← Product Code [QOF](/submissions/MI/subpart-d%E2%80%94serological-reagents/QOF) · K242526

# Visby Medical Respiratory Health Test (K242526)

_Visby Medical, Inc. · QOF · Feb 19, 2025 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/QOF/K242526

## Device Facts

- **Applicant:** Visby Medical, Inc.
- **Product Code:** [QOF](/submissions/MI/subpart-d%E2%80%94serological-reagents/QOF.md)
- **Decision Date:** Feb 19, 2025
- **Decision:** SESE
- **Submission Type:** Dual Track
- **Regulation:** 21 CFR 866.3981
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Pediatric

## Indications for Use

The Visby Medical Respiratory Health Test is a single-use (disposable), fully integrated, automated Reverse Transcription Polymerase Chain Reaction (RT-PCR) in vitro diagnostic test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, and influenza B RNA in nasopharyngeal swab and anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, and influenza B can be similar. The Visby Medical Respiratory Health Test is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, and influenza B infection if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, and influenza B viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections. Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other organisms not detected by the test. The agent(s) detected by the Visby Medical Respiratory Health Test may not be the definitive cause of disease. Negative results do not preclude SARS-CoV-2, influenza A, or influenza B infection. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

## Device Story

Single-use, fully integrated, automated RT-PCR diagnostic test; performs lysis, reverse transcription, PCR amplification, and detection within a compact device. Input: nasopharyngeal or anterior nasal swab specimens placed in buffer. Operation: sample enters lysis module; RNA reverse transcribed to cDNA; PCR reagents amplify targets (including B2M process control); biotinylated product hybridized to capture probes in flow channel; enzymatic reaction with HRP and colorimetric substrate forms purple precipitate. Output: visual color change at specific locations on device indicating presence of SARS-CoV-2, influenza A, or influenza B. Used in CLIA-waived settings (urgent/family care) by non-laboratorians. Results in ~30 minutes. Aids differential diagnosis; results combined with clinical/epidemiological data for patient management.

## Clinical Evidence

Prospective multicenter study (n=1,501) at 5 CLIA-waived sites. Compared Visby Test to FDA-cleared/authorized NAATs. Results: SARS-CoV-2 PPA 97.2%, NPA 99.0%; Influenza A PPA 96.9%, NPA 99.4%; Influenza B PPA 100%, NPA 99.8%. Study included untrained operators and diverse age range (1 month to 89 years).

## Technological Characteristics

Single-use, integrated RT-PCR device. Employs lysis module, reverse transcription, and thermal cycling for amplification. Detection via enzyme-linked colorimetric assay (streptavidin-HRP) producing purple precipitate. Includes internal process control (B2M RNA) and electronic controls. Powered by external adapter. Standards: IEC 60601-1-2, IEC 61010-1, IEC 61010-2-101, ISO 14971.

## Regulatory Identification

A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
*e.g.,* BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
*e.g.,* when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
*e.g.,* regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
*e.g.,* saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
*i.e.,* H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.

## Predicate Devices

- DiaSorin Molecular Simplexa COVID-19 & Flu A/B Direct ([K220963](/device/K220963.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY AND INSTRUMENT

## I Background Information:

A 510(k) Number

K242526

B Applicant

Visby Medical, Inc.

C Proprietary and Established Names

Visby Medical Respiratory Health Test

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  QOF | Class II | 21 CFR 866.3981 - Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The Sars-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test | MI - Microbiology  |

## II Submission/Device Overview:

### A Purpose for Submission:

To demonstrate that the performance of the Visby Medical Respiratory Health Test is equivalent to the performance of the FDA cleared DiaSorin Molecular Simplexa COVID-19 &amp; Flu A/B Direct (K220963).

### B Measurand:

- Influenza A RNA
- Influenza B RNA
- Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) RNA

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

C Type of Test:
Qualitative RT-PCR

III Intended Use/Indications for Use:

A Intended Use(s):
See Indications for Use below.

B Indication(s) for Use:
The Visby Medical Respiratory Health Test is a single-use (disposable), fully integrated, automated Reverse Transcription Polymerase Chain Reaction (RT-PCR) in vitro diagnostic test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, and influenza B RNA in nasopharyngeal swab and anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, and influenza B can be similar.

The Visby Medical Respiratory Health Test is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, and influenza B infection if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, and influenza B viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections.

Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other organisms not detected by the test. The agent(s) detected by the Visby Medical Respiratory Health Test may not be the definitive cause of disease. Negative results do not preclude SARS-CoV-2, influenza A, or influenza B infection. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For In Vitro Diagnostic Use Only

D Special Instrument Requirements:
Single-use test without special instrument requirement.

IV Device/System Characteristics:

A Device Description:
The Visby Medical Respiratory Health Test (referred to as "Visby Test" hereafter) is a single-use (disposable), fully integrated, compact device containing a reverse transcription polymerase chain reaction (RT-PCR) based assay for qualitative detection and differentiation of RNA from influenza A, influenza B, and SARS-CoV-2 viruses from nasopharyngeal swab (NPS) and anterior nasal swab (ANS) specimens.

K242526 - Page 2 of 20

{2}

The test system includes the Visby Medical Respiratory Health Test, the reusable Visby Medical power adaptor, the Visby Medical Respiratory Health Buffer Tube, and fixed-volume transfer pipette. Operators are instructed to collect the sample and elute it into the Visby Respiratory Health Buffer.

A test operator transfers the collection media (containing patient specimen) into the sample port of the device using the provided fixed-volume pipette and then slides a purple switch on the front of the device to both close the sample port and initiate the fully automated testing process. The device processes NPS and ANS samples by automatically performing all steps required to complete lysis, reverse transcription, polymerase chain reaction, and amplicon detection.

There are LED lights on the lower left corner of the device that are used to communicate the test status. Stable white light in the shape of a power symbol will appear indicating the test is plugged in. In addition, three progress lights to the right of the power light will each blink for ~10 minutes in sequence and then provide a stable white light to indicate the progression of the test. At the conclusion of the test, a green LED under "DONE" next to the white lights signals that the test run completed successfully.

When the run has completed, the operator visually interprets the status lights on the front of the device and the colorimetric output in the detection window. A valid test will have both an illuminated green LED under "DONE" and a purple spot adjacent to "RESULTS VALID" in the detection window. For valid tests, a purple spot next to the name of a pathogen indicates a positive result for that target.

In the case of a lack of a purple "RESULTS VALID" spot, the test is invalid, and the operator is instructed to repeat the test. If the subsequent test is also invalid, the operator is instructed to collect a new sample and repeat the test with a new device.

The device has built-in procedural controls. These include an internal process control and built-in electronic control. The internal control is designed to ensure that all steps in the testing process including reverse transcription, amplification of target sequences, and amplicon detection are working properly. The electronic control monitors if device is being operated outside of its temperature range or if there are any other device errors. The result of the internal process control is displayed in the detection window while the results of the electronic controls are displayed using the status lights.

