transport Buffer.
| SARS-CoV-2 Concentration (GE/mL)a | Number of Replicates Tested | Number of Replicates Detected | Detection Rate (%) |
| --- | --- | --- | --- |
| 27.0 | 20 | 4 | 20.0% |
| 55.0 | 20 | 20 | 100.0% |
| 109.0 | 20 | 20 | 100.0% |
aGE/mL = Genome Equivalent/mL
# Limit of Detection (LoD) - SARS-CoV-2 WHO International Standard in SNM
An LoD study was performed to evaluate the LoD of the Alinity m SARS-CoV-2 Assay with the World Health Organization (WHO) Internal Standard for SARS-CoV-2. The WHO $1^{\text{st}}$ International Standard for SARS-CoV-2 RNA (NIBSC code: 20/146) was diluted in negative Simulated Nasal Matrix (SNM). The dilution panel of the SARS-CoV-2 WHO standard comprised of 8 panel members at the following concentrations: 125, 100, 75, 50, 25, 10, 5 and 1 IU/mL. Twenty-four replicates per panel member were tested. The detection rates observed for each panel are summarized in Table 24. The LoD for the Alinity m SARS-CoV-2 assay with the WHO International Standard was 100 IU/mL.
Table 24. LoD Confirmation Results - WHO SARS-CoV-2 International Standard in SNM.
| Panel Member | Target Concentration (IU/mL) | Number of Replicates Detected | Number of Replicates Tested | Detection Rate (%) |
| --- | --- | --- | --- | --- |
| 1 | 125 | 24 | 24 | 100.0 |
| 2 | 100 | 24 | 24 | 100.0 |
| 3 | 75 | 21 | 24 | 87.5 |
| 4 | 50 | 20 | 24 | 83.3 |
| 5 | 25 | 16 | 24 | 66.7 |
| 6 | 10 | 7 | 24 | 29.2 |
| 7 | 5 | 1 | 24 | 4.2 |
| 8 | 1 | 0 | 24 | 0.0 |
K241580 - Page 31 of 38
{31}
# 8. Assay Cut-Off:
The assay was designed to report results through all 42 PCR cycles—the maximum number of cycles for this assay. When an amplification of specimen occurred between PCR cycles 1 to 42 (CN) and met the MR threshold (PCR efficiency-related maximum ratio (MR)) the specimen was interpreted as Positive initially. Subsequently, a CN cutoff for the Alinity m SARS-CoV-2 assay was selected at 39.57.
# 9. Carry-Over:
An analytical study was performed to assess potential carry-over or cross-contamination in the Alinity m SARS-CoV-2 assay by testing high positive and negative samples in an alternating fashion on the Alinity m System. Testing with each AMP Tray consisted of 24 replicates of high-positive sample prepared at $2.0 \times 10^{9}$ Copies/mL of SARS-CoV-2 material and 24 replicates of negative sample. Sample replicates were arranged such that high-positive and negative samples are processed in alternating reaction vessels. The carry-over contamination study tested an array of 363 high-positive samples and 361 negative samples across five Alinity m SARS-CoV-2 AMP Trays on each of three Alinity m systems. Results are presented in Table 25.
Table 25. Alinity m System Carry-Over Rate.
| Sample Type | Tested (N) | Negative (N) | Positive (N) | Carry-over Rate | 95% Confidence Interval |
| --- | --- | --- | --- | --- | --- |
| Negative | 361 | 361 | 0 | 0.0% (0/361) | (0.0%,1.1%) |
None of the 361 negative samples exhibited any evidence of carry-over contamination from the high-positive samples to SARS-CoV-2 negative samples (0/361).
# B Comparison Studies:
See Clinical Studies section below.
# 1. Matrix Comparison:
Simulated Nasal Matrix (SNM) and Clinical Matrix Equivalency Study
To evaluate the equivalence of SNM used in analytical studies with natural clinical matrix, SNM was formulated and tested side by side with clinical nasopharyngeal swab (NPS) specimens collected in UVT/UTM, clinical nasal swab (NS) specimens in collected UVT/UTM, and clinical NS specimens collected in the Universal Collection Kits (UCK) transport buffer.
K241580 - Page 32 of 38
{32}
Each matrix was evaluated by testing four panel members: a negative panel member and three positive panel members which were prepared by the spiking cultured SARS-CoV-2 virus in each type of matrix at $0.5\mathrm{x}$ LoD, 1x LoD, and 3x LoD. For each matrix and panel member 21 replicates were tested to achieve a minimum of 20 replicates. The testing was performed using Alinity m SARS-CoV-2 assay reagents and the Alinity m System. Results are summarized Table 26.