Third-party external positive and negative controls are recommended for use with this test. These controls are not provided with the device and must be purchased separately by the customer directly from Microbiologics. The labeling states that the external controls "must be tested once with each new shipment received and once for each untrained operator."

## B Principle of Operation:

The device houses all the reagents to perform the testing process. The sample enters a lysis module and rehydrates the RT enzyme and RT primers. The mixture then moves through a sample preparation module where viruses and human cells are simultaneously lysed, and RNA is reverse transcribed. The resulting fluid (containing cDNA) is then mixed with lyophilized PCR reagents containing the DNA polymerase enzyme and PCR primers. The PCR mixture (containing cDNA template and reagents) is then thermal cycled to amplify the targets, including

K242526 - Page 3 of 20

{3}

human beta-2 microglobulin (B2M) RNA, which serves as an internal process control. If present, the amplified pathogen target (SARS-CoV-2, influenza A, and/or influenza B) and the amplified internal process control hybridize to specific probes located on a flow cell. For SARS-CoV-2, primer sets are included targeting the nucleocapsid (N) gene and the open reading frame 1ab (ORF-1ab) gene. The primer sets for influenza A target the matrix protein 1 (M1) gene. For influenza B, the primer set targets the nuclear export protein (NEP) and nonstructural protein 1 (NS1) genes. Detection of the target-specific PCR product is accomplished via an enzyme-linked colorimetric assay using streptavidin bound horseradish peroxidase (HRP) and a colorimetric substrate that forms a purple precipitate. Test results can be expected in approximately 30 minutes and are visually interpreted by the operator.

V Substantial Equivalence Information:

A Predicate Device Name(s):

Simplexa COVID-19 &amp; Flu A/B Direct

B Predicate 510(k) Number(s):

K220963

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K242526 | K220963  |
| --- | --- | --- |
|  Device Trade Name | Visby Medical Respiratory Health Test | DiaSorin Molecular Simplexa COVID-19 & Flu A/B Direct  |
|  General Device Characteristic Similarities | K242526 | K220963  |
|  Product Code | Same | QOF  |
|  Regulation | Same | 21 CFR 866.3981  |
|  Prescription Use Only | Same | Yes  |
|  Technology | Same | Real-time Reverse Transcription Polymerase Chain Reaction (RT-PCR)  |
|  Detection | Same | Qualitative  |
|  Intended Use/Indications For Use | The Visby Medical Respiratory Health Test is a single-use (disposable), fully integrated, automated Reverse Transcription Polymerase Chain Reaction (RT-PCR) in vitro diagnostic test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, and influenza B RNA in | The DiaSorin Molecular Simplexa COVID-19 & Flu A/B Direct is a real-time RT-PCR assay intended for use on the LIAISON MDX instrument for the in vitro qualitative detection and differentiation of nucleic acid from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus and influenza B virus in nasopharyngeal swabs (NPS) from individuals with signs  |

K242526 - Page 4 of 20

{4}

K242526 - Page 5 of 20

|   | nasopharyngeal swab and anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, and influenza B can be similar.
The Visby Medical Respiratory Health Test is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, and influenza B infection if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, and influenza B viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections.
Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other organisms not detected by the test. The agent(s) detected by the Visby Medical Respiratory Health Test may not be the definitive cause of disease.
Negative results do not preclude SARS-CoV-2, influenza A, or influenza B infection. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. | and symptoms of respiratory tract infection.
The Simplexa COVID-19 & Flu A/B Direct assay is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A and influenza B infection. Negative results do not preclude SARS-CoV-2, influenza A or influenza B infection and should not be used as the sole basis for patient management decisions. Positive results do not rule out coinfections with other organisms. Results should be combined with clinical observations, patient history, and epidemiological information.
The Simplexa COVID-19 & Flu A/B Direct assay is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.  |
| --- | --- | --- |
|  Specimen | Nasopharyngeal Swab; Anterior Nasal Swab | Nasopharyngeal Swab  |
|  Assay Results | Same | Qualitative  |
|  Organisms Detected | Same | SARS-CoV-2, influenza A virus and influenza B virus  |

{5}

K242526 - Page 6 of 20

|  Assay Controls | Same | Internal processing control (IPC), External controls available.  |
| --- | --- | --- |
|  General Device Characteristic Differences | K242526 | K220963  |
|  Result Interpretation | Visual | Automatic  |

## VI Standards/Guidance Documents Referenced:

1. 21 CFR 866.3981 - Special Controls
2. ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] 19-36
3. IEC 60601-1-2: 2014 +Al: 2020
4. IEC TR 60601-4-2 Edition 1.0 2016-05 19-19
5. IEC 61010-1:2010
6. IEC 61010-1:2010/AMD1:2016
7. IEC 61010-1 Edition 3.1 2017-01 CONSOLIDATED VERSION 19-34
8. IEC 61010-2-101:2018
9. IEC 61010-2-010:2019
10. ANSI AAMI IEC 62304:2006/Al:2016 13-79
11. ANSI AAMI ISO 14971:2019 5-125

## VII Performance Characteristics (if/when applicable):

### A Analytical Performance:

#### 1. Precision/Reproducibility:

A precision/reproducibility study was conducted with the Visby Medical Respiratory Health Test at three CLIA-waived external sites by a total of six untrained operators (two operators at each site). Each operator tested a panel (Table 1) in triplicate each day across six non-consecutive days. The panel consisted of seven members (live virus for influenza A and influenza B, and inactivated virus for SARS-CoV-2) prepared by spiking organisms in negative pooled clinical swab matrix and applied to individual dry nasopharyngeal swabs. Three reagent lots were used in this study. Results were generated from a total of 108 replicates of each panel member (3 sites x 6 days x 2 operators/site x 3 replicates/operator/day).

Table 1. Precision/Reproducibility Study Sample Panel

|  Target | Concentration | Concentration (copies/swab)  |
| --- | --- | --- |
|  Negative | N/A | N/A  |
|  Influenza A, H1N1pdm, Brisbane/02/18 | Low (1x LoD) | 106  |
|  Influenza A, H1N1pdm, Brisbane/02/18 | Moderate (4x LoD) | 424  |
|  Influenza B Victoria, Washington/02/19 | Low (1x LoD) | 728  |
|  Influenza B Victoria, Washington/02/19 | Moderate (4x LoD) | 2912  |
|  SARS-CoV-2 USA-WA1/2020 | Low (1x LoD) | 100  |
|  SARS-CoV-2 USA-WA1/2020 | Moderate (4x LoD) | 400  |

{6}

A summary of percent agreement with expected results for each panel member by site and overall is presented in the following table. This study demonstrates that untrained users could perform the Visby Test accurately including testing samples with organism concentrations near the assay LoD.

Table 2. Precision/Reproducibility Study – Site to Site Results

|  Panel Member | % Agreement with Expected Results (Count) |   |   | Overall Agreement  |   |
| --- | --- | --- | --- | --- | --- |
|   |  Site 1 | Site 2 | Site 3 | % Agreement (count) | 95% CI  |
|  Influenza A
1x LoD | 94.4%
(34/36) | 100%
(36/36) a | 88.9%
(32/36) | 94.4%
(102/108) a | 88.4%-97.4%  |
|  Influenza A
4x LoD | 100%
(36/36) | 100%
(36/36) | 100%
(36/36) | 100.0%
(108/108) | 96.6%-100.0%  |
|  Influenza B
1x LoD | 97.2%
(35/36) | 100%
(36/36) | 91.7%
(33/36) | 96.3%
(104/108) | 90.9%-98.6%  |
|  Influenza B
4x LoD | 97.2%
(35/36) | 100%
(36/36) | 100%
(36/36) | 99.1%
(107/108) | 94.9%-99.8%  |
|  SARS-CoV-2
1x LoD | 97.2%
(35/36) | 100%
(36/36) | 88.9%
(32/36) | 95.4%
(103/108) | 89.6%-98.0%  |
|  SARS-CoV-2
4x LoD | 100%
(36/36) | 100%
(36/36) | 100%
(36/36) | 100.0%
(108/108) | 96.6%-100.0%  |
|  Negative | 100%
(36/36) | 100%
(36/36) | 100%
(36/36) | 100.0%
(108/108) | 96.6%-100.0%  |

a One low positive (1x LoD) influenza A sample was an unexpected positive for SARS-CoV-2.

Table 3. Precision/Reproducibility Study – Lot-to-Lot Results

|  Panel Member | % Agreement with Expected Results (Count) |   |   | Overall Agreement  |   |
| --- | --- | --- | --- | --- | --- |
|   |  Lot 1 | Lot 2 | Lot 3 | % Agreement (count) | 95% CI  |
|  Influenza A
1x LoD | 88.9%
(32/36) | 100.0%
(36/36) | 94.4%
(34/36) a | 94.4%
(102/108) a | 88.4%-97.4%  |
|  Influenza A
4x LoD | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(108/108) | 96.6%-100.0%  |
|  Influenza B
1x LoD | 97.2%
(35/36) | 91.7%
(33/36) | 100.0%
(36/36) | 96.3%
(104/108) | 90.9%-98.6%  |
|  Influenza B
4x LoD | 100.0%
(36/36) | 97.2%
(35/36) | 100.0%
(36/36) | 99.1%
(107/108) | 94.9%-99.8%  |
|  SARS-CoV-2
1x LoD | 88.9%
(32/36) | 97.2%
(35/36) | 100.0%
(36/36) | 95.4%
(103/108) | 89.6%-98.0%  |
|  SARS-CoV-2
4x LoD | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(108/108) | 96.6%-100.0%  |
|  Negative | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(108/108) | 96.6%-100.0%  |

a One low positive (1x LoD) influenza A sample was an unexpected positive for SARS-CoV-2.

K242526 - Page 7 of 20

{7}

a One low positive (1x LoD) influenza A sample was an unexpected positive for SARS-CoV-2.

Table 4. Precision/Reproducibility Study - Operator-to-Operator Results

|  Panel Member | Site 1: % Agreement with Expected Results (Count) |   |   | Site 2: % Agreement with Expected Results (Count) |   |   | Site 3: % Agreement with Expected Results (Count)  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Op 1 | Op 2 | Overall | Op 1 | Op 2 | Overall | Op 1 | Op 2 | Overall  |
|  Influenza A 1x LoD | 100.0% (18/18) | 88.9% (16/18) | 94.4% (34/36) | 100.0% (18/18) | 100.0% (18/18) a | 100.0% (36/36) a | 94.4% (17/18) | 83.3% (15/18) | 88.9% (32/36)  |
|  Influenza A 4x LoD | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36)  |
|  Influenza B 1x LoD | 94.4% (17/18) | 100.0% (18/18) | 97.2% (35/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36) | 94.4% (17/18) | 88.9% (16/18) | 91.7% (33/36)  |
|  Influenza B 4x LoD | 100.0% (18/18) | 94.4% (17/18) | 97.2% (35/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36)  |
|  SARS-CoV-2 1x LoD | 94.4% (17/18) | 100.0% (18/18) | 97.2% (35/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36) | 88.9% (16/18) | 88.9% (16/18) | 88.9% (32/36)  |
|  SARS-CoV-2 4x LoD | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36)  |
|  Negative | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36) | 100.0% (18/18) | 100.0% (18/18) | 100.0% (36/36)  |

a One low positive (1x LoD) influenza A sample was an unexpected positive for SARS-CoV-2.

# 2. Linearity:

Not applicable. This is a qualitative assay.

# 3. Analytical Specificity/Interference:

# a. Analytical Reactivity (Inclusivity):

# SARS-CoV-2 In Silico Analysis:

The inclusivity of the Visby Medical Respiratory Health Test was evaluated using in silico analysis of the forward primers, reverse primers, and probes for the SARS-CoV-2 target systems in relation to sequences available in the GISAID gene database. SARS-CoV-2 in silico analysis was performed using all available SARS-CoV-2 sequences in the GISAID database up to August 2, 2024 (&gt;16.8 million sequences). The SARS-CoV-2 sequences included the following lineages and variants of concern (VOC) or variants of interest (VOI) that may have important epidemiological, immunological, or pathogenic properties from a public health perspective: Delta, Alpha, Omicron BA.1, Omicron BA.2, Omicron BA.4, Omicron BA.5, Omicron BQ.1, Omicron BQ.1.1, Omicron BF.7, Gamma, Epsilon, Iota, Beta, Mu, Zeta, Kappa, Eta, Lamba, and Theta, and others.

K242526 - Page 8 of 20

{8}

All mismatches present in  $&gt;5\%$  of SARS-CoV-2 sequences (either overall or in the last 30, 60, 90, or 120 days prior to analysis) underwent further in silico analysis. None of these mismatches analyzed are predicted to significantly reduce the percent of primer/probe bound at the nominal annealing/detection temperatures of the device; the maximum predicted reduction in binding across all analyzed mismatches is  $&lt; 18\%$  in only the  $N1$  primer/probe set. These minor reductions in primer/probe binding, in combination with the redundancy of the  $N1$  or Orf1ab gene assay targets is expected to have a minimal impact on the ability of the Visby Medical Respiratory Health Test to detect variants for SARS-CoV-2.

# Wet Testing:

An analytical reactivity study was conducted to evaluate the ability of the Visby Medical Respiratory Health Test to detect multiple SARS-CoV-2, influenza A, and influenza B strains that are temporally and geographically diverse. This study evaluated ten H1N1 (pandemic 2009) strains, 13 H3N2 strains, 12 influenza B strains (five each of Victoria and Yamagata lineages and two additional strains), and six SARS-CoV-2 strains as shown in Table 5 below. Inactivated strains were used for SARS-CoV-2 and cultured virus were used for influenza strains, unless noted. Each virus was spiked into negative pooled clinical swab matrix at analyte concentrations of  $3\mathrm{x}$  LoD. This spiked material was then transferred onto a nasopharyngeal (NP) swab and tested on three devices.