Table 26. Alinity m SARS-CoV-2 Matrix Equivalency Study Results
| | | Number of Replicates | | | Positive Rate (%) |
| --- | --- | --- | --- | --- | --- |
| Matrix | Target Level | Total Tested | Negative | Positive | |
| Simulated Nasal Matrix (SNM) | Negative | 20 | 20 | 0 | 100.0% |
| | 0.5x LoD | 21 | 6 | 15 | 71.4% |
| | 1x LoD | 21 | 1 | 20 | 95.2% |
| | 3x LoD | 21 | 0 | 21 | 100.0% |
| Pooled Clinical Nasopharyngeal Swabs in UVT/UTM | Negative | 21 | 21 | 0 | 100.0% |
| | 0.5x LoD | 21 | 4 | 17 | 81.0% |
| | 1x LoD | 21 | 0 | 21 | 100.0% |
| | 3x LoD | 21 | 0 | 21 | 100.0% |
| Pooled Clinical Nasal Swabs in UVT/UTM | Negative | 21 | 21 | 0 | 100.0% |
| | 0.5x LoD | 21 | 1 | 20 | 95.2% |
| | 1x LoD | 21 | 0 | 21 | 100.0% |
| | 3x LoD | 20 | 1 | 19 | 95.0% |
| Pooled Clinical Nasal Swabs in Universal Collection Kits (UCK) Transport Buffer | Negative | 21 | 21 | 0 | 100.0% |
| | 0.5x LoD | 21 | 5 | 16 | 76.2% |
| | 1x LoD | 21 | 1 | 20 | 95.2% |
| | 3x LoD | 21 | 1 | 20 | 95.2% |
The results demonstrate the equivalency of all matrices evaluated, including the SNM.
K241580 - Page 33 of 38
{33}
# C Clinical Studies:
# Prospective Clinical Studies
The clinical performance of the Alinity m SARS-CoV-2 assay was established in two clinical studies that tested prospective clinical specimens in viral transport media (UVT or UTM) and in Universal Collection Kit (UCK) from individuals with signs and symptoms of respiratory infection.
# Composite Comparator (CC) Definition
A minimum of two and up to three highly sensitive SARS-CoV-2 molecular EUA assays were used as comparator assays to establish the composite comparator (CC) for assessing the performance of the Alinity m SARS-CoV-2 assay. A specimen was categorized as CC positive if a minimum of two comparator positive results were reported. A specimen was categorized as CC negative if two or more comparator results were negative. A specimen was categorized as indeterminate (IND) if CC could not be determined due to missing results from the comparator assays. Specimens with an Indeterminate CC result were excluded from analyses.
# I. Prospective Clinical Study - Nasopharyngeal Swab Specimens
Nasopharyngeal swabs (NPS) were prospectively collected at eight geographically distinct sites in the U.S. Specimens were collected in BD Universal Viral Transport (UVT) with a flocked swab between January 2021 and February 2021. Of the 627 NPS specimens from symptomatic subjects tested, 580 samples were Category I specimens, while 47 were Category II specimens. Of the initial 627 specimens tested, 16 were excluded as invalid with no result obtained and 76 specimens were excluded for having no composite comparator result. Five hundred thirty-five NPS specimens were included for performance analysis. Patient demographic information for the 535 specimens with valid CC results that were included in performance analysis in the prospective clinical study are presented in Table 27.
Table 27 Subject Demographics - Nasopharyngeal Specimen Prospective Clinical Study
| Demographic Characteristic | Statistic |
| --- | --- |
| Age | |
| n | 535 |
| Mean | 40 |
| Median | 39 |
| Minimum | 2 |
| Maximum | 87 |
K241580 - Page 34 of 38
{34}
| Age Group | n (%) |
| --- | --- |
| Birth to 5 years | 4 (0.7%) |
| 6 to 21 years | 67 (12.5%) |
| 22 to 59 years | 380 (71.0%) |
| ≥ 60 years | 84 (15.7%) |
A summary performance of the Alinity m SARS-CoV-2 assay with prospectively collected nasopharyngeal swab specimen in UVT is provided in Table 28.
Table 28. Summary of Performance of the Prospective Fresh (Category I) and Frozen (Category II) Nasopharyngeal Swab Specimens in UVT
| Fresh or Frozen | N | TP | FN | TN | FP | PPA (%) (95% CI) | NPA (%) (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Fresh | 489 | 152 | 6 | 316 | 15 | 96.2 (92.0-98.3) | 95.5 (92.7-97.2) |
| Frozen | 46 | 2 | 0 | 41 | 3 | 100.0 (34.2-100.0) | 93.2 (81.8-97.7) |
| All | 535 | 154 | 6 | 357 | 18 | 96.3 (92.1-98.3) | 95.2 (92.5-96.9) |
TP = true positive; FN = false negative; TN = true negative; FP = false positive, PPA = Positive Percent Agreement, NPA = Negative Percent Agreement.