Of the 41 viruses tested, 40 were successfully detected in  $3/3$  devices for the expected strain (influenza A, influenza B or SARS-CoV-2) on the Visby Medical Respiratory Health Test at  $\leq 3\mathrm{x}$  LoD. The Omicron (BA.2.3) isolate (USA/MD-HP24556/2022) was detected in  $3/3$  devices at  $6\mathrm{x}$  LoD. The results from this study demonstrate that the Visby Medical Respiratory Health Test is capable of detecting multiple strains of influenza A, influenza B and SARS-CoV2 that are circulating in humans at the concentrations tested.

Table 5. Analytical Reactivity (Inclusivity) Wet-Testing Study Results

|  Virus | Strain | Concentration Tested (copies/swab) | Detection Rate  |
| --- | --- | --- | --- |
|  Influenza A H1N1 (pandemic 2009) | A/Brownsville/39H/2009 | 318 (3x LoD) | 3/3  |
|   |  A/Hong Kong/H090-761-V1(0)/2009 |   | 3/3  |
|   |  A/Netherlands/2629/2009 |   | 3/3  |
|   |  A/Massachusetts/15/2013 |   | 3/3  |
|   |  A/Bangladesh/3002/2015 |   | 3/3  |
|   |  A/Michigan/45/2015 |   | 3/3  |
|   |  A/St. Petersburg/61/2015 |   | 3/3  |
|   |  A/Hawaii/66/2019 |   | 3/3  |
|   |  A/Indiana/02/2020 |   | 3/3  |
|   |  A/Wisconsin/588/2019 |   | 3/3  |
|  Influenza A H3N2 | A/Netherlands/22/2003 | 393 (3x LoD) | 3/3  |
|   |  A/New York/55/2004 |   | 3/3  |
|   |  A/Brisbane/10/2007 |   | 3/3  |
|   |  A/Uruguay/716/2007 |   | 3/3  |

K242526 - Page 9 of 20

{9}

|  Virus | Strain | Concentration Tested (copies/swab) | Detection Rate  |
| --- | --- | --- | --- |
|   | A/Hong Kong/H090-756-V1(0)/2009 |  | 3/3  |
|   |  A/Perth/16/2009 |   | 3/3  |
|   |  A/Victoria/361/2011 |   | 3/3  |
|   |  A/Texas/50/2012 |   | 3/3  |
|   |  A/Switzerland/9715293/2013 |   | 3/3  |
|   |  A/Alaska/232/2015 |   | 3/3  |
|   |  A/California/55/2020 |   | 3/3  |
|   |  A/Hong Kong/2671/2019 |   | 3/3  |
|   |  A/Wisconsin/04/2018 |   | 3/3  |
|  Influenza B Victoria Lineage | B/Malaysia/2506/2004 | 2,184 (3x LoD) | 3/3  |
|   |  B/St. Petersburg/14/2006 |   | 3/3  |
|   |  B/Brisbane/60/2008 |   | 3/3  |
|   |  B/Nevada/03/2011 |   | 3/3  |
|   |  B/New Jersey/1/2012 |   | 3/3  |
|  Influenza B Yamagata Lineage | B/New York/1061/2004 | 2,334 (3x LoD) | 3/3  |
|   |  B/Florida/4/2006 |   | 3/3  |
|   |  B/Texas/06/2011 |   | 3/3  |
|   |  B/Phuket/3073/2013 |   | 3/3  |
|   |  B/Guangdong-Liwan/1133/2014 |   | 3/3  |
|  Influenza B | B/Lee/1940 | 2,334 (3x LoD) | 3/3  |
|   |  B/Maryland/1/1959 |   | 3/3  |
|  SARS-CoV-2 | Alpha (B.1.1.7)
England/204820464/2020 | 100 (1x LoD) | 1/3  |
|   |   |  300 (3x LoD) | 3/3  |
|   |  Beta (B.1.351)
South Africa/ KRISP- K005325/2020 | 100 (1x LoD) | 3/3  |
|   |  Gamma (P.1)
Japan/TY7-503/2021 | 100 (1x LoD) | 2/3  |
|   |   |  300 (3x LoD) | 3/3  |
|   |  Delta (B.1.617.2)
USA/PHC658/2021 | 100 (1x LoD) | 1/3  |
|   |   |  300 (3x LoD) | 3/3  |
|   |  B.1.1.519 (BA.1)
Isolate: USA/MD-HP20874/2021 | 100 (1x LoD) | 2/3  |
|   |   |  300 (3x LoD) | 3/3  |
|   |  Omicron (BA.2.3)
Isolate: USA/MD-HP24556/2022 | 100 (1x LoD) | 1/3  |
|   |   |  300 (3x LoD) | 1/3  |
|   |   |  600 (6x LoD) | 3/3  |

b. **Cross-reactivity:**

**Wet Testing:**

The analytical specificity (cross-reactivity) of the Visby Medical Respiratory Health Test was evaluated by testing a cohort of forty-three non-targeted microorganisms (including

K242526 - Page 10 of 20

{10}

pooled human nasal wash) that may be found in a respiratory tract clinical specimen (Table 6). Panels were composed of up to five different non-target microorganisms spiked into negative pooled clinical swab matrix at  $\geq 1\times 10^{5}$  TCID $_{50}$ /mL (for viruses) and  $\geq 1\times 10^{6}$  CFU/mL (for bacteria and fungi), with the exception of Cytomegalovirus, Measles, Mumps, which were tested at  $\geq 1\times 10^{4}$  TCID $_{50}$ /mL or genome copies/mL as shown in Table 6. Each panel was tested in triplicate.

All sample groups returned 3/3 devices with the expected negative results. Overall, the data show that none of the 43 organisms (including pooled human nasal wash) cause false positive results when tested using the Visby Medical Respiratory Health Test at the concentrations tested.

Table 6. Cross-reactivity Wet-Testing Study Results

|  Organism | Concentration tested | Detection Rate  |
| --- | --- | --- |
|  Human Coronavirus 229E | 1.1 x 105TCID50/mL | 0/3  |
|  Human Coronavirus OC43 | 1.1 x 105TCID50/mL  |   |
|  Human Coronavirus HKU1*† | 1.1 x 105genome copies/mL  |   |
|  Human Coronavirus NL63*† | 1.1 x 105genome copies/mL  |   |
|  SARS-Coronavirus*† | 1.1 x 105genome copies/mL  |   |
|  MERS-Coronavirus*† | 1.1 x 105genome copies/mL | 0/3  |
|  Adenovirus strain 1, C1 Ad 71 | 1.1 x 105TCID50/mL  |   |
|  Adenovirus strain 7 | 1.1 x 105TCID50/mL  |   |
|  Cytomegalovirus | 1.1 x 104TCID50/mL  |   |
|  Epstein Barr virus | 1.1 x 105copies/mL  |   |
|  Enterovirus 68 | 1.1 x 105TCID50/mL | 0/3  |
|  Human metapneumovirus (hMPV) | 1.1 x 105TCID50/mL  |   |
|  Human parainfluenza virus 1 | 1.1 x 105TCID50/mL  |   |
|  Human parainfluenza virus 2 | 1.1 x 105TCID50/mL  |   |
|  Human parainfluenza virus 3 | 1.1 x 105TCID50/mL  |   |
|  Human parainfluenza virus 4b | 1.1 x 105TCID50/mL | 0/3  |
|  Measles | 1.1 x 104TCID50/mL  |   |
|  Mumps* | 1.1 x 104genome copies/mL  |   |
|  Respiratory syncytial virus (Strain B) | 1.1 x 105TCID50/mL  |   |
|  Human rhinovirus 1A (strain 2060) | 1.1 x 105PFU/mL  |   |
|  Bordetella pertussis | 1.1 x 106cfu/mL | 0/3  |
|  Candida albicans | 1.1 x 106cfu/mL  |   |
|  Chlamydia pneumoniae | 1.1 x 106cfu/mL  |   |
|  Corynebacterium xerosis | 1.1 x 106genomic copies/mL  |   |
|  Escherichia coli | 1.1 x 106cfu/mL  |   |

K242526 - Page 11 of 20

{11}

|  Organism | Concentration tested | Detection Rate  |
| --- | --- | --- |
|  Haemophilus influenzae | 1.1 x 10^{6} cfu/mL | 0/3  |
|  Lactobacillus brevis | 1.1 x 10^{6} genomic copies/mL  |   |
|  Legionella pneumophila | 1.1 x 10^{6} genomic copies/mL  |   |
|  Moraxella (Branhamella) catarrhalis | 1.1 x 10^{6} cfu/mL  |   |
|  Mycobacterium tuberculosis | 1.1 x 10^{6} genomic copies/mL  |   |
|  Mycoplasma pneumoniae | 1.1 x 10^{6} cfu/mL | 0/3  |
|  Neisseria meningitidis serogroup a | 1.1 x 10^{6} genomic copies/mL  |   |
|  Neisseria mucosa | 1.1 x 10^{6} genomic copies/mL  |   |
|  Pseudomonas aeruginosa | 1.1 x 10^{6} cfu/mL  |   |
|  Staphylococcus aureus | 1.1 x 10^{6} cfu/mL | 0/3  |
|  Staphylococcus epidermidis | 1.1 x 10^{6} cfu/mL  |   |
|  Streptococcus pneumoniae | 1.1 x 10^{6} genomic copies/mL  |   |
|  Streptococcus pyogenes | 1.1 x 10^{6} cfu/mL  |   |
|  Streptococcus salivarius | 1.1 x 10^{6} genomic copies/mL  |   |
|  Pneumocystis jirovecii also called Pneumocystis carinii Delanoe and Delanoe | 1.1 x 10^{6} nuclei/mL | 0/3  |
|  Bordetella parapertussis | 1.1 x 10^{6} cfu/mL  |   |
|  Mycoplasma genitalium | 1.1 x 10^{6} genomic copies/mL  |   |
|  Pooled human nasal wash | 100% (no dilution) | 0/3  |

* Purified RNA was tested for these organisms.
† Synthetic RNA was tested for these organisms.

c. Microbial Interference:

A microbial interference study was conducted to assess potential inhibitory effects of 43 non-target microorganisms (including pooled human nasal wash) that may be found in a human respiratory specimen (Table 6) on performance of the Visby Medical Respiratory Health Test. Panels were composed of up to five different non-target microorganisms spiked into pooled clinical swab matrix that contained low positive (i.e., 3x LoD) concentrations of each target organism SARS-CoV-2 (USA-WA1/2020, inactivated virus), influenza A (2009 H1N1, Brisbane/02/18, live virus), and influenza B (Victoria, Washington/02/19, live virus). Non-target organisms were spiked at $\geq 1 \times 10^{5} \mathrm{TCID}_{50} / \mathrm{mL}$ (for viruses) and $\geq 1 \times 10^{6} \mathrm{CFU} / \mathrm{mL}$ (for bacteria and fungi), with the exception of Cytomegalovirus, Measles, Mumps, which were tested at $\geq 1 \times 10^{4} \mathrm{TCID}_{50} / \mathrm{mL}$ or genome copies/mL as shown in Table 6. Each panel was tested in triplicate.

All sample groups returned 3/3 devices with the expected negative results. Overall, the data shows that none of the 43 organisms (including pooled human nasal wash) cause false negative results when tested using the Visby Medical Respiratory Health Test at the concentrations tested.

K242526 - Page 12 of 20

{12}

# d. Competitive Interference:

The potential for high concentrations (i.e.,  $&gt;1 \times 10^{5}$  copies/swab) of one target organism to interfere with detection of low concentrations (i.e.,  $3 \times \mathrm{LoD}$ ) of another target organism (i.e., co-infections with on-target analytes) was evaluated for the Visby Medical Respiratory Health Test.

Influenza A (2009 H1N1, Brisbane/02/18, live virus), influenza B (Victoria, Washington/02/19, live virus), and SARS-CoV-2 (USA-WA1/2020, inactivated virus) viruses were spiked into negative pooled clinical swab matrix at varying concentrations and then tested in triplicate. Low concentrations were prepared at 3x LoD for the respective viruses, and high concentrations were prepared as described in Table 7. No competitive interference was observed for any of the three target viruses at the concentrations tested.

Table 7. Competitive Interference Results Summary

|  Viral Targets in Sample |   |   | Detection Rate  |   |   |
| --- | --- | --- | --- | --- | --- |
|  Influenza A copies/swab) | Influenza B (copies/swab) | SARS-CoV-2 (copies/swab) | Influenza A | Influenza B | SARS-CoV-2  |
|  1.06 x 105 | 2184 | N/A | 3/3 | 3/3 | 0/3  |
|  1.06 x 105 | N/A | 300 | 3/3 | 0/3 | 3/3  |
|  318 | 7.28 x 105 | N/A | 3/3 | 3/3 | 0/3  |
|  N/A | 7.28 x 105 | 300 | 0/3 | 3/3 | 3/3  |
|  318 | N/A | 1.0 x 105 | 3/3 | 0/3 | 3/3  |
|  N/A | 2184 | 1.0 x 105 | 0/3 | 3/3 | 3/3  |

# e. Interfering Substances:

The performance of the Visby Medical Respiratory Health Test was evaluated in the presence of medically and/or physiologically relevant concentrations of potentially interfering substances that may be present in a respiratory specimen. For this study, negative swab clinical sample matrix pools were individually spiked with the potential interfering substances at the concentrations denoted in Table 8. To determine if these potential interfering substances impact test performance in the presence and absence of target analytes, both target-spiked and target-non-spiked samples were generated, respectively. Target-spiked samples were created by triple spiking influenza A (2009 H1N1, Brisbane/02/18, live virus), influenza B (Victoria, Washington/02/19, live virus), and SARS-CoV-2 (USA-WA1/2020, inactivated virus) at  $3\mathrm{x}$  LoD into negative pooled swab clinical sample matrix, containing a single potential interfering substance. Target-non-spiked samples consisted of negative pooled nasal swab clinical sample matrix and a single potential interfering substance. For each potentially interfering substance, three replicates were tested for the target-spiked and target-non-spiked samples.