Of the 627 specimens tested in the nasopharyngeal swab prospective study, 609 gave valid results with 18 initial invalid samples. The initial invalid rate was 2.9% (95% CI: 1.8% - 4.5%). Upon retest, only one sample gave a valid result with 17 samples remaining invalid for a final invalid rate of 2.7% (95% CI: 1.7% - 4.3%).
## II. Prospective Clinical Study – Anterior Nasal Swab Specimens
This prospective clinical study used anterior nasal swab (ANS) specimens (two per patient) that were self-collected under HCP supervision in the Abbott Universal Collection Kit II (UCK) and in the Universal Viral Transport (UVT). ANS samples were collected between September 2021 and January 2022. Adult and pediatric subjects, presenting signs and symptoms consistent with respiratory tract infection and/or COVID-19 were enrolled for specimen collection. Subject demographics are presented in Table 29.
Table 29. Summary of Subject Demographics - Anterior Nasal Specimen Prospective Clinical Study
| Demographic Characteristic | Statistic |
| --- | --- |
| Age | |
| n | 785 |
| Mean | 36 |
| Median | 34 |
| Minimum | 7 |
K241580 - Page 35 of 38
{35}
Of the 759 ANS UVT specimens included in the analysis with CC, 599 were tested fresh (Category I) and 160 were tested after being stored frozen (Category II). Of the 766 ANS UCK specimens included in the analysis with CC, 603 were tested fresh (Category I) and 163 were tested after being stored frozen (Category II). Performance results are presented in Table 30 for specimens collected in UVT and Table 31 for specimens collected in UCK.
Table 30. Summary of the Clinical Performance of Prospective Fresh (Category I) and Frozen (Category II) Self-Collected ANS Samples (under HCP supervision) in UVT
| Specimen Type | Fresh or Frozen | N | TP | FN | TN | FP | PPA (%) (95% CI) | NPA (%) (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Anterior Nasal Swab (Self-Collected, under HCP supervision) UVT | Fresh | 599 | 82 | 0 | 516 | 1 | 100 (95.5-100) | 99.8 (98.9-100.) |
| | Frozen | 160 | 14 | 0 | 145 | 1 | 100 (78.5-100) | 99.3 (96.2-99.9) |
| Combined | All | 759 | 96 | 0 | 661 | 2 | 100 (96.2-100) | 99.7 (98.9-99.9) |
TP = true positive; FN = false negative; TN = true negative; FP = false positive, PPA = Positive Percent Agreement, NPA = Negative Percent Agreement.
Table 31. Summary of the Clinical Performance of Prospective Fresh (Category I) and Frozen (Category II) Self-Collected ANS Samples (under HCP supervision) using UCK
| Specimen Type | Fresh or Frozen | N | TP | FN | TN | FP | PPA (%) (95% CI) | NPA (%) (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Anterior Nasal Swab (Self-Collected under HCP supervision) UCK | Fresh | 603 | 80 | 2 | 511 | 10 | 97.6 (91.5-99.3) | 98.1 (96.5-99.0) |
| | Frozen | 163 | 14 | 0 | 145 | 4 | 100 (78.5-100) | 97.3 (93.3-99.0) |
| Combined | All | 766 | 94 | 2 | 656 | 14 | 97.9 (92.7-99.4) | 97.9 (96.5-98.8) |
TP = true positive; FN = false negative; TN = true negative; FP = false positive, PPA = Positive Percent Agreement, NPA = Negative Percent Agreement.
Of the initial 787 anterior nasal specimens collected in UVT, 21 gave an invalid result. With a total of 766 valid specimens remaining, the initial invalid rate was 2.7% (95% CI: 1.8% - 4.0%). Upon retesting of the 21 invalid specimens, 20 were found to be valid with 1 specimen remaining invalid. The final invalid rate was 0.1% (95% CI: 0.0% - 0.7%).
Of the 792 anterior nasal specimens collected in UCK, 52 specimens gave an invalid result with 740 samples providing valid test results for an initial invalid rate of 6.6% (95% CI: 5.0% - 8.5%). Upon retesting of the 52 invalid specimens, 51 gave a valid result for a final invalid rate of 0.1% (95% CI: 0.0% - 0.7%).
K241580 - Page 36 of 38
{36}
D Clinical Cut-Off:
Not Appli…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.