All target-non-spiked samples were  $100\%$  negative for influenza A, influenza B, and SARS-CoV-2. All target-spiked samples were  $100\%$  positive for influenza A, influenza B, and SARS-CoV-2, except for cotton and rayon swabs. For these swab materials, two of the three target-spiked samples were positive for all targets, while one of three target-spiked samples was negative for SARS-CoV-2. Based on the results for cotton and rayon swabs, these swab

K242526 - Page 13 of 20

{13}

materials are not acceptable swab types for use with the Visby Medical Respiratory Health Test. None of the other evaluated substances interfered with performance of the device at the concentrations tested.

Table 8. Substances and Swab Materials Evaluated for Interference

|  Potential Interferant | Active Ingredient | Test Concentration | Detection Rate  |   |
| --- | --- | --- | --- | --- |
|   |   |   |  Negative Samples | 3x LoD Samples  |
|  Afrin (Nasal Spray) | Oxymetazoline | 25% (v/v) | 0/3 | 3/3  |
|  Saline (Nasal Spray) | N/A | 25% (v/v) | 0/3 | 3/3  |
|  NeoSynephrine Cold & Sinus Extra Strength Spray (Nasal Spray) | Phenylephrine | 25% (v/v) | 0/3 | 3/3  |
|  Flonase (Nasal Corticosteroid) | Fluticasone | 25% (v/v) | 0/3 | 3/3  |
|  Nasacort (Nasal Corticosteroid) | Triamcinolone | 25% (v/v) | 0/3 | 3/3  |
|  Blood, Human | N/A | 5% (v/v) | 0/3 | 3/3  |
|  Bovine Submaxillary Gland, Type I-S | Purified Mucin Protein | 1% (w/v) | 0/3 | 3/3  |
|  Hand Lotion | N/A | 5% (w/v) | 0/3 | 3/3  |
|  Hand Sanitizer | N/A | 5% (w/v) | 0/3 | 3/3  |
|  Method All Purpose Cleaner | N/A | 5% (v/v) | 0/3 | 3/3  |
|  Seventh Generation Disinfectant | N/A | 5% (v/v) | 0/3 | 3/3  |
|  Softsoap Moisturizing Hand Soap | N/A | 5% (w/v) | 0/3 | 3/3  |
|  Antibiotic, Nasal Ointment | Mupirocin | 12 mg/mL | 0/3 | 3/3  |
|  Antibiotic, Systemic | Tobramycin | 2.43 mg/mL | 0/3 | 3/3  |
|  Antiviral | Zanamivir | 5 mg/mL | 0/3 | 3/3  |
|  Biotin | N/A | 3.5 ug/mL | 0/3 | 3/3  |
|  Zicam Cold Remedy No Drip Nasal Spray (Homeopathic Allergy Relief) | Luffa Opperculata, Galphimia Glauca, Sabadilla | 25% (v/v) | 0/3 | 3/3  |
|  Swab Material Type | Active Ingredient | Test Concentration | Negative Samples | 3x LoD Samples  |
|  Cotton Swab | N/A | N/A | 0/3 | 2/3a  |
|  Flocked Polyester Swab | N/A | N/A | 0/3 | 3/3  |
|  Foam Swab | N/A | N/A | 0/3 | 3/3  |
|  Rayon Swab | N/A | N/A | 0/3 | 2/3a  |

a One device returned a positive result for influenza A and influenza B, but negative for SARS-CoV-2.

# 4. Assay Reportable Range:

Not applicable.

This is a qualitative colorimetric test with binary output (positive/negative). The intensity of the color is not proportional to the concentration of the target organisms.

# 5. Specimen Stability:

Data provided support the following stability claims for specimens eluted in Visby Respiratory Health Buffer:

K242526 - Page 14 of 20

{14}

- Room temperature (15 – 30°C) storage for up to 120 minutes (2 hours)
- Refrigerated (2 – 8°C) storage for up to 48 hours (2 days)

## 6. Detection Limit:

### a. Limit of Detection (Individual Analytes):

The purpose of this study was to determine the limit of detection (LoD) of the Visby Medical Respiratory Health Test for the detection of influenza A, influenza B, and SARS-CoV-2 as single analytes. The study assessed two live strains of influenza A (influenza A 2009 H1N1, Brisbane/02/18 and influenza A H3N2, Kansas/14/2017) and influenza B (influenza B Victoria, Washington/02/19 and influenza B Yamagata, Oklahoma/10/2018), and one inactivated strain of SARS-CoV-2 (USA-WA1/2020). Each organism was individually spiked into pooled negative clinical swab matrix and applied at the appropriate final concentration to nasopharyngeal swabs for testing per the instructions for use for the Visby Medical Respiratory Health Test. Each virus was tested in a range finding study that included five concentrations of a three-fold dilution series, where each concentration was tested with five replicates. The lowest concentration with 100% detection was estimated to be the LoD. The estimated LoD for each strain was verified by testing 20 replicates at the estimated LoD concentration and demonstrating that at least 19/20 replicates were positive. If the criteria of at least 19 out of 20 positive replicates were not met, testing was repeated with a higher concentration until at least 19/20 replicates gave a positive result.

The LoD of the Visby Medical Respiratory Health Test for influenza A, influenza B, and SARS-CoV-2 are summarized in Table 9.

|  Virus | Analytical Limit of Detection (LoD)  |   |
| --- | --- | --- |
|   |  Copies/swab | TCID_{50}/swab or FFU/swab  |
|  Influenza A 2009 H1N1, Brisbane/02/18 | 106 | 4.89 TCID_{50}/swab  |
|  Influenza A H3N2, Kansas/14/2017 | 125 | 2.01 FFU/swab  |
|  Influenza B Victoria, Washington/02/19 | 728 | 9.20 TCID_{50}/swab  |
|  Influenza B Yamagata, Oklahoma/10/2018 | 778 | 88.37 TCID_{50}/swab  |
|  SARS-CoV-2 (USA-WA1/2020) | 100 | N/A  |

FFU: Fluorescent Focus Units. Titer by Fluorescent Focus Assay in MDCK-SIAT1 Cells.

## 7. Assay Cut-Off:

Not applicable. This is a visually interpreted test where any shade of purple indicates a positive reaction.

## 8. Carry-Over:

Not applicable. This is a single-use test and is not subject to carryover from previously tested specimens.

## B Comparison Studies:

K242526 - Page 15 of 20

{15}

1. Method Comparison with Predicate Device:

See Clinical Studies below.

2. Matrix Comparison:

An LoD equivalency study was conducted to compare side-by-side testing of the same viral stocks prepared in nasal and nasopharyngeal swab matrices and eluted in Visby Respiratory Buffer. Testing was performed with a minimum of three dilutions and until the concentration was found for each virus in each matrix with at least 19/20 positive results where the dilution below had &lt; 19/20 positive results. The LoD was considered equivalent between the nasal and nasopharyngeal swab matrices if the lowest concentration of virus with at least 19/20 positives is within one three-fold dilution for each other.

Influenza A, influenza B, and inactivated SARS-CoV-2 virus were spiked into each matrix at the specified concentration for each target to prepare a triple positive sample. Three triple positive samples were made in each matrix, and they were tested with 20 devices.

The results of the testing are summarized in Table 10 below. The bolded results indicate the lowest concentrations for each virus with at least 19/20 positive results. These results support that the LoD for detection of influenza A, influenza B, and SARS-CoV-2 are equivalent for nasal and nasopharyngeal swab matrices.

Table 10. LoD Equivalency Results for Nasal and Nasopharyngeal Swab Matrices

|  Virus | Concentration (copies/swab, xLoD) | Detection Rate  |   |
| --- | --- | --- | --- |
|   |   |  Nasal Swab Matrix | Nasopharyngeal Swab Matrix  |
|  Influenza A 2009 | 318 (3x) | 20/20 | 20/20  |
|  H1N1, | 106 (1x) | 18/20 | 19/20  |
|  Brisbane/02/18 | 35 (1/3x) | 13/20 | 16/20  |
|  Influenza B | 728 (1x) | 20/20 | 20/20  |
|  Victoria, | 243 (1/3x) | 14/20 | 20/20  |
|  Washington/02/19 | 81 (1/9x) | 11/20 | 15/20  |
|  SARS-CoV-2 (USA | 333 (3.3x) | 20/20 | 20/20  |
|  WA1/2020) | 33 (1/3x) | 20/20 | 19/20  |
|   | 11 (1/9x) | 15/20 | 10/20  |

C Clinical Studies:

The clinical performance of the Visby Medical Respiratory Health Test was evaluated in a multicenter prospective study conducted at five geographically distinct clinical sites representative of CLIA-waived testing facilities. The sites were distributed across the United States and were described as urgent care and family care clinics. A total of 18 untrained operators, representative of CLIA waived users, participated in the study. Reference testing took place at a single laboratory.

The study prospectively enrolled subjects with signs and symptoms of respiratory tract infection. Subjects were enrolled from May 2022 through March 2023 and from November 2023 through

K242526 - Page 16 of 20

{16}

February 2024. The average age among study participants was 34 years, with a range between one month to 89 years of age.

One NP swab specimens or dual nostril AN swab specimen was collected from each eligible subject. The AN specimens could be self-collected under health care provider (HCP) observation (for subjects  $\geq 14$  years of age), or health care provider (HCP) collected, and all NP swabs were HCP collected. The swab specimen was placed directly in the Visby Respiratory Health Buffer and tested on-site by study operators using the Visby Medical Respiratory Health Test. The participating operators conducted the test by following the instructions in the Quick Reference Guide (QRG). The study operators were representative of operators conducting testing at a CLIA Waived setting and did not receive any training on the use of the Visby Medical Respiratory Health Test.

The specimen was also tested at the reference laboratory for comparator testing using an FDA-cleared nucleic acid amplification test (NAAT) for influenza A and influenza B targets and a highly sensitive FDA-authorized laboratory-based RT-PCR EUA assay for the SARS-CoV-2 target.

A total of 1,592 subjects were initially enrolled. Study samples were excluded from the data analysis due to lack of a valid comparator result  $(n = 53)$ , lack of a valid Visby Medical Respiratory Health Test result  $(n = 21)$ , the subject was asymptomatic  $(n = 10)$ , or for protocol deviations (i.e., failure to complete a required procedure  $(n = 4)$  or incorrectly filled out informed consent form  $(n = 2)$ ). Additionally, one subject withdrew from the study. This left 1,501 subjects for performance analysis.

The performance estimates for the Visby Medical Respiratory Health Test for the detection of SARS-CoV-2, influenza A, and influenza B were calculated as positive percent agreement (PPA) and negative percent agreement (NPA) with the comparator result, which are shown in Tables 11-13 stratified by specimen type. These results are stratified by site in Tables 14-16.

Table 11. SARS-CoV-2 Clinical Performance for the Visby Test vs. Comparator, by Specimen Type

|  Specimen Type | N | TP | FP | TN | FN | PPA (95% CI) | NPA (95% CI)  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  NPS | 829 | 130 | 7a | 687 | 5c | 96.3% (91.6-98.4%) | 99.0% (97.9-99.5%)  |
|  ANS | 672 | 116 | 5b | 549 | 2d | 98.3% (94.0-99.5%) | 99.1% (97.9-99.6%)  |
|  Total | 1501 | 246 | 12 | 1236 | 7 | 97.2% (94.4-98.7%) | 99.0% (98.3-99.5%)  |

PPA=Positive Percent Agreement with comparator; NPA=Negative Percent Agreement with comparator; TP=true positive; FP=false positive; TN=true negative; FN=false negative
${}^{a}5$  of 7 false positive NPS specimens tested positive and 2 tested negative with an alternate EUA molecular assay.
${}^{\mathrm{b}}{4}^{ \circ  }5$  false positive AN specimens tested positive and 1 tested negative with an alternate EUA molecular assay.
${}^{\mathrm{c}}{3}^{ \circ  }5$  false negative NPS specimens tested negative and 2 tested positive with an alternate EUA molecular assay.
${}^{\mathrm{d}}{10}^{ \circ  }{2}^{ \circ  }$  false negative AN specimens tested negative and 1 tested positive with an alternate EUA molecular assay.

K242526 - Page 17 of 20

{17}

Table 12. Influenza A Clinical Performance for the Visby Test vs. Comparator, by Specimen Type

|  Specimen Type | N | TP | FP | TN | FN | PPA (95% CI) | NPA (95% CI)  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  NPS | 829 | 33 | 4^{a} | 791 | 1^{c} | 97.1% (85.1-99.5%) | 99.5% (98.7-99.8%)  |
|  ANS | 672 | 61 | 5^{b} | 604 | 2^{d} | 96.8% (89.1-99.1%) | 99.2% (98.1-99.7%)  |
|  Total | 1501 | 94 | 9 | 1395 | 3 | 96.9% (91.3-98.9%) | 99.4% (98.8-99.7%)  |

PPA=Positive Percent Agreement with comparator; NPA=Negative Percent Agreement with comparator; TP=true positive; FP=false positive; TN=true negative; FN=false negative
a 3 of 4 false positive NPS specimens tested positive and 1 tested negative with an alternate EUA molecular assay.
b 5 of 5 false positive AN specimens tested positive with an alternate EUA molecular assay.
c 1 false negative NPS specimen tested positive with an alternate EUA molecular assay.
d 2 false negative AN specimens tested positive with an alternate EUA molecular assay.

Table 13. Influenza B Clinical Performance for the Visby Test vs. Comparator, by Specimen Type

|  Specimen Type | N | TP | FP | TN | FN | PPA (95% CI) | NPA (95% CI)  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  NPS | 829 | 15 | 2^{a} | 812 | 0 | 100% (79.6-100%) | 99.8% (99.1-99.9%)  |
|  ANS | 672 | 15 | 1^{b} | 656 | 0 | 100% (79.6-100%) | 99.9% (99.1-100%)  |
|  Total | 1501 | 30 | 3 | 1468 | 0 | 100% (88.7-100%) | 99.8% (99.4-99.9%)  |

PPA=Positive Percent Agreement with comparator; NPA=Negative Percent Agreement with comparator; TP=true positive; FP=false positive; TN=true negative; FN=false negative
a 2 of 2 false positive NPS specimens tested positive with an alternate EUA molecular assay.
b 1 false positive AN specimen tested positive with an alternate EUA molecular assay.

Table 14. Influenza A Clinical Performance for the Visby Test vs. Comparator, by Site

|  Site | N | TP | FP | TN | FN | PPA (95% CI) | NPA (95% CI)  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Total | 1501 | 94 | 9 | 1395 | 3 | 96.9% (91.3%-98.9%) | 99.4% (98.8%-99.7%)  |
|  Site 1 | 791 | 55 | 2 | 732 | 2 | 96.5% (88.1%-99.0%) | 99.7% (99.0%-99.9%)  |
|  Site 2 | 82 | 2 | 1 | 78 | 1 | 66.7% (20.8%-93.9%) | 98.7% (93.2%-99.8%)  |
|  Site 3 | 246 | 20 | 5 | 221 | 0 | 100% (83.9%-100%) | 97.8% (94.9%-99.1%)  |
|  Site 4 | 355 | 14 | 1 | 340 | 0 | 100% (78.5%-100%) | 99.7% (98.4%-100%)  |
|  Site 5 | 27 | 3 | 0 | 24 | 0 | 100% (43.9%-100%) | 100% (86.2%-100%)  |

K242526 - Page 18 of 20

{18}

Table 15. Influenza B Clinical Performance for the Visby Test vs. Comparator, by Site

|  Site | N | TP | FP | TN | FN | PPA (95% CI) | NPA (95% CI)  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Total | 1501 | 30 | 3 | 1468 | 0 | 100% (88.7%-100%) | 99.8% (99.4%-99.9%)  |
|  Site 1 | 791 | 14 | 0 | 777 | 0 | 100% (78.5%-100%) | 100% (99.5%-100%)  |
|  Site 2 | 82 | 3 | 0 | 79 | 0 | 100% (43.9%-100%) | 100% (95.4%-100%)  |
|  Site 3 | 246 | 1 | 0 | 245 | 0 | 100% (20.7%-100%) | 100% (98.5%-100%)  |
|  Site 4 | 355 | 11 | 2 | 342 | 0 | 100% (74.1%-100%) | 99.4% (97.9%-99.8%)  |
|  Site 5 | 27 | 1 | 1 | 25 | 0 | 100% (20.7%-100%) | 96.2% (81.1%-99.3%)  |

Table 16. SARS-CoV-2 Clinical Performance for the Visby Test vs. Comparator, by Site

|  Site | N | TP | FP | TN | FN | PPA (95% CI) | NPA (95% CI)  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Total | 1501 | 246 | 12 | 1236 | 7 | 97.2% (94.4%-98.7%) | 99.0% (98.3%-99.5%)  |
|  Site 1 | 791 | 158 | 9 | 621 | 3 | 98.1% (94.7%-99.4%) | 98.6% (97.3%-99.3%)  |
|  Site 2 | 82 | 6 | 1 | 74 | 1 | 85.7% (48.7%-97.4%) | 98.7% (92.8%-99.8%)  |
|  Site 3 | 246 | 17 | 0 | 229 | 0 | 100% (81.6%-100%) | 100.00% (98.4%-100%)  |
|  Site 4 | 355 | 60 | 2 | 290 | 3 | 95.2% (86.9%-98.4%) | 99.3% (97.5%-99.8%)  |
|  Site 5 | 27 | 5 | 0 | 22 | 0 | 100% (56.6%-100%) | 100% (85.1%-100%)  |

Of the 1,575 tests performed on the Visby Medical Respiratory Health Test, 79 (5.0%, 95% CI = 4.0-6.2%) had an initial invalid test result of which 21 (1.3%, 95% CI = 0.8-2.0%) were invalid when retested. Invalid results were more frequent for ANS (6.7% initial with 95% CI = 5.0-8.7%, 2.2% upon repeat with 95% CI = 1.3-3.5%) than for NPS (3.6% initial with 95% CI = 2.5-5.1%, 0.6% upon retest with 95% CI = 0.3-1.4%).

D Clinical Cut-Off:

Not applicable.

E Expected Values/Reference Range:

Table 17. Positivity Rate of the Visby Medical Respiratory Health Test by Age for Detection of influenza A, influenza B and SARS-CoV-2 in Respiratory Swab Specimens (NP and AN) During the Prospective Clinical Study (Enrollment from May 2022-March 2023)

|  Age | N (%) | % Positive (# Positive / # Tested)  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Influenza A | Influenza B | SARS-COV-2  |
|  0-5 | 72 | 0.0% | 0.0% | 8.3%  |

K242526 - Page 19 of 20

{19}

|  Age | N (%) | % Positive (# Positive / # Tested)  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Influenza A | Influenza B | SARS-COV-2  |
|   | (6.5%) | (0/72) | (0/72) | (6/72)  |
|  6-21 | 247 | 8.9% | 0.0% | 8.1%  |
|   |  (22.2%) | (22/247) | (0/247) | (20/247)  |
|  22-59 | 649 | 4.8% | 0.002% | 20.5%  |
|   |  (58.2%) | (31/649) | (1/649) | (133/649)  |
|  ≥ 60 | 147 | 0.01% | 0.0% | 25.9%  |
|   |  (13.2%) | (1/147) | (0/147) | (38/147)  |
|  Total | 1115 | 4.8% | 0.00% | 17.7%  |
|   |   |  (54/1115) | (1/1115) | (197/1115)  |

Table 18. Positivity Rate of the Visby Medical Respiratory Health Test by Age for Detection of influenza A, influenza B and SARS-CoV-2 in Respiratory Swab Specimens (NP and AN) During the Prospective Clinical Study (Enrollment from November 2023-February 2024)

|  Age | N (%) | % Positive (# Positive / # Tested)  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Influenza A | Influenza B | SARS-COV-2  |
|  0-5 | 13 | 15.4% | 7.7% | 0.0%  |
|   |  (3.4%) | (2/13) | (1/13) | (0/13)  |
|  6-21 | 78 | 19.2% | 12.8% | 9.0%  |
|   |  (20.1%) | (15/78) | (10/78) | (7/78)  |
|  22-59 | 234 | 11.5% | 9.0% | 16.2%  |
|   |  (60.6%) | (27/234) | (21/234) | (38/234)  |
|  ≥ 60 | 61 | 8.2% | 0.0% | 26.2%  |
|   |  (15.8%) | (5/61) | (0/61) | (16/61)  |
|  Total | 386 | 12.7% | 8.3% | 15.8%  |
|   |   |  (49/386) | (32/386) | (61/386)  |

F Other Supportive Instrument Performance Characteristics Data:

Not applicable.

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K242526 - Page 20 of 20

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/QOF/K242526](https://fda.innolitics.com/submissions/MI/subpart-d%E2%80%94serological-reagents/QOF/K242526)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
